Background: Vitamin B12 and folate measurements in serum show wide inter-methodology variability. This variability appears to be due in part to the lack of standardisation against internationally accepted reference materials. Pooled human serum, lyophilised in ampoules and designated 03/178, was therefore evaluated by 24 laboratories in seven countries for its suitability to serve as an International Standard (IS) for B12 and folate.
Methods: IS 03/178 was assayed using a range of commercial analysers, microbiological assays and, for folate, candidate reference methods based on liquid chromatography coupled to isotope-dilution tandem mass spectrometry (LC/MS/MS).
Results: Mean vitamin B12 and folate values for reconstituted 03/178 across all laboratories and methods were 480 pg/mL [coefficient of variation (CV) 12.8%] and 5.52 ng/mL (CV 17.1%), respectively. The total folate content of reconstituted 03/178, determined using LC/MS/MS, was 12.1 nmol/L (equivalent to 5.33 ng/mL), made up of 9.75 nmol/L 5-methyl tetrahydrofolic acid (5MeTHF; CV 5.5%), 1.59 nmol/L 5-formyl tetrahydrofolic acid (5FoTHF; CV 4.2%) and 0.74 nmol/L folic acid (FA; CV 31.6%). The inclusion of three serum samples in the study with different B12 and folate levels demonstrated a considerable reduction in inter-laboratory variability when the B12 and folate content of the samples was determined relative to the IS 03/178 rather than to the analyser calibration. IS 03/178 demonstrated satisfactory long-term stability in accelerated degradation studies.
Conclusions: Use of IS 03/178 to standardise serum B12 and folate assays reduced inter-laboratory variability. The World Health Organization (WHO) Expert Committee on Biological Standardisation established 03/178 as the first IS for serum vitamin B12 and serum folate, with assigned values of 480 pg/mL of vitamin B12 and 12.1 nmol/L folate when the lyophilised contents of the ampoule are reconstituted with 1 mL of water.
Clin Chem Lab Med 2007;45:380–6.
References
1. Working Party of the BCSH General Haematology Task Force. Guidelines on the investigation and diagnosis of cobalamin and folate deficiencies. Clin Lab Haematol 1994;16:101–15.Suche in Google Scholar
2. Solomon LR. Cobalamin responsive disorders in the ambulatory care setting: unreliability of cobalamin, methylmalonic acid and homocysteine testing. Blood2005;105:978–85.10.1182/blood-2004-04-1641Suche in Google Scholar PubMed
3. Goringe A, Ellis R, McDowell I, Vidal-Alaball J, Jenkins C, Butler C, et al. The limited value of methylmalonic acid, homocysteine, and holotranscobalamin in the diagnosis of early B12 deficiency. Haematologica2006;91:231–4.Suche in Google Scholar
4. Curtis AD, Mussett MV, Kennedy DA. British Standard for human serum vitamin B12. Clin Lab Haematol1986;8:135–47.10.1111/j.1365-2257.1986.tb00088.xSuche in Google Scholar PubMed
5. Gunter EW, Bowman BA, Caudill SP, Twite DB, Adams MJ, Sampson EJ. Results of an international round robin for serum and whole-blood folate. Clin Chem1996;42:1689–94.10.1093/clinchem/42.10.1689Suche in Google Scholar
6. Thorpe SJ, Sands D, Heath AB, Hamilton MS, Blackmore S, Barrowcliffe T. An International Standard for whole blood folate: evaluation of a lyophilised haemolysate in an international collaborative study. Clin Chem Lab Med2004;42:533–9.10.1515/CCLM.2004.090Suche in Google Scholar PubMed
7. Satterfield MB, Sniegoski LT, Sharpless KF, Welch MJ, Hornikova A, Zhang NF, et al. Development of a new standard reference material: SRM 1955 (homocysteine and folate in human serum). Anal Bioanal Chem2006;385:612–22.10.1007/s00216-006-0434-1Suche in Google Scholar PubMed
8. Nelson BC, Pfeiffer CM, Margolis SA, Nelson CP. Solid-phase extraction-electrospray ionization mass spectrometry for the quantification of folate in human plasma or serum. Anal Biochem2004;325:41–51.10.1016/j.ab.2003.10.009Suche in Google Scholar PubMed
9. Pfeiffer CM, Fazili Z, McCoy L, Zhang M, Gunter EW. Determination of folate vitamers in human serum by stable-isotope-dilution tandem mass spectrometry and comparison with radioassay and microbiologic assay. Clin Chem2004;50:423–32.10.1373/clinchem.2003.026955Suche in Google Scholar PubMed
10. Kelleher BP, O'Broin SD. Microbiological assay for vitamin B12 performed in 96-well microtitre plates. J Clin Pathol1991;44:592–5.10.1136/jcp.44.7.592Suche in Google Scholar PubMed PubMed Central
11. O'Broin S, Kelleher B. Microbiological assay on microtitre plates of folate in serum and red cells. J Clin Pathol1992;45:344–7.10.1136/jcp.45.4.344Suche in Google Scholar PubMed PubMed Central
12. Kirkwood TB. Predicting the stability of biological standards and products. Biometrics1977;33:736–42.10.2307/2529472Suche in Google Scholar
13. WHO. Recommendations for the preparation, characterization and establishment of international and other biological reference materials. WHO/BS/04.1995 (revised 2004). Geneva: World Health Organization.Suche in Google Scholar
14. In vitro diagnostic devices – measurement of quantities in biological samples – metrological traceability of values assigned to calibrators and control materials. EN ISO 17511. ISO, 2003.Suche in Google Scholar
©2007 by Walter de Gruyter Berlin New York
Artikel in diesem Heft
- From human genetic variations to prediction of risks and responses to drugs and the environment
- Nutrigenomics – 2006 update
- How to comprehensively analyse proteins and how this influences nutritional research
- Genotypes, obesity and type 2 diabetes – can genetic information motivate weight loss? A review
- The Gene-Diet Attica Investigation on childhood obesity (GENDAI): overview of the study design
- Polymorphisms in the APOA1/C3/A4/A5 gene cluster and cholesterol responsiveness to dietary change
- Nutri-epigenomics: lifelong remodelling of our epigenomes by nutritional and metabolic factors and beyond
- Emerging role of cathepsin S in obesity and its associated diseases
- Association analysis of hepatitis virus B infection with haplotypes of the TBX21 gene promoter region in the Chinese population
- Multiplex polymerase chain reaction on FTA cards vs. flow cytometry for B-lymphocyte clonality
- Real-time multiplex PCR assay for genotyping of three apolipoprotein E alleles and two choline acetyltransferase alleles with three hybridization probes
- Immunomagnetic CD45 depletion does not improve cytokeratin 20 RT-PCR in colorectal cancer
- Analysis of the components of hypertransaminasemia after liver resection
- Fine characterization of mitral valve glycosaminoglycans and their modification with degenerative disease
- Oxidative stress evaluated using an automated method for hydroperoxide estimation in patients with coronary artery disease
- Secretory phospholipase A2 activity and release kinetics of vascular tissue remodelling biomarkers after coronary artery bypass grafting with and without cardiopulmonary bypass
- Clustered components of the metabolic syndrome and platelet counts in Japanese females
- International Standard for serum vitamin B12 and serum folate: international collaborative study to evaluate a batch of lyophilised serum for B12 and folate content
- Multicentre physiological reference intervals for serum concentrations of immunoglobulins A, G and M, complement C3c and C4 measured with Tina-Quant® reagents systems
- In vivo and in vitro allergy diagnostics: it's time to reappraise the costs
- Experience with post-market surveillance of in-vitro diagnostic medical devices for lay use in Germany
- Evaluation of the high-sensitivity, full-range Olympus CRP OSR6199 application on the Olympus AU640®
- How to improve the teaching of urine microscopy
- In vitro determination of allergen-specific serum IgE. Comparative analysis of three methods
- Efficacy of a new blocker against anti-ruthenium antibody interference in the Elecsys free triiodothyronine assay
- Clinical indications for plasma protein assays: transthyretin (prealbumin) in inflammation and malnutrition: International Federation of Clinical Chemistry and Laboratory Medicine (IFCC): IFCC Scientific Division Committee on Plasma Proteins (C-PP)
- Meeting Report: From human genetic variations to prediction of risks and responses to drugs and the environment
Artikel in diesem Heft
- From human genetic variations to prediction of risks and responses to drugs and the environment
- Nutrigenomics – 2006 update
- How to comprehensively analyse proteins and how this influences nutritional research
- Genotypes, obesity and type 2 diabetes – can genetic information motivate weight loss? A review
- The Gene-Diet Attica Investigation on childhood obesity (GENDAI): overview of the study design
- Polymorphisms in the APOA1/C3/A4/A5 gene cluster and cholesterol responsiveness to dietary change
- Nutri-epigenomics: lifelong remodelling of our epigenomes by nutritional and metabolic factors and beyond
- Emerging role of cathepsin S in obesity and its associated diseases
- Association analysis of hepatitis virus B infection with haplotypes of the TBX21 gene promoter region in the Chinese population
- Multiplex polymerase chain reaction on FTA cards vs. flow cytometry for B-lymphocyte clonality
- Real-time multiplex PCR assay for genotyping of three apolipoprotein E alleles and two choline acetyltransferase alleles with three hybridization probes
- Immunomagnetic CD45 depletion does not improve cytokeratin 20 RT-PCR in colorectal cancer
- Analysis of the components of hypertransaminasemia after liver resection
- Fine characterization of mitral valve glycosaminoglycans and their modification with degenerative disease
- Oxidative stress evaluated using an automated method for hydroperoxide estimation in patients with coronary artery disease
- Secretory phospholipase A2 activity and release kinetics of vascular tissue remodelling biomarkers after coronary artery bypass grafting with and without cardiopulmonary bypass
- Clustered components of the metabolic syndrome and platelet counts in Japanese females
- International Standard for serum vitamin B12 and serum folate: international collaborative study to evaluate a batch of lyophilised serum for B12 and folate content
- Multicentre physiological reference intervals for serum concentrations of immunoglobulins A, G and M, complement C3c and C4 measured with Tina-Quant® reagents systems
- In vivo and in vitro allergy diagnostics: it's time to reappraise the costs
- Experience with post-market surveillance of in-vitro diagnostic medical devices for lay use in Germany
- Evaluation of the high-sensitivity, full-range Olympus CRP OSR6199 application on the Olympus AU640®
- How to improve the teaching of urine microscopy
- In vitro determination of allergen-specific serum IgE. Comparative analysis of three methods
- Efficacy of a new blocker against anti-ruthenium antibody interference in the Elecsys free triiodothyronine assay
- Clinical indications for plasma protein assays: transthyretin (prealbumin) in inflammation and malnutrition: International Federation of Clinical Chemistry and Laboratory Medicine (IFCC): IFCC Scientific Division Committee on Plasma Proteins (C-PP)
- Meeting Report: From human genetic variations to prediction of risks and responses to drugs and the environment