Abstract
This review article investigates the properties and applications of polyetheretherketone (PEEK) in the field of dental implantology. PEEK has emerged as a significant material of interest due to its mechanical strength, biocompatibility, and radiolucency. The article provides a detailed examination of PEEK’s biocompatibility and the various reinforcements that enhance its performance, including PEEK/HA, PEEK/β-TCP–TiO2, and CFR-PEEK. Focusing on dental applications, we discuss PEEK’s use in implant abutments, fixed dental prostheses, implants, and its commercial forms available for dental use. Further, the mechanical behavior of PEEK and its composites is analyzed, including its elastic behavior under various stress conditions and wear resistance. Moreover, the article conducts an integrative systematic review on the stress distribution in dental implants or abutments made from reinforced PEEK composites, assessed through finite element analysis. The aim of this review is to provide insights into the current state of research, the benefits, challenges, and future prospects of PEEK in implantology, and the biomechanical evaluation methods that underpin the development of this promising material.
1 Introduction
Advancements in dentistry involve the evolution of materials. Biocompatibility, aesthetics, low plaque affinity, and characteristics similar to dental structure are among the essential criteria of modern materials used in dentistry [1,2]. Polymers are materials that have allowed significant technological progress in the second half of the twentieth century, although they are now controversial from an environmental standpoint [3–7]. Polyetheretherketone (PEEK) is among the most innovative polymers, described as “high performance.” It was first used in the 1980s in the aerospace and automotive industries [8,9]. Then, due to its remarkable mechanical properties, biocompatibility, and modulus of elasticity similar to human bone, PEEK began to be studied in the field of orthopedic prostheses in the 1990s [6,10–12]. Today, it is expanding in most clinical disciplines of dentistry. Often shaped conventionally, meaning through extrusion, compression, injection, or machining, manufacturers are now turning to 3D printing of PEEK, or additive manufacturing [13–16].
The advancement of dental implant, abutment, and associated materials necessitates adherence to established in vitro and in vivo tests before they are fit for commercial use. Conducting in vivo tests on a substantial number of samples can be lengthy and may pose challenges to animals or patients involved. Moreover, potential inaccuracies in these assessments could lead to misleading outcomes, impacting both the materials and the results of the experiment. Therefore, the finite element method (FEM) emerges as a valuable tool for predicting potential mechanical complications using theoretical models coupled with in vitro tests [17–20]. The finite element analysis (FEA) is frequently used to assess the biomechanical behavior of implantable biomaterials in vitro, whether in orthopedics, traumatology, or oral implantology, a trend is reflected by the increasing number of publications on the subject (Figure 1). This method allows for the evaluation of stress distribution during the load transfer of an implant to the underlying bone structures and typically follows a consistent procedure.

The number of PubMed indexed publications on finite elements used in dental specialties since 2007. Search query: (peek) AND (dental implants) AND (FEA) OR (FEM).
This article delves into the latest advancements in PEEK-based biomaterials for dental applications, highlighting recent improvements in properties and mechanical testing. We will conduct an integrative systematic review focused on analyzing stress distribution across dental implants and abutments made from carbon fiber-reinforced PEEK composites, utilizing FEM studies as a core analytical tool (Figure 1). The investigation is structured around key research questions, aiming to uncover through a systematic approach:
How are PEEK and its composites applied in the dental field, especially as materials for implant abutments and the core of dental implants?
What do mechanical tests, including those for tensile, compression, bending stress, wear resistance, and particularly fatigue tests, reveal about the suitability of PEEK and its composites for dental implant applications?
How does FEA enhance our understanding of stress distribution in dental implants and abutments made from carbon fiber-reinforced PEEK composites?
2 Information sources and search strategy
A literature search on PubMed (via National Library of Medicine) and ScienceDirect (Elsevier BV) was conducted using the following search terms (Table 1):
Term 1: (peek) AND (stress analysis, mechanical) AND (dental)
Term 2: (peek) AND (dental) AND (mechanical phenomena) AND (stress, mechanical) AND (material testing)
Term 3: (peek) AND (dental) AND (finite element) AND (FEA)
Electronic search results from PubMed and ScienceDirect databases
Database | Results For | Total results | ||
---|---|---|---|---|
Term 1 | Term 2 | Term 3 | ||
PubMed | 79 | 24 | 78 | 181 |
PubMed without duplication | 40 | 13 | 55 | 108 |
ScienceDirect | 137 | 69 | 69 | 275 |
ScienceDirect without duplication | 74 | 45 | 54 | 173 |
The electronic searches described previously contributed to initial grouping:
On PubMed: 181 articles
On ScienceDirect: 275 articles
After eliminating duplicates, the titles, and abstracts of the articles resulting from the searches, only 45 publications were retained.
A complete reading of the text and an analysis of the relevance of the articles finally led to a final result grouping 25 articles (Table 2). The publication dates of the articles range between 2010 and 2023, except for 1 article.
Aim, materials, methods, and key findings derived from the chosen research papers
Articles | Type of literature and study design | Objectives | Methods | Types of biomaterials studied and their composition | Results |
---|---|---|---|---|---|
2023 | |||||
Title: Biomaterials and clinical application of dental implants in relation to bone density – A narrative review | Type: Dental/review | To assess how implant materials, designs, and surgical techniques impact bone density, focusing on titanium, zirconia, and PEEK implants, and to determine the optimal choices for varying bone densities | Analysis of literature on the biomechanical and biological effects of dental implant materials and designs on bone density |
|
Titanium is preferred for its reliability, though it presents visibility issues in thin gums and allergy concerns |
Author: Khaohoen et al. | Zirconia is chosen for aesthetics and metal sensitivity but may fracture more easily | ||||
Keywords: biomaterial, bone density, dental implants, … | PEEK is promising for its bone-like properties, yet requires further research | ||||
Reference: [21] | Implant geometry is critical, especially in low-density bone, to enhance stability | ||||
2023 | |||||
Title: Evaluation of stresses on mandible bone and prosthetic parts in fixed prosthesis by utilizing CFR-PEEK, PEKK, and PEEK frameworks | Type: Dental/FEA | The article aims to assess the impact of using polymeric frameworks instead of titanium on stress distribution within the components of fixed prostheses and surrounding bone tissue, considering the influence of spongy bone density on the success of fixed prosthesis in edentulous patients | A FEA assessed stress distribution in an edentulous mandible with fixed prostheses using polymeric materials (CFR-PEEK 60%, CFR-PEEK 30%, PEKK, PEEK) vs titanium, simulating unilateral and bilateral 300 N forces on prostheses in spongy bones of varying densities (normal, low, high) |
|
The study showed that the PEEK framework minimized stress on bone tissues and increased mucosa stress, thus reducing bone resorption risk |
Author: Shash et al. | Frameworks tested (CFR-PEEK 60%, CFR-PEEK 30%, PEKK, PEEK) worked well in normal and high spongy bone densities | ||||
Reference: [22] | |||||
2023 | |||||
Title: Evaluation of stress and strain on mandible caused by changing the bar material in hybrid prosthesis utilizing “All-on-Four” technique | Type: Dental/FEA | The research aimed to explore the viability of using ceramics, polymers, and composites as alternative materials for the bar component in hybrid prostheses, traditionally made of titanium or gold, to address full edentulism and restore the mastication mechanism | A 3D mandible model with a hybrid prosthesis was analyzed using various bar materials. The study measured von Mises stress on the prosthesis and mandible, including maximum and minimum principal stresses and strains on the bone. The simulation tested unilateral and bilateral forces to evaluate their impact on cortical and cancellous bones and mucosa stress |
|
A PEEK bar reduced von Mises stresses on cortical and cancellous bones significantly under unilateral and bilateral forces, while slightly increasing mucosa stress, remaining below pain thresholds. Despite these variations, all materials, including ceramics, polymers, and composites, kept bone stresses within safe limits, suggesting they are viable alternatives to titanium for hybrid prosthesis bars |
Author: Shash et al. | |||||
Reference: [23] | |||||
2022 | |||||
Title: Properties of polyetheretherketone (PEEK) implant abutments: A systematic review | Type: Dental/review | To know the mechanical and functional properties of PEEK abutment and to find out if it is a potential substitute for titanium abutments | The search focused on studies from 2018 to 2020 in English, documented in PROSPERO with ID 274834. Data extraction and quality evaluation adhered to revised CONSORT standards |
|
The existing data indicate that PEEK implant abutments lack the necessary biomechanical properties to serve as a permanent replacement for titanium abutments. Nonetheless, they are regarded as a suitable temporary option, particularly if placed in the anterior region |
Author: Ghazal-Maghras et al. | |||||
Keywords: Abutment, polyetheretherketone, titanium, dynamic fatigue, … | |||||
Reference: [24] | |||||
2022 | |||||
Title: Influence of different combinations of CAD-CAM crown and customized abutment materials on the force absorption capacity in implant supported restorations – In vitro study | Type: Dental/experimental study | To assess the capability of implant-supported restorations to absorb force, employing various CAD-CAM materials in the creation of crowns and customized abutments | Utilized 12 3D finite element models, 6 implant designs, applied axial loads, and analyzed displacements, stress, and strain in bone tissue |
|
Force curve progression varied by material. Zirconia abutments had highest slopes in ZZ, then EZ, VZ, PZ. PEEK abutments showed least slopes in ZP, EP. PEEK abutments increased slope loss in Zirconia, e.max crowns but not in PEEK, Vita Enamic crowns |
Author: Taha et al. | |||||
Keywords: customized abutments, force, force absorption, implant prosthetics, … | |||||
R eference: [25] | |||||
2022 | |||||
Title: In vitro assessment of PEEK and titanium implant abutments: Screw loosening and microleakage evaluations under dynamic mechanical testing | Type: Dental/experimental study | This in vitro study aimed to evaluate the mechanical and functional properties of PEEK implant abutments as an aesthetic, non-metallic alternative to traditional titanium abutments, which have aesthetic limitations and are more challenging to customize in clinical settings | In this study, 24 PEEK and 24 titanium grade 5 implant abutments fixed to MIS Implants type M4 were tested for screw loosening and microleakage in a 2% methylene blue solution, under static and dynamic conditions following ISO 14801:2016 standards. Observations at 8× magnification and statistical analysis (2-factor ANOVA, chi-square) compared the PEEK and titanium groups |
|
The study showed that titanium implant abutments performed better mechanically than PEEK abutments, with titanium experiencing around 10% torque loss compared to PEEK’s, which is up to 50%. Additionally, 91.6% of titanium abutments had no microleakage, whereas all PEEK abutments did under dynamic loading |
Author: Ortega-Martínez et al. | |||||
Reference: [26] | |||||
2022 | |||||
Title: Biomaterials and clinical applications of customized healing abutment – A narrative review | Type: Dental/review | The article aims to assess materials for customized healing abutments used in implant surgery to enhance tissue appearance and reduce treatment and prosthesis fabrication time | A literature review was conducted, searching English-language articles on customized healing abutments in Google Scholar, PubMed/MEDLINE, ScienceDirect, and Scopus databases up to August 2022 |
|
All materials—PEEK, PMMA, zirconia, resin composite, and titanium—showed promising results for customized healing abutments. However, more research is needed to compare their effects on peri-implant tissues conclusively |
Author: Chokaree et al. | |||||
Keywords: PEEK, PMMA, customized healing abutment, … | |||||
Reference: [27] | |||||
2022 | |||||
Title: PEEK biomaterial in long-term provisional implant restorations: A review | Type: Dental/review | The review article examines PEEK as a material for long-term provisional implant restorations in dental implantology, highlighting its growing use in various dental procedures | A comprehensive search across Google Scholar, ScienceDirect, PubMed/MEDLINE, and Scopus identified English-language articles on PEEK as a biomaterial for long-term provisional implant restorations, selecting relevant ones for literature review |
|
PEEK, especially when enhanced with 30–50% carbon fibers, is recognized for its broad applicability in implant dentistry, including dental implants, temporary abutments, and various prostheses, indicating its promise for both temporary and long-term provisional restorations. This suggests potential for wider clinical use |
Author: Suphangul et al. | |||||
Keywords: PEEK, carbon fiber, implant abutment, … | |||||
Reference: [28] | |||||
2021 | |||||
Title: Stress distribution around different abutments on titanium and CFR-PEEK implant with different prosthetic crowns under parafunctional loading: A 3D FEA study | Type: Dental/FEA | Clinical studies linked implant failure to bruxism, while 3D FEA evaluated stress distribution in straight and angled abutments with titanium and CFR-PEEK implants and two crown types under parafunctional loading | Twelve 3D bone block models of the maxillary right premolar with osseointegrated implants were divided into CFR-PEEK and titanium groups, each featuring three abutment angles and two crown types. Stresses from 1,000 N vertical and 500 N oblique loads were analyzed using ANSYS software |
|
Both CFR-PEEK and titanium implants yielded similar bone stress under vertical and oblique loads, with straight abutments outperforming angled ones, and PEEK crowns inducing less stress than (PFM) |
Author: Mourya et al. | |||||
Reference: [29] | |||||
2021 | |||||
Title: Biomechanical performance of Ti-PEEK dental implants in bone: An in-silico analysis | Type: Dental/experimental study and FEA | To compare Ti-PEEK composite dental implants with conventional titanium implants, assessing host bone behavior, especially under conditions of marginal bone loss | Utilized 12 3D FEMs, six implant designs, applied axial loads, and analyzed displacements, stress, and strain in bone tissue |
|
Ti-PEEK implants outperformed conventional dense titanium implants (Implant A) in non-bone loss conditions. However, in bone loss scenarios, both Implant A (made entirely of dense titanium) and Implant E (designed with titanium in the upper half and PEEK in the lower half) showed equal effectiveness |
Author: Ouldyerou et al. | |||||
Keywords: Bone, Composite, Implant, Mechanostat, Ti-PEEK | |||||
Reference: [30] | |||||
2020 | |||||
Title: Comparative evaluation of the wear resistance of two different implant abutment materials after cyclic loading | Type: Dental/experimental study | The paper aims to compare the wear resistance of titanium and PEEK abutment materials when paired with titanium implants after cyclic loading, assessing significant differences in wear resistance under repeated mechanical stress | The study involved 20 titanium implants in resin blocks, divided into groups with titanium and PEEK abutments, each undergoing 550,000 cycles of cyclic loading. Surface roughness was measured by profilometry, abutment surfaces imaged by SEM, and elemental analysis conducted using EDS, to evaluate changes at the implant-abutment interface before and after loading |
|
The study compared wear resistance of titanium and PEEK abutment materials connected to titanium implants after cyclic loading, finding no significant difference in wear patterns or surface roughness changes between the two, as confirmed by SEM and EDS analyses. Both materials are comparably effective for dental implant abutments, with PEEK offering an aesthetic alternative |
Author: Ragupathi et al. | |||||
Keywords: Abutment, Polyether ether Ketone, implant-abutment interface, … | |||||
Reference: [31] | |||||
2019 | |||||
Title: Effect of different biocompatible implant materials on the mechanical stability of dental implants under excessive oblique load | Type: Dental/FEA | The study evaluates the impact of replacing traditional titanium with carbon reinforced polyether ether ketone (CRF-PEEK) composites on stress distribution in peri-implant bone under excessive oblique loading, examining if CRF-PEEK’s bone-like mechanical properties can enhance stress distribution and decrease implant failure risks | The methodology includes constructing 3D models of a dental implant in the first mandibular molar from CT scans and creating five models using titanium, CRF-PEEK, and lithium disilicate. It employs 3D FEA to assess stress distribution at the implant-bone interface under excessive oblique loads and examines physical interactions, including friction effects between contacting surfaces |
|
The study compared wear resistance between titanium and PEEK abutment materials connected to titanium implants post-cyclic loading, revealing only slight and statistically insignificant differences in surface roughness and wear patterns as confirmed by SEM and EDS analyses. Both materials proved comparably effective for dental implant abutments, with PEEK also offering aesthetic benefits |
Author: Bataineh et al. | |||||
Keywords: CFR-PEEK, biocompatible, biomaterials, deformation, dental implant, … | |||||
Reference: [32] | |||||
2018 | |||||
Title: 3D FEA study on implant threading role on selection of implant and crown materials | Type: Dental/FEA | Examination of the impact of thread design in dental implants and the choice of material on mandibular bone response, considering two distinct crown materials: Translucent zirconia and PFM | Two single-piece dental implant designs with a proxy crown were modeled on simplified bone structures in FEA. CAD/CAM software crafted the components, which were assembled and analyzed in ANSYS, applying 100 N compressive and 50 N oblique loads |
|
In 24 cases, micro-threading slashed the dental implant’s peak Von Mises stress by 50–70% compared to conventional threading. A 50 N oblique force generated 4–5 times the stress on the implant as a 100 N vertical load. Crown material swaps had minimal impact on cortical bone stress, while titanium implants cut stress by 50–100% vs reinforced PEKK or PEEK |
Author: Wazeh et al. | |||||
Keywords: Dental implant, FEA, PEEK, PEKK, Titanium | |||||
Reference: [33] | |||||
2018 | |||||
Title: Computational modelling; damage mechanics; laminated orthopaedic devices; mechanical testing; medical grade carbon fiber reinforced PEEK | Type: Dental/FEA | The study focuses on the mechanical behavior and failure mechanisms of PEEK-OPTIMA™ Ultra-Reinforced, a unidirectional CFR-PEEK, to develop laminated orthopedic devices and predictive computational models | A series of multi-axial experimental tests were conducted, including tension, compression, in-plane shear, and fracture toughness tests, along with the development of a computational failure model that incorporates observed damage mechanisms | PEEK-OPTIMA™ Ultra-Reinforced laminates | The study identifies three key damage mechanisms: inter-laminar delamination, intra-laminar cracking, and anisotropic plasticity. The computational model effectively predicts these complex failure mechanisms, supporting the design of safe, fiber-reinforced laminated orthopaedic devices |
Author: Gallagher et al. | |||||
Keywords: computational modelling, damage mechanics, … | |||||
Reference: [34] | |||||
2018 | |||||
Title: Effect of different restorative crown and customized abutment materials on stress distribution in single implants and peripheral bone: A three-dimensional FEA study | Type: Dental/FEA | To evaluate the stress distribution effects of resin-matrix ceramic and PEEK customized abutments on dental implants and adjacent bone using FEA | The study used 3D modeling with STL data to create implant systems and abutments, and developed six models combining different restoration materials (TZI, IPS, VTE) and abutments (PEEK, zirconia). These models were subjected to vertical (200 N) and oblique (100 N) loads, analyzing stress distribution via von Mises and principal stress methods |
|
Oblique loading induced elevated stress in implants, crowns, and cortical bone, with VTE crowns showing lower stress. Zirconia abutments faced greater stress than PEEK ones. Stress patterns in implants and surrounding bone remained consistent across all models |
Author: Kaleli et al. | |||||
Reference: [35] | |||||
2016 | |||||
Title: Comparison between PEEK and Ti6Al4V concerning micro-scale abrasion wear on dental applications | Type: Dental/experimental study | The study compared the abrasive wear resistance of PEEK and titanium alloy (Ti6Al4V) under simulated three-body abrasion conditions to mimic wear from food and toothpaste during mastication and tooth brushing | The study conducted micro-scale abrasion tests on PEEK and Ti6Al4V cylinders, using varied loads and hydrated silica suspensions, measuring wear volumes and analyzing wear scars with SEM to determine wear mechanisms |
|
The study revealed that under three-body abrasion tests with hydrated silica suspensions, PEEK showed higher volume loss and lower wear resistance than Ti6Al4V, which consistently outperformed PEEK across different abrasive contents and loads |
Author: Sampaio et al. | |||||
Keywords: Bio-tribology, micro-scale abrasion, PEEK, Ti6Al4V, wear | |||||
Reference: [36] | |||||
2016 | |||||
Title: Effects of PEEK veneer thickness on the reciprocating friction and wear behavior of PEEK/Ti6Al4V structures in artificial saliva | Type: Dental/FEA | The study examines how varying thicknesses of PEEK veneers impact friction, wear rate, and contact stress on Ti6Al4V substrates, crucial for the durability and effectiveness of biomedical materials in oral applications | PEEK veneers from 0.1 to 2 mm thick were synthesized on Ti6Al4V substrates, tested for friction and wear against alumina in artificial saliva at 37°C, and analyzed via simulations to study the effect of thickness on contact stress |
|
Thinner PEEK veneers show higher coefficients of friction and wear rates due to increased contact stress, underscoring the importance of considering veneer thickness in designing biomedical devices, particularly for oral applications requiring optimal friction and wear resistance for long-term success |
Author: Sampaio et al. | |||||
Keywords: Biotribology, Ti6Al4V, PEEK, sliding wear, thickness, FEM model | |||||
Reference: [37] | |||||
2016 | |||||
Title: Pressure behavior of different PEEK materials for dental implants | Type: Dental/experimental study | Evaluate the mechanical properties of different PEEK composites under static pressure | Nine PEEK composites (11 applications biomedical, 2 applications industrial) tested under static pressure |
|
Mechanical properties: Elastic modulus, yield limit, resistance to compression |
Author: Schwitalla et al. | |||||
Reference: [38] | |||||
2015 | |||||
Title: FEA of the biomechanical effects of PEEK dental implants on the peri-implant bone | Type: Dental/FEA | Show the biomechanical behavior differences of three implants (two PEEK vs one titanium) |
|
Mechanical properties: Modulus of elasticity, coefficient of Poisson | |
Author: Schwitalla et al. | Stresses at the bone/implant: Ultimate tensile stress, von Mises, deformation and pressure on the peri-implant bone | ||||
Reference: [39] | |||||
2015 | |||||
Title: Flexural behavior of PEEK materials for dental application | Type: Dental/experimental study | Evaluate the mechanical properties of different PEEK composites by the three-point bending test | In vitro study: 150 specimens in the form of a bar made from 11 PEEK composites (9 applications biomedical, 2 applications industrial) tested by three-point bending test |
|
Mechanical properties: Bending modulus, resistance to flexion |
Author: Schwitalla et al. | |||||
Reference: [40] | |||||
2015 | |||||
Title: Applications of PEEK in oral implantology and prosthodontics | Type: Dental/review | Synthesize the results of research conducted on PEEK in dental applications and provide perspectives on the use of PEEK and its potential clinical applications | Electronic research of English publications over the last 15 years on PubMed using a combination of keywords “polyetheretherketone” and “dental” and “dentistry” |
|
Mechanical properties: Elastic modulus, resistance to traction |
Author: Najeeb et al. | |||||
Reference: [41] | |||||
2012 | |||||
Title: Stress shielding and fatigue limits of poly-ether-ether-ketone dental implants | Type: Dental/experimental study | To evaluate fatigue limits of PEEK and its effects on stress shielding in comparison to traditional titanium dental implants | The study used compressive loading tests on GFR-PEEK, CFR-PEEK, and titanium rods, with GFR-PEEK tested as per ISO 14801. FEA involved 3D models of dental implants and bone. PEEK layer coating was applied to implants for further testing and analysis |
|
GFR-PEEK’s fatigue limit reached 310 N, exceeding static strength. PEEK implants displayed elevated SED near bone. Both GFR and CFR-PEEK implants are suitable for anterior tooth replacement, potentially reducing stress shielding |
Author: Lee et al. | |||||
Reference: [42] | |||||
2010 | |||||
Title: A “metal free” material in implantology: the Biopik® | Type: Dental/Review | Presentation of the mechanical and biological properties of “Biopik®“ with a view of its application in implantology | Presentation of Biopik® | Biopik®: Tricalcium phosphate (TCP) 10% + titanium oxide (TiO2) 10% + PEEK matrix | Mechanical properties: Young’s modulus |
Author: Cougoulic et al. | In vitro study of the interface human osteogenic cells/material | ||||
Reference: [43] | In vivo study, animal experimentation on osseointegration at 4 weeks | ||||
2010 | |||||
Title: Evaluation of the stress distribution in CFR-PEEK dental implants by the three-dimensional FEM | Type: Dental/FEA | The stress dispersion around peri-implant bone was evaluated using FEM across four configurations: titanium abutment on titanium implant, CFR-PEEK abutment on titanium implant, titanium abutment on CFR-PEEK implant, and CFR-PEEK abutment on CFR-PEEK implant | Cone beam computed tomography was used to create 3D jaw models analyzed via Ansys software for FEM simulations, assessing stress distribution on various abutment-implant configurations in cortical and medullar bone, employing nonlinear simulations with rigid supports to evaluate mechanical response under varied loads |
|
The titanium implant distributes the stresses in a more homogenous manner in relation to the CFR-PEEK implant due to the smaller deformation of this material. The CFR-PEEK implant did not present any advantages in relation to the titanium implant regarding stress distribution to the peri-implant bone |
Author: Sarot et al. | |||||
Reference: [44] | |||||
2006 | |||||
Title: The long-term mechanical integrity of non-reinforced PEEK-OPTIMA polymer for demanding spinal applications: Experimental and finite-element analysis | Type: Dntal/experimental study and FEA | Compare the biomechanical performances of PEEK and titanium under constraints and analyze their diffusions | In vitro study: creep and quasi-static compression tests at one and three times in dry conditions and in saline solution (NaCl) at 37°C on cylindrical specimens of PEEK and titanium |
|
Mechanical properties: Modulus of elasticity, creep |
Author: Ferguson et al. | Stresses at the bone/implant interface: Analysis of the stress distribution after compression, bending, lateral bending, and axial rotation constraints | ||||
Keywords: PEEK, material properties, creep, cages, fusion | Finite element study: 3D modeling of vertebrae and simulation of intervertebral cage implant placement based on PEEK and titanium to compare stress diffusion under different constraints | ||||
Reference: [45] |
A selection tree of articles (as depicted in Figure 2) was therefore made from the 25 articles, including 6 literature reviews (21–24–27–28–41–43), 12 studies for FEA (22–23–29–30–32–33–34–35–37–39–44–45), and 9 in vitro experimental studies (25–26–31–36–38–40–42).

Flow diagram of the search strategy.
3 PEEK
The PEEK is a thermally stable semi-crystalline thermoplastic material. Introduced in 1977 by ICI (Imperial Chemical Industries), this material entered the market between 1978 and 1979. It gained traction in the 1980s across orthopedics, traumatology, and spinal surgery, particularly in the creation of interbody cages [46–49]. By the late 1990s, it was recognized as a premier non-metallic counterpart to titanium in these specialties [10]. The literature on implantology began highlighting this material around 1995 [50]. In April 1998, the debut PEEK implant product, branded as PEEK-OPTIMATM and compliant with both European Community and Food and Drug Administration standards, was launched [10,51]. PEEK boasts superior mechanical, physical, and chemical attributes. Notably, it has a hardness rating between 99 and 126 on the Rockwell scale [42,52] and an elasticity modulus (4 GPa) that is more akin to cortical bone (20 GPa) than titanium (110 GPa).
3.1 Biocompatibility
From a biomedical perspective, recent studies assert that the results of various experiments conducted on systemic toxicity, inflammatory response, and genotoxicity tests of PEEK and PEEK composite biomaterials support its biocompatibility [53,54,55]. This blend of biocompatibility and mechanical robustness has solidified PEEK’s standing as a preferred material in medical device manufacturing, including but not limited to, spinal fusion cages and dental implants [10]. However, Najeeb et al. [41], noted that despite the bio-inertia of unmodified PEEK, this does not present any conclusive evidence of osteo-conductive or osteo-inductive effects in vivo and in vitro unlike conventional implant materials (zirconium oxide and titanium). Thus, in its unmodified form, the long-term survival rate of PEEK implants is questionable.
To overcome this gap in osteo-integration and to enhance the biological and physical properties of PEEK, numerous nano-modifications have been explored (Table 3).
Nano-modification of PEEK to increase its bioactivity
Nano structured surfaces | Bioactive nano-composites |
---|---|
Spin-coating with nano-Hap [66,67] | TiO2/PEEK [68–70] |
Plasma-gas treatment (O2/Ar, NH4) [71,72] | |
Plasma electron beam deposition (Ti, TiO2) [54,73] | HAF/PEEK [57,70] |
Plasma ion immersion implantation (TiO2) [74,75] |
Note: TiO2: Titanium dioxide nanoparticles and HAF: Hydroxy-fluorapatite nanoparticles.
3.2 PEEK reinforcement
3.2.1 PEEK/HA
Early investigations to improve PEEK bioactivity focused on the incorporation of hydroxyapatite (HA) particles, a calcium phosphate mineral celebrated for its capacity to facilitate bone attachment and regeneration [56]. By merging HA with PEEK, the aim is to combine PEEK’s structural advantages with the bioactive capabilities of HA, thus amplifying the implant’s potential for successful and durable integration into bone structures [57]. Research around this hybrid material has yielded positive indications, with PEEK-HA dental prosthetics showing improved bone response and superior bonding than their PEEK-only counterparts [58,59,60]. While HA incorporation promotes bioactivity, it might compromise the overall mechanical strength of PEEK, a trade-off that requires careful balance [10,41,61]. Further research is pivotal in establishing the optimal ratio of HA in PEEK to maximize both bioactivity and mechanical integrity.
3.2.2 PEEK/beta-tricalcium phosphate (β-TCP)–TiO2
Studies on the incorporation of beta-tricalcium phosphate into a PEEK matrix has shown that cell proliferation was gradually inhibited when these particles were present. Moreover, the osteo-conduction on the surface of these implants, resulting from the inherent properties of β-TCP, promoted osteo-integration [10].
Cougoulic et al. [43] introduced an implant made from this material, marketed under the name SMARTPIK®, which has received EC marking after meeting the ISO 10993 standard: “Biological Evaluation of Medical Devices.”
In addition to the PEEK matrix, which provides the primary mechanical properties of the implant, and the β-TCP (17%) enhancing its osteoconductive potential, the addition of titanium dioxide (8% TiO2) provides radiopacity. A specific surface treatment optimizes its biological properties[10].
3.2.3 CFR-PEEK
From a biological perspective, CFR-PEEK composites are considered biologically stable, and it has been demonstrated that carbon fibers possess osteo-inductive activity promoting tissue integration and show no toxicity [10,62]. To achieve a strong bond to the matrix, carbon fibers undergo oxidation, allowing them to form a covalent bond robust enough to resist dislocation when a force is applied to the material [63].
The mechanical properties of CFR-PEEK composites can be controlled by the content, dimensions, and orientation of the fibers [64]. The volume percentage of carbon fiber can vary between 30 and 60%. These fibers can be either continuous or short, and they can be unidirectional (oriented parallelly) or multidirectional [38,41,42,64,65]. All authors observe that adding these fibers to PEEK results increased creep, hardness, resistance to compression and fatigue, and also considerably reduced the stress-shielding experienced by the bone since the material’s modulus of elasticity can adjust to that of the bone based on the percentage of fiber incorporated [38,41,42,64,65]. Then, the development of new implants requires adherence to established in vitro and in vivo tests before proceeding to manufacturing.
Due to their adjustable mechanical properties, researchers have suggested through their studies that CFR-PEEK composites are more suitable for use in both orthopedics and oral implantology. Indeed, they are more likely to counteract forces causing stress peaks, ensuring a more uniform distribution of load to the peri-implant bone than the gold-standard implants [38,41,61].
4 PEEK dental applications
As we delve deeper into the applications of PEEK in dentistry, we will explore the use of PEEK in various capacities: as implant abutments, a material for removable partial dentures, a foundation for fixed dental prostheses, and even as the core material for dental implants.
4.1 PEEK implant abutments
In the field of implant dentistry, an abutment serves as the bridge between the prosthetic restoration and the implant itself (Figure 3). Abutments are crafted from a variety of materials, including titanium, gold, zirconium, and ceramics [76]. Recently, there has been an increased utilization of PEEK abutments. Notably, in instances of implant screw damage, PEEK screws offer ease of removal. Studies have shown that PEEK abutments can endure intraoral chewing forces akin to those experienced by titanium abutments. The soft tissue response to PEEK is commendable, often leading to optimal gingival tissue healing [77].

Schematic representation of the structure of a dental implant.
The semi-crystalline nature of PEEK contributes to its reduced brittleness, leading to deformation rather than breakage. Research has shown that when using PEEK for abutments, while the prosthesis remains unharmed, the abutment might deform. Nevertheless, the prosthesis’ functionality can be restored by replacing the deformed abutment [78]. In a study by Koutouzis et al., a randomized controlled clinical trial found negligible differences in bone loss and tissue inflammation between PEEK and titanium abutments (Figure 4). Furthermore, the composition of oral bacteria observed was consistent with that seen with titanium, zirconia, or PMMA abutments [79].
![Figure 4
Occlusal view of implants with PEEK abutments (a) following installation and (b) at 3 months. And, occlusal view of implants with titanium abutments (c) following installation and (d) at 3 months [79].](/document/doi/10.1515/rams-2024-0031/asset/graphic/j_rams-2024-0031_fig_004.jpg)
Occlusal view of implants with PEEK abutments (a) following installation and (b) at 3 months. And, occlusal view of implants with titanium abutments (c) following installation and (d) at 3 months [79].
The distribution of occlusal forces and the manner in which stress is distributed during the implant healing period significantly impact the enduring outcomes of implant procedures. In a study by Taha et al. [25], they examined the interplay between various crown-abutment combinations and their impact on the force absorption properties of implant-supported restorations. Their findings indicated that pairing resin-based ceramic crowns with PEEK abutments did not amplify force absorption capabilities. This could be attributed to the notably reduced elastic modulus of PEEK custom abutments, leading to diminished stress within the abutment, but increased stress on the associated crown [35].
Ceramic-reinforced PEEK (RPEEK) is another variant used in implant abutments noted for its good biomechanical properties and biocompatibility, with Al-Rabab’ah et al. [80] reporting stable bone and soft tissue around implants over 2 years.
Otherwise, Al-Zordk et al. discussed the critical aspect of fixation techniques in implant-supported restorations, highlighting the advantages of cement-screws for their superior passive fit, aesthetic appeal, and force distribution [81]. Zarone et al. further asserted that cement-screw systems have a greater load-bearing capability compared to screw-retained systems [82]. Freitas et al. [83] observed that restorations retained by cement and featuring internal connections display enhanced fracture resistance, leading to their preference in Atsu et al.‘s work [84]. Yet, they caution against the potential for peri-implant diseases due to residual cement in the submucosal area, with prevalence rates ranging from 1.9 to 75% in such restorations [81]. The discussion also touches upon physiological bite forces, noting the range of occlusal forces across different teeth types. Atsu et al. reported that titanium abutments exhibit superior fracture toughness (943.67 N) compared to Zr and RPEEK (770.1 N), though the latter two materials show no significant difference in this aspect (p = 0.001). Ortega et al. presented findings on fracture toughness, with titanium abutments at 468.5 N and PEEK at 200.4 N. They noted that PEEK abutments withstand 1.2 million cycles at 140 N without failure, simulating 5 years of function, but fail under increased loads of 160 N after approximately 89,338 cycles, equating to 4–5 months of use [26]. This behavior indicates PEEK’s role as a “sacrificial material,” absorbing deformation to prevent damage to the implant or screw, in contrast to titanium where deformation affects the internal connection, potentially endangering the implant’s longevity [26]. In another study by Mourya et al. [29], the stress distribution around abutments, subjected to both vertical and oblique forces, was assessed for both titanium and carbon-fiber-reinforced PEEK (CF/PEEK) implants. The findings propose that in patients with molar teeth, using straight abutments in conjunction with PEEK crowns could mitigate intraosseous stress levels and potentially avert implant complications.
The available evidence suggests that PEEK abutments may not meet the biomechanical criteria necessary to supplant traditional titanium abutments. Among available options, zirconium is recognized for its superior biocompatibility as an abutment material [85]. Nevertheless, PEEK exhibits particular merits in applications like customized healing abutment [27], provisional restorations [24,28] for patients lacking functional deficiencies, and for shaping emergency contours during surgical interventions, especially relevant for implants in individuals subject to reduced stress loads.
Figure 5 shows the various PEEK abutments. PEEK healing abutment (Figure 5a) is used as an alternative for classic titanium healing abutment [86], and PEEK with titanium base temporary abutment (Figure 5b) is used for long-term interim restoration, especially in areas with aesthetic consideration [87].

PEEK abutments: (a) PEEK healing abutment (GM customizable healing abutment, Neodent, Curitiba, Brazil), (b) PEEK with titanium base temporary abutment (GM PRO PEEK abutment, Neodent, Curitiba, Brazil).
4.2 PEEK as fixed dental prosthesis
For fixed dental prosthetics, PEEK offers a unique solution with its metal-free crowns and bridges, known for their outstanding biocompatibility and robust mechanical attributes (Figure 6). When assessing PEEK against traditional ceramic and metal materials, it becomes evident that PEEK’s dental bridge infrastructure remains resilient, even when subjected to in vitro aging processes. For implant-linked prosthetic applications, PEEK crowns have showcased effective outcomes [88]. In terms of biocompatibility evaluations, PEEK consistently outperforms many metal-based ceramics. Nevertheless, some experts propose the integration of veneers over PEEK to achieve optimal precision [89]. Given its lightweight nature, PEEK emerges as a promising contender to replace traditional chrome-cobalt prostheses [90–92].
![Figure 6
PEEK resin-bonded fixed dental prosthesis: (a) occlusal view and (b) after fixing the teeth [94].](/document/doi/10.1515/rams-2024-0031/asset/graphic/j_rams-2024-0031_fig_006.jpg)
PEEK resin-bonded fixed dental prosthesis: (a) occlusal view and (b) after fixing the teeth [94].
PEEK restorations demonstrate satisfactory resilience against fractures, enduring anterior masticatory forces of around 300 N and forces of 600 N in posterior areas. Extended chewing simulations, replicating up to 5 years of intraoral utilization, revealed no structural damage or debonding incidents in these restorations [77].
4.3 PEEK implants
As mentioned previously, PEEK is emerging as a potential material in implant dentistry, offering a solution to both full and partial tooth loss.
Traditionally, titanium has been the go-to material for dental implants due to its osteoconductive properties. However, PEEK, having elasticity similar to human bone, is now being considered as an alternative to titanium [94]. The stability of an implant can be influenced by several elements, including the occlusal load, bone structure, and potential for bone loss [95]. Khaohoen et al. [21] highlight that PEEK is emerging as a viable option in the realm of dental implantology, countering some of the drawbacks associated with titanium such as aesthetic limitations, allergic reactions, and stress-shielding effects. Despite PEEK’s outstanding biocompatibility and absence of cytotoxic effects, its osseointegration capabilities lag behind those of titanium, marked by a reduced bone-implant contact (BIC) area, diminished osteoblast activity, and lower osteoconductivity. However, enhancing PEEK’s surface hydrophilicity and roughness through nanoscale modifications have been identified as a promising strategy to mitigate these limitations.
Ouldyerou et al. indicated that while standard titanium implants tend to concentrate stress in specific bone areas, those made of a titanium-PEEK combination distribute stress more evenly, potentially reducing bone degradation [30]. Interestingly, studies involving animals have shown that titanium-coated PEEK may support bone formation better than its uncoated counterpart [58]. Pure PEEK, given its hydrophobic nature, might deter initial cell attachment. Yet, some investigations have noted enhanced mechanical bonding between bone and implants when using porous or titanium-coated PEEK, leading to better cellular interaction and integration [96]. Additionally, many studies have reported on methods to increase the bioactivity of PEEK, and various authors have highlighted the need to incorporate bioactive inorganic ceramic particles, such as HA, HAF, β-TCP, and TiO2 to enhance the osteointegration of PEEK implants [10,67,73,97]. Therefore, while PEEK holds promise, comprehensive studies are crucial before its widespread adoption in dental implantology.
4.4 Commercial forms for dental use
Two leading commercial PEEK brands are widely recognized in the dental sector. PEEK-OPTIMA® is the preferred choice within the US, supplied by Invibio Biomaterial Solutions since 1999, and is primarily used for dental applications such as provisional prosthetic abutments and healing screws [98]. In contrast, Europe mainly utilizes BioHPP™ (Figure 7), produced by Bredent GmbH, which is specifically designed for dental applications. It is a modified PEEK material that incorporates ceramic fillers to enhance polishability and is suitable for various dental prostheses and frameworks through injection molding and CAD-CAM processing [99].
![Figure 7
PEEK abutment made by BioHPP [99].](/document/doi/10.1515/rams-2024-0031/asset/graphic/j_rams-2024-0031_fig_007.jpg)
PEEK abutment made by BioHPP [99].
5 Mechanical behavior of PEEK and its composites
In this part, we will delve into an in-depth examination of the mechanical behavior properties of PEEK and its associated composites. This analysis will provide a comprehensive understanding of the material’s strengths as well as their ability to withstand force stresses without fracturing or undergoing deformation.
5.1 Elastic behavior under tensile, compression, bending stress, and wear resistance
The tensile, compression, and flexural (bending) tests conducted on PEEK have led to determining its ability to deform after being subjected to a constantly increasing loading process.
The application of tensile forces results in a progressive elongation up to the material’s breaking point and allows for defining many of its mechanical properties. These elastic properties are derived from the interpretation of a stress–strain curve defining both an elastic and a plastic domain. They are directly dependent on the crystallinity of PEEK, reflecting its thermal treatment history [100,101].
5.1.1 Elastic modulus, flexural and tensile strengths of PEEK
In medical and dental fields, the modulus of elasticity stands as a pivotal factor influencing the long-term durability and biocompatibility of a material [102]. Studies have highlighted PEEK’s modulus of elasticity, noting that it does not match that of the human cortical bone but is closer to it than the traditional metals, which positions it as a potential candidate for medical and dental implant applications [103]. Previous studies have shown that rigid implants in Titanium or in zirconia had an elasticity modulus (110 and 210 GPa, respectively) that is 5–14 times greater than cortical bone (17 GPa). This significant difference can lead to bone resorption and loss of osseointegration due to a phenomenon known as “Stress Shielding.” The difference in mechanical behavior between these biomaterials and bone results in poor distribution and uneven diffusion of forces arising from mastication stresses [42,39]. On the other hand, authors have reported values of the modulus of elasticity of the unmodified PEEK (Table 4) ranging between 3 and 4 GPa [104] and, as mentioned previously, this modulus can be increased significantly by reinforcing PEEK with various composite and ceramic materials (hydroxyapatite, glass fiber, carbon fiber, titanium oxide, tricalcium phosphate) to achieve results even more akin to cortical bone [42,104,105,106]. This promotes better stress distribution, preventing high stress peaks at the osseous/implant interface that lead to implant failure [40,42,44,107]. The elastic modulus of CFR-PEEK can reach up to 18 GPa [104,106] while GFR-PEEK can achieve around 12 GPa [42,105]. A comprehensive list of the mechanical properties related to these materials, especially the blended PEEK types, can be found in Table 4.
Elastic modulus and the tensile strength of various materials
Material | Elastic modulus (GPa) | Tensile strength (MPa) | Ref. |
---|---|---|---|
Titanium | 110 | 954–976 | [108] |
PMMA | 3–5 | 48–76 | [109] |
PEEK | 3–4 | 80 | [104] |
CFR-PEEK | 18 | 120 | [104] |
GFR-PEEK | 12 | [110,111] | |
Cortical bone | 14 | 104–121 | [112] |
Enamel | 40–83 | 47.5 | [113] |
Dentin | 15–30 | 104 | [114,115] |
Ti6Al4V | 110–130 | 976 | [111,116] |
Schwitalla et al. [38,39] conducted studies on 11 different PEEK materials, assessing them through static compression and three-point flexural tests. Materials included grades of PEEK with various reinforcements, such as titanium dioxide, barium sulfate, carbon fibers, and glass fiber. During testing, specimens were subjected to a load from a wedge-shaped penetrator. The studies highlighted the suitability of these PEEK materials for dental implants, given their capacity to withstand high mastication pressures and compliance with the standard (EN ISO 10477).
Furthermore, Ferguson et al. [45] compared the modulus of elasticity of PEEK OPTIMA® samples through compression tests conducted both in a dry environment at room temperature (22°C) and under saline conditions (NaCl) at 37°C to simulate a natural physiological environment. The results suggested that the modulus of elasticity was 1.8% lower in the aqueous environment, indicating that temperature and humidity had a statistically significant influence on it, even though the difference was relatively small.
Also, Schwitalla et al.’s study [40] employing a three-point flexural approach revealed that PEEK’s strength surpasses the standard benchmark set for plastic materials at 65 MPa. When considering its applications in orthodontics, PEEK sets itself apart by displaying superior flexural strength and minimized creep when compared to other plastics like polyethersulfone and polyvinylidene fluoride [117]. Furthermore, with a tensile strength rating of 80 MPa [104], PEEK’s mechanical properties are close to enamel and dentin [113,114,118], making it an ideal candidate for constructing prosthodontic framework.
5.1.2 Wear resistance
Recent advancements have explored novel PEEK-based composites as potential replacements for traditional metallic materials in biomedical implants and prosthesis. However, the performance of these materials under aggressive oral conditions, particularly in the presence of abrasive particles from toothpaste, remains a critical area of research. A significant study by Sampaio et al. [38] focused on evaluating the abrasive wear resistance of PEEK and Ti6Al4V, which are prevalently utilized in the fabrication of dental prosthetics and implants. The aim of this research was to closely mimic the oral environment, characterized by frequent exposure to abrasive compounds found in both food and dental care products. Such exposure is known to induce wear on the contact surfaces between the prosthetic structures and implants, posing challenges to the longevity and reliability of these materials. The researchers prepared surfaces of both materials and subjected them to micro-scale abrasion tests under varying loads and with different concentrations of hydrated silica, mimicking the abrasive action of chewing and tooth brushing. Their findings revealed that PEEK exhibits a higher volume loss compared to Ti6Al4V under these abrasive conditions, indicating that Ti6Al4V possesses superior wear resistance. The increase in wear for both materials was proportional to the abrasive content and load applied. This suggests that while both materials are susceptible to wear from common abrasive particles in the oral cavity, Ti6Al4V may offer better durability against such challenges.
In a parallel inquiry, Ragupathi et al. [31] conducted a detailed study on the wear characteristics of PEEK in comparison to titanium, specifically focusing on dental abutments subjected to cyclic loading. This method was intended to simulate the wear that occurs over a year of mastication.
Despite the hypothesis suggesting greater wear in PEEK abutments due to their lower elastic modulus, findings revealed minimal difference in wear rates between the two materials. Advanced analytical techniques, including scanning electron microscopy (SEM), were utilized to meticulously evaluate wear patterns, demonstrating that both PEEK and titanium exhibited similar resilience to cyclic stress. This was quantitatively supported by surface roughness measurements before and after loading, indicating no statistically significant wear discrepancy between PEEK and titanium abutments. The study concluded that PEEK’s wear resistance closely matches that of titanium, making it a viable alternative for implant abutments in dental applications. These results highlight PEEK’s potential in dental implantology, although further research is encouraged to validate these findings over longer periods and in varied clinical conditions.
5.2 Mechanical testing
Mechanical testing in dental implants plays a vital role in assessing their robustness and functionality under the conditions of oral cavity. Such tests offer insights into the longevity and stability of these implants when subjected to continuous stress and environmental factors. In the following discussion, we will delve deeper into specific mechanical tests, namely, fatigue tests and wear resistance, to better understand their significance in evaluating dental implant performance.
5.2.1 Fatigue test for PEEK used in implant applications
Implant-supported prosthesis undergo stress from chewing in a wet environment and face sudden temperature changes between 0 and 65°C. Consequently, the materials chosen should meet specific standards concerning their robustness and rigidity. They should also resist against deformation and fatigue fracture [119].
During the chewing process, loads exerted can vary, sometimes remaining below the fracture load. However, even these variations can introduce weaknesses in the structure, resulting in the appearance of fissures and cracks. This is often termed as mechanical fatigue. Over time, these minor imperfections can evolve and spread, leading to a fatigue fracture [120,121]. The fatigue limit or endurance limit is the stress level below which an infinite number of loading cycles can be applied to a material without causing fatigue failure [122].
Lee et al. [42] assessed and compared the fatigue limits of PEEK polymers reinforced with carbon fibers (CFR-PEEK) and others with glass fibers (GFR-PEEK) against titanium for potential dental implant applications. These tests were conducted in accordance with ISO 14801, which outlines the application of cyclic loads under dry conditions and at room temperature on cylindrical specimens with diameters of 4 and 5 mm. This was done to closely replicate the forces transmitted to the implant during chewing: cyclic impacts on the bone/implant over a limited time period. The fatigue limit of GFR-PEEK specimens with a diameter of 4 mm was found to be 310 N, and this was higher than its maximum static compression force of 256 N. According to the authors, this vital phenomenon is due to the fact that the deformation rate can increase the elastic limit of PEEK materials by 30%. This elasticity of PEEK might also explain the high fatigue limits. The authors concluded that implants made of this material could be suitable for replacing anterior teeth where the maximum chewing forces in the incisive zone reach 140–170 N. Otherwise, the fatigue limit of CFR-PEEK specimens with a diameter of 5 mm was found to be 450 N, which was also higher than their maximum static compression force. This material was therefore favored as it appears suitable for the replacement of both anterior and posterior teeth, where the maximum chewing forces in the molar zone range from 200 to 250 N. Dental implants with PEEK polymer coatings or those based on PEEK would thus be advantageous in reducing the effects of “stress shielding.” However, the authors note that the fatigue tests were conducted at room temperature, and even though PEEK is not affected by hydrolysis, concerns are raised regarding its ability to maintain good bonding at the interface between the resin and its reinforcements. There are still too few fatigue tests in saline solution, but the initial results on CFR-PEEK samples seem to indicate good in vivo stability and long-term physical and mechanical properties of PEEK.
6 Finite elements methods
The advancement of dental implant, abutment, and associated materials necessitates adherence to established in vitro and in vivo tests before they are fit for commercial use. Conducting in vivo tests on a substantial number of samples can be lengthy and may pose challenges to animals or patients involved. Moreover, potential inaccuracies in these assessments could lead to misleading outcomes, impacting both the materials and the results of the experiment. Therefore, the FEM emerges as a valuable tool for predicting potential mechanical complications using theoretical models coupled with in vitro tests [22,23,35,39,44,123]. The FEA is frequently used to assess the biomechanical behavior of implantable biomaterials in vitro, whether in orthopedics, traumatology, or oral implantology. This method allows for the evaluation of stress distribution during the load transfer of an implant to the underlying bone structures and typically follows a consistent procedure.
The process involves obtaining a radiographic image of the bone section that will be studied. This image is then converted into a 3D model using computer-aided design (CAD) software. This model becomes the foundation for tests conducted on various models created with each implantable biomaterial.
The articles chosen for this section explore the utilization of FEA in examining the mechanical characteristics of abutments and implants. The process typically involves the creation of a 3D bone model through a CT scan, followed by importing into CAD software such as ANSYS, ABAQUS, SolidWorks, Creo Parametric, or Rhinoceros. A FEA package is then used to generate a 3D mesh which is linked to a suitable mathematical model that captures the properties of the corresponding material. While most of the studies employed ANSYS for simulation [22,23,32,33,39,44], some researchers favored ABAQUS, and CATIA [124–127]. Most of the selected studies assume a linear elastic, isotropic, and homogenous material model. In addition to using fewer parameters, in a linear elastic model, the correlation between these parameters and clinical medical imaging has been well established, which greatly enhances its potential for clinical application [127]. Whereas, an anisotropic material model is found to be more suitable for CFR-PEEK laminates [34]. CFR-PEEK laminates consist of carbon fiber-reinforced PEEK, which exhibits varying stiffness and strength properties in different directions relative to the orientation of the carbon fibers within the laminate. An anisotropic material model can accurately capture these directional dependencies, allowing for more precise simulations and prediction of mechanical behavior. This is particularly crucial in the context of dental implants and abutments, where a comprehensive understanding of material response under diverse loading conditions is essential (Figure 8).

Schematic of an implant-supported prosthesis involving implant, abutment, and a single crown.
The mechanical performance of implants, including factors such as von Mises stress, shear stress, contact stress, and stress shielding, is analyzed under varying loading conditions corresponding to clinical scenarios. An integral consideration in FEA methodologies is the assumption of forces to replicate masticatory loads encountered within the oral cavity of patients. However, defining this parameter precisely remains elusive due to its variability among individuals [39]. The magnitude of axial loading in the selected journals ranges widely from 100 to 1,000 N, while oblique loading typically spans from 50 to 500 N [29,33]. Notably, a predominant trend in these studies is the utilization of axial loading within the range of 100 to 150 N. Furthermore, the angle of oblique loading has slight variation, typically falling between 30° and 45°.
The selected journals aimed to investigate the stresses developed in dental implants and bones using various prosthesis materials, including titanium, zirconium, unfilled, and CFR-PEEK, utilized in abutments and implants, and their combinations. Oblique loading typically results in higher stresses in cortical and cancellous bone, implant components, and restorative crowns compared to axial loading across all mentioned studies. The findings of all authors are consistent with each other. A higher stress concentration is generally observed in the cortical bone near the peri-implant region, gradually decreasing towards the cancellous bone under both axial and oblique loading conditions.
Most studies indicate no significant variation in stress distribution at the implant-bone interface for different material models. Mourya et al.’s study of stress distribution in straight and angled abutments supports this conclusion, suggesting similar stress results with CFR-PEEK and titanium implants in both straight and inclined cases [29].
Another study on the performance of titanium/PEEK dental implants simulating different levels of bone loss also indicates that stresses were independent of the implant material [30]. Similarly, Sarot et al.’s work on the stress distribution of CFR-PEEK dental implants concludes that CFR-PEEK implants offer no advantages over titanium implants concerning stress distribution to the peri-implant bone [44]. Shash et al.’s study on fixed prosthesis following the “all on four” framework indicates that 60% CFR-PEEK distributes cortical bone stresses similar to the titanium framework [23] (Figure 9).
![Figure 9
The distribution of von Mises stresses (MPa) on the mucosa, cortical and spongy bones, under unilateral force, using Ti and PEEK [27].](/document/doi/10.1515/rams-2024-0031/asset/graphic/j_rams-2024-0031_fig_009.jpg)
The distribution of von Mises stresses (MPa) on the mucosa, cortical and spongy bones, under unilateral force, using Ti and PEEK [27].
Studies by Wazeh et al. [33] and Kaleli et al. [35] agree that crown material does not affect biomechanical behavior concerning stresses in implants and peripheral bone.
However, Mourya et al.‘s findings on stress distribution under parafunctional loading show that the stresses generated in bone with a PEEK crown layered with composite are lower compared to porcelain fused to metal (PFM) crown configurations [29] (Figure 10).
![Figure 10
Stress in: (a) CFR-PEEK 15° abutment under vertical loading with PFM crown, (b) CFR-PEEK straight abutment under oblique loading with (PFM) crown, (c) titanium straight abutment under vertical loading with (PFM) crown, and (d) titanium straight abutment under oblique loading with (PFM) crown [29].](/document/doi/10.1515/rams-2024-0031/asset/graphic/j_rams-2024-0031_fig_010.jpg)
Stress in: (a) CFR-PEEK 15° abutment under vertical loading with PFM crown, (b) CFR-PEEK straight abutment under oblique loading with (PFM) crown, (c) titanium straight abutment under vertical loading with (PFM) crown, and (d) titanium straight abutment under oblique loading with (PFM) crown [29].
Implants with micro threads showed a reduction of maximum von Mises stress by 50–70% in peri-implant bone compared to conventional threading [33], as demonstrated by Wazeh et al. Specifically, micro-threaded titanium implants, when used with a zirconia crown, exhibited the lowest cortical bone stress values under vertical loading, approximately 10 MPa. In contrast, implants made of CFR-PEEK with 30% carbon fibers generated notably higher stresses, around 100 MPa, on the cortical bone.
Titanium implants may induce a stress-shielding effect, leading to implant and bone loss due to their high elastic modulus (110 GPa) compared to bone (14 GPa). The stiffness mismatch between bone and implant can result in stress-shielding phenomena, potentially leading to bone resorption during the bone remodeling stage. In vivo, PEEK material possesses biomechanical properties close to human bone, reducing the risk of bone resorption and osteolysis caused by the stress-shielding effect of implants. A study by Ouldyerou et al. found that Ti-PEEK implants outperform conventional Titanium implants in reducing stress shielding and bone resorption [30]. The application of PEEK veneering on dental implants or abutments was also found to diminish the effects of stress shielding, as indicated by studies conducted by Lee et al. and Sampaio et al. [37,42]. Nevertheless, the thickness of PEEK veneering emerged as a critical factor influencing both stress distribution and the strength of hybrid abutments or implants, as highlighted in the research by Sampaio et al. [37].
Implants or abutments made of PEEK containing 30% carbon fiber exhibited increased stiffness compared to unfilled PEEK. The incorporation of such PEEK composites was found to enhance stress distribution and reduce the load concentration throughout the abutment and implant region, as suggested by Sarot et al. [44].
In weaving together, the detailed insights provided by the various studies, it becomes evident that FEA offers a robust framework for simulating and understanding the biomechanical behavior of dental implants and abutments under various conditions. Through meticulous modeling and simulation, FEA enables researchers and practitioners to predict and mitigate potential complications associated with implant materials and designs, thereby facilitating the development of more effective and safer dental restoration solutions. The consensus across studies on the efficacy of materials like CFR-PEEK and the nuances of stress distribution underlines the complexity of dental biomechanics and the indispensable role of FEA in advancing dental implantology. As the field evolves, the ongoing refinement of FEA methodologies and the exploration of new materials will undoubtedly contribute to enhancing the quality and longevity of dental implants, ultimately benefiting patients with more reliable and durable dental restorations.
7 Conclusion
The manuscript thoroughly examines the application and mechanical performance of PEEK and its composites in dental implantology, particularly regarding implant abutments and cores. Through FEA, it highlights PEEK’s pivotal role in advancing dental implant technology.
7.1 Application of PEEK and its composites in dental field
The manuscript discusses PEEK and its composites’ applications in dental implantology, citing their biomechanical compatibility and bioactivity. PEEK emerges as an alternative to titanium for abutments and dental implant cores due to its similar elasticity to bone and ease of removal. It outlines PEEK’s advantages in creating abutments, frameworks, and core materials, including its commendable soft tissue response and customization potential for provisional restorations and shaping emergency contours during surgical interventions.
7.2 Mechanical tests on PEEK
The manuscript extensively explores PEEK’s mechanical properties via tensile, compression, bending stress, wear resistance, and fatigue tests, revealing its robustness and suitability for dental implants. CFR-PEEK and GFR-PEEK composites show increased hardness, compression resistance, and fatigue endurance, reducing stress-shielding in bone. Tests, compliant with ISO 14801, demonstrate their ability to withstand chewing forces, suggesting PEEK-based materials could offer longer lasting implant solutions with more even load distribution.
7.3 Enhancement of understanding through FEA
The manuscript highlights FEA’s crucial role in advancing dental implant technology, allowing simulation of biomechanical behavior and stress distribution analysis in CFR-PEEK implants and abutments. By generating 3D models from radiographic images and using software like ANSYS and ABAQUS, FEA predicts stress distribution during load transfer, crucial for implant-bone interface evaluation. This method effectively assesses CFR-PEEK composites’ mechanical performance, reducing stress concentrations and improving load distribution, essential for implant longevity and success.
In conclusion, the widespread adoption of a new biomaterial is always a slow and cautious process. FEA emerges as a powerful tool in the development and evaluation of dental implants, providing valuable insights into material behavior and design efficacy. The ongoing evolution of this methodology, coupled with advancements in biomaterials, holds the promise of more effective, tailored, and biocompatible implant solutions, ultimately enhancing patient outcomes in dental and orthopedic implantology.
Acknowledgments
We would like to extend our sincere gratitude to the reviewers of this paper for their valuable time, insightful comments, and constructive feedback. Their expertise and meticulous evaluation have significantly contributed to enhancing the quality and clarity of our work.
-
Funding information: The authors state no funding involved.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Conflict of interest: The authors state no conflict of interest.
-
Data availability statement: Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.
References
[1] Blatz, M. B., G. Chiche, O. Bahat, R. Roblee, C. Coachman, and H. O. Heymann. Evolution of aesthetic dentistry. Journal of Dental Research, Vol. 98, 2019, pp. 1294–1304.10.1177/0022034519875450Search in Google Scholar PubMed
[2] Gazo Hanna, E., K. Younes, R. Roufayel, M. Khazaal, and Z. Fajloun. Engineering innovations in medicine and biology: Revolutionizing patient care through mechanical solutions. Heliyon, Vol. 10, No. 4.10.1016/j.heliyon.2024.e26154Search in Google Scholar PubMed PubMed Central
[3] Committee on Polymer Science and Engineering, National Research Council. Read ‘polymer science and engineering: The shifting research frontiers’ at NAP.edu, The National Academies Press, Washington, DC, 1994.Search in Google Scholar
[4] Rokaya, D., V. Srimaneepong, J. Sapkota, J. Qin, K. Siraleartmukul, and V. Siriwongrungson. Polymeric materials and films in dentistry: An overview. Journal of Advanced Research, Vol. 14, 2018, pp. 25–34.10.1016/j.jare.2018.05.001Search in Google Scholar PubMed PubMed Central
[5] Gazo-Hanna, E., A. Poitou, P. Casari, and L. Juras. Study of interply slip during thermoforming of continous fiber composite materials. In Proceedings of The18th International Conference on Composite Materials (ICCM-18), Jeju Island, Korea, 2011. https://www.iccm-central.org/Proceedings/ICCM18proceedings/data/2.%20Oral%20Presentation/Aug22(Monday)/M11%20Processing%20and%20Manufacturing%20Technologies/M11-6-IF0306.pdf.Search in Google Scholar
[6] Tigmeanu, C. V., L. C. Ardelean, L.-C. Rusu, and M. L. Negrutiu. Additive manufactured polymers in dentistry, current state-of-the-art and future perspectives-A review. Polymers, Vol. 14, 2022, id. 3658.10.3390/polym14173658Search in Google Scholar PubMed PubMed Central
[7] Gazo Hanna, E. Recycling of waste mixed plastics blends (PE/PP). Journal of Engineering Science & Technology Review, Vol. 12, No. 2, 2019, pp. 87–92.10.25103/jestr.122.12Search in Google Scholar
[8] Burg, K. J. L., and S. W. Shalaby. PES and PEEK. Encyclopedia of materials: Science and technology, K. H. J. Buschow, Cahn, R. W., Flemings, M. C., et al., (Eds.). Elsevier, Oxford, 2003, pp. 6837–6839.10.1016/B0-08-043152-6/01212-2Search in Google Scholar
[9] Ling, X., X. Jing, C. Zhang, and S. Chen. Polyether Ether Ketone (PEEK) properties and its application status. IOP Conference Series: Earth and Environmental Science, Vol. 453, 2020, id. 012080.10.1088/1755-1315/453/1/012080Search in Google Scholar
[10] Kurtz, S. M. and J. N. Devine. PEEK biomaterials in trauma, orthopedic, and spinal implants. Biomaterials, Vol. 28, 2007, pp. 4845–4869.10.1016/j.biomaterials.2007.07.013Search in Google Scholar PubMed PubMed Central
[11] Panayotov, I. V., V. Orti, F. Cuisinier, and J. Yachouh. Polyetheretherketone (PEEK) for medical applications. Journal of Materials Science. Materials in Medicine, Vol. 27, 2016, id. 118.10.1007/s10856-016-5731-4Search in Google Scholar PubMed
[12] Wei, Z., Z. Zhang, W. Zhu, and X. Weng. Polyetheretherketone development in bone tissue engineering and orthopedic surgery. Frontiers in Bioengineering and Biotechnology, Vol. 11, 2023. accessed 19 September 2023. https://www.frontiersin.org/articles/10.3389/fbioe.2023.1207277.10.3389/fbioe.2023.1207277Search in Google Scholar PubMed PubMed Central
[13] Mian, S. H., K. Moiduddin, S. M. Elseufy, and H. Alkhalefah. Adaptive mechanism for designing a personalized cranial implant and its 3D printing using PEEK. Polymers, Vol. 14, 2022, id. 1266.10.3390/polym14061266Search in Google Scholar PubMed PubMed Central
[14] Sikder, P., B. T. Challa, and S. K. Gummadi. A comprehensive analysis on the processing-structure-property relationships of FDM-based 3-D printed polyetheretherketone (PEEK) structures. Materialia, Vol. 22, 2022, id. 101427.10.1016/j.mtla.2022.101427Search in Google Scholar
[15] Hanemann, T., A. Klein, S. Baumgärtner, J. Jung, D. Wilhelm, and S. Antusch. Material extrusion 3D printing of PEEK-based composites. Polymers, Vol. 15, 2023, id. 3412.10.3390/polym15163412Search in Google Scholar PubMed PubMed Central
[16] Okdeh, N., G. Mahfouz, J. Harb, J. M. Sabatier, R. Roufayel, E. Gazo Hanna, et al. Protective role and functional engineering of neuropeptides in depression and anxiety: An overview. Bioengineering, Vol. 10, 2023, id. 258.10.3390/bioengineering10020258Search in Google Scholar PubMed PubMed Central
[17] Fabris, D., J. C. Souza, F. S. Silva, M. Fredel, J. Mesquita-Guimarães, Y. Zhang, et al. The bending stress distribution in bilayered and graded zirconia-based dental ceramics. Ceramics International, Vol. 42, 2016, pp. 11025–11031.10.1016/j.ceramint.2016.03.245Search in Google Scholar PubMed PubMed Central
[18] Fabris, D., J. C. M. Souza, F. S. Silva, M. Fredel, M. Gasik, and B. Henriques. Influence of specimens’ geometry and materials on the thermal stresses in dental restorative materials during thermal cycling. Journal of Dentistry, Vol. 69, 2018, pp. 41–48.10.1016/j.jdent.2017.08.017Search in Google Scholar PubMed
[19] Macedo, J. P., J. Pereira, J. Faria, J. Souza, J. L. Alves, J. López-López, et al. Finite element analysis of peri-implant bone volume affected by stresses around Morse taper implants: Effects of implant positioning to the bone crest. Computer Methods in Biomechanics and Biomedical Engineering, Vol. 21, 2018, pp. 655–662.10.1080/10255842.2018.1507025Search in Google Scholar PubMed
[20] Bins-Ely, L., D. Suzuki, R. Magini, C. Benfatti, W. Teughels, B. Henriques, et al. Enhancing the bone healing on electrical stimuli through the dental implant. Computer Methods in Biomechanics and Biomedical Engineering, Vol. 23, 2020, pp. 1041–1051.10.1080/10255842.2020.1785437Search in Google Scholar PubMed
[21] Khaohoen, A., T. Sornsuwan, P. Chaijareenont, et al. Biomaterials and Clinical Application of Dental Implants in Relation to Bone Density—A Narrative Review. Journal of Clinical Medicine, Vol. 12, 2023, id. 6924.10.3390/jcm12216924Search in Google Scholar PubMed PubMed Central
[22] Shash, Y. H., M. T. El-Wakad, M. A. A. El-Dosoky, et al. Evaluation of stresses on mandible bone and prosthetic parts in fixed prosthesis by utilizing CFR-PEEK, PEKK and PEEK frameworks. Scientific Reports, Vol. 13, 2023, id. 11542.10.1038/s41598-023-38288-2Search in Google Scholar PubMed PubMed Central
[23] Shash, Y. H., M. T. El-Wakad, M. A. A. Eldosoky, et al. Evaluation of stress and strain on mandible caused using ‘All-on-Four’ system from PEEK in hybrid prosthesis: finite-element analysis. Odontology / the Society of the Nippon Dental University, Vol. 111, 2023, pp. 618–629.10.1007/s10266-022-00771-zSearch in Google Scholar PubMed PubMed Central
[24] Ghazal-Maghras, R., J. Vilaplana-Vivo, F. Camacho-Alonso, et al. Properties of polyetheretheretherketone (PEEK) implant abutments: A systematic review. Journal of Clinical and Experimental Dentistry, Vol. 14, 2022, pp. e349–e358.10.4317/jced.59466Search in Google Scholar PubMed PubMed Central
[25] Taha, D., P. F. Cesar, and A. Sabet. Influence of different combinations of CAD-CAM crown and customized abutment materials on the force absorption capacity in implant supported restorations – In vitro study. Dental Materials, Vol. 38, 2022, pp. e10–e18.10.1016/j.dental.2021.12.025Search in Google Scholar PubMed
[26] Ortega-Martínez, J., L. M. Delgado, M. Ortiz-Hernández, et al. In vitro assessment of PEEK and titanium implant abutments: Screw loosening and microleakage evaluations under dynamic mechanical testing. Journal of Prosthetic Dentistry, Vol. 127, 2022, pp. 470–476.10.1016/j.prosdent.2020.09.033Search in Google Scholar PubMed
[27] Chokaree, P., P. Poovarodom, P. Chaijareenont, et al. Biomaterials and Clinical Applications of Customized Healing Abutment—A Narrative Review. Journal of Functional Biomaterials, Vol. 13, 2022, id. 291.10.3390/jfb13040291Search in Google Scholar PubMed PubMed Central
[28] Suphangul, S., D. Rokaya, C. Kanchanasobhana, et al. PEEK Biomaterial in Long-Term Provisional Implant Restorations: A Review. Journal of Functional Biomaterials, Vol. 13, 2022, id. 33.10.3390/jfb13020033Search in Google Scholar PubMed PubMed Central
[29] Mourya, A., R. Nahar, S. K. Mishra, et al. Stress distribution around different abutments on titanium and CFR-PEEK implant with different prosthetic crowns under parafunctional loading: A 3D FEA study. Journal of Oral Biology and Craniofacial Research, Vol. 11, 2021, pp. 313–320.10.1016/j.jobcr.2021.03.005Search in Google Scholar PubMed PubMed Central
[30] Ouldyerou, A., A. Merdji, L. Aminallah, et al. Biomechanical performance of Ti-PEEK dental implants in bone: An in-silico analysis. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 134, 2022, id. 105422.10.1016/j.jmbbm.2022.105422Search in Google Scholar PubMed
[31] Ragupathi, M., V. Mahadevan, N. S. Azhagarasan, et al. Comparative evaluation of the wear resistance of two different implant abutment materials after cyclic loading – An in vitro study. Contemporary Clinical Dentistry, Vol. 11, 2020, pp. 229–236.10.4103/ccd.ccd_294_19Search in Google Scholar PubMed PubMed Central
[32] Bataineh, K. and M. Al Janaideh. Effect of different biocompatible implant materials on the mechanical stability of dental implants under excessive oblique load. Clinical Implant Dentistry and Related Research, Vol. 21, 2019, pp. 1206–1217.10.1111/cid.12858Search in Google Scholar PubMed
[33] Wazeh, A. M., M. I. El-Anwar, R. M. G. Atia, et al. 3D FEA Study On Implant Threading Role on Selection of Implant and Crown Materials. Open Access Macedonian Journal of Medical Sciences, Vol. 6, 2018, pp. 1702–1706.10.3889/oamjms.2018.331Search in Google Scholar PubMed PubMed Central
[34] Gallagher, E. A., S. Lamorinière, and P. McGarry. Multi-axial damage and failure of medical grade carbon fibre reinforced PEEK laminates: Experimental testing and computational modelling. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 82, 2018, pp. 154–167.10.1016/j.jmbbm.2018.03.015Search in Google Scholar PubMed
[35] Kaleli, N., D. Sarac, S. Külünk, et al. Effect of different restorative crown and customized abutment materials on stress distribution in single implants and peripheral bone: A three-dimensional finite element analysis study. The Journal of Prosthetic Dentistry, Vol. 119, 2018, pp. 437–445.10.1016/j.prosdent.2017.03.008Search in Google Scholar PubMed
[36] Sampaio, M., M. Buciumeanu, B. Henriques, et al. Comparison between PEEK and Ti6Al4V concerning micro-scale abrasion wear on dental applications. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 60, 2016, pp. 212–219.10.1016/j.jmbbm.2015.12.038Search in Google Scholar PubMed
[37] Sampaio M., M. Buciumeanu, E. Askari, et al. Effects of poly-ether-ether ketone (PEEK) veneer thickness on the reciprocating friction and wear behavior of PEEK/Ti6Al4V structures in artificial saliva. Wear, Vol. 368–369, 2016, pp. 84–91.10.1016/j.wear.2016.09.009Search in Google Scholar
[38] Schwitalla, A. D., T. Spintig, I. Kallage, et al. Pressure behavior of different PEEK materials for dental implants. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 54, 2016, pp. 295–304.10.1016/j.jmbbm.2015.10.003Search in Google Scholar PubMed
[39] Schwitalla, A. D., M. Abou-Emara, T. Spintig, et al. Finite element analysis of the biomechanical effects of PEEK dental implants on the peri-implant bone. Journal of Biomechanics, Vol. 48, 2015, pp. 1–7.10.1016/j.jbiomech.2014.11.017Search in Google Scholar PubMed
[40] Schwitalla, A. D., T. Spintig, I. Kallage, et al. Flexural behavior of PEEK materials for dental application. Dental Materials, Vol. 31, 2015, pp. 1377–1384.10.1016/j.dental.2015.08.151Search in Google Scholar PubMed
[41] Najeeb, S., M. Zafar, Z. Sultan, K. Khurshid, and F. Siddiqui. Applications of PEEK in oral implantology and prosthodontics. Journal of Prosthodontic Research. Vol. 60, No. 1, 2015, pp. 12–19.10.1016/j.jpor.2015.10.001Search in Google Scholar PubMed
[42] Lee, W.-T., J.-Y. Koak, Y.-J. Lim, et al. Stress shielding and fatigue limits of poly-ether-ether-ketone dental implants. Journal of biomedical materials research. Part B, Applied biomaterials, Vol. 100, 2012, pp. 1044–1052.10.1002/jbm.b.32669Search in Google Scholar PubMed
[43] «metal-free» le Biopik - PDF Free Download, https://docplayer.fr/113494536-Metal-free-le-biopik.html (accessed 27 September 2023). 2010.Search in Google Scholar
[44] Sarot, J. R., C. M. M. Contar, A.C.C.da Cruz, et al. Evaluation of the stress distribution in CFR-PEEK dental implants by the three-dimensional finite element method. Journal of Materials Science. Materials in Medicine, Vol. 21, 2010, pp. 2079–2085.10.1007/s10856-010-4084-7Search in Google Scholar PubMed
[45] Ferguson, S. J., J. M. A. Visser, and A. Polikeit. The long-term mechanical integrity of non-reinforced PEEK-OPTIMA polymer for demanding spinal applications: experimental and finite-element analysis. European Spine Journal, Vol. 15, 2006, pp. 149–156.10.1007/s00586-005-0915-5Search in Google Scholar PubMed PubMed Central
[46] Cho, D.-Y., W.-R. Liau, W.-Y. Lee, et al. Preliminary experience using a polyetheretherketone (PEEK) cage in the treatment of cervical disc disease. Neurosurgery, Vol. 51, 2002, pp. 1343–1349. discussion 1349 – 1350.10.1097/00006123-200212000-00003Search in Google Scholar
[47] Costa-Palau, S., J. Torrents-Nicolas, M. Brufau-de Barberà, et al. Use of polyetheretherketone in the fabrication of a maxillary obturator prosthesis: a clinical report. Journal of Prosthetic Dentistry, Vol. 112, 2014, pp. 680–682.10.1016/j.prosdent.2013.10.026Search in Google Scholar PubMed
[48] Horák, Z., D. Pokorný, P. Fulín, et al. Polyetheretherketone (PEEK). Part I: prospects for use in orthopaedics and traumatology. Acta Chirurgiae Orthopaedicae et Traumatologiae Cechoslovaka, Vol. 77, 2010, pp. 463–469.10.55095/achot2010/077Search in Google Scholar
[49] Zhou, H., V. K. Goel, and S. B. Bhaduri. A fast route to modify biopolymer surface: A study on polyetheretherketone (PEEK). Materials Letters, Vol. 125, 2014, pp. 96–98.10.1016/j.matlet.2014.03.130Search in Google Scholar
[50] Cook, S. D. and A. M. Rust-Dawicki. Preliminary evaluation of titanium-coated PEEK dental implants. The Journal of Oral Implantology, Vol. 21, 1995, pp. 176–181.Search in Google Scholar
[51] Meningaud, J.-P., F. Spahn, and J.-M. Donsimoni. After titanium, peek?. Revue de Stomatologie et de Chirurgie Maxillo-Faciale, Vol. 113, 2012, pp. 407–410.10.1016/j.stomax.2011.12.016Search in Google Scholar PubMed
[52] Hanasono, M. M., N. Goel, and F. DeMonte. Calvarial reconstruction with polyetheretherketone implants. Annals of Plastic Surgery, Vol. 62, 2009, pp. 653–655.10.1097/SAP.0b013e318184abc7Search in Google Scholar PubMed
[53] Poulsson, A. H. C., D. Eglin, S. Zeiter, et al. Osseointegration of machined, injection moulded and oxygen plasma modified PEEK implants in a sheep model. Biomaterials, Vol. 35, 2014, pp. 3717–3728.10.1016/j.biomaterials.2013.12.056Search in Google Scholar PubMed
[54] Nakahara, I., M. Takao, T. Goto, et al. Interfacial shear strength of bioactive-coated carbon fiber reinforced polyetheretherketone after in vivo implantation. Journal of Orthopaedic Research, Vol. 30, 2012, pp. 1618–1625.10.1002/jor.22115Search in Google Scholar PubMed
[55] Nakahara, I., M. Takao, S. Bandoh, et al. In vivo implant fixation of carbon fiber-reinforced PEEK hip prostheses in an ovine model. Journal of Orthopaedic Research, Vol. 31, 2013, pp. 485–492.10.1002/jor.22251Search in Google Scholar PubMed
[56] Ma, R. and T. Tang. Current Strategies to Improve the Bioactivity of PEEK. International Journal of Molecular Sciences, Vol. 15, 2014, pp. 5426–5445.10.3390/ijms15045426Search in Google Scholar PubMed PubMed Central
[57] Wang, L., S. He, X. Wu, et al. Polyetheretherketone/nano-fluorohydroxyapatite composite with antimicrobial activity and osseointegration properties. Biomaterials, Vol. 35, 2014, pp. 6758–6775.10.1016/j.biomaterials.2014.04.085Search in Google Scholar PubMed
[58] Liao, C., Y. Li, and S. C. Tjong. Polyetheretherketone and Its Composites for Bone Replacement and Regeneration. Polymers (Basel), Vol. 12, 2020, id. 2858.10.3390/polym12122858Search in Google Scholar PubMed PubMed Central
[59] Zheng, Z., P. Liu, X. Zhang, et al. Strategies to improve bioactive and antibacterial properties of polyetheretherketone (PEEK) for use as orthopedic implants. Mater Today Bio, Vol. 16, 2022, id. 100402.10.1016/j.mtbio.2022.100402Search in Google Scholar PubMed PubMed Central
[60] Zheng, Y., C. Xiong, S. Zhang, et al. Bone-like apatite coating on functionalized poly(etheretherketone) surface via tailored silanization layers technique. Materials Science and Engineering: C, Vol. 55, 2015, pp. 512–523.10.1016/j.msec.2015.05.070Search in Google Scholar PubMed
[61] Richard-Isaac, A. Évaluation du Polyétheréthercétone (PEEK) en implantologie orale : intérêts, limites et perspectives [PhD thesis]. Revue de Littérature, 2016, id. 71.Search in Google Scholar
[62] Bonnheim, N., F. Ansari, M. Regis, et al. Effect of carbon fiber type on monotonic and fatigue properties of orthopedic grade PEEK. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 90, 2019, pp. 484–492.10.1016/j.jmbbm.2018.10.033Search in Google Scholar PubMed
[63] Qin, W., Y. Li, J. Ma, et al. Mechanical properties and cytotoxicity of hierarchical carbon fiber-reinforced poly (ether-ether-ketone) composites used as implant materials. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 89, 2019, pp. 227–233.10.1016/j.jmbbm.2018.09.040Search in Google Scholar PubMed
[64] Smółka, W., M. Dworak, B. Noworyta, M. Gubernat, J. Markowski, and M. Błażewicz. Mechanical and biological properties of carbon fiber-reinforced PEEK composite materials intended for laryngeal prostheses. Engineering of Biomaterials, Vol. 22, 2019, pp. 2–8.Search in Google Scholar
[65] Gowda, E. M., S. R. Iyer, K. Verma, and S. Murali Mohan. Evaluation of PEEK composite dental implants: A comparison of two different loading protocols. Journal of Dental Research and Reports; 1. Epub ahead of print, Vol. 1, January 2018, pp. 1–5.10.15761/JDRR.1000101Search in Google Scholar
[66] Barkarmo, S., A. Wennerberg, M. Hoffman, et al. Nano-hydroxyapatite-coated PEEK implants: A pilot study in rabbit bone. Journal of biomedical materials research. Part A, Vol. 101A, 2013, pp. 465–471.10.1002/jbm.a.34358Search in Google Scholar PubMed
[67] Najeeb, S., Z. Khurshid, J. P. Matinlinna, et al. Nanomodified Peek Dental Implants: Bioactive Composites and Surface Modification – A Review. International Journal of Dentistry, Vol. 2015, 2015, id. e381759.10.1155/2015/381759Search in Google Scholar PubMed PubMed Central
[68] Lai, Y. H., M. C. Kuo, J. C. Huang, et al. On the PEEK composites reinforced by surface-modified nano-silica. Materials Science and Engineering: A, Vol. 458, 2007, pp. 158–169.10.1016/j.msea.2007.01.085Search in Google Scholar
[69] Díez-Pascual, A. M. and A. L. Díez-Vicente. Nano-TiO2 Reinforced PEEK/PEI Blends as Biomaterials for Load-Bearing Implant Applications. ACS Applied Materials & Interfaces, Vol. 7, 2015, pp. 5561–5573.10.1021/acsami.5b00210Search in Google Scholar PubMed
[70] Rahmitasari, F., Y. Ishida, K. Kurahashi, et al. PEEK with Reinforced Materials and Modifications for Dental Implant Applications. Dent J (Basel), Vol. 5, 2017, id. 35.10.3390/dj5040035Search in Google Scholar PubMed PubMed Central
[71] Wheeler, S. L. Eight-year clinical retrospective study of titanium plasma-sprayed and hydroxyapatite-coated cylinder implants. International Journal of Oral and Maxillofacial Implants, Vol. 11, 1996, pp. 340–350.Search in Google Scholar
[72] Fu, Q., M. Gabriel, F. Schmidt, et al. The impact of different low-pressure plasma types on the physical, chemical and biological surface properties of PEEK. Dental Materials, Vol. 37, 2021, pp. e15–e22.10.1016/j.dental.2020.09.020Search in Google Scholar PubMed
[73] Almasi, D., N. Iqbal, M. Sadeghi, et al. Preparation Methods for Improving PEEK’s Bioactivity for Orthopedic and Dental Application: A Review. International Journal of Biomaterials, Vol. 2016, 2016, id. 8202653.10.1155/2016/8202653Search in Google Scholar PubMed PubMed Central
[74] Kruse, H. V., D. R. McKenzie, J. R. Clark, et al. Plasma ion implantation of 3D-printed PEEK creates optimal host conditions for bone ongrowth and mineralisation. Plasma Processes and Polymers, Vol. 18, 2021, id. 2000219.10.1002/ppap.202000219Search in Google Scholar
[75] Han, C.-M., E.-J. Lee, H.-E. Kim, et al. The electron beam deposition of titanium on polyetheretherketone (PEEK) and the resulting enhanced biological properties. Biomaterials, Vol. 31, 2010, pp. 3465–3470.10.1016/j.biomaterials.2009.12.030Search in Google Scholar PubMed
[76] Tekin, S., S. Cangül, Ö. Adıgüzel, et al. Areas for use of PEEK material in dentistry. International Dental Research, Vol. 8, 2018, pp. 84–92.10.5577/intdentres.2018.vol8.no2.6Search in Google Scholar
[77] Rupawat, D., D. Nallaswamy, J. Somasundaram, et al. An Invitro Chewing Simulation Study Comparing the Wear Resistance Behavior of Polyetheretherketone-Layered Composite Crown and Ceramic-Layered Zirconia Crown. Cureus, Vol. 15, 2023, id. e46439.10.7759/cureus.46439Search in Google Scholar PubMed PubMed Central
[78] Coray, R., M. Zeltner, and M. Özcan. Fracture strength of implant abutments after fatigue testing: A systematic review and a meta-analysis. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 62, 2016, pp. 333–346.10.1016/j.jmbbm.2016.05.011Search in Google Scholar PubMed
[79] Koutouzis, T., J. Richardson, and T. Lundgren. Comparative Soft and Hard Tissue Responses to Titanium and Polymer Healing Abutments. Journal of Oral Implantology, Vol. 37, 2011, pp. 174–182.10.1563/AAID-JOI-D-09-00102.1Search in Google Scholar PubMed
[80] Al-Rabab’ah, M., W. Hamadneh, I. Alsalem, et al. Use of High Performance Polymers as Dental Implant Abutments and Frameworks: A Case Series Report. Journal of Prosthodontics, Vol. 28, 2019, pp. 365–372.10.1111/jopr.12639Search in Google Scholar PubMed
[81] Al-Zordk, W., A. Elmisery, and M. Ghazy. Hybrid-abutment-restoration: effect of material type on torque maintenance and fracture resistance after thermal aging. International Journal of Implant Dentistry, Vol. 6, 2020, id. 24.10.1186/s40729-020-00220-ySearch in Google Scholar PubMed PubMed Central
[82] Zarone, F., R. Sorrentino, T. Traini, et al. Fracture resistance of implant-supported screw- versus cement-retained porcelain fused to metal single crowns: SEM fractographic analysis. Dental Materials, Vol. 23, 2007, pp. 296–301.10.1016/j.dental.2005.10.013Search in Google Scholar PubMed
[83] Freitas, A. C., E. A. Bonfante, E. P. Rocha, et al. Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crowns. European Journal of Oral Sciences, Vol. 119, 2011, pp. 323–330.10.1111/j.1600-0722.2011.00837.xSearch in Google Scholar PubMed
[84] Atsü, S. S., M. E. Aksan, and A. C. Bulut. Fracture Resistance of Titanium, Zirconia, and Ceramic-Reinforced Polyetheretherketone Implant Abutments Supporting CAD/CAM Monolithic Lithium Disilicate Ceramic Crowns After Aging. The International Journal of Oral & Maxillofacial Implants, Vol. 34, 2019, pp. 622–630.10.11607/jomi.7036Search in Google Scholar PubMed
[85] Rozeik, A. S., M. S. Chaar, S. Sindt, et al. Cellular properties of human gingival fibroblasts on novel and conventional implant-abutment materials. Dental Materials, Vol. 38, 2022, pp. 540–548.10.1016/j.dental.2021.12.139Search in Google Scholar PubMed
[86] Neodent. Gm Customizable Healing Abutment. Strautmann, Gm Customizable Healing Abutment, Peek/Titanium, 7.0X5.5 mm (2024).Search in Google Scholar
[87] Neodent, G. M. Pro Peek Abutment, Titanium/Peek, 6.0X5.5 mm. Strautmann, GM Pro Peek Abutment, Titanium/Peek, 6.0X5.5 mm (2024).Search in Google Scholar
[88] Wang, B., M. Huang, P. Dang, et al. PEEK in fixed dental prostheses: Application and adhesion improvement. Polymers (Basel), Vol. 14, 2022, id. 2323.10.3390/polym14122323Search in Google Scholar PubMed PubMed Central
[89] Alothman, Y. and M. S. Bamasoud. The success of dental veneers according to preparation design and material type. Open Access Macedonian Journal of Medical Sciences, Vol. 6, 2018, pp. 2402–2408.10.3889/oamjms.2018.353Search in Google Scholar PubMed PubMed Central
[90] Papathanasiou, I., P. Kamposiora, G. Papavasiliou, et al. The use of PEEK in digital prosthodontics: A narrative review. BMC Oral health, Vol. 20, 2020, id. 217.10.1186/s12903-020-01202-7Search in Google Scholar PubMed PubMed Central
[91] Mayinger, F., D. Micovic, A. Schleich, et al. Retention force of polyetheretherketone and cobalt-chrome-molybdenum removable dental prosthesis clasps after artificial aging. Clinical Oral Investigations, Vol. 25, 2021, pp. 3141–3149.10.1007/s00784-020-03642-5Search in Google Scholar PubMed PubMed Central
[92] Silva Júnior, E. V., R. T. Basting, C. P. Turssi, et al. Precision of polyether ether ketone (PEEK) or cobalt-chrome implant bar fit to implants after mechanical cycling. Acta odontológica latinoamericana : AOL, Vol. 36, 2023, pp. 71–77.10.54589/aol.36/2/71Search in Google Scholar PubMed PubMed Central
[93] Zoidis, P. and I. Papathanasiou. Modified PEEK resin-bonded fixed dental prosthesis as an interim restoration after implant placement. The Journal of Prosthetic Dentistry, Vol. 116, 2016, pp. 637–641.10.1016/j.prosdent.2016.04.024Search in Google Scholar PubMed
[94] Sarfraz, S., P.-H. Mäntynen, M. Laurila, et al. Comparison of Titanium and PEEK Medical Plastic Implant Materials for Their Bacterial Biofilm Formation Properties. Polymers (Basel), Vol. 14, 2022, id. 3862.10.3390/polym14183862Search in Google Scholar PubMed PubMed Central
[95] Pérez-Pevida, E., D. Chávarri-Prado, M. Diéguez-Pereira, et al. Consequences of Peri-Implant Bone Loss in the Occlusal Load Transfer to the Supporting Bone in terms of Magnitude of Stress, Strain, and Stress Distribution: A Finite Element Analysis. BioMed Research International, Vol. 2021, 2021, id. 3087071.10.1155/2021/3087071Search in Google Scholar PubMed PubMed Central
[96] Torstrick, F. B., A. S. P. Lin, D. Potter, et al. Porous PEEK improves the bone-implant interface compared to plasma-sprayed titanium coating on PEEK. Biomaterials, Vol. 185, 2018, pp. 106–116.10.1016/j.biomaterials.2018.09.009Search in Google Scholar PubMed
[97] Ma, R. and D. Guo. Evaluating the bioactivity of a hydroxyapatite-incorporated polyetheretherketone biocomposite. Journal of Orthopaedic Surgery and Research, Vol. 14, 2019, id. 32.10.1186/s13018-019-1069-1Search in Google Scholar PubMed PubMed Central
[98] Invibio Biomaterial Solutions - Advancing Implantable Devices. Invibio, 2024. https://invibio.com/en/ (accessed 12 October 2023).Search in Google Scholar
[99] BioHPP - Analogue and digital workflows, 2024. https://www.bredent-implants.com/products-solutions/therapies/skyonics-therapies/skyonics-biohpp-sky-elegance/ (accessed 12 October 2023).Search in Google Scholar
[100] Themes UFO. Physical and Mechanical Properties of Dental Materials. Pocket Dentistry. https://pocketdentistry.com/physical-and-mechanical-properties-of-dental-materials-2/(2020, accessed 10 October 2023)Search in Google Scholar
[101] Accelerator, A. Mechanical properties of biomaterials: Most up-to-date encyclopedia, news & reviews. . Academic Accelerator, 2007. https://academic-accelerator.com/encyclopedia/mechanical-properties-of-biomaterials(accessed 10 October 2023)Search in Google Scholar
[102] Kewalramani Kewalramani, N. Análisis in vitro de la elasticidad y la flexión del Peek. https://hdl.handle.net/20.500.14352/19990(2018, accessed 10 October 2023) Master Thesis, Universidad Complutense Madrid Search in Google Scholar
[103] Carpenter, R. D., B. S. Klosterhoff, F. B. Torstrick, et al. Effect of porous orthopaedic implant material and structure on load sharing with simulated bone ingrowth: A finite element analysis comparing titanium and PEEK. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 80, 2018, pp. 68–76.10.1016/j.jmbbm.2018.01.017Search in Google Scholar PubMed PubMed Central
[104] Sandler, J., P. Werner, M. S. P. Shaffer, et al. Carbon-nanofibre-reinforced poly(ether ether ketone) composites. Composites, Part A: Applied Science and Manufacturing, Vol. 33, 2002, pp. 1033–1039.10.1016/S1359-835X(02)00084-2Search in Google Scholar
[105] Attia, M. A., T. E. Shokry, and M. Abdel-Aziz. Effect of different surface treatments on the bond strength of milled polyetheretherketone posts. Journal of Prosthetic Dentistry, Vol. 127, 2022, pp. 866–874.10.1016/j.prosdent.2020.08.033Search in Google Scholar PubMed
[106] Wu, H., Y. Guo, and W. Guo. Effect of carbon-fiber-reinforced polyetheretherketone on stress distribution in a redesigned tumor-type knee prosthesis: a finite element analysis. Frontiers in Bioengineering and Biotechnology, Vol. 11. https://www.frontiersin.org/articles/10.3389/fbioe.2023.1243936 (2023, accessed 5 October 2023).10.3389/fbioe.2023.1243936Search in Google Scholar PubMed PubMed Central
[107] Gao, X., M. Fraulob, and G. Haïat. Biomechanical behaviours of the bone–implant interface: a review. Journal of the Royal Society Interface, Vol. 16, 2019, id. 20190259.10.1098/rsif.2019.0259Search in Google Scholar PubMed PubMed Central
[108] Mechanical Properties of Titanium Used in Dental Implants | Inside Dentistry, 2021. https://cdeworld.com/courses/5297-mechanical-properties-of-titanium-used-in-dental-implants (accessed 5 October 2023).Search in Google Scholar
[109] Vallittu, P. K. Some aspects of the tensile strength of unidirectional glass fibre-polymethyl methacrylate composite used in dentures. Journal of Oral Rehabilitation, Vol. 25, 1998, pp. 100–105.10.1046/j.1365-2842.1998.00235.xSearch in Google Scholar PubMed
[110] Comparative Study of Chemical and Mechanical Surface Treatment Effects on Shear Bond Strength of PEEK to Veneering Ceramic, 2021. http://quintpub.com/journals/ijp/abstract.php?iss2_id=1809&article_id=22252 (accessed 5 October 2023).Search in Google Scholar
[111] Fabris, D., J. P. A. Moura, M. C. Fredel, et al. Biomechanical analyses of one-piece dental implants composed of titanium, zirconia, PEEK, CFR-PEEK, or GFR-PEEK: Stresses, strains, and bone remodeling prediction by the finite element method. Journal of biomedical materials research. Part B, Applied biomaterials, Vol. 110, 2022, pp. 79–88.10.1002/jbm.b.34890Search in Google Scholar PubMed
[112] Yoon, H. G., H. K. Oh, D.-Y. Lee, et al. 3-D finite element analysis of the effects of post location and loading location on stress distribution in root canals of the mandibular 1st molar. Journal of Applied Oral Science, Vol. 26, 2018, id. e20160406.10.1590/1678-7757-2016-0406Search in Google Scholar PubMed PubMed Central
[113] Cavalli, V., M. Giannini, and R. M. Carvalho. Effect of carbamide peroxide bleaching agents on tensile strength of human enamel. Dental Materials, Vol. 20, 2004, pp. 733–739.10.1016/j.dental.2003.10.007Search in Google Scholar PubMed
[114] Sano, H., B. Ciucchi, W. G. Matthews, et al. Tensile Properties of Mineralized and Demineralized Human and Bovine Dentin. Journal of Dental Research, Vol. 73, 1994, pp. 1205–1211.10.1177/00220345940730061201Search in Google Scholar PubMed
[115] Chun, K., H. Choi, and J. Lee. Comparison of mechanical property and role between enamel and dentin in the human teeth. Journal of Dental Biomechanics, Vol. 5, 2014, id. 1758736014520809.10.1177/1758736014520809Search in Google Scholar PubMed PubMed Central
[116] Tamrakar, S. K., S. K. Mishra, R. Chowdhary, et al. Comparative analysis of stress distribution around CFR‑PEEK implants and titanium implants with different prosthetic crowns: A finite element analysis. Dent Med Probl, Vol. 58, 2021, pp. 359–367.10.17219/dmp/133234Search in Google Scholar PubMed
[117] Maekawa, M., Z. Kanno, T. Wada, et al. Mechanical properties of orthodontic wires made of super engineering plastic. Dental Materials Journal, Vol. 34, 2015, pp. 114–119.10.4012/dmj.2014-202Search in Google Scholar PubMed
[118] Martin, R. B. and J. Ishida. The relative effects of collagen fiber orientation, porosity, density, and mineralization on bone strength. Journal of Biomechanics, Vol. 22, 1989, pp. 419–426.10.1016/0021-9290(89)90202-9Search in Google Scholar PubMed
[119] García-González, M., S. Blasón-González, I. García-García, et al. Optimized Planning and Evaluation of Dental Implant Fatigue Testing: A Specific Software Application. Biology (Basel), Vol. 9, 2020, id. 372.10.3390/biology9110372Search in Google Scholar PubMed PubMed Central
[120] Materiales Dentales de Ricardo Luis Macchi | Editorial Médica Panamericana, 2007. https://www.medicapanamericana.com/co/libro/materiales-dentales-version-digital (accessed 11 October 2023).Search in Google Scholar
[121] Phillips’ Science of Dental Materials, https://www.booktopia.com.au/phillips-science-of-dental-materials-chiayi-shen/ebook/9780323697569.html (accessed 11 October 2023).Search in Google Scholar
[122] Beer, F. P., J. T. DeWolf, and E. R. Johnston Mechanics of materials, New York [u.a.] : McGraw-Hill, http://archive.org/details/mechanicsofmater00ferd_1 (1992, accessed 11 October 2023).Search in Google Scholar
[123] Shash, Y. H., M. M. Dohiem, M. T. Elwakad, and M. Eldosoky. Effect of bar material on the biomechanical performance of hybrid dental prosthesis. In 2021 3rd Novel Intelligent and Leading Emerging Sciences Conference (NILES), pp. 37–40.10.1109/NILES53778.2021.9600489Search in Google Scholar
[124] Oswal, M. M., U. N. Amasi, M. S. Oswal, et al. Influence of three different implant thread designs on stress distribution: A three-dimensional finite element analysis. The Journal of the Indian Prosthodontic Society, Vol. 16, 2016, pp. 359–365.10.4103/0972-4052.191283Search in Google Scholar PubMed PubMed Central
[125] Darvish, D., S. Khorramymehr, and M. Nikkhoo. Finite Element Analysis of the Effect of Dental Implants on Jaw Bone under Mechanical and Thermal Loading Conditions. Mathematical Problems in Engineering, Vol. 2021, 2021, id. e9281961.10.1155/2021/9281961Search in Google Scholar
[126] Dorogoy, A., D. Rittel, K. Shemtov-Yona, et al. Modeling dental implant insertion. Journal of the Mechanical Behavior of Biomedical Materials, Vol. 68, 2017, pp. 42–50.10.1016/j.jmbbm.2017.01.021Search in Google Scholar PubMed
[127] Lu, Y. 11 - Computational modelling of bone microstructure. In Computational Modelling of Biomechanics and Biotribology in the Musculoskeletal System, 2nd edn Z. Jin, Li, J., Chen, Z., (Eds.), Woodhead Publishing, pp. 251–276.10.1016/B978-0-12-819531-4.00011-0Search in Google Scholar
© 2024 the author(s), published by De Gruyter
This work is licensed under the Creative Commons Attribution 4.0 International License.
Articles in the same Issue
- Review Articles
- Effect of superplasticizer in geopolymer and alkali-activated cement mortar/concrete: A review
- Experimenting the influence of corncob ash on the mechanical strength of slag-based geopolymer concrete
- Powder metallurgy processing of high entropy alloys: Bibliometric analysis and systematic review
- Exploring the potential of agricultural waste as an additive in ultra-high-performance concrete for sustainable construction: A comprehensive review
- A review on partial substitution of nanosilica in concrete
- Foam concrete for lightweight construction applications: A comprehensive review of the research development and material characteristics
- Modification of PEEK for implants: Strategies to improve mechanical, antibacterial, and osteogenic properties
- Interfacing the IoT in composite manufacturing: An overview
- Advances in processing and ablation properties of carbon fiber reinforced ultra-high temperature ceramic composites
- Advancing auxetic materials: Emerging development and innovative applications
- Revolutionizing energy harvesting: A comprehensive review of thermoelectric devices
- Exploring polyetheretherketone in dental implants and abutments: A focus on biomechanics and finite element methods
- Smart technologies and textiles and their potential use and application in the care and support of elderly individuals: A systematic review
- Reinforcement mechanisms and current research status of silicon carbide whisker-reinforced composites: A comprehensive review
- Innovative eco-friendly bio-composites: A comprehensive review of the fabrication, characterization, and applications
- Review on geopolymer concrete incorporating Alccofine-1203
- Advancements in surface treatments for aluminum alloys in sports equipment
- Ionic liquid-modified carbon-based fillers and their polymer composites – A Raman spectroscopy analysis
- Emerging boron nitride nanosheets: A review on synthesis, corrosion resistance coatings, and their impacts on the environment and health
- Mechanism, models, and influence of heterogeneous factors of the microarc oxidation process: A comprehensive review
- Synthesizing sustainable construction paradigms: A comprehensive review and bibliometric analysis of granite waste powder utilization and moisture correction in concrete
- 10.1515/rams-2025-0086
- Research Articles
- Coverage and reliability improvement of copper metallization layer in through hole at BGA area during load board manufacture
- Study on dynamic response of cushion layer-reinforced concrete slab under rockfall impact based on smoothed particle hydrodynamics and finite-element method coupling
- Study on the mechanical properties and microstructure of recycled brick aggregate concrete with waste fiber
- Multiscale characterization of the UV aging resistance and mechanism of light stabilizer-modified asphalt
- Characterization of sandwich materials – Nomex-Aramid carbon fiber performances under mechanical loadings: Nonlinear FE and convergence studies
- Effect of grain boundary segregation and oxygen vacancy annihilation on aging resistance of cobalt oxide-doped 3Y-TZP ceramics for biomedical applications
- Mechanical damage mechanism investigation on CFRP strengthened recycled red brick concrete
- Finite element analysis of deterioration of axial compression behavior of corroded steel-reinforced concrete middle-length columns
- Grinding force model for ultrasonic assisted grinding of γ-TiAl intermetallic compounds and experimental validation
- Enhancement of hardness and wear strength of pure Cu and Cu–TiO2 composites via a friction stir process while maintaining electrical resistivity
- Effect of sand–precursor ratio on mechanical properties and durability of geopolymer mortar with manufactured sand
- Research on the strength prediction for pervious concrete based on design porosity and water-to-cement ratio
- Development of a new damping ratio prediction model for recycled aggregate concrete: Incorporating modified admixtures and carbonation effects
- Exploring the viability of AI-aided genetic algorithms in estimating the crack repair rate of self-healing concrete
- Modification of methacrylate bone cement with eugenol – A new material with antibacterial properties
- Numerical investigations on constitutive model parameters of HRB400 and HTRB600 steel bars based on tensile and fatigue tests
- Research progress on Fe3+-activated near-infrared phosphor
- Discrete element simulation study on effects of grain preferred orientation on micro-cracking and macro-mechanical behavior of crystalline rocks
- Ultrasonic resonance evaluation method for deep interfacial debonding defects of multilayer adhesive bonded materials
- Effect of impurity components in titanium gypsum on the setting time and mechanical properties of gypsum-slag cementitious materials
- Bending energy absorption performance of composite fender piles with different winding angles
- Theoretical study of the effect of orientations and fibre volume on the thermal insulation capability of reinforced polymer composites
- Synthesis and characterization of a novel ternary magnetic composite for the enhanced adsorption capacity to remove organic dyes
- Couple effects of multi-impact damage and CAI capability on NCF composites
- Mechanical testing and engineering applicability analysis of SAP concrete used in buffer layer design for tunnels in active fault zones
- Investigating the rheological characteristics of alkali-activated concrete using contemporary artificial intelligence approaches
- Integrating micro- and nanowaste glass with waste foundry sand in ultra-high-performance concrete to enhance material performance and sustainability
- Effect of water immersion on shear strength of epoxy adhesive filled with graphene nanoplatelets
- Impact of carbon content on the phase structure and mechanical properties of TiBCN coatings via direct current magnetron sputtering
- Investigating the anti-aging properties of asphalt modified with polyphosphoric acid and tire pyrolysis oil
- Biomedical and therapeutic potential of marine-derived Pseudomonas sp. strain AHG22 exopolysaccharide: A novel bioactive microbial metabolite
- Effect of basalt fiber length on the behavior of natural hydraulic lime-based mortars
- Optimizing the performance of TPCB/SCA composite-modified asphalt using improved response surface methodology
- Compressive strength of waste-derived cementitious composites using machine learning
- Melting phenomenon of thermally stratified MHD Powell–Eyring nanofluid with variable porosity past a stretching Riga plate
- Development and characterization of a coaxial strain-sensing cable integrated steel strand for wide-range stress monitoring
- Compressive and tensile strength estimation of sustainable geopolymer concrete using contemporary boosting ensemble techniques
- Customized 3D printed porous titanium scaffolds with nanotubes loading antibacterial drugs for bone tissue engineering
- Facile design of PTFE-kaolin-based ternary nanocomposite as a hydrophobic and high corrosion-barrier coating
- Effects of C and heat treatment on microstructure, mechanical, and tribo-corrosion properties of VAlTiMoSi high-entropy alloy coating
- Study on the damage mechanism and evolution model of preloaded sandstone subjected to freezing–thawing action based on the NMR technology
- Promoting low carbon construction using alkali-activated materials: A modeling study for strength prediction and feature interaction
- Entropy generation analysis of MHD convection flow of hybrid nanofluid in a wavy enclosure with heat generation and thermal radiation
- Friction stir welding of dissimilar Al–Mg alloys for aerospace applications: Prospects and future potential
- Fe nanoparticle-functionalized ordered mesoporous carbon with tailored mesostructures and their applications in magnetic removal of Ag(i)
- Study on physical and mechanical properties of complex-phase conductive fiber cementitious materials
- Evaluating the strength loss and the effectiveness of glass and eggshell powder for cement mortar under acidic conditions
- Effect of fly ash on properties and hydration of calcium sulphoaluminate cement-based materials with high water content
- Analyzing the efficacy of waste marble and glass powder for the compressive strength of self-compacting concrete using machine learning strategies
- Experimental study on municipal solid waste incineration ash micro-powder as concrete admixture
- Parameter optimization for ultrasonic-assisted grinding of γ-TiAl intermetallics: A gray relational analysis approach with surface integrity evaluation
- Producing sustainable binding materials using marble waste blended with fly ash and rice husk ash for building materials
- Effect of steam curing system on compressive strength of recycled aggregate concrete
- A sawtooth constitutive model describing strain hardening and multiple cracking of ECC under uniaxial tension
- Predicting mechanical properties of sustainable green concrete using novel machine learning: Stacking and gene expression programming
- Toward sustainability: Integrating experimental study and data-driven modeling for eco-friendly paver blocks containing plastic waste
- A numerical analysis of the rotational flow of a hybrid nanofluid past a unidirectional extending surface with velocity and thermal slip conditions
- A magnetohydrodynamic flow of a water-based hybrid nanofluid past a convectively heated rotating disk surface: A passive control of nanoparticles
- Prediction of flexural strength of concrete with eggshell and glass powders: Advanced cutting-edge approach for sustainable materials
- Efficacy of sustainable cementitious materials on concrete porosity for enhancing the durability of building materials
- Phase and microstructural characterization of swat soapstone (Mg3Si4O10(OH)2)
- Effect of waste crab shell powder on matrix asphalt
- Improving effect and mechanism on service performance of asphalt binder modified by PW polymer
- Influence of pH on the synthesis of carbon spheres and the application of carbon sphere-based solid catalysts in esterification
- Experimenting the compressive performance of low-carbon alkali-activated materials using advanced modeling techniques
- Thermogravimetric (TG/DTG) characterization of cold-pressed oil blends and Saccharomyces cerevisiae-based microcapsules obtained with them
- Investigation of temperature effect on thermo-mechanical property of carbon fiber/PEEK composites
- Computational approaches for structural analysis of wood specimens
- Integrated structure–function design of 3D-printed porous polydimethylsiloxane for superhydrophobic engineering
- Exploring the impact of seashell powder and nano-silica on ultra-high-performance self-curing concrete: Insights into mechanical strength, durability, and high-temperature resilience
- Axial compression damage constitutive model and damage characteristics of fly ash/silica fume modified magnesium phosphate cement after being treated at different temperatures
- Integrating testing and modeling methods to examine the feasibility of blended waste materials for the compressive strength of rubberized mortar
- Special Issue on 3D and 4D Printing of Advanced Functional Materials - Part II
- Energy absorption of gradient triply periodic minimal surface structure manufactured by stereolithography
- Marine polymers in tissue bioprinting: Current achievements and challenges
- Quick insight into the dynamic dimensions of 4D printing in polymeric composite mechanics
- Recent advances in 4D printing of hydrogels
- Mechanically sustainable and primary recycled thermo-responsive ABS–PLA polymer composites for 4D printing applications: Fabrication and studies
- Special Issue on Materials and Technologies for Low-carbon Biomass Processing and Upgrading
- Low-carbon embodied alkali-activated materials for sustainable construction: A comparative study of single and ensemble learners
- Study on bending performance of prefabricated glulam-cross laminated timber composite floor
- Special Issue on Recent Advancement in Low-carbon Cement-based Materials - Part I
- Supplementary cementitious materials-based concrete porosity estimation using modeling approaches: A comparative study of GEP and MEP
- Modeling the strength parameters of agro waste-derived geopolymer concrete using advanced machine intelligence techniques
- Promoting the sustainable construction: A scientometric review on the utilization of waste glass in concrete
- Incorporating geranium plant waste into ultra-high performance concrete prepared with crumb rubber as fine aggregate in the presence of polypropylene fibers
- Investigation of nano-basic oxygen furnace slag and nano-banded iron formation on properties of high-performance geopolymer concrete
- Effect of incorporating ultrafine palm oil fuel ash on the resistance to corrosion of steel bars embedded in high-strength green concrete
- Influence of nanomaterials on properties and durability of ultra-high-performance geopolymer concrete
- Influence of palm oil ash and palm oil clinker on the properties of lightweight concrete
Articles in the same Issue
- Review Articles
- Effect of superplasticizer in geopolymer and alkali-activated cement mortar/concrete: A review
- Experimenting the influence of corncob ash on the mechanical strength of slag-based geopolymer concrete
- Powder metallurgy processing of high entropy alloys: Bibliometric analysis and systematic review
- Exploring the potential of agricultural waste as an additive in ultra-high-performance concrete for sustainable construction: A comprehensive review
- A review on partial substitution of nanosilica in concrete
- Foam concrete for lightweight construction applications: A comprehensive review of the research development and material characteristics
- Modification of PEEK for implants: Strategies to improve mechanical, antibacterial, and osteogenic properties
- Interfacing the IoT in composite manufacturing: An overview
- Advances in processing and ablation properties of carbon fiber reinforced ultra-high temperature ceramic composites
- Advancing auxetic materials: Emerging development and innovative applications
- Revolutionizing energy harvesting: A comprehensive review of thermoelectric devices
- Exploring polyetheretherketone in dental implants and abutments: A focus on biomechanics and finite element methods
- Smart technologies and textiles and their potential use and application in the care and support of elderly individuals: A systematic review
- Reinforcement mechanisms and current research status of silicon carbide whisker-reinforced composites: A comprehensive review
- Innovative eco-friendly bio-composites: A comprehensive review of the fabrication, characterization, and applications
- Review on geopolymer concrete incorporating Alccofine-1203
- Advancements in surface treatments for aluminum alloys in sports equipment
- Ionic liquid-modified carbon-based fillers and their polymer composites – A Raman spectroscopy analysis
- Emerging boron nitride nanosheets: A review on synthesis, corrosion resistance coatings, and their impacts on the environment and health
- Mechanism, models, and influence of heterogeneous factors of the microarc oxidation process: A comprehensive review
- Synthesizing sustainable construction paradigms: A comprehensive review and bibliometric analysis of granite waste powder utilization and moisture correction in concrete
- 10.1515/rams-2025-0086
- Research Articles
- Coverage and reliability improvement of copper metallization layer in through hole at BGA area during load board manufacture
- Study on dynamic response of cushion layer-reinforced concrete slab under rockfall impact based on smoothed particle hydrodynamics and finite-element method coupling
- Study on the mechanical properties and microstructure of recycled brick aggregate concrete with waste fiber
- Multiscale characterization of the UV aging resistance and mechanism of light stabilizer-modified asphalt
- Characterization of sandwich materials – Nomex-Aramid carbon fiber performances under mechanical loadings: Nonlinear FE and convergence studies
- Effect of grain boundary segregation and oxygen vacancy annihilation on aging resistance of cobalt oxide-doped 3Y-TZP ceramics for biomedical applications
- Mechanical damage mechanism investigation on CFRP strengthened recycled red brick concrete
- Finite element analysis of deterioration of axial compression behavior of corroded steel-reinforced concrete middle-length columns
- Grinding force model for ultrasonic assisted grinding of γ-TiAl intermetallic compounds and experimental validation
- Enhancement of hardness and wear strength of pure Cu and Cu–TiO2 composites via a friction stir process while maintaining electrical resistivity
- Effect of sand–precursor ratio on mechanical properties and durability of geopolymer mortar with manufactured sand
- Research on the strength prediction for pervious concrete based on design porosity and water-to-cement ratio
- Development of a new damping ratio prediction model for recycled aggregate concrete: Incorporating modified admixtures and carbonation effects
- Exploring the viability of AI-aided genetic algorithms in estimating the crack repair rate of self-healing concrete
- Modification of methacrylate bone cement with eugenol – A new material with antibacterial properties
- Numerical investigations on constitutive model parameters of HRB400 and HTRB600 steel bars based on tensile and fatigue tests
- Research progress on Fe3+-activated near-infrared phosphor
- Discrete element simulation study on effects of grain preferred orientation on micro-cracking and macro-mechanical behavior of crystalline rocks
- Ultrasonic resonance evaluation method for deep interfacial debonding defects of multilayer adhesive bonded materials
- Effect of impurity components in titanium gypsum on the setting time and mechanical properties of gypsum-slag cementitious materials
- Bending energy absorption performance of composite fender piles with different winding angles
- Theoretical study of the effect of orientations and fibre volume on the thermal insulation capability of reinforced polymer composites
- Synthesis and characterization of a novel ternary magnetic composite for the enhanced adsorption capacity to remove organic dyes
- Couple effects of multi-impact damage and CAI capability on NCF composites
- Mechanical testing and engineering applicability analysis of SAP concrete used in buffer layer design for tunnels in active fault zones
- Investigating the rheological characteristics of alkali-activated concrete using contemporary artificial intelligence approaches
- Integrating micro- and nanowaste glass with waste foundry sand in ultra-high-performance concrete to enhance material performance and sustainability
- Effect of water immersion on shear strength of epoxy adhesive filled with graphene nanoplatelets
- Impact of carbon content on the phase structure and mechanical properties of TiBCN coatings via direct current magnetron sputtering
- Investigating the anti-aging properties of asphalt modified with polyphosphoric acid and tire pyrolysis oil
- Biomedical and therapeutic potential of marine-derived Pseudomonas sp. strain AHG22 exopolysaccharide: A novel bioactive microbial metabolite
- Effect of basalt fiber length on the behavior of natural hydraulic lime-based mortars
- Optimizing the performance of TPCB/SCA composite-modified asphalt using improved response surface methodology
- Compressive strength of waste-derived cementitious composites using machine learning
- Melting phenomenon of thermally stratified MHD Powell–Eyring nanofluid with variable porosity past a stretching Riga plate
- Development and characterization of a coaxial strain-sensing cable integrated steel strand for wide-range stress monitoring
- Compressive and tensile strength estimation of sustainable geopolymer concrete using contemporary boosting ensemble techniques
- Customized 3D printed porous titanium scaffolds with nanotubes loading antibacterial drugs for bone tissue engineering
- Facile design of PTFE-kaolin-based ternary nanocomposite as a hydrophobic and high corrosion-barrier coating
- Effects of C and heat treatment on microstructure, mechanical, and tribo-corrosion properties of VAlTiMoSi high-entropy alloy coating
- Study on the damage mechanism and evolution model of preloaded sandstone subjected to freezing–thawing action based on the NMR technology
- Promoting low carbon construction using alkali-activated materials: A modeling study for strength prediction and feature interaction
- Entropy generation analysis of MHD convection flow of hybrid nanofluid in a wavy enclosure with heat generation and thermal radiation
- Friction stir welding of dissimilar Al–Mg alloys for aerospace applications: Prospects and future potential
- Fe nanoparticle-functionalized ordered mesoporous carbon with tailored mesostructures and their applications in magnetic removal of Ag(i)
- Study on physical and mechanical properties of complex-phase conductive fiber cementitious materials
- Evaluating the strength loss and the effectiveness of glass and eggshell powder for cement mortar under acidic conditions
- Effect of fly ash on properties and hydration of calcium sulphoaluminate cement-based materials with high water content
- Analyzing the efficacy of waste marble and glass powder for the compressive strength of self-compacting concrete using machine learning strategies
- Experimental study on municipal solid waste incineration ash micro-powder as concrete admixture
- Parameter optimization for ultrasonic-assisted grinding of γ-TiAl intermetallics: A gray relational analysis approach with surface integrity evaluation
- Producing sustainable binding materials using marble waste blended with fly ash and rice husk ash for building materials
- Effect of steam curing system on compressive strength of recycled aggregate concrete
- A sawtooth constitutive model describing strain hardening and multiple cracking of ECC under uniaxial tension
- Predicting mechanical properties of sustainable green concrete using novel machine learning: Stacking and gene expression programming
- Toward sustainability: Integrating experimental study and data-driven modeling for eco-friendly paver blocks containing plastic waste
- A numerical analysis of the rotational flow of a hybrid nanofluid past a unidirectional extending surface with velocity and thermal slip conditions
- A magnetohydrodynamic flow of a water-based hybrid nanofluid past a convectively heated rotating disk surface: A passive control of nanoparticles
- Prediction of flexural strength of concrete with eggshell and glass powders: Advanced cutting-edge approach for sustainable materials
- Efficacy of sustainable cementitious materials on concrete porosity for enhancing the durability of building materials
- Phase and microstructural characterization of swat soapstone (Mg3Si4O10(OH)2)
- Effect of waste crab shell powder on matrix asphalt
- Improving effect and mechanism on service performance of asphalt binder modified by PW polymer
- Influence of pH on the synthesis of carbon spheres and the application of carbon sphere-based solid catalysts in esterification
- Experimenting the compressive performance of low-carbon alkali-activated materials using advanced modeling techniques
- Thermogravimetric (TG/DTG) characterization of cold-pressed oil blends and Saccharomyces cerevisiae-based microcapsules obtained with them
- Investigation of temperature effect on thermo-mechanical property of carbon fiber/PEEK composites
- Computational approaches for structural analysis of wood specimens
- Integrated structure–function design of 3D-printed porous polydimethylsiloxane for superhydrophobic engineering
- Exploring the impact of seashell powder and nano-silica on ultra-high-performance self-curing concrete: Insights into mechanical strength, durability, and high-temperature resilience
- Axial compression damage constitutive model and damage characteristics of fly ash/silica fume modified magnesium phosphate cement after being treated at different temperatures
- Integrating testing and modeling methods to examine the feasibility of blended waste materials for the compressive strength of rubberized mortar
- Special Issue on 3D and 4D Printing of Advanced Functional Materials - Part II
- Energy absorption of gradient triply periodic minimal surface structure manufactured by stereolithography
- Marine polymers in tissue bioprinting: Current achievements and challenges
- Quick insight into the dynamic dimensions of 4D printing in polymeric composite mechanics
- Recent advances in 4D printing of hydrogels
- Mechanically sustainable and primary recycled thermo-responsive ABS–PLA polymer composites for 4D printing applications: Fabrication and studies
- Special Issue on Materials and Technologies for Low-carbon Biomass Processing and Upgrading
- Low-carbon embodied alkali-activated materials for sustainable construction: A comparative study of single and ensemble learners
- Study on bending performance of prefabricated glulam-cross laminated timber composite floor
- Special Issue on Recent Advancement in Low-carbon Cement-based Materials - Part I
- Supplementary cementitious materials-based concrete porosity estimation using modeling approaches: A comparative study of GEP and MEP
- Modeling the strength parameters of agro waste-derived geopolymer concrete using advanced machine intelligence techniques
- Promoting the sustainable construction: A scientometric review on the utilization of waste glass in concrete
- Incorporating geranium plant waste into ultra-high performance concrete prepared with crumb rubber as fine aggregate in the presence of polypropylene fibers
- Investigation of nano-basic oxygen furnace slag and nano-banded iron formation on properties of high-performance geopolymer concrete
- Effect of incorporating ultrafine palm oil fuel ash on the resistance to corrosion of steel bars embedded in high-strength green concrete
- Influence of nanomaterials on properties and durability of ultra-high-performance geopolymer concrete
- Influence of palm oil ash and palm oil clinker on the properties of lightweight concrete