Performance of a multi-profile critical care testing analyzer
-
Somlak Vanavanan
Abstract
Background: Modern blood gas analyzers are often coupled to electrolyte and metabolite analyzers. We evaluated a Stat Profile Critical Care Xpress analyzer (STP CCX) for the rapid point-of-care measurement of blood gases (pH, pCO2, pO2, sO2), hematocrit (Hct), total hemoglobin (tHb), sodium (Na+), potassium (K+), chloride (Cl–), glucose (Glu), lactate (Lac), urea (BUN), ionized calcium (iCa) and ionized magnesium (iMg).
Methods: The analyzer was evaluated in terms of imprecision and recovery using the STP CCX control. Fresh blood samples were also measured to determine the between-day imprecision. Correlation was assessed by clinical sample comparison with the Nova Stat Profile Ultra C and Dimension RxL systems for Cl– and BUN. We used Deming regression, correlation coefficients, mean differences, and the Wilcoxon signed-rank test for data analysis.
Results: The coefficients of variation for all analytes were within desirable limits, ranging from 0.1% to 4.3%, and the recovery was 100%±3%. Between-day imprecision using fresh blood samples showed good results, ranging from 0.2% to 3.4%. The comparison results showed high to very high correlation. However, statistically significant mean differences with large bias were found for pCO2, pO2 and Cl–.
Conclusions: This analyzer is suitable as a simple and fast diagnostic tool in the laboratory and the critical care unit. However, users should be aware of biases that may lead to clinically significant errors in the assessment of acid-base status.
Clin Chem Lab Med 2008;46:9–14.
©2008 by Walter de Gruyter Berlin New York
Articles in the same Issue
- Editorial
- Improving healthcare through advances in point-of-care technologies
- Review
- New creatinine sensor for point-of-care testing of creatinine meets the National Kidney Disease Education Program guidelines
- Original Papers
- Performance of a multi-profile critical care testing analyzer
- Comparative study of calculated and measured total carbon dioxide
- Letter to the Editor
- Performance of immunofluorometric point-of-care assays for free pregnancy-associated plasma protein A detection in whole blood samples
- Guidelines and Recommendations
- IFCC Guideline for sampling, measuring and reporting ionized magnesium in plasma
- Reviews
- Critical evaluation of 1H NMR metabonomics of serum as a methodology for disease risk assessment and diagnostics
- The entero-insular axis: implications for human metabolism
- Genetics and Molecular Diagnostics
- Functional polymorphisms in the promoter of the matrix metalloproteinase-9 (MMP-9) gene are not linked with significant plasma MMP-9 variations in healthy subjects
- Genetic profiling in healthy subjects from the Stanislas cohort based on 24 polymorphisms: effects on biological variables
- Relationship between methionine synthase, methionine synthase reductase genetic polymorphisms and deep vein thrombosis among South Indians
- Simultaneous genotyping of the three lactose tolerance-linked polymorphisms LCT –13907C>G, LCT –13910C>T and LCT –13915T>G with Pyrosequencing™ technology
- A new high-throughput screening method for the detection of chronic lymphatic leukemia and myelodysplastic syndrome
- General Clinical Chemistry and Laboratory Medicine
- Saliva analysis by surface-enhanced laser desorption/ionization time-of-flight mass spectrometry (SELDI-TOF/MS): from sample collection to data analysis
- Mechanism of sodium loss with muscle sodium deficiency in sodium supplemented and unsupplemented subjects during hypokinesia
- Increased levels of 8-hydroxydeoxyguanosine and its relationship with lipid peroxidation and antioxidant vitamins in lung cancer
- Lipid peroxidation in stroke patients
- Resistin is present in human breast milk and it correlates with maternal hormonal status and serum level of C-reactive protein
- Waist-to-hip ratio correlates with homocysteine levels in male patients with coronary artery disease
- Isoenzymes of N-acetyl-β-hexosaminidase in human pleomorphic adenoma and healthy salivary glands: a preliminary study
- The in vivo and in vitro effects of L-carnitine supplementation on the erythrocyte membrane acetylcholinesterase, Na+, K+-ATPase and Mg2+-ATPase activities in basketball players
- Validation and Outcome Studies
- Analytical evaluation of the Optium Xido blood glucose meter
- Letters to the Editor
- New biomarkers of myocardial injury and assessment of cardiac toxicity during preparative regimen and hematopoietic cell transplantation in acute leukemia
- Analysis of serum and urinary lysophospholipase D/autotaxin in nephrotic syndrome
Articles in the same Issue
- Editorial
- Improving healthcare through advances in point-of-care technologies
- Review
- New creatinine sensor for point-of-care testing of creatinine meets the National Kidney Disease Education Program guidelines
- Original Papers
- Performance of a multi-profile critical care testing analyzer
- Comparative study of calculated and measured total carbon dioxide
- Letter to the Editor
- Performance of immunofluorometric point-of-care assays for free pregnancy-associated plasma protein A detection in whole blood samples
- Guidelines and Recommendations
- IFCC Guideline for sampling, measuring and reporting ionized magnesium in plasma
- Reviews
- Critical evaluation of 1H NMR metabonomics of serum as a methodology for disease risk assessment and diagnostics
- The entero-insular axis: implications for human metabolism
- Genetics and Molecular Diagnostics
- Functional polymorphisms in the promoter of the matrix metalloproteinase-9 (MMP-9) gene are not linked with significant plasma MMP-9 variations in healthy subjects
- Genetic profiling in healthy subjects from the Stanislas cohort based on 24 polymorphisms: effects on biological variables
- Relationship between methionine synthase, methionine synthase reductase genetic polymorphisms and deep vein thrombosis among South Indians
- Simultaneous genotyping of the three lactose tolerance-linked polymorphisms LCT –13907C>G, LCT –13910C>T and LCT –13915T>G with Pyrosequencing™ technology
- A new high-throughput screening method for the detection of chronic lymphatic leukemia and myelodysplastic syndrome
- General Clinical Chemistry and Laboratory Medicine
- Saliva analysis by surface-enhanced laser desorption/ionization time-of-flight mass spectrometry (SELDI-TOF/MS): from sample collection to data analysis
- Mechanism of sodium loss with muscle sodium deficiency in sodium supplemented and unsupplemented subjects during hypokinesia
- Increased levels of 8-hydroxydeoxyguanosine and its relationship with lipid peroxidation and antioxidant vitamins in lung cancer
- Lipid peroxidation in stroke patients
- Resistin is present in human breast milk and it correlates with maternal hormonal status and serum level of C-reactive protein
- Waist-to-hip ratio correlates with homocysteine levels in male patients with coronary artery disease
- Isoenzymes of N-acetyl-β-hexosaminidase in human pleomorphic adenoma and healthy salivary glands: a preliminary study
- The in vivo and in vitro effects of L-carnitine supplementation on the erythrocyte membrane acetylcholinesterase, Na+, K+-ATPase and Mg2+-ATPase activities in basketball players
- Validation and Outcome Studies
- Analytical evaluation of the Optium Xido blood glucose meter
- Letters to the Editor
- New biomarkers of myocardial injury and assessment of cardiac toxicity during preparative regimen and hematopoietic cell transplantation in acute leukemia
- Analysis of serum and urinary lysophospholipase D/autotaxin in nephrotic syndrome