Suitability of POC lactate methods for fetal and perinatal lactate testing: considerations for accuracy, specificity and decision making criteria

  • Jean-Luc Orsonneau , François Fraissinet , Véronique Sébille-Rivain , Daniele Dudouet and Edith Bigot-Corbel EMAIL logo
Published/Copyright: August 3, 2012

Abstract

Background: Point of care (POC) testing for fetal scalp blood lactate is a more user friendly and more successful approach compared to scalp pH for intrapartum fetal monitoring. The aim of this study was to assess the analytical specificity and clinical reliability of three POC lactate methods.

Methods: The analytical performance of three POC lactate methods was compared to Cobas 6000 (Roche Diagnostics) laboratory reference method: Lactate Pro from Arkray, GEM 4000 from Instrumentation Laboratory and StatStrip Lactate from Nova Biomedical. The clinical performance and influences on accuracy and decision making criteria for the three POC methods was assessed with umbilical cord samples and compared to the laboratory reference method. The influence of varying ranges of hemoglobin, pH and partial oxygen pressure (pO2) on the accuracy of results was assessed.

Results: Although all three POC methods showed good correlation with the reference method for the umbilical cord sample population (r=0.989, 0.973 and 0.980, respectively), Lactate Pro and Gem 4000 showed a significant negative bias compared to the reference method. The degree of bias meant a significant readjustment of decision making criteria was required for fetal lactate use. The accuracy of the Lactate Pro results was affected by hemoglobin and to a lesser extent pH.

Conclusions: The three electrochemical POC devices can measure fetal lactate reliably. StatStrip Lactate showed a closer correlation and concordance to our laboratory reference method. The results of this study indicate the requirement for predetermining the reliability of POC lactate methods before use present in fetal and perinatal settings.


Corresponding author: Edith Bigot-Corbel, Laboratoire de Biochimie Générale, Pôle de Biologie, CHU de Nantes, 9 Quai Moncousu, 44035 Nantes Cedex, France, Phone: +33 240084004, Fax: +33 240165455

Authorship: All authors confirmed they have contributed to the intellectual content of this paper and have met the following three requirements: 1) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; 2) drafting or revising the article for intellectual content; and 3) final approval of the published article. JLO, FF and DD performed research, JLO, VS and EBC analyzed data and helped write the manuscript. JLO and EBC designed and supervised research, analyzed data and wrote the manuscript. The study was supported by furniture of StatStrip Lactate by Gael Gousset and Andreii Maliic from Nova Biomedical.

Conflict of interest statement

Authors’ conflict of interest disclosure: The authors stated that there are no conflicts of interest regarding the publication of this article.

Research funding: None declared.

Employment or leadership: None declared.

Honorarium: None declared.

References

1. Strehlow MC. Early identification of shock in critically ill patients. Emerg Med Clin North Am 2010;28:57–66.10.1016/j.emc.2009.09.006Search in Google Scholar PubMed

2. Jansen TC, van Bommel J, Woodward R, Mulder PG, Bakker J. Association between blood lactate levels, Sequential Organ Failure Assessment subscores, and 28-day mortality during early and late intensive care unit stay: a retrospective observational study. Crit Care Med 2009;37:2369–74.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000268160200006&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1097/CCM.0b013e3181a0f919Search in Google Scholar PubMed

3. Arnold RC, Shapiro NI, Jones AE, Schorr C, Pope J, Casner E, et al. Multicenter study of early lactate clearance as a determinant of survival in patients with presumed sepsis. Shock 2009;32:35–9.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000267081400006&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1097/SHK.0b013e3181971d47Search in Google Scholar PubMed

4. Jansen TC, van Bommel J, Mulder P, Lima A, van der Hoven B, Rommes J, et al. Prognostic value of blood lactate levels: does the clinical diagnosis at admission matter? J Trauma 2009;66:377–85.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000263442800012&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1097/TA.0b013e3181648e2fSearch in Google Scholar PubMed

5. Ramanah R, Martin A, Riethmuller D, Maillet R, Schaal JP. Value of fetal scalp lactate sampling during labour: a comparative study with scalp pH. Gynecol Obstet Fertil 2005;33:107–12.10.1016/j.gyobfe.2005.01.004Search in Google Scholar PubMed

6. Wiberg-Itzel E, Lipponer C, Norman M, Herbst A, Prebensen D, Hansson A, et al. Determination of pH or lactate in fetal scalp blood in management of intrapartum fetal distress: randomised controlled multicentre trial. British Med J 2008;336:1284–90.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000256705900031&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Search in Google Scholar

7. East CE, Leader LR, Sheehan P, Henshall NE, Colditz PB. Intrapartum fetal scalp lactate sampling for fetal assessment in the presence of a non-reassuring fetal heart rate trace. Cochrane Database Syst Rev 2010;3:CD006174.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000275717200013&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Search in Google Scholar

8. Ridenour RV, Gada RP, Brost BC, Karon BS. Comparison and validation of point of care lactate meters as a replacement for fetal pH measurement. Clin Biochem 2008;41:1461–5.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000261558800007&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1016/j.clinbiochem.2008.08.088Search in Google Scholar PubMed

9. Dungan K, Chapman J, Braithwaite SS, Buse J. Glucose measurement: confounding issues in setting targets for inpatient management. Diabetes Care 2007;30: 403–9.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000244024100039&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.2337/dc06-1679Search in Google Scholar PubMed

10. Tang Z, Lee J, Louie R, Kost G. Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med 2000;124: 1135–40.10.5858/2000-124-1135-EODHLOSearch in Google Scholar PubMed

11. Tang Z, Du X, Louie RF, Kost GJ. Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyser. Am J Clin Path 2000;113:75–86.10.1309/QAW1-X5XW-BVRQ-5LKQSearch in Google Scholar PubMed

12. Borruto F, Comparetto C, Wegher E, Treisser A. Screening of foetal distress by assessment of umbilical cord lactate. Clin Exp Obstet Gynecol 2006;33:219–22.Search in Google Scholar

13. Gjerris AC, Staer-Jensen J, Jorgensen JS, Bergholt T, Nickelsen C. Umbilical cord blood lactate: a valuable tool in the assessment of fetal metabolic acidosis. Eur J Obstet Gynecol Reprod Biol 2008;139:16–20.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000258042600003&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1016/j.ejogrb.2007.10.004Search in Google Scholar PubMed

14. Dessolle L, Lebrec J, Daraï E. Impact of delayed arterial cord blood sampling for lactate assay: a prospective observational study. Neonatology 2009;95:224–9.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000261292700006&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1159/000165378Search in Google Scholar PubMed

15. Chevillon I, Larrose C, Moreau N, Orsonneau JL. Conservation des échantillons de sang avant analyse des paramètres biochimiques les plus courants. Ann Biol Clin 1998;56:200–4.Search in Google Scholar

16. Louie R, Tang Z, Sutton D, Lee JH, Kost GJ. Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med 2000;124:257–66.10.5858/2000-124-0257-POCGTSearch in Google Scholar PubMed

Received: 2012-03-28
Accepted: 2012-06-27
Published Online: 2012-08-03
Published in Print: 2013-02-01

©2013 by Walter de Gruyter Berlin Boston

Articles in the same Issue

  1. Letters to the Editor
  2. Performance evaluation of three different immunoassays for detection of antibodies to hepatitis B core
  3. Serum homocysteine concentrations in Chinese children with autism
  4. Interchangeability of venous and capillary HbA1c results is affected by oxidative stress
  5. Interference of hemoglobin (Hb) N-Baltimore on measurement of HbA1c using the HA-8160 HPLC method
  6. First human isolate of Mycobacterium madagascariense in the sputum of a patient with tracheobronchitis
  7. Protein S and protein C measurements should not be undertaken during vitamin K antagonist therapy
  8. α2-HS glycoprotein is an essential component of cryoglobulin associated with chronic hepatitis C
  9. An unusual interference in CK MB assay caused by a macro enzyme creatine phosphokinase (CK) type 2 in HIV-infected patients
  10. An automated technique for the measurement of the plasma glutathione reductase activity and determination of reference limits for a healthy population
  11. Is osteopontin stable in plasma and serum?
  12. Evidence-based approach to reducing perceived wasteful practices in laboratory medicine
  13. Masthead
  14. Masthead
  15. Editorials
  16. Testing volume is not synonymous of cost, value and efficacy in laboratory diagnostics
  17. Lessons from controversy: biomarkers evaluation
  18. Commercial immunoassays in biomarkers studies: researchers beware!1)
  19. Trials and tribulations in lupus anticoagulant testing
  20. Reviews
  21. Mass spectrometry: a revolution in clinical microbiology?
  22. Chronic Chagas disease: from basics to laboratory medicine
  23. General Clinical Chemistry and Laboratory Medicine
  24. Shop for quality or quantity? Volumes and costs in clinical laboratories
  25. Minor improvement of venous blood specimen collection practices in primary health care after a large-scale educational intervention
  26. Evaluation of high resolution gel β2-transferrin for detection of cerebrospinal fluid leak
  27. Serum kallikrein-8 correlates with skin activity, but not psoriatic arthritis, in patients with psoriatic disease
  28. Soluble urokinase plasminogen activator receptor (suPAR) in the assessment of inflammatory activity of rheumatoid arthritis patients in remission
  29. Bone mass density selectively correlates with serum markers of oxidative damage in post-menopausal women
  30. Validation of a fast and reliable liquid chromatography-tandem mass spectrometry (LC-MS/MS) with atmospheric pressure chemical ionization method for simultaneous quantitation of voriconazole, itraconazole and its active metabolite hydroxyitraconazole in human plasma
  31. Performance of different screening methods for the determination of urinary glycosaminoclycans
  32. Intestinal permeability and fecal eosinophil-derived neurotoxin are the best diagnosis tools for digestive non-IgE-mediated cow’s milk allergy in toddlers
  33. An internal validation approach and quality control on hematopoietic chimerism testing after allogeneic hematopoietic cell transplantation
  34. Serum levels of IgG antibodies against oxidized LDL and atherogenic indices in HIV-1-infected patients treated with protease inhibitors
  35. Cooperation experience in a multicentre study to define the upper limits in a normal population for the diagnostic assessment of the functional lupus anticoagulant assays
  36. Contribution of procoagulant phospholipids, thrombomodulin activity and thrombin generation assays as prognostic factors in intensive care patients with septic and non-septic organ failure
  37. Suitability of POC lactate methods for fetal and perinatal lactate testing: considerations for accuracy, specificity and decision making criteria
  38. Point-of-care testing on admission to the intensive care unit: lactate and glucose independently predict mortality
  39. Reference Values and Biological Variations
  40. CA125 reference values change in male and postmenopausal female subjects
  41. Distributions and ranges of values of blood and urinary biomarker of inflammation and oxidative stress in the workers engaged in office machine manufactures: evaluation of reference values
  42. Cancer Diagnostics
  43. Association of acute phase protein-haptoglobin, and epithelial-mesenchymal transition in buccal cancer: a preliminary report
  44. Comparison of diagnostic and prognostic performance of two assays measuring thymidine kinase 1 activity in serum of breast cancer patients
  45. Evaluation of the BRAHMS Kryptor® Thyroglobulin Minirecovery Test in patients with differentiated thyroid carcinoma
Downloaded on 17.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2012-0201/html
Scroll to top button