Multi-elements (aluminium, copper, magnesium, manganese, selenium and zinc) determination in serum by dynamic reaction cell-inductively coupled plasma-mass spectrometry
-
Bonnie Mei Wah Fong
Abstract
Background: Trace element determination in laboratory medicine is widely carried out by atomic absorption or emission spectroscopy. In the last decade, there has been a rapid growth in the use of inductively coupled plasma-mass spectrometry because of its strong detection power, and the possibility of multi-elements analysis in a single run.
Methods: Having the advantages of smaller sample volume and better detection limit, we developed a method for the simultaneous determinations of six trace elements by using 100 μL serum, and the assay can be accomplished within 3 min.
Results: The method developed gave recovery of the six elements ranging from 97% to 117%. The method covered a wide dynamic range with manganese in the range of nmol/L, while magnesium was in the range of mmol/L. The detection limits were 0.001 mmol/L, 0.05 μmol/L, 2.0 nmol/L, 0.2 μmol/L, 0.05 μmol/L, and 0.01 μmol/L for magnesium, aluminium, manganese, copper, zinc, and selenium, respectively. All the six elements had intra-assay imprecision of less than 5%, and inter-assay imprecision of less than 8%.
Conclusions: This fast and robust method is suitable for use in the clinical laboratory where short turnaround time is needed for managing patients with trace element deficiency or toxicity.
Clin Chem Lab Med 2009;47:75–8.
©2009 by Walter de Gruyter Berlin New York
Articles in the same Issue
- Reviews
- Molecular markers detecting circulating melanoma cells by reverse transcription polymerase chain reaction: methodological pitfalls and clinical relevance
- HIV-GB virus C co-infection: an overview
- Genetics and Molecular Diagnostics
- Relationships between C-reactive protein concentration and genotype in healthy volunteers
- SYBR Green-based real-time PCR assay for detection of VKORC1 and CYP2C9 polymorphisms that modulate warfarin dose requirement
- Functional polymorphisms in ACE and CYP11B2 genes and atrial fibrillation in patients with hypertensive heart disease
- No effect of MDR1 C3435T polymorphism on oral pharmacokinetics of telmisartan in 19 healthy Chinese male subjects
- A novel large deletion and three polymorphisms in the FECH gene associated with erythropoietic protoporphyria
- Aging and endothelial progenitor cell telomere length in healthy men
- Gln222Arg (A2317G) polymorphism in the deoxyribonuclease I gene exhibits ethnic and functional differences
- Mutation screening of the medium-chain acyl-CoA dehydrogenase (MCAD) and the ornithine transcarbamylase (OTC) genes by multiplex PCR amplification and sequencing
- General Clinical Chemistry and Laboratory Medicine
- Nitrated apolipoprotein A-I, a potential new cardiovascular marker, is markedly increased in low high-density lipoprotein cholesterol subjects
- Relationships of serum haptoglobin concentration with HbA1c and glycated albumin concentrations in Japanese type 2 diabetic patients
- Multi-elements (aluminium, copper, magnesium, manganese, selenium and zinc) determination in serum by dynamic reaction cell-inductively coupled plasma-mass spectrometry
- Urinary neutrophil gelatinase-associated lipocalin (NGAL) as an early marker of acute kidney injury in critically ill multiple trauma patients
- Validation and Outcome Studies
- Validation for quantification of immunoglobulins by Fourier transform infrared spectrometry
- Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method
- Current practices in antinuclear antibody testing: results from the Belgian External Quality Assessment Scheme
- Verification of the analytical range of a new reagent for full-range C-reactive protein determination
- Measurement uncertainty in manual differential leukocyte counting
- Letters to the Editor
- Falsely elevated carbohydrate antigen 19-9 level due to heterophilic antibody interference but not rheumatoid factor: a case report
- An unusual interference in parathormone assay caused by anti-goat IgG: a case report
Articles in the same Issue
- Reviews
- Molecular markers detecting circulating melanoma cells by reverse transcription polymerase chain reaction: methodological pitfalls and clinical relevance
- HIV-GB virus C co-infection: an overview
- Genetics and Molecular Diagnostics
- Relationships between C-reactive protein concentration and genotype in healthy volunteers
- SYBR Green-based real-time PCR assay for detection of VKORC1 and CYP2C9 polymorphisms that modulate warfarin dose requirement
- Functional polymorphisms in ACE and CYP11B2 genes and atrial fibrillation in patients with hypertensive heart disease
- No effect of MDR1 C3435T polymorphism on oral pharmacokinetics of telmisartan in 19 healthy Chinese male subjects
- A novel large deletion and three polymorphisms in the FECH gene associated with erythropoietic protoporphyria
- Aging and endothelial progenitor cell telomere length in healthy men
- Gln222Arg (A2317G) polymorphism in the deoxyribonuclease I gene exhibits ethnic and functional differences
- Mutation screening of the medium-chain acyl-CoA dehydrogenase (MCAD) and the ornithine transcarbamylase (OTC) genes by multiplex PCR amplification and sequencing
- General Clinical Chemistry and Laboratory Medicine
- Nitrated apolipoprotein A-I, a potential new cardiovascular marker, is markedly increased in low high-density lipoprotein cholesterol subjects
- Relationships of serum haptoglobin concentration with HbA1c and glycated albumin concentrations in Japanese type 2 diabetic patients
- Multi-elements (aluminium, copper, magnesium, manganese, selenium and zinc) determination in serum by dynamic reaction cell-inductively coupled plasma-mass spectrometry
- Urinary neutrophil gelatinase-associated lipocalin (NGAL) as an early marker of acute kidney injury in critically ill multiple trauma patients
- Validation and Outcome Studies
- Validation for quantification of immunoglobulins by Fourier transform infrared spectrometry
- Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method
- Current practices in antinuclear antibody testing: results from the Belgian External Quality Assessment Scheme
- Verification of the analytical range of a new reagent for full-range C-reactive protein determination
- Measurement uncertainty in manual differential leukocyte counting
- Letters to the Editor
- Falsely elevated carbohydrate antigen 19-9 level due to heterophilic antibody interference but not rheumatoid factor: a case report
- An unusual interference in parathormone assay caused by anti-goat IgG: a case report