Abstract
The self-expandable metallic stents have been widely used in tracheobronchial obstruction or fistulation, including the J-shaped and Y-shaped stents, named after the shape of the branch-stem junction of the stent. However, there is scarce data on the mechanical performance of these tracheobronchial stents, which is essential for optimal stent implantation. In this work, eight self-expandable metallic tracheobronchial stents in three types (i.e., straight, J-shaped, and Y-shaped), with or without cover, were characterized. The compression resistance of the stems was investigated through both compression and indentation tests. The bending resistance of the branches in the J-shaped and Y-shaped stents was assessed through the bending test. Our results demonstrated that the covered stents exhibited a significantly higher compression resistance and bending resistance than the uncovered ones. The branches had a minimal impact on the compression resistance of the stem. The branch of the J-shaped stent showed a significantly lower bending resistance than the Y shaped one. This work provides a testing framework for the J-shaped and Y-shaped stents,which could shed some light on the optimal design of stent with branches.
1 Introduction
Tracheobronchial obstruction or fistulation is potentially life-threatening and requires immediate therapy to restore the airflow. The traditional Y-shaped silicon stent was used for the central airway obstruction, but was limited by its noncompliant structure [1]. The self-expandable metallic stents, originating from treating the blocked artery, have become widely used for tracheobronchial obstruction and fistulation during the last two decades [2, 3, 4]. Different structures of self-expandable metallic stents were developed: straight stent without a side branch [5], J-shaped stent with one branch [6], and Y-shaped stent with two branches [7]. The deployment of the J-shaped or Y-shaped stents in the bifurcation lesions will lead to a more complex mechanical response in the lesion compared with the deployment of the straight stent [8]. These stents are usually covered with a silicon membrane to mitigate the tissue ingrowth [9], as well as to facilitate the stent removal [10]. The adverse effect of the covered stent is the higher risk of migration. The partially covered stent was designed to take advantage of the merits of both the covered stent and the bare-metal one [11].
The mechanical performance of the tracheobronchial stent is essential for serving as a scaffold to restore the airflow. The radial strength, bending resistance, and torque resistance were characterized for stent implantation in arteries [12, 13, 14]. Ratnovsky et al. measured the radial strength of the straight tracheal stents using a specially designed adaptor [15]. Regardless of scarce testing data available for the tracheobronchial stents, the characterization of the self-expandable metallic stents, either J-shaped or Y-shaped, was nonexistent, but it is essential for optimal clinical outcomes. For example, a larger bending resistance of the branches might cause abnormal stresses in the lesion subjected to the change in the tracheobronchial angle during inspiration and expiration [16]. This might be associated with long-term complications.
In this work, the mechanical behavior of eight self-expandable metallic tracheobronchial stents in three types (i.e., straight, J-shaped, and Y-shaped), with or without cover, was characterized. The influence of the branches on the compression resistance of the stem in J-shaped and Y-shaped stents was investigated using the compression test and indentation test. The bending resistance of the branches in the J-shaped and Y-shaped stents was investigated through a specially designed bending test device. The impact of the cover on the compression resistance and the bending resistance was also investigated.
2 Materials and methods
Stent information
Eight self-expandable metallic tracheobronchial stents are summarized in Table 1. Two straight stents, with and without cover, were denoted as SC and S, respectively. Three J-shaped stents with one branch (i.e., uncovered J-shaped stent, covered J-shaped stent, and covered J-shaped stent with one blocked-end branch) were denoted as J, JC, and JC&B, respectively. Three Y-shaped stents with two branches, (i.e., uncovered Y-shaped stent, covered Y-shaped stent, and covered Y-shaped stent with one blocked-end branch) were denoted as Y, YC, and YC&B, respectively.
Stent specifications
Symbol | Structural information | Photos |
---|---|---|
S | Straight, uncovered 20×50mm | ![]() |
SC | Straight, covered 20×50mm | ![]() |
J | J-shaped, uncovered Stem: 20×40mm Branch: 12×20mm | ![]() |
JC | J-shaped, covered Stem: 20×40mm Branch: 12×20mm | ![]() |
JC&B | J-shaped, covered Stem: 20×40mm Blocked branch: 12×25mm | ![]() |
Y | Y-shaped, uncovered. Stem: 20×40mm Branch: 12×20mm; 12×15mm | ![]() |
YC | Y-shaped, covered. Stem: 20×40mm Branch: 12×20mm; 12×15mm | ![]() |
Y-shaped, one side of branches blocked | ||
YC&B | Stem: 20×40mm Branch: 12×15mm Blocked branch: 12×5mm | ![]() |
The nominal diameter of the straight stent and the stem in the J-shaped and Y-shaped stents was 20 mm. The nominal diameter of the branches in the J-shaped and Y-shaped stents was 12mm. The length of the straight stent and the stem in the J-shaped and Y-shaped stents was 50 mm, 40 mm, and 45 mm, respectively. The length of the branch in all the J-shaped stents was 20 mm except the JC&B, which was 25mm. The length of the two branches in the Y-shaped stents was 20 mm and 15 mm except the YC&B, which was 5mm. The branch angle, denoted as β, was 130∘ for the J-shaped stent, and 135∘ for both branches in all the Y-shaped stents (Figure 1). The angle between both of the two branches in the Y-shaped stents was 90∘. The vertex of the angle separated the branches from the stem zone for J-shaped or Y-shaped stents.

The angle between the branches and stem in (a) J-shaped stent and (b) Y-shaped stent.
All of the stents were braided with one single Nitinol wire, while the number of the mesh along the circumferential and longitudinal direction varied in different types of the stents. The pitch angle (i.e., the angle between the wires and the circumferential direction of the stem or branch (Figure 1a) can be estimated based on the two diagonal lengths of the rectangular mesh as:
where L is the length of the stem, n1 is the number of the mesh along the longitudinal direction, D is the diameter of the stem, and n2 is the number of the mesh along the circumferential direction. The pitch angles of the straight, J-shaped, and Y-shaped stent were 35.27∘, 46.70∘, and 44.40∘, respectively. All of these stents were braided with a single Nitinol wire with a diameter of 0.23mm(Micro-Tech, Nanjing, China).
Compression test
The radial strength of the straight stent and the stem in the J-shaped and Y-shaped stents were evaluated with a parallel plates compression test. A universal testing machine (TestResources, Shakopee, MN, USA) with a load cell of 5 lbs was used. Two horizontally parallel plates with a vertical handle were firmly clamped between gripping jaws to compress the stems of the stents (Figure 2a). The dis

(a) Compression test, and (b) Indentation test of the stents.
placement control mode with a compression distance of 5 mm was adopted. The compression displacement rate was 0.1mm/s,warranting a quasi-static state. The compression process was monitored to make sure there was no slip between the stent and the plates. The resistance force was recorded during the compression. The radial strength was defined as the resistance force at the compression displacement of 5 mm. The normalized radial strength was calculated as the radial strength divided by the stem length.
Indentation test
A rectangular vertical polycarbonate plate with a thickness of 2.2 mm was firmly clamped between gripping jaws to indent the stems of stents at three locations (i.e., the center and 10 mm away from each end of the stem (Figure 2b). The vertex of the branch angle (Figure 1) served as the end of the stem in the J-shaped and Y-shaped stents. The resistance force was recorded through the indentation process, and the indentation strength was defined as the resistance force at the indentation depth of 5 mm.
2.1 Bending test
The stem portion of the stents was filled with a solid rod of 20 mm in diameter, and then fixed to a custom-designed fixture, which was clamped between lower jaws (Figure 3). An L-shaped plate was firmly clamped between upper gripping jaws to compress and bend the branches of the J-shaped and Y-shaped stents. Both the bending resistance force and the deformation of the branches were recorded in synchronization during the bending procedure. The bending angle, defined as the change in the branch angle, was measured using ImageJ [17]. The linear regression between the bending angle and the bending resistance force was performed. The coefficients and constants were obtained using the minimal least square method. The Pearson correlation coefficient was calculated as:

Bending test of the branches for both J-shaped and Y-shaped stents.
where x is denoted as the bending angle, y is denoted as the bending resistance force, and μx and μy are the means of the x and y variables. There is a strong linear dependence between x and y if r is closer to 1 or −1.
3 Results
The representative compression behavior of the straight stent, with or without cover, is demonstrated in Figure 4a. The experiment showed a good repeatability for both stents. The covered stents exhibited a higher radial resistance than the uncovered ones. The normalized radial strength for all the stents is shown in Figure 4b. As expected, the uncovered stents (S, J, and Y) had less radial strength than the covered ones. The normalized radial strength was less than 0.025 N/mm for all the uncovered stents, while it ranged from 0.027 N/mm to 0.043 N/mm for those covered ones. It is interesting to observe that the normalized radial strength of the stem in the uncovered J-shaped and Y-shaped stents was higher than the straight ones, while this trend was reversed for the covered counterparts. In addition, the blocked-end configuration of the J-shaped stent (JC&B) led to an increase in the normalized radial strength by 27.80%, compared with the JC stent. The normalized radial strength for both the YC stent and YC&B stent exhibited minimal differences (1.5%), which is also similar to the normalized radial strength of the SC stent.

(a) Representative compression behavior, and (b) Normalized radial strength of the stems.
The indentation strength of each stent at three locations, free end, center, and the other end of the stem (i.e., the end adjacent to the branches) is shown Figure 5. The indentation strength was slightly larger at the center location than that at both ends. The indentation strength at the end adjacent to the branches was higher than the free end for the types JC, JC&B, YC, and YC&B. Moreover, the indentation strength showed a similar trend as the normalized radial strength among stent types. The cover induced higher indentation strength in stents.

Indentation tests of all stems.
The bending behavior of the branch in all the J-shaped stents is shown in Figure 6a. Among the three J-shaped stents: J, JC, and JC&B, the covered stents showed a higher bending resistance than the uncovered one. The blocked end (JC&B) exhibited a less bending resistance force compared with JC. This might be due to the 5mm longer branch in the JC&B compared with the JC. The regression coefficient of the bending, also referred to as bending stiffness, was approximately doubled in the JC stent, compared with that of the J stent. A positive regression intercept indicates a local deformation at the contact region between bending plate and the branch end during the initial bending. The regression intercept for the JC and J stents were 0.0395 N and 0.018 N, respectively. The bending of the JC branch exhibited more local deformation than that of the J-shaped stent.

Bending test of (a) three J-shaped stents, (b) long branches of two Y-shaped stents; and (c) short branches of three Y-shaped stents.
The bending behavior of both branches in all the Y-shaped stents is shown in Figures 6b&c. The 20mm-long branch was denoted as “_L”, and the 15mm-long branch was denoted as “_S”. No 20mm-length branch existed in the YC&B stent. Again, the covered stents showed a higher bending resistance than the uncovered one. Both the regression coefficient and interception of the YC stent were larger than the Y stent. The bending behavior of the 15mm-long branch in both YC and YC&B exhibited a minimal difference even though the other branch was distinctly different. This indicated that the bending behavior of branches in Y-shaped stents was independent from each other.
The bending stiffness of all eight branches is also summarized in Figure 7. It is obvious that the bending stiffness of J-shaped stents was less than that of Y-shaped ones. The bending stiffness ranged from 0.0052 N/∘ to 0.0103 N/∘ for J-shaped stents, and from 0.0383 N/∘ to 0.0694 N/∘ for Y-shaped stents. This could be attributed to the configurations of the branch-stem junction. The J-shaped stent has many fewer wire connections between its branch and stem. Again, the covered stents showed a higher bending resistance than the uncovered ones. It is interesting to see that the bending resistance is more sensitive to the junction configuration, rather than the cover. In addition, the 15mm-long branch had a slightly larger bending stiffness than the 20mm-long one.

Bending stiffness of branches in both J–shaped and Y-shaped stents.
4 Discussions
The mechanical performance of the tracheobronchial stents is essential for serving as a scaffold to restore the airflow. There is limited testing data available for the tracheobronchial stents. Moreover, the characterization of J-shaped or Y-shaped self-expandable metallic stents does not exist in the literature. In this work, the compression, indentation, and bending behaviors of eight commonly used self-expandable metallic tracheobronchial stents, including the J-shaped and Y-shaped stents, were characterized and compared. The radial strength of the main stems of the stents was accessed using both compression and indentation tests. The bending resistance of side branches in the J-shaped and Y-shaped stents was assessed.
The radial strength (i.e., the scaffold capacity to open the blocked lumen) is a major consideration for stent design [8, 18]. It is clear that various configurations of side branches had minimal impact on the radial strength of the main stem of the stents. However, the covered stent could improve the radial strength, compared with the uncovered counterpart [19]. This could be explained by the load sharing capacity of the cover as well as the cover-induced extra constraints on the relative movement of wires. It has been reported that the relative sliding between wires could mitigate the radial strength of the braided stents [20].
Following the stent deployment, the mismatch between stent design and the anatomic dimensions could induce the bending of stents, especially on the side branches. To our best knowledge, this work is the first quantification on the mechanical behaviors of stents with side branches. We have observed that J-shaped stents have a much lower bending resistance than that of Y-shaped stents, regardless the cover. This was attributed to their junction configurations. Moreover, a larger bending resistance in the Y-shaped stents implied a larger wall stress on the lesion, and thus a higher risk of complications [8, 21]. Besides, the covered stent exhibited more bending resistance than its uncovered counterparts. It is interesting to see that the junction configuration, rather than the cover, has more impact on the bending resistance. In addition, a shorter branch could lead to a slightly larger bending resistance. Specifically, the partial connection from the stem to branch of the J-shaped stent along the circumferential direction is flexible than the fully connected one for the Y-shape stent.
Migration behaviors observed in clinical practices [22, 23] were not considered in this work. It has been shown that a larger radial/bending strength, correlated with higher contact force between stent and the lesion, could mitigate the migration of a straight stent [24]. The concern of migration was much relieved for both the J- and Y-shaped stents, because each junction configuration, itself, constrained the migration of the stent in a bifurcated lesion [25]. The fatigue behavior of the J-shaped and Y-shaped stents might be essential due to inspiration and expiration following stenting [26]. A sophisticated devise will be considered for the fatigue test in further work.
5 Conclusion
In this work, mechanical behaviors of the self-expandable metallic tracheobronchial stents, especially the J-shaped and Y-shaped stents, were characterized using a compression test, indentation test, and bending test. The radial strength of the stem was significantly enhanced by the cover but not affected substantially by the side branches. The bending resistance of side branches was sensitive to the junction configurations, as well as the cover. The mechanical evaluation of the stent-trachea interaction with experimental and computational methods may further validate the results from this work, help establish the standard of the test, and help the stent design.
References
[1] Dutau H. et al., The use of self-expandable metallic stents in the airways in the adult population, Expert review of respiratory medicine, 2014, 8(2), 179-190.10.1586/17476348.2014.880055Search in Google Scholar PubMed
[2] Gompelmann D. et al., Self-expanding Y stents in the treatment of central airway stenosis: a retrospective analysis, Therapeut. Adv. Respir. Dis., 2013. 7(5), 255-263.10.1177/1753465813489766Search in Google Scholar PubMed
[3] Conforti S. et al., Self-expanding Y Stent for the Treatment of Malignant Tracheobronchial Stenosis. Retrospective Study, Archivos de Bronconeumología (English Ed.), 2016. 52(11), e5-e7.10.1016/j.arbr.2016.04.006Search in Google Scholar
[4] Madan K. et al., A multicenter experience with the placement of self-expanding metallic tracheobronchial Y stents, J. Bronchol. Interv. Pulmonol., 2016, 23(1), 29- 38.10.1097/LBR.0000000000000250Search in Google Scholar PubMed
[5] Dooms C. et al., Performance of fully covered self-expanding metallic stents in benign airway strictures, Respiration, 2009, 77(4), 420-426.10.1159/000203364Search in Google Scholar PubMed
[6] Freitag L. et al., Towards individualized tracheobronchial stents: technical, practical and legal considerations, Respiration, 2017, 94(5), 442-456.10.1159/000479164Search in Google Scholar PubMed
[7] Bi Y. et al., Multiple Bifurcated Covered Self-Expanding Metallic Stents for Complex Tracheobronchial Fistulas or Stenosis, Cardiovasc. Interv. Radiol., 2018, 1-7.10.1007/s00270-018-2126-6Search in Google Scholar PubMed
[8] Gu L. et al., The relation between the arterial stress and restenosis rate after coronary stenting, J. Med. Dev., 2010, 4(3), 031005.10.1115/1.4002238Search in Google Scholar
[9] Waidmann O. et al., SEMS vs cSEMS in duodenal and small bowel obstruction: high risk of migration in the covered stent group, World J. Gastroenterol., 2013, 19(37), 6199.10.3748/wjg.v19.i37.6199Search in Google Scholar PubMed PubMed Central
[10] Langer F.B. et al., Solving the problem of diflcult stent removal due to tissue ingrowth in partially uncovered esophageal self expanding metal stents, Annals Thoracic Surg., 2010, 89(5), 16911692.10.1016/j.athoracsur.2009.07.066Search in Google Scholar PubMed
[11] Telford J.J. et al., A randomized trial comparing uncovered and partially covered self-expandable metal stents in the palliation of distal malignant biliary obstruction, Gastrointest. Endoscopy, 2010, 72(5), 907-914.10.1016/j.gie.2010.08.021Search in Google Scholar PubMed
[12] McGrath D. et al., Evaluation of cover effects on bare stent mechanical response, J. Mech. Behav. Biomed. Mat., 2016, 61, 567-580.10.1016/j.jmbbm.2016.04.023Search in Google Scholar PubMed
[13] Schmidt W. et al., A comparison of the mechanical performance characteristics of seven drug-eluting stent systems, Catheter. Cardiovasc. Interv., 2009, 73(3), 350-360.10.1002/ccd.21832Search in Google Scholar PubMed
[14] Maleckis K. et al., Comparison of femoropopliteal artery stents under axial and radial compression, axial tension, bending, and torsion deformations, J. Mech. Bech. Biomed. Mat., 2017, 75, 160-168.10.1016/j.jmbbm.2017.07.017Search in Google Scholar PubMed PubMed Central
[15] Ratnovsky A. et al., Mechanical properties of different airway stents, Med. Eng. Phys., 2015, 37(4), 408-415.10.1007/978-3-319-00846-2_231Search in Google Scholar
[16] Onoe R. et al., 3D-measurement of tracheobronchial angles on inspiratory and expiratory chest CT in COPD: respiratory changes and correlation with airflow limitation, Int. J. Chronic Obstr. Pulm. Dis., 2018, 13, 2399.10.2147/COPD.S165824Search in Google Scholar PubMed PubMed Central
[17] Rueden C.T., et al., ImageJ2: Image for the next generation of scientific image data, BMC Bioinformatics, 2017, 18(1), 529.10.1186/s12859-017-1934-zSearch in Google Scholar PubMed PubMed Central
[18] Kim J.H., Kang T.J., YuW.-R., Simulation of mechanical behavior of temperature-responsive braided stents made of shape memory polyurethanes, J. Biomech., 2010, 43(4), 632-643.10.1016/j.jbiomech.2009.10.032Search in Google Scholar PubMed
[19] Isayama H., et al., Measurement of radial and axial forces of biliary self-expandable metallic stents, Gastrointest. Endosc., 2009, 70(1), 37-44.10.1016/j.gie.2008.09.032Search in Google Scholar PubMed
[20] Zhao S., Liu X.C., Gu L., The impact of wire stent fabrication technique on the performance of stent placement, J. Med. Dev., 2012. 6(1), 011007.10.1115/1.4005788Search in Google Scholar
[21] Zhao S., Gu L., Froemming S.R., Performance of self-expanding Nitinol stent in a curved artery: impact of stent length and deployment orientation, J. Biomech. Eng., 2012, 134(7), 071007.10.1115/1.4007095Search in Google Scholar PubMed
[22] Lee H.J. et al., Airway stent complications: the role of followup bronchoscopy as a surveillance method, J. Thoracic Dis., 2017, 9(11), 4651.10.21037/jtd.2017.09.139Search in Google Scholar PubMed PubMed Central
[23] Garbey M. et al., Esophageal stent migration: Testing few hypothesis with a simplified mathematical model, Comput. Biol. Med., 2016. 79, 259-265.10.1016/j.compbiomed.2016.10.024Search in Google Scholar PubMed
[24] Mozafari H. et al., Migration resistance of esophageal stents: The role of stent design, Comput. Biol. Med., 2018, 100, 43-49.10.1016/j.compbiomed.2018.06.031Search in Google Scholar PubMed
[25] Bi Y. et al., Metallic small y stent placement at primary right carina for bronchial disease, BMC Pulmon. Med., 2018, 18(1), 182.10.1186/s12890-018-0742-1Search in Google Scholar PubMed PubMed Central
[26] Avasarala S.K., Freitag L., Mehta A.C., Metallic Endo- bronchial Stents: A Contemporary Resurrection, Chest, 2019, 155(6), 1246-1259.10.1016/j.chest.2018.12.001Search in Google Scholar PubMed
© 2019 Y. Wang et al., published by De Gruyter
This work is licensed under the Creative Commons Attribution 4.0 Public License.
Articles in the same Issue
- Research Articles
- Investigation of rare earth upconversion fluorescent nanoparticles in biomedical field
- Carbon Nanotubes Coated Paper as Current Collectors for Secondary Li-ion Batteries
- Insight into the working wavelength of hotspot effects generated by popular nanostructures
- Novel Lead-free biocompatible piezoelectric Hydroxyapatite (HA) – BCZT (Ba0.85Ca0.15Zr0.1Ti0.9O3) nanocrystal composites for bone regeneration
- Effect of defects on the motion of carbon nanotube thermal actuator
- Dynamic mechanical behavior of nano-ZnO reinforced dental composite
- Fabrication of Ag Np-coated wetlace nonwoven fabric based on amino-terminated hyperbranched polymer
- Fractal analysis of pore structures in graphene oxide-carbon nanotube based cementitious pastes under different ultrasonication
- Effect of PVA fiber on durability of cementitious composite containing nano-SiO2
- Cr effects on the electrical contact properties of the Al2O3-Cu/15W composites
- Experimental evaluation of self-expandable metallic tracheobronchial stents
- Experimental study on the existence of nano-scale pores and the evolution of organic matter in organic-rich shale
- Mechanical characterizations of braided composite stents made of helical polyethylene terephthalate strips and NiTi wires
- Mechanical properties of boron nitride sheet with randomly distributed vacancy defects
- Fabrication, mechanical properties and failure mechanism of random and aligned nanofiber membrane with different parameters
- Micro- structure and rheological properties of graphene oxide rubber asphalt
- First-principles calculations of mechanical and thermodynamic properties of tungsten-based alloy
- Adsorption performance of hydrophobic/hydrophilic silica aerogel for low concentration organic pollutant in aqueous solution
- Preparation of spherical aminopropyl-functionalized MCM-41 and its application in removal of Pb(II) ion from aqueous solution
- Electrical conductivity anisotropy of copper matrix composites reinforced with SiC whiskers
- Miniature on-fiber extrinsic Fabry-Perot interferometric vibration sensors based on micro-cantilever beam
- Electric-field assisted growth and mechanical bactericidal performance of ZnO nanoarrays with gradient morphologies
- Flexural behavior and mechanical model of aluminum alloy mortise-and-tenon T-joints for electric vehicle
- Synthesis of nano zirconium oxide and its application in dentistry
- Surface modification of nano-sized carbon black for reinforcement of rubber
- Temperature-dependent negative Poisson’s ratio of monolayer graphene: Prediction from molecular dynamics simulations
- Finite element nonlinear transient modelling of carbon nanotubes reinforced fiber/polymer composite spherical shells with a cutout
- Preparation of low-permittivity K2O–B2O3–SiO2–Al2O3 composites without the addition of glass
- Large amplitude vibration of doubly curved FG-GRC laminated panels in thermal environments
- Enhanced flexural properties of aramid fiber/epoxy composites by graphene oxide
- Correlation between electrochemical performance degradation and catalyst structural parameters on polymer electrolyte membrane fuel cell
- Materials characterization of advanced fillers for composites engineering applications
- Humic acid assisted stabilization of dispersed single-walled carbon nanotubes in cementitious composites
- Test on axial compression performance of nano-silica concrete-filled angle steel reinforced GFRP tubular column
- Multi-scale modeling of the lamellar unit of arterial media
- The multiscale enhancement of mechanical properties of 3D MWK composites via poly(oxypropylene) diamines and GO nanoparticles
- Mechanical properties of circular nano-silica concrete filled stainless steel tube stub columns after being exposed to freezing and thawing
- Arc erosion behavior of TiB2/Cu composites with single-scale and dual-scale TiB2 particles
- Yb3+-containing chitosan hydrogels induce B-16 melanoma cell anoikis via a Fak-dependent pathway
- Template-free synthesis of Se-nanorods-rGO nanocomposite for application in supercapacitors
- Effect of graphene oxide on chloride penetration resistance of recycled concrete
- Bending resistance of PVA fiber reinforced cementitious composites containing nano-SiO2
- Review Articles
- Recent development of Supercapacitor Electrode Based on Carbon Materials
- Mechanical contribution of vascular smooth muscle cells in the tunica media of artery
- Applications of polymer-based nanoparticles in vaccine field
- Toxicity of metallic nanoparticles in the central nervous system
- Parameter control and concentration analysis of graphene colloids prepared by electric spark discharge method
- A critique on multi-jet electrospinning: State of the art and future outlook
- Electrospun cellulose acetate nanofibers and Au@AgNPs for antimicrobial activity - A mini review
- Recent progress in supercapacitors based on the advanced carbon electrodes
- Recent progress in shape memory polymer composites: methods, properties, applications and prospects
- In situ capabilities of Small Angle X-ray Scattering
- Review of nano-phase effects in high strength and conductivity copper alloys
- Progress and challenges in p-type oxide-based thin film transistors
- Advanced materials for flexible solar cell applications
- Phenylboronic acid-decorated polymeric nanomaterials for advanced bio-application
- The effect of nano-SiO2 on concrete properties: a review
- A brief review for fluorinated carbon: synthesis, properties and applications
- A review on the mechanical properties for thin film and block structure characterised by using nanoscratch test
- Cotton fibres functionalized with plasmonic nanoparticles to promote the destruction of harmful molecules: an overview
Articles in the same Issue
- Research Articles
- Investigation of rare earth upconversion fluorescent nanoparticles in biomedical field
- Carbon Nanotubes Coated Paper as Current Collectors for Secondary Li-ion Batteries
- Insight into the working wavelength of hotspot effects generated by popular nanostructures
- Novel Lead-free biocompatible piezoelectric Hydroxyapatite (HA) – BCZT (Ba0.85Ca0.15Zr0.1Ti0.9O3) nanocrystal composites for bone regeneration
- Effect of defects on the motion of carbon nanotube thermal actuator
- Dynamic mechanical behavior of nano-ZnO reinforced dental composite
- Fabrication of Ag Np-coated wetlace nonwoven fabric based on amino-terminated hyperbranched polymer
- Fractal analysis of pore structures in graphene oxide-carbon nanotube based cementitious pastes under different ultrasonication
- Effect of PVA fiber on durability of cementitious composite containing nano-SiO2
- Cr effects on the electrical contact properties of the Al2O3-Cu/15W composites
- Experimental evaluation of self-expandable metallic tracheobronchial stents
- Experimental study on the existence of nano-scale pores and the evolution of organic matter in organic-rich shale
- Mechanical characterizations of braided composite stents made of helical polyethylene terephthalate strips and NiTi wires
- Mechanical properties of boron nitride sheet with randomly distributed vacancy defects
- Fabrication, mechanical properties and failure mechanism of random and aligned nanofiber membrane with different parameters
- Micro- structure and rheological properties of graphene oxide rubber asphalt
- First-principles calculations of mechanical and thermodynamic properties of tungsten-based alloy
- Adsorption performance of hydrophobic/hydrophilic silica aerogel for low concentration organic pollutant in aqueous solution
- Preparation of spherical aminopropyl-functionalized MCM-41 and its application in removal of Pb(II) ion from aqueous solution
- Electrical conductivity anisotropy of copper matrix composites reinforced with SiC whiskers
- Miniature on-fiber extrinsic Fabry-Perot interferometric vibration sensors based on micro-cantilever beam
- Electric-field assisted growth and mechanical bactericidal performance of ZnO nanoarrays with gradient morphologies
- Flexural behavior and mechanical model of aluminum alloy mortise-and-tenon T-joints for electric vehicle
- Synthesis of nano zirconium oxide and its application in dentistry
- Surface modification of nano-sized carbon black for reinforcement of rubber
- Temperature-dependent negative Poisson’s ratio of monolayer graphene: Prediction from molecular dynamics simulations
- Finite element nonlinear transient modelling of carbon nanotubes reinforced fiber/polymer composite spherical shells with a cutout
- Preparation of low-permittivity K2O–B2O3–SiO2–Al2O3 composites without the addition of glass
- Large amplitude vibration of doubly curved FG-GRC laminated panels in thermal environments
- Enhanced flexural properties of aramid fiber/epoxy composites by graphene oxide
- Correlation between electrochemical performance degradation and catalyst structural parameters on polymer electrolyte membrane fuel cell
- Materials characterization of advanced fillers for composites engineering applications
- Humic acid assisted stabilization of dispersed single-walled carbon nanotubes in cementitious composites
- Test on axial compression performance of nano-silica concrete-filled angle steel reinforced GFRP tubular column
- Multi-scale modeling of the lamellar unit of arterial media
- The multiscale enhancement of mechanical properties of 3D MWK composites via poly(oxypropylene) diamines and GO nanoparticles
- Mechanical properties of circular nano-silica concrete filled stainless steel tube stub columns after being exposed to freezing and thawing
- Arc erosion behavior of TiB2/Cu composites with single-scale and dual-scale TiB2 particles
- Yb3+-containing chitosan hydrogels induce B-16 melanoma cell anoikis via a Fak-dependent pathway
- Template-free synthesis of Se-nanorods-rGO nanocomposite for application in supercapacitors
- Effect of graphene oxide on chloride penetration resistance of recycled concrete
- Bending resistance of PVA fiber reinforced cementitious composites containing nano-SiO2
- Review Articles
- Recent development of Supercapacitor Electrode Based on Carbon Materials
- Mechanical contribution of vascular smooth muscle cells in the tunica media of artery
- Applications of polymer-based nanoparticles in vaccine field
- Toxicity of metallic nanoparticles in the central nervous system
- Parameter control and concentration analysis of graphene colloids prepared by electric spark discharge method
- A critique on multi-jet electrospinning: State of the art and future outlook
- Electrospun cellulose acetate nanofibers and Au@AgNPs for antimicrobial activity - A mini review
- Recent progress in supercapacitors based on the advanced carbon electrodes
- Recent progress in shape memory polymer composites: methods, properties, applications and prospects
- In situ capabilities of Small Angle X-ray Scattering
- Review of nano-phase effects in high strength and conductivity copper alloys
- Progress and challenges in p-type oxide-based thin film transistors
- Advanced materials for flexible solar cell applications
- Phenylboronic acid-decorated polymeric nanomaterials for advanced bio-application
- The effect of nano-SiO2 on concrete properties: a review
- A brief review for fluorinated carbon: synthesis, properties and applications
- A review on the mechanical properties for thin film and block structure characterised by using nanoscratch test
- Cotton fibres functionalized with plasmonic nanoparticles to promote the destruction of harmful molecules: an overview