Article Open Access

Mechanical characterizations of braided composite stents made of helical polyethylene terephthalate strips and NiTi wires

  • Qingli Zheng , Pengfei Dong , Zhiqiang Li , Xinwei Han , Changchun Zhou , Meiwen An EMAIL logo and Linxia Gu EMAIL logo
Published/Copyright: November 6, 2019
Become an author with De Gruyter Brill

Abstract

The novel braided composite stent (BCS), woven with both nitinol wires and polyethylene terephthalate (PET) strips, were characterized and compared with the braided nitinol stent in the same weaving pattern. Finite element models simulating the stent compression and bending were developed to quantify its radial strength and longitudinal flexibility. The interaction between the nitinol wires and the PET strips were also delineated. Results showed that the PET strips enforced more constrains on the BCS and thus enhance its radial strength especially at a larger compression load. The longitudinal flexibility of the BCS was less sensitive to the presence of the PET strips. This work suggested that the novel design of the BCS could acquire the advantage of a covered stent without compromising its mechanical performance. The fundamental understanding of the braided composite stent will facilitate a better device design.

1 Introduction

Braided nitinol stents (BNS) are increasingly popular in treating arterial occlusions, especially peripheral arterial disease [1]. Compared to the laser-cut stents, the BNS tend to have a reduced radial strength and better flexibility [2]. The radial strength of the stent was essential to keep the narrowed vessel open [3]. The flexibility of the stent was desired to move along or scaffold a curved vessel. Despite the demonstrated acute outcomes of a stented vessel, the long-term patency of the BNS remains a challenge due to tissue ingrowth through the mesh structure of stents [4].

Covered stents, i.e., stents covered by a fabric or graft material such as polytetrafluoroethylene or polyethylene terephthalate (PET), were designed to mitigate the tissue ingrowth as well as to seal the vessel perforation [5]. However, compared to a bare metallic stent, the covered stent is harder to maneuver through tortuous anatomy without the risk of kinking and migration [6, 7]. This led to undesired vessel patency as well as tissue damage and increase incidence of limb thrombosis [8]. Recently a braided composite stent (BCS) woven using PET strips and nitinol (NiTi) wires was proposed to improve the radial strength as well as the flexibility of a BNS [9]. Considering the minimal porosity of the BCS, it drove us to consider the BCS as a potential alternative to the covered stent.

Finite element method (FEM) has proven to be an effective tool for the characterization of the stent performance and its underlying mechanism. It could visualize the mechanics virtually anywhere in the computational domain, a feature that is not affected by experimental techniques. Kim et al. developed a 3D model to quantify the mechanical behavior of braided stents validated by experiments

[10]. The FEM results from Zhao et al. showed that the braided stent had a relative small radial strength than the welded or laser cut ones [2]. De Beule et al. adopted a virtual optimization algorithm to design a braided stent with less foreshortening while maintaining the radial stiffness [11]. However, no FEM studies were related to the BCS.

In this work, the mechanical behaviors of the BCS were investigated through FEM. Three-dimensional models of the BCS and the corresponding BNS in the same weaving pattern were developed to evaluate and compare their radial strength and flexibility. The BCS model was validated by the published experimental results [9]. The effects of PET strips as well as the interaction between PET strips and NiTi wires were delineated. Quantitative evaluation of the mechanical performance of the BCS and its underlying mechanisms could facilitate an optimal design of the stent.

2 Materials and methods

Three-dimensional models of the BCS as well as its counterpart BNS were constructed as shown in Figure 1. Both BCS and BNS have an outer diameter of 7 mm, length of 15mm, and a braided angle β as 65. The braided angle is the angle between the helical wire and the axial direction of the stent. The BNS was braided by 32 strands of helical NiTi wires with a wire diameter of 0.2 mm. The BCS was braided by 8 NiTi wires and 24 PET strips. The thickness of the PET strip was 0.12mm [12]. The width of the PET strip (w) was 0.57 mm, which could be calculated as following:

Figure 1 Three-dimensional models of (a) braided composite stent and (b) braided nitinol stents, (c) compression test, (d) bending test.
Figure 1

Three-dimensional models of (a) braided composite stent and (b) braided nitinol stents, (c) compression test, (d) bending test.

(1)w=Lcos(β)=πD16cos(β)

where L is the distance of the two crossover points along the circumferential direction (Figure 1a).

The superelastic constitutive model of NiTi wires was implemented through a built-in Abaqus User Material Subroutine (VUMAT) and the material parameters were adopted from literature [13]. The PET strip was characterized by an elastoplastic material model [14]. Both material parameters were summarized in Table 1. Following the mesh-sensitivity study, each NiTi wire was meshed with 240 two-node linear beam elements (B31). The PET strips were meshed with 23,040 reduced 4-node doubly curved shell elements (S4R).

Table 1

Material properties of nitinol and polyethylene terephthalate.

Nitinol
Austenite elasticity EA (MPa)50000
Martensite elasticity EM (MPa)37000
Start of transformation loading σS L (MPa)400
End of transformation loading σE L (MPa)650
Start of transformation unloading σS U (MPa)350
End of transformation unloading σE U (MPa)80
Volumetric transformation strain ϵL V0.055
Polyethylene terephthalate (PET)
Young’s modulus E (GPa)3.5
Poisson ratio ν0.4
Yield stress σs (MPa)60
Tensile stress σt (MPa)70

Compression tests were conducted in silico to quantify the radial strength of the stent (Figure 1c). Specifically, the middle section of the stent was vertically compressed by a rigid cylinder of 5 mm in diameter. The displacement control of 3.5 mm, i.e., 50% of the outer diameter of the stent, was enforced on the cylinder [15]. The reaction force of the cylinder, i.e., load applied on the stent, positively correlated with the radial strength of the stent, was monitored during the load and unload process.

The stent flexibility was quantified by the pure bending tests (Figure 1d). Both ends of the stent were fixed to a rigid plane with rotations around the X axis in opposite directions until the angle between two end surfaces reached at 60 [16]. The corresponding load, i.e., bending moment, was recorded.

A general contact algorithm was considered among all contact surfaces with friction coefficients of 0.3. The kinetic energy of each stent was kept under 5% of the internal energy to avoid the inertia effect. Abaqus/Explicit solver (Dassault Systèmes Simulia Corp.) was used due to its demonstrated capacity for complex contact problems.

3 Results

The radial strength of the BCS was compared with the published data as illustrated in Figure 2. The percentage deformation was estimated with reference to the original outer diameter of the stent. It is clear that our computational results matched well with the experiment (Model B2) [9].

Figure 2 Simulation of the braided composite stent compression in compared to experimental results.
Figure 2

Simulation of the braided composite stent compression in compared to experimental results.

The compression induced stress and strain distributions in both BCS and BNS were shown in Figure 3. The peak von Mises stress occurred on the nitinol wires for both stents. It was 581MPa and 600MPa for the BCS and BNS, respectively. Both values fall in the range of transformation loading (400 to 650 MPa) as defined in Table 1. It indicated that the deformation of nitinol wires reached at the plateau of phase transformation between austenite and martensite. The maximum tensile strain (SDV24) of the nitinol wires was 7.06% in BCS and 7.12% in BNS. Both were below the elastic limit of 8%, indicating that nitinol wires could fully recover back to its original configuration following the compression. The peak von Mises stress and strain in the BCS strips was 60.6 Mpa and 11.8%, respectively. In addition, the stress or strain distribution patterns of BCS were different with the ones of BNS. Specifically, the stress and strain of the BNS concentrated beneath the cylinder surface (Figure 3c), while the stress and strain of the BCS concentrated around its lateral region (Figure 3a&b). The braiding angle at the stress concentration area of the BCS decreases from the original 65 to 44.9 while it decreases further to 35.2 for the BNS (Figure 3a & c). This indicated the wire rotation and slippage during the loading.

Figure 3 Compression induced stress (left) and strain (right) distributions for (a) nitinol wires in braided composite stent; (b) PET strips of braided composite stent; and (c) braided nitinol stents.
Figure 3

Compression induced stress (left) and strain (right) distributions for (a) nitinol wires in braided composite stent; (b) PET strips of braided composite stent; and (c) braided nitinol stents.

The load-deformation diagrams of both BCS and BNS were depicted in Figure 4. It was clear that the BCS and BNS demonstrated a similar compressive load at lower deformation, i.e., less than 20%. Thereafter, the compressive load in BNS exhibited a plateau at approximately 4.2 N, while the compressive load of the BCS continually increased up to 6.4 N.

Figure 4 The radial compression of braided composite stent and braided nitinol stent
Figure 4

The radial compression of braided composite stent and braided nitinol stent

The slippage was estimated by monitoring the relative displacement at three representative wire intersection locations (A, B, and C, from center to the end of stent, as labeled in Figure 1). The relative displacements of these three locations subject to compression were depicted in Figure 5. It was clear that the stent deformation positively correlated with the relative deformation of wires. At the compression deformation of 50%, the relative displacementswere 0.19mm,0.34mmand 0.16mm for the location A, B and C of the BCS, respectively. For the BNS, they were

Figure 5 Relative displacement of two points between intersections of NITI wires of braided composite stent and braided nitinol stent under radial compression.
Figure 5

Relative displacement of two points between intersections of NITI wires of braided composite stent and braided nitinol stent under radial compression.

0.13 mm, 0.56mm and 0.62 mm, respectively. The results showed that the BCS showed a more uniform and smaller relative displacement than the BNS, the larger relative displacement in BNS may contribute to the unstable radial force at the larger deformation.

The stress distribution of the stents during the bending test were shown in Figure 6. At the bending angle of 37, the peak value of von Mises stress in BCS was 294MPa, which was larger than that in the BNS of 167MPa, the same trend was found at the bending angle of 60, which was 404 MPa in BCS and 260 MPa in BNS. The peak value of von Mises stress in two stents were both located in the nitinol wires. The bending moment for each stent during the bending test was shown in Figure 7. The BCS demonstrated a relative larger bending moment than the BNS at a bending angle larger than 37. At a bending angle of 60, the bending moment of the BCS was 24% larger than that of the BNS.

Figure 6 Bending behaviors of braided composite stent and braided nitinol stent at the rotation angle of 37∘ (left) and 60∘ (right).
Figure 6

Bending behaviors of braided composite stent and braided nitinol stent at the rotation angle of 37 (left) and 60 (right).

Figure 7 The bending behavior of braided composite stent and braided nitinol stent
Figure 7

The bending behavior of braided composite stent and braided nitinol stent

4 Discussions

In this work, the design of BCS was systematically characterized in terms of radial strength and longitudinal flexibility. The mechanical behaviors of the BCS were compared with the BNS with the same weaving pattern. The major difference lies in the replacement of 24 strands of NiTi wires by the wider PET strips. The BCS possessed the advantage of covered stent without compromising their mechanical performance. The BCS provided a direct barrier to the tissue underneath.Moreover, the radial strength of the BCS was larger than the BNS especially at the larger radial deformation. The bending moment was smaller than the BNS when the bending angle was less than 37 and much less than the reported ones of the covered stent. When the polyurethane or silicone covered stents were subjected to a bending angle of 30, the required forces were 11% and 44%, respectively, more than the BNS [17]. While at a bending angle of 30, the bending moment of the BCS was 9.3% smaller than that of the BNS.

The radial strength of the stent, i.e., the capability to scaffold the stenotic lumen, is the first consideration of the stent conceptual design [18, 19]. The BCS and BNS showed the same radial strength when the deformation is less than 20%. This seems different from the documented experiment [9], which stated that the radial strength of the BCS was much larger than the BNS. This could be explained by the different braided angle for the BCS and BNS. Specifically, it was 65 for BCS and 30 for BNS. The braided stent with a larger braided angle resulted in a higher radial strength [10]. As the deformation exceeded 20%, the radial strength of the BNS reached a plaque of 4.2 N, while the BCS kept increasing up to 6.4N. The different radial strength of the BCS and BNS at a large deformation could be attributed to the interaction between the PET strips and NiTi wires. In BNS, there existed the relative rotation and slippage between NiTi wires. However, the presence of wide PET stripes in BCS constrained the relative rotation and slippage of the NiTi wires. Our compression tests have shown that the alteration in the braided angle of the BNS was 48% larger than that of the BCS. This was also reflected in the relative larger displacement of the BNS than that of the BCS. A large relative movement in the woven wires was associate with the smaller radial strength and a larger longitudinal flexibility.

The longitudinal flexibility was required for a smooth maneuver through torturous arteries as well as implant inside a curved artery. Enhanced flexibility of stent was also associated with the decreased risk of kinking and incidence of limb thrombosis in tortuous anatomy [6, 7]. The BCS desired a larger bending moment when the bending angle was larger than 37. This could be explained by the larger radial strength of the BCS [9] for larger deformation.

In the present work, we focused on the role of PET strips and the interaction between the PET strips and the NiTi wires on the novel BCS performance. It should be noted that this work does not completely reflect the optimal attributes of the BCS stent, which need consider more parameters, such as the mass fraction of the wires, the braided angle, diameter of the NiTi wires, strip thickness and material properties, etc. The interaction between stent and a torturous artery could provide more complete information on the assessment of this novel stent design. Despite these limitations, the present work demonstrated the potential of the BCS as a substitute for a covered stent with improved mechanical behaviors, which might have significant clinical implications for in-stent restenosis [20].


Qingli Zheng and Pengfei Dong contributed equally to this paper

Acknowledgement

The support from the National Science Foundation CAREER award under grant No. 1254095 and the National Natural Science Foundation of China (No. 11372208, No. 11672199) was acknowledged.

References

[1] Leon L.R. Jr, Dieter R.S., Gadd C.L., Ranellone E., Mills J.L. Sr, Montero-Baker M.F., Gruessner A.C., Pacanowski J.P. Jr, Preliminary results of the initial United States experience with the Supera woven nitinol stent in the popliteal artery, J. Vasc. Surg., 2013, 57, 1014-1022.10.1016/j.jvs.2012.10.093Search in Google Scholar

[2] Zhao S., Liu X.C., Gu L., The impact of wire stent fabrication technique on the performance of stent placement, J. Med. Dev., 2012, 6, 011007.10.1115/1.4005788Search in Google Scholar

[3] Watson T., Webster M.W., Ormiston J.A., Ruygrok P.N., Stewart J.T., Long and short of optimal stent design, Open Heart, 2017, 4, e000680.10.1136/openhrt-2017-000680Search in Google Scholar

[4] Lee J.H., Krishna S.G., Singh A., Ladha H.S., Slack R.S., Ramireddy S., Raju G.S., Davila M., RossW.A., Comparison of the utility of covered metal stents versus uncovered metal stents in the management of malignant biliary strictures in 749 patients, Gastrointest. Endosc., 2013, 78, 312-324.10.1016/j.gie.2013.02.032Search in Google Scholar

[5] Al-Mukhaini M., Panduranga P., Sulaiman K., Riyami A.A., Deeb M., Riyami M.B., Coronary perforation and covered stents: an update and review, Heart views: Offic. J. Gulf Heart Assoc., 2011, 12, 63.10.4103/1995-705X.86017Search in Google Scholar

[6] Rieder E., Dunst C.,Martinec D., Cassera M., Swanstrom L., Endoscopic suture fixation of gastrointestinal stents: proof of biomechanical principles and early clinical experience, Endoscopy, 2012, 44, 1121-1126.10.1055/s-0032-1325730Search in Google Scholar

[7] Carroccio A., Faries P.L., Morrissey N.J., Teodorescu V., Burks J.A., Gravereaux E.C., Hollier L.H., Marin M.L., Predicting iliac limb occlusions after bifurcated aortic stent grafting: anatomic and device-related causes, J. Vasc. Surg., 2002, 36, 679-684.10.1016/S0741-5214(02)00117-9Search in Google Scholar

[8] Weale A., Balasubramaniam K., Hardman J., Horrocks M., Use of the Aorfix stent graft in patients with tortuous iliac anatomy, J. Cardiovasc. Surg., 2010, 51, 461-466.Search in Google Scholar

[9] Zou Q., Xue W., Lin J., Fu Y., Guan G.,Wang F.,Wang L., Mechanical characteristics of novel polyester/NiTi wires braided composite stent for the medical application, Results Phys., 2016, 6, 440-446.10.1016/j.rinp.2016.07.007Search in Google Scholar

[10] Kim J.H., Kang T.J., Yu W.-R., Mechanical modeling of selfexpandable stent fabricated using braiding technology, J. Biomech., 2008, 41, 3202-3212.10.1016/j.jbiomech.2008.08.005Search in Google Scholar PubMed

[11] De Beule M., Van Cauter S., Mortier P., Van Loo D., Van Impe R., Verdonck P., Verhegghe B., Virtual optimization of selfexpandable braided wire stents, Med. Eng. Phys., 2009, 31, 448-453.10.1016/j.medengphy.2008.11.008Search in Google Scholar PubMed

[12] Guan Y., Lin J., Dong Z., Wang L., Comparative Study of the Effect of Structural Parameters on the Flexibility of Endovascular Stent Grafts, Adv. Mat. Sci. Eng., 2018, 2018.10.1155/2018/3046576Search in Google Scholar

[13] Rebelo N., Gong X.-Y., Hall A., Pelton A.R., Duerig T.W., Finite element analysis on the cyclic properties of superelastic Nitinol, Proc. Int. Conf. Shape Memory Superelast. Technol., 2006, 157-163.Search in Google Scholar

[14] Puskas J.E., Chen Y., Biomedical application of commercial polymers and novel polyisobutylene-based thermoplastic elastomers for soft tissue replacement, Biomacromolecules, 2004, 5, 1141-1154.10.1021/bm034513kSearch in Google Scholar PubMed

[15] Liu G., Hu H., Zhang P., Wang W., Radial compressive properties of the biodegradable braided regeneration tubes for peripheral nerve repair, J. Industr. Textil., 2006, 36, 35-46.10.1177/1528083706064375Search in Google Scholar

[16] Balasubramaniam K., Hardman J., Horrocks M., Bulbulia R., The advantages of Aorfix for endovascular repair of abdominal aortic aneurysm, J. Cardiovasc. Surg. (Torino), 2009, 50, 139-143.Search in Google Scholar

[17] Isayama H., Nakai Y., Toyokawa Y., Togawa O., Gon C., Ito Y., Yashima Y., Yagioka H., Kogure H., Sasaki T., Measurement of radial and axial forces of biliary self-expandable metallic stents, Gastrointest. Endosc., 2009, 70, 37-44.10.1016/j.gie.2008.09.032Search in Google Scholar PubMed

[18] Matsumoto T., Inoue K., Tanaka S., Aoyagi Y., Matsubara Y., Matsuda D., Yoshiya K., Yoshiga R., Ohkusa T., Maehara Y., Radial forces of stents used in thoracic endovascular aortic repair and bare self-expanding nitinol stents measured ex vivo–Rapid res cue for obstruction of the innominate artery using bare selfexpanding nitinol stents, Vascular, 2017, 25, 36-41.10.1177/1708538116640131Search in Google Scholar PubMed

[19] Voűte M.T., Hendriks J.M., Van Laanen J.H., Pattynama P.M.,Muhs B.E., Poldermans D., Verhagen H.J., Radial force measurements in carotid stents: influence of stent design and length of the lesion, J. Vasc. Interv. Radiol., 2011, 22, 661-666.10.1016/j.jvir.2011.01.450Search in Google Scholar PubMed

[20] Cicha I., Lyer S., Alexiou C., Garlichs C.D., Nanomedicine in diagnostics and therapy of cardiovascular diseases: beyond atherosclerotic plaque imaging, Nanotechnology Reviews, 2013, 2, 449-472.10.1515/ntrev-2013-0009Search in Google Scholar

Received: 2019-03-04
Accepted: 2019-04-11
Published Online: 2019-11-06

© 2019 Q. Zheng et al., published by De Gruyter

This work is licensed under the Creative Commons Attribution 4.0 Public License.

Articles in the same Issue

  1. Research Articles
  2. Investigation of rare earth upconversion fluorescent nanoparticles in biomedical field
  3. Carbon Nanotubes Coated Paper as Current Collectors for Secondary Li-ion Batteries
  4. Insight into the working wavelength of hotspot effects generated by popular nanostructures
  5. Novel Lead-free biocompatible piezoelectric Hydroxyapatite (HA) – BCZT (Ba0.85Ca0.15Zr0.1Ti0.9O3) nanocrystal composites for bone regeneration
  6. Effect of defects on the motion of carbon nanotube thermal actuator
  7. Dynamic mechanical behavior of nano-ZnO reinforced dental composite
  8. Fabrication of Ag Np-coated wetlace nonwoven fabric based on amino-terminated hyperbranched polymer
  9. Fractal analysis of pore structures in graphene oxide-carbon nanotube based cementitious pastes under different ultrasonication
  10. Effect of PVA fiber on durability of cementitious composite containing nano-SiO2
  11. Cr effects on the electrical contact properties of the Al2O3-Cu/15W composites
  12. Experimental evaluation of self-expandable metallic tracheobronchial stents
  13. Experimental study on the existence of nano-scale pores and the evolution of organic matter in organic-rich shale
  14. Mechanical characterizations of braided composite stents made of helical polyethylene terephthalate strips and NiTi wires
  15. Mechanical properties of boron nitride sheet with randomly distributed vacancy defects
  16. Fabrication, mechanical properties and failure mechanism of random and aligned nanofiber membrane with different parameters
  17. Micro- structure and rheological properties of graphene oxide rubber asphalt
  18. First-principles calculations of mechanical and thermodynamic properties of tungsten-based alloy
  19. Adsorption performance of hydrophobic/hydrophilic silica aerogel for low concentration organic pollutant in aqueous solution
  20. Preparation of spherical aminopropyl-functionalized MCM-41 and its application in removal of Pb(II) ion from aqueous solution
  21. Electrical conductivity anisotropy of copper matrix composites reinforced with SiC whiskers
  22. Miniature on-fiber extrinsic Fabry-Perot interferometric vibration sensors based on micro-cantilever beam
  23. Electric-field assisted growth and mechanical bactericidal performance of ZnO nanoarrays with gradient morphologies
  24. Flexural behavior and mechanical model of aluminum alloy mortise-and-tenon T-joints for electric vehicle
  25. Synthesis of nano zirconium oxide and its application in dentistry
  26. Surface modification of nano-sized carbon black for reinforcement of rubber
  27. Temperature-dependent negative Poisson’s ratio of monolayer graphene: Prediction from molecular dynamics simulations
  28. Finite element nonlinear transient modelling of carbon nanotubes reinforced fiber/polymer composite spherical shells with a cutout
  29. Preparation of low-permittivity K2O–B2O3–SiO2–Al2O3 composites without the addition of glass
  30. Large amplitude vibration of doubly curved FG-GRC laminated panels in thermal environments
  31. Enhanced flexural properties of aramid fiber/epoxy composites by graphene oxide
  32. Correlation between electrochemical performance degradation and catalyst structural parameters on polymer electrolyte membrane fuel cell
  33. Materials characterization of advanced fillers for composites engineering applications
  34. Humic acid assisted stabilization of dispersed single-walled carbon nanotubes in cementitious composites
  35. Test on axial compression performance of nano-silica concrete-filled angle steel reinforced GFRP tubular column
  36. Multi-scale modeling of the lamellar unit of arterial media
  37. The multiscale enhancement of mechanical properties of 3D MWK composites via poly(oxypropylene) diamines and GO nanoparticles
  38. Mechanical properties of circular nano-silica concrete filled stainless steel tube stub columns after being exposed to freezing and thawing
  39. Arc erosion behavior of TiB2/Cu composites with single-scale and dual-scale TiB2 particles
  40. Yb3+-containing chitosan hydrogels induce B-16 melanoma cell anoikis via a Fak-dependent pathway
  41. Template-free synthesis of Se-nanorods-rGO nanocomposite for application in supercapacitors
  42. Effect of graphene oxide on chloride penetration resistance of recycled concrete
  43. Bending resistance of PVA fiber reinforced cementitious composites containing nano-SiO2
  44. Review Articles
  45. Recent development of Supercapacitor Electrode Based on Carbon Materials
  46. Mechanical contribution of vascular smooth muscle cells in the tunica media of artery
  47. Applications of polymer-based nanoparticles in vaccine field
  48. Toxicity of metallic nanoparticles in the central nervous system
  49. Parameter control and concentration analysis of graphene colloids prepared by electric spark discharge method
  50. A critique on multi-jet electrospinning: State of the art and future outlook
  51. Electrospun cellulose acetate nanofibers and Au@AgNPs for antimicrobial activity - A mini review
  52. Recent progress in supercapacitors based on the advanced carbon electrodes
  53. Recent progress in shape memory polymer composites: methods, properties, applications and prospects
  54. In situ capabilities of Small Angle X-ray Scattering
  55. Review of nano-phase effects in high strength and conductivity copper alloys
  56. Progress and challenges in p-type oxide-based thin film transistors
  57. Advanced materials for flexible solar cell applications
  58. Phenylboronic acid-decorated polymeric nanomaterials for advanced bio-application
  59. The effect of nano-SiO2 on concrete properties: a review
  60. A brief review for fluorinated carbon: synthesis, properties and applications
  61. A review on the mechanical properties for thin film and block structure characterised by using nanoscratch test
  62. Cotton fibres functionalized with plasmonic nanoparticles to promote the destruction of harmful molecules: an overview
Downloaded on 17.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/ntrev-2019-0016/html
Scroll to top button