Abstract
Background: Current automated immunoassays vary significantly in many aspects of their design. This study sought to establish if the theoretical advantages and disadvantages associated with different design formats of automated 25-hydroxyvitamin D (25-OHD) assays are translated into variations in assay performance in practice.
Methods: 25-OHD was measured in 1236 samples using automated assays from Abbott, DiaSorin, Roche and Siemens. A subset of 362 samples had up to three liquid chromatography-tandem mass spectrometry 25-OHD analyses performed. 25-OHD2 recovery, dilution recovery, human anti-animal antibody (HAAA) interference, 3-epi-25-OHD3 cross-reactivity and precision of the automated assays were evaluated.
Results: The assay that combined release of 25-OHD with analyte capture in a single step showed the most accurate 25-OHD2 recovery and the best dilution recovery. The use of vitamin D binding protein (DBP) as the capture moiety was associated with 25-OHD2 under-recovery, a trend consistent with 3-epi-25-OHD3 cross-reactivity and immunity to HAAA interference. Assays using animal-derived antibodies did not show 3-epi-25-OHD3 cross-reactivity but were variably susceptible to HAAA interference. Not combining 25-OHD release and capture in one step and use of biotin-streptavidin interaction for solid phase separation were features of the assays with inferior accuracy for diluted samples. The assays that used a backfill assay format showed the best precision at high concentrations but this design did not guarantee precision at low 25-OHD concentrations.
Conclusions: Variations in design among automated 25-OHD assays influence their performance characteristics. Consideration of the details of assay design is therefore important when selecting and validating new assays.
References
1. Carter GD, Carter R, Jones J, Berry J. How accurate are assays for 25-hydroxyvitamin D? Data from the international vitamin D external quality assessment scheme. Clin Chem 2004;50:2195–7.10.1373/clinchem.2004.040683Search in Google Scholar
2. Farrell CJ, Martin S, McWhinney B, Straub I, Williams P, Herrmann M. State-of-the-art vitamin D assays: comparison of automated immunoassay with liquid chromatography-tandem mass spectrometry methods. Clin Chem 2012;58:531–42.10.1373/clinchem.2011.172155Search in Google Scholar
3. Roth HJ, Schimidt-Gayk H, Weber H, Niederau C. Accuracy and clinical implications of seven 25-hydroxyvitamin D methods compared with liquid chromatography-tandem mass spectrometry as a reference. Ann Clin Biochem 2008;45:153–9.10.1258/acb.2007.007091Search in Google Scholar
4. Farrell C, Soldo J, Williams P, Herrmann M. 25-Hydroxyvitmain D testing: challenging the performance of current automated immunoassays. Clin Chem Lab Med 2012;50:1953–63.10.1515/cclm-2012-0522Search in Google Scholar
5. Cavalier E, Carlisi A, Beckaert AC, Rousselle O, Chapelle JP. Human anti-animal interference in DiaSorin Liaison total 25(OH)-vitamin D assay: towards the end of a strange story? Clin Chim Acta 2012;413:527–8.10.1016/j.cca.2011.11.014Search in Google Scholar
6. Hollis BW. Comparison and equilibrium and disequilibrium assay conditions for ergocalciferol, cholecalciferol and their major metabolites. J Steroid Biochem 1984;21:81–6.10.1016/0022-4731(84)90063-3Search in Google Scholar
7. Kricka LJ. Selected strategies for improving sensitivity and reliability of assays. Clin Chem 1994;40:347–57.10.1093/clinchem/40.3.347Search in Google Scholar
8. Holick MF, Binkley NC, Bischoff-Ferrari HA, Gordon CM, Hanley DA, Heaney RP, et al. Evaluation, treatment, and prevention of vitamin D deficiency: an Endocrine Society clinical practice guideline. J Clin Endocrinol Metab 2011;96:1911–30.10.1210/jc.2011-0385Search in Google Scholar PubMed
9. Tai SS, Bedner M, Phinney KW. Development of a candidate reference measurement procedure for the determination of 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 in human serum using isotope-dilution liquid chromatography-tandem mass spectrometry. Anal Chem 2010;82:1942–8.10.1021/ac9026862Search in Google Scholar PubMed PubMed Central
10. Stepman HC, Vanderroost A, Van Uytfanghe K, Thienpont LM. Candidate reference measurement procedures for serum 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 by using isotope-dilution liquid chromatography-tandem mass spectrometry. Clin Chem 2011;57:441–8.10.1373/clinchem.2010.152553Search in Google Scholar PubMed
11. Carter GD, Jones JC. Use of a common standard improves the performance of liquid chromatography-tandem mass spectrometry methods for serum 25-hydroxyvitmain-D. Ann Clin Biochem 2009;46:79–81.10.1258/acb.2008.008135Search in Google Scholar PubMed
12. Heijboer AC, Blankenstein MA, Kema IP, Buijs MM. Accuracy of 6 routine 25-hydroxyvitamin D assays: influence of vitamin D binding protein concentration. Clin Chem 2012;58:543–8.10.1373/clinchem.2011.176545Search in Google Scholar PubMed
13. Kamao M, Tatematsu S, Hatakeyama S, Sakaki T, Sawada N, Inouye K, et al. C-3 epimerization of vitamin D3 metabolites and further metabolism of C-3 epimers: 25-hydroxyvitamin D3 is metabolized to 3-epi-25-hydroxyvitamin D3 and subsequently metabolized though C-1alpha or C-24 hydroxylation. J Biol Chem 2004;279:15897–907.10.1074/jbc.M311473200Search in Google Scholar PubMed
14. Lensmeyer G, Poquette M, Wiebe D, Binkley N. The C-3 epimer of 25-hydroxyvitamin D(3) is present in adult serum. J Clin Endocrinol Metab 2012;97:163–8.10.1210/jc.2011-0584Search in Google Scholar PubMed
Supplemental Material
The online version of this article (DOI: 10.1515/cclm-2014-0111) offers supplementary material, available to authorized users.
©2014 by De Gruyter
Articles in the same Issue
- Frontmatter
- Editorial
- From symptoms to biomarkers: a change of paradigm
- Reviews
- Biochemical markers in early diagnosis and management of systemic amyloidoses
- The accuracy of the anti-mitochondrial antibody and the M2 subtype test for diagnosis of primary biliary cirrhosis: a meta-analysis
- Genetics and Molecular Diagnostics
- Selection of an optimal method for co-isolation of circulating DNA and miRNA from the plasma of pregnant women
- General Clinical Chemistry and Laboratory Medicine
- Harmonisation of seven common enzyme results through EQA
- Harmonization in hemolysis detection and prevention. A working group of the Catalonian Health Institute (ICS) experience
- Adopting European Network for Health Technology Assessments (EunetHTA) core model for diagnostic technologies for improving the accuracy and appropriateness of blood gas analyzers’ assessment
- Impact of assay design on test performance: lessons learned from 25-hydroxyvitamin D
- Anti-ruthenium antibodies mimic macro-TSH in electrochemiluminescent immunoassay
- The relationship between the Spine Deformity Index, biochemical parameters of bone metabolism and vascular calcifications: results from the Epidemiological VERtebral FRACtures iTalian Study (EVERFRACT) in dialysis patients
- Quantification of polyclonal free light chains in clinical samples using a single turbidimetric immunoassay
- Global coagulation tests: their applicability for measuring direct factor Xa- and thrombin inhibition and reversal of anticoagulation by prothrombin complex concentrate
- Reference Values and Biological Variations
- The importance of individual biology in the clinical use of serum biomarkers for ovarian cancer
- The quality of diagnostic testing may be impaired during shipment of lithium-heparin gel tubes
- Cancer Diagnostics
- Serum human epididymis protein 4 (HE4) as a tumor marker in men with lung cancer
- Intestinal permeability in patients with metastatic colon cancer treated with patupilone
- Cardiovascular Diseases
- European multicenter analytical evaluation of the Abbott ARCHITECT STAT high sensitive troponin I immunoassay
- Infectious Diseases
- Early indicators of severity and construction of a risk model for prognosis based upon laboratory parameters in patients with hemorrhagic fever with renal syndrome
- Corrigendum
- Automated indirect immunofluorescence antinuclear antibody analysis is a standardized alternative for visual microscope interpretation
- Letters to the Editor
- The future of the laboratory information system – what are the requirements for a powerful system for a laboratory data management?
- Hypernatraemia in disguise
- Interference of therapeutic monoclonal immunoglobulins in the investigation of M-proteins
- The impact of exercise on the variation of serum free light chains
- Response to Jacobs: N Latex FLC serum free light-chain assays in patients with renal impairment
- Reply to Berlanga et al. (DOI 10.1515/cclm-2014-0420)
- Mean platelet volume in patients with non-alcoholic fatty liver disease: is mean platelet volume ready as a surrogate marker?
- False-positive D-dimer result in a latex-enhanced immunoassay caused by interfering human anti-mouse antibodies
- D-Dimer in normal pregnancy: determination of reference values for three commercially available assays
- Paraoxonase 2 sequence variation (c.311 C>G) is associated with a modest decrease in circulating LDL size in children and adolescents
- The effect of volume of hydrochloric acid preservation on 24-hour urinary catecholamine test
- Congress Abstracts
- Congress of Clinical Chemistry and Laboratory Medicine
- The 3rd EFLM-UEMS Congress
- 46th National Congress of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC – Laboratory Medicine)
Articles in the same Issue
- Frontmatter
- Editorial
- From symptoms to biomarkers: a change of paradigm
- Reviews
- Biochemical markers in early diagnosis and management of systemic amyloidoses
- The accuracy of the anti-mitochondrial antibody and the M2 subtype test for diagnosis of primary biliary cirrhosis: a meta-analysis
- Genetics and Molecular Diagnostics
- Selection of an optimal method for co-isolation of circulating DNA and miRNA from the plasma of pregnant women
- General Clinical Chemistry and Laboratory Medicine
- Harmonisation of seven common enzyme results through EQA
- Harmonization in hemolysis detection and prevention. A working group of the Catalonian Health Institute (ICS) experience
- Adopting European Network for Health Technology Assessments (EunetHTA) core model for diagnostic technologies for improving the accuracy and appropriateness of blood gas analyzers’ assessment
- Impact of assay design on test performance: lessons learned from 25-hydroxyvitamin D
- Anti-ruthenium antibodies mimic macro-TSH in electrochemiluminescent immunoassay
- The relationship between the Spine Deformity Index, biochemical parameters of bone metabolism and vascular calcifications: results from the Epidemiological VERtebral FRACtures iTalian Study (EVERFRACT) in dialysis patients
- Quantification of polyclonal free light chains in clinical samples using a single turbidimetric immunoassay
- Global coagulation tests: their applicability for measuring direct factor Xa- and thrombin inhibition and reversal of anticoagulation by prothrombin complex concentrate
- Reference Values and Biological Variations
- The importance of individual biology in the clinical use of serum biomarkers for ovarian cancer
- The quality of diagnostic testing may be impaired during shipment of lithium-heparin gel tubes
- Cancer Diagnostics
- Serum human epididymis protein 4 (HE4) as a tumor marker in men with lung cancer
- Intestinal permeability in patients with metastatic colon cancer treated with patupilone
- Cardiovascular Diseases
- European multicenter analytical evaluation of the Abbott ARCHITECT STAT high sensitive troponin I immunoassay
- Infectious Diseases
- Early indicators of severity and construction of a risk model for prognosis based upon laboratory parameters in patients with hemorrhagic fever with renal syndrome
- Corrigendum
- Automated indirect immunofluorescence antinuclear antibody analysis is a standardized alternative for visual microscope interpretation
- Letters to the Editor
- The future of the laboratory information system – what are the requirements for a powerful system for a laboratory data management?
- Hypernatraemia in disguise
- Interference of therapeutic monoclonal immunoglobulins in the investigation of M-proteins
- The impact of exercise on the variation of serum free light chains
- Response to Jacobs: N Latex FLC serum free light-chain assays in patients with renal impairment
- Reply to Berlanga et al. (DOI 10.1515/cclm-2014-0420)
- Mean platelet volume in patients with non-alcoholic fatty liver disease: is mean platelet volume ready as a surrogate marker?
- False-positive D-dimer result in a latex-enhanced immunoassay caused by interfering human anti-mouse antibodies
- D-Dimer in normal pregnancy: determination of reference values for three commercially available assays
- Paraoxonase 2 sequence variation (c.311 C>G) is associated with a modest decrease in circulating LDL size in children and adolescents
- The effect of volume of hydrochloric acid preservation on 24-hour urinary catecholamine test
- Congress Abstracts
- Congress of Clinical Chemistry and Laboratory Medicine
- The 3rd EFLM-UEMS Congress
- 46th National Congress of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC – Laboratory Medicine)