Abstract
Although the use of minimally invasive surgery techniques has steadily increased, the development of new tools for these procedures has stagnated. Indeed a new generation of surgical instruments, with tips that have multiple degrees of freedom, has been developed. However, they are facing so many technical problems that none have been able to establish themselves in the medical market. To overcome the problems these instruments are facing, a micro hydraulic power transmission system has been developed and been presented in [1]. With these driving units it was possible to design an instrument for minimally invasive surgery with a tip which is movable in 3 degrees of freedom (DOF) and that is light in weight, small in size and powerful in movements and gripping. This paper presents the mechanical setup (including dimensions and materials), describes the theoretical basis for the control with the inverse kinematic model, discusses the external drives setup and gives first performance data of this novel hydraulically actuated laparoscopic instrument with 3 degrees of freedom.
1 Introduction
Surgical techniques have seen major developments over the last two decades, with the role of laparoscopy having an increasingly relevant place in modern medicine. The reasons for its success are the advantages for the patients, such as reduced haemorrhaging and smaller incisions, which in turn reduce pain and shorten recovery time [2, 3]. However, the tools and instruments for laparoscopic surgery have not kept up with this progress. The equipment setup in this domain has remained largely unchanged since the days in which this surgical method was introduced.
Besides some exceptions, the instrumentation consists of rigid rod-like graspers and scissors. Nonetheless, minimally invasive procedures have continued to evolve and new operation modes have been developed such as Single Port and NOTES surgery. These procedures aim to reduce complications related to port placement, but they make high demands on the instrumentation setup [4]. Even though innovative and novel instruments have been developed, such as articulating and flexible shaft instruments, none of them have been successfully introduced in surgical procedures. Due to limited force transmission, impossibility of robust grasping [5] and complicated handling [6], these mechanical multi degree-of-freedom (DOF) instruments have never really found their application [7].
By integrating electromechanical drives into the instrument’s handle to actuate the tip’s DOFs, it was possible to overcome some of the disadvantages described above. Unfortunately, this created new problems due to the weight of the motors and hence this solution failed before it could become widespread. This is supported by the fact that one of these systems has already been withdrawn from the market.
Fraunhofer PAMB, instead, proposes in [1], a new approach: laparoscopic instruments powered by hydraulics. This technology has the advantage that the actuators have a higher force and power density than most of the others [9] and, because of their dimensions, they can be located directly on the tip. The instruments featuring this technology are both small and light as their drives (the pumps producing hydraulic pressure) can be located remotely. Moreover, sensors to measure actuator parameters do not contribute to the instrument’s weight, as they can be located anywhere in the hydraulic circuit. In this paper, details about the instrument technology will be given together with information about the control concept - namely the drive set up and instrument’s inverse kinematic equations.
2 Methods
To determine the specifications of a handheld surgical instrument with multiple degrees of freedom at the tip, requirement analyses have been conducted [9].
Together with the problems existing instruments are facing, this breakdown has shown that it is impossible to develop mechanically driven instruments that fit the specifications. The main reason being the combination of holding the instrument, moving and applying force on the instrument’s tip for multi degrees of freedom.
To solve this, mechanical decoupling of the movements and forces applied by the doctor to the user interface and the movements and forces working on the instrument’s tip is necessary. This decoupling necessitates builtin drives in the instrument [5]. Hence possible drive solutions have been explored. As result, actuation with micro hydraulic systems was the only solution to fulfil the requirements listed above.
Subsequently the usability of hydraulic systems for medical applications was verified. Therefore potential hydraulic fluids compatible with use in medical devices have been examined. Then the feasibility of sterilizing hydraulic systems filled with these fluids was investigated [1].
3 Results
3.1 Mechanical setup with external drives
It was possible to develop a surgical instrument with 3 hydraulically actuated DOF shown in Figure 3. The instrument’s tip consists of three segments connected via two cylindrical joints which are o?set by 90 degrees about their longitudinal axis. The joints are connected in a serial chain and each joint is actuated by an internal mechanism consisting of two antagonistic linear hydraulic actuators.
In total, the movements of the tip are powered by 6 integrated micro hydraulic cylinders with a piston diameter of 2.45 mm. Being designed for a maximum working pressure of 200 bar, they can produce an actuator force of 94N each. Sealed with micro O-rings which are lubricated with medical white oil, their average internal friction can be kept below 2N, about 2% of the total actuator force. All the actuators and mechanical parts can be integrated in a cylindrical shape of 8 mm in diameter.
The instrument’s cylinders are powered by hydraulic syringe pumps: each cylinder of the instrument is connected with its own pump over a closed circuit. For filling and bleeding out the air, the circuit can be opened with valves. This hardware configuration has the advantage to be free from the non-linearity associated with the use of valves in hydraulic circuits. Moreover, since every syringe pump is directly connected to one of the instrument’s cylinders, equipping the syringe pump with sensors makes it possible to acquire information about the corresponding cylinder in the instrument. A position or a pressure sensor in every syringe pump makes a straightforward implementation of a closed loop control possible without increasing the instrument’s weight, as the sensors are not integral to the instrument, but located within the pump.

Picture of the 3 DOF instrument tip, actuated by 6 hydraulic cylinders 3.
3.2 Control concept and inverse kinematic model
The control concept can be summarized as follows: The control software receives the desired configuration of the tip as an input; this information is converted by the inverse kinematics module into actuator extensions and sent as a command to each syringe. The syringe pump displacement translates directly to an actuator extension with a known transmission ratio.
The first step for the implementation of this control system is the definition of the inverse kinematic equations. Given the mechanical design of the instrument, a relationship needs to be established between the actuator extensions, hence the joint rotations, and finally the tool tip position. The equations will be derived by applying inverse kinematic theory for both mechanisms: the serial chain and the internal mechanisms. The relationship between tool tip position and joint rotations can be established by applying to the serial chain the inverse serial kinematic method of Denavit-Hartenberg (DH). The relationship between the joint rotations and the actuator extensions can be determined by applying a geometrical method to the internal mechanism. The details of the derivations are described in the following paragraphs.

Image showing the instrument’s tip from two sides: a frontal face and a view of the instrument rotated by 90 degrees. The frame’s position and orientation, placed on the joints accordin to the Denavit-Hartemberg convention, are also defined.
Denavit-Hartemberg Parameters, defined for finding the relationship between instrument tool tip position and joint rotation.
| joint | theta | d | a | alpha |
|---|---|---|---|---|
| 1 | 0 | 0 | a | 0 |
| 2 | thet1 | 0 | b | 90 |
| 3 | thet2 | 0 | c | 0 |
In order to find the relationship between tool tip position and joint rotations, reference frames of the mechanism’s joints were affixed according to the Denavit-Hartemberg convention [10], as shown in Fig. 2 and the DH parameters were defined as shown in Table 1 where:
– a is the distance between zi and zi+1 measured along xi;
– α is the angle from zi to zi+1 around xi;
– d is the distance between xi−1 and xi measured along zi;
– is the angle from xi−1 to xi around zi;
Using these parameters it is then possible to define the homogeneous transformation matrix
The last column of the matrix (1) represents the vector that links the tip reference frame and the origin. Its components are defined as:
– a is the distance between reference frame and joint 1;
– b is the distance between joint 1 and joint 2;
– c is the distance between joint 2 and the tip frame;
Hence from equation (4) it is possible to deduce the relationship between end-effector jaw position and joint rotations:
With the relationship between tool tip position and joint rotation defined, a relationship can be established between the joint rotation and actuator extensions from the inverse kinematics of the internal mechanism. A schematic of this mechanism is shown in Fig. 3.

Schematic diagram showing the parallel kinematic mechanism.
Defined as: hp the x coordinate of the joint; La, Lb and, Lc the segment dimensions of the parallel kinematic; θb1 and θb2 the angle between the first and the second segment of the two branches of the mechanism; θc1 and θc2 the angle between the third segment of the two mechanism branches and a horizontal line; x1 and x2 the distance between piston and cylinder base.
A geometrical method is used to establish the following kinematic relationships:
The identities resulting from the serial and internal mechanisms are linked by the following geometrical relationship:
Where Cθ is a constant angle due to the parallel kinematic structure (Cθ is the value of θc1 when x1 = x2). (12) Defines that a rotation of the joint corresponds to an equal change of θc1 from its equilibrium position. It can be concluded that once the end-effector position (px; py and pz) is defined, θ1 and θ2 can be calculated from (5) and (6). Knowing θ1, the orientation of joint 1, it is possible to calculate θc1 from (12) and then θc2 from (11). At last θc1 and θc2 can be respectively substituted in (8) and (10), obtaining x1 and x2, the piston extensions.
4 Conclusion
A hydraulically actuated surgical instrument with a multiple DOF tip was designed. As these systems can be built as lightweight, powerful and small assemblies, they have the potential to overcome the majority of problems that similar mechanically or electrically powered instruments face.
As even small hydraulic cylinders with diameters of 2.5mm can produce forces up to 100N, it is possible to place the actuation of the tips DOF directly in the tip.
The hydraulic power can be transmitted through thin tubes, allowing the power units used to produce the hydraulic pressure to be placed beside the operating table. It is therefore possible to build powerful surgical instruments not exceeding the weight of those being actuated by hand.
Furthermore, a new control concept for a new hydraulically actuated instrument has been developed together with the mathematical description of its inverse kinematic. The pressure supply configuration, with 6 syringe pumps driving 6 hydraulic cylinders, has the advantages of a direct transmission of the pump displacement to the hydraulic actuator extension as well as enabling a straightforward implementation of closed-loop control.
– Sensor integration in the instrument cylinders to enhance safety of closed loop control via redundancy.
– Development of alternative pressure supply methods which can match the current system’s performance but improve the current solution in terms of size and cost.
Author’s Statement
Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.
References
[1] T. Cuntz, G. James, et al., “Next generation surgical instruments powered by hydraulics,” Biomedical Engineering/Biomedizinische Technik, 2013.10.1515/bmt-2013-4399Search in Google Scholar PubMed
[2] T. Carus, Operationsatlas Laparoskopische Chirurgie. Springer, 201010.1007/978-3-642-01707-0Search in Google Scholar
[3] P. Dasgupta, J. Fitzpatrick, et al., New Technologies in Urology. Springer, 2010, vol. 710.1007/978-1-84882-178-1Search in Google Scholar
[4] R. Autorino, R. J. Stein, et al., “Current status and future perspectives in laparoendoscopic single-site and natural orifice transluminal endoscopic urological surgery,” International journal of urology: official journal of the Japanese Urological Association, vol. 17, no. 5, pp. 410–31, May 2010.10.1111/j.1442-2042.2010.02497.xSearch in Google Scholar PubMed
[5] J. H. Kaouk, R. Autorino, et al., “Laparoendoscopic single-site surgery in urology: worldwide multi-institutional analysis of 1076 cases.,” Eur. Urol., vol. 60, no. 5, pp. 998–1005, Nov. 201110.1016/j.eururo.2011.06.002Search in Google Scholar PubMed
[6] F. Gaboardi, A. Gregori, et al. (2011). ‘LESS’radical prostatectomy: a pilot feasibility study with a personal original technique. BJU international, 107(3), 460-464.10.1111/j.1464-410X.2010.09475.xSearch in Google Scholar PubMed
[7] P. Dasgupta, J. Fitzpatrick, et al., “New Technologies in Urology”. Springer, 201010.1007/978-1-84882-178-1Search in Google Scholar
[8] M. De Volder, “Pneumatic and hydraulic micro actuators: a new approach for achieving high force and power densities at micro scale,” status: published, 2007.Search in Google Scholar
[9] T. Cuntz, “Untersuchungen zur Eignung mikrohydraulischer Antriebe für die minimal invasive Chirurgie”, in pressSearch in Google Scholar
[10] J. J. Craig, Introduction to Robotics, P. E. internationalSearch in Google Scholar
© 2015 by Walter de Gruyter GmbH, Berlin/Boston
This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients