Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers

  • Manuel Dethloff EMAIL logo , Marc Dangers , Boris Wilmes and Hermann Seitz
Published/Copyright: September 30, 2016

Abstract

For the development of a novel, user-friendly and low cost point-of-care diagnostic device for the detection of disease specific biomarkers a flow optimised design of the test system is investigated. For this, computational fluid dynamics and rapid prototyping methods are used. The result is a functional design, which can be used for further experimental investigations.

1 Introduction

Rapid test systems for the diagnosis of current diseases are attracting more and more attention in medical technologies. Particularly in vitro diagnostic devices play an important role in the examination of patient samples (blood, plasma, urine) for different disease specific biomarkers [1], [2].

A novel user-friendly and easy to handle diagnostic product is developed by combining the benefits of existing concepts. The basic element of the device is a line-immunoassay (LIA) with a membrane for the solid phase based on the enzyme-linked-immunosorbent-assay-technology (ELISA). The ELISA system is state-of-the-art and used for biochemical detection of several antibodies and antigens. Another speciality of the device is the transfer of the LIA-membrane into a flow optimised test case, which should move the actual execution of the test from the lab to point-of-care (POC). The resulting test system is characterised by a reduced execution period, a reduction of execution steps and an integrated waste management. A wide range of biomarkers (e.g. the cytomegalovirus) can be implemented in the test system.

In this study, the development of the flow optimised design of the diagnostic product is presented. Computational fluid dynamics (CFD) is used. CFD is a branch of fluid mechanics that uses numerical methods and algorithms to solve and analyse problems that involve fluid flows [3]. CFD studies are very suitable for the implementation of parameter variations. It is possible to change several options or vari-ables during preprocessing, and the simulations themselves can be calculated in parallel on a computing cluster. This saves time and money in comparison to experimental analyses. But although CFD is a powerful tool for the execution of flow calculations, it is necessary to validate the results with experimental analyses. Based on the CFD results, first prototypes are printed using an additive manufacturing process.

2 Material and methods

The path towards the development of a flow optimised design of the diagnostic product is based on an iterative process. First, the constructive implementation is done with the CAD tool Creo 2.0 (PTC, Needham, USA). Subsequently, the CAD design is exported to the CFD tool ANSYS CFX 16.1 (Canonsburg, PA, USA). The basis of most CFD calculations is a spatial discretization [3]. The volume occupied by the fluid is divided into discrete cells. The uniform mesh (approximately 1 million grid points) must have a sufficient resolution to get plausible results. Afterwards, the physical modelling and the boundary conditions are defined, e.g. initial conditions, inlet, outlet, etc. After preprocessing, the simulation is started on a computing cluster and the Navier-Stokes equations, which describe the motion of the viscous two-phase fluid, are solved iteratively in a transient way [4]. Then, a postprocessor is used for the analysis and visualization of the resulting solution. With the help of the visualized results, further modifications are made with respect to the geometry of the diagnostic product. Thus, the iterative process starts again until an analysed partial solution meets the requirements.

With the calculated and optimised design, first prototypes are built with an additive manufacturing process. For this case, a Poly Jet 3D-printing technique is used. (Objet Eden, Stratasys, Eden Praire, MN, USA). With this 3D-printing technique it is possible to produce components quickly and cost-efficiently.

3 Requirements for the POC device

The diagnostic product should be very user-friendly. This is achieved by using a rectangular shaped and easy to handle body. A membrane basin and a waste reservoir for the reagents waste is located within the body. A membrane strip with dried up reagents is positioned within the membrane basin (see Figure 1). The dimensions of the membrane strip are 25 × 3 mm.

Figure 1 First draft of the diagnostic product.
Figure 1

First draft of the diagnostic product.

The test principle involves successive reaction steps, where several solution steps and subsequent wash steps are necessary. After every single step a lateral movement (5°–10°, for mixing) and a subsequent tilting of the body by about 90° follow in order to carry the liquid fluid from the membrane basin into the waste reservoir. In this case, no rest fluid should remain in the membrane basin after tilting, since this would lead to erroneous results if new liquid were added. Another challenge is the ability to switch between large and small volumes during the test execution. The membrane basin must be designed in such a way, that it is big enough for large fluid volumes of about 5 ml (incubation of the patient sample), but also in such a way that small volumes of about 0.5 ml (solution and wash steps) ensure a sufficiently high liquid column over the membrane in order to ensure complete wetting. Furthermore, a mixing of waste and re-agents fluid in the membrane basin must be avoided. After finishing the test execution a total of 9 ml of liquid is delivered. So the waste reservoir must be designed for such a liquid volume. The inflow of the liquids takes place through an opening in the cover of the body. The cover is transparent in order to allow for optical evaluation of the membrane strip. A first draft of the diagnostic product can be seen in Figure 1.

4 Results

Based on the draft in Figure 1 and the dimensions of the membrane strip, the initial shape of the membrane basin is developed. During the iterative development process several shapes of the membrane basin are designed and improved. Some shapes of the membrane basin can be seen in Figure 2.

Figure 2 Different shapes of the membrane basin.
Figure 2

Different shapes of the membrane basin.

Shape 1 in Figure 2 shows a membrane basin geometry in an early stage of development. Substantial disadvantages can be seen. To realize an initial filling of 5 ml, the height of the rectangular-shaped basin is 25 mm. This kind of shape leads to a restricted optical view of the membrane strip as well as an increase of the total height of the body. Furthermore, flow simulations show that it is not possible to empty the basin at an angle of 90°. A rest fluid remains in the basin and leads to erroneous test results (see Figure 3).

Figure 3 CFD simulation of shape 1 (in Figure 2).
Figure 3

CFD simulation of shape 1 (in Figure 2).

As part of the further development, the design of the basin is changed with respect to the shape and size. Shape 2 in Figure 2 shows a wider and flatter form of the basin. The membrane strip is situated in the centred channel, which is designed for a filling volume of about 1 ml. An inclined wall makes the tilting of the fluid easier. The disadvantages of this shape are explained with the help of the CFD simulation presented in Figure 4. A small dead volume remains in the basin near the vertical walls after tilting the basin by 90°.

Figure 4 CFD simulation of shape 2 (in Figure 2).
Figure 4

CFD simulation of shape 2 (in Figure 2).

Further investigations show that the fluid column over the planar areas lateral to the membrane channel has no influence on the membrane. The height of the fluid column is essential for the biochemical processes and therefore for the effectiveness of the membrane. So in a next step, new geometrical solutions are designed, c.f. shapes 3 and 4 in Figure 2. Here, the membrane basins have sloped side walls as well as no vertical walls in the direction of tilting, so no dead volumes can occur.

Another important fact is the size and the arrangement of the waste reservoir in the body. The reservoir should be designed for a total waste of about 10 ml. In the first stage of development, the reservoir is located under the membrane basin at the bottom of the body (see Figure 5 or Figure 3). The membrane basin (shape 4 in Figure 2) is centred on the front face. The simulation study in Figure 5 shows the effect of tilting the basin forwards and then backwards again into the initial position. The membrane basin is filled with 5 ml of water in the initial position (see Figure 5A). In this study only the gravitation vector is changed. Figure 5B demonstrates the maximal tilting position of 90°. The fluid flows out of the basin. In Figure 5C and d the body is tilted backwards and the fluid flows back into the waste reservoir. Due to the inertia the fluid sloshes against the opposite wall (see Figure 5E).

Figure 5 CFD simulation study: Tilting forwards and backwards; waste reservoir under membrane basin.
Figure 5

CFD simulation study: Tilting forwards and backwards; waste reservoir under membrane basin.

If the rotating velocity of the tilting is higher than 1.5 m/s, fluid can splash back into the membrane basin. This particular case leads to incorrect results and has to be avoided. In the further course of development, the geometry of the body is changed so that the waste reservoir is located next to the membrane basin, as can be seen in Figure 6.

Figure 6 CAD model: new arrangement of basin and reservoir.
Figure 6

CAD model: new arrangement of basin and reservoir.

The width of the basin extends up to the side walls of the body. With this modification, the height of the body can be reduced by about 40% up to 25 mm. To ensure that no fluid splashes back into the basin during the tilting phase, the faces of the waste reservoir are equipped with absorbent material. This material absorbs the predominate part of the fluid. Figure 7 shows a 3D-printed prototype of the CAD model presented in Figure 6. It was printed using the Poly Jet technique. A membrane strip is located in the membrane basin (shape 3 in Figure 2). The waste reservoir next to the membrane basin is equipped with fleece material.

Figure 7 3D printed prototype of diagnostic product without cover.
Figure 7

3D printed prototype of diagnostic product without cover.

5 Conclusion

For the development of a new diagnostic product, a flow optimised design was successfully engineered. With the help of CFD, a novel, simple and user-friendly test body was designed which satisfies all requirements. It was demonstrated that for an optimised shape of the membrane basin and a suitable arrangement of the basin and waste reservoir a reliable test implementation is guaranteed. The dimensions of the body are 60 × 26 × 27 mm and thus make it easy to handle. In a next step, a cover with a click closure should be designed and further experimental investigations will be conducted. Once the biochemistry for relevant biomarkers is developed, the diagnostic product can then be tested for marketability.

Author’s Statement

Research funding: This project is funded by the Bundesministerium für Wirtschaft und Energie initiative “ZIM– Zentrales Innovationsprogramm Mittelstand”. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use complies with all the relevant national regulations, institutional policies and was performed in accordance with the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.

References

[1] Grevers G, Röcken M. Taschenatlas Allergologie – Grundlagen, Diagnostik, Klinik. Stuttgart: Thieme Verlag KG; 2008.10.1055/b-001-2118Search in Google Scholar

[2] Drescher P, Dangers M, Rübenhagen R, Seitz H. The effects of various flow velocities on the sensivity of an ELISA in a fluidic allergy diagnostic device. Point of Care: The Journal of Near-Patient Testing and Technology. 2014;2:35–40.10.1097/POC.0000000000000016Search in Google Scholar

[3] Ferziger J, Peric M. Computational methods for fluid dynamics. 3rd ed. Berlin: Springer; 2002.10.1007/978-3-642-56026-2Search in Google Scholar

[4] Oertel jr H, Böhle M, Reviol T. Strömungsmechanik. 6. Auflage. Wiesbaden: Vieweg+Teubner; 201110.1007/978-3-8348-8110-6Search in Google Scholar

Published Online: 2016-9-30
Published in Print: 2016-9-1

©2016 Manuel Dethloff et al., licensee De Gruyter.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.

Articles in the same Issue

  1. Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
  2. Novel blood protein based scaffolds for cardiovascular tissue engineering
  3. Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
  4. Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
  5. Long-term stable modifications of silicone elastomer for improved hemocompatibility
  6. The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
  7. Biocompatible wear-resistant thick ceramic coating
  8. Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
  9. Examination of dielectric strength of thin Parylene C films under various conditions
  10. Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
  11. Systemic analysis about residual chloroform in PLLA films
  12. A macrophage model of osseointegration
  13. Towards in silico prognosis using big data
  14. Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
  15. Usability evaluation of a locomotor therapy device considering different strategies
  16. Hypoxia-on-a-chip
  17. Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
  18. Fabrication of MEMS-based 3D-μECoG-MEAs
  19. High speed digital interfacing for a neural data acquisition system
  20. Bionic forceps for the handling of sensitive tissue
  21. Experimental studies on 3D printing of barium titanate ceramics for medical applications
  22. Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
  23. 3D printing – a key technology for tailored biomedical cell culture lab ware
  24. 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
  25. Biocompatibility of photopolymers for additive manufacturing
  26. Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
  27. Novel wireless measurement system of pressure dedicated to in vivo studies
  28. Portable auricular device for real-time swallow and chew detection
  29. Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
  30. Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
  31. Deep brain stimulation: increasing efficiency by alternative waveforms
  32. Prediction of immediately occurring microsleep events from brain electric signals
  33. Determining cardiac vagal threshold from short term heart rate complexity
  34. Classification of cardiac excitation patterns during atrial fibrillation
  35. An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
  36. Deriving respiration from high resolution 12-channel-ECG during cycling exercise
  37. Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
  38. Automatic detection and mapping of double potentials in intracardiac electrograms
  39. Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
  40. Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
  41. Best practice: surgeon driven application in pelvic operations
  42. Vasomotor assessment by camera-based photoplethysmography
  43. Classification of morphologic changes in photoplethysmographic waveforms
  44. Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
  45. Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
  46. Nonlinear causal influences assessed by mutual compression entropy
  47. Comparative study of methods for solving the correspondence problem in EMD applications
  48. fNIRS for future use in auditory diagnostics
  49. Semi-automated detection of fractional shortening in zebrafish embryo heart videos
  50. Blood pressure measurement on the cheek
  51. Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
  52. Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
  53. An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
  54. Real-time QRS detection using integrated variance for ECG gated cardiac MRI
  55. Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
  56. In-vitro experiments to characterize ventricular electromechanics
  57. Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
  58. Application of microwave sensor technology in cardiovascular disease for plaque detection
  59. Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
  60. Detection of microsleep events in a car driving simulation study using electrocardiographic features
  61. A method to determine the kink resistance of stents and stent delivery systems according to international standards
  62. Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
  63. Transient Euler-Lagrange/DEM simulation of stent thrombosis
  64. Automated control of the laser welding process of heart valve scaffolds
  65. Automation of a test bench for accessing the bendability of electrospun vascular grafts
  66. Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
  67. Cryopreservation of cells using defined serum-free cryoprotective agents
  68. New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
  69. Determination of the membrane hydraulic permeability of MSCs
  70. Climate retainment in carbon dioxide incubators
  71. Multiple factors influencing OR ventilation system effectiveness
  72. Evaluation of an app-based stress protocol
  73. Medication process in Styrian hospitals
  74. Control tower to surgical theater
  75. Development of a skull phantom for the assessment of implant X-ray visibility
  76. Surgical navigation with QR codes
  77. Investigation of the pressure gradient of embolic protection devices
  78. Computer assistance in femoral derotation osteotomy: a bottom-up approach
  79. Automatic depth scanning system for 3D infrared thermography
  80. A service for monitoring the quality of intraoperative cone beam CT images
  81. Resectoscope with an easy to use twist mechanism for improved handling
  82. In vitro simulation of distribution processes following intramuscular injection
  83. Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
  84. A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
  85. Smart control for functional electrical stimulation with optimal pulse intensity
  86. Tactile display on the remaining hand for unilateral hand amputees
  87. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
  88. An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
  89. Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
  90. Learning discriminative classification models for grading anal intraepithelial neoplasia
  91. Regularization of EIT reconstruction based on multi-scales wavelet transforms
  92. Assessing MRI susceptibility artefact through an indicator of image distortion
  93. EyeGuidance – a computer controlled system to guide eye movements
  94. A framework for feedback-based segmentation of 3D image stacks
  95. Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
  96. 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
  97. Inside-Out access strategy using new trans-vascular catheter approach
  98. US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
  99. Impact of different registration methods in MEG source analysis
  100. 3D segmentation of thyroid ultrasound images using active contours
  101. Designing a compact MRI motion phantom
  102. Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
  103. Classification of indirect immunofluorescence images using thresholded local binary count features
  104. Analysis of muscle fatigue conditions using time-frequency images and GLCM features
  105. Numerical evaluation of image parameters of ETR-1
  106. Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
  107. Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
  108. Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
  109. Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
  110. Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
  111. Evaluation of open-source software for the lung segmentation
  112. Automatic determination of lung features of CF patients in CT scans
  113. Image analysis of self-organized multicellular patterns
  114. Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
  115. Radiopacity assessment of neurovascular implants
  116. Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
  117. Development of an artifact-free aneurysm clip
  118. Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
  119. Rapid prototyping of replica knee implants for in vitro testing
  120. Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
  121. Advanced wettability analysis of implant surfaces
  122. Patient-specific hip prostheses designed by surgeons
  123. Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
  124. Wear of a total intervertebral disc prosthesis
  125. Digital health and digital biomarkers – enabling value chains on health data
  126. Usability in the lifecycle of medical software development
  127. Influence of different test gases in a non-destructive 100% quality control system for medical devices
  128. Device development guided by user satisfaction survey on auricular vagus nerve stimulation
  129. Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
  130. Effect of left atrial hypertrophy on P-wave morphology in a computational model
  131. Simulation of intracardiac electrograms around acute ablation lesions
  132. Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
  133. Assessment of nasal resistance using computational fluid dynamics
  134. Resistance in a non-linear autoregressive model of pulmonary mechanics
  135. Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
  136. Determination of regional lung function in cystic fibrosis using electrical impedance tomography
  137. Development of parietal bone surrogates for parietal graft lift training
  138. Numerical simulation of mechanically stimulated bone remodelling
  139. Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
  140. Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
  141. Principle study on the signal connection at transabdominal fetal pulse oximetry
  142. Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
  143. Evaluating different approaches to identify a three parameter gas exchange model
  144. Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
  145. From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
  146. Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
  147. Improved FPGA controlled artificial vascular system for plethysmographic measurements
  148. Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
  149. Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
  150. Control scheme selection in human-machine- interfaces by analysis of activity signals
  151. Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
  152. Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
  153. Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
  154. Interactive monitoring system for visual respiratory biofeedback
  155. Development of a low-cost senor based aid for visually impaired people
  156. Patient assistive system for the shoulder joint
  157. A passive beating heart setup for interventional cardiology training
Downloaded on 8.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2016-0149/html
Scroll to top button