Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
-
D. Wagner
, M. Porsch
, M. Detert , F. Bartel , S. Siedentopf , D. Schindele , JJ. Wendler , A. Lux , B. Schmidt , M. Schostak and UB. Liehr
Abstract
Adequate diagnosis of upper urinary tract (UUT) tumors is essential for successful local treatment. Organsparing approaches are technically difficult and require consistent further development. Appropriate models for investigating new diagnostic and therapeutic methods are not yet available. This study demonstrates the incorporation of a fresh sample model into five different test levels (I-V) for improving the diagnosis and therapy of ureteral tumors. In these test levels, new diagnostic and ablation techniques are evaluated for feasibility, application safety, efficacy and accuracy. An assessment of their suitability for broad preclinical and clinical application also took economic aspects into account.
1 Introduction
Adequate diagnosis of upper urinary tract (UUT) tumors is essential for successful local treatment and organ preservation. Due to the existing diagnostic uncertainty, many patients with ureteral (UR) tumors are still overtreated, which means they undergo nephroureterectomy. Nephrectomy involves dialysis at worst. Conventional imaging (magnetic resonance tomography, computed tomography, extracorporeal ultrasound) are still limited in terms of resolution and diagnostic adequacy [1]. However, diagnostic uncertainty is only part of the problem. Organ-sparing interventions in the upper urinary tract (UUT) are technically difficult and also require consistent further development. There is still a lack of appropriate models for evaluating new diagnostic and therapeutic procedures as well as practical training in their endourological application [2, 3]. This study demonstrates the incorporation of a fresh sample model, developed according to the identified demands, into the respective test levels: I – “Endourological model–UUT”, II – “Intraureteral ultrasound (IUUS)”, III - “Intraureteral laser therapy (IULT)”, IV – “First in men - safety/feasibility (IUUS/IULT)”, and V – “First in men - accuracy/efficacy (IUUS/IULT)”. An assessment of their suitability for broad preclinical and clinical application also took economic aspects into account.
2 Material and methods
2.1 Level I: endourological model–UUT
Aim: Development and production of a multifunctional endourological and practice-equivalent training model (Figure 1). Material: Transparent gel box for holding fresh samples (nephroureteral model “Deutsches Landschwein”), with two hydrogel phases for in situ simulation. The standard endourological instruments (Carl Storz, Tuttlingen, Germany; Uroskop Omnia, Siemens, Erlangen, Germany) and the validated questionnaire “Objective Structured Assessment of Technical Skills (OSATS)” were used to evaluate 15 surgeons in 3 groups with different training levels [4]. Methods: Three training blocks (OSATS and intervention time) were evaluated in the model before and after training (specialist instructions, 60 minutes): 1. Contrast-enhanced imaging of the ureter (UR) and pelvicalyceal system (PCS), placement of a safety wire and rigid ureterorenoscopy (rURS). 2. Placement of a sluice in the renal pelvis via wire, complete inspection of the PCS with a flexible ureterorenoscope (fURS). 3. Disintegration of a kidney stone by laser and removal of 2 concrements (Figure 3).

Endourological model-UUT.
2.2 Level II: intraureteral ultrasound (IUUS)
Aim: Evaluation of intravascular ultrasound (IVUS) in the fresh renal sample model for ureteral diagnosis (IUUS). Material: Volcano CORETM Mobile (Volcano Corp., USA); ultrasound catheter (Eagle Eye® Platinum ST, Visions® PV .018 and Visions® PV .035 (Volcano Corp., USA); Volcano Trak Back II, (Volcano Corp., USA). Endourological model- UUT, test specimens. Methods: IUUS with various IVUS catheters, assessment of subjective visual impression, distinguishability of individual UR wall layers and UR surroundings (test specimens) by 3 trained endourologists in a consensus procedure. Comparison with histological wall structure (Figure 2).

Assessment situation in Level I.
2.3 Level III: intraureteral laser therapy (IULT)
Aim: Evaluation of diode laser (IULT) in the endourological model-UUT for organ-sparing treatment of ureteral tumors, temperature behavior. Material: Compact Laser DIOLAS Photon (Limmer Laser, 980 nm, Berlin, Germany) TempSens72® (Opsens®, Quebec, Canada), Endourological model-UUT, test specimen. Methods: Defined UR lasering of the urothelial and deeper wall layers with various energy levels (3-30W), measurement of local energy input (temperature measurements), histological comparison.
2.4 Level IV: first in men- safety/feasibility (IUUS/IULT)
Aim: Use of an IVUS system to visualize UR wall layers (IUUS) in a clinical “all-in-one setting” with application of the Limmer Laser for IULT. Examination of feasibility and patient safety. Material: Volcano CORETM Mobile (Volcano Corp., USA); ultrasound catheter (Eagle Eye® Platinum ST, Volcano Corp., USA), Compact Laser DIOLAS Photon, Harpoon Fiber 400 µm (Limmer Laser, Berlin, Germany), endourological workplace (Uroskop Omnia, Siemens Health-care, Erlangen, Germany), flexible ureterorenoscope (Flex x2, Carl Storz, Tuttlingen, Germany), Access Sheath (10/12 Charr., Coloplast, Hamburg, Germany), standard urological instruments. Methods: To exclude infiltration, 5 patients with confirmed urothelial carcinoma in the UUT underwent IUUS as well as the standard diagnostic procedure in order to provide legitimation for an “individual curative attempt”. After confirmation, conservative IULT was performed using the Limmer Laser and Harpoon Fiber.

Experimental setup IUUS in Level II.
2.5 Level V: first in men-accuracy/efficacy (IUUS/IULT)
Aim: Examination of diagnostic (IUUS) and therapeutic (IULT) accuracy and efficacy in the context of a “second look”. Material: Analogous to section 2.4. Methods: Five patients diagnosed and treated in the context of an “individual curative attempt” (see section 2.4) again underwent IUUS as well as the standard diagnostic procedure 8-12 weeks later. After excluding infiltration, persistent residual findings were again treated with the Limmer Laser. The degree of efficacy (tumor reduction or removal) was assessed and documented by two experienced endourologists.
3 Results and discussion
3.1 Level I: endourological model –UUT
Result: In its production and application, the “endourological model-UUT” proved to be an inexpensive, simple and robust experimental setup that easily enabled stepwise evaluation of test subjects. Functionality for at least 2 days could be ensured when using fresh samples. It was assessed by all test subjects and evaluators as suitable for the subsequently planned examinations (Level II and III). The OSATS evaluation before and after training in the model revealed marked improvement in the required skills. Discussion: Available training models for the UUT are expensive and impractical [2, 3]. The “endourological model – UUT” enabled practical training in a typical endourological setting. Marked shortening of the learning curve was detectable (Figure 4). Technical improvements without great effort or expense are planned. A limiting factor was the absence of blood flow and respiratory movement in the model.
3.2 Level II: intraureteral ultrasound (IUUS)
Result: With the “endourological model-UUT” modified by test specimens, IUUS examinations could be performed for all US catheters in a robust and standardized manner (automatic US catheter pullback: 1mm/s, 15fps). The investigation focused on visualization of the urothelium and the adjacent muscle layer. Subjectively, the UR wall layers could be distinguished most precisely with the US catheter “Eagle Eye® Platinum ST”. Its US images were

Assessment of training
therefore used for histological comparison. The test specimens could be visualized in the UR surroundings in relation to device settings/catheter (Figure 5). Discussion: Our results suggest that therapeutically relevant tumor infiltrations of these layers can probably be visualized in the human UR; an investigation in a clinical setting should therefore be performed in Levels IV and V. Various study groups have already examined other intraluminal US catheters in a preclinical and clinical setting [7, 8]. Due to the small number of cases, the assessment of suitability for clinical application remained open.
3.3 Level III: intraureteral laser therapy (IULT)
Result: In the “endourological model-UUT”, IULT could be performed easily for defined energy levels (3, 10, 15, 20 watts). Laser ablation restricted to the urothelial layer succeeded subjectively with 3W. Ablations of solid tumors (5-10 mm) required 10-15W. At 20W, dosed energy output for targeted ablation was complicated (adhesions). Temperature elevations measured at the outer surface of the UR during local laser applications remained below the denaturation level (15 W, 980nm, continuous mode). Discussion: The application of laser energy with an angled Harpoon Laser tip enables subtle tissue ablation. Comparative studies of other laser types describe acceptable side effects (SE) [11]. Subjectively improved SE in our examinations therefore enabled application in a clinical setting in Levels IV/V.
3.4 Level IV: first in men- safety/feasibility (IUUS/IULT)
Result: As expected, IVUS catheter placement for IUUS succeeded easily in an endourological setting when applied in humans (5 patients) owing to the catheter dimensions and application similarity (intravascularintraureteral). Visualization of the wall layers succeeded subjectively better than in the “endourological model-UUT”; infiltration of the muscle layer could be excluded in both the IUUS and the standard biopsy (Bx)/urine cytology. Local tumor lasering was therefore carried out using the Limmer Laser (Harpoon Fiber, ’ 400 µm). Papillary tumors up to 5mm could be ablated very precisely to the muscle level of the UR (3 W, 980 nm, continuous mode). There was no intraoperative damage to surroundings. Solid findings (5-10 mm) required higher energy levels (10-15W, 980nm, continuous mode, pulsed mode) for complete ablation. The angled laser tip (30°) enables very precise layerwise ablation of tumor tissue in both continuous and pulsed mode. Both IUUS and IULT proved to be simple and safe methods for human application. The actual efficacy (tumor ablation) and accuracy (damage to surroundings) of the two methods should be established in the recommended standard endourological follow-ups (Level V). Discussion: Placement of the IVUS catheter in an IUUS setting does not differ from typical endourological procedures, and the better spatial resolution enabled a more detailed diagnosis than that achieved with currently recommended imaging. Various study groups have already tested other intraluminal US catheters (30MHz, 5F, B-mode; 12.5/20MHz, 6.2F) for detailed diagnosis of UR tumors [7, 8]. IULT management and ablation behavior made it possible to ablate even larger tumors with simultaneous hemostasis and no collateral damage. The applied 400 µm laser fiber permitted controlled flexion of the fURS for lasering lower calices. The angled laser tip offers marked advantages over other commercially available systems [5, 6].

Comparison of IUUS and histology.

Comparison of IUUS and endoscopy.
3.5 Level V: first in men- accuracy/efficacy (IUUS/IULT)
Result: The endourological follow-ups 8-12 weeks after the primary intervention (IUUS, IULT) showed good tumor control in all patients. Small papillary findings had disappeared completely in the optically regenerated urothelial layer. After lasering several large (5-10 mm) solid findings, two small residual solid tumors were seen in 1 case. Reexamination for possible infiltration into deeper wall layers after standard Bx/urine cytology and IUUS remained negative. The residual findings were again submitted to IULT with the Limmer Laser (Harpoon Fiber, Ř 400 µm, 3-10W, 980nm, continuous mode, pulsed mode). This patient was also tumor-free after another 8 weeks. Imaging of the wall layers also succeeded subjectively better after primary therapy (IULT) than in the “endourological model-UUT”. Both IUUS and IULT proved to be accurate and effective in tumor therapy. The “third look” disclosed no evidence of damage to surroundings. Discussion: Investigations on the long-term follow-up of laser-treated UR tumors are rare [9, 10]. Guidelines for IUUS diagnosis are lacking, and they are vague for IULT due to a lack of data [1]. With strict indications, good tumor control with organ preservation could be achieved in our cohorts (Figure 6). Side effects did not occur and are also assessed as minimal in the literature [11]. Limiting factors were the small number of patients and the short follow-up.
4 Conclusion
Our endourological model-UUT can be safely and effectively used as a training tool to shorten learning curves. Endourological techniques and conservative tumor treatment in the upper urinary tract is an evolving field and can be successfully used as first-line treatment for UUT tumors in selected patients. Recent technical advancements should now be consistently investigated and standardized with regard to diagnosis and therapy.
Author's Statement
Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.
References
[1] M. Roupr[eogonek]t, M. Babjuk, E. Compérat, R. Zigeuner, R. Sylvester, M. Burger, N. Cowan, A. Böhle, Van Rhijn, Bas W G, E. Kaasinen, J. Palou, and S. F. Shariat, “European guidelines on upper tract urothelial carcinomas: 2013 update,” (eng), Eur. Urol, vol. 63, no. 6, pp. 1059–1071, 2013.10.1016/j.eururo.2013.03.032Search in Google Scholar
[2] Brunckhorst O1, Aydin A1, Abboudi H1, Sahai A1, Khan MS1, Dasgupta P1, Ahmed K2. Simulation-based ureteroscopy training: a systematic review. J Surg Educ. 2015 Jan-Feb;72(1):135-43. 10.1016/j.jsurg.2014.07.003. Epub 2014 Aug 15.Search in Google Scholar
[3] Strohmaier WL1, Giese A. Porcine urinary tract as a training model for ureteroscopy. Urol Int. 2001;66(1):30-2.10.1159/000056559Search in Google Scholar
[4] Martin JA, Regehr G, Reznick R, MacRae H, Murnaghan J, Hutchison C, Brown M.Objective structured assessment of technical skill (OSATS) for surgical residents. Br J Surg. 1997 Feb; 84(2):273-8.10.1046/j.1365-2168.1997.02502.xSearch in Google Scholar
[5] Nathaniel M. Fried and Keith E. Murray. High-Power Thulium Fiber Laser Ablation of Urinary Tissues at 1.94 µm. J of Endourology. 2005, 19(1): 25-31.10.1089/end.2005.19.25Search in Google Scholar
[6] Bader MJ, Sroka R, Gratzke C, Seitz M, Weidlich P, Staehler M, Becker A, Stief CG, Reich O. Laser therapy for upper urinary tract transitional cell carcinoma: indications and management. Eur Urol. 2009 Jul;56(1):65-71.10.1016/j.eururo.2008.12.012Search in Google Scholar
[7] Matin SF1, Kamat AM, Grossman HB. High-frequency endoluminal ultrasonography as an aid to the staging of upper tract urothelial carcinoma: imaging findings and pathologic correlation. J Ultrasound Med. 2010 Sep;29(9):1277-84.10.7863/jum.2010.29.9.1277Search in Google Scholar
[8] Liu JB1, Bagley DH, Conlin MJ, Merton DA, Alexander AA, Goldberg BB. Endoluminal sonographic evaluation of ureteral and renal pelvic neoplasms. J Ultrasound Med. 1997 Aug;16(8):515-21.10.7863/jum.1997.16.8.515Search in Google Scholar
[9] Elliott DS1, Blute ML, Patterson DE, Bergstralh EJ, Segura JW. Long-term follow-up of endoscopically treated upper urinary tract transitional cell carcinoma. Urology. 1996 Jun; 47(6):819-25.10.1016/S0090-4295(96)00043-XSearch in Google Scholar
[10] Thompson RH1, Krambeck AE, Lohse CM, Elliott DS, Patterson DE, Blute ML. Endoscopic management of upper tract transitional cell carcinoma in patients with normal contralateral kidneys. Urology. 2008 Apr;71(4):713-7.10.1016/j.urology.2007.11.018Search in Google Scholar PubMed
[11] Defidio L1, De Dominicis M, Di Gianfrancesco L, Fuchs G, Patel A. First collaborative experience with thulium laser ablation of localized upper urinary tract urothelial tumors using retrograde intra-renal surgery. Arch Ital Urol Androl. 2011 Sep; 83(3):147-53.Search in Google Scholar
© 2015 by Walter de Gruyter GmbH, Berlin/Boston
This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients