A novel compact highly sensitive non-invasive microwave antenna sensor for blood glucose monitoring

  • Rabah W. Aldhaheri EMAIL logo , Jagadeesh Babu Kamili , Anveshkumar Nella and Nebras M. Sobahi
Published/Copyright: September 5, 2023

Abstract

In the present work, a novel compact and highly sensitive microwave antenna sensor at 2.45 GHz is proposed for evaluating glucose concentration in blood. The antenna is printed on an FR-4 substrate of compact dimensions 35 mm × 13.5 mm × 1.6 mm. A human finger phantom model is constructed in the EM simulation high frequency structure simulator environment consisting of skin, blood, fat, and bone layers. In the study, finger models with various shapes like rectangular, cylindrical, and ellipsoid are considered, and the results are compared. The glucose concentration is changed from 0 to 500 mg/dL, and the corresponding shift is evaluated by keeping the finger phantom at different locations near the antenna. The frequency shifts obtained in the designed experiment are used to evaluate glucose concentration in blood samples. In this work, a minimum and a maximum frequency shift of around 1.25 and 5 MHz, respectively, are observed when the finger phantom is placed at the top of the radiating element. Simulated antenna results are found to be in good agreement with the measured results. The developed method is validated with a two-antenna model by calculating time delay and isolation for different glucose concentrations. An experiment of placing a real human finger around a fabricated antenna presents good correspondence with the simulation results.

1 Introduction

Diabetes is one of the major chronic diseases mainly characterized by very high sugar levels in the blood for a long time [1,2]. The common method to measure glucose levels is using a finger prick and glucometer, which is known as an invasive method [3,4]. However, the invasive methods cause pain to the patient, as blood samples are to be suctioned/pricked each time. One solution for this problem is a non-invasive method, where microwave filters or antennas are used to monitor glucose levels in the blood samples [5,6,7,8,9,10]. Recently, microwave antennas have found various biomedical applications like body wearable devices [11].

In this non-invasive method, sensors in the form of antennas or RF filters are involved in measuring glucose concentration. In non-invasive techniques, popular methods of detecting glucose concentration levels are by using microstrip resonator (filter) technique or microstrip antenna techniques. Also, these are prominently preferred in the medical field due to no risk of ionization, deep penetration, and immune to temperatures and other disturbances.

In detecting glucose concentration in blood samples, a novel design of sensors with high sensitivity is required to be designed. Various sensors like pressure sensor [12] and a narrow band antenna sensor [13] with highly sensitive characteristics are discussed in the literature. Dielectric properties of various skin layers play an important role in the design of glucose-measuring sensors [14,15,16,17]. In the study of Nella et al. [18], a novel method of precise measurement of blood glucose using planar Yagi-Uda antenna and microstrip filter was presented. In the study of Sbrignadello et al. [19], a review of different types of novel sensors for glucose monitoring is presented in an effective manner. The use of meta-materials is another choice for detecting glucose, as detailed in the study of Islam et al. [20]. The FMCW milli-meter wave RADAR sensor is utilized for detecting glucose concentration in the blood samples [21]. Similarly, various other methods like the coplanar sensor method [22], multiple split ring resonator [23], chipless tag split ring resonator, [24] etc., are employed for glucose level monitoring. In all these methods, glucose concentration in blood samples is measured by mapping various layers of skin with different dielectric constants and thus by measuring corresponding frequency shifts, which occur due to the variation of electric fields of the sensor. In the process of glucose monitoring, sensitivity and compactness of the sensor are the main issues to be considered. Hence, this work mainly focuses on the design and development of a novel sensor with compactness and giving better sensitivity characteristics. The proposed sensor operates at 2.45 GHz without phantom loaded, and it shifts near to 5 GHz when the phantom is placed over the sensor. In the existing literature, several antennas operating at the 2.45 GHz band, covering the ISM applications, are presented [2528].

The proposed microwave sensor is taken on a low-cost FR-4 substrate of dimensions 35 mm × 13.5 mm × 1.6 mm with a dielectric constant value of 4.4 and a loss tangent of 0.02. The proposed sensor gives a quasi-linear shift in the frequency of S11 with the variation of glucose concentration, thus giving more sensitivity as the glucose differs from 0 to 500 mg/dL. The proposed antenna design and sample human finger are developed in the finite-element method (FEM) solver high frequency structure simulator (HFSS).

The rest of this article is discussed as follows. Section 2 presents an evolution of the proposed antenna design and associated glucose concentration study. Also, in this section, a four-layer finger phantom model is introduced. These layers are skin, blood, fat, and bone. Each has different thicknesses and dielectric constants. As the concentration of glucose changes, the blood dielectric constant is varied, and the S11 parameters are obtained. Section 3 discusses measured reflection coefficient parameters by placing the finger of a healthy person and the finger phantom at various positions on the antenna to validate the proposed phantom model and illustrate that the model is a replica of a real human being finger to a far extent. Section 4 concludes the work.

2 Antenna design and glucose concentration monitoring

2.1 Basic rectangular patch antenna

In this design process of a highly sensitive microwave antenna sensor, a basic rectangular patch antenna of size 50 × 40 mm2 is initially considered, as shown in Figure 1(a). Feed width (W f) is 3 mm, and slit width (W S) is considered as 1 mm. The width (W) and length (L) of the patch antenna are 40 mm and 28.15 mm, respectively. The operating resonant frequency of the antenna is determined using the design basic design equations given in ref. [29]. This antenna is formed on a low-loss FR-4 substrate with a dielectric constant of 4.4. This antenna resonates at 2.45 GHz with harmonics at 3.5 and 5.5 GHz, as shown in Figure 1(b). The distribution of the E-field is shown in Figure 1(c), from which it is observed that almost no variation of filed current is observed throughout the antenna except at the outer edges.

Figure 1 
                  Patch antenna: (a) geometry, (b) S11 parameters, and (c) E-field distribution at 2.45 GHz.
Figure 1

Patch antenna: (a) geometry, (b) S11 parameters, and (c) E-field distribution at 2.45 GHz.

2.2 Design of finger phantom for glucose monitoring

The main objective of this research work is to design an antenna that can measure glucose concentration in the blood. To achieve this, a finger phantom model is developed considering the structure of a real human finger, as shown in Figure 2 [22]. In the study of Cebedio et al. [22], it is considered that a rectangular finger phantom is more suitable for glucose concentration evaluation in blood samples. Various layers of the finger, such as skin, fat, blood, and bone, are modelled using dielectric materials of different dielectric constants. Different layers are considered in square shape, viz. bone layer as S1 of size 4 × 4 mm2, blood layer as S2 of size 9 × 9 mm2, fat layer as S3 of size 10 × 10 mm2, and skin layer as S4 of size 12 × 12 mm2. The length ( L ) of the finger phantom is considered as 30 mm with height H = 12 mm. Dimensions and dielectric constants of various layers of the finger phantom are given in Table 1 [9].

Figure 2 
                  Finger phantom model: (a) top view and (b) isometric view.
Figure 2

Finger phantom model: (a) top view and (b) isometric view.

Table 1

Dielectric constant, loss tangent, and dimensions of the finger phantom at 2.45 GHz [9]

Finger parameters Skin Fat Blood Bone
Dielectric constant value (ε r) 38 6 59 11.5
Loss tangent 0.32 0.18 0.38 0.35
Dimension (mm2) S4 (12 × 12) S3 (10 × 10) S2 (9 × 9) S1 (4 × 4)

In biomedical applications, various methods are described to model various layers in the body and to obtain the equivalent dielectric constant values. Dielectric constant values of various tissues are calculated using various models like the Debye model [30] or the Cole–Cole model, and in general, the latter model is preferred in many cases [16]. This method is proven to be the more accurate method in modelling the dielectric values of various tissue layers as this method gives more correlation between electrical properties of blood plasma and glucose concentrations. As per the Cole–Cole model, the dielectric constant value can be estimated using the following equation:

(1) ε ˆ ( w ) = ε c ( w ) j ε c ( w ) = ε + n ε n 1 + ( j ω τ n ) ( 1 α n ) + σ i j ω ϵ 0 ,

where ω is the angular frequency, ε c ( ω ) is the frequency-dependent dielectric constant, ε c ( w ) is the frequency-dependent dielectric loss, ε is the high-frequency permittivity, ∆ε n is the magnitude of the dispersion, n is the order of the Cole–Cole model, α n represents the broadening of dispersion, σ i is the ionic conductivity, and τ n is the relaxation time constant.

The finger phantom is simulated in an HFSS environment using layers of dielectric materials of various values. As mentioned in Table 1, the dielectric values at 2.45 GHz are considered for a healthy human finger. However, as the glucose concentration changes in the human blood, the dielectric values are also to be modified. The change of dielectric constant values in the finger phantom for different values of glucose concentration from 0 to 500 mg/dL is given in Table 2. As given in Figure 3, the finger phantom consists of different layers like skin, muscle, blood, fat, etc., and each layer has its own dielectric constant value. When the glucose concentration in the blood changes, obviously, its dielectric value changes owing to the variations in the blood properties. The finger phantom is placed on top of the patch antenna, as shown in Figure 3(a), and for different values of glucose concentrations from 0 to 500 mg/dL, the S11 parameters are obtained, as shown in Figure 3(b). However, no shift in the frequency is observed. In the second case, the finger phantom is placed over the top of the antenna over the slots, as shown in Figure 4, and the corresponding shifts are observed in Figure 4(b). In this case, slight shifts in the frequencies are observed due to the variation of the field near the slot. Hence, an increasing or decreasing frequency shift is observed in the entire variation of glucose from 0 to 500 mg/dL. However, the obtained frequency shifts are not suitable for evaluating glucose concentration in the blood. Hence, the basic rectangular patch is not suitable for measuring glucose concentrations in the blood. Hence, in the following section, a new antenna giving considerable shifts in the frequency is described.

Table 2

Dielectric constant values of the finger phantom at 2.45 and 5 GHz

Glucose concentration (mg/dL) ε r at 2.45 GHz ε r at 5 GHz
0 70 65
125 69 64
250 68.5 63.5
500 67 62.5
Figure 3 
                  Finger phantom placed (a) on top of the antenna and (b) S11 parameters.
Figure 3

Finger phantom placed (a) on top of the antenna and (b) S11 parameters.

Figure 4 
                  Finger phantom: (a) isometric view and (b) S11 parameters.
Figure 4

Finger phantom: (a) isometric view and (b) S11 parameters.

2.3 Slotted microwave sensor

As mentioned in the earlier subsection, the basic patch-type antenna is not suitable for glucose level measurements in the blood as no considerable shift in the frequency is observed due to less variation of the electric field at most of the antenna structure except at the slit. However, more variation of the electric field on the antenna is required for glucose level measurements. This owing to the greater variation of fringing fields on the antenna results in considerable change of the dielectric constant of the antenna, which in turn gives considerable frequency shift when the finger phantom is placed near the antenna. Hence, the basic patch antenna is modified by incorporating slots to get more variation in the electric field, as shown in Figure 5. The E-field distribution is shown in Figure 5(b), from which a good variation is observed near the slots, and this variation of the current distribution is desired for glucose monitoring, as discussed in the following sections. Various antenna dimensions are taken as W = 30 mm, W 1 = 14 mm, W 2 = 12.5 mm, W 3 = 3 mm, W S = 1 mm, L 1 = 15.5 mm, and L S = 1 mm. Two rectangular slots are employed on the antenna to increase electrical length, which causes the resonant frequency to shift towards the left [26,27]. This phenomenon is used to reduce the patch antenna dimensions to 2.45 GHz. Figure 6 describes the view of the finger phantom placed over the top of the antenna. Figure 6(b) gives the cross-section E-field variation when the finger phantom is placed over the antenna sensor. This figure clearly depicts a clear variation in the electric field owing to the placement of the phantom on the sensor. This is due to the systematic design of the slots over the antenna, resulting in more field variation and thus making the sensor suitable for glucose monitoring. The presence of slots gives more variation of the electric field, thus causing variation in dielectric values, and hence, better frequency shift values can be obtained.

Figure 5 
                  Slotted microwave antenna: (a) geometry and (b) E-field distribution at 2.45 GHz.
Figure 5

Slotted microwave antenna: (a) geometry and (b) E-field distribution at 2.45 GHz.

Figure 6 
                  Phantom at the top of the antenna: (a) top view and (b) E-field distribution with finger phantom placed on top of the antenna.
Figure 6

Phantom at the top of the antenna: (a) top view and (b) E-field distribution with finger phantom placed on top of the antenna.

The S11 parameters and their variation for different glucose concentrations are shown in Figure 7(a and b), respectively. This slotted structure also resonates at the same frequency as that of a basic patch antenna, i.e. at 2.45 GHz. However, when the finger phantom is placed over the top of the antenna, as shown in Figure 6, the resonant frequency dominates at 5 GHz with good reflection coefficient values, as shown in Figure 7(b). Hence, this particular frequency is considered to study the frequency shifts w.r.t. variations in glucose concentrations. A small change in frequency shift is observed with glucose concentration variation, as shown in Figure 7(b). This shift resulted in the variation of electric field on the antenna structure, as mentioned earlier. However, this frequency shift is not sufficient to measure glucose concentrations in the blood as much shift is not obtained, and hence this slotted sensor is further modified, resulting in an arrow-shaped microwave sensor, which is discussed in the following sub-section.

Figure 7 
                  S11-parameter: (a) slotted sensor without phantom and (b) variation with glucose concentration.
Figure 7

S11-parameter: (a) slotted sensor without phantom and (b) variation with glucose concentration.

2.4 Slotted arrow-shaped microwave sensor

Since the slotted microwave sensor mentioned in the previous sub-section does not give much frequency shift, the antenna is further modified, resulting in the slotted arrow-shaped microwave sensor, as shown in Figure 8. Various dimensions of the proposed arrow-shaped sensor are given as W = 28.95 mm, W a = 16.5 mm, W b = 14 mm, W 3 = 3 mm, W S = 1 mm, L a = 4.75 mm, L b = 1 mm, α = 75°, β = 60°. Overall substrate dimensions are maintained as 35 × 13.5 mm2. Top and bottom views of the fabricated prototype are given in Figure 9(a and b), respectively. The microwave sensor is fabricated using a photo-lithographic fabrication process.

Figure 8 
                  Proposed slotted arrow-shaped microwave antenna sensor.
Figure 8

Proposed slotted arrow-shaped microwave antenna sensor.

Figure 9 
                  Fabricated prototype: (a) top view and (b) bottom view.
Figure 9

Fabricated prototype: (a) top view and (b) bottom view.

In Figure 10(a), the simulated S11-parameters are generated up to 5.5 GHz, where a second harmonic is observed at 5 GHz. Simulated and measured S11-parameters of the slotted arrow-shaped sensor are given in Figure 10(b), where the first harmonic at 2.45 GHz is considered. The modifications of the design with slots and with the model of the finger appear to increase the resonance by about 5 GHz compared to the resonance at 2.45 GHz due to a better coupling with the feedline. A slight variation in the simulated and measured values is observed owing to the fabrication tolerances. The working of the proposed slotted arrow-shaped microwave sensor can be understood by considering the electric field distribution, as shown in Figure 11. As discussed earlier, the proposed antenna should have a considerable amount of field distribution over the entire surface of the antenna such that when the finger phantom is placed over it, the resonant frequencies of the antenna are affected. The reason is, as the phantom is placed over the sensor, its fringing fields are affected, thus affecting the dielectric value of the medium surrounding it. This in turn changes the resonant frequency of the antenna, and these changes are mapped to the changes of glucose concentration. As discussed earlier, this property is very much required to measure glucose level concentration in the blood. In order to obtain S-parameters and radiation patterns, vector network analyzer with model name Agilent N5247A:A.09.90.02 is used.

Figure 10 
                  (a) Simulated S11-parameters of the slotted arrow-shaped sensor, (b) comparison with HFSS and measured results, and (c) measurement setup.
Figure 10

(a) Simulated S11-parameters of the slotted arrow-shaped sensor, (b) comparison with HFSS and measured results, and (c) measurement setup.

Figure 11 
                  Electric field variation of the slotted arrow-shaped sensor at 2.45 GHz.
Figure 11

Electric field variation of the slotted arrow-shaped sensor at 2.45 GHz.

3 Experiment on proposed antenna’s performance for various finger positions

In this section, the finger phantom is placed at various positions like bottom and top of the proposed antenna by varying glucose concentrations, and the resulting frequency shifts are observed. The detailed analysis is discussed as follows.

3.1 Finger phantom at the bottom of the antenna

The human finger phantom discussed in the earlier sub-sections is placed at the bottom of the antenna at a spacing of 0.5 and 1.5 mm away from the feed point of the antenna, as shown in Figure 12. The cross-section of the E-field distribution on the phantom is given in Figure 12(c), from which it is observed that a considerable amount of E-field variation is observed throughout the phantom. For various levels of glucose concentration, i.e. from 0 to 500 mg/dL, the resultant S11 parameters are plotted as given in Figure 13. The graph indicates no variation in the frequency shifts, which is obvious as there is no field variation on the ground plane. However, this case is not helpful in evaluating glucose concentration as the shift is less. The next case of keeping a phantom on the top of the antenna is studied in the next sub-section.

Figure 12 
                  Phantom at the bottom of the antenna: (a) top view, (b) side view, and (c) E-field distribution with finger phantom placed.
Figure 12

Phantom at the bottom of the antenna: (a) top view, (b) side view, and (c) E-field distribution with finger phantom placed.

Figure 13 
                  S11-parameters variation with glucose concentration.
Figure 13

S11-parameters variation with glucose concentration.

3.2 Finger phantom at the top of the antenna

Initially, the finger phantom is placed at the top of the radiating element at a height of 0 mm and at 0 mm separation from the input port, as shown in Figure 14(a and b). For various levels of glucose concentration, i.e. from 0 to 500 mg/dL, the resultant S11 parameters are plotted as given in Figure 14(b). From the obtained graph, much frequency shift is not observed w.r.t. the glucose concentration variation and hence this case is not considered.

Figure 14 
                  Phantom at the top of the antenna at 0 mm height and at 0 mm distance from input port: (a) top view and (b) S11 parameters.
Figure 14

Phantom at the top of the antenna at 0 mm height and at 0 mm distance from input port: (a) top view and (b) S11 parameters.

Another case of keeping finger phantom at the top of the sensor at a height of 0 mm and at a distance of 1.5 mm from the input port is considered as shown in Figure 15(a) and the resultant S11 parameters are given in Figure 15(b). From the S parameters, it is observed that no frequency shift is observed as the phantom and radiating patch are in direct contact. Hence, this case is also not useful in determining glucose concentration in the blood samples.

Figure 15 
                  Phantom at the top of the antenna at 0 mm height and at 1.5 mm distance from the input port of antenna: (a) side view and (b) S11 parameters.
Figure 15

Phantom at the top of the antenna at 0 mm height and at 1.5 mm distance from the input port of antenna: (a) side view and (b) S11 parameters.

After performing exhaustive simulations by keeping the finger phantom at different locations on the antenna, considerable frequency shifts with change in glucose concentration are finally observed when the finger phantom is at the top of the antenna at a height of 0.5 and 1.5 mm away from the antenna excitation, as shown in Figure 16. The proposed sensor is suitable for real-time calculations of the glucose variations in the blood samples as its dimensions are compatible with the real human finger, as shown in Figure 16(b). From Figure 16(b), it is observed that the size of the proposed antenna is compatible with real human finger. This would enable us to claim that the proposed sensor can be used in the real-time applications for the blood glucose level measurements. Figure 16(c) gives the cross-section E-field variation with the finger phantom placed on the antenna, from which maximum variation of E-field compared to all the previous cases is observed when the finger phantom is placed above the antenna at a height of 0.5 and 1.5 mm away from the antenna excitation.

Figure 16 
                  Phantom at the top of the antenna at 0.5 mm height and 1.5 mm distance from the input port of antenna: (a) top view, (b) at measurement, and (c) E-field distribution with finger phantom placed.
Figure 16

Phantom at the top of the antenna at 0.5 mm height and 1.5 mm distance from the input port of antenna: (a) top view, (b) at measurement, and (c) E-field distribution with finger phantom placed.

For various levels of glucose concentration, i.e. from 0 to 500 mg/dL, the resultant S11 parameters are plotted as given in Figure 17(a). From Figure 17(a), it is observed that a clear and measurable frequency shift is observed by varying the glucose concentration levels. The obtained shifts with different glucose levels are given in Table 3, from which it is observed that good frequency shifts are obtained with changes in the glucose concentrations. When the glucose concentration is varied from 0 to 75 mg/dL, a frequency shift of 5 MHz is observed. When the concentration is varied from 75 to 125 mg/dL, a frequency shift of 4 MHz is observed. Similarly, the change of glucose concentration from 125 to 175 mg/dL gives a frequency shift of 2 MHz. From 175 to 250 mg/dL variation, a frequency shift of 3 GHz is obtained. From 250 to 375 mg/dL variation, a frequency shift of 1.75 GHz is obtained. Finally, when the concentration is in the range of 375–500 mg/dL, the obtained frequency shift is 1.25 GHz. As a clear shift in the frequency is observed for each case of glucose concentration variation, the proposed microwave sensor is a suitable device for predicting the glucose in the blood with minimum error. From the values given in Table 2, it is observed that the proposed sensor gives good sensitive values at lower glucose concentration levels, i.e. in the range of 0–250 mg/dL.

Figure 17 
                  (a) S11-parameter variation with glucose concentration, (b) comparison with measured value for 0 mg/dL case, (c) glucose concentration vs frequency shift, and (d) resonant frequency vs frequency shift.
Figure 17

(a) S11-parameter variation with glucose concentration, (b) comparison with measured value for 0 mg/dL case, (c) glucose concentration vs frequency shift, and (d) resonant frequency vs frequency shift.

Table 3

Summary of resonant frequencies obtained for different glucose concentrations

Glucose concentration (mg/dL) Resonant frequency (GHz) Bandwidth (GHz) Q-factor Frequency shift (MHz) Sensitivity (MHz/(mg/mL))
2.45 0.035 70
0 4.8710 0.2571 18.946
75 4.8760 0.2585 18.8627 5 6.66
125 4.8800 0.2596 18.7981 4 8
175 4.8820 0.2594 18.8203 2 4
200 4.8830 0.2591 18.846 1 4
225 4.8840 0.2588 18.871 1 4
250 4.8850 0.2592 18.8464 1 4
300 4.8855 0.2598 18.804 0.5 1
350 4.8860 0.2602 18.776 0.5 1
375 4.88675 0.2612 18.7088 0.75 3
400 4.8870 0.2617 18.672 0.25 1
450 4.8873 0.2620 18.653 0.3 0.6
500 4.8880 0.2622 18.6422 0.7 1.4

For the case of 0 mg/dL, S11 parameters are measured by keeping the finger of a normal person on the top of the antenna, as shown in Figure 16(b), and are closely matched with simulation values and the comparison of both simulated and measured values are given in Figure 17(b). In addition to the frequency shift values, the other important parameters to be studied in Table 2 are sensitivity values. Also, good sensitivity values are obtained using the proposed microwave sensor, and a maximum sensitivity of 8.0 MHz/mg/mL is obtained as the glucose concentration changes from 75 to 125 mg/dL, as given in Figure 17(c). Also, the variation of resonant frequency w.r.t. frequency is given in Figure 17(d). The sensitivity plots are obtained using the least-square error method, and the curve fitting is done by considering different ranges of glucose concentrations, and the plots are given in Figure 18. Figure 18(a) gives the sensitivity plot by considering glucose concentration in the range of 75–350 mg/dL, and the sensitivity is observed to be 3.5222 MHz/mg/mL, which is obtained from the slope of the curve. Similarly, Figure 18(b) gives the sensitivity value of 3.2881 MHz/mg/mL, which is plotted by considering glucose concentration in the range 125–300 mg/dL.

Figure 18 
                  Sensitivity plots obtained using the least-square error method for varying glucose concentration from (a) 75 to 350 mg/dL and (b) 125 to 300 mg/dL.
Figure 18

Sensitivity plots obtained using the least-square error method for varying glucose concentration from (a) 75 to 350 mg/dL and (b) 125 to 300 mg/dL.

In order to validate, whether the square-shaped finger phantom chosen in the present experiment is apt or not, two more finger phantom shapes, viz. cylindrical and ellipsoid, are considered, as shown in Figure 19, and the frequency shifts are generated in Figure 20. From the plotted graphs, it is observed that either cylindrical or elliptical phantoms do not give satisfactory frequency shifts, and hence, the square-shaped finger phantom chosen in the experiment is the apt choice for glucose monitoring in the blood. The details are given in Table 4, from which it is observed that the better frequency shifts are obtained using a square-shaped phantom, as the glucose changes from 0 to 500 mg/dL, whereas using cylindrical and ellipsoid phantoms, the total shifts obtained are too low making them unsuitable for glucose monitoring. Hence, for the present experiment, square-shaped phantom is the apt choice.

Figure 19 
                  Various other shapes of finger phantom: (a) cylindrical finger phantom and (b) ellipsoid finger phantom.
Figure 19

Various other shapes of finger phantom: (a) cylindrical finger phantom and (b) ellipsoid finger phantom.

Figure 20 
                  Frequency shifts obtained using: (a) cylindrical finger phantom and (b) ellipsoid finger phantom.
Figure 20

Frequency shifts obtained using: (a) cylindrical finger phantom and (b) ellipsoid finger phantom.

Table 4

Comparison of frequency shifts with various types of finger phantoms

Glucose concentration (mg/dL) Frequency shift – cylindrical phantom (MHz) Frequency shift – ellipsoid phantom (MHz) Frequency shift – square-shaped phantom (MHz)
0
125 0 0.1 9
250 0.1 0.1 5
500 0.1 0.3 3

From the results discussed in this sub-section and previous sub-section, it is concluded that the better position to place the finger to measure the variation of glucose concentrations is on the top of the antenna at a height of 0.5 mm and a distance of 1.5 mm from the input port of antenna element. This is because a clear frequency shift is observed when the finger of a healthy person or a glucose concentration varying finger is placed above the antenna.

3.3 Two antenna models for glucose monitoring

From the analysis presented in the previous sub-section, it is concluded that better frequency shifts are obtained when the finger phantom is placed over the top of the antenna at a height of 0.5 mm and a distance of 1.5 mm from the input port of the antenna element. In the present sub-section, for the first time, a two-antenna model is presented for glucose level monitoring in blood samples. The finger phantom is placed between the two antennas, as shown in Figure 21, and two important parameters, viz. group delay and isolation, are evaluated for different glucose concentrations.

Figure 21 
                  Two antenna models: (a) side view and (b) transparent view.
Figure 21

Two antenna models: (a) side view and (b) transparent view.

In Figure 22(a), the variation in group delays with respect to frequencies is shown for different glucose concentrations, viz. 0, 125, 250, and 500 mg/dL. From the plot, it is observed that group delay is decreasing when the glucose concentration increases from 0 to 500 mg/dL. This is due to the fact that when the glucose concentration increases, a dielectric constant of the finger phantom decreases. The dielectric constant consists of real and imaginary parts, and the rate of decrease in the imaginary part is higher. Hence, the slope of the curve decreases as the concentration increases. The group delay is the derivative of the angle (function of frequency), and thus, the group delay decreases with increasing glucose concentration [3]. Similarly, isolation between the two antennas for different glucose concentrations is presented in Figure 22(b). A small decrement in isolation is observed with an increase in glucose concentration. This is due to the fact that the increase in glucose concentration decreases the dielectric constant of the phantom, which is placed in between the two antennas. The lesser dielectric constant results in more surface current, giving rise to lesser coupling. However, in the present case, a small change in the isolation is observed for different glucose concentrations.

Figure 22 
                  Glucose concentration variation w.r.t. (a) group delay and (b) isolation.
Figure 22

Glucose concentration variation w.r.t. (a) group delay and (b) isolation.

The performance of the proposed microwave sensor is compared with some existing sensors in the literature that are used for glucose level monitoring and are given in Table 5. The work mentioned in the study of Turgul and Kale [31] developed a novel CPW sensor for measuring glucose concentration at 1.8 GHz operating frequency. However, the work mentions that the glucose concentrations are up to 16,000 mg/dL, which is practically not required. Also, the sensitivity values mentioned in the work are calculated from 100 to 500 mg/dL. In the study of Hasan et al. [32], a novel dielectric resonator antenna for evaluating glucose concentrations at 4.7 GHz has been presented, which also carries the calculations up to 16,000 mg/dL and getting a sensitivity value of 0.28 MHz/mg/mL. Islam et al. [33] presented the novel design of a microstrip sensor operating in the wideband of 2.5–18 GHz for measuring sugar and salt in solutions, which has the main limitation of practical reliability. Choi et al. [34] presented a novel split ring resonator for measuring glucose concentration, however, giving a sensitivity of 3.287 kHz per mM. Similarly, the design of flexible biomedical RF sensors for monitoring glucose concentration in glycerol, operating at 2.46 GHz is presented in the study of Tiwari et al. [35]. In the study of Kozak et al. [36], a patch antenna sensor operating at 2.378 GHz for the detection of frost and ice is developed. In the study of Saha et al. [37], a two-facing microstrip sensor for monitoring glucose concentration in blood is presented. This work uses transmission coefficients to evaluate the glucose concentration. In [38], a patch antenna operating at 5 GHz is presented for glucose monitoring in blood samples with a sensitivity of 0.1 MHz/mg/mL over a glucose concentration range of 0 to 250 mg/dL. In the study of Turgul and Kale [39], another single port resonator operating at 4.8 GHz, giving a sensitivity of 1.4 MHz/mg/mL, is presented. The proposed antenna sensor gives a maximum shift of 17 MHz with a sensitivity of 3.4 MHz/mg/mL, making the proposed antenna sensor a good choice for measuring the glucose level variation in blood samples.

Table 5

Performance comparison of proposed work with literature

Ref. Technology Operating frequency (GHz) Sensitivity Measuring sample
[31] Planar resonator 3.25 1.4 MHz/mg/mL 100–500 (mg/dL) (glucose variation in blood sample)
[32] Non-invasive finger placed on dielectric resonator 4.7 0.281 MHz/mg/mL 0–16,000 (mg/dL) (glucose variation in blood sample)
[33] Microstrip antenna sensor 2.5 to 18 0–80% of solution (detection of salt and sugar in water)
[34] Non-invasive wearable split ring resonator 1.5 3.287 kHz per mM 4.4–7.6 mM/L (glucose variation in microfluidic sample)
[35] Interdigitated capacitor (IDC) resonator-etched (CPW) 2.46 2 MHz/mg/mL 0–80 mg/ml (glucose concentration in glycerol)
[36] Patch antenna sensor 2.378 2–8 MHz/80–160 µL Detection of ice from 80 to 160 µL
[37] Two-facing microstrip patch antennas 60 Glucose concentration in blood up to 1.33 mmol/l
[38] Patch antenna 5 0.1 MHz/mg/mL Glucose concentration in blood from 0 to 250 mg/dL
[39] Single port resonator 4.8 1.4 MHz/mg/mL Glucose concentration in blood from 100 to 1000 mg/dL
This work Arrow-shaped compact microwave sensor 2.45 17 MHz/mg/mL Glucose concentration in blood from 0 to 500 mg/dL

4 Conclusion

A planar compact and highly sensitive arrow-shaped microwave antenna sensor is developed for measuring glucose level concentration in blood. A human finger phantom is modeled using different dielectric constants of the skin layer, and the phantom is placed at various positions around the antenna, and the corresponding frequency shifts are observed by changing the glucose concentration from 0 to 500 mg/dL. From the obtained results, it is observed that better frequency shifts are obtained if the finger is placed at the top of the radiating element at a height of 0.5 mm and at a distance of 1.5 mm from the input port. Also, a two-antenna model is presented for estimating glucose concentration by calculating time delay and isolation. Also, Q-factor and sensitivity are evaluated to understand the performance of the proposed sensor. The experiment on the human finger using an antenna-fabricated prototype also presents good correspondence with the simulation results. The developed microwave sensor is a good choice for monitoring glucose concentrations in the blood due to its compactness, high sensitivity, and affordable cost and hence can be implemented for real-time applications.

  1. Funding information: This work was funded by the Deanship of Scientific Research (DSR) at King Abdulaziz University, Jeddah, Saudi Arabia, under grant No. (RG-98-135-42).

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Conflict of interest: The authors state no conflict of interest.

References

[1] Lin X, Xu Y, Pan X, Xu J, Ding Y, Sun X, et al. Global, regional, and national burden and trend of diabetes in 195 countries and territories: An analysis from 1990 to 2025. Sci Rep. 2020;10(1):14790.10.1038/s41598-020-71908-9Search in Google Scholar PubMed PubMed Central

[2] Saeedi P, Petersohn I, Salpea P, Malanda B, Karuranga S, Unwin N, et al. Global and regional diabetes prevalence estimates for 2019 and projections for 2030 and 2045: results from the international diabetes federation diabetes atlas, 9th edition. Diabetes Res Clin Pract. 2019;157(157):107843.10.1016/j.diabres.2019.107843Search in Google Scholar PubMed

[3] Villena Gonzales W, Mobashsher A, Abbosh A. The progress of glucose monitoring—A review of invasive to minimally and non-invasive techniques, devices and sensors. Sensors. 2019;19(4):800.10.3390/s19040800Search in Google Scholar PubMed PubMed Central

[4] Juan CG, Bronchalo E, Potelon B, Quendo C, Avila-Navarro E, Sabater-Navarro JM. Concentration measurement of microliter-volume water–glucose solutions using Q factor of microwave sensors. IEEE Trans Instrum Meas. 2019;68(7):2621–34.10.1109/TIM.2018.2866743Search in Google Scholar

[5] Kumar DA, Jayanthy T. Review on non-invasive blood glucose measurement techniques. Int Conf Comm Sig Proc (ICCSP). 28–30 July 2020:1–8.10.1109/ICCSP48568.2020.9182268Search in Google Scholar

[6] Omer AE, Shaker G, Safavi-Naeini S, Alquie G, Deshours F, Kokabi H, et al. Non-Invasive real-time monitoring of glucose level using novel microwave biosensor based on triple-pole CSRR. IEEE Trans Biomed Cir Sys. 2020 Dec;14(6):1407–20.10.1109/TBCAS.2020.3038589Search in Google Scholar PubMed

[7] Yilmaz T, Ozturk T, Joof SA. Comparative study for development of microwave glucose sensors. Proc 32nd URSI GASS. 2017;1–4.Search in Google Scholar

[8] Yilmaz T, Foster R, Hao J. Radio-frequency and microwave techniques for non-invasive measurement of blood glucose levels. Diagnostics. 2019;9:1–6.10.3390/diagnostics9010006Search in Google Scholar PubMed PubMed Central

[9] Vrba J, Karch J, Vrba D. Phantoms for development of microwave sensors for noninvasive blood glucose monitoring. Int J Antennas Propag. 2015;2015:1–5.10.1155/2015/570870Search in Google Scholar

[10] Yee SK, Lim SCJ, Pong PS, Dahlan SH. Microstrip defected ground structure for determination of blood glucose concentration. Prog Electromagn Res C. 2020;99:35–48.10.2528/PIERC19110501Search in Google Scholar

[11] Januszkiewicz Ł, Barba PD, Hausman S. Optimization of wearable microwave antenna with simplified electromagnetic model of the human body. Open Phys. 2017;15(1):1055–60.10.1515/phys-2017-0133Search in Google Scholar

[12] Turgul V, Kale I. A novel pressure sensing circuit for noninvasive RF/microwave blood glucose sensor. IEEE conference Mediterrannean Microwave Symposium (MMS); 2016. p. 512–6.10.1109/MMS.2016.7803818Search in Google Scholar

[13] Megdad AR, Aldhaheri RW, Sobahi NM. A noninvasive method for measuring the blood glucose level using a narrow band microstrip antenna. ACES J. 2022;37(11):1118–30.10.13052/2022.ACES.J.371102Search in Google Scholar

[14] Deshmukh VV, Ghongade RB. Measurement of dielectric properties of aqueous glucose using planar ring resonator. 2016 Int Conf Microelectronics, Comp Comm (MicroCom); 2016:1–5.10.1109/MicroCom.2016.7522542Search in Google Scholar

[15] Bamgboje D, Christoulakis I, Smanis I, Chavan G, Shah R, Malekzadeh M, et al. Continuous non-invasive glucose monitoring via contact lenses: Current approaches and future perspectives. Biosensors, mdpi. 2021;11:189.10.3390/bios11060189Search in Google Scholar PubMed PubMed Central

[16] Karacolak T, Moreland EC, Topsakal E. Cole-cole model for glucose-dependent dielectric properties of blood plasma for continuous glucose monitoring. Microw Opt Technol Lett. 2013 Mar 23;55(5):1160–4.10.1002/mop.27515Search in Google Scholar

[17] Hays M, Wojcieszak S, Nusrat N, Secondo LE, Topsakal E. Glucose dependent cole-cole models for rat blood plasma from 500 MHz to 40 GHz for millimeter wave glucose detection. Microw Opt Technol Lett. 2020;62(9):2813–20.10.1002/mop.32371Search in Google Scholar

[18] Nella A, Aldhaheri RW, Kamili JB, Sobahi NMKA. A non-invasive method of monitoring glucose in blood using a planar Yagi-Uda antenna and microstrip filter. Elektron Elektrotech. 2023;29(1):33–9.10.5755/j02.eie.31173Search in Google Scholar

[19] Sbrignadello S, Pacini G, Tura A. Determination of glucose levels during dialysis treatment: different sensors and technologies. J Sensors, Hindawi. 2016;2016:1–8.10.1155/2016/8943095Search in Google Scholar

[20] Islam MT, Hoque A, Almutairi AF, Amin N. Left-handed metamaterial-inspired unit cell for S-band glucose sensing application. Sensors. 2019;19(1):169.10.3390/s19010169Search in Google Scholar PubMed PubMed Central

[21] Omer AE, Safavi-Naeini S, Hughson R, Shaker G. Blood glucose level monitoring using an FMCW millimeter-wave radar sensor. Remote Sens. 2020;12(3):385.10.3390/rs12030385Search in Google Scholar

[22] Cebedio MC, Rabioglio LA, Gelosi IE, Ribas RA, Uriz AJ, Moreira JC. Analysis and design of a microwave coplanar sensor for non-invasive blood glucose measurements. IEEE Sens J. 2020 Sep 15;20(18):10572–81.10.1109/JSEN.2020.2993182Search in Google Scholar

[23] Liu W, Yang X, Niu Y, Sun H. Improve planar multiple split-ring sensor for microwave detection applications. Sens Actuators A: Phys. 2019;297(1):111542.10.1016/j.sna.2019.111542Search in Google Scholar

[24] Baghelani M, Abbasi Z, Daneshmand M, Light PE. Non-invasive continuous-time glucose monitoring system using a chipless printable sensor based on split ring microwave resonators. Sci Rep. 2020;10(1):12980.10.1038/s41598-020-69547-1Search in Google Scholar PubMed PubMed Central

[25] Tawfeeq NN, Mahmood SD. Simulation study of microstrip antenna for 2.45 GHz applications based on octagon shaped. Mater Today: Proc. 2021;42(5):2448–56.10.1016/j.matpr.2020.12.560Search in Google Scholar

[26] Su, Rehman, Haider WA, Khalid K, Farooq O, Mahmood U. Design and comparative analysis of 2.45 GHz rectangular microstrip Patch Antenna. 2020 International Conference on Engineering and Emerging Technologies (ICEET). Lahore, Pakistan; 2020. p. 1–5.10.1109/ICEET48479.2020.9048228Search in Google Scholar

[27] Memon AW, de Paula IL, Malengier B, Vasile S, Van Torre P, Van Langenhove L. Breathable textile rectangular ring microstrip patch Antenna at 2.45 GHz for wearable applications. Sensors. 2021;21(5):1635.10.3390/s21051635Search in Google Scholar PubMed PubMed Central

[28] Ram P, Yadav A, Suman MK, Kumar A. Graphene based circular shaped micro strip patch antenna array for 2.45 GHz ISM band application. Wirel Pers Commun. 2021;116:1613–20.10.1007/s11277-020-07751-ySearch in Google Scholar

[29] Balanis CA. Antenna theory analysis and design. 3rd edn. New York: John Wiley & Sons; 2005.Search in Google Scholar

[30] Lumnitzer R, Tanner A, Elsherbeni AZ. Debye coefficients for biological tissues from 100 MHz to 100 GHz. ACES J. 2020;35(11):1424–25.10.47037/2020.ACES.J.351181Search in Google Scholar

[31] Turgul V, Kale I. Simulating the effects of skin thickness and fingerprints to highlight problems with non-invasive RF blood glucose sensing from fingertips. IEEE Sens J. 2017;17(2):7553–60.10.1109/JSEN.2017.2757083Search in Google Scholar

[32] Hasan MN, Tamanna S, Singh P, Nadeem MD, Rudramuni M. Cylindrical dielectric resonator antenna sensor for non-invasive glucose sensing application. 2019 6th International Conference on Signal Processing and Integrated Networks (SPIN): 2019;961–4.10.1109/SPIN.2019.8711633Search in Google Scholar

[33] Islam MT, Rahman MDN, Singh MSJ, Samsuzzaman MD. Detection of salt and sugar contents in water on the basis of dielectric properties using microstrip antenna-based sensor. IEEE Access. 2018;6:4118–26.10.1109/ACCESS.2017.2787689Search in Google Scholar

[34] Choi H, Naylon J, Luzio S, Beutler J, Birchall J, Martin C, et al. Design and in vitro interference test of microwave noninvasive blood glucose monitoring sensor. IEEE Trans Microw Theory Tech. 2015;63(10):3016–25.10.1109/TMTT.2015.2472019Search in Google Scholar PubMed PubMed Central

[35] Tiwari NK, Singh SP, Mondal D, Akhtar MJ. Flexible biomedical RF sensors to quantify the purity of medical grade glycerol and glucose concentrations. Int J Microw Wirel Technol. 2019;12(2):120–30.10.1017/S1759078719001089Search in Google Scholar

[36] Kozak R, Khorsand K, Zarifi T, Golovin K, Zarifi MH. Patch antenna sensor for wireless ice and frost detection. Sci Rep. 2021;11:13707.10.1038/s41598-021-93082-2Search in Google Scholar PubMed PubMed Central

[37] Saha S, Cano-Garcia H, Sotiriou I, Lipscombe O, Gouzouasis I, Koutsoupidou M, et al. A glucose sensing system based on transmission measurements at millimetre waves using micro strip patch antennas. Sci Rep. 2017;7:6855.10.1038/s41598-017-06926-1Search in Google Scholar PubMed PubMed Central

[38] Vrba J, Vrba D. A microwave metamaterial inspired sensor for non-invasive blood glucose monitoring. Radio Eng. 2015;24(4):877–84.10.13164/re.2015.0877Search in Google Scholar

[39] Turgul V, Kale I. Influence of fingerprints and finger positioning on accuracy of RF blood glucose measurement from fingertips. Electron Lett. 2017;53(4):218–20.10.1049/el.2016.4327Search in Google Scholar

Received: 2023-01-05
Revised: 2023-07-03
Accepted: 2023-08-09
Published Online: 2023-09-05

© 2023 the author(s), published by De Gruyter

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Regular Articles
  2. Dynamic properties of the attachment oscillator arising in the nanophysics
  3. Parametric simulation of stagnation point flow of motile microorganism hybrid nanofluid across a circular cylinder with sinusoidal radius
  4. Fractal-fractional advection–diffusion–reaction equations by Ritz approximation approach
  5. Behaviour and onset of low-dimensional chaos with a periodically varying loss in single-mode homogeneously broadened laser
  6. Ammonia gas-sensing behavior of uniform nanostructured PPy film prepared by simple-straightforward in situ chemical vapor oxidation
  7. Analysis of the working mechanism and detection sensitivity of a flash detector
  8. Flat and bent branes with inner structure in two-field mimetic gravity
  9. Heat transfer analysis of the MHD stagnation-point flow of third-grade fluid over a porous sheet with thermal radiation effect: An algorithmic approach
  10. Weighted survival functional entropy and its properties
  11. Bioconvection effect in the Carreau nanofluid with Cattaneo–Christov heat flux using stagnation point flow in the entropy generation: Micromachines level study
  12. Study on the impulse mechanism of optical films formed by laser plasma shock waves
  13. Analysis of sweeping jet and film composite cooling using the decoupled model
  14. Research on the influence of trapezoidal magnetization of bonded magnetic ring on cogging torque
  15. Tripartite entanglement and entanglement transfer in a hybrid cavity magnomechanical system
  16. Compounded Bell-G class of statistical models with applications to COVID-19 and actuarial data
  17. Degradation of Vibrio cholerae from drinking water by the underwater capillary discharge
  18. Multiple Lie symmetry solutions for effects of viscous on magnetohydrodynamic flow and heat transfer in non-Newtonian thin film
  19. Thermal characterization of heat source (sink) on hybridized (Cu–Ag/EG) nanofluid flow via solid stretchable sheet
  20. Optimizing condition monitoring of ball bearings: An integrated approach using decision tree and extreme learning machine for effective decision-making
  21. Study on the inter-porosity transfer rate and producing degree of matrix in fractured-porous gas reservoirs
  22. Interstellar radiation as a Maxwell field: Improved numerical scheme and application to the spectral energy density
  23. Numerical study of hybridized Williamson nanofluid flow with TC4 and Nichrome over an extending surface
  24. Controlling the physical field using the shape function technique
  25. Significance of heat and mass transport in peristaltic flow of Jeffrey material subject to chemical reaction and radiation phenomenon through a tapered channel
  26. Complex dynamics of a sub-quadratic Lorenz-like system
  27. Stability control in a helicoidal spin–orbit-coupled open Bose–Bose mixture
  28. Research on WPD and DBSCAN-L-ISOMAP for circuit fault feature extraction
  29. Simulation for formation process of atomic orbitals by the finite difference time domain method based on the eight-element Dirac equation
  30. A modified power-law model: Properties, estimation, and applications
  31. Bayesian and non-Bayesian estimation of dynamic cumulative residual Tsallis entropy for moment exponential distribution under progressive censored type II
  32. Computational analysis and biomechanical study of Oldroyd-B fluid with homogeneous and heterogeneous reactions through a vertical non-uniform channel
  33. Predictability of machine learning framework in cross-section data
  34. Chaotic characteristics and mixing performance of pseudoplastic fluids in a stirred tank
  35. Isomorphic shut form valuation for quantum field theory and biological population models
  36. Vibration sensitivity minimization of an ultra-stable optical reference cavity based on orthogonal experimental design
  37. Effect of dysprosium on the radiation-shielding features of SiO2–PbO–B2O3 glasses
  38. Asymptotic formulations of anti-plane problems in pre-stressed compressible elastic laminates
  39. A study on soliton, lump solutions to a generalized (3+1)-dimensional Hirota--Satsuma--Ito equation
  40. Tangential electrostatic field at metal surfaces
  41. Bioconvective gyrotactic microorganisms in third-grade nanofluid flow over a Riga surface with stratification: An approach to entropy minimization
  42. Infrared spectroscopy for ageing assessment of insulating oils via dielectric loss factor and interfacial tension
  43. Influence of cationic surfactants on the growth of gypsum crystals
  44. Study on instability mechanism of KCl/PHPA drilling waste fluid
  45. Analytical solutions of the extended Kadomtsev–Petviashvili equation in nonlinear media
  46. A novel compact highly sensitive non-invasive microwave antenna sensor for blood glucose monitoring
  47. Inspection of Couette and pressure-driven Poiseuille entropy-optimized dissipated flow in a suction/injection horizontal channel: Analytical solutions
  48. Conserved vectors and solutions of the two-dimensional potential KP equation
  49. The reciprocal linear effect, a new optical effect of the Sagnac type
  50. Optimal interatomic potentials using modified method of least squares: Optimal form of interatomic potentials
  51. The soliton solutions for stochastic Calogero–Bogoyavlenskii Schiff equation in plasma physics/fluid mechanics
  52. Research on absolute ranging technology of resampling phase comparison method based on FMCW
  53. Analysis of Cu and Zn contents in aluminum alloys by femtosecond laser-ablation spark-induced breakdown spectroscopy
  54. Nonsequential double ionization channels control of CO2 molecules with counter-rotating two-color circularly polarized laser field by laser wavelength
  55. Fractional-order modeling: Analysis of foam drainage and Fisher's equations
  56. Thermo-solutal Marangoni convective Darcy-Forchheimer bio-hybrid nanofluid flow over a permeable disk with activation energy: Analysis of interfacial nanolayer thickness
  57. Investigation on topology-optimized compressor piston by metal additive manufacturing technique: Analytical and numeric computational modeling using finite element analysis in ANSYS
  58. Breast cancer segmentation using a hybrid AttendSeg architecture combined with a gravitational clustering optimization algorithm using mathematical modelling
  59. On the localized and periodic solutions to the time-fractional Klein-Gordan equations: Optimal additive function method and new iterative method
  60. 3D thin-film nanofluid flow with heat transfer on an inclined disc by using HWCM
  61. Numerical study of static pressure on the sonochemistry characteristics of the gas bubble under acoustic excitation
  62. Optimal auxiliary function method for analyzing nonlinear system of coupled Schrödinger–KdV equation with Caputo operator
  63. Analysis of magnetized micropolar fluid subjected to generalized heat-mass transfer theories
  64. Does the Mott problem extend to Geiger counters?
  65. Stability analysis, phase plane analysis, and isolated soliton solution to the LGH equation in mathematical physics
  66. Effects of Joule heating and reaction mechanisms on couple stress fluid flow with peristalsis in the presence of a porous material through an inclined channel
  67. Bayesian and E-Bayesian estimation based on constant-stress partially accelerated life testing for inverted Topp–Leone distribution
  68. Dynamical and physical characteristics of soliton solutions to the (2+1)-dimensional Konopelchenko–Dubrovsky system
  69. Study of fractional variable order COVID-19 environmental transformation model
  70. Sisko nanofluid flow through exponential stretching sheet with swimming of motile gyrotactic microorganisms: An application to nanoengineering
  71. Influence of the regularization scheme in the QCD phase diagram in the PNJL model
  72. Fixed-point theory and numerical analysis of an epidemic model with fractional calculus: Exploring dynamical behavior
  73. Computational analysis of reconstructing current and sag of three-phase overhead line based on the TMR sensor array
  74. Investigation of tripled sine-Gordon equation: Localized modes in multi-stacked long Josephson junctions
  75. High-sensitivity on-chip temperature sensor based on cascaded microring resonators
  76. Pathological study on uncertain numbers and proposed solutions for discrete fuzzy fractional order calculus
  77. Bifurcation, chaotic behavior, and traveling wave solution of stochastic coupled Konno–Oono equation with multiplicative noise in the Stratonovich sense
  78. Thermal radiation and heat generation on three-dimensional Casson fluid motion via porous stretching surface with variable thermal conductivity
  79. Numerical simulation and analysis of Airy's-type equation
  80. A homotopy perturbation method with Elzaki transformation for solving the fractional Biswas–Milovic model
  81. Heat transfer performance of magnetohydrodynamic multiphase nanofluid flow of Cu–Al2O3/H2O over a stretching cylinder
  82. ΛCDM and the principle of equivalence
  83. Axisymmetric stagnation-point flow of non-Newtonian nanomaterial and heat transport over a lubricated surface: Hybrid homotopy analysis method simulations
  84. HAM simulation for bioconvective magnetohydrodynamic flow of Walters-B fluid containing nanoparticles and microorganisms past a stretching sheet with velocity slip and convective conditions
  85. Coupled heat and mass transfer mathematical study for lubricated non-Newtonian nanomaterial conveying oblique stagnation point flow: A comparison of viscous and viscoelastic nanofluid model
  86. Power Topp–Leone exponential negative family of distributions with numerical illustrations to engineering and biological data
  87. Extracting solitary solutions of the nonlinear Kaup–Kupershmidt (KK) equation by analytical method
  88. A case study on the environmental and economic impact of photovoltaic systems in wastewater treatment plants
  89. Application of IoT network for marine wildlife surveillance
  90. Non-similar modeling and numerical simulations of microploar hybrid nanofluid adjacent to isothermal sphere
  91. Joint optimization of two-dimensional warranty period and maintenance strategy considering availability and cost constraints
  92. Numerical investigation of the flow characteristics involving dissipation and slip effects in a convectively nanofluid within a porous medium
  93. Spectral uncertainty analysis of grassland and its camouflage materials based on land-based hyperspectral images
  94. Application of low-altitude wind shear recognition algorithm and laser wind radar in aviation meteorological services
  95. Investigation of different structures of screw extruders on the flow in direct ink writing SiC slurry based on LBM
  96. Harmonic current suppression method of virtual DC motor based on fuzzy sliding mode
  97. Micropolar flow and heat transfer within a permeable channel using the successive linearization method
  98. Different lump k-soliton solutions to (2+1)-dimensional KdV system using Hirota binary Bell polynomials
  99. Investigation of nanomaterials in flow of non-Newtonian liquid toward a stretchable surface
  100. Weak beat frequency extraction method for photon Doppler signal with low signal-to-noise ratio
  101. Electrokinetic energy conversion of nanofluids in porous microtubes with Green’s function
  102. Examining the role of activation energy and convective boundary conditions in nanofluid behavior of Couette-Poiseuille flow
  103. Review Article
  104. Effects of stretching on phase transformation of PVDF and its copolymers: A review
  105. Special Issue on Transport phenomena and thermal analysis in micro/nano-scale structure surfaces - Part IV
  106. Prediction and monitoring model for farmland environmental system using soil sensor and neural network algorithm
  107. Special Issue on Advanced Topics on the Modelling and Assessment of Complicated Physical Phenomena - Part III
  108. Some standard and nonstandard finite difference schemes for a reaction–diffusion–chemotaxis model
  109. Special Issue on Advanced Energy Materials - Part II
  110. Rapid productivity prediction method for frac hits affected wells based on gas reservoir numerical simulation and probability method
  111. Special Issue on Novel Numerical and Analytical Techniques for Fractional Nonlinear Schrodinger Type - Part III
  112. Adomian decomposition method for solution of fourteenth order boundary value problems
  113. New soliton solutions of modified (3+1)-D Wazwaz–Benjamin–Bona–Mahony and (2+1)-D cubic Klein–Gordon equations using first integral method
  114. On traveling wave solutions to Manakov model with variable coefficients
  115. Rational approximation for solving Fredholm integro-differential equations by new algorithm
  116. Special Issue on Predicting pattern alterations in nature - Part I
  117. Modeling the monkeypox infection using the Mittag–Leffler kernel
  118. Spectral analysis of variable-order multi-terms fractional differential equations
  119. Special Issue on Nanomaterial utilization and structural optimization - Part I
  120. Heat treatment and tensile test of 3D-printed parts manufactured at different build orientations
Downloaded on 9.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/phys-2023-0107/html
Scroll to top button