Hemoglobin A1c (HbA1c) reflects a time averaged blood glucose during the previous 2–3 months, and is used as the gold standard for long-term follow-up of glycemic control. The same HbA1c method was used both in the DCCT study published in 1993 [1] and the UKPDS study published in 1998 [2]. Following the development of a true reference method by IFCC [3], it was stated in an international consensus that this reference method should be used by all manufacturers for the calibration methods used in the clinical laboratory [4]. A second consensus meeting was held in 2009, where the main issue was how the results should be presented to clinicians and patients, and how they should be reported in scientific journals [5]. In short, three conclusions were reached at this meeting:
HbA1c results are to be reported by clinical laboratories worldwide in SI (Sistème Internationale) units (mmol/mol – no decimals) and derived NGSP units (% – 1 decimal), using the IFCC-NGSP master equation (DCCT units).
HbA1c conversion tables including both SI (IFCC) and NGSP units should be easily accessible for the diabetes community.
Journals are strongly recommended to require that submitted manuscripts report HbA1c in both SI (IFCC) and NGSP/DCCT units.
A third consensus meeting was held in Dubai in December 2011. At this meeting, points number 2 and 3 above were reiterated without any disagreement, and many journals are now considering how to implement dual reporting. A web calculator is available to facilitate this and the web address is http://www.hba1c.nu/eng. The participants stated that since 2009, different countries have moved in different directions regarding the double reporting of HbA1c. Laboratories in the US continue to report HbA1c results in % only (NGSP values), and clinicians are mostly unaware of SI (IFCC) units. Japanese manufactured laboratory instruments in use outside of Japan are IFCC calibrated with options to display IFCC or NGSP numbers, but Japan reports both JDS and NGSP numbers within the country. Across Europe most countries have adopted dual reporting, whereas some European countries following different time periods of dual reporting are now reporting SI units (mmol/mol) only. No new consensus was reached, but the participants agreed to the following:
The original consensus where results are reported in both SI units (mmol/mol) and in derived NGSP/DCCT percent units remains the ideal to achieve global standardization. Some countries which reported both units have now moved on to single reporting of SI units. It is particularly important to dual report for an agreed period of time if countries intend to move from percent to SI units. Some countries have decided to continue to report only NGSP/DCCT percent units as this better reflects clinical practice in that country. It is important that the whole county adopts a single approach to the reporting units so as not to cause confusion, and put patients at risk. Journals are recommended to use either HbA1c or HbA1C as abbreviations.
Editors of journals and other printed material are strongly recommended to ensure that submitted manuscripts report HbA1c in both SI (IFCC) and NGSP/DCCT units.
The 2010 statement was simultaneously published in eight journals: Annals of Clinical Biochemistry, Clinical Chemistry, Clinical Chemistry and Laboratory Medicine, Diabetes Care, Diabetes Research and Clinical Practice, Diabetic Medicine, Diabetologia and Pediatric Diabetes. These journals either require papers to be submitted in both units or are discussing how best to implement this. We urge other journals within the diabetes field to adopt the same policy.
The consensus group will meet again in 2013 at the IDF meeting in Melbourne.
Conflict of interest statement
Authors’ conflict of interest disclosure: The authors stated that there are no conflicts of interest regarding the publication of this article.
Research funding: None declared.
Employment or leadership: None declared.
Honorarium: None declared.
Appendix
Consensus Committee
ADA: Sue Kirkman, Matt Petersen; EASD: Viktor Joergens, Andrew Boulton; IDF: Jean-Claude Mbanya, Stephen Colagiuri; IFCC: Garry John, Graham Beastall; ISPAD: Ragnar Hanas, Lynda Fisher; JDS: Izumi Takei; NGSP: David Sacks.
References
1. DCCT. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977–86.10.1056/NEJM199309303291401Suche in Google Scholar
2. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33). UK Prospective Diabetes Study (UKPDS) Group. Lancet 1998;352:837–53.10.1016/S0140-6736(98)07019-6Suche in Google Scholar
3. Hoelzel W, Weykamp C, Jeppsson JO, Miedema K, Barr JR, Goodall I, etal. IFCC reference system for measurement of hemoglobin A1c in human blood and the national standardization schemes in the United States, Japan, and Sweden: a method-comparison study. Clin Chem 2004;50:166–74.10.1373/clinchem.2003.024802Suche in Google Scholar PubMed
4. Consensus statement on the worldwide standardisation of the HbA1c measurement. Diabetologia 2007;50:2042–3.10.1007/s00125-007-0789-7Suche in Google Scholar PubMed
5. Hanas R, John G. 2010 consensus statement on the worldwide standardization of the hemoglobin A1c measurement. Clin Chem Lab Med 2010;48:775–6.10.1515/CCLM.2010.029Suche in Google Scholar PubMed
©2013 by Walter de Gruyter Berlin Boston
Artikel in diesem Heft
- Masthead
- Masthead
- Editorial
- Progress towards standardization: an IFCC Scientific Division Perspective
- Review
- Quantitative Clinical Chemistry Proteomics (qCCP) using mass spectrometry: general characteristics and application
- Mini Reviews
- Analytical goals for the determination of HbA2
- Glucose meters – fit for clinical purpose
- Opinion Papers
- Metrological traceability – a concept for standardization in laboratory medicine
- Recommendations for clinical laboratory science reports regarding properties, units, and symbols: the NPU format1)
- “Good samples make good assays” – the problem of sourcing clinical samples for a standardization project
- Defining acceptable limits for the metrological traceability of specific measurands
- Original Articles
- A reference system for urinary albumin: current status
- Toward standardization of carbohydrate-deficient transferrin (CDT) measurements: III. Performance of native serum and serum spiked with disialotransferrin proves that harmonization of CDT assays is possible
- External Quality Assessment Scheme for reference laboratories – review of 8 years’ experience
- Utility of a panel of sera for the alignment of test results in the worldwide multicenter study on reference values
- Protocol and standard operating procedures for common use in a worldwide multicenter study on reference values
- Guidelines and Recommendations
- 2013 update on the worldwide standardization of the hemoglobin A1c measurement
- General Clinical Chemistry and Laboratory Medicine
- Evacuated blood-collection tubes for haematological tests – a quality evaluation prior to their intended use for specimen collection
- Refrigeration is not necessary for measurement of uric acid in patients treated with rasburicase
- Procalcitonin and mid-regional pro-adrenomedullin test combination in sepsis diagnosis
- Cysteinyl leukotrienes in exhaled breath condensate of smoking asthmatics
- Helicobacter pylori serology in autoimmune diseases – fact or fiction?
- Serum DNase I activity in systemic lupus erythematosus: correlation with immunoserological markers, the disease activity and organ involvement
- Antibodies against Nε-homocysteinylated proteins in patients on different methods of renal replacement therapy
- Reference Values and Biological Variations
- Reference values for urinary neutrophil gelatinase-associated lipocalin (NGAL) in pediatric age measured with a fully automated chemiluminescent platform
- High biological variation of serum hyaluronic acid and Hepascore, a biochemical marker model for the prediction of liver fibrosis
- Cardiovascular Diseases
- Circulating matrix Gla protein: a potential tool to identify minor carotid stenosis with calcification in a risk population
- Midregional pro-atrial natriuretic peptide in the general population/Insights from the Gutenberg Health Study
- Letters to the Editor
- Relationship between Helicobacter pylori infection and autoimmune disorders
- Rapid simultaneous genotyping of polymorphisms in ADH1B and ALDH2 using high resolution melting assay
- Comparison of biological specimens and DNA collection methods for PCR amplification and microarray analysis
- Transferrin/log(ferritin) ratio: a self-fulfilling prophecy when iron deficiency is defined by serum ferritin concentration
- Response to: Transferrin/log(ferritin) ratio: a self-fulfilling prophecy when iron deficiency is defined by serum ferritin concentration
- Determinants of homocysteine concentrations in mother and neonatal girl pairs
- Cysteine analog breaks cryoprecipitate associated with chronic hepatitis C
- Cut-off values of serum growth hormone (GH) in pharmacological stimulation tests (PhT) evaluated in short-statured children using a chemiluminescent immunometric assay (ICMA) calibrated with the International Recombinant Human GH Standard 98/574
- Normalized MEDx chart as a useful tool for evaluation of analytical quality achievements. A picture is worth a thousand words
- Harmonization of immunoassays to the all-procedure trimmed mean – proof of concept by use of data from the insulin standardization project
- Congress Abstracts
- Annual Assembly of the Swiss Society of Clinical Chemistry & International Congress of Porphyrins and Porphyrias & International Meeting of Porphyria Patients: Personalized Medicine and Rare Diseases
Artikel in diesem Heft
- Masthead
- Masthead
- Editorial
- Progress towards standardization: an IFCC Scientific Division Perspective
- Review
- Quantitative Clinical Chemistry Proteomics (qCCP) using mass spectrometry: general characteristics and application
- Mini Reviews
- Analytical goals for the determination of HbA2
- Glucose meters – fit for clinical purpose
- Opinion Papers
- Metrological traceability – a concept for standardization in laboratory medicine
- Recommendations for clinical laboratory science reports regarding properties, units, and symbols: the NPU format1)
- “Good samples make good assays” – the problem of sourcing clinical samples for a standardization project
- Defining acceptable limits for the metrological traceability of specific measurands
- Original Articles
- A reference system for urinary albumin: current status
- Toward standardization of carbohydrate-deficient transferrin (CDT) measurements: III. Performance of native serum and serum spiked with disialotransferrin proves that harmonization of CDT assays is possible
- External Quality Assessment Scheme for reference laboratories – review of 8 years’ experience
- Utility of a panel of sera for the alignment of test results in the worldwide multicenter study on reference values
- Protocol and standard operating procedures for common use in a worldwide multicenter study on reference values
- Guidelines and Recommendations
- 2013 update on the worldwide standardization of the hemoglobin A1c measurement
- General Clinical Chemistry and Laboratory Medicine
- Evacuated blood-collection tubes for haematological tests – a quality evaluation prior to their intended use for specimen collection
- Refrigeration is not necessary for measurement of uric acid in patients treated with rasburicase
- Procalcitonin and mid-regional pro-adrenomedullin test combination in sepsis diagnosis
- Cysteinyl leukotrienes in exhaled breath condensate of smoking asthmatics
- Helicobacter pylori serology in autoimmune diseases – fact or fiction?
- Serum DNase I activity in systemic lupus erythematosus: correlation with immunoserological markers, the disease activity and organ involvement
- Antibodies against Nε-homocysteinylated proteins in patients on different methods of renal replacement therapy
- Reference Values and Biological Variations
- Reference values for urinary neutrophil gelatinase-associated lipocalin (NGAL) in pediatric age measured with a fully automated chemiluminescent platform
- High biological variation of serum hyaluronic acid and Hepascore, a biochemical marker model for the prediction of liver fibrosis
- Cardiovascular Diseases
- Circulating matrix Gla protein: a potential tool to identify minor carotid stenosis with calcification in a risk population
- Midregional pro-atrial natriuretic peptide in the general population/Insights from the Gutenberg Health Study
- Letters to the Editor
- Relationship between Helicobacter pylori infection and autoimmune disorders
- Rapid simultaneous genotyping of polymorphisms in ADH1B and ALDH2 using high resolution melting assay
- Comparison of biological specimens and DNA collection methods for PCR amplification and microarray analysis
- Transferrin/log(ferritin) ratio: a self-fulfilling prophecy when iron deficiency is defined by serum ferritin concentration
- Response to: Transferrin/log(ferritin) ratio: a self-fulfilling prophecy when iron deficiency is defined by serum ferritin concentration
- Determinants of homocysteine concentrations in mother and neonatal girl pairs
- Cysteine analog breaks cryoprecipitate associated with chronic hepatitis C
- Cut-off values of serum growth hormone (GH) in pharmacological stimulation tests (PhT) evaluated in short-statured children using a chemiluminescent immunometric assay (ICMA) calibrated with the International Recombinant Human GH Standard 98/574
- Normalized MEDx chart as a useful tool for evaluation of analytical quality achievements. A picture is worth a thousand words
- Harmonization of immunoassays to the all-procedure trimmed mean – proof of concept by use of data from the insulin standardization project
- Congress Abstracts
- Annual Assembly of the Swiss Society of Clinical Chemistry & International Congress of Porphyrins and Porphyrias & International Meeting of Porphyria Patients: Personalized Medicine and Rare Diseases