Abstract
Background: We evaluated the body fluid (BF) module on the new Sysmex XN-1000 for counting blood cells.
Methods: One hundred and eighty-seven BF samples [73 cerebrospinal fluid (CSF), 48 continuous ambulatory peritoneal dialysis (CAPD), 46 ascites, and 20 pleural fluid] were used for method comparison between the XN-1000 and manual microscopy (Fuchs-Rosenthal chamber and stained cytospin slides) for counting red blood cells (RBCs) and white blood cells (WBCs) (differential).
Results: Good agreement was found for counting WBCs (y=1.06x+0.09, n=67, R2=0.96) and mononuclear cells (MNs) (y=1.04x–0.01, n=40, R2=0.93) in CSF. However, the XN-1000 systematically counted more polymorphonuclear cells (PMNs) (y=1.48x+0.18, n=40, R2=0.99) compared to manual microscopy. Excellent correlation for RBCs >1×109/L (y=0.99x+116.56, n=26, R2=0.99) in CSF was found. For other fluids (CAPD, ascites and pleural fluid) excellent agreement was found for counting WBCs (y=1.06x+0.26, n=109, R2=0.98), MNs (y=1.06x–0.41, n=93, R2=0.96), PMNs (y=1.06x+0.81, n=93, R2=0.98) and RBCs (y=1.04x+110.04, n=43, R2=0.98). By using BF XN-check, the lower limit of quantitation (LLoQ) for WBC was defined at 5×106/L. Linearity was excellent for both the WBCs (R2=0.99) and RBCs (R2=0.99) and carry-over never exceeded 0.05%.
Conclusions: The BF module on the XN-1000 is a suitable tool for fast and accurate quantification of WBC (differential) and RBC counts in CSF and other BFs in a diagnostic setting.
©2012 by Walter de Gruyter Berlin Boston
Articles in the same Issue
- Masthead
- Masthead
- Editorials
- Cancer epigenomics: moving slowly, but at a steady pace from laboratory bench to clinical practice
- Laboratory networking and sample quality: a still relevant issue for patient safety
- Reviews
- DNA methylation biomarkers and their utility for solid cancer diagnostics
- DNA methylation biomarkers in biological fluids for early detection of respiratory tract cancer
- Global DNA hypomethylation in cancer: review of validated methods and clinical significance
- Genetics and Molecular Diagnostics
- δ0-Thalassemia in cis of βKnossos globin gene: first homozygous description in thalassemia intermedia Libyans and first combination with codon 39 (C→T) in thalassemia intermedia Tunisian patients
- Genetic-based reference values for angiotensin-converting enzyme (ACE) according to I/D polymorphism in a Spanish population sample
- General Clinical Chemistry and Laboratory Medicine
- Effects of sample transportation on commonly requested laboratory tests
- Long-term (in)stability of folate and vitamin B12 in human serum
- Information comparison of the effects of drugs on laboratory tests in drug labels and Young’s book
- First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis
- Antioxidant and oligonutrient status, distribution of amino acids, muscle damage, inflammation, and evaluation of renal function in elite rugby players
- Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids
- Measuring Rivaroxaban in a clinical laboratory setting, using common coagulation assays, Xa inhibition and thrombin generation
- Cancer Diagnostics
- Increased concentrations of growth factors and activation of the EGFR system in breast cancer
- Infectious Diseases
- Incidence and significance of elevated lactate in the identification of critically ill patients
- Cardiovascular Diseases
- Lipoprotein-associated phospholipase A2 is elevated in patients with severe aortic valve stenosis without clinically overt atherosclerosis
- Diabetes
- Serum cystatin C as a marker for early detection of chronic kidney disease and grade 2 nephropathy in Japanese patients with type 2 diabetes
- Letters to the Editor
- Establishing professional phlebotomist group for quality control of the preanalytic variables of clinical laboratory
- Application of quality specification based on biological variation in planning quality control strategy
- Analytical and clinical evaluation of a new immunoassay for therapeutic drug monitoring of infliximab and adalimumab
- Comparison of relationships between FT4 and log TSH in Access DXI 800 Unicel, Modular E170 and ADVIA Centaur XP Analyzer
- Assay-specific pitfalls of catecholamine HPLC assays
- Evaluation of automated nucleated red blood cells counting on Sysmex XE5000 and Siemens ADVIA 2120
- Evaluation of the immature myeloid information (IMI) by Sysmex XE 2100® hematology analyzer in the identification of blasts of myeloid lineage
- Congress Abstracts
- Congress of Laboratory Medicine and Clinical Chemistry / 4th Annual Meeting of the Austrian Society for Laboratory Medicine and Clinical Chemistry (ÖGLMKC)
- Abstracts of the 2012 Convocation of Experts on Laboratory Quality / Embracing the Need of Continued Vigilance to Assure Quality Outcomes in the Modern Laboratory
Articles in the same Issue
- Masthead
- Masthead
- Editorials
- Cancer epigenomics: moving slowly, but at a steady pace from laboratory bench to clinical practice
- Laboratory networking and sample quality: a still relevant issue for patient safety
- Reviews
- DNA methylation biomarkers and their utility for solid cancer diagnostics
- DNA methylation biomarkers in biological fluids for early detection of respiratory tract cancer
- Global DNA hypomethylation in cancer: review of validated methods and clinical significance
- Genetics and Molecular Diagnostics
- δ0-Thalassemia in cis of βKnossos globin gene: first homozygous description in thalassemia intermedia Libyans and first combination with codon 39 (C→T) in thalassemia intermedia Tunisian patients
- Genetic-based reference values for angiotensin-converting enzyme (ACE) according to I/D polymorphism in a Spanish population sample
- General Clinical Chemistry and Laboratory Medicine
- Effects of sample transportation on commonly requested laboratory tests
- Long-term (in)stability of folate and vitamin B12 in human serum
- Information comparison of the effects of drugs on laboratory tests in drug labels and Young’s book
- First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis
- Antioxidant and oligonutrient status, distribution of amino acids, muscle damage, inflammation, and evaluation of renal function in elite rugby players
- Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids
- Measuring Rivaroxaban in a clinical laboratory setting, using common coagulation assays, Xa inhibition and thrombin generation
- Cancer Diagnostics
- Increased concentrations of growth factors and activation of the EGFR system in breast cancer
- Infectious Diseases
- Incidence and significance of elevated lactate in the identification of critically ill patients
- Cardiovascular Diseases
- Lipoprotein-associated phospholipase A2 is elevated in patients with severe aortic valve stenosis without clinically overt atherosclerosis
- Diabetes
- Serum cystatin C as a marker for early detection of chronic kidney disease and grade 2 nephropathy in Japanese patients with type 2 diabetes
- Letters to the Editor
- Establishing professional phlebotomist group for quality control of the preanalytic variables of clinical laboratory
- Application of quality specification based on biological variation in planning quality control strategy
- Analytical and clinical evaluation of a new immunoassay for therapeutic drug monitoring of infliximab and adalimumab
- Comparison of relationships between FT4 and log TSH in Access DXI 800 Unicel, Modular E170 and ADVIA Centaur XP Analyzer
- Assay-specific pitfalls of catecholamine HPLC assays
- Evaluation of automated nucleated red blood cells counting on Sysmex XE5000 and Siemens ADVIA 2120
- Evaluation of the immature myeloid information (IMI) by Sysmex XE 2100® hematology analyzer in the identification of blasts of myeloid lineage
- Congress Abstracts
- Congress of Laboratory Medicine and Clinical Chemistry / 4th Annual Meeting of the Austrian Society for Laboratory Medicine and Clinical Chemistry (ÖGLMKC)
- Abstracts of the 2012 Convocation of Experts on Laboratory Quality / Embracing the Need of Continued Vigilance to Assure Quality Outcomes in the Modern Laboratory