Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion

  • Mark Verjans EMAIL logo , Nad Siroros , Jörg Eschweiler and Klaus Radermacher
Published/Copyright: September 30, 2016

Abstract

The human shoulder is one of the most complex joints of the human body, and due to the high range of motion and the complex soft tissue apparatus prone to injuries. Surgical therapies and joint replacements often lead to unsatisfactory results. To improve the understanding of the complex biomechanics of the shoulder, experimental investigations have to be conducted. For this purpose a new shoulder simulator with an innovative muscle force generation was developed. On the basis of a modular concept six artificial pneumatic muscles were integrated to represent the functionally most important muscles of the shoulder joint, whereby a free and controlled movement of the humerus can be conducted. For each muscle individual setpoints for muscle length control based on a user defined shoulder movement for any artificial or cadaver specimen are created by manual motion “Teach-In”. Additional to muscle forces and lengths, optical tracking and a joint force measurement is used to enable different biomechanical studies of the shoulder joint. This paper describes the technical setup as well as the control strategy and first results of its experimental functional validation.

1 Introduction

The human shoulder takes on a special role in the human body due to of its specific requirements of stability and mobility [1]. In Germany, approximately 12,000 shoulder prostheses are implanted each year, making them less common than artificial knees or hips [2]. The main reason for lower numbers of shoulder prostheses is the high complexity of the human shoulder joint, whereby it is difficult to find adequate technical shoulder models for replacement. Therefore, the results of shoulder prostheses and other surgical interventions are often unsatisfactory, which motivates the endeavour to improve existing shoulder prostheses and optimize surgical interventions concerning the shoulder joint [3]. For this reason, it is necessary to enhance the knowledge of biomechanical behaviour of the shoulder joint. This knowledge forms the basis for the development of modern concepts of reconstructive surgery and arthroplasty [1]. One possibility to learn more about shoulder biomechanics is the use of an experimental shoulder testing apparatus e.g. for motion and sensitivity analysis. The majority of the simulators has focused on evaluating shoulder biomechanics while the joint is statically positioned, or when passive motions are externally applied with or without muscle loading [4]. Kedgley et al. showed that simulators using continually variable muscle forces to drive shoulder movement produce motions with higher repeatability than passive systems because the approximation to the physiological case is closer [5]. Only a few systems have investigated joint kinematics and kinetics using active muscle driven motion [4].

One limitation of most of the existing simulators is the fixed force generation for an active movement of the joint. Either an estimated fixed ratio of muscle forces is applied. Often the setpoint of the force of the deltoid muscle is set manually and the other muscles forces are scaled with a fixed ratio based on the physiological cross-sectional area (PCSA) of each muscle. The fixed ratio of muscle forces can result in an uncontrolled shoulder movement. Furthermore, alternatively an external rail is used to guide the humerus on a given path and tampers the forces by adding non physiological guidance forces.

Neither a fixed force ratio nor a guidance rail seems to generate physiological conditions to reproduce biomechanical shoulder motion behaviour. Apart from this electromyography (EMG) is used to determine muscle activation and estimate the muscle forces [5], [6]. Due to cross-talk of the muscles, the reliability of the EMG-data is questionable.

To overcome the limitations of existing shoulder simulators a new and innovative shoulder testing apparatus was developed with two main objectives:

  • A stable control algorithm for reproducible (<5°) free motion with a redundant muscle setup for cadaver specimen

  • Qualitative and quantitative (<10° deviation) accordance between passive Teach-In and active free shoulder movement

In the context of a first pilot study the potentials of the new shoulder motion simulator are investigated.

2 Material and methods

2.1 Shoulder simulator – technical setup

One synthetic shoulder and one cadaver specimen were tested on the new biomechanical shoulder simulator (Figure 1). The new shoulder simulator contains six active pneumatic muscles (DMSP, Festo, Esslingen, Germany) which are connected via UHMWPE-ropes and ball bearing pulleys to the corresponding muscle tendons. Thus the three parts of the deltoid muscle and the rotator-cuff muscles [supraspinatus, infraspinatus + teres minor (combined), subscapularis] can be actively controlled. The advantages of the used pneumatic muscles are the high force density and the inherent compliance, which is needed to fit into the elastic properties of cadaver specimen. Furthermore, two passive muscles are realized with springs (pectoralis major muscle combined with latissimus dorsi muscle and biceps brachii muscle).

Figure 1: Schematic of the biomechanical shoulder simulator (simplified to 1 artificial muscle).
Figure 1:

Schematic of the biomechanical shoulder simulator (simplified to 1 artificial muscle).

Adverse the pneumatic muscles have a highly nonlinear behaviour which makes control difficult. Therefore, a nonlinear adaptive controller for force and length was developed on the basis of Zeng and Wan [7]. With an additional compensation for the nonlinear characteristics of the used pneumatic valves (VPWP Festo, Esslingen, Germany), a precise control over a wide range of muscle forces and lengths for a changing and unknown control path, namely the specimen, is expected.

An innovative and completely new approach which is realized in the new developed shoulder simulator and control algorithm, respectively, and which is contrary to the shoulder testing apparatus found in literature, is that motion of the shoulder can be controlled by muscle lengths rather than forces. This is necessary to create controlled and free shoulder-movements but depends on detailed information on movement and specimen specific muscle lengths over time. This information is acquired by a so called “Teach-In” process, where the operator moves the humerus on a desired trajectory, while the muscles are force controlled. During this movement the muscles follow the forced movement, and the control system records the required muscle lengths for the realisation of the specific trajectory. After this procedure the system can use the measured, over the trajectory varying muscle lengths, to replay the shoulder movement without the operator’s guidance.

The shoulder simulator setup is equipped with measurement devices for muscle lengths (WS10SG, ASM GmbH, Moosinning, Germany) and forces (KM30z, ME-Messsysteme, Henningsdorf, Germany), a 6D-force torque sensor (ATI, Apex, USA) for the joint reaction forces and moments, and an optical tracking system (Polaris Spectra, NDI, Ontario, Canada) to record the arm motion. A real-time control system (MicroAutoBoxII, dSPACE, Paderborn, Germany) was used for data recording, control and communication with all included devices.

2.2 Experimental validation

To evaluate the control concept and to analyse the achievable range of motion (ROM) of different movements, an experiment with a synthetic shoulder joint was initially realized. A simplified shoulder model, made from synthetic bones and elastic rubber bands representing the capsular ligament system (glenohumeral ligament &amp; coracohumeral ligament), is attached to the shoulder simulator. This was done by fixating the inferior part of the scapula in polyurethane (PUR) foam and then mounting it to the motion simulator. Scapulae were mounted in the setup with the neutral plane of the glenoid tilted 10° superiorly, the scapula tilted 20° anteriorly, and the plane of the scapula parallel to the middle deltoid line of action [8]. The muscle and tendon attachments and pulling directions are estimated based on anatomical landmarks and the joint surfaces of the synthetic shoulder are lubricated with petrolatum.

In a second step the experiments are repeated with a human cadaver specimen (Figure 2). One fresh-frozen, upper extremity was obtained from a female donor (age: 85 years). The cadaver was free of disease, and the joint capsule was not vented. The arm was thawed at approximately 20°C for 24 h and kept hydrated with normal saline during preparation. The scapula was exposed from the medial border to the suprascapular notch before being embedded with PUR foam in a rectangular block. Then the fixed shoulder was placed into the motion simulator. The distal end of the humerus is likewise attached to a cylinder where the rigid body of the optical tracking system and the spring representing the biceps brachii muscle is fixed. The artificial muscles are connected by chirurgical suture using fibre wire cords.

Figure 2: Human cadaver shoulder in the Aachen shoulder simulator.
Figure 2:

Human cadaver shoulder in the Aachen shoulder simulator.

The desired movement of the arm was then performed manually by the operator while the artificial muscles are on force control and the variable muscle lengths are recorded. After this “Teach-In” procedure the movement of the arm was performed by the active length controlled muscles. That way both synthetic and cadaver specimen were articulated in different characteristic motions (abduction/adduction, internal/external rotation with adducted arm, anteversion/retroversion) through a physiologic ROM.

3 Results

In the context of the experiment with the synthetic shoulder model the general usability of the simulator and its novel control concept were evaluated. This setup enabled a reproducible active movement (reproducibility of active <5°) with a good agreement (<10°) between the passive “Teach-In” movement and the replayed active movement (exemplarily an abduction is shown in Figure 3).

Figure 3: Shoulder kinematics of an abduction of a synthetic shoulder model in the new shoulder simulator.
Figure 3:

Shoulder kinematics of an abduction of a synthetic shoulder model in the new shoulder simulator.

Optimal results were performed at abduction/adduction with a successful ROM from 0° to 90°. The range of internal- and external rotation was 80°–0°–25°.

With the use of a human cadaver shoulder movements could be successfully performed as well. Although, the successful ROM was smaller than with the synthetic model movements could be performed (abduction movement in Figure 4).

Figure 4: Shoulder kinematics of an abduction of a cadaver shoulder specimen in the new shoulder simulator.
Figure 4:

Shoulder kinematics of an abduction of a cadaver shoulder specimen in the new shoulder simulator.

The difference of the passive “Teach-In” and active replay movement was in general greater than with the synthetic shoulder model, because the sutured attachments of the artificial muscles were more elastic, and therefore, the elongation of the muscles was adulterated. With the human cadaver a full ROM of 35°–85° (abduction/adduction), −30° to 0° (anteversion) and −25° to 0° to 20° (internal/external rotation) could be performed.

4 Discussion and conclusion

A new and innovative motion simulator for testing the biomechanical behaviour of the shoulder joint was presented. This first study shows that repeatable movements of the shoulder can be easily performed with the new developed shoulder simulator.

The pneumatic muscles in combination with the implemented adaptive force and length controller allowed the precise control of a compliant actor in combination with an unknown control path. In combination with the novel muscle control concept free but stable and controlled movements with a redundant muscle setup of individual specimen are possible, without the use of EMG data or external rails. This enables a major improvement towards free motion simulation compared to existing biomechanical shoulder simulators. Furthermore, the new shoulder simulator offers the opportunity to analyse the needed muscle forces, instead of estimating them a-priori. In this context our cadaver experiment confirmed the thesis that there is no approximately constant ratio of the muscle forces during an abduction movement (Figure 5).

Figure 5: Applied muscle forces during abduction in relation to the combined force of the deltoid muscle (=Spinal + Acromial + Clavicular) on a cadaver specimen.
Figure 5:

Applied muscle forces during abduction in relation to the combined force of the deltoid muscle (=Spinal + Acromial + Clavicular) on a cadaver specimen.

The follow-up behaviour of the active replay movement compared to the passive “Teach-In” movement with the synthetic shoulder satisfied the qualitative and quantitative aspiration, with deviations <10°. Although the follow-up behaviour with cadaver specimen was not as good as with a synthetic model, a good qualitative accordance as well as a high reproducibility of active motion were achieved. This can be further optimized by enlarging the stiffness of the connection to the muscle tendon.

There exist some limitations coming along with the new shoulder simulator. For the pneumatic muscles, an optimization for a more physiological behaviour is one objective of our ongoing work. Because especially at turning points of the muscle length setpoints, e.g. at the changing point between abduction to adduction, the muscle tension was loose, so that no force was transmitted. This is not in accordance with the physiological ground tone of each muscle.

Our ongoing work will focus on the optimization of the length controller to solve the problem by always applying a force greater than a specified ground tone, even if it leads to deviations from the setpoint. Optimized tendon clamps and more suitable suture techniques are already under construction and evaluation. Further biomechanical studies with the shoulder simulator are planned to provide answers to specific clinical questions and to investigate different surgical strategies.

Acknowledgement

The authors would like to thank Professor A. Prescher, Institute for Anatomy and Björn Rath, Orthopaedic Clinic, RWTH Aachen University, for providing the cadaver specimen preparation. This project has been funded in parts in the framework of the START program of the Medical Faculty of the RWTH Aachen University.

Author’s Statement

Research funding: The author state no funding involved. Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent is not applicable. Ethical approval: The conducted research is not related to either human or animal use.

References

[1] Gohlke F. Biomechanik der Schulter. Der Orthopäde. 2000;29:834–44.10.1055/b-0034-7089Search in Google Scholar

[2] Loew M, editor. AE-Manual der Endoprothetik. Schulter: Springer-Verlag; 2009.10.1007/978-3-642-02854-0Search in Google Scholar

[3] Wiedemann E. Schulterendoprothetik. Der Unfallchirurg. 2006;109:1073–84.10.1007/s00113-006-1204-4Search in Google Scholar PubMed

[4] Giles JW, Ferreira LM, Athwal GS, Johnson JA. Development and performance evaluation of a multi-PID muscle loading driven in vitro active-motion shoulder simulator and application to assessing reverse total shoulder arthroplasty. J Biomech Eng. 2014;136:121007.10.1115/1.4028820Search in Google Scholar PubMed

[5] Kedgley AE, Mackenzie GA, Ferreira LM, Drosdowech DS, King GJ, Faber KJ, et al. The effect of muscle loading on the kinematics of in vitro glenohumeral abduction. J Biomech. 2007;40:2953–60.10.1016/j.jbiomech.2007.02.008Search in Google Scholar PubMed

[6] Wellmann M, Petersen W, Zantop T, Schanz S, Raschke MJ, Hurschler C. Effect of coracoacromial ligament resection on glenohumeral stability under active muscle loading in an in vitro model. Arthroscopy. 2008;24:1258–64.10.1016/j.arthro.2008.01.015Search in Google Scholar PubMed

[7] Zeng Q, Wan J. Nonlinear PID control of electronic throttle valve. In Electrical and Control Engineering (ICECE), 2011 International Conference on IEEE; 2011. p. 722–4.10.1109/ICECENG.2011.6057161Search in Google Scholar

[8] Hochschild J. Strukturen und Funktionen begreifen, Funktionelle Anatomie: Band 1: Wirbelsäule und obere Extremität. Georg Thieme Verlag; 2014.10.1055/b-003-125797Search in Google Scholar

Published Online: 2016-9-30
Published in Print: 2016-9-1

©2016 Mark Verjans et al., licensee De Gruyter.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.

Articles in the same Issue

  1. Synthesis and characterization of PIL/pNIPAAm hybrid hydrogels
  2. Novel blood protein based scaffolds for cardiovascular tissue engineering
  3. Cell adhesion and viability of human endothelial cells on electrospun polymer scaffolds
  4. Effects of heat treatment and welding process on superelastic behaviour and microstructure of micro electron beam welded NiTi
  5. Long-term stable modifications of silicone elastomer for improved hemocompatibility
  6. The effect of thermal treatment on the mechanical properties of PLLA tubular specimens
  7. Biocompatible wear-resistant thick ceramic coating
  8. Protection of active implant electronics with organosilicon open air plasma coating for plastic overmolding
  9. Examination of dielectric strength of thin Parylene C films under various conditions
  10. Open air plasma deposited antimicrobial SiOx/TiOx composite films for biomedical applications
  11. Systemic analysis about residual chloroform in PLLA films
  12. A macrophage model of osseointegration
  13. Towards in silico prognosis using big data
  14. Technical concept and evaluation of a novel shoulder simulator with adaptive muscle force generation and free motion
  15. Usability evaluation of a locomotor therapy device considering different strategies
  16. Hypoxia-on-a-chip
  17. Integration of a semi-automatic in-vitro RFA procedure into an experimental setup
  18. Fabrication of MEMS-based 3D-μECoG-MEAs
  19. High speed digital interfacing for a neural data acquisition system
  20. Bionic forceps for the handling of sensitive tissue
  21. Experimental studies on 3D printing of barium titanate ceramics for medical applications
  22. Patient specific root-analogue dental implants – additive manufacturing and finite element analysis
  23. 3D printing – a key technology for tailored biomedical cell culture lab ware
  24. 3D printing of hydrogels in a temperature controlled environment with high spatial resolution
  25. Biocompatibility of photopolymers for additive manufacturing
  26. Biochemical piezoresistive sensors based on pH- and glucose-sensitive hydrogels for medical applications
  27. Novel wireless measurement system of pressure dedicated to in vivo studies
  28. Portable auricular device for real-time swallow and chew detection
  29. Detection of miRNA using a surface plasmon resonance biosensor and antibody amplification
  30. Simulation and evaluation of stimulation scenarios for targeted vestibular nerve excitation
  31. Deep brain stimulation: increasing efficiency by alternative waveforms
  32. Prediction of immediately occurring microsleep events from brain electric signals
  33. Determining cardiac vagal threshold from short term heart rate complexity
  34. Classification of cardiac excitation patterns during atrial fibrillation
  35. An algorithm to automatically determine the cycle length coverage to identify rotational activity during atrial fibrillation – a simulation study
  36. Deriving respiration from high resolution 12-channel-ECG during cycling exercise
  37. Reducing of gradient induced artifacts on the ECG signal during MRI examinations using Wilcoxon filter
  38. Automatic detection and mapping of double potentials in intracardiac electrograms
  39. Modeling the pelvic region for non-invasive pelvic intraoperative neuromonitoring
  40. Postprocessing algorithm for automated analysis of pelvic intraoperative neuromonitoring signals
  41. Best practice: surgeon driven application in pelvic operations
  42. Vasomotor assessment by camera-based photoplethysmography
  43. Classification of morphologic changes in photoplethysmographic waveforms
  44. Novel computation of pulse transit time from multi-channel PPG signals by wavelet transform
  45. Efficient design of FIR filter based low-pass differentiators for biomedical signal processing
  46. Nonlinear causal influences assessed by mutual compression entropy
  47. Comparative study of methods for solving the correspondence problem in EMD applications
  48. fNIRS for future use in auditory diagnostics
  49. Semi-automated detection of fractional shortening in zebrafish embryo heart videos
  50. Blood pressure measurement on the cheek
  51. Derivation of the respiratory rate from directly and indirectly measured respiratory signals using autocorrelation
  52. Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
  53. An automatic systolic peak detector of blood pressure waveforms using 4th order cumulants
  54. Real-time QRS detection using integrated variance for ECG gated cardiac MRI
  55. Preprocessing of unipolar signals acquired by a novel intracardiac mapping system
  56. In-vitro experiments to characterize ventricular electromechanics
  57. Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip
  58. Application of microwave sensor technology in cardiovascular disease for plaque detection
  59. Artificial blood circulatory and special Ultrasound Doppler probes for detecting and sizing gaseous embolism
  60. Detection of microsleep events in a car driving simulation study using electrocardiographic features
  61. A method to determine the kink resistance of stents and stent delivery systems according to international standards
  62. Comparison of stented bifurcation and straight vessel 3D-simulation with a prior simulated velocity profile inlet
  63. Transient Euler-Lagrange/DEM simulation of stent thrombosis
  64. Automated control of the laser welding process of heart valve scaffolds
  65. Automation of a test bench for accessing the bendability of electrospun vascular grafts
  66. Influence of storage conditions on the release of growth factors in platelet-rich blood derivatives
  67. Cryopreservation of cells using defined serum-free cryoprotective agents
  68. New bioreactor vessel for tissue engineering of human nasal septal chondrocytes
  69. Determination of the membrane hydraulic permeability of MSCs
  70. Climate retainment in carbon dioxide incubators
  71. Multiple factors influencing OR ventilation system effectiveness
  72. Evaluation of an app-based stress protocol
  73. Medication process in Styrian hospitals
  74. Control tower to surgical theater
  75. Development of a skull phantom for the assessment of implant X-ray visibility
  76. Surgical navigation with QR codes
  77. Investigation of the pressure gradient of embolic protection devices
  78. Computer assistance in femoral derotation osteotomy: a bottom-up approach
  79. Automatic depth scanning system for 3D infrared thermography
  80. A service for monitoring the quality of intraoperative cone beam CT images
  81. Resectoscope with an easy to use twist mechanism for improved handling
  82. In vitro simulation of distribution processes following intramuscular injection
  83. Adjusting inkjet printhead parameters to deposit drugs into micro-sized reservoirs
  84. A flexible standalone system with integrated sensor feedback for multi-pad electrode FES of the hand
  85. Smart control for functional electrical stimulation with optimal pulse intensity
  86. Tactile display on the remaining hand for unilateral hand amputees
  87. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test
  88. An improved tracking framework for ultrasound probe localization in image-guided radiosurgery
  89. Improvement of a subviral particle tracker by the use of a LAP-Kalman-algorithm
  90. Learning discriminative classification models for grading anal intraepithelial neoplasia
  91. Regularization of EIT reconstruction based on multi-scales wavelet transforms
  92. Assessing MRI susceptibility artefact through an indicator of image distortion
  93. EyeGuidance – a computer controlled system to guide eye movements
  94. A framework for feedback-based segmentation of 3D image stacks
  95. Doppler optical coherence tomography as a promising tool for detecting fluid in the human middle ear
  96. 3D Local in vivo Environment (LivE) imaging for single cell protein analysis of bone tissue
  97. Inside-Out access strategy using new trans-vascular catheter approach
  98. US/MRI fusion with new optical tracking and marker approach for interventional procedures inside the MRI suite
  99. Impact of different registration methods in MEG source analysis
  100. 3D segmentation of thyroid ultrasound images using active contours
  101. Designing a compact MRI motion phantom
  102. Cerebral cortex classification by conditional random fields applied to intraoperative thermal imaging
  103. Classification of indirect immunofluorescence images using thresholded local binary count features
  104. Analysis of muscle fatigue conditions using time-frequency images and GLCM features
  105. Numerical evaluation of image parameters of ETR-1
  106. Fabrication of a compliant phantom of the human aortic arch for use in Particle Image Velocimetry (PIV) experimentation
  107. Effect of the number of electrodes on the reconstructed lung shape in electrical impedance tomography
  108. Hardware dependencies of GPU-accelerated beamformer performances for microwave breast cancer detection
  109. Computer assisted assessment of progressing osteoradionecrosis of the jaw for clinical diagnosis and treatment
  110. Evaluation of reconstruction parameters of electrical impedance tomography on aorta detection during saline bolus injection
  111. Evaluation of open-source software for the lung segmentation
  112. Automatic determination of lung features of CF patients in CT scans
  113. Image analysis of self-organized multicellular patterns
  114. Effect of key parameters on synthesis of superparamagnetic nanoparticles (SPIONs)
  115. Radiopacity assessment of neurovascular implants
  116. Development of a desiccant based dielectric for monitoring humidity conditions in miniaturized hermetic implantable packages
  117. Development of an artifact-free aneurysm clip
  118. Enhancing the regeneration of bone defects by alkalizing the peri-implant zone – an in vitro approach
  119. Rapid prototyping of replica knee implants for in vitro testing
  120. Protecting ultra- and hyperhydrophilic implant surfaces in dry state from loss of wettability
  121. Advanced wettability analysis of implant surfaces
  122. Patient-specific hip prostheses designed by surgeons
  123. Plasma treatment on novel carbon fiber reinforced PEEK cages to enhance bioactivity
  124. Wear of a total intervertebral disc prosthesis
  125. Digital health and digital biomarkers – enabling value chains on health data
  126. Usability in the lifecycle of medical software development
  127. Influence of different test gases in a non-destructive 100% quality control system for medical devices
  128. Device development guided by user satisfaction survey on auricular vagus nerve stimulation
  129. Empirical assessment of the time course of innovation in biomedical engineering: first results of a comparative approach
  130. Effect of left atrial hypertrophy on P-wave morphology in a computational model
  131. Simulation of intracardiac electrograms around acute ablation lesions
  132. Parametrization of activation based cardiac electrophysiology models using bidomain model simulations
  133. Assessment of nasal resistance using computational fluid dynamics
  134. Resistance in a non-linear autoregressive model of pulmonary mechanics
  135. Inspiratory and expiratory elastance in a non-linear autoregressive model of pulmonary mechanics
  136. Determination of regional lung function in cystic fibrosis using electrical impedance tomography
  137. Development of parietal bone surrogates for parietal graft lift training
  138. Numerical simulation of mechanically stimulated bone remodelling
  139. Conversion of engineering stresses to Cauchy stresses in tensile and compression tests of thermoplastic polymers
  140. Numerical examinations of simplified spondylodesis models concerning energy absorption in magnetic resonance imaging
  141. Principle study on the signal connection at transabdominal fetal pulse oximetry
  142. Influence of Siluron® insertion on model drug distribution in the simulated vitreous body
  143. Evaluating different approaches to identify a three parameter gas exchange model
  144. Effects of fibrosis on the extracellular potential based on 3D reconstructions from histological sections of heart tissue
  145. From imaging to hemodynamics – how reconstruction kernels influence the blood flow predictions in intracranial aneurysms
  146. Flow optimised design of a novel point-of-care diagnostic device for the detection of disease specific biomarkers
  147. Improved FPGA controlled artificial vascular system for plethysmographic measurements
  148. Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation
  149. Automated detection of alveolar arches for nasoalveolar molding in cleft lip and palate treatment
  150. Control scheme selection in human-machine- interfaces by analysis of activity signals
  151. Event-based sampling for reducing communication load in realtime human motion analysis by wireless inertial sensor networks
  152. Automatic pairing of inertial sensors to lower limb segments – a plug-and-play approach
  153. Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor
  154. Interactive monitoring system for visual respiratory biofeedback
  155. Development of a low-cost senor based aid for visually impaired people
  156. Patient assistive system for the shoulder joint
  157. A passive beating heart setup for interventional cardiology training
Downloaded on 6.3.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2016-0017/html
Scroll to top button