It is now universally acknowledged that preanalytical variability represents the leading source of problems throughout the total testing process, wherein inappropriate procedures for collection, handing and storage of blood samples jeopardize the quality of specimens and the medical decisions that may be taken when “biased” test results are obtained on otherwise unsuitable specimens [1]. It is also clear to most that spurious hemolysis, i.e., injury and breakdown of erythrocytes and other blood cells that most frequently occur during venipuncture, is by far the leading cause of unsuitable specimens in clinical laboratories worldwide [2]. The potential consequences on laboratory diagnostics of undue presence of cell-free hemoglobin in blood specimens include biological, chemical and photometrical interference, which may virtually affect all areas of testing, thus including clinical chemistry [3], immunochemistry [4], hematological [5], blood gas [6], and coagulation tests [7]. Although the biological bias is roughly unavoidable and virtually insurmountable, the chemical and photometrical interference is instead technically preventable by adjustment of sample blank, use of different wavelengths or measurement times [8], as well as by reformulation of reagents with additives so that the interference of cell-free hemoglobin and other intracellular components can be largely or completely abolished.
Several lines of evidence now attest that even major refinements of current test protocols are not effective to completely abolish the bias, so that the development of hemolysis-resistant reagent(s) represent an appealing perspective, especially for those analytes whose concentration is only influenced by chemical or photometrical (rather than biological) interference. One first example of such approach is published in this issue of Clinical Chemistry and Laboratory Medicine. Specifically, Ronda et al. developed an experimental reagent for assessment of α-amylase in hemolyzed specimens, showing that all test results could be maintained within the total allowable error for this parameter by using the experimental assay, whereas this critical threshold was exceed in the vast majority of samples (up to 85%) using a similar commercial method [9].
Besides α-amylase, other potential applications of hemolysis-resistant reagents, expectedly combined with inhibitors of leukocyte proteases (e.g., ethylenediaminetetraacetic acid, aprotinin, [4-(2-Aminoethyl) benzenesulfonyl, bestatin and leupeptin, among others] [10] that may be released after white blood cell injury, include the measurement of all those molecules that are not significantly contained in erythrocytes, white blood cells, platelets or endothelial cells, and thus would not be influenced by injury or breakdown of these cells. At the top of a virtually infinite list we obviously place stat tests, being characterized by a short turnaround time, and for which the practice of sample recollection would raise not only economic issues (e.g., raw cost of phlebotomy set, blood tubes and nurse’s occupation) and organizational concerns (e.g., disputes between emergency physicians and laboratory staff), but also clinical problems including a delay in triage, diagnosis and treatment of patients, with unavoidable overcrowding of emergency departments (EDs) [11]. One paradigmatic example is the assessment of cardiospecific troponin(s) in the emergency room, especially when using high-sensitive immunoassays [12]. Since results of most troponin immunoassays are dramatically biased by the presence of spurious hemolysis in the sample [4] and considering that the rate of hemolyzed specimens can be as high as 8% from the EDs [13, 14], it is rather obvious that the implementation of hemolysis-resistant reagents coupled with inhibitors of leukocyte proteases may be a reasonable approach for assessing an appropriate panel of stat tests (i.e., without measurement of potassium, lactate dehydrogenase and aspartate aminotransferase), thus generating remarkable advantages for the healthcare systems and contributing to prevent diagnostic delays and unnecessary expenditures. The biochemical and immunochemical analysis of biological fluids other than blood, serum or plasma is another interesting perspective of hemolysis-resistant reagents, especially for urines, peritoneal or pleural fluids, which can be frequently contaminated by the presence of injured erythrocytes.
It is rather clear that hemolysis-resistant reagents are not intended to be the panacea, but they may represent indeed another part of the puzzle for preventing interference in laboratory diagnostics along with several other aspects that are synthesized in Figure 1. First and foremost, however, further efforts should be placed to improve blood collection procedures and for enhancing compliance with existing guidelines, inasmuch as these practices are still considered a mainstay for improving healthcare operators’ education and training, and for assuring patient safety in laboratory testing [15].

Methodological and technical issues for preventing interference from spurious hemolysis in laboratory testing.
Conflict of interest statement
Authors’ conflict of interest disclosure: The authors stated that there are no conflicts of interest regarding the publication of this article.
Research funding: None declared.
Employment or leadership: None declared.
Honorarium: None declared.
References
1. Lippi G, Becan-McBride K, Behúlová D, Bowen RA, Church S, Delanghe J, et al. Preanalytical quality improvement: in quality we trust. Clin Chem Lab Med 2013;51:229–41.10.1515/cclm-2012-0597Suche in Google Scholar PubMed
2. Simundic AM, Topic E, Nikolac N, Lippi G. Hemolysis detection and management of hemolyzed specimens. Biochem Med (Zagreb) 2010;20:154–9.10.11613/BM.2010.018Suche in Google Scholar
3. Lippi G, Salvagno GL, Montagnana M, Brocco G, Guidi GC. Influence of hemolysis on routine clinical chemistry testing. Clin Chem Lab Med 2006;44:311–6.10.1515/CCLM.2006.054Suche in Google Scholar PubMed
4. Lippi G, Avanzini P, Dipalo M, Aloe R, Cervellin G. Influence of hemolysis on troponin testing: studies on Beckman Coulter UniCel Dxl 800 Accu-TnI and overview of the literature. Clin Chem Lab Med 2011;49:2097–100.10.1515/CCLM.2011.703Suche in Google Scholar PubMed
5. Lippi G, Musa R, Avanzini P, Aloe R, Pipitone S, Sandei F. Influence of in vitro hemolysis on hematological testing on Advia 2120. Int J Lab Hematol 2012;34:179–84.10.1111/j.1751-553X.2011.01378.xSuche in Google Scholar PubMed
6. Lippi G, Ippolito L, Fontana R. Prevalence of hemolytic specimens referred for arterial blood gas analysis. Clin Chem Lab Med 2011;49:931–2.10.1515/CCLM.2011.136Suche in Google Scholar PubMed
7. Lippi G, Plebani M, Favaloro EJ. Interference in coagulation testing: focus on spurious hemolysis, icterus, and lipemia. Semin Thromb Hemost 2013;39:258–66.Suche in Google Scholar
8. Frémont S, Combe AM, Mecrin M, Tronel H, Galland AV, Scholl C, et al. Influence of visible interferences on biochemical assays realised on Au 5231, AU 5223 (Olympus) and CL 7200 (Shimadzu). Ann Biol Clin (Paris) 1996;54:309–20.Suche in Google Scholar
9. Ronda L, Mozzarelli A, Aloe R, Lippi G. Development of a novel, hemolysis-resistant reagent for assessment of α-amylase in biological fluids. Clin Chem Lab Med 2013;51:1409–15.10.1515/cclm-2012-0376Suche in Google Scholar PubMed
10. Narayanan S. The preanalytic phase. An important component of laboratory medicine. Am J Clin Pathol 2000; 113:429–52.Suche in Google Scholar
11. Lippi G, Plebani M, Di Somma S, Cervellin G. Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories. Crit Rev Clin Lab Sci 2011;48:143–53.10.3109/10408363.2011.600228Suche in Google Scholar PubMed
12. Thygesen K, Mair J, Giannitsis E, Mueller C, Lindahl B, Blankenberg S, et al; Study Group on Biomarkers in Cardiology of ESC Working Group on Acute Cardiac Care. How to use high-sensitivity cardiac troponins in acute cardiac care. Eur Heart J 2012;33:2252–7.10.1093/eurheartj/ehs154Suche in Google Scholar PubMed
13. Carraro P, Servidio G, Plebani M. Hemolyzed specimens: a reason for rejection or a clinical challenge? Clin Chem 2000;46:306–7.10.1093/clinchem/46.2.306Suche in Google Scholar
14. Lippi G, Salvagno GL, Favaloro EJ, Guidi GC. Survey on the prevalence of hemolytic specimens in an academic hospital according to collection facility: opportunities for quality improvement. Clin Chem Lab Med 2009;47:616–8.10.1515/CCLM.2009.132Suche in Google Scholar PubMed
15. Plebani M. The CCLM contribution to improvements in quality and patient safety. Clin Chem Lab Med 2013;51:39–46.10.1515/cclm-2012-0094Suche in Google Scholar PubMed
©2013 by Walter de Gruyter Berlin Boston
Artikel in diesem Heft
- Masthead
- Masthead
- Editorials
- Biomarkers in the treatment of cancer: opportunities and pitfalls
- Multicenter reference intervals studies: a promising perspective for the future?
- Hemolysis-resistant reagent: another part of the puzzle for preventing errors in laboratory testing
- Review
- Exhaled breath condensate: a comprehensive update
- Mini Review
- Psoriasis, a new challenge for laboratory medicine
- Opinion Paper
- CA 19-9: handle with care
- General Clinical Chemistry and Laboratory Medicine
- Performance evaluation of human cytokines profiles obtained by various multiplexed-based technologies underlines a need for standardization
- The iron regulatory hormone hepcidin is decreased in pregnancy: a prospective longitudinal study
- Red blood cell distribution width is a potential prognostic index for liver disease
- Development of a novel, hemolysis-resistant reagent for assessment of α-amylase in biological fluids
- Performance of cassette-based blood gas analyzers to monitor blood glucose and lactate levels in a surgical intensive care setting
- Reference Values and Biological Variations
- The Asian project for collaborative derivation of reference intervals: (1) strategy and major results of standardized analytes
- The Asian project for collaborative derivation of reference intervals: (2) results of non-standardized analytes and transference of reference intervals to the participating laboratories on the basis of cross-comparison of test results
- Cancer Diagnostics
- Study of an elevated carbohydrate antigen 19-9 concentration in a large health check-up cohort in China
- Glycoproteomic identification of potential glycoprotein biomarkers in ovarian cancer proximal fluids
- Comparison of serum calcitonin and procalcitonin in detecting medullary thyroid carcinoma among patients with thyroid nodules
- Serum HER-2 predicts response and resistance to trastuzumab treatment in breast cancer
- Elevated tumor markers in patients with pulmonary alveolar proteinosis
- Serum carcinoembryonic antigen is associated with non-alcoholic fatty liver disease in healthy Korean non-smokers
- Assessment of SOX17 DNA methylation in cell free DNA from patients with operable gastric cancer. Association with prognostic variables and survival
- Sensitivity of CA 15-3, CEA and serum HER2 in the early detection of recurrence of breast cancer
- Cardiovascular Diseases
- Association of nucleotide-binding oligomerization domain-like receptor 3 inflammasome and adverse clinical outcomes in patients with idiopathic dilated cardiomyopathy
- Diabetes
- α1-Microglobulin/albumin ratio may improve interpretation of albuminuria in statin-treated patients
- Infectious Diseases
- The role of procalcitonin and IL-6 in discriminating between septic and non-septic causes of ALI/ARDS: a prospective observational study
- Letters to the Editor
- Carryover does not affect results of Beckman Coulter highly-sensitive-AccuTnI assay on Access 2
- Adding calcium to EDTA plasma samples prior to analysis could solve the compatibility issue in commercially available ELISAs that are standardized for serum
- Interference of luteinizing hormone in the urinary point-of-care human chorionic gonadotropin test
- Standardize the serum albumin assay now: calibrate it to 60% of the serum total protein assay
- Interference of M-paraprotein in automated urea assays
- A case of mistaken diagnosis of heparin-induced thrombocytopenia due to pseudothrombocytopenia
- Delta check: a must in the management of hyponatremia
- A novel cost effective algorithm for antinuclear antibody (ANA) testing in an outpatient setting
Artikel in diesem Heft
- Masthead
- Masthead
- Editorials
- Biomarkers in the treatment of cancer: opportunities and pitfalls
- Multicenter reference intervals studies: a promising perspective for the future?
- Hemolysis-resistant reagent: another part of the puzzle for preventing errors in laboratory testing
- Review
- Exhaled breath condensate: a comprehensive update
- Mini Review
- Psoriasis, a new challenge for laboratory medicine
- Opinion Paper
- CA 19-9: handle with care
- General Clinical Chemistry and Laboratory Medicine
- Performance evaluation of human cytokines profiles obtained by various multiplexed-based technologies underlines a need for standardization
- The iron regulatory hormone hepcidin is decreased in pregnancy: a prospective longitudinal study
- Red blood cell distribution width is a potential prognostic index for liver disease
- Development of a novel, hemolysis-resistant reagent for assessment of α-amylase in biological fluids
- Performance of cassette-based blood gas analyzers to monitor blood glucose and lactate levels in a surgical intensive care setting
- Reference Values and Biological Variations
- The Asian project for collaborative derivation of reference intervals: (1) strategy and major results of standardized analytes
- The Asian project for collaborative derivation of reference intervals: (2) results of non-standardized analytes and transference of reference intervals to the participating laboratories on the basis of cross-comparison of test results
- Cancer Diagnostics
- Study of an elevated carbohydrate antigen 19-9 concentration in a large health check-up cohort in China
- Glycoproteomic identification of potential glycoprotein biomarkers in ovarian cancer proximal fluids
- Comparison of serum calcitonin and procalcitonin in detecting medullary thyroid carcinoma among patients with thyroid nodules
- Serum HER-2 predicts response and resistance to trastuzumab treatment in breast cancer
- Elevated tumor markers in patients with pulmonary alveolar proteinosis
- Serum carcinoembryonic antigen is associated with non-alcoholic fatty liver disease in healthy Korean non-smokers
- Assessment of SOX17 DNA methylation in cell free DNA from patients with operable gastric cancer. Association with prognostic variables and survival
- Sensitivity of CA 15-3, CEA and serum HER2 in the early detection of recurrence of breast cancer
- Cardiovascular Diseases
- Association of nucleotide-binding oligomerization domain-like receptor 3 inflammasome and adverse clinical outcomes in patients with idiopathic dilated cardiomyopathy
- Diabetes
- α1-Microglobulin/albumin ratio may improve interpretation of albuminuria in statin-treated patients
- Infectious Diseases
- The role of procalcitonin and IL-6 in discriminating between septic and non-septic causes of ALI/ARDS: a prospective observational study
- Letters to the Editor
- Carryover does not affect results of Beckman Coulter highly-sensitive-AccuTnI assay on Access 2
- Adding calcium to EDTA plasma samples prior to analysis could solve the compatibility issue in commercially available ELISAs that are standardized for serum
- Interference of luteinizing hormone in the urinary point-of-care human chorionic gonadotropin test
- Standardize the serum albumin assay now: calibrate it to 60% of the serum total protein assay
- Interference of M-paraprotein in automated urea assays
- A case of mistaken diagnosis of heparin-induced thrombocytopenia due to pseudothrombocytopenia
- Delta check: a must in the management of hyponatremia
- A novel cost effective algorithm for antinuclear antibody (ANA) testing in an outpatient setting