Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method

  • Vratislav Chromý , Lenka Šváchová , Lukáš Novosád , Jirí Jarkovský , Pavel Sedlák , Petr Horák , Hana Dobrovolná and Barbora Hlavácová
Published/Copyright: December 31, 2008

Abstract

Background: Assay of total serum protein by the biuret method calibrated with albumin standards according to the reference method provides results with a positive bias approximately 3%–5% exceeding the total error of 3.4% allowable for total protein in serum analysis made by analysers using two-part reagents and short-term procedures.

Methods: We used two types of two-part biuret reagents utilised in a short-term measurement in analysers with albumin or serum calibrators, in which protein was attested by the Kjeldahl method.

Results: Tests with potentially interfering substances proved that serum blanking used in a short-term biuret procedure is not capable of sufficiently eliminating effects of serum interferents. A short-term blanking is evidently capable of suppressing only an absorbance caused by serum-present coloured and turbid interferents, but its capacity to transform them (oxidise, hydrolyse, saponify, etc.) to some other not-interfering substances is very low compared with a long-term blanking. Lipids and bilirubin are responsible for significant positive bias of total protein in normal serum samples (approximately 3%) and even a greater positive offset in lipaemic and icteric sera (approximately 5%). We verified that interference tests based on a normal serum spiked with endogenous lipids and bilirubin give quite false and misleading results in the biuret reaction. A pure albumin, not depending on its bovine/human origin, gives absorbance responding only to its copper complexes with protein with a biuret regent, while its absorbance with a serum also includes the absorbance of interferents present in serum. The simplest way to improve current short-term biuret procedures is the use of a human serum calibrator with total protein attested by the Kjeldahl method. A serum calibrator, behaving analogously to serum samples, compensates for a positive bias in most normal sera. Reagents with a greater concentration of active biuret components (copper and alkali, reference method included) seem to be unnecessarily aggressive to proteins and are responsible for a lower accuracy when used in short-term measurements.

Conclusions: Standard Reference Material 927c based on pure bovine albumin is still recommended and used as the primary standard for assays of total protein by colourimetric methods. The albumin calibrator is responsible for a positive bias of approximately 3%–5% in serum total protein assayed by the biuret reaction both in the reference and in current methods. Its substitution by a serum calibrator attested by the Kjeldahl method could solve this drawback.

Clin Chem Lab Med 2009;47:91–101.


Corresponding author: Prof. Dr. Vratislav Chromý, Institute of Chemistry, Faculty of Science, Masaryk University, Komenského 2, Brno, Czech Republic Phone: +420-549494555,

Received: 2008-7-3
Accepted: 2008-10-9
Published Online: 2008-12-31
Published in Print: 2009-01-01

©2009 by Walter de Gruyter Berlin New York

Articles in the same Issue

  1. Reviews
  2. Molecular markers detecting circulating melanoma cells by reverse transcription polymerase chain reaction: methodological pitfalls and clinical relevance
  3. HIV-GB virus C co-infection: an overview
  4. Genetics and Molecular Diagnostics
  5. Relationships between C-reactive protein concentration and genotype in healthy volunteers
  6. SYBR Green-based real-time PCR assay for detection of VKORC1 and CYP2C9 polymorphisms that modulate warfarin dose requirement
  7. Functional polymorphisms in ACE and CYP11B2 genes and atrial fibrillation in patients with hypertensive heart disease
  8. No effect of MDR1 C3435T polymorphism on oral pharmacokinetics of telmisartan in 19 healthy Chinese male subjects
  9. A novel large deletion and three polymorphisms in the FECH gene associated with erythropoietic protoporphyria
  10. Aging and endothelial progenitor cell telomere length in healthy men
  11. Gln222Arg (A2317G) polymorphism in the deoxyribonuclease I gene exhibits ethnic and functional differences
  12. Mutation screening of the medium-chain acyl-CoA dehydrogenase (MCAD) and the ornithine transcarbamylase (OTC) genes by multiplex PCR amplification and sequencing
  13. General Clinical Chemistry and Laboratory Medicine
  14. Nitrated apolipoprotein A-I, a potential new cardiovascular marker, is markedly increased in low high-density lipoprotein cholesterol subjects
  15. Relationships of serum haptoglobin concentration with HbA1c and glycated albumin concentrations in Japanese type 2 diabetic patients
  16. Multi-elements (aluminium, copper, magnesium, manganese, selenium and zinc) determination in serum by dynamic reaction cell-inductively coupled plasma-mass spectrometry
  17. Urinary neutrophil gelatinase-associated lipocalin (NGAL) as an early marker of acute kidney injury in critically ill multiple trauma patients
  18. Validation and Outcome Studies
  19. Validation for quantification of immunoglobulins by Fourier transform infrared spectrometry
  20. Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method
  21. Current practices in antinuclear antibody testing: results from the Belgian External Quality Assessment Scheme
  22. Verification of the analytical range of a new reagent for full-range C-reactive protein determination
  23. Measurement uncertainty in manual differential leukocyte counting
  24. Letters to the Editor
  25. Falsely elevated carbohydrate antigen 19-9 level due to heterophilic antibody interference but not rheumatoid factor: a case report
  26. An unusual interference in parathormone assay caused by anti-goat IgG: a case report
Downloaded on 28.10.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2009.011/html
Scroll to top button