Protein standardization V: value transfer. A practical protocol for the assignment of serum protein values from a Reference Material to a Target Material

  • Søren Blirup-Jensen , A. Myron Johnson , Marianne Larsen und on behalf of the IFCC Committee on Plasma Proteins
Veröffentlicht/Copyright: 29. August 2008
Veröffentlichen auch Sie bei De Gruyter Brill

Abstract

We present a practical protocol for the assignment of values to serum proteins in a Target Material using a Reference Material. This protocol is based on the model of Direct Value Transfer between serum matrices and is intended to improve the value assignment of commercial calibrators using the Reference Material CRM 470 (now labeled ERM-DA 470) or similar reference materials. The procedure describes the general as well as the practical principles involved in the value assignment (with examples). The practical transfer protocol is based on multiple assays of 6 dilutions of the Reference Material and 6 dilutions of the Target Material. The transfer protocol requires several measurements a day repeated on several days, an important prerequisite being that all reconstitutions and dilutions are controlled by weighing thus reducing uncertainty in the transfer. In open systems that allow the use of the Reference Material as calibrator and the Target Material as samples, the proportionality of the two materials (the presence or absence of matrix effects) can now be directly assessed by evaluating a single regression plot. If no matrix effects are found, the regression line will pass through zero with a slope equal to the ratio of the concentrations of the two materials. In closed systems, the dedicated commercial calibrator has to be used as such; the Reference Material and the Target Material are now assayed as samples against this calibrator. Two regression plots are therefore obtained; if no matrix effects are present among the two materials and the calibrator, both the Reference and Target Materials will show zero intercepts, and the ratio of the two slopes will equal the ratio of the concentrations.

Clin Chem Lab Med 2008;46:1470–9.


Corresponding author: S. Blirup-Jensen, DVM, PhD, Department of Clinical Chemistry, Lund University Hospital, 221 85 Lund, Sweden Phone: +46-46-173465, Fax: +46-46-130064,

Received: 2008-6-25
Published Online: 2008-08-29
Published in Print: 2008-10-01

©2008 by Walter de Gruyter Berlin New York

Artikel in diesem Heft

  1. Review
  2. Use of biochemical markers of bone turnover in the management of postmenopausal osteoporosis
  3. Minireview
  4. Is serum procalcitonin able to predict long-term kidney morbidity from urinary tract infections in children?
  5. Genetics and Molecular Diagnostics
  6. Association between the 2756A> G variant in the gene encoding methionine synthase and myocardial infarction in Tunisian patients
  7. Common 894G>T single nucleotide polymorphism in the gene coding for endothelial nitric oxide synthase (eNOS) and risk of congenital heart defects
  8. New sequence variants in BRCA1 and BRCA2 genes detected by high-resolution melting analysis in an elderly healthy female population in Croatia
  9. Enzyme-free signal amplification of analyte in a single closed tube by fluorescent hybridization chain reaction
  10. Five X-chromosome short tandem repeats in a Western Mexican population
  11. General Clinical Chemistry and Laboratory Medicine
  12. Prolyl-hydroxyproline dipeptide in non-hydrolyzed morning urine and its value in postmenopausal osteoporosis
  13. Methionine metabolism in an animal model of sepsis
  14. Hypochlorous acid is a potent inactivator of human plasminogen at concentrations secreted by activated granulocytes
  15. Association of serum γ-glutamyltransferase with C-reactive protein levels and white blood cell count in Korean adults
  16. Immunoradiometric determination of thyroglobulin in serum samples by time calibration transfer
  17. Serum interleukin 6 (IL-6) and C-reactive protein (CRP) levels in colorectal adenoma and cancer patients
  18. Measurement of 10,11-dihydro-10-hydroxy-carbamazepine in serum and plasma by high-performance liquid chromatography
  19. The relationship between usual coffee consumption and serum C-reactive protein level in a Japanese female population
  20. Reference Values and Biological Variations
  21. Analytical quality goals for parathyroid hormone based on biological variation
  22. Preanalytical effects of pneumatic tube transport on routine haematology, coagulation parameters, platelet function and global coagulation
  23. Validation and Outcome Studies
  24. Automated immunoassay methods for ferritin: recovery studies to assess traceability to an international standard
  25. An expert system for the classification of serum protein electrophoresis patterns
  26. Impact of a satellite laboratory on turnaround times for the emergency department
  27. Effect of ampicillin-sulbactam on clinical capillary zone electrophoresis of serum proteins
  28. Guidelines and Recommendations
  29. Protein standardization V: value transfer. A practical protocol for the assignment of serum protein values from a Reference Material to a Target Material
  30. Letter to the Editor
  31. Biological variability evaluation and comparison of three different methods for C-peptide measurement
  32. Abstracts
  33. 10.1515/CCLM.2008.361
Heruntergeladen am 12.9.2025 von https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2008.289/html
Button zum nach oben scrollen