Inhalation therapy has become one of the fundamental treatments for chronic obstructive pulmonary disease (COPD) and bronchial asthma (asthma).[1,2] Meanwhile, with the development of inhaled formulations and devices, inhalation therapy has shown a wide prospect of expanding indications, including respiratory diseases such as lower respiratory tract infections,[3] interstitial lung diseases, lung cancer,[4] as well as other diseases such as diabetes and depression.[5] Compared to oral and intravenous administration, the aerosolized drug is inhaled and acts directly on the airways and lungs, which has the advantages of rapid onset, sustained efficacy, and low systemic adverse effects.[6] Compared to the nebulizer, the hand-held inhaler is a better option for out-of-hospital drug delivery because the latter is easier to operate, requires shorter inhalation time, and has less environment extravasation.[6]
The commonly used inhalers include pressurized metered-dose inhalers (pMDIs), dry powder inhalers (DPIs), and soft mist inhalers (SMIs). The general requirements for using inhalers are to exhale deeply and slowly before inhaling, as well as to hold the breath preferably for about 10 s and then exhale slowly after inhalation. Also, different inhalers require specifically peak inspiratory flow against inhalers’ internal resistance (PIFr), inspiratory pattern, and actuation–inhalation coordination (Table 1).[6, 7, 8, 9, 10] It was reported that the error rates of different devices ranged from 58.9% to 87.9%, while suboptimal PIFr was common in patients with asthma or COPD.[11, 12, 13, 14, 15] Critical errors in inhaler use, including lack of synchronization, expiration in powder device before inhalation, and remaining powder in the capsule by the end were associated with an increased rate of severe COPD exacerbation.[16] Also, a systematic review identified an association between inhaler errors and poor asthma control, COPD disease stability, and greater economic burden on health.[17] Inspiratory flow rates failing to meet inhaler requirements, incorrect inspiratory patterns, or uncoordinated maneuvers would lead to insufficient drug deposition at the targeted position and inadequate drug efficacy, which in turn would increase the risk of acute exacerbation of diseases, hospitalization, emergency room visits, and economic burden.[17,18]
| Inhalers | Internal resistance levels | Minimal PIFr (L/min) | Optimal PIFr (L/min) | Inhalation maneuvers | |
|---|---|---|---|---|---|
| pMDIs | Traditional pMDI, Aerosphere® | Nearly no resistance | 10 | 30 | Inhale slowly, steadily, and deeply, synchronously actuating the inhaler after initiating inspiration |
| SMIs | Respimat® | No resistance | 10 | 30 | Actuate the inhaler and simultaneously inhale slowly, steadily, and deeply |
| DPIs | Breezhaler® Accuhaler®/Diskus®, Ellipta® Turbuhaler® (combination) | Low resistance Medium-low resistance Medium resistance | 20–50 | 30–65 | Inhale forcefully through the mouth from the very beginning until the lungs are full |
| Turbuhaler® (monotherapy), Easyhaler® C (combination) | Medium-high resistance | ||||
| HandiHaler®, Easyhaler® M (monotherapy) | High resistance |
PIFr: peak inspiratory flow against resistance; pMDIs: pressurized metered-dose inhalers; SMIs: soft mist inhalers; DPIs: dry powder inhalers.
Therefore, health-care providers should assess the patient’s inhaler technique accurately and select a proper inhaler accordingly. However, the assessment process is now mostly based on personal clinical experience, is nonhomogeneous, and lacks objective measurements. Based on the available clinical studies and pharmacological studies, we hereby propose a tentative structured assessment of inhalation capability for clinical practice and further studies. The assessment should contain patient’s inhalation maneuver (including actuation-inhalation synchronization), PIFr, and breath-holding time. The evaluation of inhalation maneuver and breath-holding time can be accomplished by clinical observation, while PIFr may require an objective quantitative measurement following a standardized procedure, besides clinical observation.
PIFr is the most important indicator of the evaluation of inhalation capability for DPIs. Clinical evidence showed that inhaling drugs with too low PIFr (< 30 L/min) could not relieve bronchospasm during an asthma attack in patients.[19] When PIFr increased from lower levels to near-optimal levels, the lung deposition rate of drugs inhaled via DPIs increased significantly.[20] However, excess inspiratory flow rates could lead to negative effects, including increased drug deposition in the central airways and decreased drug deposition in the peripheral airways.[21] The increased oropharyngeal drug deposition might lead to hoarseness and other adverse reactions. It was shown that PIFr-guided inhalation therapy (including PIFr assessment and inhaler use training) had an independent correlation with reduced risk of severe acute exacerbation of COPD.[22] Patients with suboptimal PIFr (≤ 60 L/min) had higher rates of readmission for COPD within 90 days, higher rates of all-cause readmissions, and shorter intervals of readmission caused by COPD.[23,24] The COPD patient’s PIFr at discharge could also predict all-cause readmissions within 30 and 90 days.[25] This might be related to the patient’s poor inhalation capability failing to fully exploit the drug’s efficacy in DPIs.
The clinical observation for the evaluation of PIFr includes inspiratory muscle strength and disease condition. The maximal inspiratory pressure is positively correlated with PIFr,[26] which is generated by the respiratory muscle strength and tone.[27] Age and gender were independently associated with PIFr partly by influencing inspiratory muscle strength.[27,28,29] The frailty could affect the inspiratory muscle strength and the maximum inspiratory pressure, thus decreasing PIFr.[30] People with COPD and asthma tended to have lower PIFr than healthy people.[15,31] Also, participants with active asthma or COPD exacerbation showed a lower PIFr than those without an exacerbation.[7] Individuals with suboptimal PIFr were common among COPD patients hospitalized for acute exacerbations.[23,32] Age, female sex, weak inspiratory muscle strength, frailty, pulmonary diseases such as COPD and asthma, and acute exacerbation are related to low inspiratory flow rates and failure to achieve optimal PIFr. If the patient is in these conditions, health-care providers should consider selecting an inhaler with a relatively lower requirement for inspiratory effort, and measuring PIFr objectively would ensure an optimal inhaler prescription.
The ideal device for PIFr measurement should be portable, easy to operate, user-friendly, affordable, and with a good accuracy and repeatability. In-Check DIAL™ G16 (Clement Clarke International, Harlow, UK and Alliance Tech Medical), a hand-held flow meter specifically designed to measure PIFr, is mostly widely used in clinical research and practice abroad. In-Check DIAL simulates the resistance of different inhalers by adjusting the resistance aperture, with a measurement range of 15–120 L/ min and an accuracy of ± 10% or ± 10 L/min (whichever is greater).[33,34] It is portable, inexpensive, and simple to operate, making it suitable for the detection of PIFr both in the hospital and at home. It has been proved that In-Check DIAL can help to guide inhaler choice and improve patients’ PIFr performance.[32]
Supposedly, a novel flow sensor–based spirometer, PF810® (UBREATH, Hangzhou, China) can simulate different levels of resistance, similar to In-Check DIAL, by stepper motors.[35] It uses the flow sensor to quantitatively detect PIFr and the flow waveforms during the user’s inhalation. PIFr, inspiratory time, inspiratory volume, and breath-holding time are analyzed by computer programs. However, its efficacy for clinical use needs to be validated. Moreover, it needs environmental calibration and is less portable and more expensive than In-Check Dial. Therefore, it is more suitable for use in hospitals. It should be noted that routine spirometers are not suitable for PIFr measurement due to the minimal resistance.
Accurate and reliable measurement of PIFr is dependent on the correct testing procedure. For evaluating the inhalation capability of DPI, the patient should follow the following steps: (1) sit or stand up straight and tilt head; (2) exhale slowly and as completely as possible; (3) place the mouthpiece between the lips and ensure no air leakage; (4) inhale deeply and quickly and keep the inhalation as steady as possible; (5) hold the breath about 10 s and then exhale slowly; and (6) repeat the above procedures for three times and take the maximum value to determine whether the minimum and optimal PIFr of the inhaler can be reached.[27] Incomplete exhalation at step 3 or failure to inhale deeply, rapidly, and forcefully at step 4 would lead to an underestimation of PIFr.
For instruments with calibration requirements, BTPS calibration should be performed each time the device is turned on and it should be recalibrated when the environmental conditions change greatly. Otherwise, measurement errors might happen. Extreme environments might affect not only the device, but also the patient’s inhalation capability. For example, high altitude conditions would lead to decreased inspiratory muscle strength,[36] which in turn affects the inhalation capability.
In addition to PIFr, the inspiratory pressure, inspiratory time, and inspiratory volume during inhalation could also reflect the inhalation performance. Theoretically, the inspiratory time and volume can be useful parameters to assess the inspiratory pattern, especially as dose emission varies in different inhalers. However, the study of these indicators in the guidance of clinical medication is still lacking, and their clinical correlation with drug efficacy has yet to be confirmed by further studies.
In conclusion, the assessment of inhalation capability should be performed in a structured and standardized way to guide inhaler prescriptions, combining clinical observation and objective measurements of inhalation maneuvers, PIFr, inspiratory volume, inspiratory time, and breath-holding time. The assessment would help to accurately determine whether the patient is able to use a specific inhaler with sufficient efficacy. For example, PIFr assessment would help to tell whether the patient can achieve the optimal PIFr of DPIs, and actuation–inhalation coordination evaluation would indicate whether the patient is capable of operating pMDIs. Further in-depth research will help optimize inhalation therapy strategies and improve the effectiveness of inhalation therapy.
-
Conflict of Interest
None declared.
References
1 Global Initiative for Chronic Obstructive Lung Disease. Global Strategy for the Diagnosis, Management and Prevention of Chronic Obstructive Pulmonary Disease (2023 REPORT). Available at: https://goldcopd.org/2023-gold-report-2/. Accessed Novemver 17, 2022.Suche in Google Scholar
2 Global Initiative for asthma. Global Strategy for Asthma Management and Prevention (2021 update). Available at: https://ginasthma.org/gina-reports/. Accessed June 29, 2022.Suche in Google Scholar
3 Han Y, Hua J, He B, Zhang J. Inhaled antibiotics and airway bacterial decolonization for patients with chronic obstructive pulmonary disease: The rationale and future. J Transl Med 2022;10;181–4.10.2478/jtim-2022-0005Suche in Google Scholar PubMed PubMed Central
4 Zhou QT, Tang P, Leung SS, Chan JG, Chan HK. Emerging inhalation aerosol devices and strategies: where are we headed? Adv Drug Deliv Rev 2014;75:3–17.10.1016/j.addr.2014.03.006Suche in Google Scholar PubMed
5 Rubin BK, Williams RW. Emerging aerosol drug delivery strategies: from bench to clinic. Adv Drug Deliv Rev 2014;75:141–8.10.1016/j.addr.2014.06.008Suche in Google Scholar PubMed
6 Laube BL, Janssens HM, de Jongh FH, Devadason SG, Dhand R, Diot P, et al. What the pulmonary specialist should know about the new inhalation therapies. Eur Respir J 2011;37:1308–31.10.1183/09031936.00166410Suche in Google Scholar PubMed
7 Ghosh S, Ohar JA, Drummond MB. Peak Inspiratory Flow Rate in Chronic Obstructive Pulmonary Disease: Implications for Dry Powder Inhalers. J Aerosol Med Pulm Drug Deliv 2017;30:381–7.10.1089/jamp.2017.1416Suche in Google Scholar PubMed PubMed Central
8 Haidl P, Heindl S, Siemon K, Bernacka M, Cloes RM. Inhalation device requirements for patients’ inhalation maneuvers. Respir Med 2016;118:65–75.10.1016/j.rmed.2016.07.013Suche in Google Scholar PubMed
9 Pauwels R, Newman S, Borgstrom L. Airway deposition and airway effects of antiasthma drugs delivered from metered-dose inhalers. Eur Resp J 1997;10:2127–38.10.1183/09031936.97.10092127Suche in Google Scholar PubMed
10 Sanders MJ. Guiding Inspiratory Flow: Development of the In-Check DIAL G16, a Tool for Improving Inhaler Technique. Pulm Med 2017;2017:1495867.10.1155/2017/1495867Suche in Google Scholar PubMed PubMed Central
11 Liang CY, Chen YJ, Sheu SM, Tsai CF, Chen W. Misuse of inhalers among COPD patients in a community hospital in Taiwan. Int J Chron Obstruct Pulmon Dis 2018;13:1309–16.10.2147/COPD.S158864Suche in Google Scholar PubMed PubMed Central
12 Cho-Reyes S, Celli BR, Dembek C, Yeh K, Navaie M. Inhalation Technique Errors with Metered-Dose Inhalers Among Patients with Obstructive Lung Diseases: A Systematic Review and Meta-Analysis of U.S. Studies. Chronic Obstr Pulm Dis 2019;6:267–80.10.15326/jcopdf.6.3.2018.0168Suche in Google Scholar PubMed PubMed Central
13 Navaie M, Dembek C, Cho-Reyes S, Yeh K, Celli BR. Device use errors with soft mist inhalers: A global systematic literature review and meta-analysis. Chron Respir Dis 2020;17:1479973119901234.10.1177/1479973119901234Suche in Google Scholar PubMed PubMed Central
14 Mahler DA, Ohar JA, Yun J, Moran EJ, Barnes CN, Crater GD. Suboptimal PIFR Associated with Greater Dyspnea in Chronic Obstructive Pulmonary Disease (COPD). Am J Respir Crit Care Med 2019;199:2.10.1164/ajrccm-conference.2019.199.1_MeetingAbstracts.A1132Suche in Google Scholar
15 Silva CPD, Cordeiro JSA, Britto MCA, Bezerra PGM, Andrade LB. Peak inspiratory flow in children and adolescents with asthma using dry powder inhalers: a cross-sectional study. J Bras Pneumol 2021;47:e20200473.10.36416/1806-3756/e20200473Suche in Google Scholar PubMed PubMed Central
16 Molimard M, Raherison C, Lignot S, Balestra A, Lamarque S, Chartier A, et al. Chronic obstructive pulmonary disease exacerbation and inhaler device handling: real-life assessment of 2935 patients. Eur Respir J 2017;49:1601794.10.1183/13993003.01794-2016Suche in Google Scholar PubMed
17 Usmani OS, Lavorini F, Marshall J, Dunlop WCN, Heron L, Farrington E, et al. Critical inhaler errors in asthma and COPD: a systematic review of impact on health outcomes. Respir Res 2018;19:10.10.1186/s12931-017-0710-ySuche in Google Scholar PubMed PubMed Central
18 Mahler DA, Halpin DMG. Peak Inspiratory Flow as a Predictive Therapeutic Biomarker in COPD. Chest 2021;160:491–8.10.1016/j.chest.2021.03.049Suche in Google Scholar PubMed
19 Haidl P, Schmidt F, Wiese C, Koehler D. Peak inspiratory flow rate after methacholine challenge in asthmatic patients and its impact on the effect of formoterol via different inhalers. J Aerosol Med 2006;19: 364–71.10.1089/jam.2006.19.364Suche in Google Scholar PubMed
20 Borgström L, Bondesson E, Morén F, Trofast E, Newman SP. Lung deposition of budesonide inhaled via Turbuhaler: a comparison with terbutaline sulphate in normal subjects. Eur Respir J 1994;7:69–73.10.1183/09031936.94.07010069Suche in Google Scholar PubMed
21 Usmani OS, Biddiscombe MF, Barnes PJ. Regional lung deposition and bronchodilator response as a function of beta2-agonist particle size. Am J Respir Crit Care Med 2005;172:1497–504.10.1164/rccm.200410-1414OCSuche in Google Scholar PubMed
22 Chen SY, Huang CK, Peng HC, Tsai HC, Huang SY, Yu CJ, et al. Peak-Inspiratory-Flow-Rate Guided Inhalation Therapy Reduce Severe Exacerbation of COPD. Front Pharmacol 2021;12:704316.10.3389/fphar.2021.704316Suche in Google Scholar PubMed PubMed Central
23 Loh CH, Peters SP, Lovings TM, Ohar JA. Suboptimal Inspiratory Flow Rates Are Associated with Chronic Obstructive Pulmonary Disease and All-Cause Readmissions. Ann Am Thorac Soc 2017;14:1305–11.10.1513/AnnalsATS.201611-903OCSuche in Google Scholar PubMed
24 Samarghandi A, Ioachimescu OC, Qayyum R. Association between peak inspiratory flow rate and hand grip muscle strength in hospitalized patients with acute exacerbation of chronic obstructive pulmonary disease. PLoS One 2020;15:e0227737.10.1371/journal.pone.0227737Suche in Google Scholar PubMed PubMed Central
25 Alqahtani JS, Aldabayan YS, Aldhahir AM, Al Rajeh AM, Mandal S, Hurst JR. Predictors of 30- and 90-Day COPD Exacerbation Readmission: A Prospective Cohort Study. Int J Chron Obstruct Pulmon Dis 2021;16:2769–81.10.2147/COPD.S328030Suche in Google Scholar PubMed PubMed Central
26 Janssens W, VandenBrande P, Hardeman E, De Langhe E, Philps T, Troosters T, et al. Inspiratory flow rates at different levels of resistance in elderly COPD patients. Eur Respir J 2008;31:78–83.10.1183/09031936.00024807Suche in Google Scholar PubMed
27 Clark AR, Weers JG, Dhand R. The Confusing World of Dry Powder Inhalers: It Is All About Inspiratory Pressures, Not Inspiratory Flow Rates. J Aerosol Med Pulm Drug Deliv 2020;33:1–11.10.1089/jamp.2019.1556Suche in Google Scholar PubMed PubMed Central
28 Malmberg LP, Rytilä P, Happonen P, Haahtela T. Inspiratory flows through dry powder inhaler in chronic obstructive pulmonary disease: age and gender rather than severity matters. Int J Chron Obstruct Pulmon Dis 2010;5:257–62.10.2147/COPD.S11474Suche in Google Scholar PubMed PubMed Central
29 Sun L, Chen Y. Interpretation of the key issues of expert consensus on immunomodulatory therapies for chronic obstructive pulmonary disease. J Transl Intern Med 2022;10:277–80.10.2478/jtim-2022-0069Suche in Google Scholar PubMed PubMed Central
30 Vidal MB, Pegorari MS, Santos EC, Matos AP, Pinto A, Ohara DG. Respiratory muscle strength for discriminating frailty in community-dwelling elderly: a cross-sectional study. Arch Gerontol Geriatr 2020;89:104082.10.1016/j.archger.2020.104082Suche in Google Scholar PubMed
31 Ghosh S, Pleasants RA, Ohar JA, Donohue JF, Drummond MB. Prevalence and factors associated with suboptimal peak inspiratory flow rates in COPD. Int J Chron Obstruct Pulmon Dis 2019;14:585–95.10.2147/COPD.S195438Suche in Google Scholar PubMed PubMed Central
32 Hua JL, Ye XF, Du CL, Xie N, Zhang JQ, Li M, et al. Optimizing inhalation therapy in the aspect of peak inhalation flow rate in patients with chronic obstructive pulmonary disease or asthma. BMC Pulm Med 2021;21:302.10.1186/s12890-021-01674-5Suche in Google Scholar PubMed PubMed Central
33 In-Check DIAL® G16. Clement Clarke International. Available at: https://www.haag-streit.com/clement-clarke/products/inhaler-technique/in-check-dial-g16/. Accessed June 21, 2022.Suche in Google Scholar
34 In-Check DIAL® G16. Inhaler technique training and assessment tool. Alliance Tech Medical. Available at: https://alliancetechmedical.com/products/check-dial-training-device/. Accessed June 21, 2022.Suche in Google Scholar
35 UBREATH® Multi-Function Spirometer System (PF810). Available at: https://www.e-linkcare.com/ubreath-multi-function-spirometer-system-pf810-product/ Accessed June 21, 2022.Suche in Google Scholar
36 Fasano V, Paolucci E, Pomidori L, Cogo A. High-altitude exposure reduces inspiratory muscle strength. Int J Sports Med 2007;28:426–30.10.1055/s-2006-924367Suche in Google Scholar PubMed
© 2023 Yixing Wu, Jing Zhang, published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
Artikel in diesem Heft
- Perspective
- Revising the hemodynamic criteria for pulmonary hypertension: A perspective from China
- Animal models: An essential tool to dissect the heterogeneity of chronic obstructive pulmonary disease
- Effective albumin – A novel paradigm in the management of decompensated liver cirrhosis
- Monkeypox: A real new warning or just a sign of times?
- Severe acute hepatitis of unknown origin in children: Clinical issues of concern
- Commentary
- Manipulating cell motility by Legionella: Speeding up or slowing down?
- Standardized inhalation capability assessment: A key to optimal inhaler selection for inhalation therapy
- Review Article
- Mesenchymal stem cells and connective tissue diseases: From bench to bedside
- Predictors of progression in idiopathic inflammatory myopathies with interstitial lung disease
- Original Article
- Moderate-intensity continuous training has time-specific effects on the lipid metabolism of adolescents
- Point-of-care ultrasound-guided submucosal paclitaxel injection in tracheal stenosis model
- Gas chromatography-mass spectrometry pilot study to identify volatile organic compound biomarkers of childhood obesity with dyslipidemia in exhaled breath
- Letter to Editor
- Efficacy and safety of avatrombopag in aplastic anemia patients with liver disease
- Retraction Note
- Retraction note: Hydrogel: A promising new technique for treating Alzheimer’s disease (in Volume 10 Issue 3)
Artikel in diesem Heft
- Perspective
- Revising the hemodynamic criteria for pulmonary hypertension: A perspective from China
- Animal models: An essential tool to dissect the heterogeneity of chronic obstructive pulmonary disease
- Effective albumin – A novel paradigm in the management of decompensated liver cirrhosis
- Monkeypox: A real new warning or just a sign of times?
- Severe acute hepatitis of unknown origin in children: Clinical issues of concern
- Commentary
- Manipulating cell motility by Legionella: Speeding up or slowing down?
- Standardized inhalation capability assessment: A key to optimal inhaler selection for inhalation therapy
- Review Article
- Mesenchymal stem cells and connective tissue diseases: From bench to bedside
- Predictors of progression in idiopathic inflammatory myopathies with interstitial lung disease
- Original Article
- Moderate-intensity continuous training has time-specific effects on the lipid metabolism of adolescents
- Point-of-care ultrasound-guided submucosal paclitaxel injection in tracheal stenosis model
- Gas chromatography-mass spectrometry pilot study to identify volatile organic compound biomarkers of childhood obesity with dyslipidemia in exhaled breath
- Letter to Editor
- Efficacy and safety of avatrombopag in aplastic anemia patients with liver disease
- Retraction Note
- Retraction note: Hydrogel: A promising new technique for treating Alzheimer’s disease (in Volume 10 Issue 3)