Abstract
Endovascular stents are exposed to cyclic loads resulting from daily activity and pulsatile arterial blood pressure. DIN EN ISO 25539-2 and FDA guideline 1545 recommend durability testing, exposing stents to physiological cyclic loads for a 10 year equivalent. For accelerated testing, the simulated deformation has to be comparable to physiological in-vivo deformation. A new test setup is presented to determine diameter deformation of vascular stents during fatigue testing with radial loading.
Methods
The new setup allows the investigation of stents (n = 1–10) up to a length of 200 mm using a CCD line camera independent from special configurations. For demonstration, the radial deformation of two peripheral stents (stent 1: 8.0×40 mm, stent 2: 4.5×40 mm) and coronary stents (stent 3: 2.5×22 mm, stent 4: 4.0×40 mm) is determined as a function of the longitudinal measuring position. The stents are implanted in polyurethane tubes and exposed to physiologically relevant pressure at test frequencies 100 Hz.
Results
In addition to the verification of test frequencies for fatigue testing the setup can also be used for the investigation of radial deformation performance. The results show that radial deformation may vary along the stent length. Larger radial deformation was detected at the middle of the stent. For stent 1 a maximum deformation of 0.21 ± 0.07 mm (± 2.65 %) was measured at 50 ± 40 mmHg, 90 Hz. It was also measured that the radial deformation is dependent on stent design, geometric dimension and external loading.
Conclusion
The new setup allows for test frequency verification for accelerated fatigue testing with radial loading. It is also suitable for more detailed investigation of the radial deformation performance of stents along their longitudinal axis. This is necessary for a better understanding of potential mechanical failure especially in the case of long or overlapping stents.
1 Introduction
Nowadays, stent implantation has become the primary treatment of atherosclerosis in distinction from conventional balloon angioplasty and bypass surgery, where 30 – 40 % of the patients have developed restenosis within 6 months postoperatively [1, 2]. Stents are tubular structures that are implanted into narrowed or completely closed vessels with the goal to restore the blood flow, thus, the nutrient supply of the surrounding tissue [3]. The first coronary stent was implanted in 1986 by Sigwart and Puel. With the introduction of the bare metal stent and all the follow up innovations, it was possible to decrease the restenosis rate to 20 – 30 % [4, 5]. It is known that after implantation endovascular stents are exposed to substantial mechanical loads. Implanted stents have to withstand cyclic motions such as bending, torsion, tension and compression resulting from daily activity and pulsatile arterial blood pressure. Several studies have shown that over the time these loads can lead to mechanical failure, for instance strut fractures. Stent fractures cause endothelial injuries and perforations that lead to inflammatory reactions and endothelial hyperplasia finally leading to restenosis, bleedings and thrombosis. The studies confirm the importance of fatigue performance and design/material validation of endovascular stents [6, 7]. For the commercial approval, endovascular stents have to pass several tests to demonstrate functionality and efficacy of the geometric structure, biocompatibility and among these parameters fatigue strength. In DIN EN ISO 25539-2 and FDA guideline No 1545 it is recommended to perform durability testing where the stents are exposed to simulated physiological loads that are similar to in-vivo conditions for a cyclic equivalent of 10 years (380 million cycles) at physiological (1.2 Hz) or accelerated frequency. After durability testing, the test samples have to be examined for fractures [8, 9]. It is a basic requirement for accelerated durability testing to show that the deformations caused by the simulated loads are comparable with in-vivo occurring deformations [9]. In this paper a test setup is described to determine the diameter deformation during fatigue testing at an externally applied cyclic radial load. The testing device based on a CCD line camera measurement was developed and constructed at the Institute of Biomedical Engineering (IBMT) of Rostock University.
2 Methods
The new setup allows the investigation of stents (n = 1– 10) with a diameter of 2–10 mm and up to a length of 200 mm using a CCD line camera. Therefore, the stents are expanded into mock vessels using polyurethane-tubes with known wall thickness and compliance. To expose the stents to an external dynamical pressure, the stented mock vessels are placed in the transparent pressure-proof testing chamber and surrounded by 37 ± 2°C deionized and degassed tempered water. Their inner lumen is filled with testing solution (PBS, deionized water). To allow the investigation of stents independently from its configuration two different sample holder systems were designed (Figure 1). The two systems enable the investigation of straight single or overlapping stents, as well as the investigation of bent stents with a radius of 15 mm which may be defined as a worst-case condition. The stents can be loaded with an external pressure up to 100 mmHg at test frequencies of 0-100 Hz. The external pressure load consists of dynamical and a static pressure component. The dynamic pressure component is generated by an electromechanical oscillator (S 52120, TIRA Schwingtechnik, Germany) plus amplifier. The static pressure is applied by a container at given height (hydrostatic pressure). The stent diameter is measured by optical means using a CCD line camera (LIXUSI5000, OPTOLOGIC GmbH, Germany) with 5000 pixels of the size of 7 x 7 µm. In combination with a telecentric lens (S5LPJ9325, Sill Optics GmbH & Co. KG, Germany) an optical resolution of 1.8 µm is achieved. A dimmable LED light was installed in front of the CCD line camera but behind the test chamber for back light illumination. It is mechanically connected with the camera using a drive belt mechanism. Thus, the camera and the light source are movable but adjusted in vis-á-vis position. Camera images are taken at a rate of 1.2 kHz. Real time exposure and processing of the image row data, as well as the user interface for parameter adjustment are managed by the software LIXUSI500 (developed by IBMT). The principle of the test setup is shown in Figure 2.

3D-modell of the sample holder systems for the determination of the stent diameter. System 1 (a) is designated for long peripheral stents (n = 1–2) and system 2 (b) for short coronary or peripheral stents (n = 1–10).

Schematic chart of the testing setup for the determination of the radial stent deformation.
For demonstration, the radial deformations of two peripheral (Stent 1: 8.0×40 mm, stent 2: 4.5×40 mm) and two coronary stents (Stent 3: 4.0×40 mm, stent 4: 2.5×22 mm,) are determined as a function of the measuring position along the stent axis. Three positions (proximal, middle, distal) were investigated at three pressure load modes (40 ± 30, 50 ± 30, 50 ± 40 mmHg) at test frequencies from 0 to 100 Hz. In all cases, the outer diameter of stent plus test tube was measured. To receive the actual stent diameter the tube wall thickness has to be subtracted from the measured diameter.
3 Results
DIN EN ISO 25539-2 advises the use of an optical measurement system for the diameter measurement of stents with an optical resolution of 1% of the smallest test object [9]. For the CCD camera setup a measurement uncertainty of uc95 = 6.4 µm was determined (95 % confidence level). This is better than the 1 % requirement limit (20 µm for a measuring range of 2-10 mm). Reproducible diameters were measured for all four stent types that coincide with the nominal expansion diameter of the stents under static circumstances (0 Hz). The results show that the measurement deviation increases for stents with smaller diameters, see Table 1.
The verification of test frequencies for accelerated durability testing of stents is demonstrated in Figure 3 which shows an exemplary measurement for Stent 3 at varying load frequencies (central position of the stent). Up to 40 Hz the diameter deformation is well within the 10% limit. At 50 and 60 Hz the deformations are even larger, but testing at 70-90 Hz would not meet the requirements. At these frequencies, the measured amplitudes are more than ±10 % smaller than at low frequency loading. In other words, the stent cannot complete its radial deformation and the applied stress would be not relevant. On the other hand, loading at 50, 60 and 100 Hz Stent 3 is expected to be overstressed. These observations were not confirmed for the distal and proximal end.
Similar observations were made for stent 1 and stent 4. No relevant test frequencies resulting in self-resonance of the stents were perceived.
The investigation also revealed that the radial deformation depends on the external load (radial deformation decreases when the external load is decreased). The radial deformation in the middle of the Stent 3 was measured with 8 ± 2 µm (± 0.20 %) at 40 ± 30 mmHg, 20 Hz, and 11 ± 3 µm (± 0.28 %) at 50 ± 40 mmHg, 20 Hz. Besides that the stents showed differing diameter changes along their axis. For example Stent 1 had a maximum deformation in the middle of the stent for all three pressure modes. For Stent 4 larger amplitudes were measured at the proximal and distal end for 50 ± 30 mmHg and 50 ± 40 mmHg while for 40 ± 30 mmHg larger diameter changes were measured in the central position, see Figure 4 and Table 2. Maximum deformation of Stent 3 was measured at 50 ± 40 mmHg, 20 Hz (11 ± 3 µm (± 0.28 %)). According to the detected deformation at the different pressure modes for test frequencies of 0-100 Hz there is no systematic correlation between the radial deformation and the test frequency. The measured low diameter changes are comparable to those Schmidt et al showed in their study [10]. It is an important result that the radial deformation is a function of the external load but diameter changes also depend on stent design, geometric dimension and external loading. The test setup enables diameter-frequency evaluation.

Stent 3 (4.0x40 mm) - radial deformation in the central position as function of test frequency for 50 ± 40 mmHg (exemplary). The red selected area is representing the tolerance range of radial deformation (± 10 % of amplitude at 10 Hz) at which it is reliable that the stent is performing its complete radial deformation.

Radial deformation of stent 1 (8.0×40 mm) in comparison with stent 4 (2.5×22 mm).
Detected diameter of test samples (50 ± 40 mmHg, 0 Hz) in comparison with test mandrels (diameter ± 0.001 mm) with diameters of 2.5 and 6.0 mm as reference (s: PUR-tube wall thickness).
dnominal [mm] | ddetected [mm] | ddetected −2*s [mm] | Bb Δd bb[mm] | |
---|---|---|---|---|
Stent 1 (8.0×40 mm) | 8.000 | 8.118 | 7.968 | 0.032 |
Stent 2 (4.5×40 mm) | 4.500 | 4.702 | 4.462 | 0.038 |
Stent 3 (4.0×40 mm) | 4.000 | 4.096 | 3.946 | 0.054 |
Stent 4 (2.5×22 mm) | 2.500 | 2.513 | 2.363 | 0.137 |
2.500 | 2.458 | – | 0.042 | |
Reference | 6.000 | 5.957 | – | 0.043 |
Overview of the maximum dynamic stent-radial deformation as function of the longitudinal position (proximal, middle, distal) of the peripheral stents and coronary stents (p – proximal, m – middle, d – distal).
maximum radial deformation [mm] | |||||||
---|---|---|---|---|---|---|---|
40 ± 30 mmHg | 50 ± 30 mmHg | 50 ± 40 mmHg | |||||
p | 0.046 ± 0.014 (10 Hz) | ± 0.58% | 0.028 ± 0.006 (90 Hz) | ± 0.36% | 0.038 ± 0.009 (80 Hz) | ± 0.48% | |
Stent 1 (8.0×40 mm) | m | 0.145 ± 0.017 (10 Hz) | ± 1.82% | 0.156 ± 0.017 (90 Hz) | ± 1.96% | 0.210 ± 0.072 (90 tHz) | ± 2.64% |
d | 0.054 ± 0.045 (10 Hz) | ± 0.67% | 0.066 ± 0.021 (100 Hz) | ± 0.83% | 0.092 ± 0.025 (90 Hz) | ± 1.15% | |
p | 0.007 ± 0.002 (60 Hz) | ± 0.17% | 0.008 ± 0.002 (60 Hz) | ± 0.20% | 0.011 ± 0.002 (20 Hz) | ± 0.28% | |
Stent 3 (4.0×40 mm) | m | 0.008 ± 0.002 (20 Hz) | ± 0.20% | 0.008 ± 0.001 (60 Hz) | ± 0.20% | 0.011 ± 0.003 (20 Hz) | ± 0.28% |
d | 0.006 ± 0.002 (60 Hz) | ± 0.15% | 0.008 ± 0.002 (80 Hz) | ± 0.20% | 0.008 ± 0.002 (40 Hz) | ±0.20% | |
p | 0.008 ± 0.002 (80 Hz) | ± 0.33% | 0.008 ± 0.001 (10 Hz) | ± 0.33% | 0.008 ± 0.001 (10 Hz) | ± 0.33% | |
Stent 4 (2.5×22 mm) | m | 0.010 ± 0.002 (70 Hz) | ± 0.43% | 0.008 ± 0.002 (50 Hz) | ± 0.36% | 0.009 ± 0.002 (70 Hz) | ± 0.36% |
d | 0.008 ± 0.002 (60 Hz) | ± 0.33% | 0.008 ± 0.002 (60 Hz) | ± 0.33% | 0.008 ± 0.002 (40 Hz) | ± 0.33% |
4 Conclusion
In this paper a new setup was demonstrated that allows the verification of test frequencies for fatigue testing of stents and an extended, detailed investigation of the radial stent deformation. The use of a high sample rate of the CCD line camera and the new provisions for stent placement under relevant loading conditions make the setup suitable to meet the requirements of DIN ISO EN 25539-2 and FDA guideline 1545 for fatigue testing of vascular stents.
The investigation showed that the setup allows diameter detection with a measurement inaccuracy of uc95 = 6.4 µm. The measured diameters coincide well with nominal expansion diameters of the test samples. The observation of previous studies was confirmed with respect to the radial deformation as a function of external load amplitude. Identification of test frequencies leading to effects of self-resonance of the stents or minimum/maximum radial deformation as function of longitudinal position will require further quantitative studies.
It was also found that the radial deformation depends on stent design, geometric dimension and external loading.
Conclusively, it was demonstrated that the testing set up is also suitable for more detailed investigation of the radial deformation performance of stents along their longitudinal axis. This is necessary for a better understanding of potential mechanical failure especially in the case of long or overlapping stents.
Acknowledgment
This project was partially funded by the European Regional Development Fund (ERDF) and the European Social Fund (ESF) within the collaborative research between economy and science of the state Mecklenburg-Vorpommern (V-630-F-090-2010/273; V-630-S-090-2010/272).
Author's Statement
Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.
References
[1] Holmes et al (2008). The process of bringing new drug eluting stents to market: will they see the light of day? In: JACC Cardiovasc Interv 1:533–4.10.1016/j.jcin.2008.08.001Search in Google Scholar PubMed
[2] Touze et al (2009). Systematic review of the perioperative risks of stroke or death after carotid angioplasty and stenting. Stroke 40: 683–93.10.1161/STROKEAHA.109.562041Search in Google Scholar
[3] Kim et al (2013). A Comparative Reliability and Performance Study of Different Stent Designs in Terms of Mechanical Properties: Fore- shortening, Recoil, Radial Force, and Flexibility. Artificial Organs 37(4):368–379.10.1111/aor.12001Search in Google Scholar PubMed
[4] Sandhu et al (2007). Frequency, etiology, treatment, and outcomes of drug-eluting stent thrombosis during one year of follow-up. Am J Cardiol 99: 465–9.10.1016/j.amjcard.2006.08.058Search in Google Scholar PubMed
[5] Doyle et al (2007). Outcomes of stent thrombosis and restenosis during extended follow-up of patients treated with bare-metal coronary stents. Circulation 116: 2391-8.10.1161/CIRCULATIONAHA.107.707331Search in Google Scholar PubMed
[6] Pelton et al (2008). Fatigue and durability of Nitinol stents. J. Mech. Behav. Biomed. Mater. 1, 153-164.10.1016/j.jmbbm.2007.08.001Search in Google Scholar PubMed
[7] Scheinert, et al (2005). Prevalence and clinical impact of stent fractures after femoropopliteal stenting. J Am Coll Cardiol, 45(2):312-5.10.1016/j.jacc.2004.11.026Search in Google Scholar PubMed
[8] DIN EN ISO 25539-2:2013-05. Cardiovascular Implants - Endovascular Devices – Part 2: Vascular Stents.Search in Google Scholar
[9] FDA Richtlinie Nr. 1545 (04/2010). Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems.Search in Google Scholar
[10] W. Schmidt, P. Behrens (2003). Ermüdungsanalyse von arteriellen Stents nach Europäischer Norm EN 12006-3 Prüfmethoden und Prüfparameter. Rostock-Warnemünde: Internal Report, IIB Warnemünde.Search in Google Scholar
© 2015 by Walter de Gruyter GmbH, Berlin/Boston
This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients