This work aimed at the detection of rotor centers within the atrial cavity during atrial fibrillation on the basis of phase singularities. A voxel based method was established which employs the Hilbert transform and the phase of unipolar electrograms. The method provides a 3D overview of phase singularities at the endocardial surface and within the blood volume. Mapping those phase singularities from the inside of the atria at the endocardium yielded rotor center trajectories. We discuss the results for an unstable and a more stable rotor. The side length of the areas covered by the trajectories varied from 1.5 mm to 10 mm. These results are important for cardiologists who target rotors with RF ablation in order to cure atrial fibrillation.

1 Introduction

Atrial fibrillation (AF) is the most common sustained cardiac arrhythmia [11]. During AF, the atria are typically excited more than 350 times per minute as the refractory periods of atrial cells are compressed [6]. Different excitation patterns occur. This work deals with the task of determining the centers of rotatory excitation patterns on the basis of simulated intraatrial electrograms in three-dimensional space.

The behavior of rotor filaments within ventricular and atrial tissue has already been investigated [1, 2, 5]. Especially in clinical practice, electrograms cannot be recorded within tissue. Instead, catheters are placed on the endocardium and within blood. The detection of rotor centers and their trajectory within blood generates new challenges. The signal amplitude for example decreases with increasing distance to the source on the endocardium. In this work, we introduce a method which is capable of detecting rotor centers within the whole atrial volume. Considering the rotor centers near the endocardium leads to the question, how the rotor center trajectories look like, which difficulties occur with respect to their detection in simulated signals and and in clinical data. Two rotors – an unstable and a more stable one – are discussed.

Different methods already exist for the detection of rotor centers [4]. This work focuses on a phase based approach. A phase singularity (PS) is a point in a phase map which is surrounded by the complete range of monotonically increasing phase values from −π toπ. At the PS itself, the phase is undefined [9]. In the following, rotor centers are assumed to be situated at PSs. The position of rotor centers is of interest in clinical practice [7] as creating myocardial lesions near to the rotor center is believed to alleviate or cure AF.

2 Material and methods

2.1 Simulation setup

A personalized anatomical model of the atria and simulation results were taken from earlier work [3]. The model included complete rule-based fiber orientation and was adjusted on the basis of the Courtemanche-Raimirez-Nattel model for cardiac cells [3]. The model consists of 263 x 328 x 283 voxels with a length of 0.33 mm. Extracellular potentials within blood resulted from transmembrane voltages by forward calculation. The extracellular potentials were interpreted as virtual intracardiac electrograms.

2.2 Hilbert transform

In this work, phase values of the signals were calculated with the help of the Hilbert transform which is defined as follows: Choose a filter G (f) which turns the phase of negative input frequencies forward by 90° respectively +π and the phase of positive input frequencies backward by 90° respectively −π. Constant components are suppressed by G(f). If a signal y(t) passes G(f), the Hilbert transform H{y(t)} results at the output of the filter. Calculationswere based on the MATLAB (MathWorks, Inc.) implementation of the Hilbert transform.

2.3 Phase calculation based on the Hilbert transform

For a signal with periodic properties, a measure of the signals progression [9] within a period at a certain point in time can be an aspired attribute. The instantaneous phase represents one possibility to map that position to the interval from −π to π. As the actual phase calculation with the inverse tangent is not independent of constant components, we subtracted the mean value of the signal before proceeding [8]. Let (t) be

y˜(t)=y(t)y(t)¯

with y(t)¯=const being the mean value of y(t). We defined z(t) as the analytic signal which corresponds to (t):

z(t)=y˜(t)+jH{y˜(t)}

The phase φ(t) of the complex-valued analytic signal z(t) was calculated with the help of the inverse tangent:

φ(t)=arctan(imag{z(t)}real{z(t)})

real(z(t)) and imag(z(t)) denote the real and imaginary part of z(t).

2.4 Detection of phase singularities in three-dimensional volumes

To begin with, a method was introduced which was capable of detecting PSs in 2D space. In a 2D pixel model, a square composed of four neighboring pixels had to meet two conditions if it enclosed a PS (see Figure 1):

  1. A phase jump from about −π to about +π connects two pixels which share a common edge (path A).

  2. The phase outlines a monotone course while tracking path B.

Figure 1 Two conditions define a PS in a 2D square of four pixels. Path A connects two pixels via a phase jump and the phase outlines a monotone course along path B. For reasons of graphical simplicity, each pixel is represented by a node in its center.
Figure 1

Two conditions define a PS in a 2D square of four pixels. Path A connects two pixels via a phase jump and the phase outlines a monotone course along path B. For reasons of graphical simplicity, each pixel is represented by a node in its center.

Figure 2 Nine planes can be extracted from each cube consisting of eight voxels: three planes parallel to the coordinate axes of the voxel lattice and six planes in diagonal directions. For reasons of graphical simplicity, each voxel is represented as a node in its center.
Figure 2

Nine planes can be extracted from each cube consisting of eight voxels: three planes parallel to the coordinate axes of the voxel lattice and six planes in diagonal directions. For reasons of graphical simplicity, each voxel is represented as a node in its center.

In a 3D voxel space, the task of detecting PSs became more difficult. The orientation of the rotor and thus the plane in which the PS might be located, was not predefined anymore. As the introduced model consisted of voxels, it provided a cubic structure. For each voxel, a cube was assembled with 7 neighboring voxels by walking along the coordinate axes (see Figure 2).

In order to detect PSs with unknown orientation, several planes had to be taken into account. Figure 2 shows three planes in parallel to the coordinate axes and six planes in diagonal directions. For each voxel, these nine planes were extracted and checked for the existence of a PS as described above. The resulting PSs characterize a 3D rotor, which is also called spiral wave.

2.5 From phase singularities to rotor center trajectories

The trajectory of the rotor center at the endocardium was read out after the detection of the PSs in the voxel model as described above. A triangular mesh represented the endocardium. PSs which belonged to the same rotor center filament were united into a group. The PS which was located closest to the endocardium was projected onto the endocardial surface for each group and all time steps.

3 Results

3.1 Intraatrial electrograms and corresponding phase signals

Figure 3 shows exemplarily two unipolar intraatrial electrograms. Both were recorded near a rotor center. Beneath each electrogram, the phase signal is shown which was calculated on the basis of the Hilbert transform.

Figure 3 The first and the third plot depict intraatrial electrograms (EGM). The second and the fourth plot show the corresponding phase signals.
Figure 3

The first and the third plot depict intraatrial electrograms (EGM). The second and the fourth plot show the corresponding phase signals.

The upper phase signal in Figure 3 shows the desired linear shape with phase jumps from −π to π. The lower phase signal deviates from that linear shape which is over-laid by ripples. In addition, the phase begins to oscillate near the phase jumps.

3.2 Phase singularities and rotor center trajectories

The phase values of one exemplary time step are mapped on the endocardial surface in Figure 4. The rotor center was detected with the help of the method described in section 2.4.

Rotor center trajectories of two rotors were investigated with a sample rate of 1 kHz. Rotor 1 was located in the right atrium near the tricuspid valve. Rotor 1 was observed in the time interval from 3.6 s to 5.0 s. Time is referenced to the start of the simulation at 0 s. During that time, rotor 1 rotated with a cycle length of about 145 ms and showed approximately nine and a half revolutions. The turns were regular and stable. The rotor center trajectory is depicted in Figure 5. The trajectory was restricted to a small range of approximately 1.5 mm in x-direction and approximately 3 mm in y-direction. In z-direction no significant movement occurred because the endocardium was almost orthogonal to the z-axis.

Figure 4 Excerpt of the endocardium with phase map. The colors visualize the phase at each point. The white arrow points at the rotor center.
Figure 4

Excerpt of the endocardium with phase map. The colors visualize the phase at each point. The white arrow points at the rotor center.

Figure 5 Rotor center trajectory for stable rotor 1. The rotor centers for early points in time are depicted in blue. Late points in time are depicted in yellow. The last revolution is aditionally marked with a black line which connects the rotor centers of the single time steps. An excerpt of the curved endocardium is shown in gray.
Figure 5

Rotor center trajectory for stable rotor 1. The rotor centers for early points in time are depicted in blue. Late points in time are depicted in yellow. The last revolution is aditionally marked with a black line which connects the rotor centers of the single time steps. An excerpt of the curved endocardium is shown in gray.

Rotor 2 was located at the roof of the left atrium and was obverved from 0.6 s to 2.0 s. During that time, rotor 2 rotated with a cycle length of about 190 ms and showed approximately seven revolutions. The behavior of rotor 2 was more unstable than the behavior of rotor 1. The rotor center trajectory is depicted in Figure 6 for points in time between 1.27 s and 1.60 s. For early points in time the rotor center showed stationary characteristics. Later on it moved along an oval shaped curve across the endocardium, before it returned to its initial stationary state. For approximately 3 % of those time steps the algorithm did not yield a proper rotor center. The area covered by the overall rotor center trajectory of rotor 2 measured more than 10 mm in x-direction, approximately 4 mm in y-direction and approximately 3 mm in z-direction.

Figure 6 Rotor center trajectory for unstable rotor 2. The rotor centers for early points in time are depicted in blue. Late points in time are depicted in yellow. An excerpt of the curved endocardium is shown in gray.
Figure 6

Rotor center trajectory for unstable rotor 2. The rotor centers for early points in time are depicted in blue. Late points in time are depicted in yellow. An excerpt of the curved endocardium is shown in gray.

4 Discussion

Methods for the detection of rotor center filaments in ventricular and atrial tissue had already been published [1, 2, 5]. Likewise, the rotor centers and trajectories in 2D simulations were discussed in other work [10]. This work describes and evaluates a method which is capable of detecting PSs in the atrial cavity of a realistic heart model. The point of view changes from cardiac tissue to blood, which is closer to clinical applications. Furthermore we describe the trajectories of rotor tips at the curved endocardial surface.

We showed that our phase based approach is capable of detecting rotor centers in the atrial cavity. Nevertheless, the phase signals did not always fulfill the desired linear shape. Oscillating phase values near the phase jump occurred. The algorithm which searches for PSs will fail if fluctuations become to large. The same holds for ripples which overlaid the linear part of the phase signal. Especially the criterion of a monotone course of phase values along path B (see Figure 1) was threatened by ripples. As the extent of ripples and oscillations was dependent on the shape of the electrogram, this could lead to problems in clinical applications. Further investigations need to show, whether other methods can overcome those disadvantages of the phase calculation with the inverse tangent and the Hilbert transform.

Rotor centers showed different temporal trajectories for different rotors. The algorithm did not yield rotor centers for approximately 3 % of the investigated time steps for rotor 2 which partially explains the gap in the trajectory shown in figure 6. Additionally neighboring activation patterns caused the unstable behavior of the trajectory of rotor 2. The fact that the rotor center was not fixed in space for rotor 2 needs to be paid special attention. Keeping in mind that rotor centers are targets for the creation of lesions in electrophysiological studies, the measurements should not be restricted to a single point in time. Instead the rotor center should be followed for a period of time to be sure about its trajectory.

Author’s Statement

  1. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.

References

[1] Clayton RH, Zhuchkova EA, Panfilov AV. Phase singularities and filaments: Simplifying complexity in computational models of ventricular fibrillation. Progress in Biophysics and Molecular Biology 6; 90: 378–398.10.1016/j.pbiomolbio.2005.06.011Search in Google Scholar PubMed

[2] Fenton F, Karma A. Vortex dynamics in three-dimensional continuous myocardium with fiber rotation: Filament instability and fibrillation. Chaos 1998; 1: 20–27.10.1063/1.166311Search in Google Scholar PubMed

[3] Krueger MW. Simulation of Atrial Fibrillation. In: Personalized Multi-Scale Modeling of the Atria: Heterogeneities, Fiber Architecture, Hemodialysis and Ablation Therapy. PhD thesis. Karlsruhe: KIT Scientific Publishing 2012: 205–211.Search in Google Scholar

[4] Rottmann M, Keller MW, Oesterlein T, Seemann G, Doessel O. Comparison of Different Methods and Catheter Designs to Estimate the Rotor Tip Position – A Simulation Study. Computing in Cardiology 2014; 41: 133–136.Search in Google Scholar

[5] Krummen DE, Swarup V, Narayan SM. The role of rotors in atrial fibrillation. Journal of Thoracic Disease 2015; 7: 142–151.Search in Google Scholar

[6] Mewis C, Riessen R, Spyridopoulos I. Kardiologie compact: alles für Station und Facharztprüfung. 2nd ed. Stuttgart: Thieme Georg Verlag 2006.10.1055/b-002-23562Search in Google Scholar

[7] Narayan SM, Baykaner T, Clopton P, et al. Ablation of Rotor and Focal Sources Reduces Late Recurrence of Atrial Fibrillation Compared With Trigger Ablation Alone. Journal of the American College of Cardiology 2014; 63: 1761–1768.10.1016/j.jacc.2014.02.543Search in Google Scholar PubMed PubMed Central

[8] Umapathy K, Masse, Sevaptsidis E, Asta J, Krishnan S, Nanthakumar K. Spatiotemporal Frequency Analysis of Ventricular Fibrillation in Explanted Human Hearts. IEEE Transactions on Biomedical Engineering 2009; 56: 328–335.10.1109/TBME.2008.2006031Search in Google Scholar PubMed

[9] Umapathy K, Nair K, Masse S, et al. Phase mapping of cardiac fibrillation. Circulation 2010; 3: 716–722.10.1161/CIRCEP.110.853804Search in Google Scholar PubMed

[10] Wilhelms M, Hettmann H, Maleckar MM, Koivumäki, Dössel O, Seemann G. Benchmarking electrophysiological models of human atrial myocytes. Froniers in Physiology 2013; 3: 1–15.10.3389/fphys.2012.00487Search in Google Scholar PubMed PubMed Central

[11] Wyndham CR. Atrial fibrillation: The Most Common Arrhythmia. Texas Heart Institute Journal 2000; 27: 257–267.Search in Google Scholar

Published Online: 2015-9-12
Published in Print: 2015-9-1

© 2015 by Walter de Gruyter GmbH, Berlin/Boston

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Articles in the same Issue

  1. Research Article
  2. Development and characterization of superparamagnetic coatings
  3. Research Article
  4. The development of an experimental setup to measure acousto-electric interaction signal
  5. Research Article
  6. Stability analysis of ferrofluids
  7. Research Article
  8. Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
  9. Research Article
  10. Energy harvesting for active implants: powering a ruminal pH-monitoring system
  11. Research Article
  12. New type of fluxgate magnetometer for the heart’s magnetic fields detection
  13. Research Article
  14. Field mapping of ballistic pressure pulse sources
  15. Research Article
  16. Development of a new homecare sleep monitor using body sounds and motion tracking
  17. Research Article
  18. Noise properties of textile, capacitive EEG electrodes
  19. Research Article
  20. Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
  21. Research Article
  22. Spike sorting: the overlapping spikes challenge
  23. Research Article
  24. Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
  25. Research Article
  26. Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
  27. Research Article
  28. Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
  29. Research Article
  30. Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
  31. Research Article
  32. Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
  33. Research Article
  34. Compressed sensing of multi-lead ECG signals by compressive multiplexing
  35. Research Article
  36. Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
  37. Research Article
  38. Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
  39. Research Article
  40. Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
  41. Research Article
  42. Effects of sampling rate on automated fatigue recognition in surface EMG signals
  43. Research Article
  44. Closed-loop transcranial alternating current stimulation of slow oscillations
  45. Research Article
  46. Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
  47. Research Article
  48. The role of expert evaluation for microsleep detection
  49. Research Article
  50. The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
  51. Research Article
  52. Metal artifact reduction by projection replacements and non-local prior image integration
  53. Research Article
  54. A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
  55. Research Article
  56. Processing of membranes for oxygenation using the Bellhouse-effect
  57. Research Article
  58. Inkjet printing of viable human dental follicle stem cells
  59. Research Article
  60. The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
  61. Research Article
  62. New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
  63. Research Article
  64. QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
  65. Research Article
  66. Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
  67. Research Article
  68. Next-generation vision testing: the quick CSF
  69. Research Article
  70. Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
  71. Research Article
  72. Design and control of a 3-DOF hydraulic driven surgical instrument
  73. Research Article
  74. Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
  75. Research Article
  76. Frequency based assessment of surgical activities
  77. Research Article
  78. “Hands free for intervention”, a new approach for transoral endoscopic surgery
  79. Research Article
  80. Pseudo-haptic feedback in medical teleoperation
  81. Research Article
  82. Feasibility of interactive gesture control of a robotic microscope
  83. Research Article
  84. Towards structuring contextual information for workflow-driven surgical assistance functionalities
  85. Research Article
  86. Towards a framework for standardized semantic workflow modeling and management in the surgical domain
  87. Research Article
  88. Closed-loop approach for situation awareness of medical devices and operating room infrastructure
  89. Research Article
  90. Kinect based physiotherapy system for home use
  91. Research Article
  92. Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
  93. Research Article
  94. Integrating multimodal information for intraoperative assistance in neurosurgery
  95. Research Article
  96. Respiratory motion tracking using Microsoft’s Kinect v2 camera
  97. Research Article
  98. Using smart glasses for ultrasound diagnostics
  99. Research Article
  100. Measurement of needle susceptibility artifacts in magnetic resonance images
  101. Research Article
  102. Dimensionality reduction of medical image descriptors for multimodal image registration
  103. Research Article
  104. Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
  105. Research Article
  106. Evaluation of CT capability for the detection of thin bone structures
  107. Research Article
  108. Towards contactless optical coherence elastography with acoustic tissue excitation
  109. Research Article
  110. Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
  111. Research Article
  112. Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
  113. Research Article
  114. Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
  115. Research Article
  116. Video tracking of swimming rodents on a reflective water surface
  117. Research Article
  118. MR imaging of model drug distribution in simulated vitreous
  119. Research Article
  120. Studying the extracellular contribution to the double wave vector diffusion-weighted signal
  121. Research Article
  122. Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
  123. Research Article
  124. Introducing a frequency-tunable magnetic particle spectrometer
  125. Research Article
  126. Imaging of aortic valve dynamics in 4D OCT
  127. Research Article
  128. Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
  129. Research Article
  130. Simple concept for a wide-field lensless digital holographic microscope using a laser diode
  131. Research Article
  132. Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
  133. Research Article
  134. Respiratory surface motion measurement by Microsoft Kinect
  135. Research Article
  136. Improving image quality in EIT imaging by measurement of thorax excursion
  137. Research Article
  138. A clustering based dual model framework for EIT imaging: first experimental results
  139. Research Article
  140. Three-dimensional anisotropic regularization for limited angle tomography
  141. Research Article
  142. GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
  143. Research Article
  144. Experimental computer tomograph
  145. Research Article
  146. US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
  147. Research Article
  148. Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
  149. Research Article
  150. Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
  151. Research Article
  152. Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
  153. Research Article
  154. An IHE-conform telecooperation platform supporting the treatment of dementia patients
  155. Research Article
  156. Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
  157. Research Article
  158. Identification of surgical instruments using UHF-RFID technology
  159. Research Article
  160. A generic concept for the development of model-guided clinical decision support systems
  161. Research Article
  162. Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
  163. Research Article
  164. Creating 3D gelatin phantoms for experimental evaluation in biomedicine
  165. Research Article
  166. Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
  167. Research Article
  168. Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
  169. Research Article
  170. Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
  171. Research Article
  172. Wireless medical sensors – context, robustness and safety
  173. Research Article
  174. Sequences for real-time magnetic particle imaging
  175. Research Article
  176. Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
  177. Research Article
  178. A machine learning approach for planning valve-sparing aortic root reconstruction
  179. Research Article
  180. An in-ear pulse wave velocity measurement system using heart sounds as time reference
  181. Research Article
  182. Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
  183. Research Article
  184. Multisegmental fusion of the lumbar spine a curse or a blessing?
  185. Research Article
  186. Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
  187. Research Article
  188. A muscle model for hybrid muscle activation
  189. Research Article
  190. Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
  191. Research Article
  192. An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
  193. Research Article
  194. Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
  195. Research Article
  196. Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
  197. Research Article
  198. Mesh structure-independent modeling of patient-specific atrial fiber orientation
  199. Research Article
  200. Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
  201. Research Article
  202. Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
  203. Research Article
  204. A java based simulator with user interface to simulate ventilated patients
  205. Research Article
  206. Evaluation of an algorithm to choose between competing models of respiratory mechanics
  207. Research Article
  208. Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
  209. Research Article
  210. Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
  211. Research Article
  212. Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
  213. Research Article
  214. FPGA controlled artificial vascular system
  215. Research Article
  216. Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
  217. Research Article
  218. Test setup for characterizing the efficacy of embolic protection devices
  219. Research Article
  220. Impact of electrode geometry on force generation during functional electrical stimulation
  221. Research Article
  222. 3D-based visual physical activity assessment of children
  223. Research Article
  224. Realtime assessment of foot orientation by Accelerometers and Gyroscopes
  225. Research Article
  226. Image based reconstruction for cystoscopy
  227. Research Article
  228. Image guided surgery innovation with graduate students - a new lecture format
  229. Research Article
  230. Multichannel FES parameterization for controlling foot motion in paretic gait
  231. Research Article
  232. Smartphone supported upper limb prosthesis
  233. Research Article
  234. Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
  235. Research Article
  236. Evaluation of adhesion promoters for Parylene C on gold metallization
  237. Research Article
  238. The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
  239. Research Article
  240. Increasing the visibility of thin NITINOL vascular implants
  241. Research Article
  242. Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
  243. Research Article
  244. Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
  245. Research Article
  246. Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
  247. Research Article
  248. Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
  249. Research Article
  250. Electrospun vascular grafts with anti-kinking properties
  251. Research Article
  252. Integration of temperature sensors in polyimide-based thin-film electrode arrays
  253. Research Article
  254. Use cases and usability challenges for head-mounted displays in healthcare
  255. Research Article
  256. Device- and service profiles for integrated or systems based on open standards
  257. Research Article
  258. Risk management for medical devices in research projects
  259. Research Article
  260. Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
  261. Research Article
  262. Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
  263. Research Article
  264. SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
  265. Research Article
  266. Numerical simulations of airflow in the human pharynx of OSAHS patients
Downloaded on 10.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2015-0010/html
Scroll to top button