Abstract
Electrospinning is a versatile method of producing micro- and nanofibers deposited in mats used as scaffolds for tissue engineering. Depending on the application, single or coaxial electrospinning can be used. Coaxial electrospinning enables the use of a broad spectrum of materials, the fabrication of hollow or core/shell fibers and an automatisation of the entire electrospinning process. In this regard, the design of coaxial nozzles plays a major role in a standardized as well as tailor-made scaffold fabrication. For this purpose an optimised coaxial nozzle has been designed and fabricated. Furthermore, tests have been carried out to validate the new nozzle design. With the use of the costum-made nozzle polymer concentration could be varied in a gradual manner. The variation in polymer concentration lead to fiber diameters between 0.75 to 3.25 μm. In addition, an increase in rotating velocity lead to an increase in fiber alignment as well as a slight decrease in fiber diameter. The demonstrated modifications of coaxial electrospinning proved to be a powerful tool for in-process adjustments of scaffold fabrication.
1 Introduction
Tissue engineering is known to be a promising approach to solve issues in medical treatment related to the limited availability of donor tissue [1, 2]. The process is based on the seeding of scaffolds with autologous human cells. One of the greatest challenges in fabricating scaffolds for tissue engineering deals with the imitation of the morphology of native structures [9]. Most tissues consist of multiple tissue layers and every corresponding cell type needs a specific structure or material to grow. Such complex multilayered structures can be created by electrospinning. Electrospinning is an effective and adequate way of producing tailor-made scaffolds from polymeric biomaterials. There are two commonly used setups: single and coaxial electro-spinning. Compared to single electrospinning the coaxial approach uses two cannulaes adjusted coaxially in each other [9]. This is usually done with the use of custom-made nozzles. A typical coaxial electrospinning setup is made up of two coaxially arranged spinnerets, two reservoirs with pumps, a high voltage supply and a grounded collector see Figure 1 [3, 9].

Coaxial setup with reservoirs, spinnerets, high voltage supply and collector.
The described setup has a lot of benefits regarding the fabrication of multi-layered structures due to the coaxial arrangement of the spinnerets and the independent loops for the inner and outer fluids. Furthermore, coaxial electrospinning allows the fabrication multi-fiber structures and enables the use of polymer solutions with high evaporation rates by ’coating’ the solution stream in additionally solvent. It is also possible to produce core-shell fibres [9]. This can be used for drug delivery systems by controlling the rate of diffusion of the drug releasing core fibre with the shell fibre. The aim of this study is to explore new applications for the coaxial electrospinng and to investigate different parameters on the process. More precisely the in-process adjustment of polymer concentration and the relationship between fiber diameter and polymer concentration.
1.1 Design of the coaxial nozzle
Different types of nozzles are used and described in the literature [3, 4]. The design of a new nozzle concept aimed to optimize parameters like easy and reproducible coaxial adjustment of the spinnerets, a small dead volume and a fast assembly and cleaning process. The new nozzle is built up of four parts, each made of stainless steel. They are connected via metric threads between the outer cylinder and tip as well as the outer cylinder and top screw nut (Figure 2).

Revised nozzle consisting of tip, inner cylinder, outer cylinder and top screw nut without any gaskets.
The adjustment is realised by four screws on the extent. It is possible to loosen the tip and the top screw nut to clean the nozzle, without changing the positions of the spinnerets. The dead volume is defined by the gap between tip and inner cylinder and the length of the blunt cannulaes pressed in the inner cylinder. Two holes in the top screw are acting as the solution supply. Cannulaes can be stung directly through a sealing into the press fitted blunt cannulaes. The handling and cleaning remains easy because of the small dead volume and the modular concept of the nozzle.
2 Material and methods
2.1 Scaffold fabrication
ε−polycaprolactone (PCL) was purchased from Sigma-Aldrich and the solvent 2,2,2-trifluorethanol from ABCR GmbH (Karlsruhe, Germany).
Two polymer reservoirs were connected to the coaxial nozzle. The inner fluid had a concentration of 100 mg/ml, 200 mg/ml being used as the outer fluid. The flow rates have been calculated on the following formula:
Concentrations of 100, 125, 150, 175 and 200 mg/ml were used as references. The influence of polymer concentration fiber diameter was investigated. The assumption was, that a higher polymer solution would result in a bigger fiber diameter [5]. In addition, the influence of rotation velocity on the level of anisotropy was observed. Referring to the literature, fiber alignment increases with rotation velocity [5–8]. Scaffolds were spun using a custom-made setup consisting of a high voltage supply, two syringe pumps and a rotating drum collector. Both, single and coaxial setup samples were spun with a collector-emitter distance of 250 mm, voltage of 25 kV, a temperature of 25°C, a humidity between 50% and 64% and a flow rate of 4 ml/h overall. Reference measurements were recorded for 100, 125, 150, 175 and 200 mg/ml and also for the coaxial tests. Furthermore, the collectors’ rotation speed was set to 0, 500 and 1000 rpm. For the coaxial tests, the flow rates were calculated according to [E1] and are displayed in Table 1.
Samples have been sputter coated and several images have been taken with a SEM (magnifications of 500x, 1000x, 2000x, 8000x). Carl Zeiss AxioVision software has been used to measure fiber diameters.
Flow rates of each reservoir and pump for concentrations of 100 (pump 1) and 200 (pump 2) mg/ml.
| Concentration [mg/ml] | FRtotal [ml/h] | FRpump1 [ml/h] | FRpump2 [ml/h] |
|---|---|---|---|
| 100 | 4 | 4 | 0 |
| 125 | 4 | 3 | 1 |
| 150 | 4 | 2 | 2 |
| 175 | 4 | 1 | 3 |
| 200 | 4 | 0 | 4 |
3 Results and discussion
The results of the fiber diameter analysis of the single setup are shown in Figure 3. The graph depicts the fiber diameter in relation to the concentration and the rotation speed of the collector. The mean values are located between 0.4 μm for 100 mg/ml at 1000 rpm and 3.3 μm for 200 mg/ml at 0000 rpm. For samples with a rotating velocity, fiber diameter increased from 0.7 μm to 3.3 μm with increasing polymer concentration.

Graph of fiber diameters in relation to polymer concentration. An increasing fiber diameter (0.4 μm to 3.3 μm) with increasing polymer concentrations (100 mg/ml to 200 mg/ml) was observed.
The other groups showed mean values between 0.4 μm and 3.3 μm for 0500 rpm and between 0.4 μm and 2.9 μm for 1000 rpm. Not only polymer concentration had an influence on fiber diameter. In addition rotating velocity influenced the fibers. The higher the speed, the thinner the fibers, e.g. 0.73 μm at 0000 rpm and 100 mg/ml, 0.42 μm at 0500 rpm and 0.41 μm at 1000 rpm.
In addition, a relationship between collectors’ rotation velocity and the quantity of aligned fibers could be observed. To investigate the level of anisotropy SEM-images were prepared (Figure 6). It was shown, that the anisotropy does not only depend on the rotation speed but also on the fiber diameter.

Analysis of the SEM-images, magnification ×1.00k, taken of the single set-up results, obviously increasing diameter with increasing polymer concentration for 0000 rpm and increasing alignment of the fibers from 0000 rpm to 1000 rpm.

Graph of fiber diameters in relation to polymer concentration. An increasing fiber diameter (1 μm to 1.5 μm) with increasing polymer concentrations (100 mg/ml to 200 mg/ml) was observed.

SEM-image of a sample made from a solution with a concentration of 125 mg/ml at 0000 rpm, depicting the distribution of thick and thin fibers.
The influence of the collectors’ speed decreases with increasing fiber diameter. From this, it follows that the rotational speed has to be increased along with the fiber diameter in order to achieve comparable levels of anisotropy through the different values of polymer concentration, but also that the concentration has to be increased to achieve the desired diameter.
The analysis of the coaxial tests has shown the same correlation between the fiber diameter and polymer concentration. Compared to the single setup results, there are differences in the absolute values and the gradient, but the trend is still noticeable (Figure 5). The mean values for the coaxial tests are located between 1 μm and 1.5 μm.
The distribution of thick and thin fibers in some samples (Figure 6) stands out from the other results. This occurred at the coaxial tests with 125 mg/ml at 0000 rpm, diameters between 0.29 and 1.49 μm have been measured.
4 Conclusion
Tissue engineering is dependent on the affinity of different human cells to several fiber diameters and pore sizes. To serve these needs the production of scaffolds with the right properties are indispensable. The results of this study demonstrate a solution for in-process adjustment of polymer concentration. It is not longer necessary to change solutions during the process. Based on this, the fabrication scaffolds with distributions of fibers with different diameters or multi-layered structures can be rapidly realized. The fiber diameter has been successfully influenced by the concentration of the used polymer solutions. The higher the concentration the bigger the diameter of the fibers. The level of anisotropy of the fibers has been regulated by the rotational speed of the collector. While the level has been shown to increase with increasing rotation velocity, an opposing effect of the polymer concentration has also been observed. Only for samples with 100 and 125 mg/ml polymer concentration a meaningful change in the amount of aligned fibers has been observed. At higher concentrations such as 150, 175 and 200 mg/ml the level of anisotropy has been lower than at the 100 and 125 mg/ml samples.
This study presented promising tools and methods to modify the micro- and macroscopic structures of electrospun scaffolds. These methods can be used to tailor scaffolds’ properties to specific applications in the field of tissue engineering.
Author’s Statement
Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.
References
[1] Laurencin, Cato T., et al. ”Tissue engineering: orthopedic applications.” Annual review of biomedical engineering 1.1 (1999): 19-46.10.1146/annurev.bioeng.1.1.19Suche in Google Scholar
[2] Huang, Zheng-Ming, et al. ”A review on polymer nanofibers by electrospinning and their applications in nanocomposites.” Composites science and technology 63.15 (2003): 2223-2253.10.1016/S0266-3538(03)00178-7Suche in Google Scholar
[3] Li, Fengyu, Yanlin Song, and Yong Zhao. Core-Shell nanofibers: Nano channel and capsule by coaxial electrospinning. INTECH Open Access Publisher, 2010. 419-421.10.5772/8166Suche in Google Scholar
[4] Russo, Maria Vittoria, Ilaria Fratoddi, and Iole Venditti. “Nanostructured macromolecules.” Advances in Macromolecules. Springer Netherlands, 2010. 62-63.10.1007/978-90-481-3192-1_1Suche in Google Scholar
[5] Enz, Eva: Electrospun Polymer - Liquid Crystal Composite Fibres, Naturwissenschaftliche Fakultät II der Martin-Luther-Universität Halle-Wittenberg, Diss., 2013. 27, 32.Suche in Google Scholar
[6] Teo, W. E., and S. Ramakrishna. ”A review on electrospinning design and nanofibre assemblies.” Nanotechnology 17.14 (2006): R89. R92.10.1088/0957-4484/17/14/R01Suche in Google Scholar
[7] Sundaray, Bibekananda, et al. ”Electrospinning of continuous aligned polymer fibers.” Applied physics letters 84.7 (2004): 1222-1224.10.1063/1.1647685Suche in Google Scholar
[8] Carnell, Lisa S., et al. ”Aligned mats from electrospun single fibers.” Macromolecules 41.14 (2008): 5345-5349.10.1021/ma8000143Suche in Google Scholar
[9] Szentivanyi A., Chakradeo T., Zernetsch H., Glasmacher B.: Electrospun Cellular Microenvironments: Understanding Controlled Release and Scaffold Structure. Advanced Drug Delivery Reviews, 2011; 30;63(4-5):209-2010.1016/j.addr.2010.12.002Suche in Google Scholar PubMed
© 2015 by Walter de Gruyter GmbH, Berlin/Boston
This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Artikel in diesem Heft
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients
Artikel in diesem Heft
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients