Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS

Preliminary results of model evaluation measurements
  • K. Rackebrandt EMAIL logo and H. Gehring
Published/Copyright: September 12, 2015

Abstract

The oxygenation, perfusion and metabolism of the brain - segmented in both hemispheres - can be estimated from the oxygenation and hemoglobin levels of the venous blood in the cerebral efferent vessels.

We present a phantom based model to simulate the anatomical target region which was connected to hemodynamic perfusion circuit to provide different oxygenation rates inside of the simulated target vessel (measurement cell) reproducible. A triple-wavelength (770, 808 and 850 nm) multi-distance NIRS sensor (6 photodiodes, linearly arranged, separated 6 mm each) was used to detect these different saturation levels.

The results illustrate the capability to measure the optical property variation of hemoglobin due to oxygenation and deoxygenation processes in a specific vessel. Based on these first results a series of measurements is introduced to correlate the amount of reflected light to the actual oxygen saturation of the blood.

1 Introduction

Several non-invasive near-infrared spectroscopy (NIRS) monitoring systems were developed to determine blood oxygenation in the human head. In this context the oxygen content of the cerebral efferent vessels is a physiological parameter for the estimation of the perfusion and the metabolism of the brain.

The oxygen content of the venous blood is defined as,

(1)Cvo2=(k1×tHb×Svo2)+(k1×Pvo2)

where k1 is the Hüfner’s constant, tHb is the total hemoglobin concentration, SvO2 is the venous oxygen saturation, k2 is the solubility coefficient of oxygen and PvO2 is the venous partial pressure of oxygen [1].

The NIRS systems measure the amounts of oxygenated and deoxygenated hemoglobin (O2Hb & HHb) for the estimation of tHb and SvO2. To identify these hemoglobin derivates these systems need a light source with at least three wavelengths, the first one at the isosbestic point of O2Hb and HHb (about 800 nm), the second one below and the third one above of this point. At different oxygenation levels the amount of O2Hb and HHb vary and therefore the optical properties as well. The diagnostic window ranges from 600-1000 nm (Figure 1). Only in this wavelength region light can penetrate the scalp and the skull with a sufficient depth to sense the target vessel (sigmoid sinus) [2]. For a reflective multi-distance NIRS system the irradiated light can be expected to follow a curved path through the head for each source detector pair [3]. The penetration depth of the light corresponds with the wavelength as well as the source and detector separation. The optical properties of the tissue are linked with the amount of reflected and transmitted light and they are mainly characterized by the absorption coefficient µa 1/mm, the reduced scattering coefficient µs’ 1/mm and the anisotropy g (g = <cos(θ)>; θ = Deflection angle).

Figure 1 Absorption spectra of oxygenated (O2Hb) and deoxygenated (HHb) hemoglobin (Based on [10]).
Figure 1

Absorption spectra of oxygenated (O2Hb) and deoxygenated (HHb) hemoglobin (Based on [10]).

The anatomical target region consists of four layers, the scalp (Layer1, 4-8 mm), the skull (Layer 2, 4-10 mm), the blood inside of the sigmoid sinus (Layer 3, Ř 5 mm) and the brain tissue (Layer 4) in the depth [5, 6].

2 Methods

2.1 Model setup

The model setup is based on a similar model presented by [7]. Two glass beakers made from DURAN (DURAN Group GmbH), with a transmission of 99% in the interesting wavelength region, were arranged non-concentrically. The thickness of the gap between these beakers were 8 mm (Figure 2). A measurement cell was attached to the inner side of the inner beaker, simulating the target vessel (Layer 3 – Blood inside of the cell). The cell was connected to the perfusion circuit, to generate realistic optical property changes in the sample volume.

Figure 2 Layered phantom model. Layer 1: Scalp; Layer 2: Skull; Layer 3: Blood inside the sigmoid sinus; Layer 4: Brain.
Figure 2

Layered phantom model. Layer 1: Scalp; Layer 2: Skull; Layer 3: Blood inside the sigmoid sinus; Layer 4: Brain.

The phantom model consists mainly of two intralipid based phantoms, simulating the scalp & skull (Layer1+2) and the brain (Layer 4). The scalp & skull phantom was placed in the gap (8 mm) between the glass beakers and the brain phantom inside of the inner beaker. The manufacturing process of these phantoms is presented elsewhere [8].

The layered structure and the calculated optical properties of the phantoms are visualized in Table 1. The optical properties were determined with the usage of a double integrating sphere setup and the IAD algorithm [9]. The whole setup was covered with a black box made from black POM and neoprene to avoid the interference of the ambient light.

Table 1

Phantom structure and optical properties of the different Layers.

Wavelength / nmLayer 1+2 -Scalp & skullLayer 4 -Brain
μa / 1/mm0.0320.020
770μs′ / 1/mm0.7801.220
g0.8400.800
μa / 1/mm0.0340.038
808μs′ / 1/mm0.7991.320
g0.8400.760
μa / 1/mm0.0260.019
850μs′ / 1/mm0.8601.420
g0.8200.830

2.2 Perfusion circuit

The perfusion circuit consists of a pump (Bio-Medicus 540 Bio Console centrifugal pump 120 V, 60 Hz., Medtronic GmbH, 1000 rpm) to generate a venous flow, a heating unit (D8-G, Thermo Haake) to ensure a blood temperature of about 37 °C and an oxygenator (D901 Dideco LiLLiPut, Sorin Group Italia) with an included blood reservoir, which was connected to the gas supply, to provide defined oxygenation levels. By adjusting the inflow of oxygen and nitrogen the amounts of oxygenated and deoxygenated hemoglobin could be varied and therefore the optical properties of the blood.

A schematic overview of the designed perfusion circuit is visualized in Figure 3. Two sample ports were integrated to collect the blood samples in front of and behind of the oxygenator to ensure its functionality. Additionally a reference measurement unit, consisting of a white light source (SL1, Stellarnet Inc.) and a detector (EPP2000, Stellarnet Inc.)) recording the spectrum of the blood, was attached. For the experiments erythrocyte concentrate was diluted with sodium chloride (Sterofundin ISO 1/1 E ISO, B. Braun Melsungen AG) until the total hemoglobin concentration (tHb) was equal to 12 g/dl.

Figure 3 Schematic of the perfusion circuit connected to model setup.
Figure 3

Schematic of the perfusion circuit connected to model setup.

2.3 Hardware and software setup

The hardware for the reflective sensor can be divided into a light source and a receiving unit (Figure 4). The former was equipped with three LEDs (Series 330 – 1206, OSA Opto Light GmbH) 770, 808 and 850 nm and a power of about 2 mW driven in continuous mode. The light sources were operated with 2 V DC. The latter consists of 6 photo diodes (PD) (BP 104 S, OSRAM Opto Semiconcutors GmbH) separated 6 mm each with peak sensitivity at 850 nm. Each PD had a sensitive area of 4.8 mm2.

Figure 4 Sensor configuration and target vessel position.
Figure 4

Sensor configuration and target vessel position.

The signals were acquired with a data acquisition card (USB 6259, National Instruments, TX, USA) at a sampling rate of 1 kHz. A low-pass filter (5th order Sallen-Key) with a cut-off frequency of 30 Hz was implemented in the hardware circuit to cut off noise from the power supply before further signal processing.

3 Results

To identify the needed distances of the photodiodes and the light source to measure exactly the sample volume (Layer 3), the results for all six PDs were analyzed separately. The measurements M1-M3 were separated in two parts, the inflow of oxygen and the inflow of nitrogen. In the Figures 57 the detected signals for PD3 are visualized since an SDS of 18 mm is approx. equal to a penetration depth of about 9 mm, which is within the simulated target vessel. The solid black lines in the images mark the change of the gas inflow and the dashed ones the points of the reference measurements.

Figure 5 Oxygenation changes at 770 nm.
Figure 5

Oxygenation changes at 770 nm.

Figure 6 Oxygenation changes at 808 nm.
Figure 6

Oxygenation changes at 808 nm.

Figure 7 Oxygenation changes at 850 nm.
Figure 7

Oxygenation changes at 850 nm.

The results from the series of measurements are visualized in Table 2.

4 Discussion

The results of the experiments proved the capability to detect the optical property variation of hemoglobin due to oxygenation and deoxygenation processes in the anatomical target region (Layer 3) with the presented setup.

For all three wavelengths the signal intensity changes seemed reasonable under the consideration of the associated absorption coefficients for O2Hb and HHb at these points. Since the inflow of oxygen increased the amount of O2Hb the measured signal intensity for 770 nm was heightened and for 850 nm reduced. The amount of HHb was increased by the inflow of nitrogen which raised the signal intensity at 850 nm and lowered it at 770 nm.

Table 2

Overview of the measurements, results and conclusions.

MeasurementResult and conclusion
M1_1 (770 nm, 6 l/min nitrogen)Decreased signal → High concentration of HHb → High absorption for 770 nm
M1_2 (770 nm, 4 l/min oxygen)Increasing signal → Low concentration of HHb → Low absorption for 770 nm
M2_1 (808 nm, 6 l/min nitrogen) M2_2 808 nm, 4 l/min oxygen)Nearly constant signal → No concentration changes at the isosbestic point → No changes in the absorption
M3_1 (850 nm, 6 l/min nitrogen)Increasing signal → Low concentration of O2Hb → Low absorption for 850 nm
M3_2 (850 nm, 4 l/min oxygen)Decreased signal → High concentration of O2Hb → High absorption for 850 nm

Figure 6 revealed a small intensity change for a wavelength of 808 nm, although it should be constant at the isosbestic point of both substances. This will be investigated in further measurements to verify if this signal intensity change is noise or if it’s a systematic issue linked to the setup model and the perfusion circuit. Since a small change in the tHb concentration (12-10 g/dl) was determined during the experiments with Co-Oximeter, this might be the reason for the signal variation at 808 nm.

5 Conclusion

Based on these first results a series of measurements (n=10) is planned to verify the results of the preliminary experiments. Since the three wavelength were investigated consecutively there might be a slide variation of the model setup & hemodynamic circuit properties. For the planned experiments the three LEDs will be driven in a pulsed mode to avoid any property shift during the measurements. The main objective is the correlation of the amount of reflected light to the oxygen saturation of the blood in the target volume.

The preliminary results revealed a small variation of the tHb in the perfusion circuit which will be cancelled out by increasing the pre-experimental time to get a completely mixed system upfront.

Acknowledgement

This publication is a result of the ongoing research within the LUMEN research group, which is funded by the German Federal Ministry of Education and Research (BMBF, FKZ 13EZ1140A/B). LUMEN is a joint research project of Lübeck University of Applied Sciences and Universität zu Lübeck and represents an own branch of the Graduate School for Computing in Medicine and Life Sciences of Universität zu Lübeck.

Author's Statement

  1. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.

References

[1] McLellan SA, Walsh TS. Oxygen delivery and hemoglobin. Continuing Education in Anesthesia 4 (4). 200410.1093/bjaceaccp/mkh033Search in Google Scholar

[2] Cormick W, Stewart M, Goetting MG, Dujovny M, Lewis G, Ausman JI. Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans. Stroke. pp. 596-602. 199110.1161/01.STR.22.5.596Search in Google Scholar PubMed

[3] Okada E, Delpy DT. Near-infrared light propagation in an adult head model. II. Effect of superficial tissue thickness on the sensitivity of the near-infrared spectroscopy signal. Applied Optics. Vol. 42. No. 16. 200310.1364/AO.42.002915Search in Google Scholar PubMed

[4] Takatani S, Graham MD. Theoretical analysis of diffuse reflectance from a two-layer tissue model. IEEE Trans. Biomed. Eng. BME-26. 656-664. 198710.1109/TBME.1979.326455Search in Google Scholar PubMed

[5] Moreira-Gonzalez F, Papay E, Zins JE. Cavarial Thickness and Its Relation to Cranial Bone Harvest. American Society of Plastic Surgeons. Vol. 117 (6). 200510.1097/01.prs.0000209933.78532.a7Search in Google Scholar PubMed

[6] Persons FG. The thickness of the living scalp. University of London. Anatomy LXIII. pp. 427-429. 1929Search in Google Scholar

[7] Almajidy RK, Hofmann UG. On the design of a multi-channel NIRS system to monitor functional brain activity. 16th International Conference on Near Infrared Spectroscopy. La Grande-Motte. 2013Search in Google Scholar

[8] Rackebrandt K, Gehring H. Measuring the oxygen content of the cerebral efferent vessels: First steps to a sensor design. Biomedical Engineering. Volume 59 Track L. Issue s1. pp. 832-835, October 2014Search in Google Scholar

[9] Prahl S. Optical Property Measurements using the Inverse Adding-Doubling Program. 1999Search in Google Scholar

[10] http://omlc.ogi.edu/spectra/hemoglobin/summary.htm.Search in Google Scholar

Published Online: 2015-9-12
Published in Print: 2015-9-1

© 2015 by Walter de Gruyter GmbH, Berlin/Boston

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Articles in the same Issue

  1. Research Article
  2. Development and characterization of superparamagnetic coatings
  3. Research Article
  4. The development of an experimental setup to measure acousto-electric interaction signal
  5. Research Article
  6. Stability analysis of ferrofluids
  7. Research Article
  8. Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
  9. Research Article
  10. Energy harvesting for active implants: powering a ruminal pH-monitoring system
  11. Research Article
  12. New type of fluxgate magnetometer for the heart’s magnetic fields detection
  13. Research Article
  14. Field mapping of ballistic pressure pulse sources
  15. Research Article
  16. Development of a new homecare sleep monitor using body sounds and motion tracking
  17. Research Article
  18. Noise properties of textile, capacitive EEG electrodes
  19. Research Article
  20. Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
  21. Research Article
  22. Spike sorting: the overlapping spikes challenge
  23. Research Article
  24. Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
  25. Research Article
  26. Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
  27. Research Article
  28. Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
  29. Research Article
  30. Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
  31. Research Article
  32. Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
  33. Research Article
  34. Compressed sensing of multi-lead ECG signals by compressive multiplexing
  35. Research Article
  36. Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
  37. Research Article
  38. Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
  39. Research Article
  40. Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
  41. Research Article
  42. Effects of sampling rate on automated fatigue recognition in surface EMG signals
  43. Research Article
  44. Closed-loop transcranial alternating current stimulation of slow oscillations
  45. Research Article
  46. Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
  47. Research Article
  48. The role of expert evaluation for microsleep detection
  49. Research Article
  50. The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
  51. Research Article
  52. Metal artifact reduction by projection replacements and non-local prior image integration
  53. Research Article
  54. A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
  55. Research Article
  56. Processing of membranes for oxygenation using the Bellhouse-effect
  57. Research Article
  58. Inkjet printing of viable human dental follicle stem cells
  59. Research Article
  60. The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
  61. Research Article
  62. New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
  63. Research Article
  64. QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
  65. Research Article
  66. Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
  67. Research Article
  68. Next-generation vision testing: the quick CSF
  69. Research Article
  70. Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
  71. Research Article
  72. Design and control of a 3-DOF hydraulic driven surgical instrument
  73. Research Article
  74. Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
  75. Research Article
  76. Frequency based assessment of surgical activities
  77. Research Article
  78. “Hands free for intervention”, a new approach for transoral endoscopic surgery
  79. Research Article
  80. Pseudo-haptic feedback in medical teleoperation
  81. Research Article
  82. Feasibility of interactive gesture control of a robotic microscope
  83. Research Article
  84. Towards structuring contextual information for workflow-driven surgical assistance functionalities
  85. Research Article
  86. Towards a framework for standardized semantic workflow modeling and management in the surgical domain
  87. Research Article
  88. Closed-loop approach for situation awareness of medical devices and operating room infrastructure
  89. Research Article
  90. Kinect based physiotherapy system for home use
  91. Research Article
  92. Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
  93. Research Article
  94. Integrating multimodal information for intraoperative assistance in neurosurgery
  95. Research Article
  96. Respiratory motion tracking using Microsoft’s Kinect v2 camera
  97. Research Article
  98. Using smart glasses for ultrasound diagnostics
  99. Research Article
  100. Measurement of needle susceptibility artifacts in magnetic resonance images
  101. Research Article
  102. Dimensionality reduction of medical image descriptors for multimodal image registration
  103. Research Article
  104. Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
  105. Research Article
  106. Evaluation of CT capability for the detection of thin bone structures
  107. Research Article
  108. Towards contactless optical coherence elastography with acoustic tissue excitation
  109. Research Article
  110. Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
  111. Research Article
  112. Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
  113. Research Article
  114. Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
  115. Research Article
  116. Video tracking of swimming rodents on a reflective water surface
  117. Research Article
  118. MR imaging of model drug distribution in simulated vitreous
  119. Research Article
  120. Studying the extracellular contribution to the double wave vector diffusion-weighted signal
  121. Research Article
  122. Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
  123. Research Article
  124. Introducing a frequency-tunable magnetic particle spectrometer
  125. Research Article
  126. Imaging of aortic valve dynamics in 4D OCT
  127. Research Article
  128. Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
  129. Research Article
  130. Simple concept for a wide-field lensless digital holographic microscope using a laser diode
  131. Research Article
  132. Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
  133. Research Article
  134. Respiratory surface motion measurement by Microsoft Kinect
  135. Research Article
  136. Improving image quality in EIT imaging by measurement of thorax excursion
  137. Research Article
  138. A clustering based dual model framework for EIT imaging: first experimental results
  139. Research Article
  140. Three-dimensional anisotropic regularization for limited angle tomography
  141. Research Article
  142. GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
  143. Research Article
  144. Experimental computer tomograph
  145. Research Article
  146. US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
  147. Research Article
  148. Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
  149. Research Article
  150. Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
  151. Research Article
  152. Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
  153. Research Article
  154. An IHE-conform telecooperation platform supporting the treatment of dementia patients
  155. Research Article
  156. Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
  157. Research Article
  158. Identification of surgical instruments using UHF-RFID technology
  159. Research Article
  160. A generic concept for the development of model-guided clinical decision support systems
  161. Research Article
  162. Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
  163. Research Article
  164. Creating 3D gelatin phantoms for experimental evaluation in biomedicine
  165. Research Article
  166. Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
  167. Research Article
  168. Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
  169. Research Article
  170. Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
  171. Research Article
  172. Wireless medical sensors – context, robustness and safety
  173. Research Article
  174. Sequences for real-time magnetic particle imaging
  175. Research Article
  176. Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
  177. Research Article
  178. A machine learning approach for planning valve-sparing aortic root reconstruction
  179. Research Article
  180. An in-ear pulse wave velocity measurement system using heart sounds as time reference
  181. Research Article
  182. Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
  183. Research Article
  184. Multisegmental fusion of the lumbar spine a curse or a blessing?
  185. Research Article
  186. Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
  187. Research Article
  188. A muscle model for hybrid muscle activation
  189. Research Article
  190. Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
  191. Research Article
  192. An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
  193. Research Article
  194. Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
  195. Research Article
  196. Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
  197. Research Article
  198. Mesh structure-independent modeling of patient-specific atrial fiber orientation
  199. Research Article
  200. Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
  201. Research Article
  202. Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
  203. Research Article
  204. A java based simulator with user interface to simulate ventilated patients
  205. Research Article
  206. Evaluation of an algorithm to choose between competing models of respiratory mechanics
  207. Research Article
  208. Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
  209. Research Article
  210. Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
  211. Research Article
  212. Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
  213. Research Article
  214. FPGA controlled artificial vascular system
  215. Research Article
  216. Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
  217. Research Article
  218. Test setup for characterizing the efficacy of embolic protection devices
  219. Research Article
  220. Impact of electrode geometry on force generation during functional electrical stimulation
  221. Research Article
  222. 3D-based visual physical activity assessment of children
  223. Research Article
  224. Realtime assessment of foot orientation by Accelerometers and Gyroscopes
  225. Research Article
  226. Image based reconstruction for cystoscopy
  227. Research Article
  228. Image guided surgery innovation with graduate students - a new lecture format
  229. Research Article
  230. Multichannel FES parameterization for controlling foot motion in paretic gait
  231. Research Article
  232. Smartphone supported upper limb prosthesis
  233. Research Article
  234. Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
  235. Research Article
  236. Evaluation of adhesion promoters for Parylene C on gold metallization
  237. Research Article
  238. The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
  239. Research Article
  240. Increasing the visibility of thin NITINOL vascular implants
  241. Research Article
  242. Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
  243. Research Article
  244. Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
  245. Research Article
  246. Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
  247. Research Article
  248. Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
  249. Research Article
  250. Electrospun vascular grafts with anti-kinking properties
  251. Research Article
  252. Integration of temperature sensors in polyimide-based thin-film electrode arrays
  253. Research Article
  254. Use cases and usability challenges for head-mounted displays in healthcare
  255. Research Article
  256. Device- and service profiles for integrated or systems based on open standards
  257. Research Article
  258. Risk management for medical devices in research projects
  259. Research Article
  260. Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
  261. Research Article
  262. Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
  263. Research Article
  264. SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
  265. Research Article
  266. Numerical simulations of airflow in the human pharynx of OSAHS patients
Downloaded on 25.12.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2015-0091/html
Scroll to top button