An accuracy study of the Intracavitary Electrocardiogram (IC-ECG) guided peripherally inserted central catheter tip placement among neonates

  • Lian-juan Zhou , Hong-zhen Xu EMAIL logo , Mei-fang Xu , Yan Hu und Xiao-Fang Lou
Veröffentlicht/Copyright: 22. Mai 2017

Abstract

Objective

To explore the clinical application of the intracavitary electrocardiogram (IC-ECG) guided Peripherally Inserted Central Catheter (PICC) tip placement among neonates.

Background

the ECGs of neonates are difficult to perform and their wave shapes are of doubtful accuracy due to various interfering factors

Method

115 neonates were admitted to perform PICC guided by IC-ECG. Logistic regression was performed to analyze all possible influencing factors of the accuracy from the tip placement. The puncture site of the PICC, gestational age, height, weight, basal P/R amplitude and positioning P/R amplitude might be related to the accuracy of IC-ECG location.

Result

The accuracy in the lower extremity was higher than that in the upper extremity. Multivariate logistic regression analysis showed that the weight (Odds Ratio (OR)=1.93, 95%Confidence Interval(CI):1.06-3.50) and positioning P/R amplitude (OR=32.33, 95%CI: 2.02-517.41) are statistically significant risks to the accuracy PICC tip placement.

Conclusions

Possible methods to improve the accuracy might be Catheterizing through lower extremity, keeping the neonates calm, enhancing the electrocardiogram signal and strengthening technical training. Therefore it is practical to perfrom a tip placement by the dynamic change in the P waves from an electrocardiogram (ECG) guided PICC among neonates and as reliable as using X-rays.

1 Introduction

The peripherally inserted central catheter (PICC) is a common infusion device in clinical settings that provides an ideal vein tunnel for successful treatment in critical neonates. The placement of the PICC tip plays a vital role in the effectiveness of the PICC [1]. A substantial body of research has demonstrated that locating the tip placement by the dynamic change in the P wave from an electrocardiogram (ECG) is as reliable as using X-rays [2, 3]. Studies in China have successfully used ECGs to locate the PICC tip placement in an upper extremity [4-6], but study of the intracavitary electrocardiogram (IC-ECG)-guided lower extremities is lacking. Study is underway to develop such an application, motivated by the fact that the ECGs of neonates are difficult to perform and their wave shapes are of doubtful accuracy due to various interfering factors [7, 8]. To further explore the application of the IC-ECG-guided PICC tip placement among neonates, In present study we designed to analyze the accuracy of the IC-ECG from the 115 neonates.

2 Material and methods

2.1 Neonate inclusion and exclusion criteria

Inclusion criteria: 115 neonates admitted to the Children’s Hospital affiliated to the School of Medicine, from January 2015 to July 2016 who underwent PICC placement, of which 79 were upper extremity cases and 36 lower extremity cases. Exclusion criteria: obviously abnormal body surface ECG, known severe and complex congenital heart disease, coagulation dysfunction, neck and shoulder thoracic deformity, wearing of a heart pacemaker, wearing of dynamic ECG inspection equipment, unclear basal ECG obtained during monitoring. This study was approved by the ethics committee of the hospital; all guardians or parents were told of the study objective and signed an informed consent form.

2.2 Catheterization preparation

Two nurses, both qualified to implement PICCs, inserted the catheters (Medcomp 1.9F PICC catheters) at the bedside, applying a blinding puncture, without the use of a routine guide wire.

2.3 Catheterization procedure

The neonate subjects were kept in a supine body position and their PICC placement procedures were monitored using a three-lead monitoring system (PM-8000, Mindray,China). 30 minutes before catheterization, Luminal IV bolus 5mg/kg was administered to keep the neonates calm. Then, for each subject, the vein puncture site was cleaned with pieces of cotton dampened with 75% ethanol, and after the ethanol had volatilized, three electrode pads were affixed to body surface areas below the left subclavian, the right subclavian and the lower left abdomen, respectively. A surface lead-II ECG was recorded with an output speed equal to 25mm/s and a sensitivity of 10mm/mV. Next, the surface ECG records were printed out for comparison. After successful routine puncture, 10ml 0.9% NaCl (normal saline, NS) was used to disinfect the catheter heparin cap. An adequate amount of saline via a 10-ml syringe was used to pre-rinse a scalp needle (size 7), and then half the length of the scalp needle was inserted through the heparin cap. An electrode labeled with the R logo (clip type) formerly placed in the right subclavian and topically sterilized with alcohol swabs, then was connected to the remaining half of the scalp needle. A syringe pre-loaded with 10ml NS was used to draw blood from the subject, and NS was then used to pulse-wash the catheter. After the catheter was judged to be unobstructed, a lead-II IC-ECG was recorded, observing the amplitude changes of the P wave and QRS wave group. Then, the catheter was gradually retracted until the maximum P wave occurred in the veins of the upper limb. Next, when the P wave’s amplitude was approximately 50% to 60% of the R wave’s, the catheter was fastened and an ECG was recorded and printed out. The catheter was again fastened when the P wave’s amplitude matched that of the R-wave’s in the veins of the upper limb.

If the ECG failed to appear or its wave shape exhibited disorder during the process, the catheter was withdrawn and the process to observe the changes in the ECG was repeated. Neonates who cried heavily were calmed again when necessary. After the insertion was complete, the catheter was washed and sealed with NS and a 5 U/ml dilute solution of heparin sodium in a routine manner, and an X-ray examination was also performed.

2.4 Standard of optimal tip placement

For an ECG to be judged as stable during the PICC placement, the baseline of the ECG must be constant, the waveforms of the P wave and QRS wave must be clear and easy to identify, and the P wave must have measurable amplitude. The stability of the ECGs was evaluated by a physician of the ECG Department. According to the 2016 guidelines of the American Infusion Nurses Society [9], if the veins of the catheter tip in the upper limb are placed at the junction of the lower portion of the superior vena cava (SVC) and the right atrium, the tip is regarded as optimally settled. However, if the catheters tip is located elsewhere, outside the SVC, it is defined as ectopic. If the veins of the catheter tip in the lower limb are located in or above the flat diaphragm of the inferior vena cava [10],which cannot access the atrium, this is optimal for the requirements of the central vein following the X-ray procedure; any other placement should be relocated.

2.5 Statistical analysis

All of the data analysis was performed using SPSS20.0. The t-test and chi-square test were used for univariate analysis appropriately, and logistic regression was used for multivariate analysis. Two sides of P<0.05 was selected as the significance level.

3 Results

3.1 Subject main characteristics

A total of 115 subjects met the inclusion and exclusion criteria, including 64 males (55.7%) and 51 females (44.3%). Among them, 79 were catheterized through veins of an upper extremity and 36 through veins of a lower extremity. 41 neonates were injected with 5mg/kg Luminal to maintain adequate pacification during the process. 7 cases were catheterized through the right superficial temporal vein, 2 cases through the left superficial temporal vein, 6 cases through the right external jugular vein, 8 cases through the left external jugular vein, 22 cases through the right axillary vein, 24 cases through the left axillary vein, 8 cases through the left basilica vein, 2 cases through the right basilica vein, 11 cases through the left lower extremity and 25 cases through the right lower extremity (including 1 case through the right popliteal fossa, 2 cases through the right saphenous vein and 22 cases through the right femoral vein). All PICC tips (100%) were found to be placed in the vena cava after location, and no ectopia occurred.

3.2 Accuracy rates comparison between upper and lower extremity.

The accuracy rates of the IC-ECG-guided PICC in upper and lower extremities were 59/79 (74.7%) and 33/36 (91.6%) respectively. The accuracy of the tip placement in the lower extremity in the neonates was thus higher than that in the upper extremity (χ2=4.46, P=0.035).

3.3 Factor analyses of the IC-ECG guided PICC tip placement.

The accuracy of the process was taken as the dependent variable (1= unstable,0=stable), and the gestational age, chest circumference, height, weight, body surface measurement, basal P/R amplitude, basal heart rate, length of immersion, positioning heart rate and positioning P/R amplitude were taken as the independent variables to perform a univariate analysis. The results indicated that the gestational age, height, weight, basal P/R amplitude and positioning P/R amplitude had a statistically significant effect on the IC-ECG-guided PICC tip placement (Table 1).

Table 1

Factors of subject characteristics on tip placement according to the accuracy:[Mean(Standard Deviation)/frequency(percentage)]

Factoraccuracyt/Z/X2[*]P
YESNO
Puncture site[*]Lower extremity33 (91.6)3(8.4)4.460.035
Upper extremity59 (74.7)20 (25.3)
SexFemale43 (84.3)8 (15.7)1.490.222
Male48 (75.0)16 (25.0)
Gestational age[*] (weeks)35.42(3.40)37.00(3.44)1.9820.049
Chest circumference (cm)31.41(3.502)32.71(3.391)1.6280.106
Height (cm)47.12(4.235)49.23(5.593)2.0200.046
Weight (kg)2.629(.8394)3.014(1.1091)1.8620.065
BSM (cm)11.11(3.582)10.83(3.933)0.3310.741
BasicP/R.2254(.10118).2733(.14513)1.7670.080
Immersion length (cm)11.592(3.7351)11.463(4.0998)1.4920.139
Baseline heart rate b/m130.84(17.236)129.96(10.564)0.2380.8122
Positioning heart rate b/m137.63(14.487)135.67(12.534)0.6050.546
Positioing P/R[*].4631(.20457).5976(.23975)2.6260.0099

BSM, body surface measurement;

Taking the statistically significant influencing variables identified above as independent variables to perform multiple-variable logistic regression analysis, two variables, i.e., weight and positioning P/R amplitude, were included in the final model (Table 2).

Table 2

Multivariate logistic analysis of the accuracy of IC-ECG from PICC

FactorOR 95%CL
CoefficientSEWaldPORLCIUCI
Weight.656.3054.637.0311.9271.0613.501
Positioning P/R3.4761.4156.037.01432.3312.020517.407

SE, Standard Error; OR, Odds Ratio; LCL, Lower Confidence Interval; UCL, Upper Confidence Interval

4 Discussion

4.1 Higher accuracy of the IC-ECG guided positioning technique through the lower extremity than that through the upper extremity

In this study, the univariate analysis found that the IC-ECG-guided positioning technique achieved significantly different accuracies in the upper extremity and lower extremity. The reason for the greater accuracy of the technique in the lower extremity is that, compared with the superior vena cava, the inferior vena cava is straighter, less branched and has fewer influencing variables. When positioning the puncture in the femoral veins of the lower extremity, changes in body position, such as buckling or extending, do not strongly affect the PICC tip placement; in contrast, body position changes in the upper extremity can greatly disturb the tip placement.

The P wave increases in amplitude as the PICC tip in the upper extremity approaches the junction of the SVC and the right atrium, necessitating a slight retraction of the catheter to maintain the proper position [11]; however, this repositioning is difficult to perform accurately. The ECG waves in the lower extremity differ from those of the upper extremity: as the catheter reaches a greater depth, the amplitudes of the P wave and QRS waves remain either unchanged or atactic. The latter indicates invagination of the catheter or its misplacement into the hepatic or spleen area, which necessitate pulsing with NS and observation of the change in amplitude of the QRS waves after retraction of the catheter. Regular QRS waves with the same amplitude as the body surface indicate that the catheter has not yet arrived at the inferior vena cava. After arrival of the catheter at the inferior vena cava, the QRS waves continually increase in amplitude as the catheter approaches the heart. Meanwhile, the appearance of an upright P wave indicates that the PICC tip has been placed at the level of the right atrium; in this case, the catheter should be retracted to its normal position, inside the deep inferior vena cava but outside the right atrium [12]. Wave changes in lower extremity placement are clearer, more distinct and easier to accurately manipulate than those in upper extremities.

4.2 Main influencing factors of the accuracy for PICC tip placement

Multivariate model analysis indicated that the weight of the neonates was related to the accuracy: the heavier the neonate, the lower the accuracy. This may be related to the noisiness of the neonates: premature infants have relatively low mobility and are easily calmed in a nestlike special therapy bed, especially after injection with small doses of sedatives. The heavier neonates were all term infants with relatively high mobility; theycried more readily and the effect of sedative injection in small doses was less pronounced. This is consistent with the fact that placement is more accurate in taller and younger neonates according to the invariance analysis, which may be due to the relationship between height, gestational age and weight. 70 neonates in this study had high mobility and cried heavily; they were pacified patiently and given a pacifier to suck, or fed sweetened water in advance of the procedure. 41 neonates could not be pacified in this way; they were injected with small doses of Luminal following doctors’ advice. When they became calm and their heart rates reverted to normal, the amplitudes of the P waves could be read correctly. During the actual procedure, the ECGs of the heavier neonates were easier to obtain, and their waves were also clearer, when performing the procedure in quiet surroundings.

Model analysis also indicated that the positioning P/R amplitude was related to accuracy: the higher the amplitude, the lower the accuracy. It is advisable for the tip placement to be retracted slightly when the P wave reaches maximum amplitude, especially when a small Q wave occurs [5, 13]. In our study, placement was made when P/R reached 50-60%. There were 20 cases of non-standard placement, including 12 cases where the P/R was higher than 50-60% due to the PICC tip penetrating the right atrium. During the actual procedure, it would have been preferable for an assistant to measure the stability of the P/R with a measuring ruler to provide a basis for correct placement, rather than relying on visual inspection, which no doubt caused errors.

Neonates are susceptible to Patent Ductus Arteriosus (PDA), Atrial Septal Defect (ASD) and Ventricular Septal Defect (VSD); although the body surface ECGs of affected neonates are almost normal, and they can receive ECG placement in the veins, the accuracy of placement is reduced. 16 cases in this study presented PDA, ASD or VSD upon examination, including 5 neonates who were diagnosed with congenital heart disease; the ECG baselines of these 16 remained unstable, with the P waves increasing in amplitude. 4 of the 16 cases received accurate placement, 2 cases received placement in the middle of the SVC, and 10 cases in the right atrium. The overall accuracy of only 74.7% was partly affected by the above conditions, which caused difficulties in determining the P/R amplitude. Consequently, heart B ultrasonography should be performed before PICC puncturing to exclude congenital heart disease and influencing variables of the P/R amplitude, and to increase placement accuracy in neonates.

4.3 Other influencing factors of the accuracy for PICC tip placement

One of the most important factors in ECG is the signal-transferring connection between the electrode and the steel needle. Movement of the catheter affects this connection, thus compromising the stability of the ECG recording; therefore, it is necessary to ensure slow catheter movement to obtain a stable and reliable ECG recording when the tip enters the vena cava [2, 14, 15].

Some studies have indicated that connecting a guide-wire to the electrode vein facilitates ECG recording in adults and children [7, 16-18]. Possibly, because a 1.9rPICC is relatively thin, the path length of the electrocardiographic leads is very long; thus, the electrocardio signal must travel a long distance to diffuse throughout the blood, NS, scalp needle and connection wire, and the signal is degraded by this long path length. However, neonates have a heart rate of 120-160 beats/minute, which influences the stability and clearness of the ECG records. It is necessary to keep neonates calm and minimize all sources of interference.

In this study, the concordance rate of the PICC and the X-ray in the upper extremity of the neonates was lower than that in children [11]. Extraction of the waveform is vital; when the waveform could not be extracted, we examined whether conduction was smooth, the heparin cap was filled with NS, and the needle was in contact with the NS. Furthermore, the needle tip should be perpendicular to the heparin cap and not in contact with the vessel wall, to avoid influencing conduction. The connection between the electrode and the scalp needle may influence the spread of the signals. Having connected the leads, it is inadvisable to observe and record ECGs during catheter placement because the movement of the catheter influences the stability of the lead connection, making it difficult to obtain accurate data in such a short time. Therefore, we suggest that when the catheter has entered into the SVC, the further positioning of the catheter should proceed slowly, with regular stops to examine the stability of the connection and record reliable ECG waves. It is important to monitor the basal heart rate during observation of the P wave, because a stable heart rate is necessary for a stable ECG.

5 Conclusion

During PICC tip placement, it is vital to record a stable and clear ECG. Establishing steady leads and supplying a continual injection of NS can improve the ECG signals. In clinical applications of this technique, interference from environmental signals or factors compromising the smooth operation of the procedure should be eliminated to ensure a stable vein ECG wave and increase the accuracy of tip placement. Tip placement by the dynamic change in the P waves from an electrocardiogram (ECG) guided PICC among neonates is accessible and as reliable as using X-rays.


Tel: 0086 57186670066

  1. Declaration of conflicting interests: The authors have declared that no competing interests

Aknolwdegements

We thank the members of the Department of Department of Nursing, Children’s Hospital, Zhejiang University School of Medicine who participated in the study and everyone who contributed to this project.

References

[1] rui-yi Cg-yWh-qZ. Research progress of the tip placement of peripherally inserted central venous catheters. Chinese Journal of Nursing 2011;46:1037-1039Suche in Google Scholar

[2] Smith B, Neuharth RM, Hendrix MA, McDonnall D, Michaels AD. Intravenous electrocardiographic guidance for placement of peripherally inserted central catheters. Journal of electrocardiology 2010;43:274-27810.1016/j.jelectrocard.2010.02.003Suche in Google Scholar PubMed

[3] Sette P, Azzini AM, Dorizzi RM, Castellano G. Serendipitous ECG guided PICC insertion using the guidewire as intra-cardiac electrode. J Vasc Access 2010;11:7210.1177/112972981001100116Suche in Google Scholar PubMed

[4] Hui YAO, Min S, Yuying LIU. Clinical research of intravenous electrocardiogram guided PICC tip position. Chinese Journal of Nursing 2011;46:748-750Suche in Google Scholar

[5] Zhao L, Cao X, Chen C, Wang Y, Jin X, Chen X, et al. The specific P-waveform of intracavitary ECG method for positioning the tip of valved peripherally inserted central catheters. Chinese Journal of Nursing 2015;50:1374-1378Suche in Google Scholar

[6] Hui-min FB-lS-yS-j. The changes and role of intracavitary electrocardiogram in the placement of peripherally inserted central venous catheters. Chinese Journal of Nursing 2010;45:26-28Suche in Google Scholar

[7] Rossetti F, Pittiruti M, Lamperti M, Graziano U, Celentano D, Capozzoli G. The intracavitary ECG method for positioning the tip of central venous access devices in pediatric patients: results of an Italian multicenter study. J Vasc Access 2015;16:137-14310.5301/jva.5000281Suche in Google Scholar PubMed

[8] Perin G. PICC Placement in the Neonate. New Engl J Med 2014;370:2153-215410.1056/NEJMc1404381Suche in Google Scholar PubMed

[9] Pittiruti M, Bertollo D, Briglia E, Buononato M, Capozzoli G, De Simone L, et al. The intracavitary ECG method for positioning the tip of central venous catheters: results of an Italian multicenter study. J Vasc Access 2012;13:357-36510.5301/JVA.2012.9020Suche in Google Scholar PubMed

[10] Tanabe H, Takahashi T, Murayama R, Yabunaka K, Oe M, Matsui Y, et al. Using Ultrasonography for Vessel Diameter Assessment to Prevent Infiltration. Journal of infusion nursing: the official publication of the Infusion Nurses Society 2016;39:105-11110.1097/NAN.0000000000000159Suche in Google Scholar PubMed

[11] Jain A, Deshpande P, Shah P. Peripherally inserted central catheter tip position and risk of associated complications in neonates. Journal of perinatology : official journal of the California Perinatal Association 2013;33:307-31210.1038/jp.2012.112Suche in Google Scholar PubMed

[12] Tsui BCH, Richards GJ, Van Aerde J. Umbilical vein catheterization under electrocardiogram guidance. Pediatr Anesth 2005;15:297-30010.1111/j.1460-9592.2005.01433.xSuche in Google Scholar PubMed

[13] Neubauer AP. Percutaneous Central Iv Access In the Neonate - Experience with 535 Silastic Catheters. Acta Paediatr 1995;84:756-76010.1111/j.1651-2227.1995.tb13750.xSuche in Google Scholar PubMed

[14] Ortiz-Miluy G, Sanchez-Guerra C. [Intracavitary electrocardiogram during the insertion of peripherally inserted central catheters]. Enfermeria clinica 2013;23:148-15310.1016/j.enfcli.2013.05.002Suche in Google Scholar PubMed

[15] Oliver G, Jones M. Evaluation of an electrocardiograph-based PICC tip verification system. British journal of nursing 2013;22:S24-2810.12968/bjon.2013.22.Sup9.S24Suche in Google Scholar PubMed

[16] jing Zh-gZw-hWx-jYqXg-xL. The application of a special salty electrode on the ECG positioning of a peripherally inserted central venous catheter. Journal of Nurses Training 2013;28:835-836Suche in Google Scholar

[17] Sette P, Dorizzi RM, Azzini AM. Unexpected obstacle in catheter removal could be due to knots spontaneously forming in peripherally inserted central catheters (PICC). J Vasc Access 2011;12:387-38810.5301/JVA.2011.8326Suche in Google Scholar PubMed

[18] Pittiruti M, La Greca A, Scoppettuolo G. The electrocardiographic method for positioning the tip of central venous catheters. J Vasc Access 2011;12:280-29110.5301/JVA.2011.8381Suche in Google Scholar PubMed

Received: 2016-5-10
Accepted: 2017-4-11
Published Online: 2017-5-22

© 2017 Lian-juan Zhou et al.

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.

Artikel in diesem Heft

  1. Regular Articles
  2. Intravascular treatment of left subclavian artery aneurysm coexisting with aortic coarctation in an adult patient
  3. Regular Articles
  4. Effect of electrical stimulation on blood flow velocity and vessel size
  5. Regular Articles
  6. Live birth pregnancy outcome after first in vitro fertilization treatment in a patient with Systemic Lupus Erythematosus and isolated high positive IgA anti-β2glycoprotein I antibodies: a case report
  7. Regular Articles
  8. Periodontal ligament stem cells regulate apoptosis of neutrophils
  9. Regular Articles
  10. Platelet-rich fibrin (PRF) in implants dentistry in combination with new bone regenerative flapless technique: evolution of the technique and final results
  11. Regular Articles
  12. The significance of strong ion gap for predicting return of spontaneous circulation in patients with cardiopulmonary arrest
  13. Regular Articles
  14. Clinicopathology of EpCAM and EGFR in human epithelial ovarian carcinoma
  15. Regular Articles
  16. Intraosseous lipoma of the mandibula: A case report and review of the literature
  17. Regular Articles
  18. Transurethral resection of the prostate, bladder explosion and hyponatremic encephalopathy: a rare case report of malpractice
  19. Regular Articles
  20. Brain strokes related to aortic aneurysma – the analysis of three cases
  21. Regular Articles
  22. Effect of Bicyclol tablets on drug induced liver injuries after kidney transplantation
  23. Regular Articles
  24. Plasma free fatty acids in hyperemesis gravidarum pregnancy
  25. Regular Articles
  26. Impact of chromosomal rearrangement upon DNA methylation patterns in leukemia
  27. Regular Articles
  28. Gefitinib versus docetaxel in treated non-small-cell lung cancer: a meta-analysis
  29. Regular Articles
  30. The clinical characteristics of patients with chronic idiopathic anal pain
  31. Regular Articles
  32. Bone tunnel impaction reduced the tibial tunnel enlargement
  33. Regular Articles
  34. Effects of S-1 combined with radiotherapy in the treatment of nasopharyngeal cancer: a meta-analysis based on randomized controlled trials
  35. Regular Articles
  36. Predictions and outcomes of atrial fibrillation in the patients with acute myocardial infarction
  37. Regular Articles
  38. An accuracy study of the Intracavitary Electrocardiogram (IC-ECG) guided peripherally inserted central catheter tip placement among neonates
  39. Regular Articles
  40. Serum CA125, CA199 and CEA combined detection for epithelial ovarian cancer diagnosis: A meta-analysis
  41. Regular Articles
  42. Surface coil intensity correction in magnetic resonance imaging in spinal metastases
  43. Regular Articles
  44. Muscle stem cell and physical activity: what point is the debate at?
  45. Regular Articles
  46. MicroRNA let-7g directly targets forkhead box C2 (FOXC2) to modulate bone metastasis in breast cancer
  47. Regular Articles
  48. Monitoring health inequalities at the municipal level: Lithuanian experience
  49. Regular Articles
  50. Role of Epstein-Barr virus in the development of nasopharyngeal carcinoma
  51. Regular Articles
  52. Thrombectomy combined with indwelling-catheter thrombolysis is more effective than pure thrombectomy for the treatment of lower extremity deep venous thrombosis
  53. Regular Articles
  54. Expression of Hepcidin and Neogenin in colorectal cancer
  55. Regular Articles
  56. Carnitine and adiponectin levels in breast cancer after radiotherapy
  57. Regular Articles
  58. Pathophysiology of meningioma growth in pregnancy
  59. Regular Articles
  60. Causal neuro-immune relationships at patients with chronic pyelonephritis and cholecystitis. Correlations between parameters EEG, HRV and white blood cell count
  61. Regular Articles
  62. Measuring efficiency of secondary healthcare providers in Slovenia
  63. Regular Articles
  64. Galectin-3 expression in colorectal cancer and its correlation with clinical pathological characteristics and prognosis
  65. Regular Articles
  66. Model for studying anti- allergic drugs for allergic conjunctivitis in animals
  67. Regular Articles
  68. Barriers perceived by nurses in the optimal treatment of postoperative pain
  69. Regular Articles
  70. Tumor microenvironment in treatment of glioma
  71. Regular Articles
  72. Delirium risk of dexmedetomidine and midazolam in patients treated with postoperative mechanical ventilation: A meta-analysis
  73. Regular Articles
  74. Hemangioma of the rib: a rare case report and literature review
  75. Regular Articles
  76. The diagnostic accuracy of conventional forceps biopsy compared to ESD
  77. Regular Articles
  78. Increased miR-25 expression in serum of gastric cancer patients is correlated with CA19-9 and acts as a potential diagnostic biomarker
  79. Regular Articles
  80. Therapeutic nanomedicine surmounts the limitations of pharmacotherapy
  81. Regular Articles
  82. Relationship between PD-L1 expression and clinical characteristics in patients with breast invasive ductal carcinoma
  83. Regular Articles
  84. Trypsinogen activation peptide induces HMGB1 release from rat pancreatic acinar cells
  85. Regular Articles
  86. The effective regulation of pro- and anti-inflammatory cytokines induced by combination of PA-MSHA and BPIFB1 in initiation of innate immune responses
  87. Regular Articles
  88. Cell based therapeutic approach in vascular surgery: application and review
  89. Regular Articles
  90. Clinical efficacy of alprostadil combined with α-lipoic acid in the treatment of elderly patients with diabetic nephropathy
  91. Regular Articles
  92. Professional burnout and concurrent health complaints in neonatal nursing
  93. Regular Articles
  94. Esophageal xanthoma: presence of M2 macrophages suggests association with late inflammatory and reparative processes
  95. Regular Articles
  96. Cone beam computed tomography analysis in 3D position of maxillary denture
  97. Regular Articles
  98. CK20 mRNA expression in serum as a biomarker for colorectal cancer diagnosis: A meta-analysis
  99. Regular Articles
  100. Serum AFU, 5’-NT and AFP as biomarkers for primary hepatocellular carcinoma diagnosis
  101. Regular Articles
  102. Absolute reliability and concurrent validity of hand held dynamometry and isokinetic dynamometry in the hip, knee and ankle joint: systematic review and meta-analysis
  103. Regular Articles
  104. The Fountain of Youth: A tale of parabiosis, stem cells, and rejuvenation
  105. Regular Articles
  106. Foam sclerotherapy during shunt surgery for portal hypertension and varices
  107. Regular Articles
  108. Insomnia and depression: Japanese hospital workers questionnaire survey
  109. Regular Articles
  110. Serum NF-κBp65, TLR4 as biomarker for diagnosis of preeclampsia
  111. Regular Articles
  112. Docetaxel/cisplatin therapy in myasthenia gravis with hypertension/diabetes
  113. Regular Articles
  114. Fluid resuscitation and markers of glycocalyx degradation in severe sepsis
  115. Regular Articles
  116. Modified Sauve-Kapandji procedure for patients with old fractures of the distal radius
  117. Regular Articles
  118. Bile leakage after liver transplantation
  119. Regular Articles
  120. VEGF overexpression predicts poor survival in hepatocellular carcinoma
  121. Regular Articles
  122. Galen vein aneurysm– challenge for treatment
  123. Regular Articles
  124. Retrieval of a broken sewing needle from the sacrum aided by a permanent magnet: a case report and literature review
  125. Regular Articles
  126. HIV/STI prevention interventions: A systematic review and meta-analysis
  127. Regular Articles
  128. Aortic aneurysm as a complication of myeloperoxidase-antineutrophil cytoplasmic antibody-associated vasculitis
  129. Regular Articles
  130. Real-time monitoring of contrast-enhanced ultrasound for radio frequency ablation
  131. Regular Articles
  132. Successful drug-eluting stent implantation in a male patient with dextrocardia: a case report
  133. Regular Articles
  134. Primary pleomorphic liposarcoma of fallopian tube with recurrence: a case report and review of the literature
  135. Regular Articles
  136. Color Doppler Ultrasound in Uterine Arterial Embolization
  137. Regular Articles
  138. Pattern of alcohol consumption by young people from North Eastern Portugal
  139. Regular Articles
  140. Effects of out-of-hospital continuing nursing on schizophrenia patients' rehabilitation and quality of life
Heruntergeladen am 25.1.2026 von https://www.degruyterbrill.com/document/doi/10.1515/med-2017-0019/html
Button zum nach oben scrollen