Quantification of polyclonal free light chains in clinical samples using a single turbidimetric immunoassay
-
Jeffrey M. Faint
Abstract
Background: Elevated polyclonal serum free light chain (FLC) levels have been associated with increased mortality and disease activity in many conditions. Currently, polyclonal FLC quantification requires summation of individual FLCκ and FLCλ assays. Here we present a single assay for combined FLC (cFLC, Combylite™) which reduces assay time and eliminates potential imprecision errors incurred by summating FLC assays (ΣFLC).
Methods: Sheep FLCκ- and FLCλ-specific antibodies were conjugated to latex microparticles to quantify FLCκ and FLCλ in a single assay. Combylite results were compared to ΣFLC (Freelite®) in 132 healthy controls and 1127 patient samples. The utility of cFLC for predicting all-cause mortality in a haematological referral population was evaluated.
Results: cFLC and ΣFLC results were highly concordant (Passing-Bablok equation y=0.98x–1.59 mg/L, R2=0.96). Combylite assay imprecision was low at concentrations around the upper normal range [coefficient of variation (CV) 5.5%, 54 mg/L] and the upper limit of the measuring range (CV 5.5%, 170 mg/L). cFLC levels were significantly raised in disease states compared with healthy controls. Additionally, cFLC >65 mg/L was associated with shorter overall survival in a haematological referral population (hazard ratio=4.5, p<0.001).
Conclusions: cFLC values obtained using Combylite were comparable to ΣFLC results over a wide concentration range, were elevated in diseases characterised by B cell activation and were associated with increased mortality in a haematological referral population. These observations indicate the Combylite assay has value for investigating the role of B cell activation in disparate disease groups and could be considered as a surrogate indication of B cell function.
Acknowledgments
The authors would like to thank Chris Buckley for his technical assistance in developing the Combylite assay.
Conflict of interest statement
Authors’ conflict of interest disclosure: Serum Freelite and Combylite kits were provided by The Binding Site Group, Ltd. SB, PAK, PC, DS, BKG, and CEJ have no conflicts of interest. The authors stated that there are no conflicts of interest regarding the publication of this article. Research funding played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.
Research funding: None declared.
Employment or leadership: JMF, LKA, PJS, and AM are employees of and HDC-S and SJH are directors of The Binding Site Group, Ltd.
Honorarium: None declared.
References
1. Arkfeld DG. The potential utility of B cell-directed biologic therapy in autoimmune diseases. Rheumatol Int 2008;28:205–15.10.1007/s00296-007-0471-xSuche in Google Scholar
2. Campbell KA, Lipinski MJ, Doran AC, Skaflen MD, Fuster V, McNamara CA. Lymphocytes and the adventitial immune response in atherosclerosis. Circ Res 2012;110:889–900.10.1161/CIRCRESAHA.111.263186Suche in Google Scholar
3. Waldmann TA, Strober W. Metabolism of immunoglobulins. Prog Allergy 1969;13:1–110.Suche in Google Scholar
4. Solling K. Free light chains of immunoglobulins. Scand J Clin Lab Invest Suppl 1981;157:1–83.Suche in Google Scholar
5. Abe M, Goto T, Kosaka M, Wolfenbarger D, Weiss DT, Solomon A. Differences in kappa to lambda (κ:λ) ratios of serum and urinary free light chains. Clin Exp Immunol 1998;111:457–62.10.1046/j.1365-2249.1998.00487.xSuche in Google Scholar
6. Abraham GN, Waterhouse C. Evidence for defective immunoglobulin metabolism in severe renal insufficiency. Am J Med Sci 1974;268:227–33.10.1097/00000441-197410000-00003Suche in Google Scholar
7. Epstein WV, Tan M. Increase of L-chain proteins in the sera of patients with systemic lupus erythematosus and the synovial fluids of patients with peripheral rheumatoid arthritis. Arthritis Rheum 1966;9:713–9.10.1002/art.1780090508Suche in Google Scholar
8. Brouwer J, Otting-van de Ruit M, Busking-van der Lely H. Estimation of free light chains of immunoglobulins by enzyme immunoassay. Clin Chim Acta 1985;150:267–74.10.1016/0009-8981(85)90254-2Suche in Google Scholar
9. Robinson EL, Gowland E, Ward ID, Scarffe JH. Radioimmunoassay of free light chains of immunoglobulins in urine. Clin Chem 1982;28:2254–8.10.1093/clinchem/28.11.2254Suche in Google Scholar
10. Bradwell AR, Carr-Smith HD, Mead GP, Tang LX, Showell PJ, Drayson MT, et al. Highly sensitive, automated immunoassay for immunoglobulin free light chains in serum and urine. Clin Chem 2001;47:673–80.10.1093/clinchem/47.4.673Suche in Google Scholar
11. Dispenzieri A, Kyle R, Merlini G, Miguel JS, Ludwig H, Hajek R, et al. International Myeloma Working Group guidelines for serum-free light chain analysis in multiple myeloma and related disorders. Leukemia 2009;23:215–24.10.1038/leu.2008.307Suche in Google Scholar PubMed
12. Anandram S, Assi LK, Lovatt T, Parkes J, Taylor J, MacWhannell A, et al. Elevated, combined serum free light chain levels and increased mortality: a 5-year follow-up, UK study. J Clin Pathol 2012;65:1036–42.10.1136/jclinpath-2012-200910Suche in Google Scholar PubMed
13. Dispenzieri A, Katzmann JA, Kyle RA, Larson DR, Therneau TM, Colby CL, et al. Use of nonclonal serum immunoglobulin free light chains to predict overall survival in the general population. Mayo Clin Proc 2012;87:517–23.10.1016/j.mayocp.2012.03.009Suche in Google Scholar PubMed PubMed Central
14. Aggarwal R, Sequeira W, Kokebie R, Mikolaitis RA, Fogg L, Finnegan A, et al. Serum free light chains as biomarkers for systemic lupus erythematosus disease activity. Arthritis Care Res 2011;63:891–8.10.1002/acr.20446Suche in Google Scholar PubMed
15. Kormelink TG, Tekstra J, Thurlings RM, Boumans MH, Vos K, Tak PP, et al. Decrease in immunoglobulin free light chains in patients with rheumatoid arthritis upon rituximab (anti-CD20) treatment correlates with decrease in disease activity. Ann Rheum Dis 2010;69:2137–44.10.1136/ard.2009.126441Suche in Google Scholar PubMed
16. Bibas M, Trotta MP, Cozzi-Lepri A, Lorenzini P, Pinnetti C, Rizzardini G, et al. Role of serum free light chains in predicting HIV-associated non-Hodgkin lymphoma and Hodgkin’s lymphoma and its correlation with antiretroviral therapy. Am J Hematol 2012;87:749–53.10.1002/ajh.23236Suche in Google Scholar PubMed
17. Landgren O, Goedert JJ, Rabkin CS, Wilson WH, Dunleavy K, Kyle RA, et al. Circulating serum free light chains as predictive markers of AIDS-related lymphoma. J Clin Oncol 2010;28:773–9.10.1200/JCO.2009.25.1322Suche in Google Scholar PubMed PubMed Central
18. Maurer MJ, Cerhan JR, Katzmann JA, Link BK, Allmer C, Zent CS, et al. Monoclonal and polyclonal serum free light chains and clinical outcome in chronic lymphocytic leukemia. Blood 2011;118:2821–6.10.1182/blood-2011-04-349134Suche in Google Scholar PubMed PubMed Central
19. Haynes R, Hutchison CA, Emberson J, Dasgupta T, Wheeler DC, Townend JN, et al. Serum free light chains and the risk of ESRD and death in CKD. Clin J Am Soc Nephrol 2011;6:2829–37.10.2215/CJN.03350411Suche in Google Scholar PubMed PubMed Central
20. Katzmann JA, Clark RJ, Abraham RS, Bryant S, Lymp JF, Bradwell AR, et al. Serum reference intervals and diagnostic ranges for free κ and free λ immunoglobulin light chains: relative sensitivity for detection of monoclonal light chains. Clin Chem 2002;48:1437–44.10.1093/clinchem/48.9.1437Suche in Google Scholar
21. Tholen DW, Kroll M, Astles JR, Caffo AL, Happe TM, Krouwer J, et al. Evaluation of the linearity of quantitative measurement procedures; a statistical approach, approved guideline. CLSI document EP6-A. Wayne, PA: Clinical and Laboratory Standards Institute, 2003.Suche in Google Scholar
22. Tholen DA, Kallner A, Kennedy JW, Krouwer JS, Meier K. Evaluation of precision performance of quantitative measurement methods; approved guideline. 2nd ed. CLSI document EP5-A2. Wayne, PA: Clinical and Laboratory Standards Institute, 2004.Suche in Google Scholar
23. Taylor J. An introduction to error analysis. Oxford: Oxford University Press, 1982.Suche in Google Scholar
24. McEnroe R, Burrit M, Powers D, Rheinheimer D, Wallace B. Interference testing in clinical chemistry; approved guideline. 2nd ed. CLSI document EP7-A2. Wayne, PA: Clinical and Laboratory Standards Institute, 2013.Suche in Google Scholar
25. Hoefield RA, Kalra PA, Baker P, Lane B, New JP, O’Donoghue DJ, et al. Factors associated with kidney disease progression and mortality in a referred CKD population. Am J Kidney Dis 2010;56:1072–81.10.1053/j.ajkd.2010.06.010Suche in Google Scholar PubMed
26. Jackson CE, Myles RC, Tsorlalis IK, Dalzell JR, Rocchiccioli JP, Rodgers JR, et al. Spectral microvolt T-wave alternans testing has no prognostic value in patients recently hospitalized with decompensated heart failure. Eur J Heart Fail 2013;15:1253–61.10.1093/eurjhf/hft085Suche in Google Scholar PubMed
27. Bibas M. Polyclonal serum free light chains: a biomarker of disease prognosis or of immune senescence? J Clin Oncol 2012;30:3033–4.10.1200/JCO.2012.43.4043Suche in Google Scholar
28. Hutchison CA, Landgren O. Polyclonal immunoglobulin free light chains as a potential biomarker of immune stimulation and inflammation. Clin Chem 2011;57:1387–9.10.1373/clinchem.2011.169433Suche in Google Scholar
29. Assi LK, Hughes RG, Gunson B, Webb G, Adams DH. Polyclonally raised serum free light chains in liver disease compared to markers of liver damage. Hematology Reports 2010;2:47.Suche in Google Scholar
30. Kurt IH, Yavuzer K, Batur MK. Short-term effect of levosimendan on free light chain kappa and lambda levels in patients with decompensated chronic heart failure. Heart Vessels 2010;25:392–9.10.1007/s00380-009-1216-4Suche in Google Scholar
31. Assi L, McIntyre NJ, Hutchison CA, Hughes R, Stringer S, Fluck RJ, et al. Total serum free light chains independently predict survival in patients with stage 3 chronic kidney disease. J Am Soc Nephrol 2011;22:449a.Suche in Google Scholar
32. Miettinen TA, Kekki M. Effect of impaired hepatic and renal function on [131I] Bence Jones protein catabolism in human subjects. Clin Chim Acta 1967;18:395–407.10.1016/0009-8981(67)90036-8Suche in Google Scholar
33. Waldmann TA, Strober W, Mogielnicki RP. The renal handling of low molecular weight proteins. II. Disorders of serum protein catabolism in patients with tubular proteinuria, the nephrotic syndrome, or uremia. J Clin Invest 1972;51:2162–74.10.1172/JCI107023Suche in Google Scholar PubMed PubMed Central
34. Hutchison CA, Harding S, Bevins A, Stringer SJ, Cockwell P. Polyclonal immunoglobulin free light chain concentrations and survival in patients with renal disease. J Am Soc Nephrol 2011;22:156a.Suche in Google Scholar
35. Assi L, Young P, Hughes R, Hutchison CA. Total serum free light chains are an important prognostic marker in patients with type II diabetes. J Am Soc Nephrol 2011;22:229a.Suche in Google Scholar
36. Hutchinson AT, Jones DR, Raison RL. The ability to interact with cell membranes suggests possible biological roles for free light chain. Immunol Lett 2012;142:75–7.10.1016/j.imlet.2011.10.013Suche in Google Scholar PubMed
37. Thio M, Blokhuis BR, Nijkamp FP, Redegeld FA. Free immunoglobulin light chains: a novel target in the therapy of inflammatory diseases. Trends Pharmacol Sci 2008;29:170–4.10.1016/j.tips.2008.01.004Suche in Google Scholar PubMed
38. Braber S, Thio M, Blokhuis BR, Henricks PA, Koelink PJ, Groot KT, et al. An association between neutrophils and immunoglobulin free light chains in the pathogenesis of chronic obstructive pulmonary disease. Am J Respir Crit Care Med 2012;185:817–24.10.1164/rccm.201104-0761OCSuche in Google Scholar PubMed
39. Cohen G, HÖrl WH. Free immunoglobulin light chains as a risk factor in renal and extrarenal complications. Semin Dial 2009;22:369–72.10.1111/j.1525-139X.2009.00582.xSuche in Google Scholar PubMed
40. Brebner JA, Stockley RA. Polyclonal free light chains: a biomarker of inflammatory disease or treatment target? F1000 Med Rep 2013;5:4.10.3410/M5-4Suche in Google Scholar PubMed PubMed Central
©2014 by De Gruyter
Artikel in diesem Heft
- Frontmatter
- Editorial
- From symptoms to biomarkers: a change of paradigm
- Reviews
- Biochemical markers in early diagnosis and management of systemic amyloidoses
- The accuracy of the anti-mitochondrial antibody and the M2 subtype test for diagnosis of primary biliary cirrhosis: a meta-analysis
- Genetics and Molecular Diagnostics
- Selection of an optimal method for co-isolation of circulating DNA and miRNA from the plasma of pregnant women
- General Clinical Chemistry and Laboratory Medicine
- Harmonisation of seven common enzyme results through EQA
- Harmonization in hemolysis detection and prevention. A working group of the Catalonian Health Institute (ICS) experience
- Adopting European Network for Health Technology Assessments (EunetHTA) core model for diagnostic technologies for improving the accuracy and appropriateness of blood gas analyzers’ assessment
- Impact of assay design on test performance: lessons learned from 25-hydroxyvitamin D
- Anti-ruthenium antibodies mimic macro-TSH in electrochemiluminescent immunoassay
- The relationship between the Spine Deformity Index, biochemical parameters of bone metabolism and vascular calcifications: results from the Epidemiological VERtebral FRACtures iTalian Study (EVERFRACT) in dialysis patients
- Quantification of polyclonal free light chains in clinical samples using a single turbidimetric immunoassay
- Global coagulation tests: their applicability for measuring direct factor Xa- and thrombin inhibition and reversal of anticoagulation by prothrombin complex concentrate
- Reference Values and Biological Variations
- The importance of individual biology in the clinical use of serum biomarkers for ovarian cancer
- The quality of diagnostic testing may be impaired during shipment of lithium-heparin gel tubes
- Cancer Diagnostics
- Serum human epididymis protein 4 (HE4) as a tumor marker in men with lung cancer
- Intestinal permeability in patients with metastatic colon cancer treated with patupilone
- Cardiovascular Diseases
- European multicenter analytical evaluation of the Abbott ARCHITECT STAT high sensitive troponin I immunoassay
- Infectious Diseases
- Early indicators of severity and construction of a risk model for prognosis based upon laboratory parameters in patients with hemorrhagic fever with renal syndrome
- Corrigendum
- Automated indirect immunofluorescence antinuclear antibody analysis is a standardized alternative for visual microscope interpretation
- Letters to the Editor
- The future of the laboratory information system – what are the requirements for a powerful system for a laboratory data management?
- Hypernatraemia in disguise
- Interference of therapeutic monoclonal immunoglobulins in the investigation of M-proteins
- The impact of exercise on the variation of serum free light chains
- Response to Jacobs: N Latex FLC serum free light-chain assays in patients with renal impairment
- Reply to Berlanga et al. (DOI 10.1515/cclm-2014-0420)
- Mean platelet volume in patients with non-alcoholic fatty liver disease: is mean platelet volume ready as a surrogate marker?
- False-positive D-dimer result in a latex-enhanced immunoassay caused by interfering human anti-mouse antibodies
- D-Dimer in normal pregnancy: determination of reference values for three commercially available assays
- Paraoxonase 2 sequence variation (c.311 C>G) is associated with a modest decrease in circulating LDL size in children and adolescents
- The effect of volume of hydrochloric acid preservation on 24-hour urinary catecholamine test
- Congress Abstracts
- Congress of Clinical Chemistry and Laboratory Medicine
- The 3rd EFLM-UEMS Congress
- 46th National Congress of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC – Laboratory Medicine)
Artikel in diesem Heft
- Frontmatter
- Editorial
- From symptoms to biomarkers: a change of paradigm
- Reviews
- Biochemical markers in early diagnosis and management of systemic amyloidoses
- The accuracy of the anti-mitochondrial antibody and the M2 subtype test for diagnosis of primary biliary cirrhosis: a meta-analysis
- Genetics and Molecular Diagnostics
- Selection of an optimal method for co-isolation of circulating DNA and miRNA from the plasma of pregnant women
- General Clinical Chemistry and Laboratory Medicine
- Harmonisation of seven common enzyme results through EQA
- Harmonization in hemolysis detection and prevention. A working group of the Catalonian Health Institute (ICS) experience
- Adopting European Network for Health Technology Assessments (EunetHTA) core model for diagnostic technologies for improving the accuracy and appropriateness of blood gas analyzers’ assessment
- Impact of assay design on test performance: lessons learned from 25-hydroxyvitamin D
- Anti-ruthenium antibodies mimic macro-TSH in electrochemiluminescent immunoassay
- The relationship between the Spine Deformity Index, biochemical parameters of bone metabolism and vascular calcifications: results from the Epidemiological VERtebral FRACtures iTalian Study (EVERFRACT) in dialysis patients
- Quantification of polyclonal free light chains in clinical samples using a single turbidimetric immunoassay
- Global coagulation tests: their applicability for measuring direct factor Xa- and thrombin inhibition and reversal of anticoagulation by prothrombin complex concentrate
- Reference Values and Biological Variations
- The importance of individual biology in the clinical use of serum biomarkers for ovarian cancer
- The quality of diagnostic testing may be impaired during shipment of lithium-heparin gel tubes
- Cancer Diagnostics
- Serum human epididymis protein 4 (HE4) as a tumor marker in men with lung cancer
- Intestinal permeability in patients with metastatic colon cancer treated with patupilone
- Cardiovascular Diseases
- European multicenter analytical evaluation of the Abbott ARCHITECT STAT high sensitive troponin I immunoassay
- Infectious Diseases
- Early indicators of severity and construction of a risk model for prognosis based upon laboratory parameters in patients with hemorrhagic fever with renal syndrome
- Corrigendum
- Automated indirect immunofluorescence antinuclear antibody analysis is a standardized alternative for visual microscope interpretation
- Letters to the Editor
- The future of the laboratory information system – what are the requirements for a powerful system for a laboratory data management?
- Hypernatraemia in disguise
- Interference of therapeutic monoclonal immunoglobulins in the investigation of M-proteins
- The impact of exercise on the variation of serum free light chains
- Response to Jacobs: N Latex FLC serum free light-chain assays in patients with renal impairment
- Reply to Berlanga et al. (DOI 10.1515/cclm-2014-0420)
- Mean platelet volume in patients with non-alcoholic fatty liver disease: is mean platelet volume ready as a surrogate marker?
- False-positive D-dimer result in a latex-enhanced immunoassay caused by interfering human anti-mouse antibodies
- D-Dimer in normal pregnancy: determination of reference values for three commercially available assays
- Paraoxonase 2 sequence variation (c.311 C>G) is associated with a modest decrease in circulating LDL size in children and adolescents
- The effect of volume of hydrochloric acid preservation on 24-hour urinary catecholamine test
- Congress Abstracts
- Congress of Clinical Chemistry and Laboratory Medicine
- The 3rd EFLM-UEMS Congress
- 46th National Congress of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC – Laboratory Medicine)