Home Medicine Development of a low-cost senor based aid for visually impaired people
Article Open Access

Development of a low-cost senor based aid for visually impaired people

  • Nora Loepthien EMAIL logo , Tanja Jehnichen , Josephine Hauser , Benjamin Schullcke and Knut Möller
Published/Copyright: September 30, 2016

Abstract

The aim of the project is the development of an aid for blind or visually impaired people, considering economic aspects as well as easy adaptability to various daily situations. Distance sensors were attached to a walking frame (rollator) to detect the distance to obstacles. The information from the sensors is transmitted to the user via tactile feedback. This is realized with a number of vibration motors which were located at the upper belly area of the subject. To test the functionality of the aid to the blind, a testing track with obstacles has been passed through by a number of volunteers. While passing the track five times the needed time to pass through, as well as the number of collisions, were noticed. The results showed a decline in the average time needed to pass though the testing track. This indicates a learning process of the operator to interpret the signals given by the tactile feedback.

1 Introduction

In Germany, about 645,000 people have sight disorders or are blind [1]. Various aids exist to improve their environmental perception, such as guide dogs or white canes. However, these have serious disadvantages or are quite expensive. The white cane, for example, is the most common device. It is one of the cheapest aids and easy to use and to obtain. However, it is not possible to detect any objects above the hip and one hand is permanently holding the cane, making it unsuitable for elderly people who need additional support to maintain balance or stability while walking [2].

The second generally known aid is the guide dog. High training and running expenses, pet allergies and the work that comes with a dog might prevent many blind people from using a guide dog, especially elderly.

As a result of the democratic change the elderly population will grow. There are various diseases that affect particularly elderly people in their vision. An aid to the blind that meets the needs of an elderly user is needed to ensure a safe participation in daily life.

The objective of this work is to demonstrate how common rollators can be enhanced with low-cost hardware to support elderly people with sight disorders. This should give the possibility to hold on to something and provides a more secure feeling for the user. Therefore, a rollator was equipped with a sensor system, measuring the distance to obstacles in surrounding environment. As Johnson and Higgs did in their work [3], a vibrotactile feedback system, realized with vibration motors, was used to pass environmental information to the user. This enables the users to still use their hearing sense to observe the surroundings [4]. In order to ensure high safety for the user, the system needs to be able to prevent collisions with objects in all relevant heights. The goal is to regain more independence and a save feeling for many blind people when they are outside of well-known surroundings by using the aid.

2 Methods

2.1 Control unit

Arduino is an open-source physical computing platform, consisting of a microcontroller with several I/O ports and a software component. To control the sensors and the vibration motors an Arduino Uno board is used. This type consists of 14 digital in and outputs operating at 5 V. Furthermore, there are six inputs for analogue voltage. This enables to directly connect five sensors and corresponding motors to the Arduino board.

The Arduino Uno embeds a microcontroller, providing a software to analyse the distance of an object and to control the sensors and vibration motors.

2.2 Ultrasound sensor

The principal feasibility is demonstrated using the HC-sr04 ultrasound sensor (SRF-04, Devantech Limited, England). The measuring angle of this device is 30°, which enables a good coverage with only few sensors, a distances range between 2 and 400 cm, with an accuracy of tree millimetres is considered as appropriate for the application. The power supply for the sensor is 5 V DC, making it directly applicable in combination with an Arduino Uno board. To detect an object, the HC-sr04 sends out an ultrasound impulse which is reflected by an object and then received back by the sensor. With the measured time between sending and receiving the signal, the distance d can be calculated:

d=t/2344ms

with t being the time between sending and receiving of the ultrasound pulse. The time t is measured with the internal timer of the microcontroller of the Arduino Uno platform. In the application one measuring interval is set to 0.02 s, resulting in 50 measurements per second.

2.3 Vibration motors

Vibration motors are used to transmit the information collected by the sensors to the user. The used mini vibration motors operate within an operating voltage of 2.2 V DC–3.6 V DC, thus they can be used with the Arduino Uno board without an extra power supply.

Different types of vibrations were used to signalize the distance to an object. At a distance <80 cm the corresponding motor vibrates for one second with a break of 0.02 s. As an object enters the 50 cm radius, the vibration is two seconds long with a 0.02 s break between them. Permanent vibration is used for obstacles with a distance <20 cm from the user. Each motor exclusively reacts to the linked sensor.

2.4 Complete setup

The developed aid for the blind consists of five sensors and five vibration motors. In the software each motor is directly linked to one specific sensor. The sensors are attached to a walking chair with a special developed fixture. This basic structure enables stable measured data. Three sensors are facing forward and two are additionally covering some parts of the sides (see Figure 1). The vibration motors are situated corresponding to the sensors at the upper belly area (see Figure 2). The first and fifth motor are each facing lateral, the third is in a medial position and the second and fourth are in between, which leaves a gap of approximately 10 cm between the motors. The measurements of the perception of the waist by van Erp [5] suggests, that the number of motors could be increased to enable a more accurate transmission of the information. To supply the Arduino Uno board with power, a nine-volt battery was used.

Figure 1 Top view of the walker with the registration area of the sensors.
Figure 1

Top view of the walker with the registration area of the sensors.

Figure 2 Complete set up of the aid for the blind; walker with the sensors and the attached vibration motors.
Figure 2

Complete set up of the aid for the blind; walker with the sensors and the attached vibration motors.

Figure 3 Set up of the testing track.
Figure 3

Set up of the testing track.

2.5 Experimental setup and study protocol

On a testing track with a length of 10 m, five obstacles of different sizes were positioned (see Figure 3). All the objects were made out of cardboard and about 70 cm high. 18 volunteers (10 males, average age of 23.06 years) had to pass the test track.

All volunteers did not have any sight disorder. Therefore, all had to wear a blindfold. The test persons had not seen the track beforehand. Each test person had to pass the testing track five times. For the first four trails the track has not been changed. The constellation of the objects has been changed on the fifth test to see, if the test persons had learned how to use the aid for the blind. For every test the time and number of collisions was taken.

3 Results

Figure 4 depicts the volunteers average time for to the different test numbers. The average time needed to pass though the test track for the first time was 2.02 min. The second time the test persons needed 1.26 min, the third time 1.29 min and the fourth time 1.06 min. When the arrangement of the obstacles changed, a slight increase of the average time occurred with 1.21 min. As shown in the diagram the average time decreases. The variance from the second, fourth and fifth tests gets smaller in comparison to the first one. The high variance of the third test is caused by one single volunteer.

The average collision rate of male and female test persons is depicted in Table 1. Female volunteers have a slightly higher rate. In the first test the collision rate was the highest, but no clear reduction can be seen.

Figure 4 Average time per test: Track remained unchanged for the first four trials. Last trial was performed with changed position of obstacles.
Figure 4

Average time per test: Track remained unchanged for the first four trials. Last trial was performed with changed position of obstacles.

Table 1

Average rate of collision per test divided in female and male test persons.

Test12345Ø
Male1.2000.3000.9000.5000.7000.720
Variance1.3980.4831.8530.7071.0591.196
Female1.7501.2500.8751.0000.8751.150
Variance1.3891.6690.9910.9260.8351.189

4 Discussion

Between the various test persons there are considerably differences in the needed time and number of collisions.

The decrease of the average time and variance after the first run through suggests a learning effect. During the first test the test persons quickly get used to the functionality of the aid for the blind. However, because all the participating test persons were in their early twenties, it is not clear how this test result can be transferred to elderly users. The learning time might take considerably longer. There is also a chance that the number of vibration motors overstrains elderly users. To get more specific predictions more test with people who have sight disorders and a wider range of ages are necessary.

The majority of test persons showed difficulties passing through narrow points. Most test persons were overstrained by the numerous vibration motors that signalised an object. It seemed to be difficult to give priority to one or two of the vibration motors if all five are alarming the user. To avoid that many or all vibration motors are activated at the same time, the range of the side sensors has to be decreased. To be able to react in time, the sensors in running direction have to alarm the user way ahead of an object. The sensors that detect objects on the side of the walker only need to alarm the user if they are running at the risk of scratching them.

Some of the test persons thought that a more accurate distinction of the distances would make the orientation for the user easier. It is questionable if a more distinctive differentiation makes it easier to locate an object or if it overstrains the user. The differences of the breaks between vibrations would get less noticeable and therefore the distinction more difficult. Maybe the number of differentiations can be increased as the user gets more secure with the aid of the blind and has a better feeling for the vibrations and their meaning.

To enable the user to react to different situations, there needs to be a collection of programs to select from. For example, there needs to be a program for situations that only enables a slower movement with many objects close-by. In crowds it is important, that the sensors do not react too far in advance. On the other hand, to enable a fast walking speed the sensors have to react at an earlier stage to give the user enough time to react to obstacles.

To reach the aim of a whole body protection there have to be more sensors in different angles to recognise objects in different highs. The testing showed that a sensor based aid for the blind can work. However, there are a lot of improvements, alterations and enhancements that need to take place, before a safely daily use for people with sight disorders is possible.

Author’s Statement

Research funding: This work was supported by the Bundesministerium für Bildung und Forschung BMBF under grant 01PL11008 (TREFFER). Conflict of interest: Authors state no conflict of interest. Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use complies with all the relevant national regulations, institutional policies and was performed in accordance with the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.

References

[1] Knauer C, Pfeiffer N. Erblindung in Deutschland - heute und in 2030. s.l.: Springer-Verlag; 2006.10.1007/s00347-006-1411-ySearch in Google Scholar PubMed

[2] Möller K, Möller J, Arras KO, Bach M, Schumann S, Guttmann J. Enhanced perception for visually impaired people evaluated in a real time setting. In: Dössel O, Schlegel WC, editors. World Congress on Medical Physics and Biomedical Engineering. Munich, Germany: Springer; 2009;25/4. p. 283–6.10.1007/978-3-642-03889-1_76Search in Google Scholar

[3] Johnson LA, Higgins CM. A navigation aid for the blind using tactile-visual sensory substitution. Conf Proc IEEE Eng Med Biol Soc. 2006;1:6289–92.10.1109/IEMBS.2006.259473Search in Google Scholar PubMed

[4] Dakopoulos D, Bourbakis NG. Wearable obstacle avoidance electronic travel aids for blind: a survey. Systems, Man, and Cybernetics, Part C: Applications and Reviews, IEEE Transactions on; 2010;40.10.1109/TSMCC.2009.2021255Search in Google Scholar

[5] Van Erp JBF. Vibrotactile spatial acuity on the torso: effects of location and timing parameters. In: Eurohaptics Conference, 2005 and Symposium on Haptic Interfaces for Virtual Environment and Teleoperator Systems, 2005. World Haptics 2005. First Joint; 2005. p. 80–5.Search in Google Scholar

Published Online: 2016-9-30
Published in Print: 2016-9-1

©2016 Nora Loepthien et al., licensee De Gruyter.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.

Articles in the same Issue

  1. Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
  2. Novel blood protein based scaffolds for cardiovascular tissue engineering
  3. Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
  4. Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
  5. Long-term stable modifications of silicone elastomer for improved hemocompatibility
  6. The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
  7. Biocompatible wear-resistant thick ceramic coating
  8. Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
  9. Examination of dielectric strength of thin Parylene C films under various conditions
  10. Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
  11. Systemic analysis about residual chloroform in PLLA films
  12. A macrophage model of osseointegration
  13. Towards in silico prognosis using big data
  14. Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
  15. Usability evaluation of a locomotor therapy device considering different strategies
  16. Hypoxia-on-a-chip
  17. Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
  18. Fabrication of MEMS-based 3D-μECoG-MEAs
  19. High speed digital interfacing for a neural data acquisition system
  20. Bionic forceps for the handling of sensitive tissue
  21. Experimental studies on 3D printing of barium titanate ceramics for medical applications
  22. Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
  23. 3D printing – a key technology for tailored biomedical cell culture lab ware
  24. 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
  25. Biocompatibility of photopolymers for additive manufacturing
  26. Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
  27. Novel wireless measurement system of pressure dedicated to in vivo studies
  28. Portable auricular device for real-time swallow and chew detection
  29. Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
  30. Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
  31. Deep brain stimulation: increasing efficiency by alternative waveforms
  32. Prediction of immediately occurring microsleep events from brain electric signals
  33. Determining cardiac vagal threshold from short term heart rate complexity
  34. Classification of cardiac excitation patterns during atrial fibrillation
  35. An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
  36. Deriving respiration from high resolution 12-channel-ECG during cycling exercise
  37. Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
  38. Automatic detection and mapping of double potentials in intracardiac electrograms
  39. Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
  40. Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
  41. Best practice: surgeon driven application in pelvic operations
  42. Vasomotor assessment by camera-based photoplethysmography
  43. Classification of morphologic changes in photoplethysmographic waveforms
  44. Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
  45. Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
  46. Nonlinear causal influences assessed by mutual compression entropy
  47. Comparative study of methods for solving the correspondence problem in EMD applications
  48. fNIRS for future use in auditory diagnostics
  49. Semi-automated detection of fractional shortening in zebrafish embryo heart videos
  50. Blood pressure measurement on the cheek
  51. Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
  52. Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
  53. An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
  54. Real-time QRS detection using integrated variance for ECG gated cardiac MRI
  55. Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
  56. In-vitro experiments to characterize ventricular electromechanics
  57. Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
  58. Application of microwave sensor technology in cardiovascular disease for plaque detection
  59. Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
  60. Detection of microsleep events in a car driving simulation study using electrocardiographic features
  61. A method to determine the kink resistance of stents and stent delivery systems according to international standards
  62. Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
  63. Transient Euler-Lagrange/DEM simulation of stent thrombosis
  64. Automated control of the laser welding process of heart valve scaffolds
  65. Automation of a test bench for accessing the bendability of electrospun vascular grafts
  66. Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
  67. Cryopreservation of cells using defined serum-free cryoprotective agents
  68. New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
  69. Determination of the membrane hydraulic permeability of MSCs
  70. Climate retainment in carbon dioxide incubators
  71. Multiple factors influencing OR ventilation system effectiveness
  72. Evaluation of an app-based stress protocol
  73. Medication process in Styrian hospitals
  74. Control tower to surgical theater
  75. Development of a skull phantom for the assessment of implant X-ray visibility
  76. Surgical navigation with QR codes
  77. Investigation of the pressure gradient of embolic protection devices
  78. Computer assistance in femoral derotation osteotomy: a bottom-up approach
  79. Automatic depth scanning system for 3D infrared thermography
  80. A service for monitoring the quality of intraoperative cone beam CT images
  81. Resectoscope with an easy to use twist mechanism for improved handling
  82. In vitro simulation of distribution processes following intramuscular injection
  83. Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
  84. A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
  85. Smart control for functional electrical stimulation with optimal pulse intensity
  86. Tactile display on the remaining hand for unilateral hand amputees
  87. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
  88. An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
  89. Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
  90. Learning discriminative classification models for grading anal intraepithelial neoplasia
  91. Regularization of EIT reconstruction based on multi-scales wavelet transforms
  92. Assessing MRI susceptibility artefact through an indicator of image distortion
  93. EyeGuidance – a computer controlled system to guide eye movements
  94. A framework for feedback-based segmentation of 3D image stacks
  95. Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
  96. 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
  97. Inside-Out access strategy using new trans-vascular catheter approach
  98. US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
  99. Impact of different registration methods in MEG source analysis
  100. 3D segmentation of thyroid ultrasound images using active contours
  101. Designing a compact MRI motion phantom
  102. Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
  103. Classification of indirect immunofluorescence images using thresholded local binary count features
  104. Analysis of muscle fatigue conditions using time-frequency images and GLCM features
  105. Numerical evaluation of image parameters of ETR-1
  106. Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
  107. Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
  108. Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
  109. Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
  110. Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
  111. Evaluation of open-source software for the lung segmentation
  112. Automatic determination of lung features of CF patients in CT scans
  113. Image analysis of self-organized multicellular patterns
  114. Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
  115. Radiopacity assessment of neurovascular implants
  116. Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
  117. Development of an artifact-free aneurysm clip
  118. Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
  119. Rapid prototyping of replica knee implants for in vitro testing
  120. Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
  121. Advanced wettability analysis of implant surfaces
  122. Patient-specific hip prostheses designed by surgeons
  123. Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
  124. Wear of a total intervertebral disc prosthesis
  125. Digital health and digital biomarkers – enabling value chains on health data
  126. Usability in the lifecycle of medical software development
  127. Influence of different test gases in a non-destructive 100% quality control system for medical devices
  128. Device development guided by user satisfaction survey on auricular vagus nerve stimulation
  129. Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
  130. Effect of left atrial hypertrophy on P-wave morphology in a computational model
  131. Simulation of intracardiac electrograms around acute ablation lesions
  132. Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
  133. Assessment of nasal resistance using computational fluid dynamics
  134. Resistance in a non-linear autoregressive model of pulmonary mechanics
  135. Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
  136. Determination of regional lung function in cystic fibrosis using electrical impedance tomography
  137. Development of parietal bone surrogates for parietal graft lift training
  138. Numerical simulation of mechanically stimulated bone remodelling
  139. Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
  140. Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
  141. Principle study on the signal connection at transabdominal fetal pulse oximetry
  142. Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
  143. Evaluating different approaches to identify a three parameter gas exchange model
  144. Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
  145. From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
  146. Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
  147. Improved FPGA controlled artificial vascular system for plethysmographic measurements
  148. Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
  149. Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
  150. Control scheme selection in human-machine- interfaces by analysis of activity signals
  151. Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
  152. Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
  153. Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
  154. Interactive monitoring system for visual respiratory biofeedback
  155. Development of a low-cost senor based aid for visually impaired people
  156. Patient assistive system for the shoulder joint
  157. A passive beating heart setup for interventional cardiology training
Downloaded on 6.3.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2016-0158/html
Scroll to top button