Measuring Rivaroxaban in a clinical laboratory setting, using common coagulation assays, Xa inhibition and thrombin generation

  • Pascal J. Molenaar , Jasper Dinkelaar and Anja Leyte EMAIL logo
Published/Copyright: March 30, 2012

Abstract

Background: Rivaroxaban, a direct Xa inhibitor, is one of the new oral antithrombotic agents for which laboratory monitoring is thought to be unnecessary in most cases due to predictable pharmacokinetics. Circumstances are conceivable, however, in which reliable laboratory testing of Rivaroxaban is desirable. The aim of the present in vitro study was to investigate and compare the analytical and practical use of Rivaroxaban monitoring with routine screening assays, thrombin generation and anti-Xa activity, in a clinical laboratory setting.

Methods: Rivaroxaban was added to nine normal donor plasmas and to a normal pooled plasma in concentrations up to 1000 μg/L. Prothrombin time (PT), activated partial thromboplastin time (APTT), endogenous thrombin potential (ETP) and anti-Xa activity were measured in all donor samples. Responsiveness to Rivaroxaban and imprecision of Rivaroxaban recovery were assessed.

Results: Low intra-, but high inter-individual imprecision was found for PT displaying a linear dose-response relationship. Imprecision was much lower when directly measuring anti-Xa activity. Responsiveness of ETP lag-time was high, but of total thrombin generation was low, illustrating that the main effect of Rivaroxaban Xa inhibition lies in delaying thrombin formation rather than in preventing it.

Conclusions: Despite a high inter-individual imprecision of the PT, this relatively fast and cost-friendly assay is sensitive to Rivaroxaban and integrates its effects on the global coagulant state of patients. Anti-Xa activity assays can be run to assess the actual Rivaroxaban concentration and in the future ETP could serve as a fine-tuned hemostatic balance indicator for patients using Rivaroxaban.


Corresponding author: Dr. Anja Leyte, Onze Lieve Vrouwe Gasthuis Department HKCL, Oosterparkstraat 9, 1091 AC, Amsterdam, The Netherlands Phone: +31 205993049, Fax: +31 205993150

Received: 2012-1-26
Accepted: 2012-3-8
Published Online: 2012-03-30
Published in Print: 2012-10-01

©2012 by Walter de Gruyter Berlin Boston

Articles in the same Issue

  1. Masthead
  2. Masthead
  3. Editorials
  4. Cancer epigenomics: moving slowly, but at a steady pace from laboratory bench to clinical practice
  5. Laboratory networking and sample quality: a still relevant issue for patient safety
  6. Reviews
  7. DNA methylation biomarkers and their utility for solid cancer diagnostics
  8. DNA methylation biomarkers in biological fluids for early detection of respiratory tract cancer
  9. Global DNA hypomethylation in cancer: review of validated methods and clinical significance
  10. Genetics and Molecular Diagnostics
  11. δ0-Thalassemia in cis of βKnossos globin gene: first homozygous description in thalassemia intermedia Libyans and first combination with codon 39 (C→T) in thalassemia intermedia Tunisian patients
  12. Genetic-based reference values for angiotensin-converting enzyme (ACE) according to I/D polymorphism in a Spanish population sample
  13. General Clinical Chemistry and Laboratory Medicine
  14. Effects of sample transportation on commonly requested laboratory tests
  15. Long-term (in)stability of folate and vitamin B12 in human serum
  16. Information comparison of the effects of drugs on laboratory tests in drug labels and Young’s book
  17. First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis
  18. Antioxidant and oligonutrient status, distribution of amino acids, muscle damage, inflammation, and evaluation of renal function in elite rugby players
  19. Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids
  20. Measuring Rivaroxaban in a clinical laboratory setting, using common coagulation assays, Xa inhibition and thrombin generation
  21. Cancer Diagnostics
  22. Increased concentrations of growth factors and activation of the EGFR system in breast cancer
  23. Infectious Diseases
  24. Incidence and significance of elevated lactate in the identification of critically ill patients
  25. Cardiovascular Diseases
  26. Lipoprotein-associated phospholipase A2 is elevated in patients with severe aortic valve stenosis without clinically overt atherosclerosis
  27. Diabetes
  28. Serum cystatin C as a marker for early detection of chronic kidney disease and grade 2 nephropathy in Japanese patients with type 2 diabetes
  29. Letters to the Editor
  30. Establishing professional phlebotomist group for quality control of the preanalytic variables of clinical laboratory
  31. Application of quality specification based on biological variation in planning quality control strategy
  32. Analytical and clinical evaluation of a new immunoassay for therapeutic drug monitoring of infliximab and adalimumab
  33. Comparison of relationships between FT4 and log TSH in Access DXI 800 Unicel, Modular E170 and ADVIA Centaur XP Analyzer
  34. Assay-specific pitfalls of catecholamine HPLC assays
  35. Evaluation of automated nucleated red blood cells counting on Sysmex XE5000 and Siemens ADVIA 2120
  36. Evaluation of the immature myeloid information (IMI) by Sysmex XE 2100® hematology analyzer in the identification of blasts of myeloid lineage
  37. Congress Abstracts
  38. Congress of Laboratory Medicine and Clinical Chemistry / 4th Annual Meeting of the Austrian Society for Laboratory Medicine and Clinical Chemistry (ÖGLMKC)
  39. Abstracts of the 2012 Convocation of Experts on Laboratory Quality / Embracing the Need of Continued Vigilance to Assure Quality Outcomes in the Modern Laboratory
Downloaded on 5.10.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2012-0055/html
Scroll to top button