Impact of different registration methods in MEG source analysis

  • Marie Theiß EMAIL logo , Andreas Wollbrink , Carsten H. Wolters and Heinrich Brinck
Published/Copyright: September 30, 2016

Abstract

For this study the impact of different co-registration procedures on MEG source localization of somatosensory evoked fields was evaluated. Two different co-registration procedures were used to calculate the transformation matrix which specifies how to align the MRI data to the MEG head coordinate system. In order to depict the differences, caused by the method, the Euclidian distance between the reconstructed sources was noted. It was shown that, erroneous MRI and MEG data co- registration effects source localization results. Most dipoles are located more posterior and superior when the more advanced registration procedure was applied. In conclusion the results show, that an iterative matching procedure allows an accurate knowledge of the MEG gradiometer sensor position relative to the head which is crucial to correctly reconstruct neuronal activity derived from MEG measurements.

1 Introduction

Reconstruction of brain’s activity within the cerebral cortex needs an accurate registration of magnetic resonance imaging (MRI) scans and magnetoencephalography (MEG) records. MEG is a non-invasive technique for recording neuronal activity. In contrast, MRI scans provide detailed pictures of anatomical structures. The registration process describes the spatial localization of the subject’s position inside the MEG scanner. For source analysis a volume conductor head model, representing tissue geometries and conductivities, is needed. The realistic single-shell volume conduction model approximates the tissue geometry as one shell of arbitrary (here the individual brain) shape [1].

The MRI scans are usually represented in a coordinate system (voxel space) without physical dimensions and therefore the co-registration is needed to get physical coordinates (source space coordinates) for the voxel. A frequently used technique to achieve an accurate co-registration of the MRI data to the source space relies on three anatomical landmarks (AL). The small number of landmarks which are used for the registration of MEG and MRI data seems to be problematic [2], since errors in landmark determination might lead to large registration inaccuracies. Hence, we decided to digitize points on the scalp surface with a Polhemus digitizer and to perform an iterative matching (IM), which was presented as a more precise alternative [3]. The IM can be performed by the use of the iterative closest point (ICP) algorithm [Besl 1992]. It registers the MRI scans to the MEG data by minimizing the distance between a point cloud and the head surface.

We compared the source reconstruction results of somatosensory evoked responses, obtained by means of the different registration procedures, within a single shell volume conductor head model.

2 Background

In order to take into account brain geometry and volume conduction properties the MEG data need to be registered with the MRI data. Therefore a transformation matrix for the registration of the MRI scan voxel to the source space can be calculated. This is done frequently by the help of three external AL (nasion, left and right pre-auricular points), which are used to define the source space coordinates. Their positions are marked by Gadolinium markers and identified manually within the MRI scans. During the MEG recordings, they are determined by three localization coils. Co-Registration of the data sets (MEG and MRI) means, that the marker positions are used to define the source space. That way, both data sets are represented within the same coordinate system. An accurate co-registration depends on precise determination of the AL in both data sets. This registration procedure contains several possible sources of error, since the landmark digitization has an offset of several millimeters [4]. The Gadolinium markers can slip out of place during recordings. Thus their position cannot be accurately defined within the MRI scans. As a consequence, the source space coordinates might differ slightly between MEG and MRI data. Altogether, the small number of landmarks, as well as the sources of error while localization, might lead to an imprecise co-registration of the different data sets.

In order to optimize the co-registration of MEG and MRI data sets, a Polhemus Digitizer is used to record several points of the scalp surface, the region around the eyes and the nose. In this way, registration errors are reduced, since more surface points are used. The ICP algorithm is adapted to the point cloud to fit it with the MRI-segmented scalp surface. The algorithm minimizes the distance between the points of the point cloud and their nearest neighbour on the head surface.

This study determines the differences in source localization with respect to the different co-registration methods. Considering that, somatosensory evoked responses of seven healthy subjects were analyzed using a single shell head model. MEG measurements were made in a magnetically shielded room in supine position (275 channel whole head MEG with 29 reference channels; CTF, VSM, MedTech Ltd.). A square electrical pulse was used to stimulate the median nerve of the patient’s wrist. A bandpass filter of 20 250 Hz, a baseline correction and a notch filter (for line noise voltage frequency of 50 Hz and its harmonics) were used to filter the MEG data. The trials were averaged by timelockanalysis after artificial trials and bad channels were excluded from the data.

A single dipole grid search was performed to manage the inverse problem and to find the most probable position of MEG measurements. Subsequently, a non-linear fit was used to optimize the dipole parameter. For all grid points, the squared deviation of the best fitting dipole to the measurement data was calculated to determine the dipole position with a minimal residual variance (RV) and maximal goodness of fit (GOF) value. The GOF is given as GOF = 1 – RV and only those fits with a GOF higher than 0.90 were used for further analyzes. Our results are grounded on the source reconstruction of the somatosensory evoked fields (N 30 components). All processing steps were realized using MATLAB software toolbox FieldTrip.

Figure 1 represents the constructed single shell head models for both registration procedures along with the MEG sensor positions for subject V. It is obvious that the head model is localized more frontal within the sensor cap after the application of the IM.

Figure 1 Divergent positions of the single shell volume conductor within the MEG sensor cap (black) caused by different registration methods for MRI and MEG data. Red: head model position obtained after registration based on three AL (landmark positions).
Figure 1

Divergent positions of the single shell volume conductor within the MEG sensor cap (black) caused by different registration methods for MRI and MEG data. Red: head model position obtained after registration based on three AL (landmark positions).

The Euclidian distances between the dipole locations (in mm) obtained with the different registration procedures are listed in Table 1. The results show that the method of the registration procedure has a distinct impact on the source localization. The smallest Euclidian distance amounted to approximately 4.3 mm (subject IV, Dipole B) was detected. The dipole position is visualized in the upper row of Figure 1. The largest distinction was observed for subject V, Dipole A with a value of roughly 15 mm (represented in bottom row of Figure 1). On average the distance between the dipole locations is about 8.76 mm and a value of 8.72 mm is noticeable for the median.

Table 1

Sensitivity of MEG source localization with regard to co-registration method for MEG and MRI data: Euclidian differences in dipole location in mm for N 30 components of somatosensory evoked fields in a single shell volume conductor model.

SubjectDipole ADipole B
I5.074.50
II6.876.37
III10.478.02
IV7.044.29
V14.999.43
VI12.3812.61

For each subject two sources have been reconstructed. reference: dipole location after IM.

Figure 2 illustrates the dipole positions with the smallest (subject IV, Dipole B) and the largest (subject V, Dipole A) distance with respect to the registration procedure. The red points represent the location obtained after MRI and MEG data were registered with the three AL.

Figure 2 Impact of different MEG/MRI co-registration methods on dipole location for somatosensory evoked fields. The dipole locations for the different registration methods (red: based on three AL and green: IM) are plotted on a T1w-MRI slices. The dipole location based on IM method was used for MRI slice selection and the other dipole was projected on these slices.
Figure 2

Impact of different MEG/MRI co-registration methods on dipole location for somatosensory evoked fields. The dipole locations for the different registration methods (red: based on three AL and green: IM) are plotted on a T1w-MRI slices. The dipole location based on IM method was used for MRI slice selection and the other dipole was projected on these slices.

In contrast, the green points mark the location calculated after IM of the MRI-segmented scalp surface and the digitized points of the scalp surface. It is visible, that the dipoles are furthest displaced along the sagittal axis. In addition, slight changes along the frontal and transversal axis are discernible. This becomes visual in Figure 3. The dipoles are obviously located more posterior after the IM registration was applied. This effect was also noticed for most other dipoles. Subject VII seems to be exceptions in this case here a major impact is visible along the coronal axis (changes about 12–14 mm).

Figure 3 Dipole localizations for all subjects, obtained with the different co-registration procedures (red: based on three AL and green: IM). The IM based dipoles are located more lateral and posterior in most cases.
Figure 3

Dipole localizations for all subjects, obtained with the different co-registration procedures (red: based on three AL and green: IM). The IM based dipoles are located more lateral and posterior in most cases.

3 Conclusion

A precise co-registration MEG and MRI data influences the achievable accuracy in the inverse problem. In that regard, the consequences of different MEG – MRI co-registration methods on source localization of somatosensory evoked potentials were investigated. The co-registration determines the MEG gradiometer sensor positions relative to the head, consequently it has an effect on forward and therefore the inverse solution. On the one hand the MR images were aligned to the MEG head coordinate system using the positions of three AL, on other hand the ICP algorithm was used to calculate a transformation matrix. To examine the effect of the co-registration method on the inverse solution the shift in source localization was calculated. It was shown, that errors in MEG gradiometer sensor localization has a strong impact on the inverse solution. The Euclidian distances between the dipole locations obtained with the different registration procedures were larger than 5 mm in most instances. This means that an inaccurate MEG – MRI co-registration leads to mislocalization of reconstructed sources and should be taken into account when working with somatosensory evoked potentials.

Acknowledgement

This work was supported by the Priority Program 1665 of the Deutsche Forschungsgemeinschaft (DFG) (WO1425/5-1) and the EU project ChildBrain (Marie Curie Innovative Training Networks, grant agreement no. 641652).

Author’s Statement

Research funding: The author state no funding involved. Conflict of interest: Authors state no conflict of interest. Material and methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animal use.

References

[1] Nolte G. The magnetic lead field theorem in the quasi-static approximation and its use for magnetoencephalography forward calculation in realistic volume conductors. Phys Med Biol. 2003;48:3637–52.10.1088/0031-9155/48/22/002Search in Google Scholar PubMed

[2] Lamm C, Windischberger C, Leodolter U, Moser E, Bauer H. Co-registration of EEG and MRI data using matching of spline interpolated and MRI-segmented reconstructions of the scalp surface. Brain Topogr. 2001;14:93–100. DOI: 10.1023/A:1012988728672.Search in Google Scholar

[3] Brinkmann BH, O’Brien TJ, Alex Dresner M, Lagerlund TD, Sharbrough FW, Robb RA. In Brain Topogr. 1998;10:245–53. DOI: 10.1023/A:1022266822252.Search in Google Scholar PubMed

[4] Singh KD, Holliday IE, Furlong PL, Harding GF. Evalusation of MRI-MEG/EEG co-registration strategies using Monte Carol simulation. Electroencephalogr Clin Neurophysio. 1997;102:81–5.10.1016/S0921-884X(96)96570-4Search in Google Scholar

Published Online: 2016-9-30
Published in Print: 2016-9-1

©2016 Marie Theiß et al., licensee De Gruyter.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.

Articles in the same Issue

  1. Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
  2. Novel blood protein based scaffolds for cardiovascular tissue engineering
  3. Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
  4. Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
  5. Long-term stable modifications of silicone elastomer for improved hemocompatibility
  6. The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
  7. Biocompatible wear-resistant thick ceramic coating
  8. Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
  9. Examination of dielectric strength of thin Parylene C films under various conditions
  10. Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
  11. Systemic analysis about residual chloroform in PLLA films
  12. A macrophage model of osseointegration
  13. Towards in silico prognosis using big data
  14. Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
  15. Usability evaluation of a locomotor therapy device considering different strategies
  16. Hypoxia-on-a-chip
  17. Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
  18. Fabrication of MEMS-based 3D-μECoG-MEAs
  19. High speed digital interfacing for a neural data acquisition system
  20. Bionic forceps for the handling of sensitive tissue
  21. Experimental studies on 3D printing of barium titanate ceramics for medical applications
  22. Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
  23. 3D printing – a key technology for tailored biomedical cell culture lab ware
  24. 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
  25. Biocompatibility of photopolymers for additive manufacturing
  26. Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
  27. Novel wireless measurement system of pressure dedicated to in vivo studies
  28. Portable auricular device for real-time swallow and chew detection
  29. Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
  30. Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
  31. Deep brain stimulation: increasing efficiency by alternative waveforms
  32. Prediction of immediately occurring microsleep events from brain electric signals
  33. Determining cardiac vagal threshold from short term heart rate complexity
  34. Classification of cardiac excitation patterns during atrial fibrillation
  35. An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
  36. Deriving respiration from high resolution 12-channel-ECG during cycling exercise
  37. Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
  38. Automatic detection and mapping of double potentials in intracardiac electrograms
  39. Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
  40. Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
  41. Best practice: surgeon driven application in pelvic operations
  42. Vasomotor assessment by camera-based photoplethysmography
  43. Classification of morphologic changes in photoplethysmographic waveforms
  44. Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
  45. Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
  46. Nonlinear causal influences assessed by mutual compression entropy
  47. Comparative study of methods for solving the correspondence problem in EMD applications
  48. fNIRS for future use in auditory diagnostics
  49. Semi-automated detection of fractional shortening in zebrafish embryo heart videos
  50. Blood pressure measurement on the cheek
  51. Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
  52. Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
  53. An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
  54. Real-time QRS detection using integrated variance for ECG gated cardiac MRI
  55. Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
  56. In-vitro experiments to characterize ventricular electromechanics
  57. Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
  58. Application of microwave sensor technology in cardiovascular disease for plaque detection
  59. Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
  60. Detection of microsleep events in a car driving simulation study using electrocardiographic features
  61. A method to determine the kink resistance of stents and stent delivery systems according to international standards
  62. Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
  63. Transient Euler-Lagrange/DEM simulation of stent thrombosis
  64. Automated control of the laser welding process of heart valve scaffolds
  65. Automation of a test bench for accessing the bendability of electrospun vascular grafts
  66. Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
  67. Cryopreservation of cells using defined serum-free cryoprotective agents
  68. New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
  69. Determination of the membrane hydraulic permeability of MSCs
  70. Climate retainment in carbon dioxide incubators
  71. Multiple factors influencing OR ventilation system effectiveness
  72. Evaluation of an app-based stress protocol
  73. Medication process in Styrian hospitals
  74. Control tower to surgical theater
  75. Development of a skull phantom for the assessment of implant X-ray visibility
  76. Surgical navigation with QR codes
  77. Investigation of the pressure gradient of embolic protection devices
  78. Computer assistance in femoral derotation osteotomy: a bottom-up approach
  79. Automatic depth scanning system for 3D infrared thermography
  80. A service for monitoring the quality of intraoperative cone beam CT images
  81. Resectoscope with an easy to use twist mechanism for improved handling
  82. In vitro simulation of distribution processes following intramuscular injection
  83. Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
  84. A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
  85. Smart control for functional electrical stimulation with optimal pulse intensity
  86. Tactile display on the remaining hand for unilateral hand amputees
  87. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
  88. An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
  89. Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
  90. Learning discriminative classification models for grading anal intraepithelial neoplasia
  91. Regularization of EIT reconstruction based on multi-scales wavelet transforms
  92. Assessing MRI susceptibility artefact through an indicator of image distortion
  93. EyeGuidance – a computer controlled system to guide eye movements
  94. A framework for feedback-based segmentation of 3D image stacks
  95. Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
  96. 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
  97. Inside-Out access strategy using new trans-vascular catheter approach
  98. US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
  99. Impact of different registration methods in MEG source analysis
  100. 3D segmentation of thyroid ultrasound images using active contours
  101. Designing a compact MRI motion phantom
  102. Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
  103. Classification of indirect immunofluorescence images using thresholded local binary count features
  104. Analysis of muscle fatigue conditions using time-frequency images and GLCM features
  105. Numerical evaluation of image parameters of ETR-1
  106. Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
  107. Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
  108. Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
  109. Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
  110. Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
  111. Evaluation of open-source software for the lung segmentation
  112. Automatic determination of lung features of CF patients in CT scans
  113. Image analysis of self-organized multicellular patterns
  114. Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
  115. Radiopacity assessment of neurovascular implants
  116. Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
  117. Development of an artifact-free aneurysm clip
  118. Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
  119. Rapid prototyping of replica knee implants for in vitro testing
  120. Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
  121. Advanced wettability analysis of implant surfaces
  122. Patient-specific hip prostheses designed by surgeons
  123. Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
  124. Wear of a total intervertebral disc prosthesis
  125. Digital health and digital biomarkers – enabling value chains on health data
  126. Usability in the lifecycle of medical software development
  127. Influence of different test gases in a non-destructive 100% quality control system for medical devices
  128. Device development guided by user satisfaction survey on auricular vagus nerve stimulation
  129. Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
  130. Effect of left atrial hypertrophy on P-wave morphology in a computational model
  131. Simulation of intracardiac electrograms around acute ablation lesions
  132. Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
  133. Assessment of nasal resistance using computational fluid dynamics
  134. Resistance in a non-linear autoregressive model of pulmonary mechanics
  135. Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
  136. Determination of regional lung function in cystic fibrosis using electrical impedance tomography
  137. Development of parietal bone surrogates for parietal graft lift training
  138. Numerical simulation of mechanically stimulated bone remodelling
  139. Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
  140. Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
  141. Principle study on the signal connection at transabdominal fetal pulse oximetry
  142. Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
  143. Evaluating different approaches to identify a three parameter gas exchange model
  144. Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
  145. From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
  146. Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
  147. Improved FPGA controlled artificial vascular system for plethysmographic measurements
  148. Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
  149. Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
  150. Control scheme selection in human-machine- interfaces by analysis of activity signals
  151. Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
  152. Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
  153. Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
  154. Interactive monitoring system for visual respiratory biofeedback
  155. Development of a low-cost senor based aid for visually impaired people
  156. Patient assistive system for the shoulder joint
  157. A passive beating heart setup for interventional cardiology training
Downloaded on 8.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2016-0102/html
Scroll to top button