Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
-
A. Almohallami
, B.-A. Behrens
, A. Bouguecha , M. Lerch , S. Betancur , C. Stukenborg-Colsman , I. Nolte and K. Lucas
Abstract
Aseptic loosening of the prosthesis is still a problem in artificial joint implants. The ýloosening can be caused by the resorption of the bone surrounding ýthe prosthesis according to stress shielding. A numerical model was developed and validated by means of DEXA-studies in order to ýanalyse the bone remodelling process in the periprosthetic bone. A total loss of about 3.7% of the bone density in the periprosthetic Femur with NANOS is computed. The bone remodelling calculation was validated by means of a DEXA-study with a 3-years-follow-up. The model was further developed in order to be able to calculate and consider the migration of the implants. This method was applied on the ýNANOS-implant with a computed total migration of about 0.43 mm. These calculations showed good results in comparison with a 2-year-follow-up clinical study, whereby a RSA-method was used to determine the stem migration in the bone. In order to ýstudy the mutual influence between the implant migration and the hip contact forces ý, a software is developed by our scientific group to couple a multi body simulation (MBS) of human lower limps with the FEA of the periprosthetic Femur.
1 Introduction
The implantation of artificial joints has been developed over the last decades into a routine procedure. More than 150,000 primary hip replacements are performed each year in Germany alone [1]. Despite all the technological and medical advances, the lifetime of the implants is still limited [2]. The most common reasons for implant failure are the aseptic loosening and the migration due to stress shielding [3], which leads to unstable anchoring of the prosthesis in the bone tissue [4]. In addition and due to the changes in the boundary conditions of the replaced hip joint the gait of the patient and thus the load on the hip joint varies [5]. The optimization of the prosthetic design has largely been based on the practical experience of manufacturers [8]. Extending this knowledge and experience with numerical methods can realize a time-saving, cost-effective and patient-friendly development methodology [7]. The finite element method (FEM) has been widely used in this field [6]. The aim of this study is to calculate the bone remodelling around the NANOS prosthesis in the femur and to numerically map the migration of this implant using the numerical model developed at IFUM. Furthermore, the FEA for mapping the NANOS shaft migration was coupled with a MBS of human lower limbs, which is supplied with a hip prosthesis to determine the mutual influence between the shaft migration and the hip contact forces.
2 Methods
Based on the Wolff’s law [8], different mathematical theories have been proposed with respect to bone remodelling. One of them is the model of Behrens et al., which is developed at IFUM in Hanover [9]. According to this model, the changes of the bone density depend on a stimulus ξ, which results from the time-varying strain energy density in the bone tissue. A detailed derivation of the stimulus ξ can be found in [10]. The model presented has already been used and validated for the numerical description of bone remodelling in the periprosthetic femur with different types of implants such as METHA®, Bicontact®prostheses [11]. By means of this model, the bone remodelling was numerically investigated around the NANOS prosthesis in the femur. The FE-model of the periprosthetic femur was created using HyperMesh® software (Altair, USA) with the NANOS prosthesis virtually implanted in a femur. Based on a CT data set of human patients, the femur geometry was created. In this case, an image analysis program (Yadiv®, Welfenlab, Hanover) was used to generate the 3D mesh model of the femur. Based on the HU values of CT images the calculated bone density is automatically assigned to the voxels of the finite element model. The FE model was implemented into the simulation system of MSC.Marc® (MSC.Software, Santa Ana, California, United States) to simulate the bone remodelling. The elastic modulus Eof the bone is calculated by an integrated subroutine in the simulation using the following equation 1 [12]:
The NANOS-prosthesis was modelled with a titanium alloy Ti6Al4V, which is characterized by a homogeneous and isotropic material behaviour. The physiological boundary conditions were assumed according to Speris et al. [13]. Furthermore, a reduced muscular system is added to the femur model suggested by Heller et al. [14]. The muscular system consists of the abductors, the tensor fascia lata, the vastus medialis and the vastus lateralis muscles. The resulting hip contact force was taken from a validated multi-body-simulation carried out at the IFUM in cooperation with the clinic of small animals at the University of Veterinary Medicine Hannover, Foundation [ý15]. The multi-body-model is generated using gait analysis data at the speed of 0.8 m/s.
To calculate the density of the bone in different regions of interest (ROIs), the elements were stored in individual sets, corresponding to the distribution of the ROIs (see Figure 1). The contact between the NANOS-prosthesis and the bone tissue was modelled with a no-slip condition at the coated areas of the prosthesis. In uncoated areas, a frictionless contact is assumed.
So far, the bone remodelling in the periprosthetic bone tissues is calculated without changing the bone geometry. Thus, an aim of this study is to model the prosthetic migration by means of updating the bone geometry in contact with the prosthetic as a function of the density values of the bone tissue. However, no study or equation connects the dismantling of the tissue around the implants with the reduction of bone density. Based on the equation 1, the changes in the bone density can be translated into changes in its E modulus and thus into changes in the elastic strains ε of the bone tissue as a reaction to the hip contact forces. The change in the strain can be calculated from the simulation with an appropriate subroutine, which stores the changes in the element positions and considers these changes to create new sets of elements using a remeshing function before starting the new simulation increment. Using this method it is now possible to model the changes of bone geometries numerically considering only the mechanical factors that lead to the prosthetic migration and ignoring biological influences.

The NANOS-Prosthesis and the distribution of the ROIs in the Femur.
The influence of the prosthetic migration on the hip contact forces has already been investigated and confirmed at the IFUM using the MBS model of the human lower extremities [15]. Thus, it is very important to model the implant migration considering the changes in the hip contact forces caused by the migration itself. For this purpose, online coupling software is created between the multibody simulation software LifeMOD, in which the hip contact forces will be determined, and the FE-software MSC.Marc, where the prosthesis migration can be calculated. The coupling is realized through a software written in the programming language Python. The coupling ends when the FE-calculation of bone remodelling converges and the change in position of the prosthesis is too small.
3 Results
3.1 The bone remodelling
The numerical calculation of the bone remodelling needed 48 increments to reach the state of convergence. The loss of bone density in the entire femur is about 3.7 % (see Figure 2). The FE analysis shows that the bone density in R7 is significantly reduced, where the loss is about 42.6%. Figure 3 shows the changes in the calculated bone density and the following Table 1 summarizes the calculated changes in the bone density in the individual ROIs according to the finite element analysis.

Calculation of the bone remodelling convergences in 48 increments.

Bone remodelling a) at the initial state and b) in the final state with the largest bone density loss area highlighted in red.
Calculated changes in bone density in percent.
| ROIs | 1 | 2 | 3 | 4 | 5 | 6 | 7 |
|---|---|---|---|---|---|---|---|
| Density [%] | 12.8 | 12.2 | 2.64 | −0.04 | 0.08 | 1.18 | 42.6 |
3.2 Implant migration
The numerical simulation of migration of the NANOS stem revealed that the remeshing function affects the density of the elastic strain energy in the elements of the FE model. This leads to a delayed convergence of the calculations, since the calculation of bone remodelling is based on the change in the strain energy density. The calculation converges after 52 increments. Here, the bone density of the femur decreases by 3.82%. The simulation results show a caudal (distal) shaft migration of 0.366 mm, a lateral migration of 0.17 mm and a posterior migration of 0.02 mm. The FE-calculations of the NANOS migration, taking into account the changes in hip contact forces induced by this migration, shows differences in the migration patterns to the ones calculated without the coupling with the MB-simulation. Under these circumstances, the calculated implant migration in distal direction is 0.39 mm, while it is 0.2 mm in lateral direction and 0.04 in posterior direction. Figure 4 shows the overall NANOS-migration with and without coupling with the MBS.

The progression of the NANOS-migration in the Femur bone tissue.
4 Discussion
In order to validate the bone remodelling calculations, the numerical results of the changes in the bone density were compared to those measured by the 3-Years follow-up DEXA investigations of Zeh et al. [16] on patients with NANOS hip prosthesis. The following Table 2 summarizes the results of this validation.
Calculated and measured changes in bone density in percent.
| ROIs | 1 | 2 | 3 | 4 | 5 | 6 | 7 |
|---|---|---|---|---|---|---|---|
| Density-FEM [%] | 12.8 | 12.2 | 2.64 | −0.04 | 0.08 | 1.18 | 42.6 |
| Density-DEXA [%] | 15.3 | 5.03 | 1.77 | 1.41 | 0.47 | −12.0 | 11.8 |
| Deviation | 2.5 | 7.17 | 0.87 | 1.45 | 0.39 | 13.18 | 30.8 |
| Avg. | 8.05 | ||||||
From Table 2 it can be seen that the numerical calculation shows a slight deviation from the results of the DEXA study in the regions R1, R2, R3, R4 and R5. In this case, the difference between the measured and the calculated BMD is about 2.5% in average. This increases to about 13 % in R6. Remarkably the deviation is about 30% in R7. According to the FE-calculations, the hip contact stress is induced through the prosthesis in R7 area, which leads to a high stress shielding effect and thus to bone resorption. The reason of the high deviation could be the changes in the R7 scanned area during the DEXA-investigations, which is detected by Lerch et al. with his DEXA-Study [11]. This decrease in the R7 area leads to misread the BMD value of R7 and makes it higher than it really is. However, no information about the scanned area of R7 are provided in the paper of Zeh et al. The average deviation in all regions is about 8 %, which confirms the validation of the numerical simulation of the periprosthetic femur bone remodelling with the NANOS-prosthesis. To validate the numerical investigation of shaft migration, the study of Budde et al. [17] were used. In this clinical study, the X-ray stereo photogrammetric analysis (RSA) was used to determine the NANOS-shaft migration in the femur within the first 2 years postoperative. In this study a total migration of 0.49 +/- 0.31 mm has been detected, while the numerical investigation has determined a total migration of 0.4 mm without coupling with MBS and 0.43 with coupling with MBS. This comparison confirms the validity of this numerical modelling of the NANOS-shaft migration in the femur bone tissue and emphasizes the importance of the mutual influence between the implant migration and the hip contact forces.
Acknowledgment
The study was carried out in the subproject D6 of the Collaborative Research Center 599 “Sustainable degradable and permanent implants out of metallic and ceramic materials”.
Author’s Statement
Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.
References
[1] Bundesauswertung zum Erfassungsjahr 2013.Search in Google Scholar
[2] Nakahara I, Nakamura N, Takao, M, Sakai T, Nishii T, Sugano N (2011): Eight-year wear analysis in longevity highly crosslinked polyethylene liners comparing 26- and 32-mm heads, Archives of Orthopaedic and Trauma Surgery, 131, S. 1731-1737.10.1007/s00402-011-1360-zSearch in Google Scholar PubMed
[3] Kuiper J, Huiskes R (1997): The predictive value of stress shielding for quantification of adaptive bone resorption around hip replacements,Journal of Biomechanical Engineering, 119, S. 228-231.10.1115/1.2796084Search in Google Scholar PubMed
[4] Register TSNHA, Annual Report 2004.Search in Google Scholar
[5] Escobar S, Bouguecha A, Nolte I, Stukenborg-Colsman C, Behrens B.-A.(2011): Influence of Prosthesis Migration in Human Hip Joint Using Multi-Body Simulation, 21st International Congress of Mechanical Engineering October 24-28, 2011, Natal, RN, Brazil.Search in Google Scholar
[6] Grifka, J, Kuster M (2011): Orthopädie und Unfallchirurgie. Springer Berlin Heidelberg. ISBN 13 978-3-642-13110-3.10.1007/978-3-642-13111-0Search in Google Scholar
[7] Stukenborg-Colsman C, Von der Haar-Tran A, Windhagen H, Bouguecha A, Wefstaedt P, Lerch M. Bone remodelling around a cementless straight THA stem: A prospective dual-energy Xray absorptiometry study. HIP International, Volume 22, Issue 2, March 2012, Pages 166-17110.5301/HIP.2012.9227Search in Google Scholar PubMed
[8] Wollf J (1892): Das Gesetz der Transformation der Knochen Hirschwald, BerlinSearch in Google Scholar
[9] Behrens B.-A, Nolte I, Wefstaedt P, Stukenborg-Colsman C, Bouguecha A (2009): Numerical investigations on the strain-adaptive bone remodelling in the periprosthetic femur: Influence of the boundary conditions. BioMedical Engineering OnLine. Bd. 8.10.1186/1475-925X-8-7Search in Google Scholar PubMed PubMed Central
[10] Bouguecha A, Weiglel N, Behrens B.-A, Stukenborg-Colsman C, Waizy H. Numerical simulation of strain-adaptive bone remodelling in the ankle joint, BioMedical Engineering OnLine 2011, 10:5810.1186/1475-925X-10-58Search in Google Scholar PubMed PubMed Central
[11] Lerch M, Kurtz A, Stukenborg-Colsman C, Nolte I, Weigel N, Bouguecha A, Behrens B.-A. Bone Remodeling after Total Hip Arthroplasty with a Short Stemmed Metaphyseal Loading Implant: Finite Element Analysis Validated by a Prospective DEXA Investigation. J Orthop Res. 2012 Nov; 30(11):1822-9.10.1002/jor.22120Search in Google Scholar PubMed
[12] Carter D, Hayas W (1977): The compressive behaviour of bone as a two-phase porous structure; In: J. Bone & Joint Surgery Vol.59A – No.7, S. 54-962.10.2106/00004623-197759070-00021Search in Google Scholar
[13] Speirs A, Heller M, Duda G, Taylor W (2007): Physiologically based boundary conditions in finite element modelling; In: J.Biomechanics Vol. 40 – No. 10, S. 2318-2323.10.1016/j.jbiomech.2006.10.038Search in Google Scholar PubMed
[14] Heller M, Bergmann G, Kassi J, Claes L, Haas N, Duda G (2005): Determination of muscle loading at the hip joint for use in pre-clinical testing; In: J. Biomechanics Vol. 38 – No.5, S. 1155-1163.10.1016/j.jbiomech.2004.05.022Search in Google Scholar PubMed
[15] Bouguecha A, Elgaly I, Stukenborg-Colsman C, Lerch M, Nolte I, Wefstaedt P, Behrens B.-A (2010): Numerical Investigations of the Strain-Adaptive Bone Remodeling in the Prosthetic Pelvis. IFMBE Proceedings. 2010, Bd. 29, S. 562-565.10.1007/978-3-642-13039-7_141Search in Google Scholar
[16] Zeh A, Pankow F, Röllinhoff M, Delank S, Wohlrab D, A prospective dual-energy X-ray absorptiometry study of bone remodeling after implantation of the Nanos®short-stemmed prosthesis,Acta Orthop. Belg., 2013, 79, 174-180.Search in Google Scholar
[17] Budde S, Thorey F, Schwarze M, Von Lewinski G, Hurschler C, Flörkermeier T, Windhagen H, Seehaus F. RöntgenStereometrie-Analyse (RSA) der Migration der NANOSKurzschaftprothese: 2-Jahres-Ergebnisse, Deutscher Kongress für Orthopädie und Unfallchirurgie, 10.2014, Berlin.Search in Google Scholar
© 2015 by Walter de Gruyter GmbH, Berlin/Boston
This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients
Articles in the same Issue
- Research Article
- Development and characterization of superparamagnetic coatings
- Research Article
- The development of an experimental setup to measure acousto-electric interaction signal
- Research Article
- Stability analysis of ferrofluids
- Research Article
- Investigation of endothelial growth using a sensors-integrated microfluidic system to simulate physiological barriers
- Research Article
- Energy harvesting for active implants: powering a ruminal pH-monitoring system
- Research Article
- New type of fluxgate magnetometer for the heart’s magnetic fields detection
- Research Article
- Field mapping of ballistic pressure pulse sources
- Research Article
- Development of a new homecare sleep monitor using body sounds and motion tracking
- Research Article
- Noise properties of textile, capacitive EEG electrodes
- Research Article
- Detecting phase singularities and rotor center trajectories based on the Hilbert transform of intraatrial electrograms in an atrial voxel model
- Research Article
- Spike sorting: the overlapping spikes challenge
- Research Article
- Separating the effect of respiration from the heart rate variability for cases of constant harmonic breathing
- Research Article
- Locating regions of arrhythmogenic substrate by analyzing the duration of triggered atrial activities
- Research Article
- Combining different ECG derived respiration tracking methods to create an optimal reconstruction of the breathing pattern
- Research Article
- Atrial and ventricular signal averaging electrocardiography in pacemaker and cardiac resynchronization therapy
- Research Article
- Estimation of a respiratory signal from a single-lead ECG using the 4th order central moments
- Research Article
- Compressed sensing of multi-lead ECG signals by compressive multiplexing
- Research Article
- Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry
- Research Article
- Synchronization in wireless biomedical-sensor networks with Bluetooth Low Energy
- Research Article
- Automated classification of stages of anaesthesia by populations of evolutionary optimized fuzzy rules
- Research Article
- Effects of sampling rate on automated fatigue recognition in surface EMG signals
- Research Article
- Closed-loop transcranial alternating current stimulation of slow oscillations
- Research Article
- Cardiac index in atrio- and interventricular delay optimized cardiac resynchronization therapy and cardiac contractility modulation
- Research Article
- The role of expert evaluation for microsleep detection
- Research Article
- The impact of baseline wander removal techniques on the ST segment in simulated ischemic 12-lead ECGs
- Research Article
- Metal artifact reduction by projection replacements and non-local prior image integration
- Research Article
- A novel coaxial nozzle for in-process adjustment of electrospun scaffolds’ fiber diameter
- Research Article
- Processing of membranes for oxygenation using the Bellhouse-effect
- Research Article
- Inkjet printing of viable human dental follicle stem cells
- Research Article
- The use of an icebindingprotein out of the snowflea Hypogastrura harveyi as a cryoprotectant in the cryopreservation of mesenchymal stem cells
- Research Article
- New NIR spectroscopy based method to determine ischemia in vivo in liver – a first study on rats
- Research Article
- QRS and QT ventricular conduction times and permanent pacemaker therapy after transcatheter aortic valve implantation
- Research Article
- Adopting oculopressure tonometry as a transient in vivo rabbit glaucoma model
- Research Article
- Next-generation vision testing: the quick CSF
- Research Article
- Improving tactile sensation in laparoscopic surgery by overcoming size restrictions
- Research Article
- Design and control of a 3-DOF hydraulic driven surgical instrument
- Research Article
- Evaluation of endourological tools to improve the diagnosis and therapy of ureteral tumors – from model development to clinical application
- Research Article
- Frequency based assessment of surgical activities
- Research Article
- “Hands free for intervention”, a new approach for transoral endoscopic surgery
- Research Article
- Pseudo-haptic feedback in medical teleoperation
- Research Article
- Feasibility of interactive gesture control of a robotic microscope
- Research Article
- Towards structuring contextual information for workflow-driven surgical assistance functionalities
- Research Article
- Towards a framework for standardized semantic workflow modeling and management in the surgical domain
- Research Article
- Closed-loop approach for situation awareness of medical devices and operating room infrastructure
- Research Article
- Kinect based physiotherapy system for home use
- Research Article
- Evaluating the microsoft kinect skeleton joint tracking as a tool for home-based physiotherapy
- Research Article
- Integrating multimodal information for intraoperative assistance in neurosurgery
- Research Article
- Respiratory motion tracking using Microsoft’s Kinect v2 camera
- Research Article
- Using smart glasses for ultrasound diagnostics
- Research Article
- Measurement of needle susceptibility artifacts in magnetic resonance images
- Research Article
- Dimensionality reduction of medical image descriptors for multimodal image registration
- Research Article
- Experimental evaluation of different weighting schemes in magnetic particle imaging reconstruction
- Research Article
- Evaluation of CT capability for the detection of thin bone structures
- Research Article
- Towards contactless optical coherence elastography with acoustic tissue excitation
- Research Article
- Development and implementation of algorithms for automatic and robust measurement of the 2D:4D digit ratio using image data
- Research Article
- Automated high-throughput analysis of B cell spreading on immobilized antibodies with whole slide imaging
- Research Article
- Tissue segmentation from head MRI: a ground truth validation for feature-enhanced tracking
- Research Article
- Video tracking of swimming rodents on a reflective water surface
- Research Article
- MR imaging of model drug distribution in simulated vitreous
- Research Article
- Studying the extracellular contribution to the double wave vector diffusion-weighted signal
- Research Article
- Artifacts in field free line magnetic particle imaging in the presence of inhomogeneous and nonlinear magnetic fields
- Research Article
- Introducing a frequency-tunable magnetic particle spectrometer
- Research Article
- Imaging of aortic valve dynamics in 4D OCT
- Research Article
- Intravascular optical coherence tomography (OCT) as an additional tool for the assessment of stent structures
- Research Article
- Simple concept for a wide-field lensless digital holographic microscope using a laser diode
- Research Article
- Intraoperative identification of somato-sensory brain areas using optical imaging and standard RGB camera equipment – a feasibility study
- Research Article
- Respiratory surface motion measurement by Microsoft Kinect
- Research Article
- Improving image quality in EIT imaging by measurement of thorax excursion
- Research Article
- A clustering based dual model framework for EIT imaging: first experimental results
- Research Article
- Three-dimensional anisotropic regularization for limited angle tomography
- Research Article
- GPU-based real-time generation of large ultrasound volumes from freehand 3D sweeps
- Research Article
- Experimental computer tomograph
- Research Article
- US-tracked steered FUS in a respiratory ex vivo ovine liver phantom
- Research Article
- Contribution of brownian rotation and particle assembly polarisation to the particle response in magnetic particle spectrometry
- Research Article
- Preliminary investigations of magnetic modulated nanoparticles for microwave breast cancer detection
- Research Article
- Construction of a device for magnetic separation of superparamagnetic iron oxide nanoparticles
- Research Article
- An IHE-conform telecooperation platform supporting the treatment of dementia patients
- Research Article
- Automated respiratory therapy system based on the ARDSNet protocol with systemic perfusion control
- Research Article
- Identification of surgical instruments using UHF-RFID technology
- Research Article
- A generic concept for the development of model-guided clinical decision support systems
- Research Article
- Evaluation of local alterations in femoral bone mineral density measured via quantitative CT
- Research Article
- Creating 3D gelatin phantoms for experimental evaluation in biomedicine
- Research Article
- Influence of short-term fixation with mixed formalin or ethanol solution on the mechanical properties of human cortical bone
- Research Article
- Analysis of the release kinetics of surface-bound proteins via laser-induced fluorescence
- Research Article
- Tomographic particle image velocimetry of a water-jet for low volume harvesting of fat tissue for regenerative medicine
- Research Article
- Wireless medical sensors – context, robustness and safety
- Research Article
- Sequences for real-time magnetic particle imaging
- Research Article
- Speckle-based off-axis holographic detection for non-contact photoacoustic tomography
- Research Article
- A machine learning approach for planning valve-sparing aortic root reconstruction
- Research Article
- An in-ear pulse wave velocity measurement system using heart sounds as time reference
- Research Article
- Measuring different oxygenation levels in a blood perfusion model simulating the human head using NIRS
- Research Article
- Multisegmental fusion of the lumbar spine a curse or a blessing?
- Research Article
- Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration
- Research Article
- A muscle model for hybrid muscle activation
- Research Article
- Mathematical, numerical and in-vitro investigation of cooling performance of an intra-carotid catheter for selective brain hypothermia
- Research Article
- An ideally parameterized unscented Kalman filter for the inverse problem of electrocardiography
- Research Article
- Interactive visualization of cardiac anatomy and atrial excitation for medical diagnosis and research
- Research Article
- Virtualizing clinical cases of atrial flutter in a fast marching simulation including conduction velocity and ablation scars
- Research Article
- Mesh structure-independent modeling of patient-specific atrial fiber orientation
- Research Article
- Accelerating mono-domain cardiac electrophysiology simulations using OpenCL
- Research Article
- Understanding the cellular mode of action of vernakalant using a computational model: answers and new questions
- Research Article
- A java based simulator with user interface to simulate ventilated patients
- Research Article
- Evaluation of an algorithm to choose between competing models of respiratory mechanics
- Research Article
- Numerical simulation of low-pulsation gerotor pumps for use in the pharmaceutical industry and in biomedicine
- Research Article
- Numerical and experimental flow analysis in centifluidic systems for rapid allergy screening tests
- Research Article
- Biomechanical parameter determination of scaffold-free cartilage constructs (SFCCs) with the hyperelastic material models Yeoh, Ogden and Demiray
- Research Article
- FPGA controlled artificial vascular system
- Research Article
- Simulation based investigation of source-detector configurations for non-invasive fetal pulse oximetry
- Research Article
- Test setup for characterizing the efficacy of embolic protection devices
- Research Article
- Impact of electrode geometry on force generation during functional electrical stimulation
- Research Article
- 3D-based visual physical activity assessment of children
- Research Article
- Realtime assessment of foot orientation by Accelerometers and Gyroscopes
- Research Article
- Image based reconstruction for cystoscopy
- Research Article
- Image guided surgery innovation with graduate students - a new lecture format
- Research Article
- Multichannel FES parameterization for controlling foot motion in paretic gait
- Research Article
- Smartphone supported upper limb prosthesis
- Research Article
- Use of quantitative tremor evaluation to enhance target selection during deep brain stimulation surgery for essential tremor
- Research Article
- Evaluation of adhesion promoters for Parylene C on gold metallization
- Research Article
- The influence of metallic ions from CoCr28Mo6 on the osteogenic differentiation and cytokine release of human osteoblasts
- Research Article
- Increasing the visibility of thin NITINOL vascular implants
- Research Article
- Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems
- Research Article
- Development of a bending test procedure for the characterization of flexible ECoG electrode arrays
- Research Article
- Tubular manipulators: a new concept for intracochlear positioning of an auditory prosthesis
- Research Article
- Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading
- Research Article
- Electrospun vascular grafts with anti-kinking properties
- Research Article
- Integration of temperature sensors in polyimide-based thin-film electrode arrays
- Research Article
- Use cases and usability challenges for head-mounted displays in healthcare
- Research Article
- Device- and service profiles for integrated or systems based on open standards
- Research Article
- Risk management for medical devices in research projects
- Research Article
- Simulation of varying femoral attachment sites of medial patellofemoral ligament using a musculoskeletal multi-body model
- Research Article
- Does enhancing consciousness for strategic planning processes support the effectiveness of problem-based learning concepts in biomedical education?
- Research Article
- SPIO processing in macrophages for MPI: The breast cancer MPI-SNLB-concept
- Research Article
- Numerical simulations of airflow in the human pharynx of OSAHS patients