Abstract
Objectives
The pre-analytical stability of various biochemical analytes requires careful consideration, as it can lead to the release of erroneous laboratory results. There is currently significant variability in the literature regarding the pre-analytical stability of various analytes. The aim of this study was to determine the pre-analytical stability of 65 analytes in whole blood, serum and plasma using a standardized approach.
Methods
Blood samples were collected from 30 healthy volunteers (10 volunteers per analyte) into five vacutainers; either SST, Li-heparin, K2-EDTA, or Na-fluoride/K-oxalate. Several conditions were tested, including delayed centrifugation with storage of whole blood at room temperature (RT) for 8 h, delayed centrifugation with storage of whole blood at RT or 4 °C for 24 h, and immediate centrifugation with storage of plasma or serum at RT for 24 h. Percent deviation (% PD) from baseline was calculated for each analyte and compared to the maximum permissible instability (MPI) derived from intra- and inter-individual biological variation.
Results
The majority of the analytes evaluated remained stable across all vacutainer types, temperatures, and timepoints tested. Glucose, potassium, and aspartate aminotransferase, among others, were significantly impacted by delayed centrifugation, having been found to be unstable in whole blood specimens stored at room temperature for 8 h.
Conclusions
The data presented provides insight into the pre-analytical variables that impact the stability of routine biochemical analytes. This study may help to reduce the frequency of erroneous laboratory results released due to exceeded stability and reduce unnecessary repeat phlebotomy for analytes that remain stable despite delayed processing.
-
Research ethics: Not applicable.
-
Informed consent: Not applicable.
-
Author contributions: The authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: The authors state no conflicts of interest.
-
Research funding: None declared.
-
Data availability: Not applicable.
References
1. Spithoven, EM, Bakker, SJL, Kootstra-Ros, JE, Jong, DPE, Gansevoort, RT. Stability of creatinine and cystatin C in whole blood. Clin Biochem 2013;46:1611–4. https://doi.org/10.1016/j.clinbiochem.2013.06.022.Search in Google Scholar PubMed
2. Oddoze, C, Lombard, E, Portugal, H. Stability study of 81 analytes in human whole blood, in serum and in plasma. Clin Biochem 2012;45:464–9. https://doi.org/10.1016/j.clinbiochem.2012.01.012.Search in Google Scholar PubMed
3. Monneret, D, Godmer, A, Le Guen, R, Bravetti, C, Emeraud, C, Marteau, A, et al.. Stability of routine biochemical analytes in whole blood and plasma from lithium heparin gel tubes during 6-hr storage. J Clin Lab Anal 2016;30:602–9. https://doi.org/10.1002/jcla.21909.Search in Google Scholar PubMed PubMed Central
4. Dupuy, AM, Badiou, S, Daubin, D, Bargnoux, AS, Magnan, C, Klouche, K, et al.. Comparison of Barricor™ vs. lithium heparin tubes for selected routine biochemical analytes and evaluation of post centrifugation stability. Biochem Med 2018;28:020902. https://doi.org/10.11613/bm.2018.020902.Search in Google Scholar
5. Dupuy, AM, Cristol, JP, Vincent, B, Bargnoux, AS, Mendes, M, Philibert, P, et al.. Stability of routine biochemical analytes in whole blood and plasma/serum: focus on potassium stability from lithium heparin. Clin Chem Lab Med 2018;56:413–21. https://doi.org/10.1515/cclm-2017-0292.Search in Google Scholar PubMed
6. Leino, A, Koivula, MK. Stability of chemical and immunochemical analytes in uncentrifuged plasma samples. Ann Clin Biochem 2009;46:159–61. https://doi.org/10.1258/acb.2008.008212.Search in Google Scholar PubMed
7. Hedayati, M, Razavi, SA, Boroomand, S, Kheradmand Kia, S. The impact of pre-analytical variations on biochemical analytes stability: a systematic review. J Clin Lab Anal 2020;34:e23551–1. https://doi.org/10.1002/jcla.23551.Search in Google Scholar PubMed PubMed Central
8. Guder, WG, da Fonseca-Wollheim, F, Heil, W, Schmitt, Y, Töpfer, G, Wisser, H, et al.. The quality of diagnostic samples. Recommendations of the working group on preanalytical quality of the German Society for Clinical Chemistry and the German Society for Laboratory Medicine, in DGKL. Oxford, UK: Becton Dickinson (BD); 2010: 3–78 pp.Search in Google Scholar
9. Foucher, B, Pina, G, Desjeux, G, Prevosto, JM, Chaulet, JF, Cheminel, V. Influence of temperature and delayed centrifugation: stability studies of 28 analytes currently analysed. Ann Biol Clin 2005;63:93–100.Search in Google Scholar
10. Tanner, M, Kent, N, Smith, B, Fletcher, S, Lewer, M. Stability of common biochemical analytes in serum gel tubes subjected to various storage temperatures and times pre-centrifugation. Ann Clin Biochem 2008;45:375–9. https://doi.org/10.1258/acb.2007.007183.Search in Google Scholar PubMed
11. Zwart, SR, Wolf, M, Rogers, A, Rodgers, S, Gillman, PL, Hitchcox, K, et al.. Stability of analytes related to clinical chemistry and bone metabolism in blood specimens after delayed processing. Clin Biochem 2009;42:907–10. https://doi.org/10.1016/j.clinbiochem.2009.02.010.Search in Google Scholar PubMed
12. Cornes, M, Simundic, A-M, Cadamuro, J, Costelloe, JS, Baird, G, Kristensen, BBG, et al.. The CRESS checklist for reporting stability studies: on behalf of the European federation of clinical chemistry and laboratory medicine (EFLM) working group for the preanalytical phase (WG-PRE). Clin Chem Lab Med 2021;59:59–69. https://doi.org/10.1515/cclm-2020-0061.Search in Google Scholar PubMed
13. Gomez-Rioja, R, Meyer, VA, Cornes, M, Costelloe, S, Vermeersch, P, Simundic, A-M, et al.. Recommendation for the design of stability studies on clinical specimens. Clin Chem Lab Med 2023;61:1708–18. https://doi.org/10.1515/cclm-2023-0221.Search in Google Scholar PubMed
14. Aarsand, AK, Fernandez-Calle, P, Webster, C, Coskun, A, Gonzales-Lao, E, Diaz-Garzon, J, et al.. The EFLM biological variation database. Available at: https://biologicalvariation.eu/.Search in Google Scholar
15. Ho, KMC, Chen, C, Setoh, WSJ, Yap, WTW, Hawkins, CWR. Optimization of hemolysis, icterus and lipemia interference thresholds for 35 clinical chemistry assays. Pract Lab Med 2021;25:e00232. https://doi.org/10.1016/j.plabm.2021.e00232.Search in Google Scholar PubMed PubMed Central
16. Karin, A, Higgins, V, Miller, J, Brinc, D, Kulasingam, V, Selvaratnam, R. Evaluation of hemolysis, lipemia, and icterus interference with common clinical immunoassays. Clin Chem Lab Med 2023;61:1035–45. https://doi.org/10.1515/cclm-2022-0924.Search in Google Scholar PubMed
17. Gómez-Rioja, R, Amaro, SM, Diaz-Garzón, J, Bauçà, MJ, Espartosa, MD, Fernández-Calle, P. A protocol for testing the stability of biochemical analytes. Technical document. Clin Chem Lab Med 2019;57:1829–36. https://doi.org/10.1515/cclm-2019-0586.Search in Google Scholar PubMed
18. Henriksen, LO, Faber, NR, Moller, MF, Nexo, E, Hansen, AB. Stability of 35 biochemical and immunological routine tests after 10 hours storage and transport of human whole blood at 21 °C. Scand J Clin Lab Invest 2014;74:603–10. https://doi.org/10.3109/00365513.2014.928940.Search in Google Scholar PubMed PubMed Central
19. Boyanton, BL, Blick, KE. Stability studies of twenty-four analytes in human plasma and serum. Clin Chem 2002;48:2242–7. https://doi.org/10.1093/clinchem/48.12.2242.Search in Google Scholar
20. Stahl, M, Brandslund, I. Controlled storage conditions prolong stability of biochemical components in whole blood. Clin Chem Lab Med 2005;43:210–5. https://doi.org/10.1515/cclm.2005.036.Search in Google Scholar
21. Zavorsky, GS, van Wijk, XMR, Gasparyan, S, Stollenwerk, NS, Brooks, RA. Stability of whole blood electrolyte specimens at room temperature vs. Slushed ice conditions. J Appl Lab Med 2022;7:541–54. https://doi.org/10.1093/jalm/jfab089.Search in Google Scholar PubMed
22. Gobinet-Georges, A, Valli, N, Filliatre, H, Dubernet, M-F, Dedeystere, O, Bordenave, L. Stability of brain natriuretic peptide (BNP) in human whole blood and plasma. Clin Chem Lab Med 2000;38:519–23. https://doi.org/10.1515/cclm.2000.076.Search in Google Scholar
23. Chakera, JA, Mcdonald, JT, Knight, AB, Vaidya, B, Jones, GA. Current laboratory requirements for adrenocorticotropic hormone and renin/aldosterone sample handling are unnecessarily restrictive. J Clin Med 2017;17:18–21. https://doi.org/10.7861/clinmedicine.17-1-18.Search in Google Scholar PubMed PubMed Central
24. Sofronescu, GA, Loebs, T, Zhu, Y. Effects of temperature and light on the stability of bilirubin in plasma samples. Clin Chim Acta 2012;413:463–6. https://doi.org/10.1016/j.cca.2011.10.036.Search in Google Scholar PubMed
25. Sandhu, N, Karlsen, AM, Høgdall, C, Laursen, AI, Christensen, JI, Høgdall, SVE. Stability of HE4 and CA125 in blood samples from patients diagnosed with ovarian cancer. Scand J Clin Lab Invest 2014;74:477–84. https://doi.org/10.3109/00365513.2014.903430.Search in Google Scholar PubMed
26. Canki, E, Schuurbiers, MM, Linders, CT, Korse, MC, Heuvel, DVMM, Herwaarden, VEA, et al.. Pre-analytical stability of the CEA, CYFRA 21.1, NSE, CA125 and HE4 tumor markers. Tumor Biol 2023:1–11. https://doi.org/10.3233/tub-220013.Search in Google Scholar PubMed
27. Diver, JM, Hughes, GJ, Hutton, LJ, West, RC, Hipkin, JL. The long-term stability in whole blood of 14 commonly-requested hormone analytes. Ann Clin Biochem 1994;31:561–5. https://doi.org/10.1177/000456329403100606.Search in Google Scholar PubMed
28. Sealey, EJ, Blumenfeld, J, Laragh, HJ. Prorenin cryoactivation as a possible cause of normal renin levels in patients with primary aldosteronism. J Hypertens 2005;23:459–60. https://doi.org/10.1097/00004872-200502000-00030.Search in Google Scholar PubMed
29. Carter, S, Owen, JL, Kerstens, NM, Dullaart, FPR, Keevil, GB. A liquid chromatography tandem mass spectrometry assay for plasma renin activity using online solid-phase extraction. Ann Clin Biochem 2012;49:570–9. https://doi.org/10.1258/acb.2012.011186.Search in Google Scholar PubMed
30. Hepburn, S, Munday, C, Taylor, K, Halsall, JD. Stability of direct renin concentration and plasma renin activity in EDTA whole blood and plasma at ambient and refrigerated temperatures from 0 to 72 hours. Clin Chem Lab Med 2022;60:1384–92. https://doi.org/10.1515/cclm-2022-0375.Search in Google Scholar PubMed
31. Khalil, H, Borai, A, Dakhakhni, M, Bahijri, S, Faizo, H, Bokhari, FF, et al.. Stability and validity of intact parathyroid hormone levels in different sample types and storage conditions. J Clin Lab Anal 2021;35:1–8. https://doi.org/10.1002/jcla.23771.Search in Google Scholar PubMed PubMed Central
32. Wielders, MPJ, Wijnberg, AF. Preanalytical stability of 25(OH)–vitamin D3 in human blood or serum at room temperature: solid as a rock. Clin Chem 2009;55:1584–5. https://doi.org/10.1373/clinchem.2008.117366.Search in Google Scholar PubMed
33. Gao, Y-C, Yang, Y-D, Yuan, Z-B, Lu, H-K. Serum thyroglobulin stability for immunoassay. Lab Med 2007;38:618–20. https://doi.org/10.1309/hubpyxqku0jadfn2.Search in Google Scholar
34. Spithoven, EM, Bakker, SJL, Kootstra-Ros, JE, de Jong, PE, Gansevoort, RT, Investigators, DC. Stability of creatinine and cystatin C in whole blood. Clin Biochem 2013;46:1611–4. https://doi.org/10.1016/j.clinbiochem.2013.06.022.Search in Google Scholar PubMed
35. Hartog, H, van der Graaf, WT, Wesseling, J, van der Veer, E, Boezen, HM. Measurement of insulin-like growth factor-1 and insulin-like growth factor binding protein-3 after delayed separation of whole blood samples. Clin Biochem 2008;41:636–9. https://doi.org/10.1016/j.clinbiochem.2008.02.001.Search in Google Scholar PubMed
36. Elmlinger, MW, Zwirner, M, Kuhnel, W. Stability of insulin-like growth factor (IGF)-I and IGF binding protein (IGFBP)-3 measured by the IMMULITE automated chemiluminescence assay system in different blood specimens. Clin Lab 2005;51:145–52.Search in Google Scholar
37. Dromigny, J-A, Robert, E. Stability of blood potassium: effects of duration, temperature and transport during 10 hours storage of human whole blood in serum and plasma. Ann Biol Clin 2017;75:369–74. https://doi.org/10.1684/abc.2017.1255.Search in Google Scholar PubMed
38. Hanon, AE, Sturgeon, MC, Lamb, JE. Sampling and storage conditions influencing the measurement of parathyroid hormone in blood samples: a systematic review. Clin Chem Lab Med 2013;51:1925–41. https://doi.org/10.1515/cclm-2013-0315.Search in Google Scholar PubMed
39. Nielsen, BK, Frederiksen, T, Friis-Hansen, L, Larsen, PB. Post-analytical stability of 23 common chemistry and immunochemistry analytes in incurred samples. Clin Biochem 2017;50:1175–82. https://doi.org/10.1016/j.clinbiochem.2017.08.003.Search in Google Scholar PubMed
40. Dai, Z, Zhang, J-W, Lin, J, Zhu, J-H, Lin, Y-Z, Wang, X-X, et al.. The stability of intact parathyroid hormone (PTH) in different types of blood collection tubes. Clin Lab 2022;68. https://doi.org/10.7754/clin.lab.2021.210421.Search in Google Scholar
41. Bowen, ARR, Remaley, TA. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med 2014;24:31–44. https://doi.org/10.11613/bm.2014.006.Search in Google Scholar PubMed PubMed Central
42. Gawria, G, Tillmar, L, Landberg, E. A comparison of stability of chemical analytes in plasma from the BD Vacutainer® Barricor™ tube with mechanical separator versus tubes containing gel separator. J Clin Lab Anal 2020;34:1–8. https://doi.org/10.1002/jcla.23060.Search in Google Scholar PubMed PubMed Central
43. Erickson, AJ, Chiang, F-I, Walker, MC, Genzen, RJ, Doyle, K. Comparison of two chromogranin A assays and investigation of nonlinear specimens. Pract Lab Med 2022;32:e00299. https://doi.org/10.1016/j.plabm.2022.e00299.Search in Google Scholar PubMed PubMed Central
44. Pedersen, L, Nybo, M. Preanalytical factors of importance for measurement of Chromogranin A. Clin Chim Acta 2014;436:41–4. https://doi.org/10.1016/j.cca.2014.04.026.Search in Google Scholar PubMed
45. Thirkettle, S, Blaszczyk, P, Evans, R, Wheatley, M, Abbas, M, Russell, J, et al.. Stability assessment of serum tumour markers: calcitonin, chromogranin A, thyroglobulin and anti-thyroglobulin antibodies. Ann Clin Biochem 2023;60:212–6. https://doi.org/10.1177/00045632231160371.Search in Google Scholar PubMed
46. Cauliez, B, Guignery, J, Marinier, S, Mariau, I, Lavoinne, A. Two-year stability of NT-proBNP in frozen samples using the Roche Elecsys system. Ann Clin Biochem 2008;45:318–9. https://doi.org/10.1258/acb.2007.007187.Search in Google Scholar PubMed
47. Ellis, JM, Livesey, HJ, Evans, JM. Hormone stability in human whole blood. Clin Biochem 2003;36:109–12. https://doi.org/10.1016/s0009-9120(02)00440-x.Search in Google Scholar PubMed
48. Parra-Robert, M, Rico-Santana, N, Alcaraz-Quiles, J, Sandalinas, S, Fernández, E, Falcón, I, et al.. Improvement in the stability of serum samples stored in an automated refrigerated module. Clin Biochem 2016;49:1396–8. https://doi.org/10.1016/j.clinbiochem.2016.10.012.Search in Google Scholar PubMed
49. Cuhadar, S, Atay, A, Koseoglu, M, Dirican, A, Hur, A. Stability studies of common biochemical analytes in serum separator tubes with or without gel barrier subjected to various storage conditions. Biochem Med 2012;22:202–14. https://doi.org/10.11613/bm.2012.023.Search in Google Scholar PubMed PubMed Central
50. Balboni, F, Burbui, S, Lippi, G. Glucose variation in centrifuged serum and lithium-heparin gel tubes stored for up to 96 hours at room temperature or 4 °C. Scand J Clin Lab Invest 2018;78:546–50. https://doi.org/10.1080/00365513.2018.1517221.Search in Google Scholar PubMed
51. Graaf, DOA, Jonge, DR, Velmans, HM, Swaanenburg, CJMJ, Janssen, JWM. Evaluation of blood collection tubes specific for homocysteine measurement. Clin Chem Lab Med 2008;46:1652–4. https://doi.org/10.1515/cclm.2008.324.Search in Google Scholar PubMed
52. Narayanan, S. The preanalytic phase. Am J Clin Pathol 2000;113:429–52. https://doi.org/10.1309/c0nm-q7r0-ll2e-b3uy.Search in Google Scholar
53. Li, G, Cabanero, M, Wang, Z, Wang, H, Huang, T, Alexis, H, et al.. Comparison of glucose determinations on blood samples collected in three types of tubes. Ann Clin Lab Sci 2013;43:278–84.Search in Google Scholar
54. Peake, JM, Bruns, ED, Sacks, BD, Horvath, RA. It’s time for a better blood collection tube to improve the reliability of glucose results. Diabetes Care 2013;36:e2–2. https://doi.org/10.2337/dc12-1312.Search in Google Scholar PubMed PubMed Central
55. López, GM, Roca, RN, Velasco, SM, Martín, AJM, Soto, BA, Rioja, GR. Stability of lactate dehydrogenase in plasma at different temperatures: post-analytical storage. Adv Lab Med 2020;1:20200077. https://doi.org/10.1515/almed-2020-0077.Search in Google Scholar PubMed PubMed Central
56. Bidlingmaier, M, Friedrich, N, Emeny, TR, Spranger, J, Wolthers, DO, Roswall, J, et al.. Reference intervals for insulin-like growth factor-1 (IGF-I) from birth to senescence: results from a multicenter study using a new automated chemiluminescence IGF-I immunoassay conforming to recent international recommendations. J Clin Endocrinol 2014;99:1712–21. https://doi.org/10.1210/jc.2013-3059.Search in Google Scholar PubMed
57. Sandberg, S, Fraser, GC, Horvath, RA, Jansen, R, Jones, G, Oosterhuis, W, et al.. Defining analytical performance specifications: consensus statement from the 1st strategic conference of the European federation of clinical chemistry and laboratory medicine. Clin Chem Lab Med 2015;53:833–5. https://doi.org/10.1515/cclm-2015-0067.Search in Google Scholar PubMed
58. Evans, MJ, Livesey, JH, Ellis, MJ, Yandle, TG. Effect of anticoagulants and storage temperatures on stability of plasma and serum hormones. Clin Biochem 2001;34:107–12. https://doi.org/10.1016/s0009-9120(01)00196-5.Search in Google Scholar PubMed
59. El-Khoury, MJ, Wang, S. Stability of 1,25-dihydroxyvitamin D2 and 1,25-dihydroxyvitamin D3 in human serum. Clin Biochem 2012;45:707–8. https://doi.org/10.1016/j.clinbiochem.2012.03.016.Search in Google Scholar PubMed
60. Colak, A, Toprak, B, Dogan, N, Ustuner, F. Effect of sample type, centrifugation and storage conditions on vitamin D concentration. Biochem Med 2013;23:321–5. https://doi.org/10.11613/bm.2013.039.Search in Google Scholar PubMed PubMed Central
Supplementary Material
This article contains supplementary material (https://doi.org/10.1515/cclm-2023-1192).
© 2024 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Editorial
- Analytical performance specifications – moving from models to practical recommendations
- Opinion Papers
- What the Milan conference has taught us about analytical performance specification model definition and measurand allocation
- The role of analytical performance specifications in international guidelines and standards dealing with metrological traceability in laboratory medicine
- How clinical laboratories select and use Analytical Performance Specifications (APS) in Italy
- Outcome-based analytical performance specifications: current status and future challenges
- Analytical performance specifications based on biological variation data – considerations, strengths and limitations
- State-of-the-art model for derivation of analytical performance specifications: how to define the highest level of analytical performance technically achievable
- Analytical performance specifications for combined uncertainty budget in the implementation of metrological traceability
- When bias becomes part of imprecision: how to use analytical performance specifications to determine acceptability of lot-lot variation and other sources of possibly unacceptable bias
- Using analytical performance specifications in a medical laboratory
- Issues in assessing analytical performance specifications in healthcare systems assembling multiple laboratories and measuring systems
- Applying the Milan models to setting analytical performance specifications – considering all the information
- Guidelines and Recommendations
- Recommendations for blood sampling in emergency departments from the European Society for Emergency Medicine (EUSEM), European Society for Emergency Nursing (EuSEN), and European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for the Preanalytical Phase. Executive summary
- General Clinical Chemistry and Laboratory Medicine
- Assessment of accuracy of laboratory testing results, relative to peer group consensus values in external quality control, by bivariate z-score analysis: the example of D-Dimer
- The stability of 65 biochemistry analytes in plasma, serum, and whole blood
- Assessment of the 2023 European Kidney Function Consortium (EKFC) equations in a Chinese adult population
- In vitro, in vivo metabolism and quantification of the novel synthetic opioid N-piperidinyl etonitazene (etonitazepipne)
- Quantification of blood glial fibrillary acidic protein using a second-generation microfluidic assay. Validation and comparative analysis with two established assays
- Association of prehospital lactate levels with base excess in various emergencies – a retrospective study
- Reference Values and Biological Variations
- Stability of ten serum tumor markers after one year of storage at −18°C
- Variations in tumor growth, intra-individual biological variability, and the interpretation of changes
- Cancer Diagnostics
- N-linked glycosylation of the M-protein variable region: glycoproteogenomics reveals a new layer of personalized complexity in multiple myeloma
- Cardiovascular Diseases
- Reference intervals for high sensitivity cardiac troponin I and N-terminal pro-B-type natriuretic peptide in children and adolescents on the Siemens Atellica
- Infectious Diseases
- Fetal chronic hypoxia does not affect urinary presepsin levels in newborns at birth
- Letters to the Editor
- AWMF statement on medical services in laboratory diagnostics and pathology with regard to the IVDR
- An outline of measurement uncertainty of total protein in urine estimated according to the ISO Technical Specification 20914
- Early diagnosis of severe illness in an outpatient – the Sysmex XN’s neutrophil reactivity parameter
- Analytical and diagnostic performance of Theradiag i-Tracker assays on IDS-iSYS for infliximab and adalimumab therapeutic drug monitoring
- Improving the diagnosis of AATD with aid of serum protein electrophoresis: a prospective, multicentre, validation study
- Cerebrospinal fluid kappa free light chains in patients with tumefactive demyelination
- Diagnostic value of quantitative chemiluminescence immunoassay for anti-gp210 and anti-sp100 antibodies in primary biliary cholangitis
Articles in the same Issue
- Frontmatter
- Editorial
- Analytical performance specifications – moving from models to practical recommendations
- Opinion Papers
- What the Milan conference has taught us about analytical performance specification model definition and measurand allocation
- The role of analytical performance specifications in international guidelines and standards dealing with metrological traceability in laboratory medicine
- How clinical laboratories select and use Analytical Performance Specifications (APS) in Italy
- Outcome-based analytical performance specifications: current status and future challenges
- Analytical performance specifications based on biological variation data – considerations, strengths and limitations
- State-of-the-art model for derivation of analytical performance specifications: how to define the highest level of analytical performance technically achievable
- Analytical performance specifications for combined uncertainty budget in the implementation of metrological traceability
- When bias becomes part of imprecision: how to use analytical performance specifications to determine acceptability of lot-lot variation and other sources of possibly unacceptable bias
- Using analytical performance specifications in a medical laboratory
- Issues in assessing analytical performance specifications in healthcare systems assembling multiple laboratories and measuring systems
- Applying the Milan models to setting analytical performance specifications – considering all the information
- Guidelines and Recommendations
- Recommendations for blood sampling in emergency departments from the European Society for Emergency Medicine (EUSEM), European Society for Emergency Nursing (EuSEN), and European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for the Preanalytical Phase. Executive summary
- General Clinical Chemistry and Laboratory Medicine
- Assessment of accuracy of laboratory testing results, relative to peer group consensus values in external quality control, by bivariate z-score analysis: the example of D-Dimer
- The stability of 65 biochemistry analytes in plasma, serum, and whole blood
- Assessment of the 2023 European Kidney Function Consortium (EKFC) equations in a Chinese adult population
- In vitro, in vivo metabolism and quantification of the novel synthetic opioid N-piperidinyl etonitazene (etonitazepipne)
- Quantification of blood glial fibrillary acidic protein using a second-generation microfluidic assay. Validation and comparative analysis with two established assays
- Association of prehospital lactate levels with base excess in various emergencies – a retrospective study
- Reference Values and Biological Variations
- Stability of ten serum tumor markers after one year of storage at −18°C
- Variations in tumor growth, intra-individual biological variability, and the interpretation of changes
- Cancer Diagnostics
- N-linked glycosylation of the M-protein variable region: glycoproteogenomics reveals a new layer of personalized complexity in multiple myeloma
- Cardiovascular Diseases
- Reference intervals for high sensitivity cardiac troponin I and N-terminal pro-B-type natriuretic peptide in children and adolescents on the Siemens Atellica
- Infectious Diseases
- Fetal chronic hypoxia does not affect urinary presepsin levels in newborns at birth
- Letters to the Editor
- AWMF statement on medical services in laboratory diagnostics and pathology with regard to the IVDR
- An outline of measurement uncertainty of total protein in urine estimated according to the ISO Technical Specification 20914
- Early diagnosis of severe illness in an outpatient – the Sysmex XN’s neutrophil reactivity parameter
- Analytical and diagnostic performance of Theradiag i-Tracker assays on IDS-iSYS for infliximab and adalimumab therapeutic drug monitoring
- Improving the diagnosis of AATD with aid of serum protein electrophoresis: a prospective, multicentre, validation study
- Cerebrospinal fluid kappa free light chains in patients with tumefactive demyelination
- Diagnostic value of quantitative chemiluminescence immunoassay for anti-gp210 and anti-sp100 antibodies in primary biliary cholangitis