Analytical evaluation of the Optium Xido blood glucose meter
-
Bogdan Solnica
Abstract
Background: The main prevention of chronic diabetic complications is maintaining near normoglycemia. In this study, evaluation of analytical performance of the Optium Xido blood glucose meter designed for patients' self-monitoring of glycemia was carried out.
Methods: Glucose concentrations were measured in 118 EDTA blood samples using the Xido glucose meter and the GOD/PAP method on a Modular P clinical analyzer. The results obtained were used for the assessment of accuracy, precision and linearity. Performance of the Xido glucose meter was also assessed using four different reagent strip lots, and the between-lot variation was calculated.
Results: The within-run imprecision coefficient of variation (CV) amounted to 4.2%. Good response linearity was found in glucose concentrations of 31–444 mg/dL (1.7–24.7 mmol/L). In the concentration range studied, the glucose meter error was 5.14% and the linear regression equation was y=0.91x+6.2, (r=0.984). The Passing-Bablok agreement test indicated good concordance of results. However, for glucose concentrations <100 mg/dL (5.6 mmol/L) (n=69) the error was 6.82% with regression equation y=0.86x+5.9 (r=0.757). Between-lot differences amounted to 0.7%–18.2%.
Conclusions: The Xido glucose meter has good precision and accuracy when compared to the laboratory method and meets the quality recommendations of the National Committee of Clinical Laboratory Standards (NCCLS, currently the Clinical Laboratory Standards Institute), the National Academy of Clinical Biochemistry (NACB) and the International Organization for Standardization (ISO). However, between-lot variability may result in differences between day-to-day results when the device is continuously used. Therefore, the validation of new lots of reagent strips with a laboratory method is recommended.
Clin Chem Lab Med 2008;46:143–7.
©2008 by Walter de Gruyter Berlin New York
Articles in the same Issue
- Editorial
- Improving healthcare through advances in point-of-care technologies
- Review
- New creatinine sensor for point-of-care testing of creatinine meets the National Kidney Disease Education Program guidelines
- Original Papers
- Performance of a multi-profile critical care testing analyzer
- Comparative study of calculated and measured total carbon dioxide
- Letter to the Editor
- Performance of immunofluorometric point-of-care assays for free pregnancy-associated plasma protein A detection in whole blood samples
- Guidelines and Recommendations
- IFCC Guideline for sampling, measuring and reporting ionized magnesium in plasma
- Reviews
- Critical evaluation of 1H NMR metabonomics of serum as a methodology for disease risk assessment and diagnostics
- The entero-insular axis: implications for human metabolism
- Genetics and Molecular Diagnostics
- Functional polymorphisms in the promoter of the matrix metalloproteinase-9 (MMP-9) gene are not linked with significant plasma MMP-9 variations in healthy subjects
- Genetic profiling in healthy subjects from the Stanislas cohort based on 24 polymorphisms: effects on biological variables
- Relationship between methionine synthase, methionine synthase reductase genetic polymorphisms and deep vein thrombosis among South Indians
- Simultaneous genotyping of the three lactose tolerance-linked polymorphisms LCT –13907C>G, LCT –13910C>T and LCT –13915T>G with Pyrosequencing™ technology
- A new high-throughput screening method for the detection of chronic lymphatic leukemia and myelodysplastic syndrome
- General Clinical Chemistry and Laboratory Medicine
- Saliva analysis by surface-enhanced laser desorption/ionization time-of-flight mass spectrometry (SELDI-TOF/MS): from sample collection to data analysis
- Mechanism of sodium loss with muscle sodium deficiency in sodium supplemented and unsupplemented subjects during hypokinesia
- Increased levels of 8-hydroxydeoxyguanosine and its relationship with lipid peroxidation and antioxidant vitamins in lung cancer
- Lipid peroxidation in stroke patients
- Resistin is present in human breast milk and it correlates with maternal hormonal status and serum level of C-reactive protein
- Waist-to-hip ratio correlates with homocysteine levels in male patients with coronary artery disease
- Isoenzymes of N-acetyl-β-hexosaminidase in human pleomorphic adenoma and healthy salivary glands: a preliminary study
- The in vivo and in vitro effects of L-carnitine supplementation on the erythrocyte membrane acetylcholinesterase, Na+, K+-ATPase and Mg2+-ATPase activities in basketball players
- Validation and Outcome Studies
- Analytical evaluation of the Optium Xido blood glucose meter
- Letters to the Editor
- New biomarkers of myocardial injury and assessment of cardiac toxicity during preparative regimen and hematopoietic cell transplantation in acute leukemia
- Analysis of serum and urinary lysophospholipase D/autotaxin in nephrotic syndrome
Articles in the same Issue
- Editorial
- Improving healthcare through advances in point-of-care technologies
- Review
- New creatinine sensor for point-of-care testing of creatinine meets the National Kidney Disease Education Program guidelines
- Original Papers
- Performance of a multi-profile critical care testing analyzer
- Comparative study of calculated and measured total carbon dioxide
- Letter to the Editor
- Performance of immunofluorometric point-of-care assays for free pregnancy-associated plasma protein A detection in whole blood samples
- Guidelines and Recommendations
- IFCC Guideline for sampling, measuring and reporting ionized magnesium in plasma
- Reviews
- Critical evaluation of 1H NMR metabonomics of serum as a methodology for disease risk assessment and diagnostics
- The entero-insular axis: implications for human metabolism
- Genetics and Molecular Diagnostics
- Functional polymorphisms in the promoter of the matrix metalloproteinase-9 (MMP-9) gene are not linked with significant plasma MMP-9 variations in healthy subjects
- Genetic profiling in healthy subjects from the Stanislas cohort based on 24 polymorphisms: effects on biological variables
- Relationship between methionine synthase, methionine synthase reductase genetic polymorphisms and deep vein thrombosis among South Indians
- Simultaneous genotyping of the three lactose tolerance-linked polymorphisms LCT –13907C>G, LCT –13910C>T and LCT –13915T>G with Pyrosequencing™ technology
- A new high-throughput screening method for the detection of chronic lymphatic leukemia and myelodysplastic syndrome
- General Clinical Chemistry and Laboratory Medicine
- Saliva analysis by surface-enhanced laser desorption/ionization time-of-flight mass spectrometry (SELDI-TOF/MS): from sample collection to data analysis
- Mechanism of sodium loss with muscle sodium deficiency in sodium supplemented and unsupplemented subjects during hypokinesia
- Increased levels of 8-hydroxydeoxyguanosine and its relationship with lipid peroxidation and antioxidant vitamins in lung cancer
- Lipid peroxidation in stroke patients
- Resistin is present in human breast milk and it correlates with maternal hormonal status and serum level of C-reactive protein
- Waist-to-hip ratio correlates with homocysteine levels in male patients with coronary artery disease
- Isoenzymes of N-acetyl-β-hexosaminidase in human pleomorphic adenoma and healthy salivary glands: a preliminary study
- The in vivo and in vitro effects of L-carnitine supplementation on the erythrocyte membrane acetylcholinesterase, Na+, K+-ATPase and Mg2+-ATPase activities in basketball players
- Validation and Outcome Studies
- Analytical evaluation of the Optium Xido blood glucose meter
- Letters to the Editor
- New biomarkers of myocardial injury and assessment of cardiac toxicity during preparative regimen and hematopoietic cell transplantation in acute leukemia
- Analysis of serum and urinary lysophospholipase D/autotaxin in nephrotic syndrome