This issue of Clinical Chemistry and Laboratory Medicine includes a report by Pradella and colleagues regarding a cooperative multicentre study to define the upper limits of normal for several tests used in the diagnostic assessment of lupus anticoagulant (LA) [1]. In total, 200 normal samples were assessed, comprising 40 from each of five centres, for a total of six functional LA assays. Each centre used the same tests, dilute Russell viper venom time (dRVVT) (screen and confirm), and silica clotting time (SCT), the same reagents and the same type of instrument from a single manufacturer. This is a notable report for several reasons, apart an interesting observational exercise around the concepts of standardisation (use of common products, procedures, processes, and practices) and harmonisation (the process whereby different analytic systems are determined to provide clinically similar results) of LA testing.
The study explores an interesting approach to a common laboratory problem – undertaking studies with sufficient numbers of normal individuals to establish a critical cut-off value that will help define patients ‘with’ or ‘without’ LA. This is an important component of the laboratory evaluation of the antiphospholipid syndrome (APS) [2, 3], in turn having significant implications for patient management [2–4]. In brief, patients identified clinically to have APS, as diagnostically aided using laboratory assessment by LA, can be subjected to long-term anticoagulant therapy, which has significant implications for both lifestyle changes as well as bleeding and thrombosis risk. In essence, an incorrect diagnosis will have significant clinical and social implications; a false-positive diagnosis of APS may subject a patient to unnecessarily long-term anticoagulant therapy with risk of bleeding, whereas a false-negative diagnosis may mean the patient is not appropriately anti-coagulated and therefore at future risk of thrombosis.
Current LA guidelines [5] suggest that laboratories use ‘at least 40 adult healthy donors’ and take ‘the cut-off as the value above the 99th percentile of the distribution’. This has caused some contention in the field, given that a statistically valid 99th percentile evaluation of a non-Gaussian distributed normal population would require a minimum of nearly 400 samples [6]. Pradella and colleagues [1] comparatively evaluated the data obtained in each assay, as well as assay ratios, in each center, as well as the composite of all study data. Notably, when assay data was expressed in seconds, some test results showed a normal Gaussian distribution, whereas others did not, and there were also some statistically significant differences in results obtained between centers. In contrast, when results were expressed as normalised ratios, the test results were normally distributed and there were no longer any statistically significant differences in data between centres. Thus, better agreement between centres, and in essence standardisation and harmonisation, was obtained when data was interpreted in terms of normalised ratios.
In LA testing, there is actually very limited consensus in terms of inter-laboratory processes. Thus, different laboratories use different test procedures [e.g., based on dRVVT, SCT, activated partial thromboplastin time (APTT), kaolin clotting time (KCT) and/or other tests], different test panels (e.g., APTT±dRVVT±KCT±SCT, etc.), different test approaches [testing of neat plasma vs. testing of mixed plasma (i.e., patient plus normal), sometimes using different test plasma:normal plasma mixtures], different ways of interpreting test results (e.g., test ratios [division], test differences [subtraction], percent corrections, Rosner Index), and different cut-off values [6–9].
Of interest, Pradella and colleagues [1] determined a cut-off value of 1.22 for the dRVVT normalised ratio. This value is actually very close to that value ‘nominally’ indicated on many dRVVT product information sheets, as well as generally used by laboratories. For example, Figure 1A shows data from the Australasian College of Pathologists of Australasia (RCPA) Haematology Quality Assurance Program (QAP) [10]. Most laboratories report a negative LA interpretation when they identify a normalised LA ratio below 1.2, and most report an LA positive finding for LA ratios above 1.2. There is some inter-laboratory variation, however, in interpretations of LA positivity vs. negativity when ratio values are very close to the cut-off value of 1.2; thus, some laboratories will call these negative, and some will call these positive or equivocal (‘borderline’ or ‘indeterminant’).
![Figure 1
Representative data from the RCPA QAP to illustrate important points of discussion; data capture years=2009–2011 inclusive, for all or select representative cross-laboratory tested samples.
(A) Comparison of participant reported normalised dRVVT ratio vs. interpretation (all samples). (B) Participant reported normalised dRVVT ratios for select samples representing clearly LA positive (LUP9-03b), clearly LA negative (LUP10-08a), or weak LA positive samples providing equivocal findings from participants (LUP9-08a and LUP10-03a). (C) Participant reported interpretations for samples identified in Figure B. (D) Inter-laboratory CVs vs. median of reported normalised dRVVT ratios for LA testing (all samples). (E) Inter-laboratory CVs vs. median of reported values for aCL IgG testing contrasted to normalised dRVVT ratio for LA testing (all samples). For additional detail regarding findings from the RCPA QAP in this area of testing see elsewhere [10, 11].](/document/doi/10.1515/cclm-2012-0578/asset/graphic/j_cclm-2012-0578_fig_001.jpg)
Representative data from the RCPA QAP to illustrate important points of discussion; data capture years=2009–2011 inclusive, for all or select representative cross-laboratory tested samples.
(A) Comparison of participant reported normalised dRVVT ratio vs. interpretation (all samples). (B) Participant reported normalised dRVVT ratios for select samples representing clearly LA positive (LUP9-03b), clearly LA negative (LUP10-08a), or weak LA positive samples providing equivocal findings from participants (LUP9-08a and LUP10-03a). (C) Participant reported interpretations for samples identified in Figure B. (D) Inter-laboratory CVs vs. median of reported normalised dRVVT ratios for LA testing (all samples). (E) Inter-laboratory CVs vs. median of reported values for aCL IgG testing contrasted to normalised dRVVT ratio for LA testing (all samples). For additional detail regarding findings from the RCPA QAP in this area of testing see elsewhere [10, 11].
Indeed, the cut-off value – what value is assigned, how it is determined and subsequently the interpretation of test results close to the cut-off value – often causes the greatest problems in LA testing. This can be illustrated by Figures 1B and 1C, again using data from the RCPA QAP [10]. Thus, samples with a normalised dRVVT ratio above 1.3 are invariably considered LA positive by participants of this program, although the grade of positivity designated may differ between laboratories (see data for sample LUP9-03b as representative example); similarly, samples with a normalised dRVVT ratio below 1.1 are invariably considered LA negative by participants of this program (see data for sample LUP10-08a as representative example). However, when a sample with a normalised dRVVT ratio close to 1.2 is tested by participants, a mixture of interpretations (LA-negative and LA-positive) is always returned; this is dependant on inter-laboratory assay variation, the normalised dRVVT ratio actually obtained by the laboratory, and whether this value is above or below their own designated cut-off value (albeit recognising that this value will be close to 1.2 for most laboratories; see data for samples LUP9-08a and LUP10-03a as representative examples).
This brings this discussion to two other critical considerations in LA testing, and namely inter-assay variability and standardisation/harmonisation. As noted at the start of this commentary, the report from Pradella and colleagues [1] identified a statistically significant inter-laboratory difference in some test times when these were expressed in seconds, but not when expressed as ratios. In this study, all laboratories used the same tests, the same reagents and the same type of instrument. Notably, when laboratories use different tests or different reagents or instruments, one would expect a larger variation in test results, reflecting a relative lack of standardisation or harmonisation. When results are expressed as a ratio, such as a normalised dRVVT ratio, the resultant inter-laboratory coefficients of variation (CVs) range from approximately 5%–20% (see Figure 1D for examples from the RCPA QAP). Interestingly, comparative data for solid phase antiphospholipid antibody (aPL) testing, which represents another facet of testing in APS [2, 3, 7], shows substantially worse inter-laboratory CVs ranging from approximately 30% to over 200% (see Figure 1E for examples from the RCPA QAP; [11]).
All of the discussed elements above suggest that although LA testing is considered not well standardised or harmonised [12], it is on comparably better standardised than solid phase aPL testing. This may also help to explain why LA is more strongly associated with adverse clinical events, such as thrombosis than other aPL, such as anticardiolipin or anti-beta-2-glycoprotein I antibodies [13, 14]. In essence, this means that we can be better assured about the test outcome and clinical ‘quality’ of LA testing by dRVVT than that of aPL testing by solid phase assays. Nevertheless, much remains to be done. For example, we need to train clinicians better in terms of test requests; to avoid inappropriate testing that is both costly and which may lead to false-positive diagnosis of APS and adverse patient treatments [15, 16]. We also need to develop better assays for APS diagnosis [12]. Additionally, but certainly not finally, we need to continue to improve harmonisation and standardisation in laboratory testing for all APS associated tests including LA [17].
Conflict of interest statement
Author’s conflict of interest disclosure: The author states that there are no conflicts of interest regarding the publication of this article.
Research funding: None declared.
Employment or leadership: None declared.
References
1. Pradella P, Azzarini G, Santarossa L, Caberlotto L, Bardin C, Poz A, et al. Cooperation experience in a multicenter study to define the upper limits in a normal population for the diagnostic assessment of the functional Lupus Anticoagulant assays. Clin Chem Lab Med 2013;51:379–85.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000314999000029&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1515/cclm-2012-0382Suche in Google Scholar PubMed
2. Favaloro EJ, Wong RC. The antiphospholipid syndrome: a large elephant with many parts or an elusive chameleon disguised by many colours? Autoimmunity Highlights 2010;1:5–14.10.1007/s13317-010-0003-7Suche in Google Scholar PubMed PubMed Central
3. Favaloro EJ, Wong RC. The antiphospholipid syndrome: diagnosis, pathogenesis, laboratory testing, and management. Semin Thromb Hemost 2012;38:299–304.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600001&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Suche in Google Scholar PubMed
4. Les I, Ruiz-Irastorza G, Khamashta MA. Intensity and duration of anticoagulation therapy in antiphospholipid syndrome. Semin Thromb Hemost 2012;38:339–47.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600006&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Suche in Google Scholar PubMed
5. Pengo V, Tripodi A, Reber G, Rand JH, Ortel TL, Galli M, et al. Subcommittee on lupus anticoagulant/antiphospholipid antibody of the scientific and standardisation committee of the international society on thrombosis and haemostasis. update of the guidelines for lupus anticoagulant detection. subcommittee on lupus anticoagulant/antiphospholipid antibody of the scientific and standardisation committee of the international society on thrombosis and haemostasis. J Thromb Haemost 2009;7:1737–40.Suche in Google Scholar PubMed
6. Kershaw G, Suresh S, Orellana D, Nguy YM. Laboratory identification of lupus anticoagulants. Semin Thromb Hemost 2012;38:375–84.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600010&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1055/s-0032-1311991Suche in Google Scholar PubMed
7. Favaloro EJ, Wong R. Laboratory testing for the Antiphospholipid Syndrome: making sense of antiphospholipid antibody assays. Clin Chem Lab Med 2011;49:447–61.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000288238500013&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1515/CCLM.2011.064Suche in Google Scholar PubMed
8. Tripodi A. To mix or not to mix in lupus anticoagulant testing? That is the question. Semin Thromb Hemost 2012;38:385–9.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600011&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Suche in Google Scholar PubMed
9. Favaloro EJ, Wong RC. Current clinical and laboratory practice for the investigation of the antiphospholipid syndrome: Findings from the 2008 Australasian antiphospholipid antibody survey. Pathology 2009;41:666–75.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000273212200012&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.3109/00313020903257731Suche in Google Scholar PubMed
10. Favaloro EJ, Bonar R, Marsden K. Internal quality control and external quality assurance in testing for antiphospholipid antibodies: Part II – Lupus Anticoagulant. Semin Thromb Hemost 2012;38:404–11.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600013&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1055/s-0032-1311993Suche in Google Scholar PubMed
11. Favaloro EJ, Wheatland L, Jovanovich S, Roberts-Thomson P, Wong RC. Internal quality control and external quality assurance in testing for antiphospholipid antibodies: Part I: anti-cardiolipin and anti-β2-glycoprotein I antibodies. Semin Thromb Hemost 2012;38:390–403.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600012&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1055/s-0032-1311990Suche in Google Scholar PubMed
12. de Groot PG, Urbanus RT. The future of antiphospholipid antibody testing. Semin Thromb Hemost 2012;38: 412–20.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600014&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1055/s-0032-1304715Suche in Google Scholar PubMed
13. Pengo V, Banzato A, Bison E, Bracco A, Denas G, Ruffatti A. What have we learned about antiphospholipid syndrome from patients and antiphospholipid carrier cohorts? Semin Thromb Hemost 2012;38:322–7.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600003&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1055/s-0032-1304719Suche in Google Scholar PubMed
14. Galli M. Interpretation and recommended testing for antiphospholipid antibodies. Semin Thromb Hemost 2012;38:348–52.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600007&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Suche in Google Scholar PubMed
15. Mameli A, Barcellona D, Vannini ML, Marongiu F. High frequency of inadequate test requests for antiphospholipid antibodies in daily clinical practice. Clin Chem Lab Med 2011;49:695–8.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000289464200018&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1515/CCLM.2011.111Suche in Google Scholar PubMed
16. Favaloro EJ, Reben R, Mohammed S, Koutts J. A clinical audit of antiphospholipid antibody testing in tertiary practice. Towards improved relevance in thrombophilia investigations? Intern Med 2012;42:427–34.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000302796000020&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Suche in Google Scholar
17. Willis R, Harris EN, Pierangeli SS. Current international initiatives in antiphospholipid antibody testing. Semin Thromb Hemost 2012;38:360–74.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304352600009&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1055/s-0032-1313564Suche in Google Scholar PubMed
©2013 by Walter de Gruyter Berlin Boston
Artikel in diesem Heft
- Letters to the Editor
- Performance evaluation of three different immunoassays for detection of antibodies to hepatitis B core
- Serum homocysteine concentrations in Chinese children with autism
- Interchangeability of venous and capillary HbA1c results is affected by oxidative stress
- Interference of hemoglobin (Hb) N-Baltimore on measurement of HbA1c using the HA-8160 HPLC method
- First human isolate of Mycobacterium madagascariense in the sputum of a patient with tracheobronchitis
- Protein S and protein C measurements should not be undertaken during vitamin K antagonist therapy
- α2-HS glycoprotein is an essential component of cryoglobulin associated with chronic hepatitis C
- An unusual interference in CK MB assay caused by a macro enzyme creatine phosphokinase (CK) type 2 in HIV-infected patients
- An automated technique for the measurement of the plasma glutathione reductase activity and determination of reference limits for a healthy population
- Is osteopontin stable in plasma and serum?
- Evidence-based approach to reducing perceived wasteful practices in laboratory medicine
- Masthead
- Masthead
- Editorials
- Testing volume is not synonymous of cost, value and efficacy in laboratory diagnostics
- Lessons from controversy: biomarkers evaluation
- Commercial immunoassays in biomarkers studies: researchers beware!1)
- Trials and tribulations in lupus anticoagulant testing
- Reviews
- Mass spectrometry: a revolution in clinical microbiology?
- Chronic Chagas disease: from basics to laboratory medicine
- General Clinical Chemistry and Laboratory Medicine
- Shop for quality or quantity? Volumes and costs in clinical laboratories
- Minor improvement of venous blood specimen collection practices in primary health care after a large-scale educational intervention
- Evaluation of high resolution gel β2-transferrin for detection of cerebrospinal fluid leak
- Serum kallikrein-8 correlates with skin activity, but not psoriatic arthritis, in patients with psoriatic disease
- Soluble urokinase plasminogen activator receptor (suPAR) in the assessment of inflammatory activity of rheumatoid arthritis patients in remission
- Bone mass density selectively correlates with serum markers of oxidative damage in post-menopausal women
- Validation of a fast and reliable liquid chromatography-tandem mass spectrometry (LC-MS/MS) with atmospheric pressure chemical ionization method for simultaneous quantitation of voriconazole, itraconazole and its active metabolite hydroxyitraconazole in human plasma
- Performance of different screening methods for the determination of urinary glycosaminoclycans
- Intestinal permeability and fecal eosinophil-derived neurotoxin are the best diagnosis tools for digestive non-IgE-mediated cow’s milk allergy in toddlers
- An internal validation approach and quality control on hematopoietic chimerism testing after allogeneic hematopoietic cell transplantation
- Serum levels of IgG antibodies against oxidized LDL and atherogenic indices in HIV-1-infected patients treated with protease inhibitors
- Cooperation experience in a multicentre study to define the upper limits in a normal population for the diagnostic assessment of the functional lupus anticoagulant assays
- Contribution of procoagulant phospholipids, thrombomodulin activity and thrombin generation assays as prognostic factors in intensive care patients with septic and non-septic organ failure
- Suitability of POC lactate methods for fetal and perinatal lactate testing: considerations for accuracy, specificity and decision making criteria
- Point-of-care testing on admission to the intensive care unit: lactate and glucose independently predict mortality
- Reference Values and Biological Variations
- CA125 reference values change in male and postmenopausal female subjects
- Distributions and ranges of values of blood and urinary biomarker of inflammation and oxidative stress in the workers engaged in office machine manufactures: evaluation of reference values
- Cancer Diagnostics
- Association of acute phase protein-haptoglobin, and epithelial-mesenchymal transition in buccal cancer: a preliminary report
- Comparison of diagnostic and prognostic performance of two assays measuring thymidine kinase 1 activity in serum of breast cancer patients
- Evaluation of the BRAHMS Kryptor® Thyroglobulin Minirecovery Test in patients with differentiated thyroid carcinoma
Artikel in diesem Heft
- Letters to the Editor
- Performance evaluation of three different immunoassays for detection of antibodies to hepatitis B core
- Serum homocysteine concentrations in Chinese children with autism
- Interchangeability of venous and capillary HbA1c results is affected by oxidative stress
- Interference of hemoglobin (Hb) N-Baltimore on measurement of HbA1c using the HA-8160 HPLC method
- First human isolate of Mycobacterium madagascariense in the sputum of a patient with tracheobronchitis
- Protein S and protein C measurements should not be undertaken during vitamin K antagonist therapy
- α2-HS glycoprotein is an essential component of cryoglobulin associated with chronic hepatitis C
- An unusual interference in CK MB assay caused by a macro enzyme creatine phosphokinase (CK) type 2 in HIV-infected patients
- An automated technique for the measurement of the plasma glutathione reductase activity and determination of reference limits for a healthy population
- Is osteopontin stable in plasma and serum?
- Evidence-based approach to reducing perceived wasteful practices in laboratory medicine
- Masthead
- Masthead
- Editorials
- Testing volume is not synonymous of cost, value and efficacy in laboratory diagnostics
- Lessons from controversy: biomarkers evaluation
- Commercial immunoassays in biomarkers studies: researchers beware!1)
- Trials and tribulations in lupus anticoagulant testing
- Reviews
- Mass spectrometry: a revolution in clinical microbiology?
- Chronic Chagas disease: from basics to laboratory medicine
- General Clinical Chemistry and Laboratory Medicine
- Shop for quality or quantity? Volumes and costs in clinical laboratories
- Minor improvement of venous blood specimen collection practices in primary health care after a large-scale educational intervention
- Evaluation of high resolution gel β2-transferrin for detection of cerebrospinal fluid leak
- Serum kallikrein-8 correlates with skin activity, but not psoriatic arthritis, in patients with psoriatic disease
- Soluble urokinase plasminogen activator receptor (suPAR) in the assessment of inflammatory activity of rheumatoid arthritis patients in remission
- Bone mass density selectively correlates with serum markers of oxidative damage in post-menopausal women
- Validation of a fast and reliable liquid chromatography-tandem mass spectrometry (LC-MS/MS) with atmospheric pressure chemical ionization method for simultaneous quantitation of voriconazole, itraconazole and its active metabolite hydroxyitraconazole in human plasma
- Performance of different screening methods for the determination of urinary glycosaminoclycans
- Intestinal permeability and fecal eosinophil-derived neurotoxin are the best diagnosis tools for digestive non-IgE-mediated cow’s milk allergy in toddlers
- An internal validation approach and quality control on hematopoietic chimerism testing after allogeneic hematopoietic cell transplantation
- Serum levels of IgG antibodies against oxidized LDL and atherogenic indices in HIV-1-infected patients treated with protease inhibitors
- Cooperation experience in a multicentre study to define the upper limits in a normal population for the diagnostic assessment of the functional lupus anticoagulant assays
- Contribution of procoagulant phospholipids, thrombomodulin activity and thrombin generation assays as prognostic factors in intensive care patients with septic and non-septic organ failure
- Suitability of POC lactate methods for fetal and perinatal lactate testing: considerations for accuracy, specificity and decision making criteria
- Point-of-care testing on admission to the intensive care unit: lactate and glucose independently predict mortality
- Reference Values and Biological Variations
- CA125 reference values change in male and postmenopausal female subjects
- Distributions and ranges of values of blood and urinary biomarker of inflammation and oxidative stress in the workers engaged in office machine manufactures: evaluation of reference values
- Cancer Diagnostics
- Association of acute phase protein-haptoglobin, and epithelial-mesenchymal transition in buccal cancer: a preliminary report
- Comparison of diagnostic and prognostic performance of two assays measuring thymidine kinase 1 activity in serum of breast cancer patients
- Evaluation of the BRAHMS Kryptor® Thyroglobulin Minirecovery Test in patients with differentiated thyroid carcinoma