Workflow restrictions on hematology analyzers XE-2100 and XS-800 when assaying pediatric samples with limited volume

  • Klas Böer and Thomas Deufel
Published/Copyright: May 1, 2009

Abstract

Background: Samples with limited volume is a common problem in laboratories receiving samples from pediatric patients. Also, pediatric samples may contain nucleated red blood cells (NRBC) which distorts the white blood cell (WBC) count and which can only be measured by some automated cell counting systems. Differential counts are sometimes required, posing the question of validity of flagging depending on age of the patients and on predilutions.

Methods: We evaluated the hematology analyzers XE-2100 and XS-800 for their suitability in measuring hematological parameters in such samples.

Results: With the exception of the MCHC and partly the MCH, we observed very good agreement between complete blood counts (CBC) in diluted and undiluted samples. Dilution did not impair sensitivity in the clinically relevant range nor, accuracy of the NRBC count on XE-2100. Flagging was ineffective in undiluted samples from children <1 year of age and in all diluted samples when measuring differential counts.

Conclusions: In summary, while automated measurement of CBC and NRBC is possible in diluted samples, measurement of differential counts is restricted by loss of flagging efficiency. In addition, flagging is also ineffective in children <1 year of age using the analyzers evaluated and should, for diagnostic purposes, be performed manually.

Clin Chem Lab Med 2009;47:607–11.


Corresponding author: Dr. Klas Böer, Institute for Clinical Chemistry and Laboratory Medicine, Friedrich-Schiller-University Jena, 07740 Jena, Germany

Received: 2008-11-13
Accepted: 2009-3-1
Published Online: 2009-05-01
Published in Print: 2009-05-01

©2009 by Walter de Gruyter Berlin New York

Articles in the same Issue

  1. Review
  2. Pharmacogenetics of vitamin K antagonists: useful or hype?
  3. Genetics and Molecular Diagnostics
  4. Illumina DNA test panel-based genotyping of whole genome amplified-DNA extracted from hair samples: performance and agreement with genotyping results from genomic DNA from buccal cells
  5. Association of interleukin-18 gene promoter polymorphisms with risk of acute myocardial infarction in northern Chinese Han population
  6. MYC quantitation in cell-free plasma DNA by real-time PCR for gastric cancer diagnosis
  7. Predictive value of CA 125 and CA 72-4 in ovarian borderline tumors
  8. Anti DNA antibodies are not restricted to a specific pattern of fluorescence on HEp2 cells
  9. General Clinical Chemistry and Laboratory Medicine
  10. Determination of volatile organic compounds in exhaled breath of patients with lung cancer using solid phase microextraction and gas chromatography mass spectrometry
  11. Determination of urinary stone composition based on stone morphology: a prospective study of 325 consecutive patients in an emerging country
  12. Determination of branched chain amino acids, methionine, phenylalanine, tyrosine and α-keto acids in plasma and dried blood samples using HPLC with fluorescence detection
  13. Hematocytometry analysis as discriminative marker for asthma phenotypes
  14. A routine method for the quantification of the novel antimycotic drug posaconazole in plasma using liquid chromatography-tandem mass spectrometry
  15. Reference Values and Biological Variations
  16. Serum β2-microglobulin measured by immunonephelometry: expression patterns and reference intervals in healthy adults
  17. Serum enzyme profile characteristics of victims following the Wenchuan earthquake in China
  18. Stability of selected serum proteins after long-term storage in the Janus Serum Bank
  19. Spuriously high thyrotropin values due to anti-thyrotropin antibodies in adult patients
  20. Validation and Outcome Studies
  21. Workflow restrictions on hematology analyzers XE-2100 and XS-800 when assaying pediatric samples with limited volume
  22. Letters to the Editor
  23. Carbamylated hemoglobin remains a critical issue in HbA1c measurements
  24. Detection of a monoclonal gammopathy by serum index measurement
  25. Survey on the prevalence of hemolytic specimens in an academic hospital according to collection facility: opportunities for quality improvement
  26. Quantitative scanning analysis of a cryoglobulin ring detected in sera of patients with hepatitis C using a cooling gel diffusion method
  27. Significant reduction of erythrocyte glucose-6-phosphate dehydrogenase activity in soccer-players during play. Evidence for catecholamine mediated enzyme inhibition
Downloaded on 5.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2009.135/html
Scroll to top button