Analytical evaluation of the Optium Xido blood glucose meter

  • Bogdan Solnica , Jerzy Naskalski and Wojciech Gernand
Published/Copyright: November 22, 2007

Abstract

Background: The main prevention of chronic diabetic complications is maintaining near normoglycemia. In this study, evaluation of analytical performance of the Optium Xido blood glucose meter designed for patients' self-monitoring of glycemia was carried out.

Methods: Glucose concentrations were measured in 118 EDTA blood samples using the Xido glucose meter and the GOD/PAP method on a Modular P clinical analyzer. The results obtained were used for the assessment of accuracy, precision and linearity. Performance of the Xido glucose meter was also assessed using four different reagent strip lots, and the between-lot variation was calculated.

Results: The within-run imprecision coefficient of variation (CV) amounted to 4.2%. Good response linearity was found in glucose concentrations of 31–444 mg/dL (1.7–24.7 mmol/L). In the concentration range studied, the glucose meter error was 5.14% and the linear regression equation was y=0.91x+6.2, (r=0.984). The Passing-Bablok agreement test indicated good concordance of results. However, for glucose concentrations <100 mg/dL (5.6 mmol/L) (n=69) the error was 6.82% with regression equation y=0.86x+5.9 (r=0.757). Between-lot differences amounted to 0.7%–18.2%.

Conclusions: The Xido glucose meter has good precision and accuracy when compared to the laboratory method and meets the quality recommendations of the National Committee of Clinical Laboratory Standards (NCCLS, currently the Clinical Laboratory Standards Institute), the National Academy of Clinical Biochemistry (NACB) and the International Organization for Standardization (ISO). However, between-lot variability may result in differences between day-to-day results when the device is continuously used. Therefore, the validation of new lots of reagent strips with a laboratory method is recommended.

Clin Chem Lab Med 2008;46:143–7.


Corresponding author: Bogdan Solnica, Department of Clinical Biochemistry, Jagiellonian University Medical College, ul. Kopernika 15b, 31-501 Krakow, Poland Phone/Fax: +48-12-4248361,

Received: 2007-7-6
Accepted: 2007-9-10
Published Online: 2007-11-22
Published in Print: 2008-01-01

©2008 by Walter de Gruyter Berlin New York

Articles in the same Issue

  1. Editorial
  2. Improving healthcare through advances in point-of-care technologies
  3. Review
  4. New creatinine sensor for point-of-care testing of creatinine meets the National Kidney Disease Education Program guidelines
  5. Original Papers
  6. Performance of a multi-profile critical care testing analyzer
  7. Comparative study of calculated and measured total carbon dioxide
  8. Letter to the Editor
  9. Performance of immunofluorometric point-of-care assays for free pregnancy-associated plasma protein A detection in whole blood samples
  10. Guidelines and Recommendations
  11. IFCC Guideline for sampling, measuring and reporting ionized magnesium in plasma
  12. Reviews
  13. Critical evaluation of 1H NMR metabonomics of serum as a methodology for disease risk assessment and diagnostics
  14. The entero-insular axis: implications for human metabolism
  15. Genetics and Molecular Diagnostics
  16. Functional polymorphisms in the promoter of the matrix metalloproteinase-9 (MMP-9) gene are not linked with significant plasma MMP-9 variations in healthy subjects
  17. Genetic profiling in healthy subjects from the Stanislas cohort based on 24 polymorphisms: effects on biological variables
  18. Relationship between methionine synthase, methionine synthase reductase genetic polymorphisms and deep vein thrombosis among South Indians
  19. Simultaneous genotyping of the three lactose tolerance-linked polymorphisms LCT –13907C>G, LCT –13910C>T and LCT –13915T>G with Pyrosequencing technology
  20. A new high-throughput screening method for the detection of chronic lymphatic leukemia and myelodysplastic syndrome
  21. General Clinical Chemistry and Laboratory Medicine
  22. Saliva analysis by surface-enhanced laser desorption/ionization time-of-flight mass spectrometry (SELDI-TOF/MS): from sample collection to data analysis
  23. Mechanism of sodium loss with muscle sodium deficiency in sodium supplemented and unsupplemented subjects during hypokinesia
  24. Increased levels of 8-hydroxydeoxyguanosine and its relationship with lipid peroxidation and antioxidant vitamins in lung cancer
  25. Lipid peroxidation in stroke patients
  26. Resistin is present in human breast milk and it correlates with maternal hormonal status and serum level of C-reactive protein
  27. Waist-to-hip ratio correlates with homocysteine levels in male patients with coronary artery disease
  28. Isoenzymes of N-acetyl-β-hexosaminidase in human pleomorphic adenoma and healthy salivary glands: a preliminary study
  29. The in vivo and in vitro effects of L-carnitine supplementation on the erythrocyte membrane acetylcholinesterase, Na+, K+-ATPase and Mg2+-ATPase activities in basketball players
  30. Validation and Outcome Studies
  31. Analytical evaluation of the Optium Xido blood glucose meter
  32. Letters to the Editor
  33. New biomarkers of myocardial injury and assessment of cardiac toxicity during preparative regimen and hematopoietic cell transplantation in acute leukemia
  34. Analysis of serum and urinary lysophospholipase D/autotaxin in nephrotic syndrome
Downloaded on 13.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2008.022/html
Scroll to top button