Article Open Access

Next-generation vision testing: the quick CSF

  • Michael Dorr EMAIL logo , Manuel Wille , Tiberiu Viulet , Edward Sanchez , Peter J Bex , Zhong-Lin Lu and Luis Lesmes
Published/Copyright: September 12, 2015

Abstract

The Contrast Sensitivity Function relates the spatial frequency and contrast of a spatial pattern to its visibility and thus provides a fundamental description of visual function. However, the current clinical standard of care typically restricts assessment to visual acuity, i.e. the smallest stimulus size that can be resolved at full contrast; alternatively, tests of contrast sensitivity are typically restricted to assessment of the lowest visible contrast for a fixed letter size. This restriction to one-dimensional subspaces of a two-dimensional space was necessary when stimuli were printed on paper charts and simple scoring rules were applied manually. More recently, however, computerized testing and electronic screens have enabled more flexible stimulus displays and more complex test algorithms. For example, the quick CSF method uses a Bayesian adaptive procedure and an information maximization criterion to select only informative stimuli; testing times to precisely estimate the whole contrast sensitivity function are reduced to 2-5 minutes. Here, we describe the implementation of the quick CSF method in a medical device. We make several usability enhancements to make it suitable for use in clinical settings. A first usability study shows excellent results, with a mean System Usability Scale score of 86.5.

1 Introduction

The current gold standard for the assessment of visual function goes back to 1862, when Herman Snellen developed a standard letter chart that has remained in use with only small modifications ever since. By showing progressively smaller letters at full contrast, the Snellen chart assesses visual acuity, i.e. the visual system’s resolving power. However, the visual system’s ability to detect a pattern depends not only on the size of the pattern, but also on its contrast; the Contrast Sensitivity Function (CSF) [4], which relates spatial frequency (size) to threshold contrast, therefore is a more comprehensive descriptor of visual function. Importantly, neurologic and ophthalmic pathologies may affect contrast sensitivity before they affect acuity [8, 13].

Despite its recognized clinical utility, routine assessment of contrast sensitivity in clinical trials and regular care is hampered by practical limitations. Compared to one-dimensional acuity, the two-dimensional CSF necessitates testing at many combinations of spatial frequency and contrast, for which paper charts lack resolution and flexibility. Simple scoring heuristics such as “stop at the last line with three out of five correct letter identifications” can be evaluated manually, but also lack computational efficiency to precisely capture the probabilistic nature of psychometric responses. By randomizing the sequence of tested optotypes, computerized vision testing alleviates some of the shortcomings of paper charts, but test scoring is typically implemented following the same simple rules as in non-computerized testing.

The quick CSF method [9] was originally developed for behavioural testing in psychophysical laboratories and uses a Bayesian adaptive procedure and an information maximization rule to test only informative combinations of spatial frequency and contrast, based on the full trial history. In this paper, we describe the implementation of the quick CSF method in a medical device that may be used to rapidly and precisely estimate visual function in clinical settings.

2 The quick CSF method

The quick CSF method exploits the observation that the human CSF can be accurately described using only four parameters [10, 12]: peak frequency and gain, bandwidth, and a low-frequency truncation parameter. After each trial of a test, the probability distribution p(Θ) over a set of possible CSFs described by four-parameter tuples Θ (in our implementation, about 12 million CSFs) can be computed, given the full trial history. Before each trial, the expected information gain is calculated for all possible stimuli (here, 19 spatial frequencies at 128 contrast levels each; 2432 stimuli overall) over a sampled subset of Θ, and only informative stimuli are chosen for presentation. By avoiding uninformative regions of the search space, the quick CSF method in [9] already estimated the whole CSF in 50-100 trials. For these results, the authors in [9] used sinewave gratings of two possible orientations as stimuli. Gratings are very frequency-selective and correspond well to neuronal receptive fields in primary visual cortex [6], but the small number of alternatives leads to a high guessing rate (50%) and thus lower statistical efficiency. Therefore, we implemented the use of ten Sloan letters, bandpass-filtered with a raised cosine window with peak frequency 4 cycles per letter, as proposed in [5]; as a consequence, the number of trials can be further reduced to 25 for rapid testing times, or 50 for very high precision.

2.1 Usability enhancements

We implemented two additional usability enhancements of the quick CSF. First, in classical forced-choice tasks, observers are asked to guess a response even if they do not perceive the stimulus. While even highly experienced psychophysical observers still can have strong biases when guessing [7], this is particularly difficult for inexperienced observers such as patients; we therefore added an “I don’t know” to the set of responses. Second, the lack of a (visible) stimulus on the screen can be confusing to observers; therefore, three letters are always presented simultaneously in a horizontal line, with the middle and left letters displayed at two and four times the contrast of the right letter, respectively. Because the contrast of the right letter is chosen by the quick CSF method and is usually near threshold contrast, it is very likely that at least one of the letters is easily recognizable by the observer, which helps observers to determine the location and size of the test letters.

3 Hardware

The hardware setup of our quick CSF implementation mainly comprises a small-form factor PC (Intel i5 CPU, 4 GB RAM) and a large-format screen for stimulus display. At 46” diagonal with a resolution of 1920 by 1080 pixels and at a viewing distance of 400 cm, the screen allows the display of stimuli in a spatial frequency range from 1.4 to 36.2 cycles per degree, which includes the whole set of frequencies mandated by the FDA (1.5 to 18 cpd). Also according to FDA standards, mean screen luminance is calibrated to 85 cd/m2. All device functionality including power control is accessed through a handheld tablet device; an NFC reader provides an authentication mechanism through smartcards. External interfacing for data export is provided by an Ethernet and a USB port.

Figure 1 Picture of a device prototype in demo mode, where particularly visible (medium spatial frequency, high contrast) stimuli are shown.
Figure 1

Picture of a device prototype in demo mode, where particularly visible (medium spatial frequency, high contrast) stimuli are shown.

4 Software

4.1 Main computer

The small-form factor PC runs a regular Linux operating system. However, the OS is transparent to the user because all user interaction is performed on the tablet remote control, and the vision test software permanently runs in fullscreen mode. The PC-side software is written in C++; the graphics display further uses OpenGL shaders and implements spatio-temporal dithering to increase bit depth of the screen.

For convenience, the software automatically calculates several features of the CSF: threshold sensitivities at five individual spatial frequencies mandated by the FDA [14]; CSF acuity, the intersection of the CSF with the x-axis (i.e. spatial frequency where contrast threshold is 100%); and a summary statistic, the area under the log CSF in the range from 1.5 to 18 cpd [1]. For an example screenshot of the results display, see Fig. 3.

Figure 2 Screenshot of remote control tablet during a test session. The trial history (correct answers, triangles; mistakes, crosses; ’no answer’ responses, slashes) and the current best estimate of the CSF are shown in the top panel, with spatial frequency on the x-axis and contrast sensitivity on the y-axis. The current trial is highlighted in blue.
Figure 2

Screenshot of remote control tablet during a test session. The trial history (correct answers, triangles; mistakes, crosses; ’no answer’ responses, slashes) and the current best estimate of the CSF are shown in the top panel, with spatial frequency on the x-axis and contrast sensitivity on the y-axis. The current trial is highlighted in blue.

Raw data, such as the trial history and the full distribution of posterior parameters fmax, ymax, β, δ, are stored in a database and can be exported via a web-based database interface or the USB port.

4.2 Tablet remote control

The tablet remote control is based on a customized Android image that always runs the remote control app in the foreground. During the test, the patient reads out the letters on the screen. The examiner, who is presented with the ground truth, codes these responses as correct or incorrect; an “I don’t know” response can also be coded; see Fig. 2 for a screenshot of the user interface. In order to reduce spatial uncertainty, each stimulus presentation is preceded by markers indicating the spatial position and scale of the upcoming stimulus; the examiner can repeat this marker presentation using the “Prompt” button. After response entry, the examiner can initiate the next trial; if necessary, previous responses can be undone as well.

Figure 3 Screenshot of remote control tablet, showing test results after a session. The solid line denotes median sensitivity estimate for the posterior distribution of possible CSFs, with the shaded region denoting the 66% confidence interval.
Figure 3

Screenshot of remote control tablet, showing test results after a session. The solid line denotes median sensitivity estimate for the posterior distribution of possible CSFs, with the shaded region denoting the 66% confidence interval.

5 Usability evaluation

Protoype systems are currently in use in several clinics in the United States and Germany. After extended use, we asked the technicians who operate the device on a regular basis to anonymously fill out system usability scale (SUS) questionnaires [3]. SUS comprises ten questions (five positive, five negative statements) that require a rating on a scale from 1 (“strongly disagree”) to 5 (“strongly agree”).

Fig. 4 shows the response distributions for each of the 10 SUS questions obtained from nine participants. For visualization purposes, we converted scores to a range of ’-’ (score 1 on odd-numbered questions, score 5 on even-numbered questions) to ’++’ (scores 5 and 1 for odd- and even-numbered, respectively), so that darker colours in Fig. 4 are better. Numerically, the mean SUS was 86.5 with a s.d. of 10. The strongest agreement was found with item 3, “I thought the system was easy to use” (mean 4.78, s.d. of .44).

6 Discussion

Moore’s law and the increase in available computational power have made it possible to apply complex algorithms even during brief inter-trial intervals in behavioural testing. Adaptive testing that rests on the exact computation of probability distributions in large multi-dimensional spaces goes beyond manually-scored heuristics and provides greater precision with reduced testing times. We here describe how an algorithm that was developed for psychophysics laboratories has been translated into an easy-to-use medical device for clinical settings. We implemented usability enhancements for the patient, and integrated all device interaction in a tablet-based remote control for the technician’s convenience. A usability study demonstrated that users rated the device highly, with a mean score that corresponds to an “excellent” rating in a large-scale study of SUS scores [2].

Figure 4 Results for System Usability Scale questionnaires. For visualization, (dis-)agreement with positive (negative) questions on a scale from 1 to 5 was converted to a scale from ’- -’ to ’++’.
Figure 4

Results for System Usability Scale questionnaires. For visualization, (dis-)agreement with positive (negative) questions on a scale from 1 to 5 was converted to a scale from ’- -’ to ’++’.

A first clinical study has already shown that quick CSF test results correlate better with patient-reported outcomes than established far and near visual acuity measures [11]. Further studies are currently ongoing to demonstrate the quick CSF’s sensitivity to track subtle changes in visual function due to disease progression or treatment effects.

Author’s Statement

  1. Conflict of interest: The authors declare personal financial interests in and/or employment with Adaptive Sensory Technology, a company commercializing the technology presented here. Material and Methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animals use.

References

[1] R Applegate, G Hilmantel, and H Howland. Area under the log contrast sensitivity function: A concise method of following changes in visual performance. OSA Technical Digest Series, 1:98–101, 1997.10.1364/VSIA.1997.SaB.4Search in Google Scholar

[2] Aaron Bangor, Philip Kortum, and James Miller. Determining what individual SUS scores mean: Adding an adjective rating scale. Journal of Usability Studies, 4(3):114–123, 2009.Search in Google Scholar

[3] John Brooke. SUS – a ’quick and dirty’ usability scale. Usability evaluation in industry, 189(194):4–7, 1996.Search in Google Scholar

[4] F. W. Campbell and D. G. Green. Optical and retinal factors affecting visual resolution. J Physiol, 181(3):576–593, Dec 1965.10.1113/jphysiol.1965.sp007784Search in Google Scholar

[5] Fang Hou, Luis Lesmes, Peter Bex, Michael Dorr, and ZhongLin Lu. Using 10AFC to further improve the efficiency of the quick CSF method. Journal of Vision, 15(2):1–18, 2015.10.1167/15.9.2Search in Google Scholar

[6] D H Hubel and T N Wiesel. Functional architecture of macaque monkey visual cortex. Proc R Soc Lond, 198:1–59, 1977.10.1098/rspb.1977.0085Search in Google Scholar

[7] Frank Jäkel and Felix A Wichmann. Spatial four-alternative forced-choice method is the preferred psychophysical method for naive observers. Journal of Vision, 6:1307–22, 2006.10.1167/6.11.13Search in Google Scholar

[8] L F Jindra and V Zemon. Contrast sensitivity testing: a more complete assessment of vision. J Cataract Refract Surg, 15(2):141–148, Mar 1989.10.1016/S0886-3350(89)80002-1Search in Google Scholar

[9] Luis Andres Lesmes, Zhong-Lin Lu, Jongsoo Baek, and Thomas D. Albright. Bayesian adaptive estimation of the contrast sensitivity function: The quick CSF method. Journal of Vision, 10(3), 2010.10.1167/10.3.17Search in Google Scholar PubMed PubMed Central

[10] A. M. Rohaly and C. Owsley. Modeling the contrast-sensitivity functions of older adults. J Opt Soc Am A, 10(7):1591–1599, Jul 1993.10.1364/JOSAA.10.001591Search in Google Scholar PubMed

[11] J P Stellmann, K L Young, J Pöttgen, M Dorr, and C Heesen. Introducing a new method to assess vision: Computer-adaptive contrast-sensitivity testing predicts visual functioning better than charts in multiple sclerosis patients. Multiple Sclerosis Journal: Experimental, Translational and Clinical, 1(2055217315596184):1–8, 2015.10.1177/2055217315596184Search in Google Scholar PubMed PubMed Central

[12] Andrew B Watson and Albert J Ahumada. A standard model for foveal detection of spatial contrast. Journal of Vision, 5(9):717–740, 2005.10.1167/5.9.6Search in Google Scholar PubMed

[13] R L Woods and J M Wood. The role of contrast sensitivity charts and contrast letter charts in clinical practice. Clinical & Experimental Optometry, 78(2):43–57, 1995.10.1111/j.1444-0938.1995.tb00787.xSearch in Google Scholar

[14] American National Standards Institute Committee Z80. American National Standard for Ophthalmics: Multifocal Intraocular Lenses. Optical Laboratories Association, 11096 Lee Highway, Fairfax, VA, 2007.Search in Google Scholar

Published Online: 2015-9-12
Published in Print: 2015-9-1

© 2015 by Walter de Gruyter GmbH, Berlin/Boston

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Articles in the same Issue

  1. Research Article
  2. Development and characterization of superparamagnetic coatings
  3. Research Article
  4. The development of an experimental setup to measure acousto-electric interaction signal
  5. Research Article
  6. Stability analysis of ferrofluids
  7. Research Article
  8. Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
  9. Research Article
  10. Energy harvesting for active implants: powering a ruminal pH-monitoring system
  11. Research Article
  12. New type of fluxgate magnetometer for the heart’s magnetic fields detection
  13. Research Article
  14. Field mapping of ballistic pressure pulse sources
  15. Research Article
  16. Development of a new homecare sleep monitor using body sounds and motion tracking
  17. Research Article
  18. Noise properties of textile, capacitive EEG electrodes
  19. Research Article
  20. Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
  21. Research Article
  22. Spike sorting: the overlapping spikes challenge
  23. Research Article
  24. Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
  25. Research Article
  26. Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
  27. Research Article
  28. Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
  29. Research Article
  30. Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
  31. Research Article
  32. Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
  33. Research Article
  34. Compressed sensing of multi-lead ECG signals by compressive multiplexing
  35. Research Article
  36. Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
  37. Research Article
  38. Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
  39. Research Article
  40. Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
  41. Research Article
  42. Effects of sampling rate on automated fatigue recognition in surface EMG signals
  43. Research Article
  44. Closed-loop transcranial alternating current stimulation of slow oscillations
  45. Research Article
  46. Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
  47. Research Article
  48. The role of expert evaluation for microsleep detection
  49. Research Article
  50. The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
  51. Research Article
  52. Metal artifact reduction by projection replacements and non-local prior image integration
  53. Research Article
  54. A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
  55. Research Article
  56. Processing of membranes for oxygenation using the Bellhouse-effect
  57. Research Article
  58. Inkjet printing of viable human dental follicle stem cells
  59. Research Article
  60. The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
  61. Research Article
  62. New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
  63. Research Article
  64. QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
  65. Research Article
  66. Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
  67. Research Article
  68. Next-generation vision testing: the quick CSF
  69. Research Article
  70. Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
  71. Research Article
  72. Design and control of a 3-DOF hydraulic driven surgical instrument
  73. Research Article
  74. Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
  75. Research Article
  76. Frequency based assessment of surgical activities
  77. Research Article
  78. “Hands free for intervention”, a new approach for transoral endoscopic surgery
  79. Research Article
  80. Pseudo-haptic feedback in medical teleoperation
  81. Research Article
  82. Feasibility of interactive gesture control of a robotic microscope
  83. Research Article
  84. Towards structuring contextual information for workflow-driven surgical assistance functionalities
  85. Research Article
  86. Towards a framework for standardized semantic workflow modeling and management in the surgical domain
  87. Research Article
  88. Closed-loop approach for situation awareness of medical devices and operating room infrastructure
  89. Research Article
  90. Kinect based physiotherapy system for home use
  91. Research Article
  92. Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
  93. Research Article
  94. Integrating multimodal information for intraoperative assistance in neurosurgery
  95. Research Article
  96. Respiratory motion tracking using Microsoft’s Kinect v2 camera
  97. Research Article
  98. Using smart glasses for ultrasound diagnostics
  99. Research Article
  100. Measurement of needle susceptibility artifacts in magnetic resonance images
  101. Research Article
  102. Dimensionality reduction of medical image descriptors for multimodal image registration
  103. Research Article
  104. Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
  105. Research Article
  106. Evaluation of CT capability for the detection of thin bone structures
  107. Research Article
  108. Towards contactless optical coherence elastography with acoustic tissue excitation
  109. Research Article
  110. Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
  111. Research Article
  112. Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
  113. Research Article
  114. Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
  115. Research Article
  116. Video tracking of swimming rodents on a reflective water surface
  117. Research Article
  118. MR imaging of model drug distribution in simulated vitreous
  119. Research Article
  120. Studying the extracellular contribution to the double wave vector diffusion-weighted signal
  121. Research Article
  122. Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
  123. Research Article
  124. Introducing a frequency-tunable magnetic particle spectrometer
  125. Research Article
  126. Imaging of aortic valve dynamics in 4D OCT
  127. Research Article
  128. Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
  129. Research Article
  130. Simple concept for a wide-field lensless digital holographic microscope using a laser diode
  131. Research Article
  132. Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
  133. Research Article
  134. Respiratory surface motion measurement by Microsoft Kinect
  135. Research Article
  136. Improving image quality in EIT imaging by measurement of thorax excursion
  137. Research Article
  138. A clustering based dual model framework for EIT imaging: first experimental results
  139. Research Article
  140. Three-dimensional anisotropic regularization for limited angle tomography
  141. Research Article
  142. GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
  143. Research Article
  144. Experimental computer tomograph
  145. Research Article
  146. US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
  147. Research Article
  148. Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
  149. Research Article
  150. Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
  151. Research Article
  152. Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
  153. Research Article
  154. An IHE-conform telecooperation platform supporting the treatment of dementia patients
  155. Research Article
  156. Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
  157. Research Article
  158. Identification of surgical instruments using UHF-RFID technology
  159. Research Article
  160. A generic concept for the development of model-guided clinical decision support systems
  161. Research Article
  162. Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
  163. Research Article
  164. Creating 3D gelatin phantoms for experimental evaluation in biomedicine
  165. Research Article
  166. Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
  167. Research Article
  168. Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
  169. Research Article
  170. Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
  171. Research Article
  172. Wireless medical sensors – context, robustness and safety
  173. Research Article
  174. Sequences for real-time magnetic particle imaging
  175. Research Article
  176. Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
  177. Research Article
  178. A machine learning approach for planning valve-sparing aortic root reconstruction
  179. Research Article
  180. An in-ear pulse wave velocity measurement system using heart sounds as time reference
  181. Research Article
  182. Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
  183. Research Article
  184. Multisegmental fusion of the lumbar spine a curse or a blessing?
  185. Research Article
  186. Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
  187. Research Article
  188. A muscle model for hybrid muscle activation
  189. Research Article
  190. Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
  191. Research Article
  192. An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
  193. Research Article
  194. Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
  195. Research Article
  196. Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
  197. Research Article
  198. Mesh structure-independent modeling of patient-specific atrial fiber orientation
  199. Research Article
  200. Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
  201. Research Article
  202. Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
  203. Research Article
  204. A java based simulator with user interface to simulate ventilated patients
  205. Research Article
  206. Evaluation of an algorithm to choose between competing models of respiratory mechanics
  207. Research Article
  208. Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
  209. Research Article
  210. Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
  211. Research Article
  212. Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
  213. Research Article
  214. FPGA controlled artificial vascular system
  215. Research Article
  216. Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
  217. Research Article
  218. Test setup for characterizing the efficacy of embolic protection devices
  219. Research Article
  220. Impact of electrode geometry on force generation during functional electrical stimulation
  221. Research Article
  222. 3D-based visual physical activity assessment of children
  223. Research Article
  224. Realtime assessment of foot orientation by Accelerometers and Gyroscopes
  225. Research Article
  226. Image based reconstruction for cystoscopy
  227. Research Article
  228. Image guided surgery innovation with graduate students - a new lecture format
  229. Research Article
  230. Multichannel FES parameterization for controlling foot motion in paretic gait
  231. Research Article
  232. Smartphone supported upper limb prosthesis
  233. Research Article
  234. Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
  235. Research Article
  236. Evaluation of adhesion promoters for Parylene C on gold metallization
  237. Research Article
  238. The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
  239. Research Article
  240. Increasing the visibility of thin NITINOL vascular implants
  241. Research Article
  242. Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
  243. Research Article
  244. Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
  245. Research Article
  246. Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
  247. Research Article
  248. Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
  249. Research Article
  250. Electrospun vascular grafts with anti-kinking properties
  251. Research Article
  252. Integration of temperature sensors in polyimide-based thin-film electrode arrays
  253. Research Article
  254. Use cases and usability challenges for head-mounted displays in healthcare
  255. Research Article
  256. Device- and service profiles for integrated or systems based on open standards
  257. Research Article
  258. Risk management for medical devices in research projects
  259. Research Article
  260. Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
  261. Research Article
  262. Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
  263. Research Article
  264. SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
  265. Research Article
  266. Numerical simulations of airflow in the human pharynx of OSAHS patients
Downloaded on 25.12.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2015-0034/html
Scroll to top button