Home Medicine Blood pressure measurement on the cheek
Article Open Access

Blood pressure measurement on the cheek

  • Sarah Schneider EMAIL logo , Klaus Affeld , Claudia Kopic and Ulrich Kertzscher
Published/Copyright: September 30, 2016

Abstract

In a large group of patients, it is impossible to measure blood pressure using an upper arm cuff. An alternative, non-invasive method of blood pressure measurement is required for patients with severe limb deformities or obesity, for amputees, and in the emergency medicine. The device proposed here measures blood pressure in the cheek using a small pressure pad and a pump to occlude the cheek artery – arteria facialis – and assesses blood flow with an infrared light source and a detector. The infrared light signal is analysed to assess the systolic and diastolic blood pressure of the patient. Manual evaluation of the light intensity signal showed a good agreement between cheek blood pressure measurement and a reference measurement using an upper arm cuff.

1 Objective

Blood pressure is one of the most important factors for the diagnosis of cardiovascular disease [1]. In 2014, 39% of all deaths in Germany were associated with diseases of the cardiovascular system [2]. High blood pressure (hyper-tension) often goes unnoticed by the patient, and becomes apparent only when serious events such as heart attack, stroke or kidney failure occur. These complications can be prevented or at least delayed by an early treatment of hypertension. Therefore, blood pressure measurement is an important routine examination and should be performed at home on a daily basis.

In a large group of patients, it is not possible to administer the customary, non-invasive blood pressure measurement using an upper arm or wrist cuff. These are patients with severe limb deformities, such as people damaged by the drug thalidomide (“Contergan”), but also obese patients and amputees. Additionally, in emergency medicine, there is often limited access to the upper arm or wrist as required to measure blood pressure.

A new blood pressure measurement device is presented here which does not depend on a patient’s limbs; instead, a small pressure pad is applied to the cheek and blood flow is detected using a photoplethysmographic sensor.

2 Material and methods

The measurement device consists of a U-shaped frame which is to be placed in the corner of the mouth as in Figure 1. The first leg of the frame is situated in the oral cavity, holding an infrared light detector at the inner cheek. An infrared light-emitting diode is located on the second leg of the frame at the outer cheek. A pressure pad is situated between the frame and the light diode at the outer cheek.

Figure 1: Schematic of the measurement device. A light diode, a light detector, and the pressure pad are mounted on a frame which is placed in the corner of the mouth.
Figure 1:

Schematic of the measurement device. A light diode, a light detector, and the pressure pad are mounted on a frame which is placed in the corner of the mouth.

Principle: Infrared light travels through the pressure pad, cheek tissue and the facial artery (Arteria facialis), and a fraction of the light is detected by the photo-detector at the inner cheek. The transmitted light intensity changes according to the cheek’s blood content. This intensity signal is used to detect the blood pulse in the facial artery.

The light intensity signal demonstrates a specific behavior during the inflation and deflation of pad pressure, as seen in Figure 2. Blood flow in the facial artery will decrease with increasing pad pressure, reaching a global maximum at the diastolic blood pressure.

Figure 2: Light intensity signal (top) and pad pressure signal (bottom) on the cheek of a subject during a measurement. Systolic (SP) and diastolic blood pressure (DP) can be determined from the characteristic progression of the light intensity signal. The height-adjusted blood pressure from this measurement was 130/83 mm Hg during inflation and 112/64 mm Hg during deflation of the pressure pad.
Figure 2:

Light intensity signal (top) and pad pressure signal (bottom) on the cheek of a subject during a measurement. Systolic (SP) and diastolic blood pressure (DP) can be determined from the characteristic progression of the light intensity signal. The height-adjusted blood pressure from this measurement was 130/83 mm Hg during inflation and 112/64 mm Hg during deflation of the pressure pad.

Eventually, the facial artery will be completely emptied of blood when the systolic pressure is reached in the pad. The situation is reversed with decreasing pad pressure: blood flow will return to the facial artery, resulting in a recurring light intensity signal when systolic pressure is reached. The two global maxima of the intensity signal indicate the location of the diastolic pressure in the pad.

Evaluation: Light intensity signals (LI signals) were examined manually by practiced investigators. Data was processed using MATLAB R2013b (The MathWorks, Inc., Natick, MA, USA). The evaluation is based on the following assumptions:

  1. Systolic pressure is detected where LI signals disappear and no clear pulsations are found any longer in the LI signal, and

  2. Diastolic pressure is detected at the global maxima of the LI signal.

For the evaluation, the upper and lower envelopes of the LI signal were computed. The lower envelope is used to generate a positive corrected intensity signal, as shown in Figure 3.

Figure 3: Light intensity signal with envelopes (top) and corrected intensity signal with upper envelope (bottom).
Figure 3:

Light intensity signal with envelopes (top) and corrected intensity signal with upper envelope (bottom).

Implementation: The measurement starts by inflating the pressure pad with a piston pump (PHD Ultra, Harvard Apparatus, Holliston, MA, USA) to a suprasystolic pressure. Once the peak pressure is reached, the pressure pad is slowly deflated. Pad pressure and transmitted infrared light signals were recorded using LabVIEW (National Instruments, Austin, TX, USA).

A prototype of the device is shown in Figure 4. The frame is made from polylactic acid (PLA) and printed using a 3D printer (Ultimaker 2, Geldermalsen, The Netherlands). The pressure pad is made from polyurethane foil (100 μm thickness). The diode emits infrared light with a wavelength of 940 nm.

Figure 4: Prototype of the new blood pressure measurement device.
Figure 4:

Prototype of the new blood pressure measurement device.

The device was tested on 10 voluntary, healthy subjects in comparison with an M500 automatic blood pressure monitor (Omron Healthcare, Kyoto, Japan). The inner cheek area of the cheek device was covered with a protective sleeve (Premium Plus Dental Supplies Inc., NY, USA). Both devices were positioned correctly on the cheek and upper arm, and subjects remained seated during the measurement. The height difference between cheek and upper arm was measured for a correct calibration of the pressure results. The arm cuff measurement was activated manually during the inflation and deflation phases of the cheek measurement. Measurements were performed both at rest and after light physical activity.

3 Results

Fifty eight measurements were performed on 10 subjects (five male, five female, average age: 26.2 ± 5.1 years). Forty seven measurements were performed at rest and 11 measurements were performed after the subject performed light physical activity. Each measurement resulted in two analyzable parts (inflation and deflation phases of pressure) and two upper arm measurements to compare with the cheek measurements. Slopes of the pressure curve were 2.73 ± 0.04 mm Hg ascending and 2.53 ± 0.05 mm Hg descending.

Manual detection: The manual detection was performed using the raw light intensity signal as shown in Figure 3 (top) and at the corrected LI signal, shown in Figure 3 (bottom).

The differences between the manual detection and the reference measurement at the upper arm using the raw LI signal were systolic −15.93 ± 8.88 mm Hg for inflation and −19 ± 6.43 mm Hg for deflation and diastolic −1.28 ± 7.19 mm Hg for inflation and −11.4 ± 12.39 mm Hg for deflation. Bland-Altman plots are shown in Figure 5.

Figure 5: Bland-Altman plots of manual detection (based on raw LI signal) vs. reference blood pressure measurement with upper arm cuff (Omron) for systolic and diastolic inflation and deflation phases.
Figure 5:

Bland-Altman plots of manual detection (based on raw LI signal) vs. reference blood pressure measurement with upper arm cuff (Omron) for systolic and diastolic inflation and deflation phases.

The differences between the manual detection and the reference measurement at the upper arm using the corrected LI signal were systolic −3.62 ± 16.24 mm Hg during inflation and −10.09 ± 9.66 mm Hg during deflation and diastolic −0.33 ± 13.33 mm Hg during inflation and −12.85 ± 13.03 mm Hg during deflation. Bland-Altman plots are shown in Figure 6.

Figure 6: Bland-Altman plots of manual detection (based on corrected LI signal) vs. reference blood pressure measurement with upper arm cuff (Omron) for systolic and diastolic inflation and deflation phases.
Figure 6:

Bland-Altman plots of manual detection (based on corrected LI signal) vs. reference blood pressure measurement with upper arm cuff (Omron) for systolic and diastolic inflation and deflation phases.

4 Discussion

The results of the manual detection show a good agreement between manual detection and reference measurement for systole and diastole during the inflation phase. Deviations are assessed as acceptable if they meet the criteria of the American Association for the Advancement of Medical Instrumentation (AAMI), [3]: the test device must not differ from the reference method (mercury standard in the AAMI protocol) by a mean difference >5 mm Hg or a standard deviation >8 mm Hg. This is true for all of the manual detection methods during inflation except for the systolic inflation using raw light intensity signal.

During the deflation phase, a systemic error occurs for both systole and diastole: lower absolute pressures show a higher difference from reference measurements than high absolute pressure. The standard deviation is high in all measurements.

There are various reasons for the pressure differences:

  1. Measurement errors in height difference result in pressure difference errors of 0.73 mm Hg/cm,

  2. The initial or final pulse of the LI signal could not be clearly differentiated from noisy signals during occlusion, resulting in temporal allocation errors of approximately 2.63 mm Hg/s,

  3. Cheek blood pressure is actually lower than upper arm blood pressure because the vessel is smaller,

  4. The sensor was not positioned correctly and the facial artery was not in the center of infrared light.

  5. The reference measurement using the upper arm cuff has an error which should be below 5 ± 8 mm Hg (to meet the criteria of the American Association for the Advancement of Medical Instrumentation, [3]).

There are differences between the manual detection methods (raw LI signal vs. corrected LI signal) of 12.30 ± 13.33 mm Hg for systole during inflation, 8.91 ± 9.95 mm Hg for systole during deflation, 0.95 ± 9.22 mm Hg for diastole during inflation, and –1.49 ± 5.64 mm Hg for diastole during deflation. This shows that systolic values are better detected using the corrected LI signal.

5 Conclusion

A new technique was developed to measure blood pressure in patients with severe limb deformation. Pressure and infrared light signals could be measured on the cheek using the device.

The manual detection of systolic and diastolic blood pressure from light intensity signals is possible, and sufficiently comparable to measurements on the upper arm. The system has been evaluated in normotensive subjects. The performance of the device will be tested for hyper- and hypotensive subjects and thalidomide patients in the future.

Currently, algorithms are being evaluated for the automatic detection of blood pressure values from infrared light signals.

Author’s Statement

Research funding: Funded by “BIH Technology Transfer Fund Medical Devices”, Berlin Institute of Health (BIH). Conflict of interest: The authors declare no conflict of interest. Material and Methods: Informed consent has been obtained from all individuals included in the experiments. Ethical approval: The research related to human use complies with all the relevant national regulations, institutional policies and was performed in accordance with the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.

References

[1] Hypertonie – Gesundheitsberichterstattung des Bundes. Robert-Koch-Institut; 2008.Search in Google Scholar

[2] Gesundheit – Todesursachen in Deutschland 2014, Statistisches Bundesamt, Wiesbaden.Search in Google Scholar

[3] O’Brien E, Waeber B, Parati G, Staessen J, Myers MG. Blood pressure measuring devices: recommendations of the European Society of Hypertension. Br Med J. 2001;322:531–6.10.1136/bmj.322.7285.531Search in Google Scholar PubMed PubMed Central

Published Online: 2016-9-30
Published in Print: 2016-9-1

©2016 Sarah Schneider et al., licensee De Gruyter.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.

Articles in the same Issue

  1. Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
  2. Novel blood protein based scaffolds for cardiovascular tissue engineering
  3. Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
  4. Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
  5. Long-term stable modifications of silicone elastomer for improved hemocompatibility
  6. The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
  7. Biocompatible wear-resistant thick ceramic coating
  8. Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
  9. Examination of dielectric strength of thin Parylene C films under various conditions
  10. Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
  11. Systemic analysis about residual chloroform in PLLA films
  12. A macrophage model of osseointegration
  13. Towards in silico prognosis using big data
  14. Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
  15. Usability evaluation of a locomotor therapy device considering different strategies
  16. Hypoxia-on-a-chip
  17. Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
  18. Fabrication of MEMS-based 3D-μECoG-MEAs
  19. High speed digital interfacing for a neural data acquisition system
  20. Bionic forceps for the handling of sensitive tissue
  21. Experimental studies on 3D printing of barium titanate ceramics for medical applications
  22. Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
  23. 3D printing – a key technology for tailored biomedical cell culture lab ware
  24. 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
  25. Biocompatibility of photopolymers for additive manufacturing
  26. Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
  27. Novel wireless measurement system of pressure dedicated to in vivo studies
  28. Portable auricular device for real-time swallow and chew detection
  29. Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
  30. Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
  31. Deep brain stimulation: increasing efficiency by alternative waveforms
  32. Prediction of immediately occurring microsleep events from brain electric signals
  33. Determining cardiac vagal threshold from short term heart rate complexity
  34. Classification of cardiac excitation patterns during atrial fibrillation
  35. An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
  36. Deriving respiration from high resolution 12-channel-ECG during cycling exercise
  37. Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
  38. Automatic detection and mapping of double potentials in intracardiac electrograms
  39. Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
  40. Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
  41. Best practice: surgeon driven application in pelvic operations
  42. Vasomotor assessment by camera-based photoplethysmography
  43. Classification of morphologic changes in photoplethysmographic waveforms
  44. Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
  45. Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
  46. Nonlinear causal influences assessed by mutual compression entropy
  47. Comparative study of methods for solving the correspondence problem in EMD applications
  48. fNIRS for future use in auditory diagnostics
  49. Semi-automated detection of fractional shortening in zebrafish embryo heart videos
  50. Blood pressure measurement on the cheek
  51. Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
  52. Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
  53. An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
  54. Real-time QRS detection using integrated variance for ECG gated cardiac MRI
  55. Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
  56. In-vitro experiments to characterize ventricular electromechanics
  57. Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
  58. Application of microwave sensor technology in cardiovascular disease for plaque detection
  59. Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
  60. Detection of microsleep events in a car driving simulation study using electrocardiographic features
  61. A method to determine the kink resistance of stents and stent delivery systems according to international standards
  62. Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
  63. Transient Euler-Lagrange/DEM simulation of stent thrombosis
  64. Automated control of the laser welding process of heart valve scaffolds
  65. Automation of a test bench for accessing the bendability of electrospun vascular grafts
  66. Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
  67. Cryopreservation of cells using defined serum-free cryoprotective agents
  68. New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
  69. Determination of the membrane hydraulic permeability of MSCs
  70. Climate retainment in carbon dioxide incubators
  71. Multiple factors influencing OR ventilation system effectiveness
  72. Evaluation of an app-based stress protocol
  73. Medication process in Styrian hospitals
  74. Control tower to surgical theater
  75. Development of a skull phantom for the assessment of implant X-ray visibility
  76. Surgical navigation with QR codes
  77. Investigation of the pressure gradient of embolic protection devices
  78. Computer assistance in femoral derotation osteotomy: a bottom-up approach
  79. Automatic depth scanning system for 3D infrared thermography
  80. A service for monitoring the quality of intraoperative cone beam CT images
  81. Resectoscope with an easy to use twist mechanism for improved handling
  82. In vitro simulation of distribution processes following intramuscular injection
  83. Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
  84. A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
  85. Smart control for functional electrical stimulation with optimal pulse intensity
  86. Tactile display on the remaining hand for unilateral hand amputees
  87. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
  88. An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
  89. Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
  90. Learning discriminative classification models for grading anal intraepithelial neoplasia
  91. Regularization of EIT reconstruction based on multi-scales wavelet transforms
  92. Assessing MRI susceptibility artefact through an indicator of image distortion
  93. EyeGuidance – a computer controlled system to guide eye movements
  94. A framework for feedback-based segmentation of 3D image stacks
  95. Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
  96. 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
  97. Inside-Out access strategy using new trans-vascular catheter approach
  98. US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
  99. Impact of different registration methods in MEG source analysis
  100. 3D segmentation of thyroid ultrasound images using active contours
  101. Designing a compact MRI motion phantom
  102. Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
  103. Classification of indirect immunofluorescence images using thresholded local binary count features
  104. Analysis of muscle fatigue conditions using time-frequency images and GLCM features
  105. Numerical evaluation of image parameters of ETR-1
  106. Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
  107. Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
  108. Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
  109. Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
  110. Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
  111. Evaluation of open-source software for the lung segmentation
  112. Automatic determination of lung features of CF patients in CT scans
  113. Image analysis of self-organized multicellular patterns
  114. Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
  115. Radiopacity assessment of neurovascular implants
  116. Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
  117. Development of an artifact-free aneurysm clip
  118. Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
  119. Rapid prototyping of replica knee implants for in vitro testing
  120. Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
  121. Advanced wettability analysis of implant surfaces
  122. Patient-specific hip prostheses designed by surgeons
  123. Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
  124. Wear of a total intervertebral disc prosthesis
  125. Digital health and digital biomarkers – enabling value chains on health data
  126. Usability in the lifecycle of medical software development
  127. Influence of different test gases in a non-destructive 100% quality control system for medical devices
  128. Device development guided by user satisfaction survey on auricular vagus nerve stimulation
  129. Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
  130. Effect of left atrial hypertrophy on P-wave morphology in a computational model
  131. Simulation of intracardiac electrograms around acute ablation lesions
  132. Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
  133. Assessment of nasal resistance using computational fluid dynamics
  134. Resistance in a non-linear autoregressive model of pulmonary mechanics
  135. Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
  136. Determination of regional lung function in cystic fibrosis using electrical impedance tomography
  137. Development of parietal bone surrogates for parietal graft lift training
  138. Numerical simulation of mechanically stimulated bone remodelling
  139. Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
  140. Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
  141. Principle study on the signal connection at transabdominal fetal pulse oximetry
  142. Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
  143. Evaluating different approaches to identify a three parameter gas exchange model
  144. Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
  145. From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
  146. Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
  147. Improved FPGA controlled artificial vascular system for plethysmographic measurements
  148. Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
  149. Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
  150. Control scheme selection in human-machine- interfaces by analysis of activity signals
  151. Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
  152. Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
  153. Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
  154. Interactive monitoring system for visual respiratory biofeedback
  155. Development of a low-cost senor based aid for visually impaired people
  156. Patient assistive system for the shoulder joint
  157. A passive beating heart setup for interventional cardiology training
Downloaded on 28.2.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2016-0053/html
Scroll to top button