Automated immunoassay methods for ferritin: recovery studies to assess traceability to an international standard
-
Sheena Blackmore
Abstract
Background: Ferritin standardisation is problematical due to the heterogeneity of ferritin isoforms and the antibodies used in its immunoassay, and the lack of a reference measurement procedure. We investigated the performance of the 1st (liver), 2nd (spleen) and 3rd (recombinant) International Standards (ISs) for ferritin in major assays.
Methods: The ferritin in a serum pool ‘spiked’ with either the 2nd or 3rd IS for ferritin was measured by 52 laboratories using five automated methods and the recovery of the target values calculated. A smaller serum pool was ‘spiked’ with the 1st IS for a limited recovery exercise. The ferritin values of five serum samples were also measured and recalculated relative to the ISs.
Results: Recoveries of each of the 2nd and 3rd ISs were 90%–110% for four of five methods; recoveries of the 1st IS were 104% and 111% for two of three methods claiming traceability to this IS. One method significantly over-recovered each of the IS (124%–155%). Recalculating the ferritin values of the serum samples relative to the IS reduced the overall inter-method agreement, largely because of the anomalous over-recovery of the IS by one method.
Conclusions: The use of the 3rd IS to standardise assays will minimise assay drift due to manufacturers adopting a ‘harmonisation’ approach in which the calibration is adjusted to conform to overall mean values. Standardisation against the current IS also ensures compliance with the European Union In-Vitro Diagnostic Directive which requires traceability of assay calibrators to reference materials of a higher order. Assay drift may result in poor sensitivity and specificity in the diagnosis of iron status, and would require laboratories to continually re-evaluate reference intervals.
Clin Chem Lab Med 2008;46:1450–7.
©2008 by Walter de Gruyter Berlin New York
Articles in the same Issue
- Review
- Use of biochemical markers of bone turnover in the management of postmenopausal osteoporosis
- Minireview
- Is serum procalcitonin able to predict long-term kidney morbidity from urinary tract infections in children?
- Genetics and Molecular Diagnostics
- Association between the 2756A> G variant in the gene encoding methionine synthase and myocardial infarction in Tunisian patients
- Common 894G>T single nucleotide polymorphism in the gene coding for endothelial nitric oxide synthase (eNOS) and risk of congenital heart defects
- New sequence variants in BRCA1 and BRCA2 genes detected by high-resolution melting analysis in an elderly healthy female population in Croatia
- Enzyme-free signal amplification of analyte in a single closed tube by fluorescent hybridization chain reaction
- Five X-chromosome short tandem repeats in a Western Mexican population
- General Clinical Chemistry and Laboratory Medicine
- Prolyl-hydroxyproline dipeptide in non-hydrolyzed morning urine and its value in postmenopausal osteoporosis
- Methionine metabolism in an animal model of sepsis
- Hypochlorous acid is a potent inactivator of human plasminogen at concentrations secreted by activated granulocytes
- Association of serum γ-glutamyltransferase with C-reactive protein levels and white blood cell count in Korean adults
- Immunoradiometric determination of thyroglobulin in serum samples by time calibration transfer
- Serum interleukin 6 (IL-6) and C-reactive protein (CRP) levels in colorectal adenoma and cancer patients
- Measurement of 10,11-dihydro-10-hydroxy-carbamazepine in serum and plasma by high-performance liquid chromatography
- The relationship between usual coffee consumption and serum C-reactive protein level in a Japanese female population
- Reference Values and Biological Variations
- Analytical quality goals for parathyroid hormone based on biological variation
- Preanalytical effects of pneumatic tube transport on routine haematology, coagulation parameters, platelet function and global coagulation
- Validation and Outcome Studies
- Automated immunoassay methods for ferritin: recovery studies to assess traceability to an international standard
- An expert system for the classification of serum protein electrophoresis patterns
- Impact of a satellite laboratory on turnaround times for the emergency department
- Effect of ampicillin-sulbactam on clinical capillary zone electrophoresis of serum proteins
- Guidelines and Recommendations
- Protein standardization V: value transfer. A practical protocol for the assignment of serum protein values from a Reference Material to a Target Material
- Letter to the Editor
- Biological variability evaluation and comparison of three different methods for C-peptide measurement
- Abstracts
- 10.1515/CCLM.2008.361
Articles in the same Issue
- Review
- Use of biochemical markers of bone turnover in the management of postmenopausal osteoporosis
- Minireview
- Is serum procalcitonin able to predict long-term kidney morbidity from urinary tract infections in children?
- Genetics and Molecular Diagnostics
- Association between the 2756A> G variant in the gene encoding methionine synthase and myocardial infarction in Tunisian patients
- Common 894G>T single nucleotide polymorphism in the gene coding for endothelial nitric oxide synthase (eNOS) and risk of congenital heart defects
- New sequence variants in BRCA1 and BRCA2 genes detected by high-resolution melting analysis in an elderly healthy female population in Croatia
- Enzyme-free signal amplification of analyte in a single closed tube by fluorescent hybridization chain reaction
- Five X-chromosome short tandem repeats in a Western Mexican population
- General Clinical Chemistry and Laboratory Medicine
- Prolyl-hydroxyproline dipeptide in non-hydrolyzed morning urine and its value in postmenopausal osteoporosis
- Methionine metabolism in an animal model of sepsis
- Hypochlorous acid is a potent inactivator of human plasminogen at concentrations secreted by activated granulocytes
- Association of serum γ-glutamyltransferase with C-reactive protein levels and white blood cell count in Korean adults
- Immunoradiometric determination of thyroglobulin in serum samples by time calibration transfer
- Serum interleukin 6 (IL-6) and C-reactive protein (CRP) levels in colorectal adenoma and cancer patients
- Measurement of 10,11-dihydro-10-hydroxy-carbamazepine in serum and plasma by high-performance liquid chromatography
- The relationship between usual coffee consumption and serum C-reactive protein level in a Japanese female population
- Reference Values and Biological Variations
- Analytical quality goals for parathyroid hormone based on biological variation
- Preanalytical effects of pneumatic tube transport on routine haematology, coagulation parameters, platelet function and global coagulation
- Validation and Outcome Studies
- Automated immunoassay methods for ferritin: recovery studies to assess traceability to an international standard
- An expert system for the classification of serum protein electrophoresis patterns
- Impact of a satellite laboratory on turnaround times for the emergency department
- Effect of ampicillin-sulbactam on clinical capillary zone electrophoresis of serum proteins
- Guidelines and Recommendations
- Protein standardization V: value transfer. A practical protocol for the assignment of serum protein values from a Reference Material to a Target Material
- Letter to the Editor
- Biological variability evaluation and comparison of three different methods for C-peptide measurement
- Abstracts
- 10.1515/CCLM.2008.361