Abstract
There is an urgent need for in vitro dissolution test setups for intramuscularly applied dosage forms. Especially biorelevant methods are needed to predict the in vivo behavior of newly developed dosage forms in a realistic way. There is a lack of knowledge regarding critical in vivo parameters influencing the release and absorption behavior of an intramuscularly applied drug. In the presented work the focus was set on the simulation of blood perfusion and muscle tissue. A solid agarose gel, being incorporated in an open-pored foam, was used to mimic the gel phase of muscle tissue and implemented in a flow through cell. An aqueous solution of fluorescein sodium was injected. Compared to recently obtained in vivo results the distribution of the model substance was very slow. Furthermore an agarose gel of lower viscosity an open-pored foam and phosphate buffer saline pH 7.4 were implemented in a multi-channel-ceramic membrane serving as a holder for the muscle imitating material. Blood simulating release medium was perfused through the ceramic membrane including filling materials. Transport of the dissolved fluorescein sodium was, in case of the gel, not only determined by diffusion but also by convective transport processes. The more realistic the muscle simulating materials were constituted the less reproducible results were obtained with the designed test setups.
1 Introduction
Long acting depot formulations for subcutaneous or intramuscular application are getting into the focus of pharmaceutical industry as drug delivery systems. They offer the advantage of a constant drug release over a long period of time and therewith less fluctuating plasma levels compared to a daily application of other dosage forms. Furthermore, the gastrointestinal pathway can be circumvented which is very useful for drugs with stability problems in the gastrointestinal tract or a small absorption window in the small intestine. Nevertheless, there is a lack of in vitro dissolution methods for parenterally administered dosage forms [1]. This might be caused by the great variability of parenterally applied formulations but also by the lack of data describing most critical parameters influencing drug absorption in vivo. In the following work the focus was set on intramuscularly applied dosage forms. It is expected that, next to other parameters, blood flow within the muscle tissue, the formation of a depot and the exact location of the injection within the muscle tissue pose critical variables on drug absorption into the systemic circulation [2]. Within the presented in vitro study these parameters were taken into consideration. Thus it was the aim of the study to simulate the process of injection into the muscle tissue, the blood flow through the muscle tissue (and injected depot) and the muscle tissue itself.
2 Material and methods
For the imitation of the gel phase of the muscle tissue being embedded between the muscle fibres an open-pored foam (Carpenter®, Germany) in combination with an agarose gel 2% was used. The thermoreversible agarose gel was heated to 90°C and consequently incorporated in the foam before cooling and solidification. The foam comprised four radial arranged cavities. Before solidification of the gel phase the foam was mounted on metal rods that were attached over a thread to a stainless steel plate. This plate was fitted into a flow through cell (FTC) which possessed a conical part that was shortened by 10 mm in favor of the cylindrical part, compared to the compendial FTC for tablets (see Figure 1).

Prepared gel-foam-block with injected depot in FTC. (Arrow indicates the flow direction of the release medium).
After the gel became solid the metal rods were removed and 30 μl of an aqueous solution of fluorescein sodium were injected centrally, using an injection stencil, into the gel-foam-block. Consequently, the block was perfused (35 ml/min) within the FTC by the release medium (phosphate buffer saline pH 7.4, PBS, 37°C) operating in a closed system. At predetermined time points samples were taken from the reservoir and replaced by fresh buffer. The fluorescein contents of samples were quantified using a fluorescence reader (Varioskan Flash Multimode Reader, Thermo Fisher Scientific, USA) with calibration samples on each well-plate. Furthermore, the shape of the depot was observed at certain time points after the injection of an aqueous solution of methylene blue by cutting the block in half.
Besides the gel-foam-block the same foam saturated with PBS solution and agarose gels with lower viscosities were tested as muscle tissue imitating materials reflecting in vivo conditions in a potentially more realistic way. The release test setup had to allow for the implementation of the non-solid agarose gel. For this purpose a porous multi-channel-ceramic membrane (inopor®, Germany) saturated with the release medium was used as a holder. Afterwards 30 μl of an aqueous solution of fluorescein sodium were injected centrally into the filled inner cavity. Besides the foam and agarose gel 0.05%, PBS was used as filling material. A plug was set on the cavity which was provided with a tube (see Figure 2). The prepared membrane was then placed in a beaker filled with 500 ml PBS. A peristaltic pump led the medium out of the beaker via the tube through the inner cavity of the ceramic membrane. The flow rate was set to 2 ml/min. Due to its porous structure the medium passed the membrane and was recirculated into the reservoir. At certain time points samples were taken from the reservoir and replaced with fresh buffer. The quantification of the substance in the samples was performed using fluorescence spectroscopy, as stated above.

Prepared ceramic membrane with injected depot. (Arrow indicates the flow direction of the release medium).
3 Results
The gel-foam-blocks allowed the injection of an aqueous depot. The depot spread over the time and was removed slowly after perfusion with PBS (see Figure 3).

Spreading of an aqueous depot of methylene blue centrally injected into an agarose-foam-block with four radial arranged cavities and perfused at 35 ml/min in a modified FTC.
About 30 h after the injection of an aqueous solution of fluorescein sodium 100% of the injected amount were found in the release medium. The standard deviations within the first 12 h were remarkable (see Figure 4).

Distribution of fluorescein sodium into the release medium after the injection of an aqueous solution into the agarose-foam-block located in the FTC perfused at 35 ml/min over the time. (n =3; MW ± SD).
Also the used open-pored foam, which was saturated in PBS allowed the injection of an aqueous solution and the formation of a depot. Nevertheless, the depot was only stable for a few minutes before spreading throughout the whole foam. It was found that an agarose gel at a concentration of 0.05% without a foam structure yielded the formation of a depot. The spreading process was slower and more controlled than in the PBS soaked foam (results not shown) but faster than in the gel-foam-block (Figure 5).

Spreading of an aqueous depot (methylene blue) being injected in an agarose gel 0.05% over the time.
Figure 6 shows the curves describing the distribution of the injected flourescein sodium from PBS, agarose gel 0.05%, and open-pored foam all located within the ceramic membrane into the release medium over the time. The fastest distribution occurred after the injection into PBS reaching a plateau after about 15 min. A slower distribution process, taking about 60 min, was observed using the foam as well as the agarose gel. The distribution of the substance from the agarose gel was not completed within that time and showed a plateau between 10 and 40 min. The standard deviations in case of the foam and especially in case of the agarose gel were remarkable. Filling the inner cavity of the ceramic membrane with PBS led to more reproducible results.

Distribution of fluorescein sodium into the release medium after the injection of an aqueous solution into the inner cavity of the ceramic membrane filled with PBS pH 7.4, Agarose gel 0.05% or an open-pored foam being perfused at 2 ml/min over the time. (n =3; MW ± SD).
4 Discussion
It is expected, that the extent of the influence of the different physiological parameters that are affecting drug release and absorption after intramuscular application is not constant but depends on the formulation. In the presented study the focus was set on simulating blood perfusion, which is thought to be relevant especially in fast releasing dosage forms [3], and the simulation of the muscle tissue, having an impact on the formation of a depot and on the time for removal of the depot from the muscle tissue [2]. The application of a FTC-system allows the simulation of blood perfusion through the muscle tissue. The flow rate can be easily adapted. Volumes of the release medium were chosen with respect to the maintenance of sink-conditions. Though in vivo no sink-conditions are present it is crucial to maintain them in vitro in order to achieve reproducible results. In general, muscle tissue consists of muscle fibres surrounded by connective tissue. The connective tissue consists of a framework of collagen fibres in which a gel phase is embedded. Since an injection into the surrounding connective tissue is very likely a simulation of this tissue for the imitation of the muscle tissue seems to be a promising approach. In vitro an agarose gel was used as a hydrogel. Since it was the aim to implement and perfuse the imitated muscle tissue in a FTC it had to be of solid consistency. Injecting into a solid gel led to a backflow along the puncture channel. Consequently the gel was incorporated in an open-pored foam interrupting the tight framework of the gel and enabling the forming of a depot. Since a diffusion of the solved model substance through the framework had to be possible an open-pored foam had to be used. A recently described in vivo study in rats [2] showed, that an aqueous depot is removed from the muscle tissue within 2 – 6 h whilst maximal plasma levels were already achieved within about 30 min after the injection. Thus, absorption processes were finished at the latest after 2 h, when no drug was quantified in the blood anymore [2]. The spreading and the removal of the depot in the gel-foam-block took much longer time. The time that was needed until the whole amount of the injected model substance was found in the release medium was about 30 h. Thus, the diffusion process through the solid gel was slower than the diffusion process of the drug molecules in the muscle tissue. Furthermore the standard deviations between the single trials were very high. One reason for that can be an uneven texture of the gel-foam-block. Another reason might be a non reproducible injection technique although a standardization of the injection depth and angle was assured by using an injection stencil.
In vivo the transport of a dissolved substance is thought not only to be determined by its diffusion but also by convective transport processes. Furthermore, the viscosity of the gel phase of the muscle tissue is considerably lower than of a solid hydrogel. For this reason another test setup was applied. An agarose gel of low viscosity, allowing the injection and forming of a depot without incorporating the gel in a foam, and the foam without the use of a gel were tested as muscle imitating materials. Since the agarose gel 0.05% is in a liquid state an implementation into the FTC was not possible. A porous ceramic membrane was used as a holder for this material. In order to perfuse the imitated muscle tissue, under consideration of convective transport mechanisms, the release medium was pumped through the membrane filled with either the foam, the agarose gel 0.05% or PBS, after the injection of the aqueous solution of fluorescein sodium. PBS was used as filling material in order to investigate the influence of the ceramic membrane as a barrier. The new test setup accelerated the time needed for a complete distribution of the injected model substance into the release medium to 10–60 min. This reflects the time needed for complete absorption in vivo in a better way. In the case of the foam or the low viscosity gel were filled in the inner cavity of the membrane, the times to reach a plateau were increased compared to a filling with PBS. Thus the filling materials have a greater effect on the distribution of flourescein than the ceramic itself. Furthermore, the standard deviations increased greatly using muscle imitating materials. The results were not reproducible anymore. In case of the agarose gel the reason for that may be different dilution processes of the gel due to the perfusion of the ceramic membrane. It cannot be excluded that liquid nests were formed influencing the transport time of the dissolved model substance. In the case of PBS and the foam only convective transport processes may play a role. The use of an agarose gel requires transport via diffusion and convective processes, but in a non reproducible way.
In summary, the presented test setups reflect in vivo conditions in some parts. The more biorelevant the method is designed the more problems emerge in creating reproducible results. Nevertheless, it may give a hint that the observed high variability in plasma levels in vivo after intramuscular injection may result, from the constitution of the muscle tissue and therefore be dependent on the injection site.
Acknowledgement
The authors thank inopor® and Carpenter® for generously providing ceramic membranes and open-pored foams.
Author’s Statement
Research funding: Financial support by the Federal Ministry of Education and Research (BMBF) within Remedis is gratefully acknowledged. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animal use.
References
[1] Martinez MN, Rathbone MJ, Burgess D, Huynh M. Breakout session summary from AAPS/CRS joint workshop on critical variables in the in vitro and in vivo performance of parenteral sustained release products. J Control Release. 2010;142:2–7.10.1016/j.jconrel.2009.09.028Search in Google Scholar PubMed
[2] Probst M, Kühn J-P, Scheuch E, Seidlitz A, Hadlich S, Evert K, et al. Weitschies, Simultaneous magnetic resonance imaging and pharmacokinetic analysis of intramuscular depots. J Control Release. 2016;227:1–12.10.1016/j.jconrel.2016.02.029Search in Google Scholar PubMed
[3] Soni SD, Wiles D, Schiff A, Bamrah J. Plasma levels of fluphenazine decanoate. Effects of site of injection, massage and muscle activity. Br J Psychiatry. 1998;153:382–4.10.1192/bjp.153.3.382Search in Google Scholar PubMed
©2016 Anne Seidlitz et al., licensee De Gruyter.
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.
Articles in the same Issue
- Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
- Novel blood protein based scaffolds for cardiovascular tissue engineering
- Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
- Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
- Long-term stable modifications of silicone elastomer for improved hemocompatibility
- The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
- Biocompatible wear-resistant thick ceramic coating
- Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
- Examination of dielectric strength of thin Parylene C films under various conditions
- Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
- Systemic analysis about residual chloroform in PLLA films
- A macrophage model of osseointegration
- Towards in silico prognosis using big data
- Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
- Usability evaluation of a locomotor therapy device considering different strategies
- Hypoxia-on-a-chip
- Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
- Fabrication of MEMS-based 3D-μECoG-MEAs
- High speed digital interfacing for a neural data acquisition system
- Bionic forceps for the handling of sensitive tissue
- Experimental studies on 3D printing of barium titanate ceramics for medical applications
- Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
- 3D printing – a key technology for tailored biomedical cell culture lab ware
- 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
- Biocompatibility of photopolymers for additive manufacturing
- Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
- Novel wireless measurement system of pressure dedicated to in vivo studies
- Portable auricular device for real-time swallow and chew detection
- Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
- Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
- Deep brain stimulation: increasing efficiency by alternative waveforms
- Prediction of immediately occurring microsleep events from brain electric signals
- Determining cardiac vagal threshold from short term heart rate complexity
- Classification of cardiac excitation patterns during atrial fibrillation
- An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
- Deriving respiration from high resolution 12-channel-ECG during cycling exercise
- Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
- Automatic detection and mapping of double potentials in intracardiac electrograms
- Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
- Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
- Best practice: surgeon driven application in pelvic operations
- Vasomotor assessment by camera-based photoplethysmography
- Classification of morphologic changes in photoplethysmographic waveforms
- Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
- Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
- Nonlinear causal influences assessed by mutual compression entropy
- Comparative study of methods for solving the correspondence problem in EMD applications
- fNIRS for future use in auditory diagnostics
- Semi-automated detection of fractional shortening in zebrafish embryo heart videos
- Blood pressure measurement on the cheek
- Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
- Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
- An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
- Real-time QRS detection using integrated variance for ECG gated cardiac MRI
- Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
- In-vitro experiments to characterize ventricular electromechanics
- Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
- Application of microwave sensor technology in cardiovascular disease for plaque detection
- Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
- Detection of microsleep events in a car driving simulation study using electrocardiographic features
- A method to determine the kink resistance of stents and stent delivery systems according to international standards
- Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
- Transient Euler-Lagrange/DEM simulation of stent thrombosis
- Automated control of the laser welding process of heart valve scaffolds
- Automation of a test bench for accessing the bendability of electrospun vascular grafts
- Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
- Cryopreservation of cells using defined serum-free cryoprotective agents
- New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
- Determination of the membrane hydraulic permeability of MSCs
- Climate retainment in carbon dioxide incubators
- Multiple factors influencing OR ventilation system effectiveness
- Evaluation of an app-based stress protocol
- Medication process in Styrian hospitals
- Control tower to surgical theater
- Development of a skull phantom for the assessment of implant X-ray visibility
- Surgical navigation with QR codes
- Investigation of the pressure gradient of embolic protection devices
- Computer assistance in femoral derotation osteotomy: a bottom-up approach
- Automatic depth scanning system for 3D infrared thermography
- A service for monitoring the quality of intraoperative cone beam CT images
- Resectoscope with an easy to use twist mechanism for improved handling
- In vitro simulation of distribution processes following intramuscular injection
- Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
- A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
- Smart control for functional electrical stimulation with optimal pulse intensity
- Tactile display on the remaining hand for unilateral hand amputees
- Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
- An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
- Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
- Learning discriminative classification models for grading anal intraepithelial neoplasia
- Regularization of EIT reconstruction based on multi-scales wavelet transforms
- Assessing MRI susceptibility artefact through an indicator of image distortion
- EyeGuidance – a computer controlled system to guide eye movements
- A framework for feedback-based segmentation of 3D image stacks
- Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
- 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
- Inside-Out access strategy using new trans-vascular catheter approach
- US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
- Impact of different registration methods in MEG source analysis
- 3D segmentation of thyroid ultrasound images using active contours
- Designing a compact MRI motion phantom
- Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
- Classification of indirect immunofluorescence images using thresholded local binary count features
- Analysis of muscle fatigue conditions using time-frequency images and GLCM features
- Numerical evaluation of image parameters of ETR-1
- Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
- Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
- Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
- Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
- Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
- Evaluation of open-source software for the lung segmentation
- Automatic determination of lung features of CF patients in CT scans
- Image analysis of self-organized multicellular patterns
- Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
- Radiopacity assessment of neurovascular implants
- Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
- Development of an artifact-free aneurysm clip
- Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
- Rapid prototyping of replica knee implants for in vitro testing
- Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
- Advanced wettability analysis of implant surfaces
- Patient-specific hip prostheses designed by surgeons
- Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
- Wear of a total intervertebral disc prosthesis
- Digital health and digital biomarkers – enabling value chains on health data
- Usability in the lifecycle of medical software development
- Influence of different test gases in a non-destructive 100% quality control system for medical devices
- Device development guided by user satisfaction survey on auricular vagus nerve stimulation
- Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
- Effect of left atrial hypertrophy on P-wave morphology in a computational model
- Simulation of intracardiac electrograms around acute ablation lesions
- Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
- Assessment of nasal resistance using computational fluid dynamics
- Resistance in a non-linear autoregressive model of pulmonary mechanics
- Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
- Determination of regional lung function in cystic fibrosis using electrical impedance tomography
- Development of parietal bone surrogates for parietal graft lift training
- Numerical simulation of mechanically stimulated bone remodelling
- Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
- Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
- Principle study on the signal connection at transabdominal fetal pulse oximetry
- Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
- Evaluating different approaches to identify a three parameter gas exchange model
- Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
- From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
- Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
- Improved FPGA controlled artificial vascular system for plethysmographic measurements
- Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
- Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
- Control scheme selection in human-machine- interfaces by analysis of activity signals
- Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
- Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
- Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
- Interactive monitoring system for visual respiratory biofeedback
- Development of a low-cost senor based aid for visually impaired people
- Patient assistive system for the shoulder joint
- A passive beating heart setup for interventional cardiology training
Articles in the same Issue
- Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
- Novel blood protein based scaffolds for cardiovascular tissue engineering
- Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
- Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
- Long-term stable modifications of silicone elastomer for improved hemocompatibility
- The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
- Biocompatible wear-resistant thick ceramic coating
- Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
- Examination of dielectric strength of thin Parylene C films under various conditions
- Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
- Systemic analysis about residual chloroform in PLLA films
- A macrophage model of osseointegration
- Towards in silico prognosis using big data
- Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
- Usability evaluation of a locomotor therapy device considering different strategies
- Hypoxia-on-a-chip
- Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
- Fabrication of MEMS-based 3D-μECoG-MEAs
- High speed digital interfacing for a neural data acquisition system
- Bionic forceps for the handling of sensitive tissue
- Experimental studies on 3D printing of barium titanate ceramics for medical applications
- Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
- 3D printing – a key technology for tailored biomedical cell culture lab ware
- 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
- Biocompatibility of photopolymers for additive manufacturing
- Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
- Novel wireless measurement system of pressure dedicated to in vivo studies
- Portable auricular device for real-time swallow and chew detection
- Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
- Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
- Deep brain stimulation: increasing efficiency by alternative waveforms
- Prediction of immediately occurring microsleep events from brain electric signals
- Determining cardiac vagal threshold from short term heart rate complexity
- Classification of cardiac excitation patterns during atrial fibrillation
- An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
- Deriving respiration from high resolution 12-channel-ECG during cycling exercise
- Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
- Automatic detection and mapping of double potentials in intracardiac electrograms
- Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
- Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
- Best practice: surgeon driven application in pelvic operations
- Vasomotor assessment by camera-based photoplethysmography
- Classification of morphologic changes in photoplethysmographic waveforms
- Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
- Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
- Nonlinear causal influences assessed by mutual compression entropy
- Comparative study of methods for solving the correspondence problem in EMD applications
- fNIRS for future use in auditory diagnostics
- Semi-automated detection of fractional shortening in zebrafish embryo heart videos
- Blood pressure measurement on the cheek
- Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
- Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
- An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
- Real-time QRS detection using integrated variance for ECG gated cardiac MRI
- Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
- In-vitro experiments to characterize ventricular electromechanics
- Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
- Application of microwave sensor technology in cardiovascular disease for plaque detection
- Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
- Detection of microsleep events in a car driving simulation study using electrocardiographic features
- A method to determine the kink resistance of stents and stent delivery systems according to international standards
- Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
- Transient Euler-Lagrange/DEM simulation of stent thrombosis
- Automated control of the laser welding process of heart valve scaffolds
- Automation of a test bench for accessing the bendability of electrospun vascular grafts
- Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
- Cryopreservation of cells using defined serum-free cryoprotective agents
- New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
- Determination of the membrane hydraulic permeability of MSCs
- Climate retainment in carbon dioxide incubators
- Multiple factors influencing OR ventilation system effectiveness
- Evaluation of an app-based stress protocol
- Medication process in Styrian hospitals
- Control tower to surgical theater
- Development of a skull phantom for the assessment of implant X-ray visibility
- Surgical navigation with QR codes
- Investigation of the pressure gradient of embolic protection devices
- Computer assistance in femoral derotation osteotomy: a bottom-up approach
- Automatic depth scanning system for 3D infrared thermography
- A service for monitoring the quality of intraoperative cone beam CT images
- Resectoscope with an easy to use twist mechanism for improved handling
- In vitro simulation of distribution processes following intramuscular injection
- Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
- A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
- Smart control for functional electrical stimulation with optimal pulse intensity
- Tactile display on the remaining hand for unilateral hand amputees
- Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
- An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
- Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
- Learning discriminative classification models for grading anal intraepithelial neoplasia
- Regularization of EIT reconstruction based on multi-scales wavelet transforms
- Assessing MRI susceptibility artefact through an indicator of image distortion
- EyeGuidance – a computer controlled system to guide eye movements
- A framework for feedback-based segmentation of 3D image stacks
- Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
- 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
- Inside-Out access strategy using new trans-vascular catheter approach
- US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
- Impact of different registration methods in MEG source analysis
- 3D segmentation of thyroid ultrasound images using active contours
- Designing a compact MRI motion phantom
- Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
- Classification of indirect immunofluorescence images using thresholded local binary count features
- Analysis of muscle fatigue conditions using time-frequency images and GLCM features
- Numerical evaluation of image parameters of ETR-1
- Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
- Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
- Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
- Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
- Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
- Evaluation of open-source software for the lung segmentation
- Automatic determination of lung features of CF patients in CT scans
- Image analysis of self-organized multicellular patterns
- Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
- Radiopacity assessment of neurovascular implants
- Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
- Development of an artifact-free aneurysm clip
- Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
- Rapid prototyping of replica knee implants for in vitro testing
- Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
- Advanced wettability analysis of implant surfaces
- Patient-specific hip prostheses designed by surgeons
- Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
- Wear of a total intervertebral disc prosthesis
- Digital health and digital biomarkers – enabling value chains on health data
- Usability in the lifecycle of medical software development
- Influence of different test gases in a non-destructive 100% quality control system for medical devices
- Device development guided by user satisfaction survey on auricular vagus nerve stimulation
- Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
- Effect of left atrial hypertrophy on P-wave morphology in a computational model
- Simulation of intracardiac electrograms around acute ablation lesions
- Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
- Assessment of nasal resistance using computational fluid dynamics
- Resistance in a non-linear autoregressive model of pulmonary mechanics
- Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
- Determination of regional lung function in cystic fibrosis using electrical impedance tomography
- Development of parietal bone surrogates for parietal graft lift training
- Numerical simulation of mechanically stimulated bone remodelling
- Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
- Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
- Principle study on the signal connection at transabdominal fetal pulse oximetry
- Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
- Evaluating different approaches to identify a three parameter gas exchange model
- Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
- From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
- Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
- Improved FPGA controlled artificial vascular system for plethysmographic measurements
- Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
- Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
- Control scheme selection in human-machine- interfaces by analysis of activity signals
- Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
- Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
- Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
- Interactive monitoring system for visual respiratory biofeedback
- Development of a low-cost senor based aid for visually impaired people
- Patient assistive system for the shoulder joint
- A passive beating heart setup for interventional cardiology training