FPGA controlled artificial vascular system

  • D. Laqua EMAIL logo , C. Brieskorn , J. H. Koch , M. Rothmayer , S. Zeiske , M. Böttrich , S. Ley and P. Husar
Published/Copyright: September 12, 2015

Abstract

Monitoring the oxygen saturation of an unborn child is an invasive procedure, so far. Transabdominal fetal pulse oximetry is a promising method under research, used to estimate the oxygen saturation of a fetus noninvasively. Due to the nature of the method, the fetal information needs to be extracted from a mixed signal. To properly evaluate signal processing algorithms, a phantom modeling fetal and maternal blood circuits and tissue layers is necessary. This paper presents an improved hardware concept for an artificial vascular system, utilizing an FPGA based CompactRIO System from National Instruments. The experimental model to simulate the maternal and fetal blood pressure curve consists of two identical hydraulic circuits. Each of these circuits consists of a pre-pressure system and an artificial vascular system. Pulse curves are generated by proportional valves, separating these two systems. The dilation of the fetal and maternal artificial vessels in tissue substitutes is measured by transmissive and reflective photoplethysmography. The measurement results from the pressure sensors and the transmissive optical sensors are visualized to show the functionality of the pulse generating systems. The trigger frequency for the maternal valve was set to 1 per second, the fetal valve was actuated at 0.7 per second for validation. The reflective curve, capturing pulsations of the fetal and maternal circuit, was obtained with a high power LED (905 nm) as light source. The results show that the system generates pulse curves, similar to its physiological equivalent. Further, the acquired reflective optical signal is modulated by the alternating diameter of the tubes of both circuits, allowing for tests of signal processing algorithms.

1 Introduction

A lack of oxygen can lead to serious damage in brain development of the fetus before or during labor and delivery. There is still no procedure to determine the fetal oxygen saturation non-invasively. Invasive methods are a risk for mother and fetus and only will be acceptable in case of a strong suspicion of a risk for the fetus.

There is no reliable description in literature for the light modulation when passing the pulsating vessels of mother and fetus. This however is an essential premise for separating the pulse wave curve from mother and fetus.

Due to these uncertain conditions, a realistic phantom is needed to analyze the nature of transabdominal fetal pulse oximetry in detail. A phantoms allows to generate predefined and deterministic conditions and can be used to evaluate signal processing algorithms. Two separate artificial blood circuits are needed to simulate fetal and maternal pulse curves and further analyze photoplethysmo-graphic signals.

Complex phantoms with various interdependent actuators and sensors require a suitable, real-time controlling system. It is difficult to meet properties of real-time systems with computer-based solutions. Embedded systems or field-programmable gate arrays (FPGA) are an appropriate solution for such a task. The architecture has a high real-time performance for controlling peripheral components.

Oura et al. [1] and Parlikar et al. [2] developed an extremely compact mock circulation system, which can simulate artery blood circulation and generate a pulse wave with very small amounts of blood. The patent for a blood vessel model for medical training from Umezu et al. [3] includes an artificially fabricated tube as simulated vessel partly buried in a base. The patent of Holdsworth et al. [4] for a computer controlled positive displacement pump produces a physiological flow stimulation. In a previous investigation, we tried to generate pulsatile waveforms with a roller pump and a stepping motor [5]. The resulting waveforms contained an unacceptable oscillation, induced by the roller pump. An alternative approach is described by Laqua et al. [9]. A pre-pressure system in combination with controllable valves allows to generate the pulse curves without oscillation, close to the physiological equivalent.

The work of Jurovata et al. [6] describes the light transport, photon trajectory and the interaction in tissue. Therefore, a Monte Carlo simulation has been implemented in Matlab to evaluate the light distribution and the penetration depth without patients.

In order to examine the optical properties of tissue, Suzaki et al. [7] investigated the relation between optical density (defined as O.D.) of whole blood and hematocrit by transmission spectrophotometry during pulsating blood perfusion. The optical characteristics of artificial tissue were investigated by Ley et al. [8].

2 Material and methods

The block diagram as shown in Fig. 1 describes the improved hardware concept using a CompactRIO from National Instruments. The system includes a processor and an FPGA and can be customized for various tasks adding different hardware modules. In the case of the phantom, we use two different types of modules. Firstly, NI 9401 digital I/O-modules, which are connected to all the controllers symbolized by red lines and secondly, NI 9239 analog modules used for data acquisition (green lines).

However, lets take a closer look at the single components. For regulating the stepper motors two controllers (SMCI12 from Nanotec Electronic GmbH & Co KG) are used. Programming of the controllers is done via a RS-485 interface and allows to adjust the driving profiles of the motor. The CompactRIO-System triggers the driver by two digital signals as higher level controller.

The proportional valves are used to generate the pulse curves. They are controlled by PWM signals, generated by two control modules (Fig. 2). These modules are connected to the CompactRIO using an RS-232 interface.

Figure 1 Improved hardware concept using a CompactRIO from National Instruments including an embedded processor and an FPGA; components: (1) Reservoir, (2) Roller pump, (3) Pressure sensor, (4) Pressure tank, (5) Proportional valve, (6) Compliance, (7) Artifi-cial tissue with fetal (lower) and maternal (upper) circuit, (8) Resistance, (9) Catch tank
Figure 1

Improved hardware concept using a CompactRIO from National Instruments including an embedded processor and an FPGA; components: (1) Reservoir, (2) Roller pump, (3) Pressure sensor, (4) Pressure tank, (5) Proportional valve, (6) Compliance, (7) Artifi-cial tissue with fetal (lower) and maternal (upper) circuit, (8) Resistance, (9) Catch tank

Figure 2 Hydraulic control for the proportional valves to generate the pulse curves and the control for the stepping motor modules with the roller pumps; Data acquisition for the pressure sensors and the optical sensor elements
Figure 2

Hydraulic control for the proportional valves to generate the pulse curves and the control for the stepping motor modules with the roller pumps; Data acquisition for the pressure sensors and the optical sensor elements

The experimental model to simulate the maternal and fetal blood pressure curve consists of two identical circuits. Each of these circuits is split into a pre-pressure system and into an artificial vascular system. Two proportional valves control the flow and separate the pre-pressure system from the secondary sub-system. The nonlinear extension characteristic of the aorta, the so called Windkessel function, is reproduced by the compliance in the artificial vascular system. To simulate the total peripheral resistance of the Windkessel model a resistance element is used. For the determination of the outcome flow two different types of clamps are used for the maternal and fetal circuit. An explicit description of the two sub-systems and the elements of the systems can be found in a previous work [9]. The tissue phantom consists of transparent silicone tubes with an internal diameter of 5 mm and a wall thickness of 0.5 mm, which are molded in the artificial tissue made of silicone.

The measurement system in detail is described by Laqua et al. [9]. The dilation of the artificial fetal and maternal vessels in tissue substitute is measured by transmissive photoplethysmography. Two LEDs are used to irradiate the artificial vessels with light in the infrared range, respectively. A photo detector acquires the transmitted optical signal, modulated by the alternating tube. In addition, pressure sensors are integrated, for gathering the maternal and fetal pressure curves as a reference for the plethysmo-graphic measurements.

Figure 3 Artificial pulse waves for mother and fetus; the upper part of the diagram shows the maternal and fetal pressure measurements; In the lower part of the diagram the reflective and transmissive plethysmographic measurements are depicted
Figure 3

Artificial pulse waves for mother and fetus; the upper part of the diagram shows the maternal and fetal pressure measurements; In the lower part of the diagram the reflective and transmissive plethysmographic measurements are depicted

The signals from the pressure sensors of the fetal and maternal vascular system, as well as of the pre-pressure system is acquired by the analog input module NI 9239. Furthermore, a second module is used for photoplethysmographic data acquisition.

To reproduce the optical properties of the maternal and the fetal tissue, tissue substitutes have been developed. Therefore the different types of tissues, like skin, muscle or fat, are represented by one layer, so that there is one maternal and one fetal layer. The amniotic liquor is modeled by an extra layer. RTV silicone rubber is used as a solid carrier material, because its refractive index is similar to the refractive index of human soft tissue. Moreover the mechanical properties can be varied by changing the mixture of the RTV silicone rubber. Light scattering and light absorption reproduced by adding Titanium dioxide and cometic powder, mixed into the carrier material. The maternal and fetal tissues differ in the concentrations of the additives. The studies of the optical properties of the additives were conducted in a previous work. [10]

Figure 4 Artificial pulse waves with solely the fetal circuit active; the upper part of the diagram shows the pressure measurement of the fetal pulsatile curve; In the lower part of the diagram the reflective and transmissive plethysmographic measurements are depicted
Figure 4

Artificial pulse waves with solely the fetal circuit active; the upper part of the diagram shows the pressure measurement of the fetal pulsatile curve; In the lower part of the diagram the reflective and transmissive plethysmographic measurements are depicted

The individual layers, which represent the maternal and fetal tissues and the amniotic liquor, are poured into a wooden box, so that the tissue phantom is measuring 215 mm x 200 mm x 52 mm (L x W x H).

3 Results

Fig. 3 shows the measured simulated pulse waves for mother and fetus. The upper part of the diagram shows the maternal and fetal pressure measurements, which range from 5 to 45 kPa. The trigger frequency for the maternal valve was set to 1 per second, and the fetal valve to 0.7 per second. It can be seen that a peak is reached after a short while which decays over a longer time period. A dicrotic pulse is visible in both curves. The curve therefore resembles the human arterial blood pressure curve. In the lower part of the diagram the plethysmographic measurements are depicted. The data was superimposed by high frequency disturbances and was therefore filtered with a Chebyshev low-pass filter with a pass frequency of 5 Hz and stop frequency of 25 Hz before plotting. As expected, the transmissive measurement corresponds qualitatively to the fetal pulse curve, as it was measured in the same circuit. The peaks in both measurements occur at the same time.

The reflective curve, which was obtained with a high power LED as constant light source with a wavelength of 905 nm, is obviously more complex. It is modulated by the diameter changes of both artificial vessels. The curve rises and decays according to the pressure in both vascular systems. It is noticeable that the influence of the maternal pulses is higher than the fetal. This is due to the fact that the maternal tube is closer to the surface of the tissue substitute and consequently closer to the light source and the sensor respectively. Yet the possibility is given to decompose the reflective curve to gain the pulse wave curves of the single systems. To assess the influence of a single pressure curve a measurement with solely the fetal circuit active was conducted. This measurement can be found in Fig. 4. In the regions after the peak it can be seen that the reflective measurement is very sensitive to pressure changes.

4 Conclusion

In this contribution a phantom for producing plethysmo-graphic curves, which is controlled by an FPGA based CompactRIO System from National Instruments, has been developed. It contains two independent liquid closed loops, which reproduce the maternal and fetal pulse waves in tissue substitutes. The data acquisition and recording of the sensors is executed by the FPGA-system. The sensor system and the tubes were poured into tissue substitute made of silicone. Beside the optical sensor system, a capacitive sensor system has also been integrated. However, the connection to the cRIO-system is still outstanding, but will be realized as future work. As a future perspective an artificial blood substitute shall also be applied. At the moment this blood substitute is still under development. Furthermore, a new artificial tissue substitute phantom should be built, which reproduces the bending of the belly during pregnancy.


Author: D. Laqua: Technische Universität Ilmenau/Biosignal Processing Group, Ilmenau, Germany, +49 3677 69-2860

Author’s Statement

  1. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.

References

[1] Oura, M., Kobayashi, N., Takeda, S., Iwasaki, K., Umezu, M. Development of a compact mock circulation system and a new flow-cell model for pulse spectrophotometry. Engineering in Medicine and Biology Society, 30th Annual International Conference of the IEEE 2008; 670–673. http://doi.org/10.1109/IEMBS.2008.464924110.1109/IEMBS.2008.4649241Search in Google Scholar PubMed

[2] Parlikar, T. A. Modeling and Monitoring of Cardiovascular Dynamics for Patients in Critical Care. Department of Electrical Engineering and Computer Science, Cambride 2007; Retrieved from http://hdl.handle.net/1721.1/40859Search in Google Scholar

[3] Umezu, M., Park, Y. K. Blood vessel model for medical training and method for manufactoring same. Patent 2012Search in Google Scholar

[4] Holdsworth, D., Rickey, D., Drangova, M., Fenster, A. Computer controlled positive displacement pump for physiological flow simulation. Patent 199210.1007/BF02446086Search in Google Scholar PubMed

[5] Weyrich, S., Sprenger, S., Böttrich, M., Schmidt, P., Laqua, D., Ley, S., Husar, P. Development of a phantom to modulate the maternal and fetal pulse curve for pulse oximetry measurements. Biomedical Engineering 2012; 57 Suppl 1, 803–806. http://doi.org/10.1515/bmt-2012-420410.1515/bmt-2012-4204Search in Google Scholar PubMed

[6] Jurovata, D., Kurnatova, J., Ley, S., Laqua, D., Vazan, P., Husar, P. Simulation of photon propagation in tissue using MATLAB. International Doctoral Seminar 2013 Dubrovnik, Croatia 2013; http://doi.org/10.2478/rput-2013-000610.2478/rput-2013-0006Search in Google Scholar

[7] Suzaki, H., Takeda, S., Kobayashi, N., Kubota, H., Aomi, T., Nagaoka, T., Uchiyama, A. Relation between hematocrit and optical density in pulse oximetry -In vitro study with Waseda mock circulatory system-. Engineering in Medicine and Biology Society 2005; 2626–2629. http://doi.org/10.1109/IEMBS.2005.161700810.1109/IEMBS.2005.1617008Search in Google Scholar PubMed

[8] Ley, S., Laqua, D., Husar, P. Conception of a sensor circuit for non-invasive fetal pulse oximetry based on investigations of the optical transmission properties of biological tissues World Congress on Medical Physics and Biomedical Engineering, IFMBE Proceedings 39 Beijing 2012; 1431–1434. http://doi.org/10.1007/978-3-642-29305-410.1007/978-3-642-29305-4Search in Google Scholar

[9] Laqua, D., Pollnow, S., Fischer, J., Ley, S., Husar, P. A phantom with pulsating artificial vessels for non-invasive fetal pulse oximetry. 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Chicago, IL, USA 2014; 5631–5634. IEEE. http://doi.org/10.1109/EMBC.2014.694490410.1109/EMBC.2014.6944904Search in Google Scholar PubMed

[10] Ley, S., Stadthalter, M., Link, D., Laqua, D., Husar, P. Phantom materials mimicking the optical properties in the near infrared range for non-invasive fetal pulse oximetry. 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society . Chicago, IL, USA 2014; 1432–1435. IEEE. http://doi.org/10.1109/EMBC.2014.694386910.1109/EMBC.2014.6943869Search in Google Scholar PubMed

Published Online: 2015-9-12
Published in Print: 2015-9-1

© 2015 by Walter de Gruyter GmbH, Berlin/Boston

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Articles in the same Issue

  1. Research Article
  2. Development and characterization of superparamagnetic coatings
  3. Research Article
  4. The development of an experimental setup to measure acousto-electric interaction signal
  5. Research Article
  6. Stability analysis of ferrofluids
  7. Research Article
  8. Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
  9. Research Article
  10. Energy harvesting for active implants: powering a ruminal pH-monitoring system
  11. Research Article
  12. New type of fluxgate magnetometer for the heart’s magnetic fields detection
  13. Research Article
  14. Field mapping of ballistic pressure pulse sources
  15. Research Article
  16. Development of a new homecare sleep monitor using body sounds and motion tracking
  17. Research Article
  18. Noise properties of textile, capacitive EEG electrodes
  19. Research Article
  20. Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
  21. Research Article
  22. Spike sorting: the overlapping spikes challenge
  23. Research Article
  24. Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
  25. Research Article
  26. Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
  27. Research Article
  28. Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
  29. Research Article
  30. Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
  31. Research Article
  32. Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
  33. Research Article
  34. Compressed sensing of multi-lead ECG signals by compressive multiplexing
  35. Research Article
  36. Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
  37. Research Article
  38. Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
  39. Research Article
  40. Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
  41. Research Article
  42. Effects of sampling rate on automated fatigue recognition in surface EMG signals
  43. Research Article
  44. Closed-loop transcranial alternating current stimulation of slow oscillations
  45. Research Article
  46. Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
  47. Research Article
  48. The role of expert evaluation for microsleep detection
  49. Research Article
  50. The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
  51. Research Article
  52. Metal artifact reduction by projection replacements and non-local prior image integration
  53. Research Article
  54. A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
  55. Research Article
  56. Processing of membranes for oxygenation using the Bellhouse-effect
  57. Research Article
  58. Inkjet printing of viable human dental follicle stem cells
  59. Research Article
  60. The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
  61. Research Article
  62. New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
  63. Research Article
  64. QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
  65. Research Article
  66. Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
  67. Research Article
  68. Next-generation vision testing: the quick CSF
  69. Research Article
  70. Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
  71. Research Article
  72. Design and control of a 3-DOF hydraulic driven surgical instrument
  73. Research Article
  74. Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
  75. Research Article
  76. Frequency based assessment of surgical activities
  77. Research Article
  78. “Hands free for intervention”, a new approach for transoral endoscopic surgery
  79. Research Article
  80. Pseudo-haptic feedback in medical teleoperation
  81. Research Article
  82. Feasibility of interactive gesture control of a robotic microscope
  83. Research Article
  84. Towards structuring contextual information for workflow-driven surgical assistance functionalities
  85. Research Article
  86. Towards a framework for standardized semantic workflow modeling and management in the surgical domain
  87. Research Article
  88. Closed-loop approach for situation awareness of medical devices and operating room infrastructure
  89. Research Article
  90. Kinect based physiotherapy system for home use
  91. Research Article
  92. Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
  93. Research Article
  94. Integrating multimodal information for intraoperative assistance in neurosurgery
  95. Research Article
  96. Respiratory motion tracking using Microsoft’s Kinect v2 camera
  97. Research Article
  98. Using smart glasses for ultrasound diagnostics
  99. Research Article
  100. Measurement of needle susceptibility artifacts in magnetic resonance images
  101. Research Article
  102. Dimensionality reduction of medical image descriptors for multimodal image registration
  103. Research Article
  104. Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
  105. Research Article
  106. Evaluation of CT capability for the detection of thin bone structures
  107. Research Article
  108. Towards contactless optical coherence elastography with acoustic tissue excitation
  109. Research Article
  110. Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
  111. Research Article
  112. Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
  113. Research Article
  114. Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
  115. Research Article
  116. Video tracking of swimming rodents on a reflective water surface
  117. Research Article
  118. MR imaging of model drug distribution in simulated vitreous
  119. Research Article
  120. Studying the extracellular contribution to the double wave vector diffusion-weighted signal
  121. Research Article
  122. Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
  123. Research Article
  124. Introducing a frequency-tunable magnetic particle spectrometer
  125. Research Article
  126. Imaging of aortic valve dynamics in 4D OCT
  127. Research Article
  128. Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
  129. Research Article
  130. Simple concept for a wide-field lensless digital holographic microscope using a laser diode
  131. Research Article
  132. Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
  133. Research Article
  134. Respiratory surface motion measurement by Microsoft Kinect
  135. Research Article
  136. Improving image quality in EIT imaging by measurement of thorax excursion
  137. Research Article
  138. A clustering based dual model framework for EIT imaging: first experimental results
  139. Research Article
  140. Three-dimensional anisotropic regularization for limited angle tomography
  141. Research Article
  142. GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
  143. Research Article
  144. Experimental computer tomograph
  145. Research Article
  146. US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
  147. Research Article
  148. Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
  149. Research Article
  150. Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
  151. Research Article
  152. Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
  153. Research Article
  154. An IHE-conform telecooperation platform supporting the treatment of dementia patients
  155. Research Article
  156. Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
  157. Research Article
  158. Identification of surgical instruments using UHF-RFID technology
  159. Research Article
  160. A generic concept for the development of model-guided clinical decision support systems
  161. Research Article
  162. Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
  163. Research Article
  164. Creating 3D gelatin phantoms for experimental evaluation in biomedicine
  165. Research Article
  166. Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
  167. Research Article
  168. Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
  169. Research Article
  170. Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
  171. Research Article
  172. Wireless medical sensors – context, robustness and safety
  173. Research Article
  174. Sequences for real-time magnetic particle imaging
  175. Research Article
  176. Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
  177. Research Article
  178. A machine learning approach for planning valve-sparing aortic root reconstruction
  179. Research Article
  180. An in-ear pulse wave velocity measurement system using heart sounds as time reference
  181. Research Article
  182. Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
  183. Research Article
  184. Multisegmental fusion of the lumbar spine a curse or a blessing?
  185. Research Article
  186. Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
  187. Research Article
  188. A muscle model for hybrid muscle activation
  189. Research Article
  190. Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
  191. Research Article
  192. An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
  193. Research Article
  194. Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
  195. Research Article
  196. Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
  197. Research Article
  198. Mesh structure-independent modeling of patient-specific atrial fiber orientation
  199. Research Article
  200. Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
  201. Research Article
  202. Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
  203. Research Article
  204. A java based simulator with user interface to simulate ventilated patients
  205. Research Article
  206. Evaluation of an algorithm to choose between competing models of respiratory mechanics
  207. Research Article
  208. Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
  209. Research Article
  210. Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
  211. Research Article
  212. Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
  213. Research Article
  214. FPGA controlled artificial vascular system
  215. Research Article
  216. Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
  217. Research Article
  218. Test setup for characterizing the efficacy of embolic protection devices
  219. Research Article
  220. Impact of electrode geometry on force generation during functional electrical stimulation
  221. Research Article
  222. 3D-based visual physical activity assessment of children
  223. Research Article
  224. Realtime assessment of foot orientation by Accelerometers and Gyroscopes
  225. Research Article
  226. Image based reconstruction for cystoscopy
  227. Research Article
  228. Image guided surgery innovation with graduate students - a new lecture format
  229. Research Article
  230. Multichannel FES parameterization for controlling foot motion in paretic gait
  231. Research Article
  232. Smartphone supported upper limb prosthesis
  233. Research Article
  234. Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
  235. Research Article
  236. Evaluation of adhesion promoters for Parylene C on gold metallization
  237. Research Article
  238. The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
  239. Research Article
  240. Increasing the visibility of thin NITINOL vascular implants
  241. Research Article
  242. Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
  243. Research Article
  244. Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
  245. Research Article
  246. Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
  247. Research Article
  248. Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
  249. Research Article
  250. Electrospun vascular grafts with anti-kinking properties
  251. Research Article
  252. Integration of temperature sensors in polyimide-based thin-film electrode arrays
  253. Research Article
  254. Use cases and usability challenges for head-mounted displays in healthcare
  255. Research Article
  256. Device- and service profiles for integrated or systems based on open standards
  257. Research Article
  258. Risk management for medical devices in research projects
  259. Research Article
  260. Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
  261. Research Article
  262. Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
  263. Research Article
  264. SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
  265. Research Article
  266. Numerical simulations of airflow in the human pharynx of OSAHS patients
Downloaded on 25.12.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2015-0107/html
Scroll to top button