Abstract
Pseudo-haptic feedback is a haptic illusion based on a mismatch of haptic and visual perception. It is well known from applications in virtual environments. In this work, we discuss the usabiliy of the principle of pseudo-haptic feedback for teleoperation. Using pseudo-haptic feedback can ease the design of haptic medical tele-operation systems.
Thereby a user’s grasping force at an isometric user interface is used to control the closing angle of an end effector of a surgical robot. To provide a realistic haptic feedback, the coupling characteristic of grasping force and end effector closing angle is changed depending on acting end effector interaction forces.
With an experiment, we show the usability of pseudo-haptic feedback for discriminating compliances, comparable to the mechanical characteristic of muscles relaxed and contracted. The provided results base upon the data of 10 subjects, and 300 trails.
1 Motivation
Haptic feedback can improve the performance of telerobotic surgery in terms of decreasing the time of a surgery and the trauma [1, 2]. Also, providing haptic feedback for teleoperation allows a surgeon to use his haptic sensation, instead of using vision only, to estimate the interaction forces of an end effector in situ. This is supposed to reduce cognitive load of the surgeon.
State of the art teleoperation systems including haptic feedback are commonly based on an impedance-admittance system architecture [3]. This implies a masterslave structure. A user controls the slave robot by interacting with the master unit. The user sets a position or velocity at the master. The slave follows the user’s input. The force between the end effector and tissue, resulting from the performed manipulation in situ, is measured. The measured forces are used to reflect haptic feedback to a user at the master unit.
In this case, it is required that the dexterity of the master unit fits the dexterity of the slave robot. At least one sensor at the slave and one actuator at the master unit is required to provide haptic feedback for each degree of freedom of the teleoperation system. The integration of force sensors to the end effector and implementing up to 7 actuators to the master input device, that are needed to provide full dexterity, enhances the complicity of a system. Also, the requirements concerning the dynamic behavior of the actuators rise with rising dexterity and workspace of the telerobotic system.
With our work, we introduce the principle of pseudo-haptic feedback as an option of reflecting haptic feedback of grasping forces of an end effector without the necessity of a moving or active part in the master unit.
2 Application
Pseudo haptic feedback is well known from applications in virtual environments [4–6]. Different kinds of interaction and haptic behavior of virtual object can be simulated and fed back to the user by for example changing the resulting moving characteristics of a mouse curser dependent on special tasks or interaction. The haptic sensation of pseudo haptics bases on a haptic illusion resulting from the mismatch of haptic- and visual perception [4–6].
For the pseudo-haptic teleoperation, we assume a stiff interface as input device (Figure 1). The input device is sensitive to the grasping force of the user (Fg). The closing angle of the end effector (φ) is controlled by the user’s grasping force. The acting interaction force at the end effector (Fe) is measured.
The relation between the grasping force and the end effector is designed as a linear spring for Fe = 0 to present the characteristics of an surgical forceps (Figure 2). Based on interaction forces acting at the end effector, the coupling characteristic of grasping force and end effector closing angle is altered. For example the interaction with a linear compliance (Fe = 1/n · φ) results in a drop of the slope of the characteristic curve. Dependent on mechanical properties, the interaction with any kind of tissue leads to unique characteristic curves. The change of the coupling characteristic, depending on the case of interaction, causes a realistic haptic feedback.

Parts of a teleoperation system concerning the grasping degree of freedom. Principle of an isometric user interface in the upper part of the figure and the principle of the end effector in the lower part of the figure.

Characteristic curve of the relation of the grasping force and the end effector closing angle. Basis couple characteristic for idle case (Fe = 0 N) (solid), Interaction with a linear compliance (Fe = 1/n · φ) and interaction with any environment (Fe(φ, environment)) (dashed).
The haptic feedback perceived by the user results from the perceived difference of grasping forces at the user interface needed to reach the same end effector closing angle. The intensity and the rate of realism of the perceived sensation can be altered by using different coupling characteristics [8].
3 Experiments
A compliance identification experiment is performed to verify the principle of pseudo-haptic teleoperation. An adjustable compliance is used as specimen to simulate different properties of tissue (Figure 3, b, c). For the experiment the specimen’s compliance is set to n1 = 8 mm/N, n2 = 4 mm/N and n3 = 2 mm/N. The values are chosen with respect to the properties of muscles, relaxed (n ≈ 8 mm/N) and contracted (n ≈ 2mm/N) [7].
A load cell is used as isometric user interface (Figure 3, a). The end effector displacement is controlled by the user’s force excitation at the load cell. The end effector bases on a linear moving axis with a force sensor at its tip. The signals of the force sensor at the end effector is used to change the couple characteristic between the user’s grasping force and the end effector displacement.

Experimental setup. Isometric user-interface based on a load cell (a). End effector in interaction with an adjustable compliance, simulating the stiffness of relaxed and contracted muscles (b, c).
While performing the experiment, the subjects were asked to identify the three compliances (n1 . . . n3) only by analyzing their excited grasping forces to the load cell and the visual information of the resulting end effector movement at the specimen.
The experiment was performed with ten subjects, all of them were unprejudiced to what pseudo-haptic feedback is. Every subject was asked to perform 30 trails. The sequence of presented compliances was chosen randomly. The results of the experiment are presented in a confusion matrix (Table 1). The confusion matrix describes the number of perceived values in respect to the presented stimuli [9, 10].
Results of the experiment presented as confusion matrix.
| Teleoperation | ||||
|---|---|---|---|---|
| Presented | ||||
| n1 | n2 | n3 | ||
| Perceived | n1 | 78 | 14 | 0 |
| n2 | 22 | 75 | 18 | |
| n3 | 0 | 14 | 79 | |
The experiment shows that 77.3 % of all presented stimuli were perceived correctly. 47 % of the wrong identified stimuli were perceived as weaker and 53 % as harder than the stimuli presented. All subjects were able to differ between n1 and n3 without any mistake.
In addition to the compliance identification experiment with the teleoperation system, all subjects were asked to repeat the experiment by using their finger directly at the compliance, to derive a reference value. Therefore, the subjects touched and displaced the compliance (Fig. 3 b, c) by their finger directly, instead of the end effector. With direct interaction 99.3 % of all presented stimuli were perceived correctly [11]. The result of the identification experiment without any technical devices in the perceptual channel, can be seen as gold standard for the maximum reachable performance for the identification of compliances with the presented setup.
4 Discussion and further work
The experiment shows the usability of pseudo haptic feedback for the discrimination of compliances. The compliances, chosen as stimuli in the experiment, derived from muscles compliances, allow the conclusion that the discrimination of different tissue is possible, too. While for first experiments the chosen setup with defined and time constant values is best to reach meaningful results, future experiments under more realistic conditions with tissue need to be performed.
Compliance values of muscles were used for first experiments. In future studies weaker material like liver or spleen with compliances up to 15 mm/N should be taken into account to derive information about the usability of pseudo-haptic feedback for very compliant tissue [12]. To give a surgeon the possibility to differ between characteristic material or tissue, the coupling characteristic of the grasping force at the user interface and the resulting end effector closing angle can be adapted.
Since the visual feedback is a very important factor of providing pseudo-haptic feedback, future studies need to take the intra operative sight in situ into account. In contrast to the presented setup of this work, during real surgery the end effector might be covered for some time. Also different kinds of viewing angles and a scaling of displacements need to be considered.
Funding source: Deutsche Forschungsgemeinschaft
Award Identifier / Grant number: WE 2308/13-1
Funding statement: The work on this project is funded by the Deutsche Forschungsgemeinschaft (DFG) under grant WE 2308/13-1
Funding
The work on this project is funded by the Deutsche Forschungsgemeinschaft (DFG) under grant WE 2308/13-1
Author's Statement
Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.
References
[1] King, C. H.; Culjat, M. O.; Franco M. L.; Lewis, C. E.; Dutson, E. P.; Grundfest, W. S. & Bisley, J. W.: Tactile feedback induces reduced grasping force in robot-assisted surgery, Haptics, IEEE Transactions on, vol. 2, pp. 103-110, 2009.10.1109/TOH.2009.4Search in Google Scholar
[2] Okamura, A. M.: Methods for haptic feedback in teleoperated robotassisted surgery, Industrial Robot: An International Journal, vol. 31, pp. 499-508, 2004.10.1108/01439910410566362Search in Google Scholar
[3] Matich, S.; Neupert, C.; Kirschniak, A.; Werthschützky, R.; Schlaak, H. F. & Pott, P. P.: Teleoperation system with haptic feedback for single insicion surgery - concept ans system design, In: proceedings of cars, 2013.Search in Google Scholar
[4] Lecuyer, A.: Simulating haptic feedback using vision: A survey of research and applications of pseudo-haptic feedback, Presence: Teleoperators and Virtual Environments, MIT Press, 2009, 18, 39-53.10.1162/pres.18.1.39Search in Google Scholar
[5] Lecuyer, A.; Coquillart, S.; Kheddar, A.; Richard, P. & Coiffet, P.: Pseudo-haptic feedback: can isometric input devices simulate force feedback? Virtual Reality, 2000. Proceedings. IEEE, 2000, 83-90.10.1109/VR.2000.840369Search in Google Scholar
[6] Lecuyer, A.; Cuquillart, S. & Coiffet, P.: Simulating haptic information with haptic illusions in virtual environments, DTIC Document, 2001.Search in Google Scholar
[7] Zheng, Y.-P.; Mak, A. F. & Lue, B.: Objective assessment of limb tissue elasticity: development of a manual indentation procedure, Journal of rehabilitation research and development, Rehabilitation Research and Development Service, US Department of Veterans Affairs, 1999.Search in Google Scholar
[8] Neupert, C.; Hatzfeld, C.; Matich, S.: Teleoperationssystem mit intrinsischem haptischen Feedback durch dynamische Kennlinienanpassung für Greifkraft und Endeffektorkoordinaten, patent pending (2015011915374800DE), 2015.Search in Google Scholar
[9] Jones, L. A. & Tan, H. Z.: Application of psychophysical techniques to haptic research, Haptics, IEEE Transactions on, 2013, 6, 268-284.10.1109/TOH.2012.74Search in Google Scholar
[10] Hatzfeld, C. & Kern, T. A.: Engineering Haptic Devices, A Beginners Guide, Second Edition Springer, 2014.10.1007/978-1-4471-6518-7Search in Google Scholar
[11] Neupert, C.; Matich, S.; Hatzfeld, C.; Kupnik, M. & Werthschützky, R.: Investigation of the Usability of Pseudo-Haptic Feedback in Teleoperation, IEEE Worldhaptics Conference 2015, Evanston, USA, 2015.10.1109/WHC.2015.7177759Search in Google Scholar
[12] Carter, F. J.; Frank, T. G.; Davis, P. J.; McLean, D. & Cuschieri, A.: Measurements and modelling of the compliance of human and porcine organs. Medical Image Analysis 5.4, 2001, 231-236.10.1016/S1361-8415(01)00048-2Search in Google Scholar
© 2015 by Walter de Gruyter GmbH, Berlin/Boston
This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients