Startseite The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
Artikel Open Access

The effect of thermal treatment on the mechanical properties of PLLA tubular specimens

  • Daniela Arbeiter EMAIL logo , K. Schümann , O. Sahmel , T. Eickner , K.-P. Schmitz und N. Grabow
Veröffentlicht/Copyright: 30. September 2016
Veröffentlichen auch Sie bei De Gruyter Brill

Abstract

Conventional permanent stent systems for vascular applications are associated with long-term risks, such as restenosis and thrombosis. To overcome these limitations, novel approaches using various biodegradable materials for stent construction have been investigated. In this context, thermal treatment of polymer materials is investigated to adjust the mechanical properties of biodegradable stents. In this work polymeric tubular specimens of biodegradable poly(L-lactide) (PLLA) were extruded and subjected to a molding process using different temperatures above glass transition temperature TG. Physicochemical properties of the molded samples were analyzed using DSC measurements and uniaxial tensile tests. The molding process resulted in a weakening of the PLLA tubular specimens with a simultaneous increase in the degree of crystallinity (χ).

1 Introduction

Stents made of biodegradable polymeric materials are currently being investigated as a future alternative to permanent metallic stents. Insufficient mechanical properties of biodegradable polymer stents are a major challenge, as they may lead to mechanical failure or high elastic recoil after implantation into the vessel [1]. Glass transition TG and crystalline microstructure of the applied polymeric materials influence the mechanical properties of the stents. Polymers with TG higher than human body temperature may show little or no plastic deformation prior to failure, although they potentially offer high stiffness. In opposition, polymers with TG below human body temperature could exhibit high plastic deformation associated with low stiffness. The crystallinity and crystalline microstructure of polymers, and thus the mechanical properties, can be altered by subjecting the polymers to deformation processes [2], which was the focus of the present study.

PLLA mini tubes were extruded and expanded with a blow molding process using different temperatures above TG. This temperature was necessary to get the samples more flexible during expansion. The morphology of the specimens was investigated, the degree of crystallinity χ and TG were determined, and the mechanical properties were subsequently assessed at 37°C in uniaxial tensile tests.

2 Material and methods

2.1 Sample preparation

Stent tubes were manufactured from PLLA (Resomer L210, Evonik Industries, Germany, TG = 60°C, TM = 178°C) with a molecular weight of ∼400.000 g/mol.

2.1.1 Extrusion

PLLA tube samples (nominal diameter d = 1.7 mm, wall thickness 300 μm) were extruded using a MiniLab II HAAKE (Thermo Fisher Scientific, Karlsruhe, Germany). For extrusion a temperature of T = 215°C and a pressure of p = 3.5 bar were adjusted.

2.1.2 Blow molding

Blow molding was performed by means of an air current with a preset pressure of p = 6 bar in a tubular mold to a nominal diameter of d = 2.3 mm.

During blow molding, heating of the PLLA tubes above glass transition temperature TG was performed to improve flexibility of the PLLA molecules. Different temperatures above TG (67/77/87°C) were realized (PLLA 67/77/87). Holding time was t = 30 s. Subsequent to the blow molding process the expanded tubes were quenched to a temperature of T = 25°C.

2.2 Thermal characterization

Melting behaviour of the blow molded samples was measured using a DSC 1 system (Mettler Toledo, Switzerland). During the scans, the specimens were exposed to nitrogen gas. The sample weights were in the range of 2–5 mg, and the heating rate was 10 K/min. The data were analyzed with respect to glass transition (TG), melting temperature (TM) and degree of crystallinity (χ) (see eq. 1). For the latter a value of ΔH100 = 93.7 J/g [3] for totally crystalline PLLA was used.

(1)χS=ΔHSΔH100100

2.3 Mechanical characterization

Uniaxial tensile testing was performed with a Zwicki ZN 2.5 (Zwick, Ulm, Germany). Tests were conducted with a 100 N load cell and a crosshead speed of 50 mm/min. During the procedure samples were kept at a constant temperature of 37°C. The tensile force as a function of elongation was measured. Based on these data the elastic modulus (E) was calculated in the range of 0.05–0.25%. Furthermore, the elongation at break (εB) was determined.

3 Results

3.1 Sample treatment

PLLA tubular specimens were succesfully expanded with all three temperature settings. Figure 1 illustrates the extruded PLLA samples before (top) and after (middle) blow molding. The expanded tubes showed a clear enlargement in radial direction (Figure 1, bottom).

Figure 1: PLLA tubes manufactured with extrusion process (top) and after expansion with blow molding (middle) at T = 77°C (PLLA 77). Bottom: enlarged image of the samples.
Figure 1:

PLLA tubes manufactured with extrusion process (top) and after expansion with blow molding (middle) at T = 77°C (PLLA 77). Bottom: enlarged image of the samples.

No geometric imperfections or other detrimental defects on the PLLA specimens could be observed.

3.2 Effect of blow molding on physicochemical properties of PLLA tubes

3.2.1 Thermal characterization

DSC measurements were conducted with extruded specimens without molding treatment (PLLA original) in comparison to the blow molded samples PLLA 67, PLLA 77, and PLLA 87 (Figure 2). From the extruded PLLA tubes a degree of crystallinity of χ = 18.2 ± 1.0% was determined. Glass transition temperature TG of PLLA original was TG = 60.0 ± 0.3°C. In all samples cold crystallization was observed (data not shown). The crystallinity of PLLA 67 and PLLA 77 did not differ from χ of the PLLA sample without treatment.

Figure 2: Degree of crystallinity and glass transition temperature of extruded PLLA tubes and subsequently blow molded specimens as measured with DSC.
Figure 2:

Degree of crystallinity and glass transition temperature of extruded PLLA tubes and subsequently blow molded specimens as measured with DSC.

However, for PLLA 87 a χ = 32.4 ± 3.0% was observed, an almost two-fold increase compared to PLLA original. No change in TG was detected in any sample.

3.2.2 Mechanical characterization

From the uniaxial tensile tests stress strain curves were derived as shown in Figure 3. Without thermal treatment using blow molding, an elongation at break of εB = 20 ± 3% for PLLA could be determined. With increase of the heating temperature of blow mold specimens the value of εB increased.

Figure 3: Stress strain curves of blow molded PLLA samples compared with original extruded PLLA samples measured with uniaxial tensile test.
Figure 3:

Stress strain curves of blow molded PLLA samples compared with original extruded PLLA samples measured with uniaxial tensile test.

PLLA 67 reached a value of εB = 56 ± 5% and the value of εB of PLLA 87 increased to εB = 120 ± 11%. However, a slight tendency to advanced values in the elastic modulus could be seen. With increasing of the blow mold temperature the modulus increases (Table 1).

Table 1:

Physicochemical and mechanical properties of the tested PLLA specimens.

Modulus (MPa)εB (%)TG (°C)χ (%)
PLLA original2166 ± 15820 ± 360.2 ± 0.318.2 ± 1.0
PLLA 672271 ± 16856 ± 562.5 ± 0.117.1 ± 0.4
PLLA 772494 ± 4679 ± 862.0 ± 0.117.7 ± 0.3
PLLA 872591 ± 72120 ± 1162.3 ± 0.332 ± 3

4 Discussion

PLLA is a semicrystalline polymer consisting of amorphous and crystalline regions. Renouf-Glauser et al. showed that εB of the amorphous part is much higher than in the crystalline phase [4]. Hence, the amorphous areas are responsible for flexible behavior. In the study we expanded tubular specimens above TG. Thus, the amorphous regions of the samples were in the liquid phase, and the tubes could be molded without the formation of cracks.

The increase of elongation at break εB may be caused by the orientation of the molecular chains during the expansion process [2]. This is confirmed by the DSC measurements where no change in crystallinity occurred for any of the blow molded samples (Table 1).

Nevertheless, the modulus increased, indicating an increased stiffness of the specimens. The thermal properties also confirmed this interpretation, as the crystallinity is higher along with elevated heating temperatures. Here, annealing of the samples predominates, so crystal growth of the PLLA molecular chains occurred (Table 1).

With the extrusion and blow molding process cold crystallization could not be eliminated. Manufacturing of PLLA tubes with reproducible mechanical properties remains challenging.

Acknowledgement

Financial support by the Federal Ministry of Education and Research (BMBF) within RESPONSE “Partnership for Innovation in Implant Technology” is gratefully acknowledged.

Author’s Statement

Research funding: Financial support by the Federal Ministry of Education and Research (BMBF) within RESPONSE “Partnership for Innovation in Implant Technology”. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animal use.

References

[1] Soares JS, Moore, JE. Biomechanical challenges to polymeric biodegradable stents. Ann Biomed Eng. 2016;44:560–79.10.1007/s10439-015-1477-2Suche in Google Scholar PubMed

[2] Kumaraswamy G, Kornfield JA, Yeh F, Hsiao BS. Shear-enhanced crystallization in isotactic polypropylene. 3. Evidence for a kinetic pathway to nucleation. Macromolecules. 2002;35:1762–9.10.1021/ma0114180Suche in Google Scholar

[3] Fischer EW, Sterzel HJ, Wegner G. Investigation of the structure of solution grown crystals of lactide copolymers by means of chemical reactions. Kolloid.-Z und Z Polymer. 1973;251:980–90.10.1007/978-3-642-47050-9_24Suche in Google Scholar

[4] Renouf-Glauser AC, Rose J, Farrar DF, Cameron RE. The effect of crystallinity on the deformation mechanism and bulk mechanical properties of PLLA. Biomaterials. 2005;26:5771–82.10.1016/j.biomaterials.2005.03.002Suche in Google Scholar PubMed

Published Online: 2016-9-30
Published in Print: 2016-9-1

©2016 Daniela Arbeiter et al., licensee De Gruyter.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.

Artikel in diesem Heft

  1. Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
  2. Novel blood protein based scaffolds for cardiovascular tissue engineering
  3. Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
  4. Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
  5. Long-term stable modifications of silicone elastomer for improved hemocompatibility
  6. The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
  7. Biocompatible wear-resistant thick ceramic coating
  8. Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
  9. Examination of dielectric strength of thin Parylene C films under various conditions
  10. Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
  11. Systemic analysis about residual chloroform in PLLA films
  12. A macrophage model of osseointegration
  13. Towards in silico prognosis using big data
  14. Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
  15. Usability evaluation of a locomotor therapy device considering different strategies
  16. Hypoxia-on-a-chip
  17. Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
  18. Fabrication of MEMS-based 3D-μECoG-MEAs
  19. High speed digital interfacing for a neural data acquisition system
  20. Bionic forceps for the handling of sensitive tissue
  21. Experimental studies on 3D printing of barium titanate ceramics for medical applications
  22. Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
  23. 3D printing – a key technology for tailored biomedical cell culture lab ware
  24. 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
  25. Biocompatibility of photopolymers for additive manufacturing
  26. Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
  27. Novel wireless measurement system of pressure dedicated to in vivo studies
  28. Portable auricular device for real-time swallow and chew detection
  29. Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
  30. Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
  31. Deep brain stimulation: increasing efficiency by alternative waveforms
  32. Prediction of immediately occurring microsleep events from brain electric signals
  33. Determining cardiac vagal threshold from short term heart rate complexity
  34. Classification of cardiac excitation patterns during atrial fibrillation
  35. An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
  36. Deriving respiration from high resolution 12-channel-ECG during cycling exercise
  37. Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
  38. Automatic detection and mapping of double potentials in intracardiac electrograms
  39. Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
  40. Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
  41. Best practice: surgeon driven application in pelvic operations
  42. Vasomotor assessment by camera-based photoplethysmography
  43. Classification of morphologic changes in photoplethysmographic waveforms
  44. Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
  45. Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
  46. Nonlinear causal influences assessed by mutual compression entropy
  47. Comparative study of methods for solving the correspondence problem in EMD applications
  48. fNIRS for future use in auditory diagnostics
  49. Semi-automated detection of fractional shortening in zebrafish embryo heart videos
  50. Blood pressure measurement on the cheek
  51. Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
  52. Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
  53. An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
  54. Real-time QRS detection using integrated variance for ECG gated cardiac MRI
  55. Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
  56. In-vitro experiments to characterize ventricular electromechanics
  57. Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
  58. Application of microwave sensor technology in cardiovascular disease for plaque detection
  59. Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
  60. Detection of microsleep events in a car driving simulation study using electrocardiographic features
  61. A method to determine the kink resistance of stents and stent delivery systems according to international standards
  62. Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
  63. Transient Euler-Lagrange/DEM simulation of stent thrombosis
  64. Automated control of the laser welding process of heart valve scaffolds
  65. Automation of a test bench for accessing the bendability of electrospun vascular grafts
  66. Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
  67. Cryopreservation of cells using defined serum-free cryoprotective agents
  68. New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
  69. Determination of the membrane hydraulic permeability of MSCs
  70. Climate retainment in carbon dioxide incubators
  71. Multiple factors influencing OR ventilation system effectiveness
  72. Evaluation of an app-based stress protocol
  73. Medication process in Styrian hospitals
  74. Control tower to surgical theater
  75. Development of a skull phantom for the assessment of implant X-ray visibility
  76. Surgical navigation with QR codes
  77. Investigation of the pressure gradient of embolic protection devices
  78. Computer assistance in femoral derotation osteotomy: a bottom-up approach
  79. Automatic depth scanning system for 3D infrared thermography
  80. A service for monitoring the quality of intraoperative cone beam CT images
  81. Resectoscope with an easy to use twist mechanism for improved handling
  82. In vitro simulation of distribution processes following intramuscular injection
  83. Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
  84. A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
  85. Smart control for functional electrical stimulation with optimal pulse intensity
  86. Tactile display on the remaining hand for unilateral hand amputees
  87. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
  88. An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
  89. Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
  90. Learning discriminative classification models for grading anal intraepithelial neoplasia
  91. Regularization of EIT reconstruction based on multi-scales wavelet transforms
  92. Assessing MRI susceptibility artefact through an indicator of image distortion
  93. EyeGuidance – a computer controlled system to guide eye movements
  94. A framework for feedback-based segmentation of 3D image stacks
  95. Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
  96. 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
  97. Inside-Out access strategy using new trans-vascular catheter approach
  98. US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
  99. Impact of different registration methods in MEG source analysis
  100. 3D segmentation of thyroid ultrasound images using active contours
  101. Designing a compact MRI motion phantom
  102. Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
  103. Classification of indirect immunofluorescence images using thresholded local binary count features
  104. Analysis of muscle fatigue conditions using time-frequency images and GLCM features
  105. Numerical evaluation of image parameters of ETR-1
  106. Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
  107. Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
  108. Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
  109. Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
  110. Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
  111. Evaluation of open-source software for the lung segmentation
  112. Automatic determination of lung features of CF patients in CT scans
  113. Image analysis of self-organized multicellular patterns
  114. Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
  115. Radiopacity assessment of neurovascular implants
  116. Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
  117. Development of an artifact-free aneurysm clip
  118. Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
  119. Rapid prototyping of replica knee implants for in vitro testing
  120. Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
  121. Advanced wettability analysis of implant surfaces
  122. Patient-specific hip prostheses designed by surgeons
  123. Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
  124. Wear of a total intervertebral disc prosthesis
  125. Digital health and digital biomarkers – enabling value chains on health data
  126. Usability in the lifecycle of medical software development
  127. Influence of different test gases in a non-destructive 100% quality control system for medical devices
  128. Device development guided by user satisfaction survey on auricular vagus nerve stimulation
  129. Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
  130. Effect of left atrial hypertrophy on P-wave morphology in a computational model
  131. Simulation of intracardiac electrograms around acute ablation lesions
  132. Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
  133. Assessment of nasal resistance using computational fluid dynamics
  134. Resistance in a non-linear autoregressive model of pulmonary mechanics
  135. Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
  136. Determination of regional lung function in cystic fibrosis using electrical impedance tomography
  137. Development of parietal bone surrogates for parietal graft lift training
  138. Numerical simulation of mechanically stimulated bone remodelling
  139. Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
  140. Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
  141. Principle study on the signal connection at transabdominal fetal pulse oximetry
  142. Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
  143. Evaluating different approaches to identify a three parameter gas exchange model
  144. Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
  145. From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
  146. Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
  147. Improved FPGA controlled artificial vascular system for plethysmographic measurements
  148. Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
  149. Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
  150. Control scheme selection in human-machine- interfaces by analysis of activity signals
  151. Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
  152. Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
  153. Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
  154. Interactive monitoring system for visual respiratory biofeedback
  155. Development of a low-cost senor based aid for visually impaired people
  156. Patient assistive system for the shoulder joint
  157. A passive beating heart setup for interventional cardiology training
Heruntergeladen am 8.9.2025 von https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2016-0009/html
Button zum nach oben scrollen