Home Biocompatible wear-resistant thick ceramic coating
Article Open Access

Biocompatible wear-resistant thick ceramic coating

Characterization of the hardness and porosity
  • Nicola Vogt , Katarzyna Wozniak , Armando Salito and Michael de Wild EMAIL logo
Published/Copyright: September 30, 2016

Abstract

Sensitisation to immunologically active elements like chromium, cobalt or nickel and debris particle due to wear are serious problems for patients with metallic implants. We tested the approach of using a hard and thick ceramic coating as a wear-resistant protection of titanium implants, avoiding those sensitisation and foreign body problems. We showed that the process parameters strongly influence the coating porosity and, as a consequence, also its hardness.

1 Introduction

Sensitisation to chromium (1%), cobalt (2.4–3%), nickel (13.1%) of patients who received endoprosthetic surgery is a growing problem [1]. The ionic release of these immunologically active elements from the implant material can lead to allergic responses of the host like eczema, swelling, sterile osteomyelitis, aseptic loosening and in worst cases also to the loss of the implant [2], [3]. Furthermore, particular debris of articulating joints can lead to wear problem in metal-on-metal [4] but also in metal-on-polymer systems [5]. CoCr alloys are used for endoprosthetic devices since the sixties [6] and today 80% of the artificial joints are made of CoCr. It cannot simply be replaced by titanium because its wear-resistivity is not sufficient [7], even though titanium is biocompatible and its mechanical strength is high enough [8]. One approach is to compensate the wear resistance by applying a hard ceramic coating onto the titanium parts, in particular onto the sliding surface.

Advanced plasma spray (PS) method allows coating of complex geometries and is not limited to small implants sizes. Coatings with thicknesses of several hundreds of microns are feasible while other coating technologies like PVD or thermal oxidation [9] are limited to few microns. Whereas commercial implant surfaces in direct contact with bone have been coated by titanium-spraying (TPS) since decades and show good long-term survival rates [10], plasma sprayed coatings of the Al-oxide- and Ti-oxide-based system are so far not used as slide bearing of implants but for research purpose [11] or photocatalytic application [12]. In this study, the morphology, hardness and the porosity of biocompatible advanced plasma spray coating consisting of a mix of oxides were investigated.

2 Material and methods

2.1 Sample production

Discs (ø25 mm, height 6 mm) made of CoCr were sandblasted with corundum and coated by advanced plasma spraying. Different process parameters were use in order to optimize the coating towards porosity and hardness.

2.2 Sample preparation

The coated samples were prepared for subsequent microscopic and materialographic characterization. The specimens were fixed in a polymeric 2-component cold embedding material. Cross sections were then produced on the TegraPol 21 grinding and polishing system with a Tegra Force 5 head from Struers. SiC grinding paper from BUEHLER followed by polishing paper from the same manufacturer has been used.

2.3 SEM/EDX analysis

The scanning electron microscope (SEM) TM3030 Plus (acceleration voltage U0 = 15 kV, backscattered electron and secondary electron detectors, Hitachi) with the integrated energy dispersive X-ray spectroscopy (EDX) system (Quantax 70, Bruker) enables the materialographic analysis.

2.4 Roughness analysis

The FORM Talysurf PGI 800 device from Taylor Hobson, equipped with a standard 2 μm, 90° diamond tip, a cut-off λC = 2.5 mm/0.08 mm and a measuring length l = 15 mm has been used for roughness determination.

2.5 Analysis of the porosity

The porosity of the coating was optically determined in polished cross sections according to ASTM F1854 [13] with light microscope Axiovert 40 MAT ZEISS, AxioCam MRc CCD camera, Epiplan x10/0.20HD ocular, AxioVision 40 V.4.8.2.0 Software used for image analysis. Five microscopic images of every sample were analysed (np = 5).

2.6 XRD analysis

The crystallographic phases were characterized with the diffractometer D2 PHASER (Bruker) and the Diffrac EVA software V.4.0, applying Co-Kα radiation (λCo = 1.79 Å), angular θ-range 20–120°, power 300 W, sample φ-rotation of 15°/min. The ICDD database served for the identification of the phases.

2.7 Hardness testing

The hardness of the coating was determined by the Zwick Roell ZHV 10 system at a load of 2.94 N and a holding time of 15 s. Four spots were measured for each specimen (nh = 4).

3 Results

3.1 SEM/EDX analysis

While the native PS surface shows a rough morphology, see Figure 1A, the polished surface is smooth and only a few pores can be seen, see Figure 1B. The coating homogeneously consists of titanium, aluminium and oxygen, see EDX mapping in Figure 1C. No foreign elements were detected. The elementary composition is given in Table 1.

Figure 1: SEM image of PS coating (A) before and (B) after polishing. (C) EDX mapping of the native surface with elements: Ti (pink), Al (yellow) and O (red).
Figure 1:

SEM image of PS coating (A) before and (B) after polishing. (C) EDX mapping of the native surface with elements: Ti (pink), Al (yellow) and O (red).

Table 1:

Elementary composition of the PS coating determined by EDX.

Element[at.%]
Ti49.9 ± 0.2
Al7.9 ± 0.1
O42.2 ± 0.2

The cross section through the coating is shown in Figure 2. The coating in tight contact with the rough substrate shows a layered microstructure originating from the deposition process and a low porosity. A thickness of about 100 μm was estimated.

Figure 2: Cross section of the PS coating on a metallic substrate.
Figure 2:

Cross section of the PS coating on a metallic substrate.

3.2 Roughness analysis

The roughness values of the native PS coating and the polished specimen are shown in Table 2.

Table 2:

Roughness of the native and polished PS coating.

RoughnessNative [μm] (n=9)Polished [μm] (n=9)
Ra8.15 ± 2.070.016 ± 0.005
Rt65.54 ± 13.980.67 ± 0.33
Rz53.72 ± 10.750.13 ± 0.04

3.3 XRD analysis

The crystallographic phases of the coating were identified as different modifications of TiO2 (Brookite, Anatase, Rutile) and Al2O3. No signals from the metallic substrate were observed.

3.4 Hardness testing

The impressions of the pyramidal diamond tip had a diagonal of 26.9 ± 3.3 μm, see Figure 3. The values of the Vickers hardness are quantified in Figure 4.

Figure 3: SEM image of hardness impression in PS cross section.
Figure 3:

SEM image of hardness impression in PS cross section.

Figure 4: Comparison between hardness and porosity of the ceramic coating (np = 5, nh = 4).
Figure 4:

Comparison between hardness and porosity of the ceramic coating (np = 5, nh = 4).

3.5 Analysis of the porosity

The optical quantification of the porosity and its relation to the observed hardness is illustrated in Figure 4. Porosity values between 0.43% and 12.95% were found, depending on the applied process parameters.

4 Conclusion

The investigation of the hard plasma spray process coatings indicate a trend between the hardness and the porosity as expected. According to [14] the hardness is related to the abrasive wear resistance. The quality of the coating can be optimized using specific process parameters to reduce surface roughness and porosity while simultaneously increasing the hardness, both important prerequisites to prevent friction. The proven feasibility of polishing the hard coating to surface roughness in the sub-micrometer range and expose a low-porous sliding surface opens perspectives for new biomedical applications. This novel biocompatible hard-coating could enhance the wear resistivity of titanium implants that are free from Co, Cr and Ni.

Author’s Statement

Research funding: This study was supported by the Hightech Zentrum Aargau AG. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The conducted research is not related to either human or animal use.

References

[1] Schafer T, Bohler E, Ruhdorfer S, Weigl L, Wessner D, Filipiak B, et al. Epidemiology of contact allergy in adults. Allergy. 2001;56:1192–6.10.1034/j.1398-9995.2001.00086.xSearch in Google Scholar

[2] Pacheco KA. Allergy to surgical implants. J Allergy Clin Immunol Pract. 2015;3:683–95.10.1016/j.jaip.2015.07.011Search in Google Scholar

[3] Guenther D, Thomas P, Kendoff D, Omar M, Gehrke T, Haasper C. Allergic reactions in arthroplasty: myth or serious problem? Int Orthop. 2016;40:239–44.10.1007/s00264-015-3001-6Search in Google Scholar

[4] Vendittoli PA, Roy A, Mottard S, Girard J, Lusignan D, Lavigne M. Metal ion release from bearing wear and corrosion with 28 mm and large-diameter metal-on-metal bearing articulations: a follow-up study. J Bone Joint Surg Br. 2010;92:12–9.10.1302/0301-620X.92B1.22226Search in Google Scholar

[5] Stahnke JT, Sharpe KP. Pseudotumor formation in a metal-on-polyethylene total hip arthroplasty due to trunnionosis at the head-neck taper. Surg Technol Int. 2015;27:245–50.Search in Google Scholar

[6] Morscher EW. Endoprosthetics, Springer Verlag 1995.10.1007/978-3-642-79306-6Search in Google Scholar

[7] Budinski KG. Tribological properties of titanium alloys. Wear. 1991;151:203–17.10.1016/0043-1648(91)90249-TSearch in Google Scholar

[8] ASTM F67 – 13, Standard Specification for Unalloyed Titanium, for Surgical Implant Applications, 2013.Search in Google Scholar

[9] Wen M, Wen C, Hodgson P, Li Y. Improvement of the biomedical properties of titanium using SMAT and thermal oxidation. Colloids Surf B Biointerfaces. 2014;116:658–65.10.1016/j.colsurfb.2013.10.039Search in Google Scholar PubMed

[10] Becker ST, Beck-Broichsitter BE, Rossmann CM, Behrens E, Jochens A, Wiltfang J. Long-term survival of Straumann dental implants with TPS surfaces: a retrospective study with a follow-up of 12 to 23 years. Clin Implant Dent Relat Res. 2015; doi: 10.1111/cid.12334.Search in Google Scholar PubMed

[11] Wang Y, Tian W, Yang Y. Preparation and characterization of rare earth modified nanocrystalline Al2O3/13 wt%TiO2 feedstock for plasma spraying. J Nanosci Nanotechnol. 2009;9:1445–8.10.1166/jnn.2009.C175Search in Google Scholar

[12] Stengl V, Ageorges H, Ctibor P, Murafa N. Atmospheric plasma sprayed (APS) coatings of Al2O3-TiO2 system for photocatalytic application. Photochem Photobiol Sci. 2009;8:733–8.10.1039/b817199hSearch in Google Scholar PubMed

[13] ASTM F1854-15, Standard Test Method for Stereological Evaluation of Porous Coatings on Medical Implants, 2015.Search in Google Scholar

[14] Härteprüfverfahren nach Vickers DIN EN 843-4:2005.Search in Google Scholar

Published Online: 2016-9-30
Published in Print: 2016-9-1

©2016 Michael de Wild et al., licensee De Gruyter.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.

Articles in the same Issue

  1. Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
  2. Novel blood protein based scaffolds for cardiovascular tissue engineering
  3. Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
  4. Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
  5. Long-term stable modifications of silicone elastomer for improved hemocompatibility
  6. The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
  7. Biocompatible wear-resistant thick ceramic coating
  8. Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
  9. Examination of dielectric strength of thin Parylene C films under various conditions
  10. Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
  11. Systemic analysis about residual chloroform in PLLA films
  12. A macrophage model of osseointegration
  13. Towards in silico prognosis using big data
  14. Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
  15. Usability evaluation of a locomotor therapy device considering different strategies
  16. Hypoxia-on-a-chip
  17. Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
  18. Fabrication of MEMS-based 3D-μECoG-MEAs
  19. High speed digital interfacing for a neural data acquisition system
  20. Bionic forceps for the handling of sensitive tissue
  21. Experimental studies on 3D printing of barium titanate ceramics for medical applications
  22. Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
  23. 3D printing – a key technology for tailored biomedical cell culture lab ware
  24. 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
  25. Biocompatibility of photopolymers for additive manufacturing
  26. Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
  27. Novel wireless measurement system of pressure dedicated to in vivo studies
  28. Portable auricular device for real-time swallow and chew detection
  29. Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
  30. Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
  31. Deep brain stimulation: increasing efficiency by alternative waveforms
  32. Prediction of immediately occurring microsleep events from brain electric signals
  33. Determining cardiac vagal threshold from short term heart rate complexity
  34. Classification of cardiac excitation patterns during atrial fibrillation
  35. An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
  36. Deriving respiration from high resolution 12-channel-ECG during cycling exercise
  37. Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
  38. Automatic detection and mapping of double potentials in intracardiac electrograms
  39. Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
  40. Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
  41. Best practice: surgeon driven application in pelvic operations
  42. Vasomotor assessment by camera-based photoplethysmography
  43. Classification of morphologic changes in photoplethysmographic waveforms
  44. Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
  45. Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
  46. Nonlinear causal influences assessed by mutual compression entropy
  47. Comparative study of methods for solving the correspondence problem in EMD applications
  48. fNIRS for future use in auditory diagnostics
  49. Semi-automated detection of fractional shortening in zebrafish embryo heart videos
  50. Blood pressure measurement on the cheek
  51. Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
  52. Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
  53. An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
  54. Real-time QRS detection using integrated variance for ECG gated cardiac MRI
  55. Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
  56. In-vitro experiments to characterize ventricular electromechanics
  57. Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
  58. Application of microwave sensor technology in cardiovascular disease for plaque detection
  59. Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
  60. Detection of microsleep events in a car driving simulation study using electrocardiographic features
  61. A method to determine the kink resistance of stents and stent delivery systems according to international standards
  62. Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
  63. Transient Euler-Lagrange/DEM simulation of stent thrombosis
  64. Automated control of the laser welding process of heart valve scaffolds
  65. Automation of a test bench for accessing the bendability of electrospun vascular grafts
  66. Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
  67. Cryopreservation of cells using defined serum-free cryoprotective agents
  68. New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
  69. Determination of the membrane hydraulic permeability of MSCs
  70. Climate retainment in carbon dioxide incubators
  71. Multiple factors influencing OR ventilation system effectiveness
  72. Evaluation of an app-based stress protocol
  73. Medication process in Styrian hospitals
  74. Control tower to surgical theater
  75. Development of a skull phantom for the assessment of implant X-ray visibility
  76. Surgical navigation with QR codes
  77. Investigation of the pressure gradient of embolic protection devices
  78. Computer assistance in femoral derotation osteotomy: a bottom-up approach
  79. Automatic depth scanning system for 3D infrared thermography
  80. A service for monitoring the quality of intraoperative cone beam CT images
  81. Resectoscope with an easy to use twist mechanism for improved handling
  82. In vitro simulation of distribution processes following intramuscular injection
  83. Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
  84. A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
  85. Smart control for functional electrical stimulation with optimal pulse intensity
  86. Tactile display on the remaining hand for unilateral hand amputees
  87. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
  88. An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
  89. Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
  90. Learning discriminative classification models for grading anal intraepithelial neoplasia
  91. Regularization of EIT reconstruction based on multi-scales wavelet transforms
  92. Assessing MRI susceptibility artefact through an indicator of image distortion
  93. EyeGuidance – a computer controlled system to guide eye movements
  94. A framework for feedback-based segmentation of 3D image stacks
  95. Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
  96. 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
  97. Inside-Out access strategy using new trans-vascular catheter approach
  98. US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
  99. Impact of different registration methods in MEG source analysis
  100. 3D segmentation of thyroid ultrasound images using active contours
  101. Designing a compact MRI motion phantom
  102. Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
  103. Classification of indirect immunofluorescence images using thresholded local binary count features
  104. Analysis of muscle fatigue conditions using time-frequency images and GLCM features
  105. Numerical evaluation of image parameters of ETR-1
  106. Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
  107. Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
  108. Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
  109. Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
  110. Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
  111. Evaluation of open-source software for the lung segmentation
  112. Automatic determination of lung features of CF patients in CT scans
  113. Image analysis of self-organized multicellular patterns
  114. Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
  115. Radiopacity assessment of neurovascular implants
  116. Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
  117. Development of an artifact-free aneurysm clip
  118. Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
  119. Rapid prototyping of replica knee implants for in vitro testing
  120. Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
  121. Advanced wettability analysis of implant surfaces
  122. Patient-specific hip prostheses designed by surgeons
  123. Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
  124. Wear of a total intervertebral disc prosthesis
  125. Digital health and digital biomarkers – enabling value chains on health data
  126. Usability in the lifecycle of medical software development
  127. Influence of different test gases in a non-destructive 100% quality control system for medical devices
  128. Device development guided by user satisfaction survey on auricular vagus nerve stimulation
  129. Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
  130. Effect of left atrial hypertrophy on P-wave morphology in a computational model
  131. Simulation of intracardiac electrograms around acute ablation lesions
  132. Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
  133. Assessment of nasal resistance using computational fluid dynamics
  134. Resistance in a non-linear autoregressive model of pulmonary mechanics
  135. Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
  136. Determination of regional lung function in cystic fibrosis using electrical impedance tomography
  137. Development of parietal bone surrogates for parietal graft lift training
  138. Numerical simulation of mechanically stimulated bone remodelling
  139. Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
  140. Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
  141. Principle study on the signal connection at transabdominal fetal pulse oximetry
  142. Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
  143. Evaluating different approaches to identify a three parameter gas exchange model
  144. Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
  145. From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
  146. Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
  147. Improved FPGA controlled artificial vascular system for plethysmographic measurements
  148. Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
  149. Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
  150. Control scheme selection in human-machine- interfaces by analysis of activity signals
  151. Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
  152. Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
  153. Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
  154. Interactive monitoring system for visual respiratory biofeedback
  155. Development of a low-cost senor based aid for visually impaired people
  156. Patient assistive system for the shoulder joint
  157. A passive beating heart setup for interventional cardiology training
Downloaded on 10.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2016-0010/html
Scroll to top button