Abstract
The estimation of respiratory rates from contineous respiratory signals is commonly done using either fourier transformation or the zero-crossing method. This paper introduces another method which is based on the autocorrelation function of the respiratory signal. The respiratory signals can be measured either directly using a flow sensor or chest strap or indirectly on the basis of the electrocardiogram (ECG). We compare our method against other established methods on the basis of real-world ECG signals and use a respiration-based breathing frequency as a reference. Our method achieved the best agreement between respiration rates derived from directly and indirectly measured respiratory signals.
1 Introduction
The continuous measurement of the respiratory rate is particularly important in the medical environment as well as in the area of sports applications. Here, the most widely used approach is the measurement of the movements due to breathing using a chest strap or an impedance-based pneumographic device [1], [2]. However, there are also algorithms that derive the respiratory frequency noninvasively from other biosignals. It has been shown that the amplitude of the QRS-complex is modulated by chest movements during breathing, which results in a changing R-peak amplitude that correlates highly with the actual respiratory signal (taken from a chest strap or a flow sensor) [3]. From this signal one can estimate the breathing rate using a great variety of algorithms. The most commonly used methods are the Fast-Fourier Transform (FFT-method) and the zero-crossing method. The former estimates the rate from the strongest frequency component in the respiration signal, whereas the latter is based on determining the average duration of each breath [3], [4].
Interferences in the estimated respiration signal can lead to deviations in the estimated respiration rate. In this paper we propose a new method that allows a less interference-prone estimation of the respiratory rate. We applied the method to respiration signals derived both from ECG and a chest strap and evaluated the results in comparison to FFT and zero-crossing based respiration rate estimation methods.
2 Methods
2.1 Estimation of the respiratory rate
As a first step the respiratory signal is extracted from the ECG-signal sampled with the sampling frequency fs. For this purpose, a Hamilton-Tompkins algorithm [5] is used to detect all QRS-complexes in the ECG signal (comp. Figure 1). Next, the amplitude of the R-to-S peak is computed, which provides a robust estimation of the signal amplitude with respect to the baseline [6]. These extracted amplitudes are then interpolated by means of cubic splines to obtain the respiratory signal with the same sampling frequency as the ECG signal. An example for an extracted respiratory signal can be seen in Figure 1. Aside from a phase shift there is a high correlation between the extracted and the reference signal.

Top: baseline-corrected ECG signal with marked R- and S-peaks for the estimation of the R-to-S amplitude; bottom: reference respiratory signal (blue) compared to ECG-derived respiratory signal (orange). Amplitudes are normalized, arbitrary units.
In order to estimate the respiration rate, the fundamental frequency of the extracted respiratory signal has to be calculated. As mentioned above, this can be done using methods like the FFT in the frequency domain [4] or the zero-crossing method in the time domain [3]. In this paper we present a different approach which is based on the autocorrelation function of the estimated respiratory signal. The autocorrelation function Ψ(i) of the respiratory signal xresp is defined in Eq. 1 [7]:
Here, N and i denote the number of samples and the displacement of the signal towards itself, respectively. Generally, the autocorrelation function describes the similarity of a signal to a shifted version of itself. In the case of periodic signals, this function shows periodic maxima. The displacements i at which those maxima occur correspond to integer multiples of the fundamental period of the signal (here, i.e. the respiratory signal). Employing the Wiener-Khinchin theorem [8] allows an efficient calculation of Eq. 1 on the basis of an FFT. Next, the intervals ti of all neighbouring local maxima of the autocorrelation function are determined. Disturbances in the extracted respiratory signal or its non-stationarity may cause additional local maxima. Figure 2 shows one such example. Here it can be seen that high-frequency disturbances of a respiratory signal lead to local maxima (denoted in red) in the autocorrelation function that are unrelated to the fundamental period of the respiratory signal. To suppress the influence of these additional maxima an empirically estimated threshold is used so that only maxima exceeding this threshold were included in subsequent calculations. To further weaken the influence of highly deviating time intervals their median is usesd as the final estimate for the duration of a breath cycle. The respiratory frequency in breaths per minute (bpm) can then be calculated using Eq. 2.

Top: ECG-derived respiratory signal (example); bottom: local maxima belonging to the fundamental period (green) and local maxima caused by artifacts (red).
The first reference method to estimate respiratory rates is the FFT. It is used to calculate the spectrum of the respiratory signal. The frequency component fmax with the highest amplitude is thaken as the respiratory frequency. This frequencies can be converted to the respiratory rate in bpm using Eq. 3.
For the zero-crossing method, i.e. the second reference method, the first step is to center the signal at zero by subtracting the mean respiratory signal from each sample. After that, all zero-crossings ti are detected. The respiratory rate can finally be estimated by calculating the mean of all detected zero-crossings using Eq. 4.
2.2 Simulation setup
Simultaneous recordings of ECG and respiration were used to evaluate the method. The recordings of 13 subjects with a length of 30 min each were taken from the Combined Measurement of ECG, Breathing and Seismocardiogram (CEBS) database [9], [10]. All datasets contained an ECG-signal as well as a respiratory signal measured by a chest strap. The datasets were converted to a MATLAB-readable format using the WFDB toolbox [11]. From each dataset ECG and respiratory signal were obtained and resampled to a sampling frequency of fs = 250 Hz. Both signals were subdivided into sliding windows of length Tmeas = 30 s and an overlap of 15 s. Baseline removal was applied to the ECG-signal [12]. Furthermore, the 50 Hz powerline interference was removed using an IIR-Notch-Filter. The respiratory signal was smoothed using an FIR-lowpass-filter (fcut = 2 Hz). The method described in section 2.1 was used for the estimation of the respiratory rate by means of directly measured respiratory signals (chest strap) and ECG signals. Subsequently, the results are then compared to the ones obtained for the FFT-method and the zero-crossing method.
3 Results
Figure 3 shows the raw results of all evaluated methods. The horizontal axis represents estimated respiratory rate based on the ECG and the vertical axis represents the respiratory rate based on the chest strap signal. All methods show a high correlation between respiratory rates estimated from ECG and chest strap. Both the autocorrelation method (r = 0.86, p < 10−3) and the zero-crossing method (r = 0.88, p < 10−3) exhibit a strong linear relationship between both estimated respiratory rates whereas this relation is much weaker for the FFT-method (r = 0,6, p < 10−3). The statistical significance of the correlation was tested using a t-test. It can be seen that the reference-based respiratory rate estimated by means of the FFT-method shows discrete values. This is due to the finite frequency resolution. The ECG-based respiration frequency does not show this effect. This is due to the spline interpolation of the R-to-S peak amplitudes leading to slightly varying sample-sizes for each time window and thus to varying frequency resolutions depending on the length of the time window Tmeas.

Correlation between respiration-derived respiratory rate and ECG-derived estimated respiratory rate using the autocorellation method (A), FFT-method (B) and zero-crossing method (C).
The error between ECG- and chest-strap-based respiration rates for each method is displayed in Figure 4. A Kolmogorov-Smirnov Test was used to compare the error distibutions. It turned out that all distributions were statistically significantly different (p < 10−3). While the median of the autocorrelation method is located around zero, the median of the FFT-method lies above zero and shows a higher spread. The median of the zero-crossing method is also close to zero. However, its error distribution exhibits an asymmetric deviation from the median. Values for mean and variance of the error are shown in Table 1. It can be seen that the autocorrelation method has a smaller error in average than any of the reference methods. In contrast to the zero-crossing method there is a huge difference regarding the variance of the error between autocorrelation- and FFT-method.

Comparison of the error of respiration-based and ecg-based estimation of the respiratory rate using all methods considered. Boxes show interquartile range (IQR = Q0.75 − Q0.25) and median. Outliers are depicted as dots outside of the boxes.
Mean and variance of the error distributions shown in Figure 4.
| μ (err) | Var (err) | |
|---|---|---|
| Autocorrelation | −0.15 | 3.68 |
| FFT | −0.8 | 17.54 |
| Zero-crossing | −0.44 | 2.7 |
An inter-subject analysis emphasizes these findings. It can bee seen in Figure 5 that the error of the zero-crossing method shows a higher spread than the error of the autocorrelation method. We have omitted the FFT-method since we consider it unsuited here due to its discrete-valued results.

Error of the autocorrelation- and zero-crossing method on an inter-subject level. Boxes show interquartile range (IQR = Q0.75 − Q0.25) and median. Outliers are depicted as dots outside of the boxes.
4 Discussion
We compared ECG-derived and chest-strap-derived respiratory rates using three different methods, i.e. FFT, zero-crossing and the newly proposed autocorrelation method. It became apparent that there is a high correlation between the results of zero-crossing and the autocorrelation. Yet, the correlation between the FFT-method and the two others is considerably lower. The results also show a low variance of the error for zero-crossing and autocorrelation as opposed to the FFT-method. However, the ECG-based respiratory rate estimates by means of zero-crossing are systematically lower than the chest-based estimation. Apparently, the autocorrelation method is less prone to interferences in the respiratory signal than the other methods. These interferences can be caused by either artifacts in the ECG signal or as a result of natural changes of the breath frequency in the analysed time interval. Another disadvantage of the FFT-method is its finite frequency resolution in the time window under consideration. It is also possible that an insufficient baseline removal leads to low-frequency components in the derived respiratory signal and therefore to a larger error. This effect was reduced by calculating the amplitude of the R-peak with respect to the S-peak but can not be completely eliminated.
Algorithms used for preprocessing can also have a deteriorating effect on the estimation of the respiratory frequency. FIR-filters used to enhance the signal-to-noise ratio almost always also remove important signal components, which in turn can lead to changed R-peak amplitudes. Other techniques (i.e. wavelet filters) might help to mitigiate this effect.
We have shown that the proposed autocorrelation method is generally feasible for the estimation of the respiration rate from both ECG-based and chest strap based respiratory signals. It is applicable to the long-term monitoring of patients in a clinical environment without directly measuring the respiratory signal. This is particularly useful, e.g. during rehabilitation after a surgical intervention. The used data exhibits a rather low variability in terms of respiratory frequency. An evaluation of our method using very low and very high respiratory frequencies is necessary to further substantiate its usefulness. An extension of our method to PPG signals may also be conceivable.
Author’s Statement
Research funding: This work was funded by the Federal Ministry of Education and Research (BMBF) (FKZ 03FH032IX5). Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The conducted research is not related to either human or animal use.
References
[1] Grenvik A, Ballou S, McGinley E, Millen JE, Cooley WL, Safar P. Impedance pneumography: comparison between chest impedance changes and respiratory volumes in 11 healthy volunteers. Chest J. 1972;62:439–43.10.1378/chest.62.4.439Suche in Google Scholar
[2] Gupta AK. Respiration rate measurement basetd on impedance pneumography. Texas Instruments, Tech. Rep. February, 2011.Suche in Google Scholar
[3] Moody GB, Mark RG, Zoccola A, Mantero S. Derivation of respiratory signals from multi-lead ECGs. Compu Cardiol. 1985;12:113–6.Suche in Google Scholar
[4] Karlen W, Mattiussi C, Florean D. Sleep and wake classification with ECG and respiratory effort signals. IEEE Trans Bio-Med Circuits Syst. 2009;3:71–8.10.1109/TBCAS.2008.2008817Suche in Google Scholar
[5] Hamilton P, Tompkins WJ. Quantitative investigation of QRS detection rules using the MIT/BIH arrhythmia database. IEEE Trans Bio-Med Eng. 1986;33:1157–65.10.1109/TBME.1986.325695Suche in Google Scholar
[6] Dobrev D, Daskalov I. Two-electrode telemetric instrument for infant heart rate and apnea monitoring. Med Eng Phys. 1999;20:729–34.10.1016/S1350-4533(98)00091-5Suche in Google Scholar
[7] McLeod P, Wyvill G. A smarter way to find pitch. In: Proceedings of International Computer Music Conference, ICMC; 2005.Suche in Google Scholar
[8] Box GE, Jenkins GM, Reinsel GC. Autocorrelation function and spectrum of stationary processes. In: Time series analysis. 4th ed. John Wiley & Sons, Inc., 2013, ch. Autocorrel, p. 19–47.Suche in Google Scholar
[9] García-González M, Argelagós A, Fernández-Chimeno M, Ramos-Castro J. Differences in QRS locations due to ECG lead: relationship with breathing. In: XIII Mediterranean Conference on Medical and Biological Engineering and Computing 2013, 2014;13:962–4.10.1007/978-3-319-00846-2_238Suche in Google Scholar
[10] Goldberger AL, Amaral LA, Glass L, Hausdorff JM, Ivanov PC, Mark RG, et al. Physiobank, physiotoolkit, and physionet components of a new research resource for complex physiologic signals. Circulation. 2000;101:215–20.10.1161/01.CIR.101.23.e215Suche in Google Scholar
[11] Silva I, Moody GB. An open-source toolbox for analysing and processing physionet databases in MATLAB and Octave. J Open Res Softw. 2014;2:e27.10.5334/jors.biSuche in Google Scholar PubMed PubMed Central
[12] Zhang F, Chen S, Zhang H, Zhang X, Li G. Bioelectric signal detrending using smoothness prior approach. Med Eng Phys. 2014;36:1007–13.10.1016/j.medengphy.2014.05.009Suche in Google Scholar PubMed
©2016 Fabian Schrumpf et al., licensee De Gruyter.
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.
Artikel in diesem Heft
- Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
- Novel blood protein based scaffolds for cardiovascular tissue engineering
- Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
- Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
- Long-term stable modifications of silicone elastomer for improved hemocompatibility
- The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
- Biocompatible wear-resistant thick ceramic coating
- Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
- Examination of dielectric strength of thin Parylene C films under various conditions
- Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
- Systemic analysis about residual chloroform in PLLA films
- A macrophage model of osseointegration
- Towards in silico prognosis using big data
- Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
- Usability evaluation of a locomotor therapy device considering different strategies
- Hypoxia-on-a-chip
- Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
- Fabrication of MEMS-based 3D-μECoG-MEAs
- High speed digital interfacing for a neural data acquisition system
- Bionic forceps for the handling of sensitive tissue
- Experimental studies on 3D printing of barium titanate ceramics for medical applications
- Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
- 3D printing – a key technology for tailored biomedical cell culture lab ware
- 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
- Biocompatibility of photopolymers for additive manufacturing
- Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
- Novel wireless measurement system of pressure dedicated to in vivo studies
- Portable auricular device for real-time swallow and chew detection
- Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
- Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
- Deep brain stimulation: increasing efficiency by alternative waveforms
- Prediction of immediately occurring microsleep events from brain electric signals
- Determining cardiac vagal threshold from short term heart rate complexity
- Classification of cardiac excitation patterns during atrial fibrillation
- An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
- Deriving respiration from high resolution 12-channel-ECG during cycling exercise
- Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
- Automatic detection and mapping of double potentials in intracardiac electrograms
- Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
- Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
- Best practice: surgeon driven application in pelvic operations
- Vasomotor assessment by camera-based photoplethysmography
- Classification of morphologic changes in photoplethysmographic waveforms
- Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
- Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
- Nonlinear causal influences assessed by mutual compression entropy
- Comparative study of methods for solving the correspondence problem in EMD applications
- fNIRS for future use in auditory diagnostics
- Semi-automated detection of fractional shortening in zebrafish embryo heart videos
- Blood pressure measurement on the cheek
- Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
- Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
- An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
- Real-time QRS detection using integrated variance for ECG gated cardiac MRI
- Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
- In-vitro experiments to characterize ventricular electromechanics
- Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
- Application of microwave sensor technology in cardiovascular disease for plaque detection
- Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
- Detection of microsleep events in a car driving simulation study using electrocardiographic features
- A method to determine the kink resistance of stents and stent delivery systems according to international standards
- Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
- Transient Euler-Lagrange/DEM simulation of stent thrombosis
- Automated control of the laser welding process of heart valve scaffolds
- Automation of a test bench for accessing the bendability of electrospun vascular grafts
- Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
- Cryopreservation of cells using defined serum-free cryoprotective agents
- New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
- Determination of the membrane hydraulic permeability of MSCs
- Climate retainment in carbon dioxide incubators
- Multiple factors influencing OR ventilation system effectiveness
- Evaluation of an app-based stress protocol
- Medication process in Styrian hospitals
- Control tower to surgical theater
- Development of a skull phantom for the assessment of implant X-ray visibility
- Surgical navigation with QR codes
- Investigation of the pressure gradient of embolic protection devices
- Computer assistance in femoral derotation osteotomy: a bottom-up approach
- Automatic depth scanning system for 3D infrared thermography
- A service for monitoring the quality of intraoperative cone beam CT images
- Resectoscope with an easy to use twist mechanism for improved handling
- In vitro simulation of distribution processes following intramuscular injection
- Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
- A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
- Smart control for functional electrical stimulation with optimal pulse intensity
- Tactile display on the remaining hand for unilateral hand amputees
- Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
- An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
- Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
- Learning discriminative classification models for grading anal intraepithelial neoplasia
- Regularization of EIT reconstruction based on multi-scales wavelet transforms
- Assessing MRI susceptibility artefact through an indicator of image distortion
- EyeGuidance – a computer controlled system to guide eye movements
- A framework for feedback-based segmentation of 3D image stacks
- Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
- 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
- Inside-Out access strategy using new trans-vascular catheter approach
- US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
- Impact of different registration methods in MEG source analysis
- 3D segmentation of thyroid ultrasound images using active contours
- Designing a compact MRI motion phantom
- Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
- Classification of indirect immunofluorescence images using thresholded local binary count features
- Analysis of muscle fatigue conditions using time-frequency images and GLCM features
- Numerical evaluation of image parameters of ETR-1
- Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
- Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
- Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
- Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
- Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
- Evaluation of open-source software for the lung segmentation
- Automatic determination of lung features of CF patients in CT scans
- Image analysis of self-organized multicellular patterns
- Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
- Radiopacity assessment of neurovascular implants
- Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
- Development of an artifact-free aneurysm clip
- Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
- Rapid prototyping of replica knee implants for in vitro testing
- Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
- Advanced wettability analysis of implant surfaces
- Patient-specific hip prostheses designed by surgeons
- Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
- Wear of a total intervertebral disc prosthesis
- Digital health and digital biomarkers – enabling value chains on health data
- Usability in the lifecycle of medical software development
- Influence of different test gases in a non-destructive 100% quality control system for medical devices
- Device development guided by user satisfaction survey on auricular vagus nerve stimulation
- Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
- Effect of left atrial hypertrophy on P-wave morphology in a computational model
- Simulation of intracardiac electrograms around acute ablation lesions
- Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
- Assessment of nasal resistance using computational fluid dynamics
- Resistance in a non-linear autoregressive model of pulmonary mechanics
- Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
- Determination of regional lung function in cystic fibrosis using electrical impedance tomography
- Development of parietal bone surrogates for parietal graft lift training
- Numerical simulation of mechanically stimulated bone remodelling
- Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
- Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
- Principle study on the signal connection at transabdominal fetal pulse oximetry
- Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
- Evaluating different approaches to identify a three parameter gas exchange model
- Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
- From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
- Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
- Improved FPGA controlled artificial vascular system for plethysmographic measurements
- Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
- Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
- Control scheme selection in human-machine- interfaces by analysis of activity signals
- Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
- Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
- Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
- Interactive monitoring system for visual respiratory biofeedback
- Development of a low-cost senor based aid for visually impaired people
- Patient assistive system for the shoulder joint
- A passive beating heart setup for interventional cardiology training
Artikel in diesem Heft
- Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
- Novel blood protein based scaffolds for cardiovascular tissue engineering
- Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
- Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
- Long-term stable modifications of silicone elastomer for improved hemocompatibility
- The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
- Biocompatible wear-resistant thick ceramic coating
- Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
- Examination of dielectric strength of thin Parylene C films under various conditions
- Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
- Systemic analysis about residual chloroform in PLLA films
- A macrophage model of osseointegration
- Towards in silico prognosis using big data
- Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
- Usability evaluation of a locomotor therapy device considering different strategies
- Hypoxia-on-a-chip
- Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
- Fabrication of MEMS-based 3D-μECoG-MEAs
- High speed digital interfacing for a neural data acquisition system
- Bionic forceps for the handling of sensitive tissue
- Experimental studies on 3D printing of barium titanate ceramics for medical applications
- Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
- 3D printing – a key technology for tailored biomedical cell culture lab ware
- 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
- Biocompatibility of photopolymers for additive manufacturing
- Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
- Novel wireless measurement system of pressure dedicated to in vivo studies
- Portable auricular device for real-time swallow and chew detection
- Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
- Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
- Deep brain stimulation: increasing efficiency by alternative waveforms
- Prediction of immediately occurring microsleep events from brain electric signals
- Determining cardiac vagal threshold from short term heart rate complexity
- Classification of cardiac excitation patterns during atrial fibrillation
- An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
- Deriving respiration from high resolution 12-channel-ECG during cycling exercise
- Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
- Automatic detection and mapping of double potentials in intracardiac electrograms
- Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
- Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
- Best practice: surgeon driven application in pelvic operations
- Vasomotor assessment by camera-based photoplethysmography
- Classification of morphologic changes in photoplethysmographic waveforms
- Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
- Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
- Nonlinear causal influences assessed by mutual compression entropy
- Comparative study of methods for solving the correspondence problem in EMD applications
- fNIRS for future use in auditory diagnostics
- Semi-automated detection of fractional shortening in zebrafish embryo heart videos
- Blood pressure measurement on the cheek
- Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
- Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
- An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
- Real-time QRS detection using integrated variance for ECG gated cardiac MRI
- Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
- In-vitro experiments to characterize ventricular electromechanics
- Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
- Application of microwave sensor technology in cardiovascular disease for plaque detection
- Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
- Detection of microsleep events in a car driving simulation study using electrocardiographic features
- A method to determine the kink resistance of stents and stent delivery systems according to international standards
- Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
- Transient Euler-Lagrange/DEM simulation of stent thrombosis
- Automated control of the laser welding process of heart valve scaffolds
- Automation of a test bench for accessing the bendability of electrospun vascular grafts
- Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
- Cryopreservation of cells using defined serum-free cryoprotective agents
- New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
- Determination of the membrane hydraulic permeability of MSCs
- Climate retainment in carbon dioxide incubators
- Multiple factors influencing OR ventilation system effectiveness
- Evaluation of an app-based stress protocol
- Medication process in Styrian hospitals
- Control tower to surgical theater
- Development of a skull phantom for the assessment of implant X-ray visibility
- Surgical navigation with QR codes
- Investigation of the pressure gradient of embolic protection devices
- Computer assistance in femoral derotation osteotomy: a bottom-up approach
- Automatic depth scanning system for 3D infrared thermography
- A service for monitoring the quality of intraoperative cone beam CT images
- Resectoscope with an easy to use twist mechanism for improved handling
- In vitro simulation of distribution processes following intramuscular injection
- Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
- A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
- Smart control for functional electrical stimulation with optimal pulse intensity
- Tactile display on the remaining hand for unilateral hand amputees
- Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
- An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
- Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
- Learning discriminative classification models for grading anal intraepithelial neoplasia
- Regularization of EIT reconstruction based on multi-scales wavelet transforms
- Assessing MRI susceptibility artefact through an indicator of image distortion
- EyeGuidance – a computer controlled system to guide eye movements
- A framework for feedback-based segmentation of 3D image stacks
- Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
- 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
- Inside-Out access strategy using new trans-vascular catheter approach
- US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
- Impact of different registration methods in MEG source analysis
- 3D segmentation of thyroid ultrasound images using active contours
- Designing a compact MRI motion phantom
- Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
- Classification of indirect immunofluorescence images using thresholded local binary count features
- Analysis of muscle fatigue conditions using time-frequency images and GLCM features
- Numerical evaluation of image parameters of ETR-1
- Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
- Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
- Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
- Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
- Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
- Evaluation of open-source software for the lung segmentation
- Automatic determination of lung features of CF patients in CT scans
- Image analysis of self-organized multicellular patterns
- Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
- Radiopacity assessment of neurovascular implants
- Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
- Development of an artifact-free aneurysm clip
- Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
- Rapid prototyping of replica knee implants for in vitro testing
- Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
- Advanced wettability analysis of implant surfaces
- Patient-specific hip prostheses designed by surgeons
- Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
- Wear of a total intervertebral disc prosthesis
- Digital health and digital biomarkers – enabling value chains on health data
- Usability in the lifecycle of medical software development
- Influence of different test gases in a non-destructive 100% quality control system for medical devices
- Device development guided by user satisfaction survey on auricular vagus nerve stimulation
- Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
- Effect of left atrial hypertrophy on P-wave morphology in a computational model
- Simulation of intracardiac electrograms around acute ablation lesions
- Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
- Assessment of nasal resistance using computational fluid dynamics
- Resistance in a non-linear autoregressive model of pulmonary mechanics
- Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
- Determination of regional lung function in cystic fibrosis using electrical impedance tomography
- Development of parietal bone surrogates for parietal graft lift training
- Numerical simulation of mechanically stimulated bone remodelling
- Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
- Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
- Principle study on the signal connection at transabdominal fetal pulse oximetry
- Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
- Evaluating different approaches to identify a three parameter gas exchange model
- Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
- From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
- Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
- Improved FPGA controlled artificial vascular system for plethysmographic measurements
- Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
- Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
- Control scheme selection in human-machine- interfaces by analysis of activity signals
- Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
- Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
- Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
- Interactive monitoring system for visual respiratory biofeedback
- Development of a low-cost senor based aid for visually impaired people
- Patient assistive system for the shoulder joint
- A passive beating heart setup for interventional cardiology training