Home Medicine A reference system for urinary albumin: current status
Article Publicly Available

A reference system for urinary albumin: current status

  • John C. Lieske EMAIL logo , Olga Bondar , W. Greg Miller , Lorin M. Bachmann , Andrew S. Narva , Yoshihisa Itoh , Ingrid Zegers , Heinz Schimmel , Karen Phinney and David M. Bunk
Published/Copyright: December 12, 2012

Abstract

Background: Increased urinary excretion of albumin reflects kidney damage and is a recognized risk factor for progression of renal and cardiovascular disease. Considerable inter-method differences have been reported for both albumin and creatinine measurement results, and therefore the albumin-to-creatinine ratio. Measurement accuracy is unknown and there are no independent reference measurement procedures for albumin and no reference materials for either measurand in urine.

Methods: The National Kidney Disease Education Program (NKDEP) Laboratory Working Group and the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) have initiated joint projects to facilitate standardization of urinary albumin and creatinine measurement.

Results: A candidate LC-MS/MS reference measurement procedure for urinary albumin and candidate reference materials for urinary albumin and creatinine has been developed. The status of validations of these reference system components is reported.

Conclusions: The development of certified reference materials and reference measurement procedures for urinary albumin will enable standardization of this important measurand.

Introduction

Increased urinary excretion of albumin reflects kidney damage and is a widely recognized risk factor for progression of renal and cardiovascular disease (CVD). Considerable inter-method differences have been reported for both albumin and creatinine measurement procedures [and, therefore, the albumin-to-creatinine ratio (ACR)], but accuracy is unknown and there are no independent reference measurement procedures for albumin and no reference materials for either measurand in urine [1]. At present most IVD manufacturers provide calibrators with values that are traceable either to the serum reference material ERM-DA470k/IFCC, certified for its albumin content by immunoassay [2], or to purified albumin preparations. Only recently a highly purified human serum albumin reference material was prepared specifically for the standardization of urinary albumin measurements [3].

The National Kidney Disease Education Program (NKDEP) Laboratory Working Group and the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) have initiated joint projects to facilitate standardization of urinary albumin measurement. Over the last 4 years much progress has been made towards development of a candidate reference measurement procedure for urinary albumin that employs liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS), and preparation and evaluation of urinary albumin and creatinine reference materials. A large-scale assessment of harmonization among routine measurement procedures was also initiated which aims to evaluate the comparability of urinary albumin measurement procedures, the effects of interferences, and the suitability of existing (serum) reference materials and candidate reference materials. This report describes the status of these efforts.

Results

Initial impressions of agreement among routine measurement procedures and commutability of diluted ERM-DA470k/IFCC

To facilitate the development of recommendations for standardization of routine urinary albumin measurement procedures, the joint Working Group identified the following goals: 1) obtain and evaluate objective data on the current state of agreement among different routine measurement procedures for urinary albumin; 2) identify potential urine matrix and albumin forms for which routine measurement procedures are expected to provide good performance; 3) evaluate the equivalence of immunoassay measurement procedures based on comparison with the candidate LC-MS/MS reference measurement procedure; and 4) evaluate the commutability characteristics of two candidate reference materials for urinary albumin (highly-purified HSA from the Japanese Society of Clinical Chemistry and dilutions of ERM-DA470k/IFCC), to determine the effectiveness of these materials for standardization efforts.

To address these goals, 342 residual native urine samples were obtained from patients for whom urinary albumin measurements were made as part of their routine medical care. The fresh urine samples were shipped overnight at 2–8°C for testing by 16 commercially available quantitative routine urinary albumin measurement procedures. A series of centrally-prepared dilutions of ERM-DA470k/IFCC were also distributed to participants for blinded analysis with patient samples. In addition, each participant prepared dilutions of the ERM-DA470k/IFCC and the JSCC reference materials.

Preliminary evaluation of agreement among routine measurement procedures suggested that medians of the measured concentrations of patient urine samples varied approximately 40% for urinary albumin concentrations of 12–1270 mg/L. Preliminary visual inspection of the relationship of results between the routine measurement procedures and the candidate reference measurement procedure suggested that diluted ERM-DA470k material was likely to be commutable with patient results for at least 11 of 16 routine measurement procedures. The commutability characteristics of the ERM-DA470k materials for the remaining five methods were indeterminate by visual inspection. The data generated from this study will be analyzed in detail and used to determine the feasibility of improving standardization of results among routine measurement procedures through use of currently available reference materials and the candidate reference measurement procedure.

Urinary albumin and creatinine certified reference materials (CRMs)

ERM-DA470 and ERM-DA470k/IFCC

Most routine measurement procedures for albumin in serum produce results that are traceable to the certified reference material ERM-DA470k/IFCC or to its predecessor ERM-DA470. ERM-DA470 was produced in the early 1990s after extensive feasibility studies [4]. The result was a material that, although extensively stabilized, delipidated and freeze-dried, had properties close enough to serum samples to be commutable for most proteins that were value assigned. Quantity values were assigned to the material with routine clinical laboratory immunoassays that were further optimized. In 2008 this batch of material was exhausted and ERM-DA470k/IFCC was released to replace it. The value for albumin in ERM-DA40k/IFCC was again assigned by immunoassays, using ERM-DA470 as calibrator [2].

These reference materials have been used by a majority of manufacturers for assigning values to their product calibrators for routine serum measurement procedures. External quality assessment schemes show that in general measurements of serum albumin are under control, with between-laboratory coefficient of variations (CVs) around 5%–6% for all measurement procedures, including methods based on turbidimetry, nephelometry, and the use of dye-binding chromophores [results from the Referenzinstitut für Bioanalytik (http://www.dgkl-rfb.de/)in 2011 and 2012]. The success of ERM-DA470 for serum measurements is to a considerable extent due to the development of very carefully controlled procedures for transferring values from the CRM to manufacturers’ master calibrators and from master calibrators to product calibrators [5]. There are still specific cases in which a significant bias between measurement procedures exists, and it has been postulated that this bias could be due to problems with the transfer of values along the traceability chain [6] or to use of different dye binding chromophores.

The use of ERM-DA470k/IFCC for the calibration of urinary albumin measurements requires that dilutions of the CRM are commutable for the different measurement procedures to be calibrated. Such dilutions are not necessarily trivial, as the urine matrix is quite different from diluted serum. Therefore, the commutability of dilutions of ERM-DA470k/IFCC for urinary albumin measurements is the subject of a large-scale commutability study (discussed previously). If dilutions prepared according to controlled protocols are found to be sufficiently commutable, their use for calibration may lead to harmonized urinary albumin results.

Pure HSA CRM

NIST is in the process of developing a primary certified reference material for albumin (NIST SRM 2925, ‘Human serum albumin solution’). The intended use of this certified reference material will be the preparation of calibration solutions for the reference measurement procedure of albumin in human urine. Recombinant human serum albumin (rHSA) to prepare this CRM has been purchased from a commercial source and evaluated at NIST to assess protein purity and molecular structure heterogeneity. The target albumin concentration of NIST SRM 2925 is approximately 1 g/L. NIST SRM 2925 will be supplied in approximately 1 mL aliquots of frozen solution, stored at −80°C. The certified HSA concentration of SRM 2925 will be determined at NIST by amino acid analysis using isotope dilution LC-MS/MS [7].

The Japanese Society of Clinical Chemistry (JSCC) has proposed highly purified, monomeric, human serum albumin in an aqueous buffer as a secondary reference material for urinary albumin [3]. It was prepared by gel-filtration from commercially-available highly purified monomeric HSA (JSCC in preparation). The assigned value was transferred from HSA in ERM-DA470. The assigned value was 225.1±9.11 mg/L (mean±expanded uncertainty with coverage factor of 2) when reconstituted with 3.00 mL of purified water. This monomeric HSA secondary reference material is intended for use as a calibrator or a control in albumin and total protein measurement procedures for urine.

15N-labeled albumin CRM

The candidate reference measurement procedures for albumin in urine developed at the Mayo Clinic, and under development at NIST, require a series of reagents. These reagents should also be accessible to other reference laboratories. Currently, there are no commercial sources of the 15N-labeled recombinant HSA which is used as the internal standard in the reference measurement procedure. NIST is searching for a commercial laboratory to evaluate approaches for the production of the 15N-labeled recombinant HSA in sufficient quantity and with sufficient isotopic incorporation to be available as a CRM.

Urinary creatinine CRM

To support the clinical measurement of the ACR in urine samples, NIST has produced SRM 3667 (‘Creatinine in Frozen Human Urine’). NIST SRM 3667 was prepared from pooled human urine, collected from a minimum of 10 male and female donors in good health (as self-reported). The urine pool was blended, filtered, and dispensed into amber glass bottles, with a fill volume of approximately 10 mL each, and stored at −80°C. The certification of creatinine in SRM 3667 was performed at NIST using isotope dilution LC-MS in a measurement procedure similar to that used at NIST to certify the concentration of creatinine in serum [8]. The measured concentration of creatinine in SRM 3667 is 5.5 mmol/L (62 mg/dL), and within the typical reference interval. NIST SRM 3667 is expected to be available by the end of 2012.

Urinary albumin reference measurement procedure

Materials

A candidate reference measurement procedure has been developed in the Mayo Clinic Renal Function Laboratory that employs trypsin digestion of whole urine followed by LC-MS/MS [9]. A stock solution of purified human serum albumin (8.74 g/L HSA, #A3782, Sigma Aldrich, St. Louis, MO, USA) was prepared in water. The concentration was determined by ultraviolet absorption spectroscopy with a molar absorptivity at 280 nm of 38533 L/(mol·cm) [10]. Calibrators were prepared by adding 572 µL of the stock solution to a charcoal-stripped urine matrix (BioChemed Services, Winchester, VA, USA) to a final volume of 25 mL and concentration of 200 mg/L. Serial dilutions were made into the charcoal stripped urine (12.5 mL each) to achieve concentrations of 100, 50, 25, 12.5, 6.25, and 3.13 mg/L. Low (10 mg/L), medium (40 mg/L), and high (106 mg/L) quality controls (QC) were prepared by adding the HSA stock solution to a waste human urine sample with endogenous concentration of 0.8 mg/L albumin. Recombinant 15N-labeled human albumin (15NrHSA) for use as an internal standard was synthesized in Pichia pastoris as described in the supplemental portion of a previous publication [11]. All calibrators and controls were divided into 0.5 mL aliquots and frozen at –80°C until used.

Urine samples (40 µL) were diluted with ammonium bicarbonate (0.1 mol/L, 135 µL, Sigma-Aldrich) to normalize sample pH, and then the 15NrHSA internal standard was added (10 µL of a 161 mg/L stock). Samples were reduced by the addition of dithiothreitol (final concentration 10 mmol/L, Sigma-Aldrich) for 60 min at 60°C, and then alkylated using iodoacetamide (50 mmol/L, Sigma-Aldrich) for 30 min at 25°C in the dark. Next L-(tosylamido-2-phenyl) ethyl chloromethyl ketone-treated trypsin (5 µL, 1 g/L Sigma Aldrich, in 0.1 mol/L ammonium bicarbonate) was added and the sample was digested for 1 h at 37°C. Finally all samples were acidified with 2 µL of concentrated formic acid to result in a final sample volume of 204 µL.

LC-MS/MS

Three tryptic peptides that were reproducibly detected in samples of trypsin-digested HSA and distributed throughout the protein were selected for measurement of multiple reaction monitoring (MRM) transitions: 42LVNEVTEFAK51, 526QTALVELVK534 and 13DLGEENFK20 (Figure 1). For quantification of: 13DLGEENFK20 the doubly charged precursor ion at m/z 476.2 was selected in Q1, and 3 singly charged transitions were monitored in Q3: m/z 723.3; 229.1; 201.1; for 42LVNEVTEFAK51 the doubly charged precursor ion at m/z 575.3 was selected in Q1, and 5 singly charged transitions were monitored in Q3: m/z 185.2; 213.2; 595.3; 694.4; 937.5; and for 526QTALVELVK534 the doubly charged precursor ion at m/z 500.8 was selected in Q1, and 3 singly charged transitions were monitored in Q3: m/z 700.5; 587.4; 147.3. Tryptic peptides from the internal standard 15NrHSA were monitored using the same conditions listed above for the native peptides. Quantification was performed by using a Thermo Scientific Aria TLX2 LC system coupled to an Applied Biosystems API 5500 triple-quadrupole mass spectrometer. For each run a total of 20 µL of sample digest was injected onto a 50×2.1 mm TARGA C18 column (Higgins Analytical). Chromatography was performed at a flow rate of 250 µL/min. Total run time was 30 min. Mobile phases consisted of solvent A (0.1% formic acid in water) and solvent B (0.1% formic acid in methanol).

Figure 1 Amino acid sequence of human serum albumin.Underlined sequences are peptides whose masses have been observed by LC-MS/MS analysis. Those highlighted in red have been reproducibly observed in patient urine samples and used for quantification.
Figure 1

Amino acid sequence of human serum albumin.

Underlined sequences are peptides whose masses have been observed by LC-MS/MS analysis. Those highlighted in red have been reproducibly observed in patient urine samples and used for quantification.

Each final albumin concentration represents the mean value of all transitions for each selected peptide.

Validation to date

Between-day precision of the candidate reference measurement procedure was evaluated by measuring the seven concentrations of the standard curve over 20 days (Table 1). Within-day precision was validated by measuring the low, medium and high QC material over 20 replicates on the same day (Table 2) by each of three tryptic peptides over 20 different days. All had CVs <5%. Recovery was evaluated by adding stock HSA to normal patient urine with an endogenous concentration of albumin 0.8 mg/L to achieve concentrations of 10, 40 and 100 mg/L (Table 3). Recovery was between 97% and 108% with an average of 103%. To assess linearity, waste urine samples containing high concentrations of albumin were serially diluted with 2× charcoal-stripped urine matrix to produce samples with values within the dynamic range of the assay (Table 4). Linearity, determined by comparing observed values with expected values, was established for dilutions up to 1:4.

Table 1

Between-day imprecision of the candidate reference measurement procedure for urine albumin over 20 days.

Expected HSA concentration in urine for calibrator or QC materials, mg/LHSA, mg/L n=20 days%CVHSA, mg/L n=20 days%CVHSA, mg/L n=20 days%CVHSA, mg/L n=20 days%CV
LVNEVTEFAK peptideQTALVELVK peptideDLGEENFK peptideTotal of 3 peptides
3.133.102.93.10.183.420.213.144.31
6.256.071.46.330.225.470.176.144.70
12.512.61.512.60.1411.40.2212.443.54
2525.50.624.90.0624.60.0924.991.68
5052.91.550.90.3349.90.3951.441.89
10095.31.595.60.83102.00.796.432.25
2002020.7204.20.371990.33202.880.88
10.8 low QC11.39.911.510.211.89.511.5510.48
40.8 medium QC41.56.839.956.440.66.440.285.97
106.8 high QC105.14.4101.83.7103.34.1103.24.07
Table 2

Within-day precision evaluated based on the measurement of QC samples (20 replicates).

PeptidesHSA, mg/L n=20 timesCV%HSA, mg/L n=20 timesCV%HSA, mg/L n=20 timesCV%
Low QCMedium QCHigh QC
LVNEVTEFAK12.81.344.54.0111.92.5
QTALVELVK12.91.143.41.3111.50.6
DLGEENFK13.51.044.11.3111.42.0
Total of 3 peptides13.12.8544.14.111.62.24
Table 3

Recovery of HSA (10, 40, and 100 mg/L) spiked into human urine (0.8 mg/L).

PeptidesExpected mean, mg/LCalculated mean, mg/L% Recovery
Sample 1LVNEVTEFAK10.810.4697
QTALVELVK10.810.97102
DLGEENFK10.811.63108
Sample 2LVNEVTEFAK40.841.5102
QTALVELVK40.841.3101
DLGEENFK40.842.5104
Sample 3LVNEVTEFAK100.8104103
QTALVELVK100.8105104
DLGEENFK100.8105104
Table 4

Linearity of albumin measurement in three human urine samples diluted with 2X-charcoal stripped urine.

PeptidesSample 1Sample 2Sample 3
DilutionExpected concentration, mg/LCalculated concentration, mg/L% RecoveryExpected concentration, mg/LCalculated concentration, mg/L% RecoveryExpected concentration, mg/LCalculated concentration, mg/L% Recovery
LVNEVTEFAK
 0110110.8101166170.2103182181.2100
 25558.21068385.51039190.7100
 427.524.0874242.410145.546101
 813.756.75492121.71032315.567
QTALVELVK
 0110110.7101166155.093182178.398
 25551.4938381.1989186.3095
 427.522.7824240.29645.543.696
 813.755.70412120.80992313.659
DLGEENFK
 0110110.0100166165.7100182187.7103
 25557.31048383.41009191.6101
 427.524.2884241.79945.546.3102
 813.756.8492122.01052315.668

A total of 334 human urine samples that covered the range of clinical values (5–1270 mg/L) were measured in duplicate with the LC-MS/MS assay as part of a clinical validation study. This data was analyzed to determine the influence that the number of peptides and resulting transitions that were measured had on the ultimate reproducibility of the value obtained (Table 5). When one peptide and one transition were measured, the CV was 3.21%. The CV improved to 2.21% when two transitions were averaged for each peptide. Reproducibility improved marginally when results from any two peptides were averaged (CV 1.94%). Using two transitions for each peptide further improved the average CV to 1.80%. Addition of three or more peptides did not improve precision, regardless of the number of transitions used. These results suggest that averaging results from two peptides and two transitions each will provide maximal precision for this measurement procedure.

The albumin concentration indicated by each fragment was also compared across these 334 human urine samples, and the mean% bias between each fragment pair assessed by Bland-Altman plots. Indeed, there was a subtle bias such that a peptide fragment towards the amino terminus (13DLGEENK20) gave values 1.7% lower than a more centrally located fragment (138YLYEIAR144), while the fragment nearest the carboxyl terminus (526QTALVELVK534) gave even lower values (3.6% lower than 138YLYEIAR144, Figure 2). These biases could be consistent with a minor amount of degradation of albumin molecules at both ends while in human urine.

Table 5

Influence of number of peptides/transitions on within run precision for human urine samples.

1 peptideMS/MS transitions2 peptidesMS/MS transitions3 peptidesMS/MS transitions4 peptidesMS/MS transitions5 peptidesMS/MS transitions
Transitions1212121212
DLGEENFK2.732.09DL2.001.57DLQ2.021.88DLQA1.991.90DLQAY2.011.91
LVNEVTEFAK3.942.08DQ1.911.69DLA2.011.97DLQY2.021.61
QTALVELVK4.532.04DA1.891.80DLY2.061.94LQAY2.091.96
AEFAEVSKLVTDLTK2.342.58DY1.281.88LQA2.071.96
YLYEIAR2.512.24LQ2.021.96LQY2.121.95
LA1.941.93QAY2.192.01
LY1.971.89
QA2.041.93
QY2.071.72
AY2.321.60
Average %CV3.212.211.941.802.081.952.031.822.011.91

The quantification of intact albumin in urine by trypsin digestion followed by LC-MS/MS could potentially be influenced by smaller endogenous fragments of albumin in the urine sample. To evaluate this possibility, five samples with albumin concentrations within the dynamic range of the assay and five samples with concentrations of albumin more than 1000 mg/L were selected for study. The waste urine sample (200 µL) was loaded on a 10 kD cut Amicon filter to remove small fragments. The retentate (approx. 20 µL) was reconstituted with charcoal-stripped urine matrix (200 µL) and concentrated again. Once again the retentate (20 µL now without small fragments <10 kDa) was reconstituted with charcoal-stripped urine matrix to a final volume of 200 µL. The concentration of albumin in all five samples before and after concentration was similar (Table 6). Therefore, our studies do not suggest that endogenous albumin fragments interfere with the measurements.

Figure 2 Bias between urinary albumin peptide measurements.For each of 334 human urine samples, the albumin concentration assessed by each of five trypsin fragments was compared (2 transitions each). The mean % bias (95% CI) between each fragment pair was assessed by Bland-Altman plots. There was a subtle bias with the overall lowest values for the fragment nearest the carboxyl terminus.
Figure 2

Bias between urinary albumin peptide measurements.

For each of 334 human urine samples, the albumin concentration assessed by each of five trypsin fragments was compared (2 transitions each). The mean % bias (95% CI) between each fragment pair was assessed by Bland-Altman plots. There was a subtle bias with the overall lowest values for the fragment nearest the carboxyl terminus.

NIST validation of LC-MS/MS assay

For a reference measurement procedure to be useful and robust, it should be independently validated in a laboratory setting other than the one in which it was developed. As such, the candidate reference measurement procedure developed at the Mayo Clinic is also being implemented at NIST for method validation. The implementation at NIST has maintained the basic measurement strategy of the Mayo LC-MS/MS method; however the NIST measurements have used instrumentation and chromatographic media from manufacturers different than those used by Mayo in order to evaluate the robustness of the methods based on LC-MS. A bilateral measurement comparison of patient samples is planned between NIST and the Mayo Clinic.

Table 6

Measured albumin concentration before and after filtration to remove fragments <10 kDa.

QTALVELVKLVNEVTEFAKDLGEENFK
Albumin, mg/L in patient samplesAlbumin, mg/L in 10 kDa cut samples% RecoveryAlbumin, mg/L in patient samplesAlbumin, mg/L in 10 kD cut samples% RecoveryAlbumin, mg/L in patient samplesAlbumin, mg/L in 10 kD cut samples% Recovery
Sample 147.547.610447.548.410746.247.2101
Sample 283.072.09284.773.89385.97498
Sample 388.878.08689.37884887689
Sample 420.520.3932222942121109
Sample 5110.7105.4951111069611010695
Sample 624422392962442243010023002313101
Sample 7258322817926982385802542248088
Sample 8442539358944373997914337385388
Sample 9630673106650682106603657109
Sample 10425240259343204088954125392396

In addition to implementing the Mayo Clinic’s candidate reference measurement procedure for albumin in urine, NIST has also expanded the method to provide qualitative information on the structure of albumin in urine. Capitalizing on the multiplexed measurement capabilities of modern mass spectrometers, NIST has expanded the LC-MS/MS measurement of three proteolytic peptides from albumin in the Mayo Clinic method to include eight additional proteolytic peptides, 11 peptides in total. The 11 peptides measured include peptides derived from albumin’s N- and C-terminal regions as well as several peptides from the central portion of albumin’s protein chain. Measurement of these peptides should provide qualitative information on the ‘intactness’ of the molecular form of albumin in urine samples to compliment the quantitative information on the concentration of albumin. The ability to evaluate potential molecular heterogeneity of albumin in urine samples while simultaneously quantifying albumin in urine will be very useful in future commutability studies of certified reference materials which will also include measurement of sets of individual patient samples. The qualitative measurement capability will help to identify structural differences in the albumin in patient samples that can be useful in the interpretation of commutability study results and in the assessment of the molecular forms that should be measured by routine clinical laboratory procedures. The number of peptides incorporated into the final candidate reference measurement procedure will be determined once the performance of each of these eight additional peptides has been assessed.

Discussion

This report describes progress towards a reference measurement system for urinary albumin, a key biomarker of kidney disease. Studies are in progress to assess the current state of the art, including agreement amongst current commercially available routine measurement procedures and commutability of available candidate reference materials. Efforts have also been directed towards development of a candidate reference measurement procedure for urinary albumin, and validation of candidate reference materials for urinary creatinine and albumin. The influence of adsorption of albumin to container surfaces is also being investigated by the working group, although those results are not described here.

Other preanalytic and analytic factors that can also influence use of urinary albumin as a biomarker of renal disease and that may benefit from standardized approaches are summarized below based on the report from the joint working group in 2009 [1].

Biologic/physiologic factors

Time of urine collection (first morning, second morning, random, or 24 h) is important. Twenty-four hour collections may be theoretically ideal, but are not always practical and may be incomplete which will increase variability. Among random urine collections, a first morning void is preferred, since this collection will decrease biologic variability [1]. Importantly, different decision limits may be needed for random vs. first morning collections, or other standardized collection time(s) [12].

Stability and interference

The influence of blood (menstrual or urinary bleeding) seminal fluid and other physiologic contaminants of urine remain to be examined. The molecular forms of albumin in freshly voided urine, and their correlation to kidney disease, has not been fully investigated. Clarification regarding variation of urinary matrix composition and its effects on urinary albumin measurement procedures is needed, including the degree to which albumin degrades under various conditions of storage.

Interpretation

The ACR varies with age, gender and ethnicity. Decision thresholds suitable for these subgroups need further investigation. The commonly used 30 mg/g or 3.4 mg/mmol decision limits may not be adequate for each subgroup in terms of diagnostic sensitivity. Risk of chronic kidney disease and CVD are continuous functions of urinary albumin. The appropriate reference intervals for given populations (e.g., general population or high-risk groups such as diabetes, hypertension or CVD) need to be examined. It would be interesting to examine if age- and gender-specific equations to convert ACR to an estimated albumin excretion rate could be useful to enable using a single reference limit.

Conclusions

Urinary albumin excretion is a critical marker of kidney disease. Since the initial report of the NKDEP and IFCC joint working group in 2009 [1], much progress toward standardizing measurement of urine albumin has been made, including the development of candidate reference materials and a candidate LC-MS/MS reference measurement procedure.


Corresponding author: John C. Lieske, Mayo Clinic Renal Function Laboratory, Department of Laboratory Medicine and Pathology, Mayo Clinic Division of Nephrology and Hypertension, Department of Internal Medicine, Rochester, MN 55905, USA

The authors appreciate the contribution of members of the joint working group of the National Kidney Disease Education Program (NKDEP) Laboratory Working Group and the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Working Group for Standardization of Albumin in Urine.

Conflict of interest statement

Authors’ conflict of interest disclosure: The authors stated that there are no conflicts of interest regarding the publication of this article.

Research funding: None declared.

Employment or leadership: None declared.

Honorarium: None declared.

References

1. Miller WG, Bruns DE, Hortin GL, Sandberg S, Aakre KM, McQueen MJ, et al. Current issues in measurement and reporting of urinary albumin excretion. Clin Chem 2009;55:24–38.10.1373/clinchem.2008.106567Search in Google Scholar

2. Zegers I, Keller T, Schreiber W, Sheldon J, Albertini R, Blirup-Jensen S, et al. Characterization of the new serum protein reference material ERM-DA470k/IFCC: value assignment by immunoassay. Clin Chem 2011;56:1880–8.10.1373/clinchem.2010.148809Search in Google Scholar

3. Itoh Y, Ichihara K, Kishi K, Hosogaya S, Yamada T. Preparation of highly purified monomeric human serum albumin as secondary reference material for standardization of urinary albumin immunoassays. Clin Chim Acta 2012;413:175–81.10.1016/j.cca.2011.09.016Search in Google Scholar

4. Baudner S, Bienvenu J, Blirup-Jensen S, Carlström A, Johnson AM, Milfor A, editors. The certification of a matrix reference material for immunochemical measurement of 14 human serum proteins CRM470EUR 15423 EN and 16882 N. Luxembourg: European Community, 1993.Search in Google Scholar

5. Blirup-Jensen S, Johnson AM, Larsen M. Protein standardization V: value transfer. A practical protocol for the assignment of serum protein values from a Reference Material to a Target Material. Clin Chem Lab Med 2008;46:1470–9.10.1515/CCLM.2008.289Search in Google Scholar

6. Infusino I, Braga F, Valente C, Panteghini M. Commutability of the ERM-DA470k reference material for two assays measuring serum albumin using immunochemical principles. Clin Chem Lab Med 2011;49:1383–4.10.1515/CCLM.2011.619Search in Google Scholar

7. Bunk DM, Lowenthal MS. Isotope dilution liquid chromatography-tandem mass spectrometry for quantitative amino acid analysis. Methods Mol Biol 2012;828:29–38.10.1007/978-1-61779-445-2_3Search in Google Scholar

8. Huskova R, Chrastina P, Adam T, Schneiderka P. Determination of creatinine in urine by tandem mass spectrometry. Clin Chim Acta 2004;350:99–106.10.1016/j.cccn.2004.07.007Search in Google Scholar

9. Seegmiller JC, Barnidge DR, Burns BE, Larson TS, Lieske JC, Kumar R. Quantification of urinary albumin by using protein cleavage and LC-MS/MS. Clin Chem 2009;55:1100–7.10.1373/clinchem.2008.115543Search in Google Scholar

10. Gill SC, von Hippel PH. Calculation of protein extinction coefficients from amino acid sequence data. Anal Biochem 1989;182:319.10.1016/0003-2697(89)90602-7Search in Google Scholar

11. Singh R, Crow FW, Babic N, Lutz WH, Lieske JC, Larson TS, et al. A liquid chromatography-mass spectrometry method for the quantification of urinary albumin using a novel 15N-isotopically labeled albumin internal standard. Clin Chem 2007;53:540–2.10.1373/clinchem.2006.078832Search in Google Scholar PubMed

12. Witte EC, Lambers Heerspink HJ, de Zeeuw D, Bakker SJ, de Jong PE, Gansevoort R. First morning voids are more reliable than spot urine samples to assess microalbuminuria. J Am Soc Nephrol 2009;20:436–43.10.1681/ASN.2008030292Search in Google Scholar PubMed PubMed Central

Received: 2012-5-30
Accepted: 2012-11-9
Published Online: 2012-12-12
Published in Print: 2013-05-01

©2013 by Walter de Gruyter Berlin Boston

Articles in the same Issue

  1. Masthead
  2. Masthead
  3. Editorial
  4. Progress towards standardization: an IFCC Scientific Division Perspective
  5. Review
  6. Quantitative Clinical Chemistry Proteomics (qCCP) using mass spectrometry: general characteristics and application
  7. Mini Reviews
  8. Analytical goals for the determination of HbA2
  9. Glucose meters – fit for clinical purpose
  10. Opinion Papers
  11. Metrological traceability – a concept for standardization in laboratory medicine
  12. Recommendations for clinical laboratory science reports regarding properties, units, and symbols: the NPU format1)
  13. “Good samples make good assays” – the problem of sourcing clinical samples for a standardization project
  14. Defining acceptable limits for the metrological traceability of specific measurands
  15. Original Articles
  16. A reference system for urinary albumin: current status
  17. Toward standardization of carbohydrate-deficient transferrin (CDT) measurements: III. Performance of native serum and serum spiked with disialotransferrin proves that harmonization of CDT assays is possible
  18. External Quality Assessment Scheme for reference laboratories – review of 8 years’ experience
  19. Utility of a panel of sera for the alignment of test results in the worldwide multicenter study on reference values
  20. Protocol and standard operating procedures for common use in a worldwide multicenter study on reference values
  21. Guidelines and Recommendations
  22. 2013 update on the worldwide standardization of the hemoglobin A1c measurement
  23. General Clinical Chemistry and Laboratory Medicine
  24. Evacuated blood-collection tubes for haematological tests – a quality evaluation prior to their intended use for specimen collection
  25. Refrigeration is not necessary for measurement of uric acid in patients treated with rasburicase
  26. Procalcitonin and mid-regional pro-adrenomedullin test combination in sepsis diagnosis
  27. Cysteinyl leukotrienes in exhaled breath condensate of smoking asthmatics
  28. Helicobacter pylori serology in autoimmune diseases – fact or fiction?
  29. Serum DNase I activity in systemic lupus erythematosus: correlation with immunoserological markers, the disease activity and organ involvement
  30. Antibodies against Nε-homocysteinylated proteins in patients on different methods of renal replacement therapy
  31. Reference Values and Biological Variations
  32. Reference values for urinary neutrophil gelatinase-associated lipocalin (NGAL) in pediatric age measured with a fully automated chemiluminescent platform
  33. High biological variation of serum hyaluronic acid and Hepascore, a biochemical marker model for the prediction of liver fibrosis
  34. Cardiovascular Diseases
  35. Circulating matrix Gla protein: a potential tool to identify minor carotid stenosis with calcification in a risk population
  36. Midregional pro-atrial natriuretic peptide in the general population/Insights from the Gutenberg Health Study
  37. Letters to the Editor
  38. Relationship between Helicobacter pylori infection and autoimmune disorders
  39. Rapid simultaneous genotyping of polymorphisms in ADH1B and ALDH2 using high resolution melting assay
  40. Comparison of biological specimens and DNA collection methods for PCR amplification and microarray analysis
  41. Transferrin/log(ferritin) ratio: a self-fulfilling prophecy when iron deficiency is defined by serum ferritin concentration
  42. Response to: Transferrin/log(ferritin) ratio: a self-fulfilling prophecy when iron deficiency is defined by serum ferritin concentration
  43. Determinants of homocysteine concentrations in mother and neonatal girl pairs
  44. Cysteine analog breaks cryoprecipitate associated with chronic hepatitis C
  45. Cut-off values of serum growth hormone (GH) in pharmacological stimulation tests (PhT) evaluated in short-statured children using a chemiluminescent immunometric assay (ICMA) calibrated with the International Recombinant Human GH Standard 98/574
  46. Normalized MEDx chart as a useful tool for evaluation of analytical quality achievements. A picture is worth a thousand words
  47. Harmonization of immunoassays to the all-procedure trimmed mean – proof of concept by use of data from the insulin standardization project
  48. Congress Abstracts
  49. Annual Assembly of the Swiss Society of Clinical Chemistry & International Congress of Porphyrins and Porphyrias & International Meeting of Porphyria Patients: Personalized Medicine and Rare Diseases
Downloaded on 26.2.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2012-0768/html
Scroll to top button