In vitro simulation of distribution processes following intramuscular injection

  • , , and EMAIL logo
Published/Copyright: September 30, 2016

Abstract

There is an urgent need for in vitro dissolution test setups for intramuscularly applied dosage forms. Especially biorelevant methods are needed to predict the in vivo behavior of newly developed dosage forms in a realistic way. There is a lack of knowledge regarding critical in vivo parameters influencing the release and absorption behavior of an intramuscularly applied drug. In the presented work the focus was set on the simulation of blood perfusion and muscle tissue. A solid agarose gel, being incorporated in an open-pored foam, was used to mimic the gel phase of muscle tissue and implemented in a flow through cell. An aqueous solution of fluorescein sodium was injected. Compared to recently obtained in vivo results the distribution of the model substance was very slow. Furthermore an agarose gel of lower viscosity an open-pored foam and phosphate buffer saline pH 7.4 were implemented in a multi-channel-ceramic membrane serving as a holder for the muscle imitating material. Blood simulating release medium was perfused through the ceramic membrane including filling materials. Transport of the dissolved fluorescein sodium was, in case of the gel, not only determined by diffusion but also by convective transport processes. The more realistic the muscle simulating materials were constituted the less reproducible results were obtained with the designed test setups.

1 Introduction

Long acting depot formulations for subcutaneous or intramuscular application are getting into the focus of pharmaceutical industry as drug delivery systems. They offer the advantage of a constant drug release over a long period of time and therewith less fluctuating plasma levels compared to a daily application of other dosage forms. Furthermore, the gastrointestinal pathway can be circumvented which is very useful for drugs with stability problems in the gastrointestinal tract or a small absorption window in the small intestine. Nevertheless, there is a lack of in vitro dissolution methods for parenterally administered dosage forms [1]. This might be caused by the great variability of parenterally applied formulations but also by the lack of data describing most critical parameters influencing drug absorption in vivo. In the following work the focus was set on intramuscularly applied dosage forms. It is expected that, next to other parameters, blood flow within the muscle tissue, the formation of a depot and the exact location of the injection within the muscle tissue pose critical variables on drug absorption into the systemic circulation [2]. Within the presented in vitro study these parameters were taken into consideration. Thus it was the aim of the study to simulate the process of injection into the muscle tissue, the blood flow through the muscle tissue (and injected depot) and the muscle tissue itself.

2 Material and methods

For the imitation of the gel phase of the muscle tissue being embedded between the muscle fibres an open-pored foam (Carpenter®, Germany) in combination with an agarose gel 2% was used. The thermoreversible agarose gel was heated to 90°C and consequently incorporated in the foam before cooling and solidification. The foam comprised four radial arranged cavities. Before solidification of the gel phase the foam was mounted on metal rods that were attached over a thread to a stainless steel plate. This plate was fitted into a flow through cell (FTC) which possessed a conical part that was shortened by 10 mm in favor of the cylindrical part, compared to the compendial FTC for tablets (see Figure 1).

Figure 1 Prepared gel-foam-block with injected depot in FTC. (Arrow indicates the flow direction of the release medium).
Figure 1

Prepared gel-foam-block with injected depot in FTC. (Arrow indicates the flow direction of the release medium).

After the gel became solid the metal rods were removed and 30 μl of an aqueous solution of fluorescein sodium were injected centrally, using an injection stencil, into the gel-foam-block. Consequently, the block was perfused (35 ml/min) within the FTC by the release medium (phosphate buffer saline pH 7.4, PBS, 37°C) operating in a closed system. At predetermined time points samples were taken from the reservoir and replaced by fresh buffer. The fluorescein contents of samples were quantified using a fluorescence reader (Varioskan Flash Multimode Reader, Thermo Fisher Scientific, USA) with calibration samples on each well-plate. Furthermore, the shape of the depot was observed at certain time points after the injection of an aqueous solution of methylene blue by cutting the block in half.

Besides the gel-foam-block the same foam saturated with PBS solution and agarose gels with lower viscosities were tested as muscle tissue imitating materials reflecting in vivo conditions in a potentially more realistic way. The release test setup had to allow for the implementation of the non-solid agarose gel. For this purpose a porous multi-channel-ceramic membrane (inopor®, Germany) saturated with the release medium was used as a holder. Afterwards 30 μl of an aqueous solution of fluorescein sodium were injected centrally into the filled inner cavity. Besides the foam and agarose gel 0.05%, PBS was used as filling material. A plug was set on the cavity which was provided with a tube (see Figure 2). The prepared membrane was then placed in a beaker filled with 500 ml PBS. A peristaltic pump led the medium out of the beaker via the tube through the inner cavity of the ceramic membrane. The flow rate was set to 2 ml/min. Due to its porous structure the medium passed the membrane and was recirculated into the reservoir. At certain time points samples were taken from the reservoir and replaced with fresh buffer. The quantification of the substance in the samples was performed using fluorescence spectroscopy, as stated above.

Figure 2 Prepared ceramic membrane with injected depot. (Arrow indicates the flow direction of the release medium).
Figure 2

Prepared ceramic membrane with injected depot. (Arrow indicates the flow direction of the release medium).

3 Results

The gel-foam-blocks allowed the injection of an aqueous depot. The depot spread over the time and was removed slowly after perfusion with PBS (see Figure 3).

Figure 3 Spreading of an aqueous depot of methylene blue centrally injected into an agarose-foam-block with four radial arranged cavities and perfused at 35 ml/min in a modified FTC.
Figure 3

Spreading of an aqueous depot of methylene blue centrally injected into an agarose-foam-block with four radial arranged cavities and perfused at 35 ml/min in a modified FTC.

About 30 h after the injection of an aqueous solution of fluorescein sodium 100% of the injected amount were found in the release medium. The standard deviations within the first 12 h were remarkable (see Figure 4).

Figure 4 Distribution of fluorescein sodium into the release medium after the injection of an aqueous solution into the agarose-foam-block located in the FTC perfused at 35 ml/min over the time. (n =3; MW ± SD).
Figure 4

Distribution of fluorescein sodium into the release medium after the injection of an aqueous solution into the agarose-foam-block located in the FTC perfused at 35 ml/min over the time. (n =3; MW ± SD).

Also the used open-pored foam, which was saturated in PBS allowed the injection of an aqueous solution and the formation of a depot. Nevertheless, the depot was only stable for a few minutes before spreading throughout the whole foam. It was found that an agarose gel at a concentration of 0.05% without a foam structure yielded the formation of a depot. The spreading process was slower and more controlled than in the PBS soaked foam (results not shown) but faster than in the gel-foam-block (Figure 5).

Figure 5 Spreading of an aqueous depot (methylene blue) being injected in an agarose gel 0.05% over the time.
Figure 5

Spreading of an aqueous depot (methylene blue) being injected in an agarose gel 0.05% over the time.

Figure 6 shows the curves describing the distribution of the injected flourescein sodium from PBS, agarose gel 0.05%, and open-pored foam all located within the ceramic membrane into the release medium over the time. The fastest distribution occurred after the injection into PBS reaching a plateau after about 15 min. A slower distribution process, taking about 60 min, was observed using the foam as well as the agarose gel. The distribution of the substance from the agarose gel was not completed within that time and showed a plateau between 10 and 40 min. The standard deviations in case of the foam and especially in case of the agarose gel were remarkable. Filling the inner cavity of the ceramic membrane with PBS led to more reproducible results.

Figure 6 Distribution of fluorescein sodium into the release medium after the injection of an aqueous solution into the inner cavity of the ceramic membrane filled with PBS pH 7.4, Agarose gel 0.05% or an open-pored foam being perfused at 2 ml/min over the time. (n =3; MW ± SD).
Figure 6

Distribution of fluorescein sodium into the release medium after the injection of an aqueous solution into the inner cavity of the ceramic membrane filled with PBS pH 7.4, Agarose gel 0.05% or an open-pored foam being perfused at 2 ml/min over the time. (n =3; MW ± SD).

4 Discussion

It is expected, that the extent of the influence of the different physiological parameters that are affecting drug release and absorption after intramuscular application is not constant but depends on the formulation. In the presented study the focus was set on simulating blood perfusion, which is thought to be relevant especially in fast releasing dosage forms [3], and the simulation of the muscle tissue, having an impact on the formation of a depot and on the time for removal of the depot from the muscle tissue [2]. The application of a FTC-system allows the simulation of blood perfusion through the muscle tissue. The flow rate can be easily adapted. Volumes of the release medium were chosen with respect to the maintenance of sink-conditions. Though in vivo no sink-conditions are present it is crucial to maintain them in vitro in order to achieve reproducible results. In general, muscle tissue consists of muscle fibres surrounded by connective tissue. The connective tissue consists of a framework of collagen fibres in which a gel phase is embedded. Since an injection into the surrounding connective tissue is very likely a simulation of this tissue for the imitation of the muscle tissue seems to be a promising approach. In vitro an agarose gel was used as a hydrogel. Since it was the aim to implement and perfuse the imitated muscle tissue in a FTC it had to be of solid consistency. Injecting into a solid gel led to a backflow along the puncture channel. Consequently the gel was incorporated in an open-pored foam interrupting the tight framework of the gel and enabling the forming of a depot. Since a diffusion of the solved model substance through the framework had to be possible an open-pored foam had to be used. A recently described in vivo study in rats [2] showed, that an aqueous depot is removed from the muscle tissue within 2 – 6 h whilst maximal plasma levels were already achieved within about 30 min after the injection. Thus, absorption processes were finished at the latest after 2 h, when no drug was quantified in the blood anymore [2]. The spreading and the removal of the depot in the gel-foam-block took much longer time. The time that was needed until the whole amount of the injected model substance was found in the release medium was about 30 h. Thus, the diffusion process through the solid gel was slower than the diffusion process of the drug molecules in the muscle tissue. Furthermore the standard deviations between the single trials were very high. One reason for that can be an uneven texture of the gel-foam-block. Another reason might be a non reproducible injection technique although a standardization of the injection depth and angle was assured by using an injection stencil.

In vivo the transport of a dissolved substance is thought not only to be determined by its diffusion but also by convective transport processes. Furthermore, the viscosity of the gel phase of the muscle tissue is considerably lower than of a solid hydrogel. For this reason another test setup was applied. An agarose gel of low viscosity, allowing the injection and forming of a depot without incorporating the gel in a foam, and the foam without the use of a gel were tested as muscle imitating materials. Since the agarose gel 0.05% is in a liquid state an implementation into the FTC was not possible. A porous ceramic membrane was used as a holder for this material. In order to perfuse the imitated muscle tissue, under consideration of convective transport mechanisms, the release medium was pumped through the membrane filled with either the foam, the agarose gel 0.05% or PBS, after the injection of the aqueous solution of fluorescein sodium. PBS was used as filling material in order to investigate the influence of the ceramic membrane as a barrier. The new test setup accelerated the time needed for a complete distribution of the injected model substance into the release medium to 10–60 min. This reflects the time needed for complete absorption in vivo in a better way. In the case of the foam or the low viscosity gel were filled in the inner cavity of the membrane, the times to reach a plateau were increased compared to a filling with PBS. Thus the filling materials have a greater effect on the distribution of flourescein than the ceramic itself. Furthermore, the standard deviations increased greatly using muscle imitating materials. The results were not reproducible anymore. In case of the agarose gel the reason for that may be different dilution processes of the gel due to the perfusion of the ceramic membrane. It cannot be excluded that liquid nests were formed influencing the transport time of the dissolved model substance. In the case of PBS and the foam only convective transport processes may play a role. The use of an agarose gel requires transport via diffusion and convective processes, but in a non reproducible way.

In summary, the presented test setups reflect in vivo conditions in some parts. The more biorelevant the method is designed the more problems emerge in creating reproducible results. Nevertheless, it may give a hint that the observed high variability in plasma levels in vivo after intramuscular injection may result, from the constitution of the muscle tissue and therefore be dependent on the injection site.

Acknowledgement

The authors thank inopor® and Carpenter® for generously providing ceramic membranes and open-pored foams.

Author’s Statement

Research funding: Financial support by the Federal Ministry of Education and Research (BMBF) within Remedis is gratefully acknowledged. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animal use.

References

[1] Martinez MN, Rathbone MJ, Burgess D, Huynh M. Breakout session summary from AAPS/CRS joint workshop on critical variables in the in vitro and in vivo performance of parenteral sustained release products. J Control Release. 2010;142:2–7.10.1016/j.jconrel.2009.09.028Search in Google Scholar PubMed

[2] Probst M, Kühn J-P, Scheuch E, Seidlitz A, Hadlich S, Evert K, et al. Weitschies, Simultaneous magnetic resonance imaging and pharmacokinetic analysis of intramuscular depots. J Control Release. 2016;227:1–12.10.1016/j.jconrel.2016.02.029Search in Google Scholar PubMed

[3] Soni SD, Wiles D, Schiff A, Bamrah J. Plasma levels of fluphenazine decanoate. Effects of site of injection, massage and muscle activity. Br J Psychiatry. 1998;153:382–4.10.1192/bjp.153.3.382Search in Google Scholar PubMed

Published Online: 2016-9-30
Published in Print: 2016-9-1

©2016 Anne Seidlitz et al., licensee De Gruyter.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.

Articles in the same Issue

  1. Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
  2. Novel blood protein based scaffolds for cardiovascular tissue engineering
  3. Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
  4. Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
  5. Long-term stable modifications of silicone elastomer for improved hemocompatibility
  6. The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
  7. Biocompatible wear-resistant thick ceramic coating
  8. Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
  9. Examination of dielectric strength of thin Parylene C films under various conditions
  10. Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
  11. Systemic analysis about residual chloroform in PLLA films
  12. A macrophage model of osseointegration
  13. Towards in silico prognosis using big data
  14. Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
  15. Usability evaluation of a locomotor therapy device considering different strategies
  16. Hypoxia-on-a-chip
  17. Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
  18. Fabrication of MEMS-based 3D-μECoG-MEAs
  19. High speed digital interfacing for a neural data acquisition system
  20. Bionic forceps for the handling of sensitive tissue
  21. Experimental studies on 3D printing of barium titanate ceramics for medical applications
  22. Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
  23. 3D printing – a key technology for tailored biomedical cell culture lab ware
  24. 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
  25. Biocompatibility of photopolymers for additive manufacturing
  26. Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
  27. Novel wireless measurement system of pressure dedicated to in vivo studies
  28. Portable auricular device for real-time swallow and chew detection
  29. Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
  30. Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
  31. Deep brain stimulation: increasing efficiency by alternative waveforms
  32. Prediction of immediately occurring microsleep events from brain electric signals
  33. Determining cardiac vagal threshold from short term heart rate complexity
  34. Classification of cardiac excitation patterns during atrial fibrillation
  35. An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
  36. Deriving respiration from high resolution 12-channel-ECG during cycling exercise
  37. Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
  38. Automatic detection and mapping of double potentials in intracardiac electrograms
  39. Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
  40. Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
  41. Best practice: surgeon driven application in pelvic operations
  42. Vasomotor assessment by camera-based photoplethysmography
  43. Classification of morphologic changes in photoplethysmographic waveforms
  44. Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
  45. Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
  46. Nonlinear causal influences assessed by mutual compression entropy
  47. Comparative study of methods for solving the correspondence problem in EMD applications
  48. fNIRS for future use in auditory diagnostics
  49. Semi-automated detection of fractional shortening in zebrafish embryo heart videos
  50. Blood pressure measurement on the cheek
  51. Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
  52. Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
  53. An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
  54. Real-time QRS detection using integrated variance for ECG gated cardiac MRI
  55. Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
  56. In-vitro experiments to characterize ventricular electromechanics
  57. Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
  58. Application of microwave sensor technology in cardiovascular disease for plaque detection
  59. Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
  60. Detection of microsleep events in a car driving simulation study using electrocardiographic features
  61. A method to determine the kink resistance of stents and stent delivery systems according to international standards
  62. Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
  63. Transient Euler-Lagrange/DEM simulation of stent thrombosis
  64. Automated control of the laser welding process of heart valve scaffolds
  65. Automation of a test bench for accessing the bendability of electrospun vascular grafts
  66. Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
  67. Cryopreservation of cells using defined serum-free cryoprotective agents
  68. New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
  69. Determination of the membrane hydraulic permeability of MSCs
  70. Climate retainment in carbon dioxide incubators
  71. Multiple factors influencing OR ventilation system effectiveness
  72. Evaluation of an app-based stress protocol
  73. Medication process in Styrian hospitals
  74. Control tower to surgical theater
  75. Development of a skull phantom for the assessment of implant X-ray visibility
  76. Surgical navigation with QR codes
  77. Investigation of the pressure gradient of embolic protection devices
  78. Computer assistance in femoral derotation osteotomy: a bottom-up approach
  79. Automatic depth scanning system for 3D infrared thermography
  80. A service for monitoring the quality of intraoperative cone beam CT images
  81. Resectoscope with an easy to use twist mechanism for improved handling
  82. In vitro simulation of distribution processes following intramuscular injection
  83. Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
  84. A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
  85. Smart control for functional electrical stimulation with optimal pulse intensity
  86. Tactile display on the remaining hand for unilateral hand amputees
  87. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
  88. An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
  89. Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
  90. Learning discriminative classification models for grading anal intraepithelial neoplasia
  91. Regularization of EIT reconstruction based on multi-scales wavelet transforms
  92. Assessing MRI susceptibility artefact through an indicator of image distortion
  93. EyeGuidance – a computer controlled system to guide eye movements
  94. A framework for feedback-based segmentation of 3D image stacks
  95. Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
  96. 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
  97. Inside-Out access strategy using new trans-vascular catheter approach
  98. US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
  99. Impact of different registration methods in MEG source analysis
  100. 3D segmentation of thyroid ultrasound images using active contours
  101. Designing a compact MRI motion phantom
  102. Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
  103. Classification of indirect immunofluorescence images using thresholded local binary count features
  104. Analysis of muscle fatigue conditions using time-frequency images and GLCM features
  105. Numerical evaluation of image parameters of ETR-1
  106. Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
  107. Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
  108. Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
  109. Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
  110. Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
  111. Evaluation of open-source software for the lung segmentation
  112. Automatic determination of lung features of CF patients in CT scans
  113. Image analysis of self-organized multicellular patterns
  114. Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
  115. Radiopacity assessment of neurovascular implants
  116. Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
  117. Development of an artifact-free aneurysm clip
  118. Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
  119. Rapid prototyping of replica knee implants for in vitro testing
  120. Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
  121. Advanced wettability analysis of implant surfaces
  122. Patient-specific hip prostheses designed by surgeons
  123. Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
  124. Wear of a total intervertebral disc prosthesis
  125. Digital health and digital biomarkers – enabling value chains on health data
  126. Usability in the lifecycle of medical software development
  127. Influence of different test gases in a non-destructive 100% quality control system for medical devices
  128. Device development guided by user satisfaction survey on auricular vagus nerve stimulation
  129. Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
  130. Effect of left atrial hypertrophy on P-wave morphology in a computational model
  131. Simulation of intracardiac electrograms around acute ablation lesions
  132. Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
  133. Assessment of nasal resistance using computational fluid dynamics
  134. Resistance in a non-linear autoregressive model of pulmonary mechanics
  135. Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
  136. Determination of regional lung function in cystic fibrosis using electrical impedance tomography
  137. Development of parietal bone surrogates for parietal graft lift training
  138. Numerical simulation of mechanically stimulated bone remodelling
  139. Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
  140. Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
  141. Principle study on the signal connection at transabdominal fetal pulse oximetry
  142. Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
  143. Evaluating different approaches to identify a three parameter gas exchange model
  144. Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
  145. From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
  146. Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
  147. Improved FPGA controlled artificial vascular system for plethysmographic measurements
  148. Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
  149. Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
  150. Control scheme selection in human-machine- interfaces by analysis of activity signals
  151. Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
  152. Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
  153. Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
  154. Interactive monitoring system for visual respiratory biofeedback
  155. Development of a low-cost senor based aid for visually impaired people
  156. Patient assistive system for the shoulder joint
  157. A passive beating heart setup for interventional cardiology training
Downloaded on 21.4.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2016-0085/html
Scroll to top button