Surface coil intensity correction in magnetic resonance imaging in spinal metastases

  • Hong Ren , Wei Lin und Xianjun Ding EMAIL logo
Veröffentlicht/Copyright: 20. Mai 2017

Abstract

Objective

To evaluate the clinical application of phased-array surface coil intensity correction in magnetic resonance imaging (MRI) in spinal metastases.

Methods

3 phantoms and 50 patients with a corresponding total number of 80 spinal metastases were included in this study. Fast spin echo T1- and T2- weighted MRI with and without surface coil intensity correction was routinely performed for all phantoms and patients. Phantoms were evaluated by means of variance to mean ratio of signal intensity on both T1- and T2- weighted MRI obtained with and without surface coil intensity correction. Spinal metastases were evaluated by image quality scores; reading time per case on both T1- and T2- weighted MRI obtained with and without surface coil intensity correction.

Results

Spinal metastases were diagnosed more successfully on MRI with surface coil intensity correction than on MRI with conventional surface coil technique. The variance to mean ratio of signal intensity was 53.36% for original T1-weighted MRI and 53.58% for original T2-weighted MRI. The variance to mean ratio of signal intensity was reduced to 18.99% for T1-weighted MRI with surface coil intensity correction and 22.77% for T2-weighted MRI with surface coil intensity correction. The overall image quality scores (interface conspicuity of lesion and details of lesion) were significantly higher than those of the original MRI. The reading time per case was shorter for MRI with surface coil intensity correction than for MRI without surface coil intensity correction.

Conclusions

Phased-array surface coil intensity correction in MRIs of spinal metastases provides improvements in image quality that leads to more successfully detection and assessment of spinal metastases than original MRI.

1 Introduction

Magnetic resonance images offer accurate spatial information about human anatomy. A quantitative measurement of sizes and volumes requires a careful segmentation of the different tissues to visualize the lesions or organs of interest. A surface coil provides a very high signal-to-noise ratio (SNR) close to the region where it is used, but it suffers from decreased signal far from the coil. The phased-array surface coil is utilized to overcome the limited coverage of the surface coil by combining the images from two or more coils to produce a signal image. However, the non-uniformity of profile reception of phased-array surface coils still hampers their application. We have evaluated the coil sensitivity modulation from surface coil intensity correction (SCIC), which is used to improve the image homogeneity of magnetic resonance imaging when a phased-array surface coil is used for reception. In this study, SCIC was evaluated for both fast spin echo (FSE) T1-weighted and T2-weighted on a phantom study and on a spinal metastases MRI. Its efficiency was illustrated with the aid of both phantom and spinal metastases MRI.

2 Material and methods

2.1 Phantom studies

The SCIC method [1,2] was tested with a 1.5 Tesla SIGNA CV/i scanner (GE Medical Systems). To statistically evaluate the effect of the correction method, a homogeneous MRI phantom filled with aqueous copper sulfate solution (diameter: 10 cm, 16 cm and 26 cm) was imaged. The MR examination position and protocol of FSE T1- and T2-weighted MRI were constant. The T1-weighted MRIs of the spine were obtained with a 450-500 ms TR range, a 15-20 ms TE , a 192x256 matrix one acquisition, a 3 ms excitation, a 30 cm field of view, a 3 mm slice thickness, a 1mm interslice gap, and an echo train length (ETL) of two. T2-weighted MRIs of the spine were obtained with a 2000-3000 ms TR range, a 80-120 ms TE range, a 192x256 matrix, one acquisition, a 3 ms excitation, a 30 cm field of view, a 3 mm slice thickness, a 1 mm intersection gap, and an echo train length (ETL) of two.

We analyzed the mean and variance of image signal intensities of the uniform phantom taken with and without SCIC. To calculate these values, the signal intensities of the phantom were measured by placing regions of interest (ROIs). The ROIs for each image was placed at an identical position and size for MRIs with and without SCIC. The variance to mean ratio of signal intensity was used to determine the significance of differences between T1- and T2-weighted MRIs with and without SCIC. For methods, see references [1,2].

2.2 Clinical experiments

We retrospectively evaluated the image records of 50 patients with 80 spinal metastases among them who were referred to our department of radiology between July 2011 and June 2014, and who signed an informed consent after full explanation of MR imaging technique. The study had been approved by the ethics committee of Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University. T1- and T2-weighted MRIs with and without SCIC were performed. The study group, including variable spinal metastases, consists of 26 men and 24 women, mean age 61years (range: 37-80 years). The origins of spinal metastases were identified as being lung cancer in 27 patients, breast cancer in 10 patients, prostate cancer in 6 patients, renal cancer in 4 patients, lymphoma in 1 patient, and ovarian cancer in 2 patents. 50 patients were diagnosed by pathology and biopsy. Two different methods of MRI were obtained at the same time in each patient.

FSE T1- and T2-weighted MRIs of human spine were performed with the same machine and same acquisition parameters as the phantom. Patients were placed in a supine position on the table platform with a standard phased-array surface coil. The pre-saturation technique was used with FSE images to reduce motion artifacts. We selected the No Phase Wrap (NPW) from the imaging option window to prevent aliasing of anatomy inside a prescribed field of view (FOV); Flow Compensation (FC) was used for gradient moment nulling to decrease flow artifacts.

MRI findings of both groups were evaluated in a random order. The two observers were blind to the patient's clinical information. If there was any disagreement of interpretation, a conference was held to reach a consensus. In the image analysis, MR imaging with and without SCIC were independently compared by two experienced radiologists. They assigned one of five numerical scores (0, non-visualization; 1, poor; 2, average; 3, good; 4, excellent) for the following two image characteristics: interface conspicuity of lesion, and details of lesion on MRI with and without SCIC. The Student’s t-test was used to determine the significance of differences of MRI with and without SCIC in numerical scores. A significance level of α=0.001 was selected.

Two experienced radiologists independently recorded the reading time per case from 50 cases. A conference was held to reach a consensus. Statistical differences were evaluated with the independent t-test; a P value of less than 0.05 was considered significant.

3 Results

3.1 Phantom studies

To more effectively assess the behavior of the SCIC method, a homogenous phantom was imaged. The original image without SCIC is inhomogeneous in signal intensity from the left to the right of the image, which is demonstrated by lower intensity values in the left region of the image than those in the right region, (Fig.1. A and C). The variance to mean ratio of signal intensity was 53.36% for original T1- weighted MRI and 53.58% for original T2- weighted MRI (Table 1, 2). The intensity variation from areas right to left is almost suppressed in the image with SCIC (Fig.1. B and D). The variance to mean ratio of signal intensity is reduced to 18.99% for T1-weighted MRI with SCIC and 22.77% for T2-weighted MRI with SCIC, (Table 1, 2). Thus, SCIC does improve the homogeneity of these images.

Figure 1 All these images were displayed at the same window and level, diameter 16cm phantom. (A) FSE T1-weighted MRI without SCIC. (B) T1-weighted MRI with SCIC. (C) FSE T2-weighted MRI without SCIC. (D) T2-weighted MRI with SCIC.
Figure 1

All these images were displayed at the same window and level, diameter 16cm phantom. (A) FSE T1-weighted MRI without SCIC. (B) T1-weighted MRI with SCIC. (C) FSE T2-weighted MRI without SCIC. (D) T2-weighted MRI with SCIC.

Table 1

Analysis of phantom T1-weighted MRI Quality

Image numberImage with SCIC MeanVarianceImage without SCIC MeanVariance
12677.58515.472665.151244.11
2699.51107.12701.47449.03
3289.4473.70290.52258.40
Mean1222.18232.101219.05650.51
Var/mean%18.99%53.36%

Image number 1, diameter 10 cm phantom; Image number 2, diameter 16 cm phantom; Image number 3, diameter 26 cm phantom.

Table 2

Analysis of phantom T2-weighted MRI Quality

Image numberImage with SCIC MeanVarianceImage without SCIC MeanVariance
11123.79271.781153.85546.00
2307.9550.24310.26193.00
3123.5032.10125.36112.58
Mean518.41118.04529.82283.86
Var/mean%22.77%53.58%

Image number 1, diameter 10 cm phantom; Image number 2, diameter 16 cm phantom; Image number 3, diameter 26 cm phantom.

3.2 Clinical experiments

Quality scores of MRI with and without SCIC are summarized in Table 3. Both scores of interface conspicuity and details of lesion with SCIC are significantly higher than those without SCIC. MR images with SCIC decrease intensity near the phased-array surface coil, reducing the areas of bright spots, (Fig.2-4).

Table 3

Comparison quality scores of MRI with and without SCIC (n=80 lesions)

Image characteristicsT1-weighted MRI

With SCIC
Without SCICT2-weighted MRI

With SCIC
Without SCIC
Interface conspicuity of lesion2.34±0.87[*]1.20±0.892.28±0.86[*]1.23±0.84
Details of lesion2.20±0.86[*]1.08±0.812.18±0.87[*]1.08±0.81

Grading system: 0, non-visualization; 1, poor; 2, average; 3, good; 4, excellent.

Figure 2 A 55-year-old man suffered from renal cell cancer with spinal metastasis. (A) In conventional FSE T1-weighted MRI, spinous process of thoracic vertebra (arrow) shows increased signal intensity, but the spinal metastasis margin and its extension is not clear. (B) In FSE T1-weighted MRI with SCIC, the margin and its extension of lesion is clearer than conventional MRI (arrow). (C) In conventional FSE T2-weighted MRI, lesion (arrow) is not clear. (D) In T2-weighted MRI with SCIC, the image clearly shows lesion in spinous process and soft tissue lesion around the vertebrae with more details (arrow)
Figure 2

A 55-year-old man suffered from renal cell cancer with spinal metastasis. (A) In conventional FSE T1-weighted MRI, spinous process of thoracic vertebra (arrow) shows increased signal intensity, but the spinal metastasis margin and its extension is not clear. (B) In FSE T1-weighted MRI with SCIC, the margin and its extension of lesion is clearer than conventional MRI (arrow). (C) In conventional FSE T2-weighted MRI, lesion (arrow) is not clear. (D) In T2-weighted MRI with SCIC, the image clearly shows lesion in spinous process and soft tissue lesion around the vertebrae with more details (arrow)

Figure 3 A 58-year-old man suffered from lung cancer with spinal metastasis. (A) In conventional FSE T1-weighted MRI, spinous process of thoracic vertebra (arrow) shows increased signal intensity, but the margin and its extension of lesion is not clear. (B) In T1-weighted MRI with SCIC, the image clearly shows spinal metastasis margin and its extension (arrow).
Figure 3

A 58-year-old man suffered from lung cancer with spinal metastasis. (A) In conventional FSE T1-weighted MRI, spinous process of thoracic vertebra (arrow) shows increased signal intensity, but the margin and its extension of lesion is not clear. (B) In T1-weighted MRI with SCIC, the image clearly shows spinal metastasis margin and its extension (arrow).

Figure 4 A 59-year-old man suffered from lung cancer with spinal metastasis. (A) In conventional FSE T1-weighted MRI, L1 vertebral body, spinous process and soft tissue lesion around the vertebrae (arrow) shows increased signal intensity, but the spinal metastases margin, extension and details are not clear. (B) In FSE T1-weighted MRI with SCIC, the margin and its details of lesion are clearer than conventional MRI (arrow). (C) In conventional FSE T2-weighted MRI, lesion (arrow) is not clear. (D) In T2-weighted MRI with SCIC, the image clearly shows spinal metastasis margin and its extension with more details (arrow).
Figure 4

A 59-year-old man suffered from lung cancer with spinal metastasis. (A) In conventional FSE T1-weighted MRI, L1 vertebral body, spinous process and soft tissue lesion around the vertebrae (arrow) shows increased signal intensity, but the spinal metastases margin, extension and details are not clear. (B) In FSE T1-weighted MRI with SCIC, the margin and its details of lesion are clearer than conventional MRI (arrow). (C) In conventional FSE T2-weighted MRI, lesion (arrow) is not clear. (D) In T2-weighted MRI with SCIC, the image clearly shows spinal metastasis margin and its extension with more details (arrow).

Compared with MRI without surface coil intensity correction, the reading time per case was shorter for MRI with surface coil intensity correction (Fig.5). The difference between them is significant.

Figure 5 The reading time per case. Compared with MRI without surface coil intensity correction, the reading time per case is shorter for MRI with surface coil intensity correction (p<0.05).
Figure 5

The reading time per case. Compared with MRI without surface coil intensity correction, the reading time per case is shorter for MRI with surface coil intensity correction (p<0.05).

4 Discussion

Development of high-field MRI scanners with excellent tissue contrast, non-invasive character, high spatial resolution, and easy slice selection at any orientation is a powerful imaging technique for studying spine disorders. However, the signal intensity variation, high signal intensity near the phased-array surface coil, and low signal intensity far from the surface coil limit their use in clinical applications [2-7]. There are several facts, including coil tuning, short repetition and echo times, RF coils with poor RF uniformity and number of echoes in a multi-echo sequence, can have significant effects on image non-uniformity, but the differences between transverse, sagittal and coronal non-uniformity are small [3]. The image quality in MRI is linked to the spatial homogeneity of the electromagnetic field produced by radio frequency (RF) coils. The presence of the non-uniformity in RF magnetic fields is a significant obstacle to quantitative analysis and image processing. Such non-uniformity can be caused by the instrument itself rather than by the objects being examined. One of the major sources of image non-uniformity in the high field MR scanners is the radiofrequency (RF) coil non-uniformity. It degrades conspicuity of lesion(s) and surrounding tissues in the MRI and reduces accuracy of image [4]. The phased-array surface coil image non-uniformity results from narrowing the signal intensity distribution due to reduction of signal drop-off as the distance from the coil increase. SCIC produces broadening of the intensity distribution of homogeneous phantom. The uniformity of MRI can be improved in scans of any orientation, including oblique scans at arbitrary angles [5]. Although the general development of the hardware of the MR scanners has reduced the non-uniformity, the problem remains unsolved. In this paper, we assessed the phantom and spine as exemplars in sagittal FSE T1- and T2-weighted MRI with and without SCIC on phased-array surface coil. The SCIC feature is an automatic post-processing technique that reduces noise, and enhances contrast in the image. It will produce a series of filtered images with decreased intensity near the coil, reducing the areas of bright spots. With SCIC, each scanned series produces two reconstructed series: one original and the other corrected using SCIC. SCIC was supported by all imaging planes and pulse sequences.

To better assess the behavior of the correction method, phantom was scanned to test the performance. A uniform spherical phantom filled with aqueous copper sulfate solution was used with phased-array surface coil. The original phantom T1-weighted MRI (Fig.1. A) and T2-weighted MRI (Fig.1. C) shows that the signal intensity variation in the image is visible. There is a diminution of signal intensity from areas near phased-array surface coil to areas far from surface coil (from right to left in image). Both measurements were taken with regions of interest of same size. The coefficient of variation of signal intensity of an image is an important index of image quality. The coefficient of variation of signal intensity with SCIC is lower than original image without correction. The variance to mean ratio of signal intensity is 53.36% for original T1- weighted MRI and 53.58% for original T2- weighted MRI. The intensity variation from areas right to left was almost suppressed in the image with SCIC (Fig.1. B and D). The variance to mean ratio of signal intensity is reduced to 18.99% for T1- weighted MRI with SCIC and 22.77% for T2- weighted MRI with SCIC.

This intensity non-uniformity image is attributed to poor radio frequency (RF) coil uniformity. In phased-array surface coil images, the spread of the distribution will vary with the depth from the coil due to signal drop-off.

This intensity non-uniformity image not only makes the image interpretation difficult, but also restricts the use of phased-array surface coils in some areas of interest, such as display spinous process and soft tissue lesion around the vertebrae. In addition, this intensity fall-off causes distracting artifacts and makes it difficult to film images at a single window/level setting. Images without correction show non-uniformity and a high signal level from subcutaneous tissue near the coils [6]. The SCIC was found to significantly improve the image homogeneity near high contrast edges when a phased-array surface coil is used for reception. The intensity correction algorithm should remove the effect of the variance due to signal drop-off. Phased-array surface coil images with correction demonstrated greater anatomic detail [7].

The phased-array surface coil has an inherently non-uniformity reception profile that leads to a variation in image brightness across the spine. This significantly degrades the utility of the images for evaluation of anatomy in the spine and can also impair automated segmentation of spine structures. For example, the uncorrected T1-weighted images, the signal in the spine near the coils is much higher than spine far from the phased-array surface coils. This wide variation in image intensity makes most of the anatomy difficult to visualize with a single window and level setting (Fig.2-4). The MRI without SCIC was morphologically unacceptable before SCIC. After SCIC, the difference between the two results with and without SCIC is evident (Fig.2-4). There is significant difference between the original images and the SCIC images. The corrected phased-array surface coil images showed areas both near and far away from the surface coil with relatively constant contrast and higher SNR for area far away from the surface coil compared to the original non- corrected images. The signal intensity near surface coil was more homogeneous than and not as bright as the original one. In corrected surface coil image, a diminution of the SNR near the surface coil will not hamper the quantitative analysis of the superficial tissues of the spine, because it is too bright originally. In other words, the non-uniformity correction will make easier the automatic segmentation of different tissues near and far from the surface coil.

The comparison between MRI with and without SCIC is summarized in Table 3. The overall image quality scores (interface conspicuity of lesion and details of lesion) are significantly higher than those original MRI. The average reading time per case is shorter for MRI with SCIC than for MRI without SCIC. The advantages to the SCIC method include: a. MRI with SCIC shows the spinal lesions more clearly than the original one; b. MRI with SCIC does not have any additional time; c. The reading time per case is shorter for MRI with SCIC than those without SCIC, while its disadvantage is that MRI is susceptible to complex patient motion [8-15].

In conclusion, the results in this study showed that spinal metastases were better demonstrated on MRI with SCIC than on MRI with conventional surface coil technique. Their overall image quality scores were significantly higher than those of original MRI. The reading time per case was shorter for MRI with SCIC than for MRI without SCIC. Therefore, we consider that the phased-array surface coil intensity correction could replace or supplement conventional technology in MRI of spinal metastases.

Acknowledgement

This study was partly supported by the grants from the Medical and Health Research Project of Zhejiang Province (No.2013KYA109 and No.2015KYB210), Zhejiang Province Natural Fund (No.LY16H060002).

  1. Conflict of interest: No authors report any conflict of interest.

References

[1] Han C, Hatsukami TS, Yuan C. A multi-scale method for automatic correction of intensity non-uniformity in MR images. J Magn Reson Imaging 2001; 13:428-43610.1002/jmri.1062Suche in Google Scholar PubMed

[2] Yun S, Kyriakos WE, Chung JY, Han Y, Yoo SS, Park H. Projection-based estimation and nonuniformity correction of sensitivity profiles in phased-array surface coils. J Magn Reson Imaging 2007; 25: 588-59710.1002/jmri.20826Suche in Google Scholar PubMed

[3] Ishimori Y, Yamada K, Kimura H, Fujiwara Y, Yamaguchi I, Monma M, et al. Correction of inhomogeneous RF field using multiple SPGR signals for high-field spin-echo MRI. Magn Reson Med Sci 2007; 6: 67-7310.2463/mrms.6.67Suche in Google Scholar PubMed

[4] Wang D, Doddrell DM. Method for a detailed measurement of image intensity nonuniformity in magnetic resonance imaging. Med Phys 2005; 32: 952-96010.1118/1.1869572Suche in Google Scholar PubMed

[5] Wicks DA, Barker GJ, Tofts PS. Correction of intensity nonuniformity in MRI of any orientation. Magn Reson Imaging 1993; 11:183-19610.1016/0730-725X(93)90023-7Suche in Google Scholar

[6] Tofts PS, Barker GJ, Simmons A, MacManus DG, Thorpe J, Gass A, et al. Correction of nonuniformity in images of the spine and optic nerve from fixed receive-only surface coils at 1.5 T. J Comput Assist Tomogr 1994; 18:997-100310.1097/00004728-199411000-00033Suche in Google Scholar PubMed

[7] Lin FH, Chen YJ, Belliveau JW, Wald LL. A wavelet-based approximation of surface coil sensitivity profiles for correction of image intensity inhomogeneity and parallel imaging reconstruction. Hum Brain Mapp 2003; 19:96-11110.1002/hbm.10109Suche in Google Scholar PubMed PubMed Central

[8] Martin CT, Morcuende J, Buckwalter JA, Miller BJ. Prereferral MRI use in patients with musculoskeletal tumors is not excessive. Clin Orthop Relat Res 2012; 470: 3240-324510.1007/s11999-012-2394-4Suche in Google Scholar PubMed PubMed Central

[9] Vokurka EA, Watson NA, Watson Y, Thacker NA, Jackson A. Improved high resolution MR imaging for surface coils using automated intensity non-uniformity correction: feasibility study in the orbit. J Magn Reson Imaging 2001; 14:540-54610.1002/jmri.1217Suche in Google Scholar PubMed

[10] Cohen MS, DuBois RM, Zeineh MM. Rapid and effective correction of RF inhomogeneity for high field magnetic resonance imaging. Hum Brain Mapp 2000; 10:204-21110.1002/1097-0193(200008)10:4<204::AID-HBM60>3.0.CO;2-2Suche in Google Scholar

[11] Zhou LQ, Zhu YM, Bergot C, Laval-Jeantet AM, Bousson V, Laredo JD, et al. A method of radio-frequency inhomogeneity correction for brain tissue segmentation in MRI. Comput Med Imaging Graph 2001; 25:379-38910.1016/S0895-6111(01)00006-4Suche in Google Scholar PubMed

[12] Gispert JD, Reig S, Pascau J, Vaquero JJ, García-Barreno P, Desco M. Method for bias field correction of brain T1-weighted magnetic resonance images minimizing segmentation error. Hum Brain Mapp 2004; 22: 133-14410.1002/hbm.20013Suche in Google Scholar PubMed PubMed Central

[13] Noterdaeme O, Anderson M, Gleeson F, Brady SM. Intensity correction with a pair of spoiled gradient recalled echo images. Phys Med Biol 2009; 54: 3473-348910.1088/0031-9155/54/11/013Suche in Google Scholar PubMed

[14] Collewet G, Davenel A, Toussaint C, Akoka S. Correction of intensity nonuniformity in spin-echo T(1)-weighted images. Magn Reson Imaging 2002; 20:365-37310.1016/S0730-725X(02)00502-7Suche in Google Scholar PubMed

[15] Sosna J, Pedrosa I, Dewolf WC, Mahallati H, Lenkinski RE, Rofsky NM. MR imaging of the prostate at 3 Tesla: comparison of an external phased-array coil to imaging with anendorectal coil at 1.5 Tesla. Acad Radiol 2004; 11: 857-86210.1016/j.acra.2004.04.013Suche in Google Scholar PubMed

Received: 2017-2-21
Accepted: 2017-3-24
Published Online: 2017-5-20

© 2017 Hong Ren et al.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.

Artikel in diesem Heft

  1. Regular Articles
  2. Intravascular treatment of left subclavian artery aneurysm coexisting with aortic coarctation in an adult patient
  3. Regular Articles
  4. Effect of electrical stimulation on blood flow velocity and vessel size
  5. Regular Articles
  6. Live birth pregnancy outcome after first in vitro fertilization treatment in a patient with Systemic Lupus Erythematosus and isolated high positive IgA anti-β2glycoprotein I antibodies: a case report
  7. Regular Articles
  8. Periodontal ligament stem cells regulate apoptosis of neutrophils
  9. Regular Articles
  10. Platelet-rich fibrin (PRF) in implants dentistry in combination with new bone regenerative flapless technique: evolution of the technique and final results
  11. Regular Articles
  12. The significance of strong ion gap for predicting return of spontaneous circulation in patients with cardiopulmonary arrest
  13. Regular Articles
  14. Clinicopathology of EpCAM and EGFR in human epithelial ovarian carcinoma
  15. Regular Articles
  16. Intraosseous lipoma of the mandibula: A case report and review of the literature
  17. Regular Articles
  18. Transurethral resection of the prostate, bladder explosion and hyponatremic encephalopathy: a rare case report of malpractice
  19. Regular Articles
  20. Brain strokes related to aortic aneurysma – the analysis of three cases
  21. Regular Articles
  22. Effect of Bicyclol tablets on drug induced liver injuries after kidney transplantation
  23. Regular Articles
  24. Plasma free fatty acids in hyperemesis gravidarum pregnancy
  25. Regular Articles
  26. Impact of chromosomal rearrangement upon DNA methylation patterns in leukemia
  27. Regular Articles
  28. Gefitinib versus docetaxel in treated non-small-cell lung cancer: a meta-analysis
  29. Regular Articles
  30. The clinical characteristics of patients with chronic idiopathic anal pain
  31. Regular Articles
  32. Bone tunnel impaction reduced the tibial tunnel enlargement
  33. Regular Articles
  34. Effects of S-1 combined with radiotherapy in the treatment of nasopharyngeal cancer: a meta-analysis based on randomized controlled trials
  35. Regular Articles
  36. Predictions and outcomes of atrial fibrillation in the patients with acute myocardial infarction
  37. Regular Articles
  38. An accuracy study of the Intracavitary Electrocardiogram (IC-ECG) guided peripherally inserted central catheter tip placement among neonates
  39. Regular Articles
  40. Serum CA125, CA199 and CEA combined detection for epithelial ovarian cancer diagnosis: A meta-analysis
  41. Regular Articles
  42. Surface coil intensity correction in magnetic resonance imaging in spinal metastases
  43. Regular Articles
  44. Muscle stem cell and physical activity: what point is the debate at?
  45. Regular Articles
  46. MicroRNA let-7g directly targets forkhead box C2 (FOXC2) to modulate bone metastasis in breast cancer
  47. Regular Articles
  48. Monitoring health inequalities at the municipal level: Lithuanian experience
  49. Regular Articles
  50. Role of Epstein-Barr virus in the development of nasopharyngeal carcinoma
  51. Regular Articles
  52. Thrombectomy combined with indwelling-catheter thrombolysis is more effective than pure thrombectomy for the treatment of lower extremity deep venous thrombosis
  53. Regular Articles
  54. Expression of Hepcidin and Neogenin in colorectal cancer
  55. Regular Articles
  56. Carnitine and adiponectin levels in breast cancer after radiotherapy
  57. Regular Articles
  58. Pathophysiology of meningioma growth in pregnancy
  59. Regular Articles
  60. Causal neuro-immune relationships at patients with chronic pyelonephritis and cholecystitis. Correlations between parameters EEG, HRV and white blood cell count
  61. Regular Articles
  62. Measuring efficiency of secondary healthcare providers in Slovenia
  63. Regular Articles
  64. Galectin-3 expression in colorectal cancer and its correlation with clinical pathological characteristics and prognosis
  65. Regular Articles
  66. Model for studying anti- allergic drugs for allergic conjunctivitis in animals
  67. Regular Articles
  68. Barriers perceived by nurses in the optimal treatment of postoperative pain
  69. Regular Articles
  70. Tumor microenvironment in treatment of glioma
  71. Regular Articles
  72. Delirium risk of dexmedetomidine and midazolam in patients treated with postoperative mechanical ventilation: A meta-analysis
  73. Regular Articles
  74. Hemangioma of the rib: a rare case report and literature review
  75. Regular Articles
  76. The diagnostic accuracy of conventional forceps biopsy compared to ESD
  77. Regular Articles
  78. Increased miR-25 expression in serum of gastric cancer patients is correlated with CA19-9 and acts as a potential diagnostic biomarker
  79. Regular Articles
  80. Therapeutic nanomedicine surmounts the limitations of pharmacotherapy
  81. Regular Articles
  82. Relationship between PD-L1 expression and clinical characteristics in patients with breast invasive ductal carcinoma
  83. Regular Articles
  84. Trypsinogen activation peptide induces HMGB1 release from rat pancreatic acinar cells
  85. Regular Articles
  86. The effective regulation of pro- and anti-inflammatory cytokines induced by combination of PA-MSHA and BPIFB1 in initiation of innate immune responses
  87. Regular Articles
  88. Cell based therapeutic approach in vascular surgery: application and review
  89. Regular Articles
  90. Clinical efficacy of alprostadil combined with α-lipoic acid in the treatment of elderly patients with diabetic nephropathy
  91. Regular Articles
  92. Professional burnout and concurrent health complaints in neonatal nursing
  93. Regular Articles
  94. Esophageal xanthoma: presence of M2 macrophages suggests association with late inflammatory and reparative processes
  95. Regular Articles
  96. Cone beam computed tomography analysis in 3D position of maxillary denture
  97. Regular Articles
  98. CK20 mRNA expression in serum as a biomarker for colorectal cancer diagnosis: A meta-analysis
  99. Regular Articles
  100. Serum AFU, 5’-NT and AFP as biomarkers for primary hepatocellular carcinoma diagnosis
  101. Regular Articles
  102. Absolute reliability and concurrent validity of hand held dynamometry and isokinetic dynamometry in the hip, knee and ankle joint: systematic review and meta-analysis
  103. Regular Articles
  104. The Fountain of Youth: A tale of parabiosis, stem cells, and rejuvenation
  105. Regular Articles
  106. Foam sclerotherapy during shunt surgery for portal hypertension and varices
  107. Regular Articles
  108. Insomnia and depression: Japanese hospital workers questionnaire survey
  109. Regular Articles
  110. Serum NF-κBp65, TLR4 as biomarker for diagnosis of preeclampsia
  111. Regular Articles
  112. Docetaxel/cisplatin therapy in myasthenia gravis with hypertension/diabetes
  113. Regular Articles
  114. Fluid resuscitation and markers of glycocalyx degradation in severe sepsis
  115. Regular Articles
  116. Modified Sauve-Kapandji procedure for patients with old fractures of the distal radius
  117. Regular Articles
  118. Bile leakage after liver transplantation
  119. Regular Articles
  120. VEGF overexpression predicts poor survival in hepatocellular carcinoma
  121. Regular Articles
  122. Galen vein aneurysm– challenge for treatment
  123. Regular Articles
  124. Retrieval of a broken sewing needle from the sacrum aided by a permanent magnet: a case report and literature review
  125. Regular Articles
  126. HIV/STI prevention interventions: A systematic review and meta-analysis
  127. Regular Articles
  128. Aortic aneurysm as a complication of myeloperoxidase-antineutrophil cytoplasmic antibody-associated vasculitis
  129. Regular Articles
  130. Real-time monitoring of contrast-enhanced ultrasound for radio frequency ablation
  131. Regular Articles
  132. Successful drug-eluting stent implantation in a male patient with dextrocardia: a case report
  133. Regular Articles
  134. Primary pleomorphic liposarcoma of fallopian tube with recurrence: a case report and review of the literature
  135. Regular Articles
  136. Color Doppler Ultrasound in Uterine Arterial Embolization
  137. Regular Articles
  138. Pattern of alcohol consumption by young people from North Eastern Portugal
  139. Regular Articles
  140. Effects of out-of-hospital continuing nursing on schizophrenia patients' rehabilitation and quality of life
Heruntergeladen am 5.11.2025 von https://www.degruyterbrill.com/document/doi/10.1515/med-2017-0021/html
Button zum nach oben scrollen