Abstract
Carbon fiber reinforced polyetheretherketone (CFR-PEEK) has similar mechanical properties to human bone and is considered as the best alternative material to substitute titanium for spine cage implants. To compensate its poor osteogenic properties and limited bioinertness, CFR-PEEK was coated with a thin film of titanium. In the study, we investigated the biological response in vitro of titanium coated CFR-PEEK with different vacuum plasma pretreatments. The so modified surface revealed first hints for a good cell response by excellent cell adhesion and morphology of human osteoblast – like cells MG 63 (ATXX:’CRL-1427). Thus, the findings show that surface roughness of CFR-PEEK material has a profound effect on the biological activity via vacuum plasma treatment.
1 Introduction
Up to 80% of the population suffer dorsal pain at some time during life. Most of back problems get better without a surgical procedure. There are many different causes of back pain and various types of treatments. In case of posttraumatic stress disorder i.e. spine instability or degenerative disk disease, defined as abnormal movement between one vertebrae to another, it cannot be always cured by conventional therapy. Thus, a lumbar spinal fusion surgery is performed to stop the motion of the painful vertebral segment and to strengthen the vertebral body. One approach is to remove the disk and vertebral bodies and replace a vertebrae with spinal cage implant (Figure 1). The requirements for those implants are biocompatibility, cell adhesion, sufficient biomechanical behavior (e.g. modulus of elasticity) and osteoconductivity. Titanium is the material mostly used for such implants [1]. However, titanium has disadvantages for this application. Its stiffness, which is ten times higher than human bones, can lead to unbalances between the implant and spine and results in loosening of the implants due to stress shielding effect. Another disadvantage is the metal-induced artefact in computed tomography (CT) and magnetic resonance imaging (MRI) that consequently reduces drastically the ingrowth control of the implant. To overcome those drawbacks, polyetheretherketone (PEEK) is becoming increasingly popular as an alternative material for implants. This thermoplastic polymer possesses excellent biocompatibility, ductility, elongation and fatigue resistance, and its elastic modulus is close to human bones. Furthermore, PEEK does not induce image artefacts and allows a better control of implant using MRI [2], [3]. However, deficient osteogenic properties limits its fields of application. PEEK is not used in the lumbar spine region due to the extreme high loads in this area, which may cause e.g. pseudarthrosis and loosening of the implant. Carbon fiber reinforced (CFR) PEEK is a possible material to overcome the mechanical weakness. Added carbon fibers improve the mechanical resistance and ductility compared to pure PEEK material. Surface modification of CFR PEEK promotes the osteoblast attachment and growth. A functional titanium coating is applied and offers the best possible osseointegration of CFR PEEK implants.

Cage for posterior lumbal interbody fusion PLIF (Silony Medical GmbH, Germany).
This study assesses the in vitro osseointegration of CFR PEEK implants coated with titanium. Before coating, various vacuum plasma treatments were carried out. The aim is to evaluate the effect of surface properties i.e. surface roughness to reduce the osseointegration period for the proposed spinal cage implant [4].
2 Methods
Different surface treatments in vacuo were performed before titanium deposition. Cylindrically formed CFR PEEK substrates were specially prepared for this study with a diameter of 30 mm and a thickness of 4 mm. For cell tests the cylindrical substrates were cut into quarters. The CFR PEEK substrates were provided by Silony Medical GmbH, Germany.
2.1 Surface roughness analysis
To characterize the roughness before and after surface treatments a roughness tester PCR-RT 1200 (PCE Instruments, Germany) was used. The roughness tester works according to the piezoelectric micro probe principle like the highly accurate laboratory measuring instrument. The roughness tester scans the surface within seconds and shows digitally either the value Ra, Rz, Rq or Rt.
2.2 Face milling
The CFR PEEK substrates were milled to get a suitable surface roughness before plasma treatment and deposition steps. The milling process was done with a DECKEL FP1 machine (Friedrich Deckel AG, Germany). Following milling parameter were set to reach a common surface roughness of Ra = 0.9 μm.
The milling speed applied was 600 rotation per min, feed rate 50 mm/min, number of teeth 6, and diameter 40 mm.
2.3 Plasma treatment and coating
Plasma treatment and the titanium deposition were done with a combined Physical Vapor Deposition and Plasma Enhanced Chemical Vapor Deposition (PVD&PECVD) machine Porta 350 (Plasma Electronic GmbH, Germany). In this study, we performed two different plasma treatments on CFR PEEK substrates before titanium deposition. Samples were placed in the vacuum chamber and fixed on a sample holder connected to a radio frequency (RF) generator and pumped down to 2 × 10−3 Pa. The first plasma treatment was performed using an argon plasma with 80 sccm gas flow at 0.37 Pa. As alternative plasma treatment, we carried out an argon and oxygen gas mixture at 0.51 Pa with 70 sccm and 30 sccm, respectively. Substrates were treated for 3 h at 330 W and 350 V bias. After plasma treatment, the substrates were coated in the same vacuum chamber without breaking the vacuum.
The deposition method of the titanium films was unbalanced magnetron direct current (DC) sputtering. The background pressure was maintained at 2 × 10−3 Pa. Titanium growth was deposited with an argon flow of 30 sccm and at a pressure of 0.11 Pa during 30 min at 2000 W DC power and target voltage of 410 V. A 240 nm thin titanium layer was deposited on treated and untreated CFR-PEEK samples. Two different bias voltages were applied 0 and 200 V during titanium deposition.
2.4 Coating adhesion
The film adhesion was quantified by cross-cut tests in accordance with standard DIN EN ISO 2409:2007. It describes a test method for assessing the resistance of coatings to separation from substrates when a right-angle lattice pattern is cut into the coating, penetrating through to the substrate. Then coating adhesion was measured by tape test method. A specific adhesive tape was applied to the pattern and pulled off at constant speed at an angle of approximately 60 degrees.
2.5 Cell adhesion & morphology
In order to collect first hints for a good tissue integration of the modified CFR PEEK implant, cell adhesion and morphology of human osteoblast-like MG-63 P91 cells, were seeded at different treated CFR PEEK substrates.
After sterilization of the substrates by incubation in 70% Ethanol for > 1 h they were placed in a cell culture 12 well plate and seeded with MG-63 cells with a cell density of 20,000 cells/cm2. The duration of cell adhesion was 20 h at 37°C and 5% CO2 under static conditions. As culture medium Dulbecco‘s Modified Eagle Medium (DMEM; high glucose) supplemented with 10% fetal calf serum, 10,000 U/ml penicillin and 10,000 μg/ml streptomycin was used.
After cell seeding an actin staining (= cytoskelet-staining) with phalloidin oregon green (Thermo Fisher Scientific) was performed to monitor cell morphology of the adhered osteoblast-like cells. Between all staining steps substrates were washed at least once with phosphate buffered saline (PBS). All steps were realized at room temperature. To fix the cells substrates were incubated with 4% paraformaldehyde/PBS for 30 min. Since actin is a part of the cytoskelet cells were permeabilized with 0.02% TitonX100/dH2O for 15 min to get access to the antigen. For blocking unspecific binding of the phalloidin the fixed and permeabilized cells were treated with a solution of 1% bovine serum albumin/PBS for 45 min. Finally samples were stained with phalloidin oregon green/PBS solution (Dilution: 1:50) for 3 h. In order to stain cell nuclei additionally a second staining with 5 μg/ml 4′,6-Diamidin-2-phenylindol (DAPI)/PBS for 10 min was performed.
To evaluate the adhesion-rate and morphology of the adhered cells, the so prepared substrates were investigated by fluorescence microscopy.
3 Results and discussion
Milling tests showed, that carbon reinforced PEEK is machinable with standard milling processes. After the milling step, we measured surface roughness (Ra) of substrates between 0.5 μm to 1 μm. The milling process creates groove shapes on the substrate surface. These undesirable grooves are responsible for the inhomogeneous surface roughness. It has been determined that standard machining process reached his limits and it is not possible to treat undercuts, contrary to plasma processes. With an argon or argon/oxygen plasma treatment it is possible to adjust a perfect and homogeneous surface roughness, even in undercuts. After 3 h plasma treatment, surface roughness (Ra) between 1.5 μm and 1.8 μm are achieved. Samples 1–9 were treated with Ar/O2 plasma. Samples 1–5 and 6–9 were treated with a generator power of 500 W and 750 W, so plasma treatment increases the roughness up to 100% (Figure 2).

Roughness before and after plasma treatment. Sample 1–5 were etched by 500 V; Sample 5–9 at 750 V.
After argon and oxygen mixed plasma treatment, samples showed no adhesion of the titanium layer to the CFR PEEK for 0 and 200 V bias, as highlighted in Figure 3-left. This might be due to strong and irreversible surface reactions between PEEK and oxygen plasma.

Substrate treated with Ar/O2 plasma and 200 V bias during titanium deposition (left) and treated with pure argon plasma and titanium coated (right) CFR PEEK.
Substrates treated under argon plasma and coated with 0 V bias showed very good adhesion according to DIN EN ISO 2409:2007. After tape test, no delamination was observed, as shown in Figure 4.

Titanium coated substrate after cross cutting test.
The results showed, that low bias voltage at 0 V yield better layer adhesions. However high bias voltage at 200 V increase drastically the surface roughness. However, the influence of bias voltages during deposition needs to be clarified.
After titanium deposition onto CFR-PEEK substrates, cell tests were carried out to evaluate the adhesion and morphology of seeded MG-63 osteoblastic-like cells. For this test, two different substrates and one standard cell culture plastic as positive control were populated with cells. One substrate without milling and plasma treatment, another substrate milled and plasma treated with a surface roughness of Ra = 1.9 μm after treatment. The results showed that cells adhere on the plasma treated and coated CFR-PEEK substrates.
MG-63 cells showed a comparable adhesion-rate to the positive control and a typical well spread morphology of viable MG-63 cells, as depicted in Figure 5 on right side. In Figure 5 (left side), the untreated substrate showed reduced adhesion rate of about 30% compared to the plasma treated CFR-PEEK sample and the cell culture reference. Furthermore, cells settled down on untreated substrate showed a narrow cell morphology due to preferable orientation on the milling grooves.

Fluorescence microscopy pictures of cell adhesion and morphology of MG-63 cells on different titan-coated CFR-PEEK substrates (left: without plasma treatment; right: with plasma treatment. Green: actin, stained with phalloidin Oregon green, Blue: cell nuclei stained with DAPI. The dark areas correspond with the milling grooves.
4 Summary
In this study, we showed that a homogeneous surface roughness of CFR-PEEK can be tuned using argon plasma treatment. It was also shown that adhesive thin titanium films could be deposited on CFR PEEK using pure argon plasma treatment.
Biological activity of titanium coated CFR-PEEK substrates was proven with human osteoblastic-like cells MG-63. It was demonstrated that cells adhere well on the treated and coated implant surface. The cells proliferate very well on surfaces with a roughness around 2 μm. The optimal surface roughness and layer thickness have to be investigated in further tests. It is also necessary to conduct toxicity tests at the optimized CFR PEEK surfaces.
Implants of this kind are massively loaded mechanically while implantation. So it is also important to test the layer adhesion under high loads. Tensile and peel tests can be carried out to characterize the layer adhesion.
Author’s Statement
Research funding: The author state no funding involved. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animal use.
References
[1] Kunzler TP, Drobek T, Schuler M, Spencer ND. Systematic study of osteoblast and fibroblast response to roughness by means of surface-morphology gradients. Biomaterials. 2007;28: 2175–82.10.1016/j.biomaterials.2007.01.019Search in Google Scholar PubMed
[2] Kurtz SM, Devine JN. PEEK Biomaterials in Trauma, Orthopedic, and Spinal Implants. Biomaterials. 2007;28:4845–69.10.1016/j.biomaterials.2007.07.013Search in Google Scholar PubMed PubMed Central
[3] Stübinger S, Drechsler A, Bürki A, Klein K, Kronen P, von Rechenberg B. Titanium and hydroxyapatite coating of polyetheretherketone and carbon fiber−reinforced polyetheretherketone: a pilot study in sheep. J Biomed Mater Res B Appl Biomater. 2016;104:1182–91.10.1002/jbm.b.33471Search in Google Scholar PubMed
[4] Albrektsson T, Wennerberg A. Oral implant surfaces: part 1–review focusing on topographic and chemical properties of different surfaces and in vivo responses to them. Int J Prosthodont. 2004;17:536–43.Search in Google Scholar
©2016 Michael Banghard et al., licensee De Gruyter.
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.
Articles in the same Issue
- Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
- Novel blood protein based scaffolds for cardiovascular tissue engineering
- Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
- Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
- Long-term stable modifications of silicone elastomer for improved hemocompatibility
- The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
- Biocompatible wear-resistant thick ceramic coating
- Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
- Examination of dielectric strength of thin Parylene C films under various conditions
- Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
- Systemic analysis about residual chloroform in PLLA films
- A macrophage model of osseointegration
- Towards in silico prognosis using big data
- Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
- Usability evaluation of a locomotor therapy device considering different strategies
- Hypoxia-on-a-chip
- Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
- Fabrication of MEMS-based 3D-μECoG-MEAs
- High speed digital interfacing for a neural data acquisition system
- Bionic forceps for the handling of sensitive tissue
- Experimental studies on 3D printing of barium titanate ceramics for medical applications
- Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
- 3D printing – a key technology for tailored biomedical cell culture lab ware
- 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
- Biocompatibility of photopolymers for additive manufacturing
- Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
- Novel wireless measurement system of pressure dedicated to in vivo studies
- Portable auricular device for real-time swallow and chew detection
- Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
- Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
- Deep brain stimulation: increasing efficiency by alternative waveforms
- Prediction of immediately occurring microsleep events from brain electric signals
- Determining cardiac vagal threshold from short term heart rate complexity
- Classification of cardiac excitation patterns during atrial fibrillation
- An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
- Deriving respiration from high resolution 12-channel-ECG during cycling exercise
- Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
- Automatic detection and mapping of double potentials in intracardiac electrograms
- Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
- Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
- Best practice: surgeon driven application in pelvic operations
- Vasomotor assessment by camera-based photoplethysmography
- Classification of morphologic changes in photoplethysmographic waveforms
- Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
- Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
- Nonlinear causal influences assessed by mutual compression entropy
- Comparative study of methods for solving the correspondence problem in EMD applications
- fNIRS for future use in auditory diagnostics
- Semi-automated detection of fractional shortening in zebrafish embryo heart videos
- Blood pressure measurement on the cheek
- Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
- Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
- An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
- Real-time QRS detection using integrated variance for ECG gated cardiac MRI
- Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
- In-vitro experiments to characterize ventricular electromechanics
- Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
- Application of microwave sensor technology in cardiovascular disease for plaque detection
- Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
- Detection of microsleep events in a car driving simulation study using electrocardiographic features
- A method to determine the kink resistance of stents and stent delivery systems according to international standards
- Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
- Transient Euler-Lagrange/DEM simulation of stent thrombosis
- Automated control of the laser welding process of heart valve scaffolds
- Automation of a test bench for accessing the bendability of electrospun vascular grafts
- Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
- Cryopreservation of cells using defined serum-free cryoprotective agents
- New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
- Determination of the membrane hydraulic permeability of MSCs
- Climate retainment in carbon dioxide incubators
- Multiple factors influencing OR ventilation system effectiveness
- Evaluation of an app-based stress protocol
- Medication process in Styrian hospitals
- Control tower to surgical theater
- Development of a skull phantom for the assessment of implant X-ray visibility
- Surgical navigation with QR codes
- Investigation of the pressure gradient of embolic protection devices
- Computer assistance in femoral derotation osteotomy: a bottom-up approach
- Automatic depth scanning system for 3D infrared thermography
- A service for monitoring the quality of intraoperative cone beam CT images
- Resectoscope with an easy to use twist mechanism for improved handling
- In vitro simulation of distribution processes following intramuscular injection
- Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
- A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
- Smart control for functional electrical stimulation with optimal pulse intensity
- Tactile display on the remaining hand for unilateral hand amputees
- Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
- An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
- Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
- Learning discriminative classification models for grading anal intraepithelial neoplasia
- Regularization of EIT reconstruction based on multi-scales wavelet transforms
- Assessing MRI susceptibility artefact through an indicator of image distortion
- EyeGuidance – a computer controlled system to guide eye movements
- A framework for feedback-based segmentation of 3D image stacks
- Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
- 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
- Inside-Out access strategy using new trans-vascular catheter approach
- US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
- Impact of different registration methods in MEG source analysis
- 3D segmentation of thyroid ultrasound images using active contours
- Designing a compact MRI motion phantom
- Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
- Classification of indirect immunofluorescence images using thresholded local binary count features
- Analysis of muscle fatigue conditions using time-frequency images and GLCM features
- Numerical evaluation of image parameters of ETR-1
- Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
- Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
- Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
- Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
- Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
- Evaluation of open-source software for the lung segmentation
- Automatic determination of lung features of CF patients in CT scans
- Image analysis of self-organized multicellular patterns
- Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
- Radiopacity assessment of neurovascular implants
- Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
- Development of an artifact-free aneurysm clip
- Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
- Rapid prototyping of replica knee implants for in vitro testing
- Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
- Advanced wettability analysis of implant surfaces
- Patient-specific hip prostheses designed by surgeons
- Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
- Wear of a total intervertebral disc prosthesis
- Digital health and digital biomarkers – enabling value chains on health data
- Usability in the lifecycle of medical software development
- Influence of different test gases in a non-destructive 100% quality control system for medical devices
- Device development guided by user satisfaction survey on auricular vagus nerve stimulation
- Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
- Effect of left atrial hypertrophy on P-wave morphology in a computational model
- Simulation of intracardiac electrograms around acute ablation lesions
- Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
- Assessment of nasal resistance using computational fluid dynamics
- Resistance in a non-linear autoregressive model of pulmonary mechanics
- Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
- Determination of regional lung function in cystic fibrosis using electrical impedance tomography
- Development of parietal bone surrogates for parietal graft lift training
- Numerical simulation of mechanically stimulated bone remodelling
- Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
- Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
- Principle study on the signal connection at transabdominal fetal pulse oximetry
- Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
- Evaluating different approaches to identify a three parameter gas exchange model
- Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
- From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
- Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
- Improved FPGA controlled artificial vascular system for plethysmographic measurements
- Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
- Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
- Control scheme selection in human-machine- interfaces by analysis of activity signals
- Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
- Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
- Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
- Interactive monitoring system for visual respiratory biofeedback
- Development of a low-cost senor based aid for visually impaired people
- Patient assistive system for the shoulder joint
- A passive beating heart setup for interventional cardiology training
Articles in the same Issue
- Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
- Novel blood protein based scaffolds for cardiovascular tissue engineering
- Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
- Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
- Long-term stable modifications of silicone elastomer for improved hemocompatibility
- The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
- Biocompatible wear-resistant thick ceramic coating
- Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
- Examination of dielectric strength of thin Parylene C films under various conditions
- Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
- Systemic analysis about residual chloroform in PLLA films
- A macrophage model of osseointegration
- Towards in silico prognosis using big data
- Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
- Usability evaluation of a locomotor therapy device considering different strategies
- Hypoxia-on-a-chip
- Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
- Fabrication of MEMS-based 3D-μECoG-MEAs
- High speed digital interfacing for a neural data acquisition system
- Bionic forceps for the handling of sensitive tissue
- Experimental studies on 3D printing of barium titanate ceramics for medical applications
- Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
- 3D printing – a key technology for tailored biomedical cell culture lab ware
- 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
- Biocompatibility of photopolymers for additive manufacturing
- Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
- Novel wireless measurement system of pressure dedicated to in vivo studies
- Portable auricular device for real-time swallow and chew detection
- Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
- Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
- Deep brain stimulation: increasing efficiency by alternative waveforms
- Prediction of immediately occurring microsleep events from brain electric signals
- Determining cardiac vagal threshold from short term heart rate complexity
- Classification of cardiac excitation patterns during atrial fibrillation
- An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
- Deriving respiration from high resolution 12-channel-ECG during cycling exercise
- Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
- Automatic detection and mapping of double potentials in intracardiac electrograms
- Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
- Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
- Best practice: surgeon driven application in pelvic operations
- Vasomotor assessment by camera-based photoplethysmography
- Classification of morphologic changes in photoplethysmographic waveforms
- Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
- Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
- Nonlinear causal influences assessed by mutual compression entropy
- Comparative study of methods for solving the correspondence problem in EMD applications
- fNIRS for future use in auditory diagnostics
- Semi-automated detection of fractional shortening in zebrafish embryo heart videos
- Blood pressure measurement on the cheek
- Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
- Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
- An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
- Real-time QRS detection using integrated variance for ECG gated cardiac MRI
- Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
- In-vitro experiments to characterize ventricular electromechanics
- Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
- Application of microwave sensor technology in cardiovascular disease for plaque detection
- Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
- Detection of microsleep events in a car driving simulation study using electrocardiographic features
- A method to determine the kink resistance of stents and stent delivery systems according to international standards
- Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
- Transient Euler-Lagrange/DEM simulation of stent thrombosis
- Automated control of the laser welding process of heart valve scaffolds
- Automation of a test bench for accessing the bendability of electrospun vascular grafts
- Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
- Cryopreservation of cells using defined serum-free cryoprotective agents
- New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
- Determination of the membrane hydraulic permeability of MSCs
- Climate retainment in carbon dioxide incubators
- Multiple factors influencing OR ventilation system effectiveness
- Evaluation of an app-based stress protocol
- Medication process in Styrian hospitals
- Control tower to surgical theater
- Development of a skull phantom for the assessment of implant X-ray visibility
- Surgical navigation with QR codes
- Investigation of the pressure gradient of embolic protection devices
- Computer assistance in femoral derotation osteotomy: a bottom-up approach
- Automatic depth scanning system for 3D infrared thermography
- A service for monitoring the quality of intraoperative cone beam CT images
- Resectoscope with an easy to use twist mechanism for improved handling
- In vitro simulation of distribution processes following intramuscular injection
- Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
- A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
- Smart control for functional electrical stimulation with optimal pulse intensity
- Tactile display on the remaining hand for unilateral hand amputees
- Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
- An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
- Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
- Learning discriminative classification models for grading anal intraepithelial neoplasia
- Regularization of EIT reconstruction based on multi-scales wavelet transforms
- Assessing MRI susceptibility artefact through an indicator of image distortion
- EyeGuidance – a computer controlled system to guide eye movements
- A framework for feedback-based segmentation of 3D image stacks
- Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
- 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
- Inside-Out access strategy using new trans-vascular catheter approach
- US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
- Impact of different registration methods in MEG source analysis
- 3D segmentation of thyroid ultrasound images using active contours
- Designing a compact MRI motion phantom
- Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
- Classification of indirect immunofluorescence images using thresholded local binary count features
- Analysis of muscle fatigue conditions using time-frequency images and GLCM features
- Numerical evaluation of image parameters of ETR-1
- Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
- Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
- Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
- Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
- Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
- Evaluation of open-source software for the lung segmentation
- Automatic determination of lung features of CF patients in CT scans
- Image analysis of self-organized multicellular patterns
- Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
- Radiopacity assessment of neurovascular implants
- Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
- Development of an artifact-free aneurysm clip
- Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
- Rapid prototyping of replica knee implants for in vitro testing
- Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
- Advanced wettability analysis of implant surfaces
- Patient-specific hip prostheses designed by surgeons
- Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
- Wear of a total intervertebral disc prosthesis
- Digital health and digital biomarkers – enabling value chains on health data
- Usability in the lifecycle of medical software development
- Influence of different test gases in a non-destructive 100% quality control system for medical devices
- Device development guided by user satisfaction survey on auricular vagus nerve stimulation
- Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
- Effect of left atrial hypertrophy on P-wave morphology in a computational model
- Simulation of intracardiac electrograms around acute ablation lesions
- Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
- Assessment of nasal resistance using computational fluid dynamics
- Resistance in a non-linear autoregressive model of pulmonary mechanics
- Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
- Determination of regional lung function in cystic fibrosis using electrical impedance tomography
- Development of parietal bone surrogates for parietal graft lift training
- Numerical simulation of mechanically stimulated bone remodelling
- Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
- Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
- Principle study on the signal connection at transabdominal fetal pulse oximetry
- Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
- Evaluating different approaches to identify a three parameter gas exchange model
- Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
- From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
- Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
- Improved FPGA controlled artificial vascular system for plethysmographic measurements
- Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
- Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
- Control scheme selection in human-machine- interfaces by analysis of activity signals
- Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
- Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
- Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
- Interactive monitoring system for visual respiratory biofeedback
- Development of a low-cost senor based aid for visually impaired people
- Patient assistive system for the shoulder joint
- A passive beating heart setup for interventional cardiology training