Abstract

We processed 317 samples from healthy adult volunteers for a complete blood count, including leukocyte differentials and reticulocyte parameters, through five new-generation haematology analysers: Abx Pentra 120 Retic, Coulter Gen-S, Sysmex SE 9500, Abbott Cell Dyn 4000 and Bayer Advia 120. From these data nonparametric 2.5–97.5 percentile reference intervals were calculated for all parameters on all analysers.

Some differences were found compared with previously reported reference intervals. Reference intervals for platelet parameters and reticulocytes agreed with these usually accepted. For red blood cell parameters, including haemoglobin and haematocrit, and white blood cell count, including absolute white blood cell differentials, our calculated reference intervals were in agreement with less frequently cited earlier reports, but were lower compared to the usually accepted reference intervals.

:
Published Online: 2005-06-01
Published in Print: 2002-01-29

Copyright © 2002 by Walter de Gruyter GmbH & Co. KG

Artikel in diesem Heft

  1. Editorial
  2. Growing Significance of Myeloperoxidase in Non-infectious Diseases
  3. Serum Paraoxonase Activity and the Extent of Lipid Peroxidation Are not Affected by Increased Levels of Human Apolipoprotein A-I: Studies in Transgenic Mice
  4. Relationship between the Sialic Acid Content of Low-Density Lipoprotein (LDL) and Autoantibodies to Oxidized LDL in the Plasma of Healthy Subjects and Patients with Atherosclerosis
  5. The Presence of Heparin-Platelet Factor 4 Antibodies as a Marker of Hypercoagulability during Hemodialysis
  6. Clinical Evaluation of Changes in the Morphology of Eosinophils
  7. The n-3 and n-6 Polyunsaturated Fatty Acid Composition of Plasma Phospholipids in Pregnant Women and Their Infants. Relationship with Maternal Linoleic Acid Intake
  8. Quantification of Sirolimus by Liquid Chromatography-Tandem Mass Spectrometry Using On-Line Solid-Phase Extraction
  9. The Effects of Affinity-Purified Anti-DNA Antibodies Derived from Patients with Systemic Lupus Erythematosus on the Fluorescent Antinuclear Antibody Assay Using HEp-2 Cells
  10. Relationships between Serum Markers of Monocyte/Macrophage Activation in Type 1 Gaucher's Disease
  11. Gender Differences in C-Reactive Protein Concentrations - Confirmation with Two Sensitive Methods
  12. Serum Cholinesterase Activity in Patients with Burns
  13. Serum Lipids and Apolipoproteins in Patients with Psoriasis
  14. Reference Intervals for a Complete Blood Count Determined on different Automated Haematology Analysers: Abx Pentra 120 Retic, Coulter Gen-S, Sysmex SE 9500, Abbott Cell Dyn 4000 and Bayer Advia 120
  15. Science, Alchemy and Light: Paintings by Joseph Wright of Derby
  16. Approved IFCC Reference Method for the Measurement of HbA1c in Human Blood
  17. Meetings and Awards
Heruntergeladen am 25.11.2025 von https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2002.014/html?lang=de
Button zum nach oben scrollen