Stability analysis of ferrofluids

  • Duda Katharina EMAIL logo and Lüdtke-Buzug Kerstin
Published/Copyright: September 12, 2015

Abstract

Superparamagnetic iron oxides (SPIOs) are used as tracer for the new imaging technique Magnetic Particle Imaging. The stability of ferrofluids for medical application has a great importance, in addition to the particle size. The shell material, which protects the iron core prior from agglomeration and sedimentation, can be degraded by various processes. Another important aspect of stability is the constant performance of magnetisation. Therefore, the measurement of the magnetisation of the particles must be controlled in order to ensure the stability of the samples.

1 Introduction

Magnetic Particle Imaging (MPI) denotes a new imaging method, which is attributable to the tracer method. The tracers are administered to the patient intravenously and an excited external magnetic field detects the with the measurement of the magnetic response of the particles. The aim of this method is an accurate and real-time imaging of the spatial distribution of the nanoparticles. The possible applications in medical imaging are the visualization of organs and a 3D real-time monitoring of the cellular level of metabolic processes. Due to the versatile possibilities for the use of MPI, special demands are placed on the tracer material. Crucial for the quality of imaging is the quality of the used nanoparticles.

The properties, which means the size of the core, the hole particle and also the stability of the nanoparticles influences the imaging significantly. Important paramaters such as the hydrodynamic diameter, core size and particle magnetization, can be investigated and controlled by the photon cross-correlation spectroscopy (PCCS) and magnet particle spectrometry (MPS). The samples were stored and measured by 15 °C. Therefore in these work the samples for the MPS measurement series were measured at room temperature. In order to evaluate the behavior of the SPIOs the samples were measured over several weeks and statistically evaluated.

The factor of stability at the superparamagnetic iron oxide particles plays an increasingly important role. The synthesis of the particles is a time consuming expensive process, therfore so therefore the stability of the particles should be as large as possible. The particles are considered stable if they neither agglomerate nor sediment. A shell of dextran or other polymers used to protect the particles from agglomeration. When the shell material degrades and becomes thinner, it may come to dipole-dipole interactions between the particles. These particles may build bigger agglomerates and are not longer stable and they sediment. Changes in the geometry of the particles can be examined by the PCCS and MPS measurement. If changes are measured, it can be assumed that the suspensions of the ferrofluids are no longer stable [9] [8] [5].

2 Methods

2.1 Photon cross-correlation spectroscopy (PCCS)

The PCCS is able to analyzes highly concentrated suspensions. A laser beam is split into two laser beams with the same intensity. The rays are refracted so that they intersect in the same scattering volume. It results two equivalent speckle pattern. The identical components in the speckle patterns resulting from simply scattered light, the differences resulting from multiple scattering. The intensity changes are detected by two detectors. The light of equivalent speckle is cross-correlated in a correlator to filter out the identical shares.

The response of the cross-correlation function contains two parameters. On the one hand, the slope of the function is related to the diffusion rate and also for particle size. On the other hand, the amplitude gives information about how much the transparency of the sample changes during the measurement [1] [2] [3] [4].

2.2 Magnet particle spectrometry (MPS)

For the characterization of superparamagnetic iron oxide nanoparticles, also in this contribution the magnetic particle spectrometry is used. Using an oscillating excitation field, which is generated by a drive- field coil, the iron oxide nanoparticles experience a time-varying magnetization. The change in the magnetization induces a voltage in a receiver coil, which generates a homogeneous magnetic field. By the homogeneous field all particles have the same magnetization. A receiving coil records the change of the magnetization. Due to the non-linearity of the magnetization-curve the recorded spectrum has harmonics which decreases with increasing frequency. A linear decrease of the harmonics indicates an optimum particle quality with ideal particle magnetization. The magnetization curve can be mathematically described by the Langevin function ξ(t)=πdk3Msμ0H(t)kBTa is the saturation magnetization, H(t) is the time-varying magnetization, μ0 is the magnetic field constant, kB is the Boltzmann constant and Ta is the absolute temperature. With this function it is possible to specify a particle size distribution. [5] [4] [6] [7] [8].

3 Experimental set up

To investigate the hydrodynamic diameter dH of the SPIOs, four samples of the same batch of particles were prepared. The undiluted solution KLB_063 used has an iron concentration of 0.33 mol/L. Four cuvettes were filled with 3 mL of different concentrated solutions of KLB_063. Samples A1 and B1 were mixed with 500 μL of undiluted iron oxide solution and 2500 μL of distilled water and samples A2 and B2 with 250 μL of undiluted iron oxide solution and 2750 μL distilled water.

To investigate the iron core diameter of the SPIOs, two samples were prepared. A dilution of the samples was not necessary for the measurement by the MPS. Each sample had a volume of 10 μL.

All samples were stored at 15 °C. The samples for the PCCS were also measured at 15 °C. Each sample was measured five times in order to avoid fluctuations. The samples were analyzed over a period of eleven weeks, and measured at regular intervals.

In order to evaluate the measured values of the hydrodynamic diameter, it is necessary to know the device-specific deviation. Measurements have shown that the deviation of the PCCS is about 8 %.

The particle size distribution is realized by the Langevin function. The function does not account on relaxation effects or non-spherical particles. Therefore, it is possible that certain solutions, like KLB_063, were calculated incorrectly. For this reason, it is impossible to calculate a device-specific deviation in this case.

4 Results and conclusion

4.1 Hydrodynamic diameter of the SPIOs

All measured values were normalized to the initial. Therefore, the first measurement functions as standard.

Table 1

Hydrodynamic Diameter dH is the starting value for the following measurements

samplesdHinnmM
A1109114120108110112
B1123116126117117120
A2105108103106107106
B2102106106107108106

The samples were measured continuously for eleven weeks in order to detect possible variations in the hydrodynamic diameter. The averaged hydrodynamic diameter of sample A is 116 nm and of sample B 106 nm. The average difference of about 10 nm can be explained by the high values of sample B1. During the filling of the cuvette it may comes to impurities.

Figure 1 Development of the averaged hydrodynamic diameter of the sample A1 and B1 over eleven weeks. The red lines are the upper and lower bounds. The bounds are based on the device-specific deviations of the PCCS. The blue line represents the average diameter of the first measurement.
Figure 1

Development of the averaged hydrodynamic diameter of the sample A1 and B1 over eleven weeks. The red lines are the upper and lower bounds. The bounds are based on the device-specific deviations of the PCCS. The blue line represents the average diameter of the first measurement.

The average diameter exhibit variations, the measurements No. 17 and 19 are below the lower bound of 8 % deviation. In addition, the deviation has increased during the cycle of measurement. Which suggests that the hydrodynamic diameter of the SPIOs has changed from sample A1 and B1. The reason for the difference could be a possible contamination of the sample B1, or the lower concentration of SPIOs in the first sample. It is to be noted that the measured hydrodynamic diameter has been reduced in the first sample during the measurement campaign. Initially, the average was still 116 nm, after eleven weeks it is only 109 nm. The reduction of the particle diameter confirmed the hypothesis that large, maybe some larger nano particles sediment in the medium and therefore they can no longer be detected by the PCCS. To confirm this assumption, the second sample is examined, which has a lower concentration of iron oxide.

Figure 2 Development of the averaged hydrodynamic diameter dHoof the second sample over eleven weeks. The red lines are the upper and lower bounds. The bounds are based on the device-specific deviations of the PCCS. The blue line represents the average diameter of the initial measurement of A2 and B2.
Figure 2

Development of the averaged hydrodynamic diameter dHoof the second sample over eleven weeks. The red lines are the upper and lower bounds. The bounds are based on the device-specific deviations of the PCCS. The blue line represents the average diameter of the initial measurement of A2 and B2.

The deviations of the averaged hydrodynamic diameter of the lower concentrated sample is lower than of the sample A1 and B1, which has a higher concentration. The variations are below the device-specific deviation of the PCCS. Overall, the average values of the second sample are very close to the standard value. The lower deviation is due to the smaller default value that has been calculated at the beginning. The second sample has a lower iron concentration. Therefore it can be assumed that the iron oxide particle content is largely responsible for the stability of a suspension. The lower concentration of 2.5 μL / 3 mL iron oxide has a positive impact on the stability of the suspension. The mean deviation after eleven weeks is 0.02742. After eleven weeks, no specific changes in the hydrodynamic diameter of the second sample can be detected. The calculated difference is even less than the calculated measuring inaccuracy of the PCCS. The black line do not touch the r ed upper and lower bounds. Accordingly, the samples who have a lower iron concentration, own the better stability properties at 15 °C. Because there have been no sedimentation in the medium and no large fluctuations of the hydrodynamic diameter.

4.2 Dore diameter of the SPIOs

To be able to control the outcome of the PCCS investigation, the results of the MPS measurement need to be controlled. The MPS offers the possibility to measure the iron oxide core diameter, we used therefore undiluted particle solutions, to get a high response of the measured particles. It is important to control the iron oxide core diameter to examine it in terms of stability. All measured values were normalized to the first. Therefore, the first measurement functions as standard.

Table 2

core diameter dk of the first MPS measurement

samplesdkinnmM
A16.015.916.015.815.915.92
B15.715.615.615.715.515.62

The samples were measured continuously for eleven weeks in order to detect possible variations in the core diameter. The averaged core diameter of sample A and B is 15.8 nm.

Figure 3 Development of the averaged core diameter of sample A and B over eleven weeks. The red lines are the upper and lower bounds of 5 % deviation. The blue line represents the average core diameter of the first measurement.
Figure 3

Development of the averaged core diameter of sample A and B over eleven weeks. The red lines are the upper and lower bounds of 5 % deviation. The blue line represents the average core diameter of the first measurement.

The measurement results of the averaged core diameter over a period of eleven weeks shows only very small deviations. There are almost all of the calculated deviations below 5%. This results an average relative deviation of 0.0349.

The control of the iron oxide core diameter can not support the measurement result from the investigation of the hydrodynamic diameter fully. Due to the fact that PCCS and MPS use different measurement methods, the MPS may only provide additional insight into the particle quality of the solution during the measurement period. As just presented, the deviation of the iron oxide core diameter is very slight. This means that a stable iron diameter also points to a constant particle quality. At a storage temperature of 15 °C the undiluted iron solution can be regarded as stable over a long period, in this work shown for nearly 3 month.

5 Outlook

The measurements have shown that the storage of the samples at 15 °C has a positive effect on the stability of the iron oxide particles. In addition, it is advantageous to store samples with a lower iron concentration. To make the aspect of stability available for medical applications, it is useful to store the iron oxide particles in fluids which are similar to the pH value of blood. The PCCS also offers the opportunity to study opaque suspensions, so it is possible to take a measurement with iron oxide particles which are dissolved in porcine or bovine blood. In the future the tracer materials could be solved and stored in autologous blood in order to conduct an investigation with the MPI. The bearings of the tracer in the autologous blood could improve the compatibility of the injection and thus the MPI signal. For this purpose, additional tests must be performed to deal with the particle stability at the temperature elevation from 15 to 37 °C body temperature [10] [11].

Author's Statement

  1. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.

References

[1] Frauenhofer Institut Fertigungstechnik Materialforschung: Partikelgrössen- und Stabilitätsanalyse von Nanoteilchen, Produktblatt. http://www.ifam.fraunhofer.de/content/dam/ifam/de/documents/IFAM-Bremen/2804/fachinfo/infoblaetter/de/oe411/Produktblatt-2804-DE-Klebtechnik-Partikelgroessen-_und-Stabilitaetsanalyse_von_Nanoteilchen.pdf.- abgerufen am 15.05.2014.Search in Google Scholar

[2] Sympatec GmbH: Photonenkreuzkorrelationsspektroskopie. http://www.sympatec.com/DE/PCCS/PCCS.html.- abgerufen am 09.10.2014.Search in Google Scholar

[3] W. Lämmle, Particle size and stability analysis in turbid suspensions and emulsions with Photon Cross Correlation Spectroscopy. Sympatec GmbH, System-Partikel-Technik, ClausthalZellerfeld.Search in Google Scholar

[4] S. Bregenzer, ViskositÄt und Diffusion in Ferrofluiden. http://kops.uni-konstanz.de/bitstream/handle/123456789/9296/Bregenzer.pdf?sequence=1&isAllowed=y, (2004). - Wissenschaftliche Arbeit, abgerufen am 29.04.2014Search in Google Scholar

[5] S. Biederer, Entwicklung eines Spektrometers zur Analyse superparagmagnetischer Eisenoxid-Nanopartikel für Magnetic-Particle-Imaging. Vieweg+Teubner Verlag, Wiesbaden. (2012).10.1007/978-3-8348-2407-3_3Search in Google Scholar

[6] S. Biederer, S. Kren, T.F. Sattel, M. Erbe, T. Knopp, K. LüdtkeBuzug, T.M. Buzug, Ein magnetisches Partikel-Spektrometer zur Messung der Magnetisierung von Nanopartikeln unter der Verwendung von AC- und DC-Feldern. In: Proceedings Biomedizinische Technik / Biomedical Engineering 55, (Suppl.1), Rostock (2010). 10.1515/BMT.2010.295.Search in Google Scholar

[7] S. Biederer, T. Knopp, T. F. Sattel, K. Lüdtke-Buzug, B. Gleich, J. Weizenecker, J. Borgert, T. M. Buzug, Magnetization response spectroscopy of superparamagnetic nanoparticles for magnetic particle imaging. In: Journal of Physics D: Applied Physics 42 (2009), Nr. 20. 10.1088/0022-3727/42/20/205007Search in Google Scholar

[8] K. Lüdtke-Buzug, S. Biederer, T.M. Buzug, Analyse des Seperationsergebnisses bei der Herstellung Superparamagnetischer Eisenoxid-Nanopartiklel für Magnetic-Particle-Imaging. In: Proceedings Biomedizinische Technik / Biomedical Engineering 55,(Suppl.1), Rostock (2010). 10.1515/BMT.2010.640Search in Google Scholar

[9] K. Lüdtke-Buzug, Von der Synthese zur klinischen Anwendung Magnetische Nanopartikel. In: Chemie unserer Zeit, Wiley-VCH Verlag GmbH & Co KGaA, Weinheim 46 (2012), S. 32-39. 10.1002/ciuz.201200558.Search in Google Scholar

[10] A.Oberle, K. Lüdtke-Buzug, Stability Analysis of Superparamagnetic Iron Oxide Nanoparticles (SPIONs) at 37 °C. Conference Paper BMT. 2013.10.1515/bmt-2013-4099Search in Google Scholar PubMed

[11] C. Borchers, K.Lüdtke-Buzug, Stability Evaluation of Super-paramagnetic Iron Oxide Nanoparticles in Different Media for Magnetic Particle Imaging. Abstract. 2011.10.1142/9789814324687_0006Search in Google Scholar

Published Online: 2015-9-12
Published in Print: 2015-9-1

© 2015 by Walter de Gruyter GmbH, Berlin/Boston

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Articles in the same Issue

  1. Research Article
  2. Development and characterization of superparamagnetic coatings
  3. Research Article
  4. The development of an experimental setup to measure acousto-electric interaction signal
  5. Research Article
  6. Stability analysis of ferrofluids
  7. Research Article
  8. Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
  9. Research Article
  10. Energy harvesting for active implants: powering a ruminal pH-monitoring system
  11. Research Article
  12. New type of fluxgate magnetometer for the heart’s magnetic fields detection
  13. Research Article
  14. Field mapping of ballistic pressure pulse sources
  15. Research Article
  16. Development of a new homecare sleep monitor using body sounds and motion tracking
  17. Research Article
  18. Noise properties of textile, capacitive EEG electrodes
  19. Research Article
  20. Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
  21. Research Article
  22. Spike sorting: the overlapping spikes challenge
  23. Research Article
  24. Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
  25. Research Article
  26. Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
  27. Research Article
  28. Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
  29. Research Article
  30. Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
  31. Research Article
  32. Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
  33. Research Article
  34. Compressed sensing of multi-lead ECG signals by compressive multiplexing
  35. Research Article
  36. Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
  37. Research Article
  38. Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
  39. Research Article
  40. Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
  41. Research Article
  42. Effects of sampling rate on automated fatigue recognition in surface EMG signals
  43. Research Article
  44. Closed-loop transcranial alternating current stimulation of slow oscillations
  45. Research Article
  46. Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
  47. Research Article
  48. The role of expert evaluation for microsleep detection
  49. Research Article
  50. The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
  51. Research Article
  52. Metal artifact reduction by projection replacements and non-local prior image integration
  53. Research Article
  54. A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
  55. Research Article
  56. Processing of membranes for oxygenation using the Bellhouse-effect
  57. Research Article
  58. Inkjet printing of viable human dental follicle stem cells
  59. Research Article
  60. The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
  61. Research Article
  62. New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
  63. Research Article
  64. QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
  65. Research Article
  66. Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
  67. Research Article
  68. Next-generation vision testing: the quick CSF
  69. Research Article
  70. Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
  71. Research Article
  72. Design and control of a 3-DOF hydraulic driven surgical instrument
  73. Research Article
  74. Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
  75. Research Article
  76. Frequency based assessment of surgical activities
  77. Research Article
  78. “Hands free for intervention”, a new approach for transoral endoscopic surgery
  79. Research Article
  80. Pseudo-haptic feedback in medical teleoperation
  81. Research Article
  82. Feasibility of interactive gesture control of a robotic microscope
  83. Research Article
  84. Towards structuring contextual information for workflow-driven surgical assistance functionalities
  85. Research Article
  86. Towards a framework for standardized semantic workflow modeling and management in the surgical domain
  87. Research Article
  88. Closed-loop approach for situation awareness of medical devices and operating room infrastructure
  89. Research Article
  90. Kinect based physiotherapy system for home use
  91. Research Article
  92. Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
  93. Research Article
  94. Integrating multimodal information for intraoperative assistance in neurosurgery
  95. Research Article
  96. Respiratory motion tracking using Microsoft’s Kinect v2 camera
  97. Research Article
  98. Using smart glasses for ultrasound diagnostics
  99. Research Article
  100. Measurement of needle susceptibility artifacts in magnetic resonance images
  101. Research Article
  102. Dimensionality reduction of medical image descriptors for multimodal image registration
  103. Research Article
  104. Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
  105. Research Article
  106. Evaluation of CT capability for the detection of thin bone structures
  107. Research Article
  108. Towards contactless optical coherence elastography with acoustic tissue excitation
  109. Research Article
  110. Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
  111. Research Article
  112. Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
  113. Research Article
  114. Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
  115. Research Article
  116. Video tracking of swimming rodents on a reflective water surface
  117. Research Article
  118. MR imaging of model drug distribution in simulated vitreous
  119. Research Article
  120. Studying the extracellular contribution to the double wave vector diffusion-weighted signal
  121. Research Article
  122. Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
  123. Research Article
  124. Introducing a frequency-tunable magnetic particle spectrometer
  125. Research Article
  126. Imaging of aortic valve dynamics in 4D OCT
  127. Research Article
  128. Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
  129. Research Article
  130. Simple concept for a wide-field lensless digital holographic microscope using a laser diode
  131. Research Article
  132. Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
  133. Research Article
  134. Respiratory surface motion measurement by Microsoft Kinect
  135. Research Article
  136. Improving image quality in EIT imaging by measurement of thorax excursion
  137. Research Article
  138. A clustering based dual model framework for EIT imaging: first experimental results
  139. Research Article
  140. Three-dimensional anisotropic regularization for limited angle tomography
  141. Research Article
  142. GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
  143. Research Article
  144. Experimental computer tomograph
  145. Research Article
  146. US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
  147. Research Article
  148. Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
  149. Research Article
  150. Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
  151. Research Article
  152. Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
  153. Research Article
  154. An IHE-conform telecooperation platform supporting the treatment of dementia patients
  155. Research Article
  156. Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
  157. Research Article
  158. Identification of surgical instruments using UHF-RFID technology
  159. Research Article
  160. A generic concept for the development of model-guided clinical decision support systems
  161. Research Article
  162. Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
  163. Research Article
  164. Creating 3D gelatin phantoms for experimental evaluation in biomedicine
  165. Research Article
  166. Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
  167. Research Article
  168. Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
  169. Research Article
  170. Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
  171. Research Article
  172. Wireless medical sensors – context, robustness and safety
  173. Research Article
  174. Sequences for real-time magnetic particle imaging
  175. Research Article
  176. Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
  177. Research Article
  178. A machine learning approach for planning valve-sparing aortic root reconstruction
  179. Research Article
  180. An in-ear pulse wave velocity measurement system using heart sounds as time reference
  181. Research Article
  182. Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
  183. Research Article
  184. Multisegmental fusion of the lumbar spine a curse or a blessing?
  185. Research Article
  186. Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
  187. Research Article
  188. A muscle model for hybrid muscle activation
  189. Research Article
  190. Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
  191. Research Article
  192. An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
  193. Research Article
  194. Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
  195. Research Article
  196. Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
  197. Research Article
  198. Mesh structure-independent modeling of patient-specific atrial fiber orientation
  199. Research Article
  200. Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
  201. Research Article
  202. Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
  203. Research Article
  204. A java based simulator with user interface to simulate ventilated patients
  205. Research Article
  206. Evaluation of an algorithm to choose between competing models of respiratory mechanics
  207. Research Article
  208. Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
  209. Research Article
  210. Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
  211. Research Article
  212. Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
  213. Research Article
  214. FPGA controlled artificial vascular system
  215. Research Article
  216. Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
  217. Research Article
  218. Test setup for characterizing the efficacy of embolic protection devices
  219. Research Article
  220. Impact of electrode geometry on force generation during functional electrical stimulation
  221. Research Article
  222. 3D-based visual physical activity assessment of children
  223. Research Article
  224. Realtime assessment of foot orientation by Accelerometers and Gyroscopes
  225. Research Article
  226. Image based reconstruction for cystoscopy
  227. Research Article
  228. Image guided surgery innovation with graduate students - a new lecture format
  229. Research Article
  230. Multichannel FES parameterization for controlling foot motion in paretic gait
  231. Research Article
  232. Smartphone supported upper limb prosthesis
  233. Research Article
  234. Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
  235. Research Article
  236. Evaluation of adhesion promoters for Parylene C on gold metallization
  237. Research Article
  238. The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
  239. Research Article
  240. Increasing the visibility of thin NITINOL vascular implants
  241. Research Article
  242. Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
  243. Research Article
  244. Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
  245. Research Article
  246. Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
  247. Research Article
  248. Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
  249. Research Article
  250. Electrospun vascular grafts with anti-kinking properties
  251. Research Article
  252. Integration of temperature sensors in polyimide-based thin-film electrode arrays
  253. Research Article
  254. Use cases and usability challenges for head-mounted displays in healthcare
  255. Research Article
  256. Device- and service profiles for integrated or systems based on open standards
  257. Research Article
  258. Risk management for medical devices in research projects
  259. Research Article
  260. Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
  261. Research Article
  262. Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
  263. Research Article
  264. SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
  265. Research Article
  266. Numerical simulations of airflow in the human pharynx of OSAHS patients
Downloaded on 7.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2015-0003/html
Scroll to top button