Abstract
Background: HbA1c is a key biomarker for the monitoring of glycemic balance in diabetic patients. It may be measured by various methods, including HPLC and immunoassays. Here we report the results of the first evaluation of Capillarys 2 Flex Piercing®, a new analyzer using capillary electrophoresis for the separation and the quantification of HbA1c.
Methods: We have evaluated the analytical performances of the method as well as the influence of the most frequent analytical interferences regarding HbA1c assays (i.e., labile HbA1c, carbamylated hemoglobin, high HbF and hemoglobin variants).
Results: Intra-assay and between-assays CVs were respectively lower than 1.98% and 2.68% on the pre-market version of the instrument, and lower than 1.62% and 1.45% on the commercial version. The linearity was excellent for HbA1c values ranging from 19 mmol/mol (3.9%) to 161 mmol/mol (16.9%) (r=0.999). The results were well correlated with those obtained by the HPLC method routinely used in the laboratory (Variant II® NU Kit – Bio-Rad): HbA1c[Capillarys 2]=0.9452×HbA1c[Variant II]+1.7279 (r=0.994, n=500). The use of titrated quality control samples indicated a good accuracy of the method. Neither the presence of HbF (until 15%), labile HbA1c or carbamylated hemoglobin, nor that of some typical hemoglobin variants, such as hemoglobin S, D, C and E, affected HbA1c measurement.
Conclusions: This evaluation showed that the analytical performances of Capillarys 2 Flex Piercing® analyzer for HbA1c assay fulfilled quality criteria requested for clinical use, and allowed to recommend its implementation in clinical chemistry laboratories for routine practice.
©2012 by Walter de Gruyter Berlin Boston
Articles in the same Issue
- Masthead
- Masthead
- Editorials
- Cancer epigenomics: moving slowly, but at a steady pace from laboratory bench to clinical practice
- Laboratory networking and sample quality: a still relevant issue for patient safety
- Reviews
- DNA methylation biomarkers and their utility for solid cancer diagnostics
- DNA methylation biomarkers in biological fluids for early detection of respiratory tract cancer
- Global DNA hypomethylation in cancer: review of validated methods and clinical significance
- Genetics and Molecular Diagnostics
- δ0-Thalassemia in cis of βKnossos globin gene: first homozygous description in thalassemia intermedia Libyans and first combination with codon 39 (C→T) in thalassemia intermedia Tunisian patients
- Genetic-based reference values for angiotensin-converting enzyme (ACE) according to I/D polymorphism in a Spanish population sample
- General Clinical Chemistry and Laboratory Medicine
- Effects of sample transportation on commonly requested laboratory tests
- Long-term (in)stability of folate and vitamin B12 in human serum
- Information comparison of the effects of drugs on laboratory tests in drug labels and Young’s book
- First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis
- Antioxidant and oligonutrient status, distribution of amino acids, muscle damage, inflammation, and evaluation of renal function in elite rugby players
- Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids
- Measuring Rivaroxaban in a clinical laboratory setting, using common coagulation assays, Xa inhibition and thrombin generation
- Cancer Diagnostics
- Increased concentrations of growth factors and activation of the EGFR system in breast cancer
- Infectious Diseases
- Incidence and significance of elevated lactate in the identification of critically ill patients
- Cardiovascular Diseases
- Lipoprotein-associated phospholipase A2 is elevated in patients with severe aortic valve stenosis without clinically overt atherosclerosis
- Diabetes
- Serum cystatin C as a marker for early detection of chronic kidney disease and grade 2 nephropathy in Japanese patients with type 2 diabetes
- Letters to the Editor
- Establishing professional phlebotomist group for quality control of the preanalytic variables of clinical laboratory
- Application of quality specification based on biological variation in planning quality control strategy
- Analytical and clinical evaluation of a new immunoassay for therapeutic drug monitoring of infliximab and adalimumab
- Comparison of relationships between FT4 and log TSH in Access DXI 800 Unicel, Modular E170 and ADVIA Centaur XP Analyzer
- Assay-specific pitfalls of catecholamine HPLC assays
- Evaluation of automated nucleated red blood cells counting on Sysmex XE5000 and Siemens ADVIA 2120
- Evaluation of the immature myeloid information (IMI) by Sysmex XE 2100® hematology analyzer in the identification of blasts of myeloid lineage
- Congress Abstracts
- Congress of Laboratory Medicine and Clinical Chemistry / 4th Annual Meeting of the Austrian Society for Laboratory Medicine and Clinical Chemistry (ÖGLMKC)
- Abstracts of the 2012 Convocation of Experts on Laboratory Quality / Embracing the Need of Continued Vigilance to Assure Quality Outcomes in the Modern Laboratory
Articles in the same Issue
- Masthead
- Masthead
- Editorials
- Cancer epigenomics: moving slowly, but at a steady pace from laboratory bench to clinical practice
- Laboratory networking and sample quality: a still relevant issue for patient safety
- Reviews
- DNA methylation biomarkers and their utility for solid cancer diagnostics
- DNA methylation biomarkers in biological fluids for early detection of respiratory tract cancer
- Global DNA hypomethylation in cancer: review of validated methods and clinical significance
- Genetics and Molecular Diagnostics
- δ0-Thalassemia in cis of βKnossos globin gene: first homozygous description in thalassemia intermedia Libyans and first combination with codon 39 (C→T) in thalassemia intermedia Tunisian patients
- Genetic-based reference values for angiotensin-converting enzyme (ACE) according to I/D polymorphism in a Spanish population sample
- General Clinical Chemistry and Laboratory Medicine
- Effects of sample transportation on commonly requested laboratory tests
- Long-term (in)stability of folate and vitamin B12 in human serum
- Information comparison of the effects of drugs on laboratory tests in drug labels and Young’s book
- First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis
- Antioxidant and oligonutrient status, distribution of amino acids, muscle damage, inflammation, and evaluation of renal function in elite rugby players
- Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids
- Measuring Rivaroxaban in a clinical laboratory setting, using common coagulation assays, Xa inhibition and thrombin generation
- Cancer Diagnostics
- Increased concentrations of growth factors and activation of the EGFR system in breast cancer
- Infectious Diseases
- Incidence and significance of elevated lactate in the identification of critically ill patients
- Cardiovascular Diseases
- Lipoprotein-associated phospholipase A2 is elevated in patients with severe aortic valve stenosis without clinically overt atherosclerosis
- Diabetes
- Serum cystatin C as a marker for early detection of chronic kidney disease and grade 2 nephropathy in Japanese patients with type 2 diabetes
- Letters to the Editor
- Establishing professional phlebotomist group for quality control of the preanalytic variables of clinical laboratory
- Application of quality specification based on biological variation in planning quality control strategy
- Analytical and clinical evaluation of a new immunoassay for therapeutic drug monitoring of infliximab and adalimumab
- Comparison of relationships between FT4 and log TSH in Access DXI 800 Unicel, Modular E170 and ADVIA Centaur XP Analyzer
- Assay-specific pitfalls of catecholamine HPLC assays
- Evaluation of automated nucleated red blood cells counting on Sysmex XE5000 and Siemens ADVIA 2120
- Evaluation of the immature myeloid information (IMI) by Sysmex XE 2100® hematology analyzer in the identification of blasts of myeloid lineage
- Congress Abstracts
- Congress of Laboratory Medicine and Clinical Chemistry / 4th Annual Meeting of the Austrian Society for Laboratory Medicine and Clinical Chemistry (ÖGLMKC)
- Abstracts of the 2012 Convocation of Experts on Laboratory Quality / Embracing the Need of Continued Vigilance to Assure Quality Outcomes in the Modern Laboratory