Home Medicine Wireless medical sensors – context, robustness and safety
Article Open Access

Wireless medical sensors – context, robustness and safety

  • Christian Bollmeyer EMAIL logo , Mathias Pelka , Hartmut Gehring and Horst Hellbrück
Published/Copyright: September 12, 2015

Abstract

Wireless medical sensors are an emerging technology. Wireless sensors form networks and are placed in an unknown environment. For indoor scenarios context detection of medical sensors, e.g. removal of sensors from a specific room, is important. Current algorithms for context detection of wireless sensors are based on RF signals, but RF signal propagation and room location show only a weak correlation. Recent approaches with RSSI-measurements are based on prior fingerprinting and therefore costly. In our approach, we equip wireless sensor nodes with a barometric sensor to measure pressure disturbances that occur, when doors of rooms are opened or closed. By signal processing of these disturbances our proposed algorithm detects rooms and estimates distances without prior knowledge in an unknown environment. Based on these measurement we automatically build a topology graph representing the room context and distances for indoor environment in a model for buildings. We evaluate our algorithm within a wireless sensor network and show the performance of our solution.

1 Introduction

In the past the number of medical sensors has increased with the advent of new types of sensors that measure various physiological parameters of patient. However, cabling of these sensors is still an open issue. Wireless interfaces have the potential to replace cables but create new issues. How can one be sure that the values displayed on the bedside monitor are really from the wireless sensor attached to the patient and not from the neighboring room? Therefore, the physical location of a wireless medical sensor within the medical environment is an important aspect of context-awareness in the sense of [11]. Context-aware medical sensors are helpful in medical applications. Bardram presents benefits of context-aware medical equipment e.g. context-aware hospital bed in [1]. In [4] we suggest a wireless barometric sensor to track a reference point during surgery for correction of the invasive measured blood pressure. However, detection if a wireless sensor is accidentally removed from a specific area e.g. a operational theater is still a problem. Context-aware medical sensors solve this problem. However, medical sensors require small, easy to deploy and low-cost indoor localization. In our work we assume a scenario where a number of wireless sensor nodes with a barometric sensor are deployed in an unknown medical indoor environment. We show how wireless medical sensors equipped with barometric pressure sensors can establish a common room context and recognize spatial relationships without any infrastructure. The contributions of the paper are as follows:

  1. We present an algorithm to detect same rooms or neighboring rooms by correlation of pressure disturbances.

  2. We build a topology graph representing the context for indoor environment in a model for buildings.

  3. We estimate distance between wireless nodes based on cross–correlation.

The rest of this paper is organized as follows: Section 2 introduces related work and motivates the need for a new approach. We will introduce our approach in combination with the implementation in Section 3. Preliminary results are provided in Section 4. In Section 5 we summarize our work and given an outlook to future work.

2 Related work

Several methods have been suggested in the past to create or establish context and localize wireless sensors. RF-based distance estimations include two-way ranging and receive signal strength indicator (RSSI) based measurements [2, 5, 6, 9]. Systems utilizing RSSI or other RF-based technologies require line of sight or experience difficulties with multipath propagation in indoor environment. In contrast, barometric sensors investigated here, provide measurements without line of sight requirements and are independent of other participants in the network. Usage of barometric sensors for context-detection and for altitude measurement was already investigated in the past. Bianchi et al. showed barometric pressure and triaxial accelerometry based fall detection in [3]. Sankaran et al. focus on context-detection (with the states: IDLE, WALKING and VEHICLE) of a person [10] with a barometric sensor. In [4] it was shown that differential pressure measurement improves accuracy compared to absolute measurement with a single sensor. In [8] Patel et al. detected movement of humans in a heating, ventilation, and air conditioning (HVAC) system with 75-80% accuracy. Our work here shows, that wireless sensors can also take advantages of barometric disturbances for distance estimation and context-detection.

Related work shows that context-detection in terms of spatial relationship without prior knowledge of the environment is not feasible with current approaches. To the best of our knowledge, we are the first to propose pressure sensors for distance estimation and room context-detection.

3 Approach and implementation

This section introduces the decentralized approach for context-aware wireless medical sensors without prior knowledge or infrastructure. Our approach is based on barometric disturbances which occur when doors are opened ore closed. This is also valid for rooms with over-pressure e.g. operational theater, since opening of a door leads to decrease of pressure.

Figure 1 shows a barometric pressure measurement of two wireless sensor nodes over time. If a door between two rooms is opened, pressure increases in one room and decreases in the other room. At t = 1 s a door was opened which decreases barometric pressure in the room of approximately 0.2 mbar. In our experiments we found out, that barometric disturbances have an duration of ≈ 400 ms. In our approach we use the last N samples for signal processing. Time synchronization between nodes ensures simultaneous detection of barometric disturbances. First, barometric disturbances are detected by analysis of the normalized auto correlation function ACF(τ). If values of the ACF at τ ≠ 0 are larger than the given confidence intervals [7] a barometric disturbance is detected.

Figure 1 Two wireless sensors in the same room experiencing barometric disturbances caused by opening a door at t = 1 s
Figure 1

Two wireless sensors in the same room experiencing barometric disturbances caused by opening a door at t = 1 s

If a disturbance is detected, we perform cross–correlation to calculate the delay τdelay of samples between barometric measurements of two spatially separated nodes. Therefore, distance can be calculated by knowing τdelay, the sampling interval Tsample and the speed of sound v. However, applying a moving window with N samples enables multiple distance estimations di within one barometric disturbance. Therefore, we suggest the Pearson’s correlation coefficient ri as quality indicator of a single measurement. The correlation coefficient ri is a measure of linear relationship between series of values. The result is in the interval [−1, 1] where −1 and 1 indicate a total negative or total positive correlation, respectively. Values of |ri| close to unity indicate strong linear dependency and therefore a reliable barometric measurement. Positive values of ri indicates that sensors are placed in the same room. For the example shown in Figure 1rmax = 0.98. Negative correlation indicates that sensors are in neighboring rooms. Distance d between two wireless sensor nodes is estimated by weighting the estimated distances di of a single barometric disturbance by following equation:

(1)d=(Ri|τdelay|)Tsamplev

where Ri stand for the normalized weights i.e. ΣRi = 1. The proposed barometric measurement is implemented at our proposed medical sensor [4]. For our practical evaluation we selected the 24 bit MS5803-01BA pressure sensor from Measurement Specialties with a sampling interval of T = 24 ms and N = 20 samples ( 0.48 s) to fully cover a single barometric disturbance.

4 Evaluation

In our evaluation four wireless sensors are distributed in different rooms on one floor, as it might be the case in a hospital. We demonstrate context detection of wireless medical sensors without prior knowledge about environment or infrastructure. Evaluation of our approach shows that removed sensors from a specific area e.g. the operational theater can be detected. Additionally, we perform distance measurements based on barometric disturbances. Our experimental setup is shown in Figure 2.

Figure 2 Placement of wireless sensor nodes on the floor
Figure 2

Placement of wireless sensor nodes on the floor

For the first experiment two nodes are placed in neighboring rooms separated by a door. Figure 3 shows the result of relative barometric pressure measurement from time 4 s to 9 s for node 4 and node 1 for an experiment. Opening of the door happened between 4 s and 5 s and between 8 s and 9 s. The maximum change of barometric pressure caused by opening a door is −0.08 mbar and −0.1 mbar for node 4 and node 1 respectively. Barometric disturbances caused by closing the door are not visible in this figure, since we closed the door gently. The negative correlation coefficient ri = −0.88 indicates that sensors 4 and node 1 are in neighboring rooms.

Figure 3 Relative barometric pressure values for node 4 and node 1 in neighboring rooms with opening and closing a door in between
Figure 3

Relative barometric pressure values for node 4 and node 1 in neighboring rooms with opening and closing a door in between

Additionally, we evaluated the performance of distance estimation between node 2 and node 3. Barometric disturbances are created by closing a door on this hallway. Distances between nodes are varied between 0 m and 24 m. Figure 4 shows the barometric pressure measurement of two nodes in the same room with a distance of d = 24 m. Maximum change of barometric pressure is −0.1 mbar for both nodes. Additionally, pressure curves are displaced against each other as assumed in Section 3.

Figure 4 Relative barometric pressure measurement results for two nodes with a distance d = 24 m at a floor
Figure 4

Relative barometric pressure measurement results for two nodes with a distance d = 24 m at a floor

Our preliminary results are summarized in Table 1 and show that the estimated distance corresponds with the true distance d and has a maximum error of ≈ 3 m. One reason for distance errors is the limited sampling interval Tsample = 0.024 ms which leads to limited resolution within our cross–correlation.

Table 1

Distance estimation of our algorithm for two nodes in the same room with varying distances

Distance [m]Estimated Distance [m]Max. Corr. Coefficient
000.92
41.070.92
85.850.92
1211.200.86
1615.720.94
2021.510.91
2423.540.89

Figure 5 shows the resulting topology graph which is automatically built with the metrics distance and correlation coefficient for the edges. A topology graph shows spatial relationships between medical sensors and specific rooms.

Context of medical sensors is estimated by the correlation coefficient, estimated distance between nodes and the detection of events like opening and closing doors. Our approach enables medical sensors in an unknown environment to detect neighboring rooms, same rooms and estimate distances between other medical sensors. This en ables a low-cost and easy to deploy solution to detect e.g. removal of sensors from an operational theater. Hence, it increases robustness and safety while having the benefits of wireless sensors. Additionally, our approach detects events like opening doors as well as open doors.

Figure 5 Undirected topology graph containing edges correlation coeflcient and distance for a distribution of nodes
Figure 5

Undirected topology graph containing edges correlation coeflcient and distance for a distribution of nodes

5 Conclusion and future work

In this paper we presented a novel approach for automated context-detection for wireless medical sensors in an unknown environment without infrastructure. We showed that barometric pressure disturbances caused by opening and closing doors can be measured. Our approach processes pressure signals between wireless medical sensors and enables detection of same rooms, neighboring rooms and estimation of distances between wireless medical sensors. This enables medical sensors to detect if they are accidentally removed from a specific area e.g. the operational theater. Furthermore we can create a topology graph of distributed wireless medical sensors with our approach. Preliminary results show that distance estimation has an maximum error of ≈ 3 m. In the next step, we will increase the sampling rate of barometric pressure measurement to increase resolution of distance estimation.

Author's Statement

  1. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animals use.

Funding

This publication is a result of the ongoing research within the LUMEN research group, which is funded by the German Federal Ministry of Education and Research (BMBF, FKZ 13EZ1140A/B). LUMEN is a joint research project of Lübeck University of Applied Sciences and Universität zu Lübeck and represents an own branch of the Graduate School for Computing in Medicine and Life Sciences of Universität zu Lübeck. This publication is also a result of the research work of the Center of Excellence CoSA in the project m:flo, which is funded by Bundesministerium für Wirtschaft und Energie (BMWi, FKZ KF3177201ED3).

References

[1] Jakob E Bardram. Applications of context-aware computing in hospital work: examples and design principles. In Proceedings of the 2004 ACM symposium on Applied computing. ACM, 2004.10.1145/967900.968215Search in Google Scholar

[2] K. Benkic, M. Malajner, P. Planinsic, and Z. Cucej. Using RSSI value for distance estimation in wireless sensor networks based on ZigBee. In Systems, Signals and Image Processing, 2008. IWSSIP 2008. 15th International Conference on, June 2008.10.1109/IWSSIP.2008.4604427Search in Google Scholar

[3] Federico Bianchi, Stephen J Redmond, Michael R Narayanan, Sergio Cerutti, and Nigel H Lovell. Barometric pressure and triaxial accelerometry-based falls event detection. IEEE Transactions on Neural Systems and Rehabilitation Engineering, (6), 2010.10.1109/TNSRE.2010.2070807Search in Google Scholar PubMed

[4] Christian Bollmeyer, Mathias Pelka, Hartmut Gehring, and Horst Hellbrück. Evaluation of Radio Based, Optical and Barometric Localization for Indoor Altitude Estimation in Medical Applications. In The Fifth International Conference on Indoor Positioning and Indoor Navigation 2014 (IPIN 2014), Busan, Korea, 2014.10.1109/IPIN.2014.7275477Search in Google Scholar

[5] Matteo Cypriani, Frédéric Lassabe, Philippe Canalda, and François Spies. Open wireless positioning system: A wi-fibased indoor positioning system. In Vehicular Technology Conference Fall (VTC 2009-Fall), 2009 IEEE 70th, 2009.10.1109/VETECF.2009.5378966Search in Google Scholar

[6] A Faheem, R. Virrankoski, and M. Elmusrati. Improving RSSI based distance estimation for 802.15.4 wireless sensor networks. In Wireless Information Technology and Systems (ICWITS), 2010 IEEE International Conference on, 2010.10.1109/ICWITS.2010.5611813Search in Google Scholar

[7] Simon S Haykin. Adaptive filter theory. Pearson Education India, 2008.Search in Google Scholar

[8] Shwetak N Patel, Matthew S Reynolds, and Gregory D Abowd. Detecting human movement by differential air pressure sensing in hvac system ductwork: An exploration in infrastructure mediated sensing. In Pervasive Computing.10.1007/978-3-540-79576-6_1Search in Google Scholar

[9] Christof Rohrig and M Muller. Localization of sensor nodes in a wireless sensor network using the nanoLOC TRX transceiver. In Vehicular Technology Conference, 2009. VTC Spring 2009. IEEE 69th. IEEE, 2009.10.1109/VETECS.2009.5073650Search in Google Scholar

[10] Kartik Sankaran, Minhui Zhu, Xiang Fa Guo, Akkihebbal L Ananda, Mun Choon Chan, and Li-Shiuan Peh. Using mobile phone barometer for low-power transportation context detection. In Proceedings of the 12th ACM Conference on Embedded Network Sensor Systems, 2014.10.1145/2668332.2668343Search in Google Scholar

[11] Bill N Schilit and Marvin M Theimer. Disseminating active map information to mobile hosts. Network, IEEE, 8(5), 1994.10.1109/65.313011Search in Google Scholar

Published Online: 2015-9-12
Published in Print: 2015-9-1

© 2015 by Walter de Gruyter GmbH, Berlin/Boston

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Articles in the same Issue

  1. Research Article
  2. Development and characterization of superparamagnetic coatings
  3. Research Article
  4. The development of an experimental setup to measure acousto-electric interaction signal
  5. Research Article
  6. Stability analysis of ferrofluids
  7. Research Article
  8. Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
  9. Research Article
  10. Energy harvesting for active implants: powering a ruminal pH-monitoring system
  11. Research Article
  12. New type of fluxgate magnetometer for the heart’s magnetic fields detection
  13. Research Article
  14. Field mapping of ballistic pressure pulse sources
  15. Research Article
  16. Development of a new homecare sleep monitor using body sounds and motion tracking
  17. Research Article
  18. Noise properties of textile, capacitive EEG electrodes
  19. Research Article
  20. Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
  21. Research Article
  22. Spike sorting: the overlapping spikes challenge
  23. Research Article
  24. Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
  25. Research Article
  26. Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
  27. Research Article
  28. Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
  29. Research Article
  30. Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
  31. Research Article
  32. Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
  33. Research Article
  34. Compressed sensing of multi-lead ECG signals by compressive multiplexing
  35. Research Article
  36. Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
  37. Research Article
  38. Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
  39. Research Article
  40. Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
  41. Research Article
  42. Effects of sampling rate on automated fatigue recognition in surface EMG signals
  43. Research Article
  44. Closed-loop transcranial alternating current stimulation of slow oscillations
  45. Research Article
  46. Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
  47. Research Article
  48. The role of expert evaluation for microsleep detection
  49. Research Article
  50. The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
  51. Research Article
  52. Metal artifact reduction by projection replacements and non-local prior image integration
  53. Research Article
  54. A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
  55. Research Article
  56. Processing of membranes for oxygenation using the Bellhouse-effect
  57. Research Article
  58. Inkjet printing of viable human dental follicle stem cells
  59. Research Article
  60. The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
  61. Research Article
  62. New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
  63. Research Article
  64. QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
  65. Research Article
  66. Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
  67. Research Article
  68. Next-generation vision testing: the quick CSF
  69. Research Article
  70. Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
  71. Research Article
  72. Design and control of a 3-DOF hydraulic driven surgical instrument
  73. Research Article
  74. Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
  75. Research Article
  76. Frequency based assessment of surgical activities
  77. Research Article
  78. “Hands free for intervention”, a new approach for transoral endoscopic surgery
  79. Research Article
  80. Pseudo-haptic feedback in medical teleoperation
  81. Research Article
  82. Feasibility of interactive gesture control of a robotic microscope
  83. Research Article
  84. Towards structuring contextual information for workflow-driven surgical assistance functionalities
  85. Research Article
  86. Towards a framework for standardized semantic workflow modeling and management in the surgical domain
  87. Research Article
  88. Closed-loop approach for situation awareness of medical devices and operating room infrastructure
  89. Research Article
  90. Kinect based physiotherapy system for home use
  91. Research Article
  92. Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
  93. Research Article
  94. Integrating multimodal information for intraoperative assistance in neurosurgery
  95. Research Article
  96. Respiratory motion tracking using Microsoft’s Kinect v2 camera
  97. Research Article
  98. Using smart glasses for ultrasound diagnostics
  99. Research Article
  100. Measurement of needle susceptibility artifacts in magnetic resonance images
  101. Research Article
  102. Dimensionality reduction of medical image descriptors for multimodal image registration
  103. Research Article
  104. Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
  105. Research Article
  106. Evaluation of CT capability for the detection of thin bone structures
  107. Research Article
  108. Towards contactless optical coherence elastography with acoustic tissue excitation
  109. Research Article
  110. Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
  111. Research Article
  112. Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
  113. Research Article
  114. Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
  115. Research Article
  116. Video tracking of swimming rodents on a reflective water surface
  117. Research Article
  118. MR imaging of model drug distribution in simulated vitreous
  119. Research Article
  120. Studying the extracellular contribution to the double wave vector diffusion-weighted signal
  121. Research Article
  122. Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
  123. Research Article
  124. Introducing a frequency-tunable magnetic particle spectrometer
  125. Research Article
  126. Imaging of aortic valve dynamics in 4D OCT
  127. Research Article
  128. Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
  129. Research Article
  130. Simple concept for a wide-field lensless digital holographic microscope using a laser diode
  131. Research Article
  132. Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
  133. Research Article
  134. Respiratory surface motion measurement by Microsoft Kinect
  135. Research Article
  136. Improving image quality in EIT imaging by measurement of thorax excursion
  137. Research Article
  138. A clustering based dual model framework for EIT imaging: first experimental results
  139. Research Article
  140. Three-dimensional anisotropic regularization for limited angle tomography
  141. Research Article
  142. GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
  143. Research Article
  144. Experimental computer tomograph
  145. Research Article
  146. US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
  147. Research Article
  148. Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
  149. Research Article
  150. Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
  151. Research Article
  152. Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
  153. Research Article
  154. An IHE-conform telecooperation platform supporting the treatment of dementia patients
  155. Research Article
  156. Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
  157. Research Article
  158. Identification of surgical instruments using UHF-RFID technology
  159. Research Article
  160. A generic concept for the development of model-guided clinical decision support systems
  161. Research Article
  162. Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
  163. Research Article
  164. Creating 3D gelatin phantoms for experimental evaluation in biomedicine
  165. Research Article
  166. Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
  167. Research Article
  168. Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
  169. Research Article
  170. Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
  171. Research Article
  172. Wireless medical sensors – context, robustness and safety
  173. Research Article
  174. Sequences for real-time magnetic particle imaging
  175. Research Article
  176. Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
  177. Research Article
  178. A machine learning approach for planning valve-sparing aortic root reconstruction
  179. Research Article
  180. An in-ear pulse wave velocity measurement system using heart sounds as time reference
  181. Research Article
  182. Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
  183. Research Article
  184. Multisegmental fusion of the lumbar spine a curse or a blessing?
  185. Research Article
  186. Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
  187. Research Article
  188. A muscle model for hybrid muscle activation
  189. Research Article
  190. Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
  191. Research Article
  192. An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
  193. Research Article
  194. Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
  195. Research Article
  196. Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
  197. Research Article
  198. Mesh structure-independent modeling of patient-specific atrial fiber orientation
  199. Research Article
  200. Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
  201. Research Article
  202. Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
  203. Research Article
  204. A java based simulator with user interface to simulate ventilated patients
  205. Research Article
  206. Evaluation of an algorithm to choose between competing models of respiratory mechanics
  207. Research Article
  208. Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
  209. Research Article
  210. Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
  211. Research Article
  212. Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
  213. Research Article
  214. FPGA controlled artificial vascular system
  215. Research Article
  216. Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
  217. Research Article
  218. Test setup for characterizing the efficacy of embolic protection devices
  219. Research Article
  220. Impact of electrode geometry on force generation during functional electrical stimulation
  221. Research Article
  222. 3D-based visual physical activity assessment of children
  223. Research Article
  224. Realtime assessment of foot orientation by Accelerometers and Gyroscopes
  225. Research Article
  226. Image based reconstruction for cystoscopy
  227. Research Article
  228. Image guided surgery innovation with graduate students - a new lecture format
  229. Research Article
  230. Multichannel FES parameterization for controlling foot motion in paretic gait
  231. Research Article
  232. Smartphone supported upper limb prosthesis
  233. Research Article
  234. Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
  235. Research Article
  236. Evaluation of adhesion promoters for Parylene C on gold metallization
  237. Research Article
  238. The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
  239. Research Article
  240. Increasing the visibility of thin NITINOL vascular implants
  241. Research Article
  242. Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
  243. Research Article
  244. Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
  245. Research Article
  246. Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
  247. Research Article
  248. Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
  249. Research Article
  250. Electrospun vascular grafts with anti-kinking properties
  251. Research Article
  252. Integration of temperature sensors in polyimide-based thin-film electrode arrays
  253. Research Article
  254. Use cases and usability challenges for head-mounted displays in healthcare
  255. Research Article
  256. Device- and service profiles for integrated or systems based on open standards
  257. Research Article
  258. Risk management for medical devices in research projects
  259. Research Article
  260. Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
  261. Research Article
  262. Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
  263. Research Article
  264. SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
  265. Research Article
  266. Numerical simulations of airflow in the human pharynx of OSAHS patients
Downloaded on 25.12.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2015-0086/html
Scroll to top button