Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence

  • Thomas Pollack EMAIL logo , Marc Dangers and Hermann Seitz
Published/Copyright: September 12, 2015

Abstract

The drying and resolving processes of surface-bound proteins were analysed with a stereoscopic microscope combined with laser-induced fluorescence (LIF) with the aim to assist the advancement of a semiautomated point-of-care allergy diagnostic device. The results shown in this paper present the use of the LIF-method for concentration calibration and the analysis of drying and resolving dyed proteins. The method is established successfully and delivers precise results.

1 Introduction

The allergy diagnostic device is the point-of-care test FCP20 (DST GmbH, Germany). It is a membrane-based ELISA test for the semi-quantitative detection of allergenspecific IgE [1]. It uses allergen extracts, reagents, and blood, serum or plasma of a patient; after 30 minutes the results of all 20 allergens can be read semi-quantitatively by the naked eye. The reagents of the current product are stored as liquids in reagent tubes and have to be cooled.

To simplify the transport and operation the test procedure had to be semi-automated. Therefore, a separate case with a flow passage and dried-up proteins within and a pump unit with a reservoir of rinsing solution are added. The advantages of this are a better transportability due to the fact that no cooling and no bottles are needed, as well as a less time consuming procedure for the user. The additional pump unit controls the whole process and ensures that the results of the analysis are independent from the user. The whole system will consist of three components; two disposable cartridges (detector and protein resolve system) and one stationary (pump unit and reservoir). Figure 1 shows the described devices in a schematic diagram.

Figure 1 Schematic overview of the studied test setup.
Figure 1

Schematic overview of the studied test setup.

Subject of the present paper is the establishment of the LIF-method, which is used to describe the release kinetics of surface-bound proteins [2]. In order to describe the resolve process of dried-up proteins in fluid (e.g. NaCl 0.9%) different analyses were made. First of all, the concept had to be proven. In other words, the possibility to rediscover the resolved quantity of protein in relation to the brought in had to be demonstrated. Secondly the dye-protein combination which works best had to be found and the lowest concentration, at which the microscope-LIF-system works, had to be determined. Thirdly the ideal volume and concentration of dried-up proteins which is needed to transport the dissolved proteins to the scope had to be determined experimentally. Finally, it was important to observe whether the resolved protein in the volume flow was homogenous, granular, in strings or something different.

The analyses were made with an optical study method named laser-induced fluorescence (LIF) [3]. The system detects if dyed labelled proteins emit light when illuminated with light of a specific wavelength. The LIF-system is combined with a stereo-microscope and observes and measures the processes in the microfluidic protein storage and resolve device.

2 Methods

The investigations are divided into three stages. First of all, a calibration for the concentration is performed in order to assign the measured light intensities to a concentration. After this, the drying and resolving processes of a drop of dye-protein solution are studied. The investigations are then completed by measuring the volume flow of the resolved protein streaming towards the detector.

The stereo-microscope Discovery V20 is manufactured by Carl Zeiss (Göttingen, Germany), the LIF-system and the imaging application software were provided by LaVision (Göttingen, Germany). The aperture is wide open. The light source is a mercury lamp operated at full intensity. Zoom is 46x with a zoom lens of 10x. The filter set has the number 43 Cy3/Rhod/RFP (D) from Carl Zeiss (EX BP 545/25, BS FT 565, EM BP 606/70). The camera system “Imager sCMOS” has a sCMOS-chip and a resolution of 2560x2160 pixels. Images are made in single-frame mode with an exposure time of 100 ms. The resolving fluid is NaCl 0.9%.

Every displayed figure is characterised by its colour resolution or, more specifically, its range of counts, where the lowest measured concentrations are blue or nearly black and the highest concentrations are displayed in red. Figures which belong together have the same colour resolution to ensure comparability (e.g. 0 – 25.000 counts).

2.1 Calibration process

A great number of dye-protein combinations are available on the market. On the one hand, a qualified combination has to meet the requirements of the POCT allergy diagnostic and be a standard of comparison for the proposed analyses. On the other hand, the LIF-system has its own settings and requirements (e.g. filters, beam splitter, light source and emission spectrum).

AlexaFluor 532 and 546 from Life Technologies (Darmstadt, Germany) fulfil both the mentioned requirements. Both combinations are compared by means of the calibration process and their intensity of luminescence. This intensity depends linearly on the concentration of the dye-solvent-mix. A greater intensity leads to a more sensitive detection of lower concentrations.

Within the calibration process, different known concentrations are observed and processed in order to obtain a valid calibration for later testing. The measured light intensity in later analyses will then be processed and calculated in the same way and compared with the calibration. This way, the measured light emission results in a value for the concentration

2.2 Drying and resolve process

At this point the release kinetics is observed in a static way. In other words, a known amount of concentration of a protein-dye solution is dried-up in a flow passage. Then the dry-process is observed during the following hours. After this, the same volume (10 µl) of a NaCl-solution is added. The resolving process is observed and is displayed in a time lapse of about half an hour.

The investigations are executed without any volume flow. Only capillary forces and the resolving process lead to any observed mixing.

2.2 Flow process

The resolve process is examined under known conditions, i.e. for a given dye-protein combination and known values of the concentration, volume of the dried-up solution and the volume flow. The fluid (NaCl 0.9%) resolves the thin film of dye-protein and takes an unknown amount of concentration with it.

The following figure presents an example of the analysed flow passages. The channel has a length of 6 cm, a width of 2 mm and a depth of 0.3 mm. The cover and the channel are both made of polycarbonate. The focus of the microscope is in the centre of the figure during the observation of the flow.

Figure 2 Example of a flow passage (intensity: 350–550 counts).
Figure 2

Example of a flow passage (intensity: 350–550 counts).

3 Results

3.1 Calibration process

At this point, the dye-protein combinations (AF532 and AF546) are analysed concerning their suitability for further testing. The results are shown in the following figures. To ensure comparability, the figures have the same colour resolution (0 – 25.000 counts). As previously mentioned, lower concentrations are blue or dark, whereas the highest concentration is shown in red.

Figure 3 AF532 calibration seq. (0.01, 0.10, 0.25, 1.00 mg/ml.)
Figure 3

AF532 calibration seq. (0.01, 0.10, 0.25, 1.00 mg/ml.)

Figure 4 AF546 calibration seq. (0.01, 0.10, 0.25, 1.00 mg/ml).
Figure 4

AF546 calibration seq. (0.01, 0.10, 0.25, 1.00 mg/ml).

Comparing Figures 3 and 4 reveals that AF546 has a higher light intensity. The absorption and emission fit better than for AF532. The following dry and resolve observation are made with AF532. The flow process is observed with AF546.

Figure 5 AF546 calibration seq. (0.01, 0.10, 0.25, 1.00 mg/ml).
Figure 5

AF546 calibration seq. (0.01, 0.10, 0.25, 1.00 mg/ml).

Figure 5 shows the relationship between the sample concentration and the measured light intensity. A mixture of 0.01 mg dye-protein per millilitre stands for a concentration of 1 %, a mixture of 1.00 mg per ml for 100%.

3.2 Drying and resolve process

Figure 6 illustrates the drying process of a 0.5 mg/ml dye-protein-solvent solution. The time lapse shows a decrease in light intensity. The initial concentration of about 50% decreases to at least 2% due to the drying process. The volume of the homogenous solution from the left picture evaporates. Parts of the fluorescent stick to the edge of the flow passage or cake at the bottom (right picture). The volume of 10 µl has disappeared and the protein-dye solution has become a thin film.

Figure 6 AF532 drying process (solvent, after 2h, 3.5h and 5h)
Figure 6

AF532 drying process (solvent, after 2h, 3.5h and 5h)

The colour resolution of the Figures 6 and 7 was equalized (400 – 6400 counts) to ensure comparability.

Figure 7 AF532 resolving process (after 1, 5, 10 and 30 min).
Figure 7

AF532 resolving process (after 1, 5, 10 and 30 min).

Figure 7 shows the resolving process. Little grainy fragments appear immediately and flow to the centre due to capillary forces. A homogenous solution appears within 30 minutes and the concentration increases to at least 6%.

3.3 Flow process

The green, red and white lit area in Figure 8 is a dried-up drop of dye-protein solution. The little black and dark-blue point in its centre is a marker, made with a pencil and pointed onto the transparent cover of the channel.

Figure 8 Example of a flow passage (intensity: 350–1250 counts)
Figure 8

Example of a flow passage (intensity: 350–1250 counts)

The time lapses in Figures 9 and 10 show a volume flow streaming from left to right.

Figure 9 AF546 resolving process dynamic æ1 (0, 3, 6, 12 sec).
Figure 9

AF546 resolving process dynamic æ1 (0, 3, 6, 12 sec).

The first image in Figure 9 displays the frontline of the flow field and a volume of circa 1µl. The concentration is nearly 10% and seems uniform except for some strings in the centre of the flow. The second image, which is recorded 3 seconds later, shows a concentration which is at least 4 times higher. In the third image, recorded 6 seconds after the flow started and protein began to resolve, the concentration is increased by at least 15% and seems to distribute itself homogeneously. The last image, taken 12 seconds after the delivery started, shows a decrease in concentration and a more and more grainy solution.

The second time lapse in Figure 10 shows another flow passage. The conditions are the same as above and the pictures are recorded at the same time steps.

Unlike the process in Figure 9, this resolving process is non-uniform and less homogeneous. In the first image the concentration is about 7% and shows strings and fragments. The concentration in the middle of the flow in the second image is approximately 18% and contains grains as well. The third image shows even more particles and grainy fragments. Finally, after 12 seconds of streaming, the concentration decreases to at least 7% with several highlighted fragments.

The resolving occurs within the first second of wetting in all analysed flow passages. Within a few seconds, the concentration rises to a relatively high level and decreases afterwards. However, the flow has no constant level of concentration and appears in strings or grainy fragments in certain cases. Although, the initial conditions were identical, the maximum level of concentration in one of the considered cases is only the half of the other one.

Figure 10 AF546 resolving process dynamic æ2 (0, 3, 6, 12 sec).
Figure 10

AF546 resolving process dynamic æ2 (0, 3, 6, 12 sec).

4 Conclusion

The method of measuring the drying and resolving process with the use of a stereo-microscope combined with the LIF-system works accurately and provides satisfying results, as expected. Whether in a static or dynamic way, the concentration of the observed processes can be measured precisely and visualized in high resolution. Low concentration levels and how the dye-protein solution is resolved in the flow (e.g. homogenously, granular, in strings or something else) can be measured and observed.

Further studies have to be performed in order to get a better understanding of the resolving process and the mixing of the dye-protein in the flow [4]. The high deviation of the peak-concentration and the time curve of the concentration will be another aspect of interest. Its cause is currently suspected in the dispensing system. When the solvent dye-protein solution is brought onto the flow passage, the fluid volume is forced by capillary forces and the dried-up drop forms a more or less random formed thin-film (see Figure 8). These thin-films have different thicknesses, areas and form.

Funding: This project is funded by the Bundesministerium für Bildung und Forschung (FKZ: 03WKCC 10; Innovative Regional Growth Cores “Centifluidic Technologies”) and is part of the BMBF-initiative “Entrepreneurial Regions” “The BMBF Innovation Initiative for the New German Länder“.

Author's Statement

  1. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animals use.

References

[1] Drescher P, et al. The effects of various flow velocities on the sensitivity of an enzyme-linked immunosorbent assay in a fluidic allergy diagnostic device. Point of Care: The Journal of Near-Patient Testing & Technology 2014; 13: 35-40.10.1097/POC.0000000000000016Search in Google Scholar

[2] Atkins PW, de Paula J. Atkins’ Physical chemistry. 10th ed. Oxford. Oxford University Press 2014.Search in Google Scholar

[3] Zare RN. My life with LIF: a personal account of developing laser-induced fluorescence. Annual review of analytical chemistry 2012; 5: 1-14.10.1146/annurev-anchem-062011-143148Search in Google Scholar PubMed

[4] Dash S, et al. Kinetic modeling on drug release from controlled drug delivery systems.Acta poloniae pharmaceutica 2010; 67: 217-223Search in Google Scholar

Published Online: 2015-9-12
Published in Print: 2015-9-1

© 2015 by Walter de Gruyter GmbH, Berlin/Boston

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Articles in the same Issue

  1. Research Article
  2. Development and characterization of superparamagnetic coatings
  3. Research Article
  4. The development of an experimental setup to measure acousto-electric interaction signal
  5. Research Article
  6. Stability analysis of ferrofluids
  7. Research Article
  8. Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
  9. Research Article
  10. Energy harvesting for active implants: powering a ruminal pH-monitoring system
  11. Research Article
  12. New type of fluxgate magnetometer for the heart’s magnetic fields detection
  13. Research Article
  14. Field mapping of ballistic pressure pulse sources
  15. Research Article
  16. Development of a new homecare sleep monitor using body sounds and motion tracking
  17. Research Article
  18. Noise properties of textile, capacitive EEG electrodes
  19. Research Article
  20. Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
  21. Research Article
  22. Spike sorting: the overlapping spikes challenge
  23. Research Article
  24. Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
  25. Research Article
  26. Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
  27. Research Article
  28. Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
  29. Research Article
  30. Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
  31. Research Article
  32. Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
  33. Research Article
  34. Compressed sensing of multi-lead ECG signals by compressive multiplexing
  35. Research Article
  36. Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
  37. Research Article
  38. Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
  39. Research Article
  40. Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
  41. Research Article
  42. Effects of sampling rate on automated fatigue recognition in surface EMG signals
  43. Research Article
  44. Closed-loop transcranial alternating current stimulation of slow oscillations
  45. Research Article
  46. Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
  47. Research Article
  48. The role of expert evaluation for microsleep detection
  49. Research Article
  50. The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
  51. Research Article
  52. Metal artifact reduction by projection replacements and non-local prior image integration
  53. Research Article
  54. A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
  55. Research Article
  56. Processing of membranes for oxygenation using the Bellhouse-effect
  57. Research Article
  58. Inkjet printing of viable human dental follicle stem cells
  59. Research Article
  60. The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
  61. Research Article
  62. New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
  63. Research Article
  64. QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
  65. Research Article
  66. Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
  67. Research Article
  68. Next-generation vision testing: the quick CSF
  69. Research Article
  70. Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
  71. Research Article
  72. Design and control of a 3-DOF hydraulic driven surgical instrument
  73. Research Article
  74. Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
  75. Research Article
  76. Frequency based assessment of surgical activities
  77. Research Article
  78. “Hands free for intervention”, a new approach for transoral endoscopic surgery
  79. Research Article
  80. Pseudo-haptic feedback in medical teleoperation
  81. Research Article
  82. Feasibility of interactive gesture control of a robotic microscope
  83. Research Article
  84. Towards structuring contextual information for workflow-driven surgical assistance functionalities
  85. Research Article
  86. Towards a framework for standardized semantic workflow modeling and management in the surgical domain
  87. Research Article
  88. Closed-loop approach for situation awareness of medical devices and operating room infrastructure
  89. Research Article
  90. Kinect based physiotherapy system for home use
  91. Research Article
  92. Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
  93. Research Article
  94. Integrating multimodal information for intraoperative assistance in neurosurgery
  95. Research Article
  96. Respiratory motion tracking using Microsoft’s Kinect v2 camera
  97. Research Article
  98. Using smart glasses for ultrasound diagnostics
  99. Research Article
  100. Measurement of needle susceptibility artifacts in magnetic resonance images
  101. Research Article
  102. Dimensionality reduction of medical image descriptors for multimodal image registration
  103. Research Article
  104. Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
  105. Research Article
  106. Evaluation of CT capability for the detection of thin bone structures
  107. Research Article
  108. Towards contactless optical coherence elastography with acoustic tissue excitation
  109. Research Article
  110. Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
  111. Research Article
  112. Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
  113. Research Article
  114. Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
  115. Research Article
  116. Video tracking of swimming rodents on a reflective water surface
  117. Research Article
  118. MR imaging of model drug distribution in simulated vitreous
  119. Research Article
  120. Studying the extracellular contribution to the double wave vector diffusion-weighted signal
  121. Research Article
  122. Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
  123. Research Article
  124. Introducing a frequency-tunable magnetic particle spectrometer
  125. Research Article
  126. Imaging of aortic valve dynamics in 4D OCT
  127. Research Article
  128. Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
  129. Research Article
  130. Simple concept for a wide-field lensless digital holographic microscope using a laser diode
  131. Research Article
  132. Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
  133. Research Article
  134. Respiratory surface motion measurement by Microsoft Kinect
  135. Research Article
  136. Improving image quality in EIT imaging by measurement of thorax excursion
  137. Research Article
  138. A clustering based dual model framework for EIT imaging: first experimental results
  139. Research Article
  140. Three-dimensional anisotropic regularization for limited angle tomography
  141. Research Article
  142. GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
  143. Research Article
  144. Experimental computer tomograph
  145. Research Article
  146. US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
  147. Research Article
  148. Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
  149. Research Article
  150. Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
  151. Research Article
  152. Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
  153. Research Article
  154. An IHE-conform telecooperation platform supporting the treatment of dementia patients
  155. Research Article
  156. Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
  157. Research Article
  158. Identification of surgical instruments using UHF-RFID technology
  159. Research Article
  160. A generic concept for the development of model-guided clinical decision support systems
  161. Research Article
  162. Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
  163. Research Article
  164. Creating 3D gelatin phantoms for experimental evaluation in biomedicine
  165. Research Article
  166. Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
  167. Research Article
  168. Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
  169. Research Article
  170. Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
  171. Research Article
  172. Wireless medical sensors – context, robustness and safety
  173. Research Article
  174. Sequences for real-time magnetic particle imaging
  175. Research Article
  176. Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
  177. Research Article
  178. A machine learning approach for planning valve-sparing aortic root reconstruction
  179. Research Article
  180. An in-ear pulse wave velocity measurement system using heart sounds as time reference
  181. Research Article
  182. Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
  183. Research Article
  184. Multisegmental fusion of the lumbar spine a curse or a blessing?
  185. Research Article
  186. Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
  187. Research Article
  188. A muscle model for hybrid muscle activation
  189. Research Article
  190. Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
  191. Research Article
  192. An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
  193. Research Article
  194. Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
  195. Research Article
  196. Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
  197. Research Article
  198. Mesh structure-independent modeling of patient-specific atrial fiber orientation
  199. Research Article
  200. Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
  201. Research Article
  202. Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
  203. Research Article
  204. A java based simulator with user interface to simulate ventilated patients
  205. Research Article
  206. Evaluation of an algorithm to choose between competing models of respiratory mechanics
  207. Research Article
  208. Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
  209. Research Article
  210. Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
  211. Research Article
  212. Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
  213. Research Article
  214. FPGA controlled artificial vascular system
  215. Research Article
  216. Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
  217. Research Article
  218. Test setup for characterizing the efficacy of embolic protection devices
  219. Research Article
  220. Impact of electrode geometry on force generation during functional electrical stimulation
  221. Research Article
  222. 3D-based visual physical activity assessment of children
  223. Research Article
  224. Realtime assessment of foot orientation by Accelerometers and Gyroscopes
  225. Research Article
  226. Image based reconstruction for cystoscopy
  227. Research Article
  228. Image guided surgery innovation with graduate students - a new lecture format
  229. Research Article
  230. Multichannel FES parameterization for controlling foot motion in paretic gait
  231. Research Article
  232. Smartphone supported upper limb prosthesis
  233. Research Article
  234. Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
  235. Research Article
  236. Evaluation of adhesion promoters for Parylene C on gold metallization
  237. Research Article
  238. The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
  239. Research Article
  240. Increasing the visibility of thin NITINOL vascular implants
  241. Research Article
  242. Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
  243. Research Article
  244. Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
  245. Research Article
  246. Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
  247. Research Article
  248. Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
  249. Research Article
  250. Electrospun vascular grafts with anti-kinking properties
  251. Research Article
  252. Integration of temperature sensors in polyimide-based thin-film electrode arrays
  253. Research Article
  254. Use cases and usability challenges for head-mounted displays in healthcare
  255. Research Article
  256. Device- and service profiles for integrated or systems based on open standards
  257. Research Article
  258. Risk management for medical devices in research projects
  259. Research Article
  260. Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
  261. Research Article
  262. Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
  263. Research Article
  264. SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
  265. Research Article
  266. Numerical simulations of airflow in the human pharynx of OSAHS patients
Downloaded on 8.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2015-0084/html
Scroll to top button